Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.ricelake.com/training
or obtained by calling 715-234-9171 and asking for the training department.
Contents
About This Manual ................................................................................................................................... 1
420 Plus Limited Warranty .................................................................................................................... 54
ii420 Plus Installation Manual
Page 5
About This Manual
WARNING
WARNING
Important
CAUTION
WARNING
This manual is intended for use by service technicians
responsible for installing and servicing
420 Plus HMI
digital weight indicators. This manual applies to
indicators using Version 1.14 of the
420 Plus software.
Configuration and calibration of the indicator can be
accomp
EDP command set, or the
lished using the indicator front panel keys, the
Revolution
®
configuration
utility. See Section 3.1 on page 15 for information
about configuration methods.
Safety
Safety Symbol Definitions
Indicates a potentially hazardous situation that, if not avoided, could result in serious injury or death, and
includes hazards that are exposed when guards are removed.
Indicates a potentially hazardous situation that, if not avoided may result in minor or moderate injury.
Indicates information about procedures that, if not observed, could
corruption to and loss of data.
Safety Precautions
Do not operate or work on this equipment unless you have read and understand the instructions and
warnings in this Manual. Failure to follow the instructions o
death. Contact any Rice Lake Weighing Systems dealer for replacement manuals. Proper care is your
responsibility.
Some procedures described in this m
are to be performed by qualified service personnel only.
anual require work inside the indicator enclosure. These procedures
This manual can be viewed from the Rice
Lake Weighing Systems distributor site at
www.ricelake.com.
The Operator Car
basic operating instructions for users of the
d included with this manual provides
420 Plus.
Please leave the Operator Card with the indicator
when installation and configuration are complete.
result in damage to equipment or
r heed the warnings could result in injury or
General Safety
Failure to heed may result in serious injury or death.
DO NOT allow minors (children) or inexper
DO NOT operate without all shields and guards in place.
DO NOT step on the unit.
DO NOT jump up and down on the scale.
DO NOT use for purposes other than weight taking.
DO NOT place fingers into slots or possible pinch points.
DO NOT use any load-bearing component that is wo
DO NOT use this product if any of the
DO NOT exceed the rated load limit of the unit.
DO NOT make alterations or modifications to the unit.
DO NOT remove or obscure warning labels.
DO NOT use near water.
Before opening the unit, ensure the power cord
Keep hands, feet and loose clothing away from moving parts.
ienced persons to operate this unit.
rn beyond 5% of the original dimension.
components are cracked.
is disconnected from the outlet.
420 Plus Installation Manual - Safety1
Page 6
1.0Introduction
The 420 Plus is a single-channel digital weight
indicator housed in a NEMA 4X/IP66-rated stainless
steel enclosure. The indicator front panel consists of a
large (.8 in, 20 mm), six-digit, seven-segment LED
display and twenty-one-button keypad. Features
include:
•Drives up to eight 350or sixteen 700 load
cells
•Supports 4- and 6-wire load cell connections
•Two configurable digital inputs
•Two configurable digital outputs
•Electronic data processing (EDP) port for full
duplex, RS-232 communications at up to
38400 bps
•Printer port for output-only RS-232 or 20 mA
current loop communications at up to 38400
bps
•Optional analog output module provides 0–10
VDC or 0–20/4–20 mA tracking of gross or net
weight values
•Available in 115 VAC and 230 VAC versions
•Available in DC power version
The 420 Plus is NTEP-certified and pending
Measurement Canada approval for Classes III, III HD,
and III L at 10,000 divisions. See
Section 7.12 on
page 53 for detailed specifications.
1.1Operating Modes
The 420 Plus has four modes of operation:
Normal (Primary) mode
Normal mode is the “default” mode of the
indicator. The indicator displays gross or net
weights as required, using the LED annunciators
described in
status and the type of weight value displayed. Once
configuration is complete and a legal seal is affixed
to the back of the indicator, this is the primary
mode in which the
Piece Count (Secondary) Mode
In piece count mode, the indicator display shows
the number of parts on the scale rather than the
weight of those parts. Piece count mode has two
submodes:
•Count display mode displays the current parts
count and allows ticket printing using the
CFMT print format.
•Sample acquisition mode is used to calibrate
the indicator for parts counting.
Operator access to piece count mode is disabled
when the indicator is shipped from the factory.
Setup mode
Most of the procedures described in this manual
require the indicator to be in setup mode, including
configuration and calibration.
To enter setup mode, remove the large fillister head
screw from the bottom of the enclosure. Insert a
screwdriver or a similar tool into the access hole
and press the setup switch once. The indicator
display changes to show the word
Section 1.3 on page 3 to indicate scale
420 Plus can operate.
CONFIG.
2420 Plus Installation Manual
Test mode
Test mode provides a number of diagnostic
functions for the
420 Plus indicator. Like setup
mode, test mode is entered using the setup switch.
Section 7.9 on page 51 for more information
See
about entering and using test mode.
Page 7
1.2Front Panel Keypad
lb
kg
Count
Gross
Net
DISPLAY
TARE
SETPOINT TIME/DATE
POWER
I/O
2
3 4 5
7 8 9 0 6
CLR
ZERO GROSS
NET
TARE
ENTER MODE SAMPLE
UNITS PRINT
Figure 1-1 shows the 420 Plus LED annunciators,
keypad, and normal mode key functions.
The symbols shown above the keys (representing up,
down, enter, left, right) describe the key functions
assigned in setup mode. In setup mode, the keys are
used to navigate through menus, select digits within
numeric values, and increment/decrement values. See
Section 3.1.3 on page 16 for information about using
the front panel keys in setup mode.
Figure 1-1. 420 Plus Front Panel, Showing LED Annunciators and Normal Mode Key Functions
1.3LED Annunciators
The 420 Plus display uses a set of eight LED
annunciators to provide additional information about
the value being displayed:
Gross and Net annunciators are lit to show
•
whether the displayed weight is a gross or net
weight.
): Gross weight is within
): Scale is at standstill or
•Center of zero (
0.25 graduations of zero. This annunciator
lights whe
•Standstill (
within the specified motion band. Some
operations, including tare functions
printing, can only be done when the standstill
symbol is shown.
lb and kg annunciators indicate the units
•
associated with the displayed value:
lb=pounds, kg=kilograms.
n the scale is zeroed.
and
•The displayed units can also be set to short
tons (tn
), metric tons (t), ounces (oz), grams
(g), or NONE (no units information displayed).
lb and kg LEDs function as primary and
The
secondary units annunciators for some
combinations of primary and secondary units.
If neither primary nor secondary units are lb or
kg the
for secondary units.
•The
indicator is in piece count mode.
•The Tare Acquired (
lb annunciator is lit for primary units, kg
Count annunciator is lit to show that the
) lights to show that a
tare value was entered.
420 Plus Installation Manual - Introduction3
Page 8
Table 1-1 shows which annunciators are used for all combinations of configured primary and secondary units. For
Note
example:
•If the primary unit is pounds (lb) and the
kg for secondary units.
units,
•If the primary unit is pounds (lb) and the secondary unit
secondary unit is kilograms (kg), the lb LED is lit for primary
is short tons (tn), the lb LED is lit for primary
units, kg for secondary units. There is not an LED for short tons, so the kg LED is used as the secondary
units annunciator.
•If the primary unit is short tons (tn) and the
(tn), and
kg is lit for secondary units (lb). Because there is no LED for short tons, the lb and kg LEDs are
secondary unit is pounds (lb), the lb LED is lit for primary units
used as primary and secondary units annunciators.
See Section on page 19 for more information about configuring primary and secondary display units.
Table 1-1. Units Annunciators, Showing Primary / Secondary LEDs Used for All Configurations
lbkgozgtntnone
1.4Indicator Operations
Basic 420 Plus operations are summarized below.
1.4.1Toggle Gross/Net Mode/Piece Count
Press the GROSS/NET key to switch the display mode
from gross to net, or from net to gross. If a tare value
has been entered or acquired, the net value is the gross
weight minus the tare.
Gross mode is shown by the
mode is shown by the
1.4.2Toggle Units
Net annunciator.
Gross annunciator; net
Press the UNITS key to switch between primary and
secondary units. The appropriate units LED to the right
of the display is lit.
1.4.3Zero Scale
1. In gross mode, remove all weight from the
scale and wait for the standstill annunciator
(
2. Press the
).
ZERO key. The center of zero ()
annunciator lights to indicate the scale is
zeroed.
1.4.4Acquire Tare
1. Place container on scale and wait for the
standstill annunciator (
).
2. Press the
TARE key to acquire the tare weight
of the container. Net weight is displayed and
the (
) annunciator lights to show the tare
value was entered
1.4.5Enter Tare (Keyed Tare)
1. Use the numeric keypad to enter the tare value,
then press the TARE key.
2. Net weight is displayed and the (
annunciator lights to show the tare value was
entered.
1.4.6Remove Stored Tare Value
1. Remove all weight from the scale and wait for
the standstill annunciator (
2. Press the
TAR E key. The () annunciator goes
off, indicating the tare value has been removed.
Indicators with the REGULA parameter set to
NONE or NTEP (see Section 3.2.5 on page 23)
can clear a stored tare value using the
following pr
ocedure:
1. Press DISPLAY TARE to show the stored tare
value.
2. Press the
CLEAR key twice to remove the
stored tare.
)
).
4420 Plus Installation Manual
Page 9
1.4.7Acquire Parts Sample
Note
1. Place empty parts container on scale. Wait for
the standstill annunciator (
TARE to acquire the tare weight of the
), then press
container.
2. Press
MODE (GROSS/NET) key to enter piece
count mode.
Press the
3. Press the
CLEAR key to exit.
SAMPLE (UNITS) key to enter sample
acquisition mode.
The indicator display shows the message
where
nnn is the sample quantity to be placed on the
Addnnn,
scale. You can do one of the following:
•Add the number of parts shown.
•Choose a different sample size. Press the
SAMPLE key to scroll through the
selectable sample quantities (5, 10, 20, 50,
100) or use the numeric keypad to specify
a custom sample size.
•Specify a known piece weight. Press the
SAMPLE key to scroll through the
selectable sample quantities until the
WGT prompt is shown. Use the numeric
PC
keypad to enter the piece weight.
4. Once the sample quantity is
ENTER to calibrate the indicator for counting
on the scale, press
the new parts. If a sample size was specified,
the indicator display shows the message –CNT–
as it acquires the sample weight, then switches
to count display mode and shows the part
quantity. If a known piece weight was
specified, the display switches to count display
mode immediately.
1.4.8Display Part Weight
To view gross and net weight parts, press MODE to
switch from count display mode to normal weighing
mode. To view the current piece weight while in count
mode, press
1.4.9Display Accumulator
DISPLAY TARE key
Hold the (MODE) GROSS/NET key for three seconds to
display the accumulated value if enabled in
configuration.
The accumulated value will be displayed for about 10
seconds.
To clear the accumulator, press the
CLR key
twice while the accumulated value is being displayed.
To set the time, press the
TIME/DATE key twice. Use the
numeric keypad to enter the time in 24-hour format,
then press the
1.4.11Display or Change Setpoint Value
ENTER key.
The time and date is backed up with the
battery. If the main power is interrupted, time/
date should not
be lost.
To display a setpoint value, use the numeric keypad to
enter the setpoint number, then press the
SETPOINT key.
Or, you can display a setpoint value by pressing the
SETPOINT key a number of times equal to the setpoint
number. For example, to display the value of setpoint 2,
press the
The current value will display
to enter the new value and press the
SETPOINT key two times.
, use the numeric keypad
ENTER key. This
will bring you back to the TRIP submenu. To exit and
save, press the
ZERO ( ) and GROSS/NET ( ) to navigate to the
CONFIG menu. Press
UNITS ( ) and PRINT ( ) keys and
to exit and save. Use
Figure 3-11, “Setpoint Menu,” on page 25 to navigate
the menus.
1.4.12Turn Setpoint On or Off
To turn a setpoint on or off at the front panel, press the
SETPOINT key a number of times equal to the setpoint
number (for example, for setpoint number 2 press the
SETPOINT key two times). Press TAR E to exit value
input mode and go left to
arrow key to select
ENABLE and use the down
On or Off. At this point if the
setpoint is on, you can turn it off by using the right or
left arrow keys.
1.4.13Print Ticket
1. Wait for standstill annunciator ().
2. Press the
PRINT key to send data to the serial
port.
1.4.14Enter New ID
1. Ensure digital input 1 or digital input 2 is
configured for NEWID.
2. Activate the digital input.
3. Once activated, the digital input will
go into ID
mode.
4. Enter the ID using the nume
ric keypad and
press TARE .
1.4.10Display or Change Time
To display the date, press the TIME/DATE key once;
TIME/DATE a second time to display the time.
press
To set the date, press the
numeric keypad to enter the date, then press the
TIME/DATE key once. Use the
ENTER
key. Use the numeric keypad to enter the date in the
same format configured for the indicator: MMDDYY, DDMMYY, or YYMMDD.
420 Plus Installation Manual - Introduction5
Page 10
2.0Installation
CAUTION
WARNING
Note
Setup Switch
Access Screw
Communications Access
Cord Grip (Plugged)
Power Cord
Front of
Indicator
Backplate
Bottom View
Load Cell Cable
Cord Grip (Open)
This section describes procedures for connecting load
cells, digital inputs, and serial communications cables
to the
420 Plus indicator. Instructions for field
installation of the analog output option and
replacement of the CPU board are included, along with
assembly drawings and parts lists for the service
technician.
•Use a wrist strap to ground yourself and protect
components from electrostatic discharge (ESD)
when working inside the indicator enclosure.
•This unit uses line fusing which could create an
electric shock hazard.
inside the indicator must be performed by
qualified service personnel only.
•The supply cord serves as the main power
disc
onnect for the
supplying the indicator must be installed near the
unit and be easily accessible
Procedures requiring work
420 Plus. The power outlet
2.1Unpacking and Assembly
Immediately after unpacking, visually inspect the 420
to ensure all components are included and
Plus
undamaged. The shipping carton should contain the
indicator with attached tilt stand, this manual, and a
parts kit. If any parts were damaged in shipment, notify
Rice Lake Weighing Systems and the shipper
immediately.
The parts kit (PN 85219) contains the items listed
low:
be
•Two, six-position screw terminals (PN 70599)
for connector
screw terminals (PN 71125) for connectors J2
and J3, and one, four-position screw terminal
(PN 71126) for connector J6 (see figure 2-4).
•Two 8-32NC x 7/16 fillister head screws (PN
623).
30
•Four 8-32NC x 3/8 machine screws (PN
862) for the indicator backplate (see #1 in
14
Figure 2-8 13).
•Six neoprene washers (PN 45042) for
ba
ckplate screws included in the parts kit.
•Four rubber bumpers (“feet”) for the tilt stand,
(PN 42
149).
•Three reducing glands (PN 15664).
•One capacity label (PN 42350).
•Three each of grounding clamps (PN 53075),
external too
kep nuts (PN 14626) for cable shield
grounding against the enclosure.
s J4 & J1, two, three-position
th lock washers (PN 15133), and
•One SEC C (section cap) and CLC
(Concentrated Load) (PN 8
The indicator enclosure must be opened to connect
cables for load cells, communications, digital inputs,
and analog output.
The 420 Plus has an on/off switch for the
load cells and processor functions. Before
disconnected from the power outlet. The power outlet
opening the
must be located near the indicator to allow the operator to
easily disconnect power to the unit.
unit, ensure the power cord is
Ensure power to the indicator is disconnected, then
place the indicator face-down on an antistatic work
mat. Remove the screws that hold the backplate to the
enclosure body, then lift the backplate away from the
enclosure and set it aside.
2.3Cable Connections
The 420 Plus provides four cord grips for cabling into
the indicator: one for the power cord, three to
accommodate load cell, communications, digital
inputs, and analog output cables. Two of the three free
cord grips come with a plug installed to prevent
moisture from entering the enclosure. Depending on
your application, remove the plug from any cord grip
that will be used and install cables as required.
The unit will keep the date and time as long as
it is plugged in, Even if display and load cells
are turned off.
will lose date and time information.
Figure 2-1 shows the recommended assignments for
420 Plus cord grips.
the
Figure 2-1. Recommended Cord Grip Assignments
When the unit is unplugged, it
6420 Plus Installation Manual
Page 11
2.3.1Cable Grounding
Insulated cable
Foil (silver side out)
Grounding clamp
Shield wir e (cut)
Length of foil before folding
back on cable insulation
Cut insulation here
for foil-shielded cables
Braid
Cut insulation here
for braided cables
NOTE: Install lockwashers
first, against backplate,
under grounding clamp
Except for the power cord, all cables routed through the
cord grips should be grounded against the indicator
enclosure. Do the following to ground shielded cables:
•Use the lockwashers, clamps, and kep nuts
vided in the parts kit to install grounding
pro
clamps on the enclosure studs adjacent to cord
grips. Install grounding clamps only for cord
grips that will be used; do not tighten nuts.
•Route cables through cord grips and grounding
clamps to
determine cable lengths required to
reach cable connectors. Mark cables to remove
insulation and shield as described below:
•For cables with foil shielding, strip insulation
foil from the cable half an inch (15 mm)
and
past the grounding clamp (see Figure 2-2).
Fold the foil shield back on the ca
ble where the
cable passes through the clamp. Ensure silver
(conductive) side of foil is turned outward for
contact with the grounding clamp.
•For cables with braided shielding, strip cable
insulation and
braided shield from a point just
past the grounding clamp. Strip another half
inch (15 mm) of insulation only to expose the
braid where the cable passes through the clamp
(see Figure 2-2).
•For load cell cables, cut the shield wire just
ast the grounding clamp. Shield wire function
p
is provided by contact between the cable shield
and the grounding clamp.
•Route stripped cables th
rough cord grips and
clamps. Ensure shields contact grounding
clamps as shown in Figure 2-2. Tighten
grounding clamp nuts.
•Finish installation using cable mounts and ties
cure cables inside of indicator enclosure.
to se
Figure 2-2. Grounding Clamp Attachment for Foil-Shielded
and Braided Cabling
2.3.2Bypass Power Button
If jumper JMP1 is set for SW position, the power
switch on the front panel of the indicator will turn the
unit ON or OFF. If it is not in the SW position, the
indicator will power up as soon as the AC is applied.
This allows the front panel overlay power control
switch to be bypassed. Label “F” in Figure 2-3 shows
the location of JMP1. Figure 2-4 on page 8 shows
entire board.
Figure 2-3. 420 Plus CPU and Power Supply Board JMP1
DC power wiring to the indicator should be 18
AWG to 14 AWG for DC+, DC-, and earth
round conductors.
g
In a mobile application, earth ground (chassis) should
be terminated to the vehicle chassis.
In longer power cable runs, voltage drop over the
r conductor needs to be considered. See table and
powe
formula below to compute voltage drop.
= (2.85A)(x / 1000ft)(length of run in ft.)
V
DROP
2.85A = maximum current draw from DC/DC power
supply
x = Oh
ms from Table 2-1
Wire Gauge
(AWG)
14
15
16
17
18
Impedance
(OHMS/1000ft)
2.252
3.184
4.016
5.064
6.385
Table 2-1. Cable Impedance
Cable
DC voltage supplied to DC/DC power supply
should not be less than 9VDC. Using larger
gauge wire will result
in less voltage drop.
Example:
100ft run with 18 AWG wire
= (2.85A)(6.385 / 1000ft)(100 ft)
V
V
DROP
= 1.82V
DROP
12VDC will drop to 10.18V after 100ft run.
420 Plus Installation Manual - Installation9
Page 14
2.3.4Load Cells
Note
To attach cable from a load cell or junction box,
remove connector J1 from the board. The connector
plugs into a header on the board as shown in Figure 2-4
8.
Using one of the 6-position connectors, provided
in the
parts kit, wire the load cell cable from the load cell or
junction box to connector J1 on the CPU board (See
Figure 2-4 8). If using six-wire load cell cable (with
sense wires), remove jumpers JP1 and JP2 before
reinstalling co
four-wire installation, leave jumpers
nnector J1 (see Figure 2-4). For
JP1 and JP2 on.
When connections are complete, reinstall connector J1
onto the header so that it snaps securely into place. Use
two cable ties to secure the load cell cable to the inside
of the enclosure.
J1 PinFunction
1
2
3
4
5
6
+SIG
–SIG
+SENSE
–SENSE
+EXC
–EXC
Table 2-2. J1 Pin Assignments
Use grounding procedures described in
Section 2.3.1 on page 7
For 6-wire connections, remove jumpers JP1
and JP2
For 4-wire connections, leave jumpers JP1 and
JP2 on
2.3.5Serial Communications
Using one of the six-position connectors, provided in
the parts kit, wire the serial communications cables to
J4. Connector J3 provides connections for the EDP/
RS-232 port. Connect communications cables to
connectors J3 and J4 as shown in Table 2-3.
Once cables are attached, reconnec
t J3 and J4 to the
headers on the board (see Figure 2-4). Use cable ties to
secure serial cables to the inside of the enclosure.
The EDP port supports full duplex RS-232
communications only; the
serial port provides either
active 20 mA output or duplex RS-232 transmission.
Both ports are configured using the SERIAL menu. See
Section 3.0 on page 15 for configuration information.
PortConnector PinLabel
Serial PortJ4
1
TxD
2
RxD
3
Gnd
4
20mA+
5
20mA–
6
Gnd
Table 2-3. J3 and J4 Pin Assignments
2.3.6Digital I/O
Digital inputs can be set to provide several indicator
functions, including all keypad functions. The inputs
are active (on) with low voltage (0 VDC) and can be
driven by TTL or 5V logic without additional
hardware. Use the DIG IN menu to configure the
digital inputs. LED’s on the CPU board light when
digital inputs are active.
Digital outputs are typically used to
control relays that
drive other equipment. Outputs are designed to sink not
source, switching current. Each output is a normally
open connector circuit, capable of sinking 250 mA
when active. Digital outputs are wired to switch relays
when the digital output is active (low, 0 VDC) with
reference to 5 VDC supply. LEDs on the CPU board
light when the digital outputs are active.
PortConnector PinLabel
Digital InputJ2
Digital
Output
J6
Table 2-4. J2 and J6 Pin Assignments
1DI 1
2DI 2
3Gnd
1Gnd
2DO 1
3DO 2
4+5V
PortConnector PinLabel
EDP/RS-232J3
Table 2-3. J3 and J4 Pin Assignments
10420 Plus Installation Manual
1
TxD
2
RxD
3
Gnd
Page 15
2.3.7Analog Output
1
6
9
4
2
5
7
10
Torque Pattern
8
3
NTEP - Approval
Fastener
If the optional analog output module is installed, attach
the output cable to connector J1 on the analog output
board. Tab l e 2-5 lists the analog output pin
assignments.
Use the ALGOUT menu to configure and calibrate the
analog out
Section 2.4 for information about installing the
put module when cabling is complete. See
analog
output module.
PinSignal
1+ Current Out
2– Current Out
3+ Voltage Out
4– Voltage Out
Table 2-5. Analog Output Module Pin Assignments
2.4Analog Output Module Installation
To install or replace the analog output module (PN
85659), follow the steps listed in Section 2.2 on page 6
for opening the
420 Plus enclosure.
Mount the analog output module on its standoffs in the
location shown in Figure 2-4 8 and plug the module
input into connector J9
on the 420 Plus board. Connect
output cable to the analog output module as shown in
Table 2-5, then reassemble the enclosure (Section 2.5).
See Figure 7.8 50 for analog output calibration
procedures.
2.5Enclosure Reassembly
Once cabling is complete, position the backplate over
the enclosure and reinstall the backplate screws. Use
the torque pattern shown in Figure 2-6 to prevent
distorting the backplate gasket. Torque screws to 15
in-lb (1.7
N-m).
Figure 2-7. NTEP Sealing
2.6Board Removal
If you must remove the 420 Plus CPU board, use the
following procedure:
1. Disconnect power to the indicator. Remove
backplate as describe
page 6.
2. Disconnect power supply cable from connector
J7 on
the 420 Plus CPU board.
3. Unplug connectors J1 (load cell cable), J2
(digital inputs), J3 (EDP/RS-232), J4 (serial
communications), J6 (digital outputs), and J1
& J11 (keypad ribbon cables). If an analog
output board is installed, disconnect the analog
output cable. See Figure 2-4 8 for connector
locations.
4. Remove the five screws from the CPU board,
then lift t
he board out of the enclosure.
To replace the CPU board, reverse the above
procedure. Be sure to reinstall cable ties to secure all
ca
bles inside the indicator enclosure.
d in Section 2.2 on
0
Figure 2-6. 420 Plus Enclosure Backplate
Torqued screws may become less tight as the gasket is
compressed during torque pattern, therefore a second
torque is required using the same pattern and torque
value.
420 Plus Installation Manual - Installation11
Page 16
2.7Replacement Parts
Table 2-6 lists replacement parts for the 420 Plus, including all parts referenced in Figures 2-8 and 2-9.
Screws 8-32NC x 3/8 (8)Figure 2-9 14
Sealing washers (8)
Backplate (1)
backplate gasket (1)
Screws 6-32NC x 1/4 (9)
CPU and display board assembly (1)
Enclosure (1)
Nylon washers 1/4x1x1/16 (2)
Four-cornered wing knobs for tilt stand (2)
Tilt stand (1)
Sealing washer (1)
Screws 1/4-28NF X 1/4 (2)
Cable grip plugs (2)
Cable grips (3)
Nylon seal rings for cable grips (3)
Power cord assembly – 115 VAC
85203
Power cord assembly – 220 VAC
Power supply bracket (1)
Power supply switch (1)
85554
Power supply, DC/DC converter
Overlay panel (1)Figure 2-8 13
Rice Lake nameplate (1)
Power supply ribbon cable(1)
Earth ground label (1)
Lock washer, No 8, Type A (3)
Ground wire 4 in, No. 8 (1)
Kep nuts, 8-32NC Hex (5)
Locknuts (3)
Protective cover (1)Figure 2-9 14
Standoffs, male - female (6)
85791
Fuse, 2.5 Amp 5x20mm
12420 Plus Installation Manual
Table 2-6. Replacement Parts
Page 17
20
21
22
To J7 header on PC Board (6)
From Power Cord (16)
23
28
5-pin overlay
8-pin overlay
Ground Wire Detail
25
27
24
To backplate (3)
From power cord (16)
24
25
27
Figure 2-8. 420 Plus Overlay and Power Supply
420 Plus Installation Manual - Installation13
Page 18
1
2
3
4
5
6
7
8
9
10
12
11
13
14
15
16
17
18
32
30
14420 Plus Installation Manual
Figure 2-9. 420 Plus Enclosure, Backplate and CPU Board
Page 19
3.0Configuration
To configure the 420 Plus indicator, the indicator must
be placed in setup mode. The setup switch is accessed
by removing the large fillister head screw on the
enclosure bottom. Switch position is changed by
inserting a screwdriver or similar tool into the access
hole and pressing the setup switch.
When the indicator is placed in setup mode, the word
CONFIG is shown on the display. The CONFIG menu is
the first of ten main menus used to configure the
indicator. Detailed descriptions of these menus are
given in Section 3.2 on page 17. When configuration is
complete, return to the CONF
IG menu and press the
(ZERO) key to exit setup mode, then replace the setup
switch access screw.
3.1Configuration Methods
The 420 Plus indicator can be configured by using the
front panel keys to navigate through a series of
configuration menus or by sending commands or
configuration data to the EDP port. Configuration
using the menus is described in Section 3.1.3 on
page 16.
Configuration using the EDP port can be accomplished
ing either the EDP command set described in
us
Section 5.0 or the
3.1.1Revolution Configuration
The Revolution configuration utility provides the
preferred method for configuring the
indicator. Revolution runs on a personal computer to
set configuration parameters for the indicator. When
Revolution configuration is complete, configuration
data is downloaded to the indicator.
Revolution® software.
420 Plus
Revolution
supports both uploading and downloading
of indicator configuration data. This capability allows
configuration data to be retrieved from one indicator,
edited, then downloaded to another.
To use
Revolution, do the following:
1. Install the
Revolution module on an
IBM-compatible personal computer running
®
Windows
98 or later. Minimum system
requirements are 4MB of extended memory
and at least 5MB of available hard disk space.
2. With both indicator and PC powered off,
connect
the PC serial port to the indicator EDP
port.
3. Power up the PC and the indicator. Use the
setup switch
to place the indicator in setup
mode.
4. Start the
Figure 3-1 shows an example of one of the
Revolution program.
Revolution
configuration displays.
Revolution provides online help for each of its
configuration displays. Parameter descriptions
provided in this manual for front panel configuration
can also be used when configuring the indicator using
Revolution: the interface is different, but the
parameters set are the same.
3.1.2EDP Command Configuration
The EDP command set can be used to configure the
420 Plus indicator using a personal computer, terminal,
or remote keyboard. Like
Revolution, EDP command
configuration sends commands to the indicator EDP
port; unlike
Revolution, EDP commands can be sent
using any external device capable of sending ASCII
characters over a serial connection.
EDP commands duplicate the fu
nctions available using
the indicator front panel and provide some functions
not otherwise available. EDP commands can be used to
simulate pressing front panel keys, to configure the
indicator, or to dump lists of parameter settings. See
Section 5.0 on page 33 for more information about
using the EDP command set.
When moving thr ough values below the first menu level, press to r etur n to the level
above. Press or to move to the next parameter on the level above
1st Level
Parameter
2
nd
Level
Parameter
2
nd
Level
Parameter
Value
Value
The 420 Plus indicator can be configured using a series of menus accessed through the indicator front panel when
the indicator is in setup mode. Tab le 3-1 summarizes the functions of each of the main menus.
MenuMenu Function
CONFIGConfigurationConfigure grads, zero tracking, zero range, motion ban
and digital filtering parameters.
FORMATFormatSet format of primary and secondary units, display rate.
CALIBRCalibrationCalibrate indicator. See Section 4.0 on page 30 for calibration procedures.
SERIALSerialConfigure EDP and printer serial ports.
PROGRMProgramSet power-up mode, regulatory mode,
and consecutive number values.
PFORMTPrint FormatSet print format used for gross and net tickets. See Section 6.0 on page 39 for more
information.
SETPNTSetpointConfigure Setpoints and digital outputs.
DIG INDigital InputAssign digital input functions.
ALGOUTAnalog OutputConfigure analog output module. Used only if
analog output option is installed.
VERSVersionDisplay installed software version number.
Table 3-1. 420 Plus Menu Summary
d, overload, tare function, sample rate,
Figure 3-2. Front Panel Key Functions in Setup Mode
Four front panel keys are used as directional keys to navigate through the menus in setup mode (see Figure 3-2).
UNITS ( ) and PRINT ( ) keys scroll left and right (horizontally) on the same menu level; ZERO ( ) and
The
GROSS/NET ( ) move up and down (vertically) to different menu levels. The TAR E key ( ) serves as an Enter key
for selecting parameter values within the me
nus. A label above each of these keys identifies the direction provided
by the key when navigating through the setup menus.
Figure 3-3. Setup Mode Menu Navigation
16420 Plus Installation Manual
Page 21
To select a parameter, press or to scroll left or
When editing numeric values, press ort o change the
digit selected. Press or to increment or decrement the
value of the selected digit, or use the numeric keypad.
Press to save the value entered and return to the level above.
XXXXXXX XXXXXXX XXXXXXX XXXXXXX
GRADS
10000
0
number
ZTRKBN ZRANGE
1.900000
MOTBAN
FS+2%
FS+9D
FS+1D
FS
OVRLOA
1
4
2
8
DIGFL1
32
16
64
number
1
4
2
8
DIGFL2
32
16
64
1
4
2
8
DIGFL3
32
16
64
BOTH
PBTARE
NOT ARE
KEYED
T AREFN
ALGOUT DIG IN SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
8OUT
32OUT
16OUT
64OUT
DFSENS DFTHRH
2OUT
128OUT
4OUT
2DD
NONE
5DD
20DD
10DD
50DD
200DD
100DD
250DD
15HZ
60HZ
30HZ
7.5HZ
SMPRAT
VERS
number
1
number
right until the desired menu group appears on the
displa
y, then press
to move down to the submenu or
parameter you want. When moving through the menu
parameters, the default or pre
viously selected value
appears first on the display.
To change a parameter value, scroll left or right to view
the values
appears on the display, press
for that parameter. When the desired value
to select the value and
move back up one level. To edit numerical values, use
e navigation keys to select the digit and to increment
th
or decrement the value or use the numeric keypad. (see
Figure 3-4. Editing Procedure for Numeric Values
Figure 3-4).
3.2Menu Structures and Parameter Descriptions
The following sections provide graphic representations of the 420 Plus menu structures. In the actual menu
structure, the settings you choose under each parameter are arranged horizontally. To save page space, menu
choices are shown in vertical columns. The factory default setting appears at the top of each column in bold type.
Most menu diagrams are accompanied by a table that describes all parameters
with that menu. Default parameter values are shown in bold type.
3.2.1Configuration Menu
and parameter values associated
Figure 3-5. Configuration Menu
420 Plus Installation Manual - Configuration17
Page 22
CONFIG Menu
Note
ParameterChoices
Level 2 submenus
GRADS10000
number
Description
Graduations. Specifies the number
the range 1–100 000 and should be consistent with legal requirements and environmental
limits on system resolution.
of full scale graduations. The value entered must be in
ZTRKBND0
number
ZRANGE1.900000
number
MOTBAND1
number
OVRLOAFS+2%
FS+1D
FS+9D
FS
SMPRAT15HZ
30HZ
60HZ
7.5HZ
DIGFL1
DIGFL2
DIGFL3
DFSENS8OUT
2
4
8
16
32
64
1
16OUT
32OUT
64OUT
128OUT
2OUT
4OUT
To calculate GRADS, use the formula, GRADS = Capaci
Display divisions for primary and sec
Automatically zeroes the scale when within the
within the ZRANGE and scale is at standstill. Specify the zero tracking band in ± display
divisions. Maximum legal value varies depending on local regulations.
For scales using linear calibration, do not set the zero tracking band to
a value greater than that specified for the first linearization point.
Selects the range within which the scale can be
1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use the default value for legal-for-trade applications.
Sets the level, in display divisions, at
detected for 1 second or more, the standstill symbol lights. Some operations, including
print, tare, and zero, require the scale to be at standstill. Maximum legal value varies
depending on local regulations.
If this parameter is set to 0, the standstill annunci
operations including zero, print, and tare will be performed regardless of scale motion. If 0
is selected, ZTRKBND must also be set to 0.
Overload. Determines the point at whi
message is displayed. Maximum legal value varies depending on local regulations.
Sample rate. Selects measurement rate, in samples per second, of the analogconverter. Lower sample rate values provide greater signal noise immunity.
Digital filtering. Selects the digital filtering rate used
vibration from the immediate area of the scale.
Choices indicate the number of A/D
displayed reading. A higher number gives a more accurate display by minimizing the
effect of a few noisy readings, but slows down the settling rate of the indicator. See
Section 7.7 on page 49 for more information on digital filtering.
Digital filter cutout sensitivity. Specifies the numbe
outside the filter threshold (DFTHRH parameter) before digital filtering is suspended. If
NONE is selected, the filter is always enabled.
cifies the filter threshold, in display divisions. When a
specified number of consecutive scale readings (DFSENS parameter) fall outside of this
threshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.
Tare function. Enables or disables push-button and keyed tares. Possible values are:
BOTH:Both push-button and keyed tares are enabled
Table 3-2. Configuration Menu Parameters (Continued)
FORMAT Menu
ParameterChoicesDescription
Level 2 submenus
PRIMARDECPNT
DSPDIV
UNITS
SECNDRDECPNT
DSPDIV
UNITS
MULT
Figure 3-6. Format Menu
Specifies the decimal position, display divisions, and
Level 3 submenu parameter descriptions.
Specifies the decimal position, display divisions, units, and conversion
secondary units. See Level 3 submenu parameter descriptions.
Table 3-3. Format Menu Parameters
420 Plus Installation Manual - Configuration19
units used for the primary units. See
multiplier used for the
Page 24
Note
Note
FORMAT Menu
ParameterChoicesDescription
DSPRAT250MS
500MS
750MS
1SEC
1.5SEC
2SEC
2.5SEC
3SEC
4SEC
6SEC
8SEC
Level 3 submenus
Primary Units (PRIMAR Parameter)
DECPNT888888
888880
8.88888
88.8888
888.888
8888.88
88888.8
DSPDIV1D
2D
5D
UNITSLB
KG
OZ
TN
T
G
NONE
Secondary Units (SECNDR Parameter)
DECPNT88888.8
888888
888880
8.88888
88.8888
888.888
8888.88
DSPDIV5D
1D
2D
UNITS KG
OZ
TN
T
G
NONE
LB
MULT0.453592
Enter other
ces via
choi
keyboard
Display rate. Sets the update rate for displayed values. Values are in milliseconds (MS) or
seconds (SEC).
Decimal point location. Specifies t
primary unit display. Value should be consistent with local legal requirements.
Display divisions. Selects the minimum di
Specifies primary units for displayed and printed weight. Values are: LB=pound;
KG=kilogr
Decimal point location. Determines the location of th
secondary unit display.
Display divisions. Selects the val
Specifies secondary units for displayed and pr
OZ=ounce; TN=short ton; T=metric ton; G=gram; LB=pound.
Multiplier. Specifies the conversion factor by whi
the secondary units. The default is 0.453592, which is the conversion factor for changing
pounds to kilograms. See Section 7.6 on page 48 for a list of multipliers.
To toggle between primary and secondary units, press the UNITS
am; OZ=ounce; TN=short ton; T=metric ton; G=gram
I
ndicators sold outside North America are configured with KG for both
primary and secondary units.
Multipliers are pre-configured within the indicator. Manual entry is only
necessary when NONE is selected under UNITS.
he location of the decimal point or dummy zeroes in the
vision size for the primary units displayed weight.
ue of minimum division size of the displayed weight.
Table 3-3. Format Menu Parameters (Continued)
e decimal point or dummy zeros in the
inted weight. Values are: KG=kilogram;
ch the primary units are multiplied to obtain
key.
20420 Plus Installation Manual
Page 25
3.2.3Calibration Menu
XXXXXXX XXXXXXX XXXXXXX XXXXXXX
WZERO WV AL WSPAN
WLIN
ALGOUT DIGI N SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
REZERO
VERS
WPT–> 1
WPT–> 2 WPT–> 3 WPT–> 4 WPT–> 5
Same as PT -> 1
*CAL*
Display and edit
span calibration
A/D count value
See Section 4.0 on page 30 for Calibration procedures.
Figure 3-7. Calibration Menu
CALIBR Menu
ParameterChoicesDescription
Level 2 submenus
WZERO—Display and edit the zero calibration A/D count value.
DO NOT adjust this value after WSPAN has been set!
WVAL—Display and edit the test weight value.
WSPAN—Display and edit the span calibration A/D count value.
WLINWPT->1 —
->5
WPT
Press ENTER to display and edit test weight value. Pressing ENTER again will calibrate and
display the raw A/D value. Pressing ENTER a third time will move to the next calibration point.
For millivolt calibration, press ENTER to display a
again to display and edit the millivolt value for that weight. Press ENTER a third time to
calibrate and display the raw A/D value. Press ENTER the fourth time to move to the next
point.
REZERO—Press Enter to remove an offset value from the zer
are being used during calibration.
Always use this parameter after WZERO and WSP
value. See Section 4.1 on page 30 for more information about using this parameter.
See Section 7.3 on page 44 for information about the 420 Plus serial data format.
Figure 3-8. Serial Menu
SERIAL Menu
ParameterChoicesDescription
Level 2 submenus
EDPBAUD
BITS
TERMIN
EOLDLY
ECHO
PRINTBAUD
BITS
TERMIN
EOLDLY
ECHO
STREAMOFF
EDP
PRN
BOTH
STRRTEINDUST
LFT
PRNDESEDP
PRN
BOTH
PRNMSGOFF
ON
Level 3 SubmenusEDP Port and Printer Port
BAUD9600
300
600
1200
2400
4800
19200
38400
BITS8NONE
22420 Plus Installation Manual
7EVEN
7ODD
Specifies settings for baud rate, data bits, termination
used by the EDP port.
Specifies settings for baud rate, data bits, termination
used by the printer port.
Selects the serial port used for continuous transmission. Streaming can be set for the EDP port,
Printer port, or both ports simultaneously. See Section 7.3 on page 44 for information about the
420 Plus continuous data format.
Specifies stream rate. Stream rate can be set to industrial or legal for trade.
Print destination. Selects the EDP port, printer port, or both ports sim
transmission when the PRINT key is pressed or the KPRINT EDP command is sent.
Print message. Default will be OFF. When the print key
PRINT is momentarily displayed on the remote display.
Baud rate. Selects the transmission speed for the EDP
Selects number of data bits and parity of data tran
Table 3-5. Serial Menu Parameters
characters, end-of-line delay and echo
characters, end-of-line delay and echo
ultaneously for data
is pressed and data is sent out, the word
or printer port.
smitted from the EDP or printer port.
Page 27
XXXXXXX XXXXXXX
PWRUPM
GO
OIML
REGULA
000000
number
CONSNU
DELAY
SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
000000
number
CONSTU
NTEP
CANADA
1
UID
VERS DIGIN ALGOUT
DATFMT
MMDDYY
DATE
DDMMYY
YYMMDD
YYDDMM
DATSEP
SLASH
DASH
SEMI
TIMFMT
12HOUR
TIME
24HOUR
TIMSEP
COLON
COMA
NONE
COUNT
ACCESS
SPLISIZ
ACCUM
ON/OFF
ON
ON
RTZGRD
0
Note
SERIAL Menu
ParameterChoicesDescription
S BITS1 STOP
Stop bits. Sets the number of stop bits to 1 or 2.
2 STOP
TERMINCR/LF
Termination character. Selects termination character
CR
EOLDLY000000
number
End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line is
terminated to
the beginning of the next formatted serial output. Value specified must be in the
range 0-255, in tenths of a second (10 = 1 second).
An EOL may be required for continuous transmission at slower baud rates to
ensure the receiving buffer is empty before another string is transmitted
ECHOOFF
This command enables or disables echoing of the serial commands sent to the indicator.
ON
Table 3-5. Serial Menu Parameters (Continued)
3.2.5Program Menu
for data sent from the EDP or printer port.
Figure 3-9. Program Menu
PROGRM Menu
ParameterChoicesDescription
Level 2 submenus
PWRUPMGO
DELAY
COUNTACCESS
SPLSIZ
Power up mode. In GO mode, the indicator goes into ope
up display test.
In DELAY mode, the indica
tor performs a power up display test, then enters a 30-second
warm up period. If no motion is detected during the warm-up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset and
the warm up period repeated.
Specifies whether operator has access to piece count mode and the
for parts counting. See level three submenu for parameter descriptions.
Table 3-6. Program Menu Parameters
ration immediately after a brief power
default sample size used
420 Plus Installation Manual - Configuration23
Page 28
PROGRM Menu
ParameterChoicesDescription
REGULANTEP
OIML
CANADA
NONE
CONSNU000000
number
CONSTU000000
number
UID1Specifies the unit identification number (any numeric value up to six digits).
ACCUMON/OFF
RTZGRD
DATEDATFMT
DATSEP
TIMETIMFMT
TIMSEP
Level 3 submenus
ACCESSDISABLE
ENABLE
SPLSIZ10
20
50
100
5
PCWGT
ON/OFFON
OFF
RTZGRD0
Number
DATFMTMMDDYY
DDMMYY
YYMMDD
DATSEPSLASH
DASH
SEMI
TIMFMT24HOUR
12HOUR
TIMSEPCOLAN
COMMA
Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.
OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than zero.
NONE allows tares to be acquired at any weight value.
OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at no
load. NONE allows tares to be cleared at any weight value.
NTEP and OIML modes allow a new tare to be acquired even if a tare is already present. In
CANADA mode, the previous tare must be cleared before a new tare can be acquired.
NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net mode as
long as the current weight is within the specified ZRANGE. In OIML mode, the scale must be in
gross mode before it can be zeroed; pressing the ZERO key in net mode clears the tare.
Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.
The initial value of this parameter is set to the start up value specified on the CONSTU
parameter. Changing either CONSTU or CONSNU immediately resets the consecutive number
used for printing.
Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value
used when the consecutive number is reset by sending CLRCN digital input.
Turns the accumulator on and off. Stores the count, date and time of last accumulation.
Return to zero grads to re-arm the accumulator. Default = 0.
Allows selection of date format and date separator. See Level three parameter for descriptions.
Allows selection of time format and separator. See level three parameter for descriptions.
Operator access to piece count mode. Specify DISABLE if piece count mode will not be used.
With access disabled, pressing the GROSS/NET (MODE) key toggles between gross and net
modes only.
Sample size. Specify the default size used for counting scale operations. Sample size can be
changed in counting mode during sample acquisition.
Turns the accumulator on and off.
This determines the number of grads away from 0 that it has to return to re-arm the
accumulator between weighments.
Specifies the format used to display or print the date.
Specifies the date separator character.
Specifies the format used to display or print the time.
Specifies the time separator character.
24420 Plus Installation Manual
Table 3-6. Program Menu Parameters (Continued)
Page 29
3.2.6Print Format Menu
XXXXXXX XXXXXXX XXXXXXX XXXXXXX
GFMT
Display first 6
characters of format
Same as GFMT
NFMT
ALGOUT DIGIN SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
Display and edit
active character and
ASCII value
Delete active
character
Scroll left in formatting string
Increment ASCII value of active character Decrement ASCII value of active character
Press to insert a space
before the active character
VERS
Same as GFMT
CFMT
Scroll right in formatting string
NOTE: To change the active character,
use the numeric keypad to enter the new
ASCII value and press the ENTER (TARE) key.
GROSS
NET
SETPT1
Same as SETPT1
XXXXXXX XXXXXXX
XXXXXXX XXXXXXX
VERS ALGOUT DIG IN
PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
XXXXXXX
SETPNT
KIND
number
V ALUE
HIGHER
LOWER
TRIP
number
BNDVAL
INBAND
OUTBAND
number
HYSTER
OFF
ON
ENABLE
OFF
ON
ACCESS
If TRIP = HIGHER/
LOWER
SETPT2
See Section 6.0 for information about custom print formatting.
Figure 3-10. Print Format Menu
3.2.7Setpoint Menu
Figure 3-11. Setpoint Menu
420 Plus Installation Manual - Configuration25
Page 30
SETPNT Menu
ParameterChoicesDescription
Level 2 submenus
SETPT1
SETPT2
Level 3 submenus
ENABLEOFF
KINDGROSS
VALUEnumberDisplay and edit the setpoint value
TRIPHIGHER
BNDVALnumberThe band value for either INBAND or OUTBND trip setpoints. BNDVAL is ignored unless trip is
HYSTERnumberSpecifies a band around the setpoint value that must be
ACCESSOFF
ENABLE
KIND
VALUE
TRIP
BNDVAL
ON
NET
LOWER
INBAND
OUTBND
ON
Specify settings for setpoint enable, kind, weight value,
setpoint
Turn setpoint ON or OFF.
Specifies the setpoint kind and determines wheth
weight.
GROSS or NET setpoint kinds can be used.
Trips the setpoint when the weight
outside of the band value. LOWER means the output is active until you reach weight. If trip is
HIGHER, the output is active when the setpoint is met or exceeded. INBAND means the
output is active if the weight is within the band value. If trip is OUTBND, the output is active
when the weight is outside of the band value.
set to IN
off, can trip on.
Front panel access to setpoints. Specify OFF if setpoints will not be used. ON enables
operato
during weigh mode.
BAND or OUTBND.
r to turn setpoints on/off, change value, or change BANDVAL via front panel entry
is higher or lower than the setpoint value, or is within or
trip, and band value used by the
er function is based on GROSS or NET
exceeded before the setpoint, once
Table 3-7. Setpoint Menu Parameters
26420 Plus Installation Manual
Page 31
3.2.8Digital Input Menu
XXXXXXX XXXXXXX XXXXXXX XXXXXXX
DIGIN1
OFF
DIGIN2
ZERO
T ARE
NT/GRS
UNITS
DSPTAR
PRINT
CLRCN
OFF
ZERO
T ARE
NT/GRS
UNITS
DSPTAR
PRINT
CLRCN
ALGOUT DIGI N SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
HOLD HOLD
KBDLOC KBDLOC
VERS
CLRTAR CLRTAR
CLRACC CLRACC
ACCUM
NEWID
ACCUM
NEWID
Figure 3-12. Digital Input Menu
DIG IN Menu
ParameterChoicesDescription
Level 2 submenus
DIGIN1
DIGIN2
OFF
ZERO
TAR E
NT/GRS
UNITS
DSPTAR
PRINT
CLRCN
KBDLOC
HOLD
CLRTAR
ACCUM
CLRACC
NEWID
Specifies the function activated by digital inputs 1 and
• ZERO, NT/GRS (net/gross mode toggle), TARE,
functions as the front panel keys.
• DSPTAR displays the current tare value.
• CLRCN resets the consecutive number to the value specified on th
(PROGRM menu).
• KBDLOC disables the keypad while the digital input is held low.
• HOLD holds the current display. Releasing this i
• CLRTAR clears the tare.
• ACCUM displays the accumulator.
• CLRACC clears the accumulator.
• NEWID enters the ID number.
Table 3-8. Digital Input Menu Parameters
420 Plus Installation Manual - Configuration27
2.
UNITS, and PRINT provide the same
e CONSTU parameter
nput clears the running average filter.
Page 32
3.2.9Analog Output Menu
XXXXXXX XXXXXXX ALGOUT DIGIN SETPNT XXXXXXX PROGRM PFORMT SERIAL CALIBR XXXXXXX CONFIG FORMAT
SOURCE
GROSS
0%
20%
OFFSET
ERRACT
FULLSC
HOLD
000000
number
MIN
10000
number
MAX
TWZERO
NET
TWSPAN
ZEROSC
VERS
MINNEG
OFF
ON
MAXNEG
OFF
ON
The ALGOUT menu is used only if the analog output option is installed. If the analog output option is installed,
configure all other indicator functions and calibrate the indicator (see Section 4.0) before configuring the analog
output. See Section 7.8 on page 50 for analog output calibration procedures.
Figure 3-13. Analog Output Menu
ALG OUT Menu
ParameterChoicesDescription
Level 2 submenus
SOURCEGROSS
NET
OFFSET0%
20%
ERRACTFULLSC
HOLD
ZEROSC
MIN000000
number
MINNEGOFF
ON
MAX010000
number
MAXNEGOFF
ON
TWZERO—Tweak zero. Adjust the analog output zero calibration
TWSPAN—Tweak span. Adjust the analog outp
Specifies the source tracked by the analog output.
Zero offset. Selects whether the analog output supplies voltage (0–10 V) or curr
ent (4–20 mA)
output. Select 0% for 0–10 V output; select 20% for 4–20 mA output.
Error action. Specifies how the analog output responds to system error conditions.
Possible values are:
FULLSCSet to full value (10 V or 20 mA)
HOLD:Hold current value
ZEROSC:Set to zero value (0 V or 4 mA)
Specifies the minimum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0–999 999.
Turn ON or OFF.
Specifies the maximum weight value tracked by th
e analog output. Specify a weight value (in
primary units) in the range 0–999 999.
Turn ON or OFF.
. Use a multimeter to monitor the analog
output value. Press and hold
or to adjust the output. Press to save the new value.
ut span calibration. Use a multimeter to monitor the analog
output value. Press and hold or to adjust the output. Press to save the new value.
28420 Plus Installation Manual
Table 3-9. Analog Output Menu Parameters
Page 33
3.2.10Version Menu
XXXXXXX ALGOUT XXXXXXX XXXXXXX
Software
version
VERS
DIGI N SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
The VERS menu is used to check the software version and reg version installed in the indicator. You can also check
the indicator model.
Figure 3-14. Version Menu
420 Plus Installation Manual - Configuration29
Page 34
4.0Calibration
XXXXXXX XXXXXXX XXXXXXX XXXXXXX
WZERO WV AL WSPAN
WLIN
ALGOUT DIGI N SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT
REZERO
VERS
WPT–> 1
WPT–> 2 WPT–> 3 WPT–> 4 WPT–> 5
Same as PT -> 1
*CAL*
Display and edit
span calibration
A/D count value
The 420 Plus can be calibrated using the front panel, EDP commands, or the Revolution® configuration utility.
Each method consists of the following steps:
•Zero calibration
•Entering the test weight value
The following sections describe the
calibration procedure for each of the calibration methods.
•Span calibration
•Optional rezero calibration for test weights
u
sing hooks or chains.
4.1Front Panel Calibration
To calibrate the indicator using the front panel,
1. Place the indicator in setup mode (display
reads
CONFIG) and remove all weight from the
scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.
2. Press
Figure 4-1). Press
(WZERO).
3. With
zero. The indicator displays
calibration is in progress. When complete, the
A/D count for the zero calibration is displayed.
DO NOT adjust this value after WSPAN has
been set! Press
calibration value and go to the next prompt
WVAL).
(
4. With
stored calibration weight. Use the procedure
shown in Figure 3-4 15 to enter the actual
value of the test weights to be used for the span
calibration. Press
value and go to span calibration (WSPAN).
5. With
the scale and press
indicator displays
progress. When complete, the A/D count for
the span calibration is displayed. Press
to store the span calibration value and go to the
ne
until the display reads CALIBR (see
WZERO displayed, press to calibrate
again to save the zero
WVAL displayed, press to show the
to store the entered WVAL
WSPAN displayed, place test weights on
*CAL* while calibration is in
xt prompt (
REZERO).
Figure 4-1. Calibration (CALIBR) Menu
to go to zero calibration
*CAL* while
to calibrate span. The
again
6. Five-point linearization (using the WLIN
parameter) provides increased scale a
ccuracy
by calibrating the indicator at up to five
additional points between the zero and span
calibrations.
Linearization is optional: if you choose not to
perform linearization, skip the WLIN
parameter; if linearization
values have
previously been entered, these values are reset
to zero during calibration. To perform
linearization, follow the procedure below:
Standard Calibration: With
Press
PT-> 1). Place test weights on the scale and
(
press
to go to the first linearization point
or ENTER.
Enter the actual test weight va
WLIN displayed,
lue, then press
to calibrate. The indicator displays *CAL* while
calibration is in progress. When complete, the
A/D count for the linear calibration is
displayed. Press
again to save the calibration
value and go to the next prompt (PT-> 2).
Repeat for up to five linearization points. To
exit the linearization parameters, press
to
return to WLIN.
Millivolt Calibration: If millivolt calibration is
enabled press
weight value, press
the millivolt value for that weight, press
to display and edit the test
again to display and edit
again to calibrate and display the raw A/D
count, press
again to move to the next
calibration point.
30420 Plus Installation Manual
Page 35
7. The rezero function is used to remove a
Note
calibration offset when hooks or chains are
used to hang the test weights.
• If no other apparatus was used to hang the test
weights during calibration, remove the test
weights and press to return to the CALIBR
menu.
• If hooks or chains were used during
calibration, re
move these and the test weights
from the scale. With all weight removed, press
to rezero the scale. This function adjusts the
zero and span calibration values. The indicator
lays
disp
*CAL* while the zero and span
calibrations are adjusted. When complete, the
adjusted A/D count for the zero calibration is
displayed. Press
to save the value, then press
to return to the CALIBR menu.
8. Press
press
until the display reads CONFIG, then
to exit setup mode.
4.2EDP Command Calibration
To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.3.5 on page 10 for
EDP port pin assignments; see Section 5.0 on page 33
for more information about using EDP commands.
Once the indicator is connected to
1. Place the indicator in setup mode (display
reads
CONFIG) and remove all weight from the
scale platform. If your test weights require
hooks or chains, place them on the scale for
zero calibration.
2. Send the WZERO EDP command to calibrate
. The indicator displays *CAL* while
zero
calibration is in progress.
3. Place test weights on the scale and use the
AL command to enter the test weight value
WV
in the following format:
WVAL=nnnnnn<CR>
4. Send the WSPAN EDP command to calibrate
span. The indicator displays
calibration is in progress.
5. Up to five linearization points can be
calibra
ted between the zero and span
calibration values. Use the following
commands to set and calibrate a single
linearization point:
WLIN.V1=nnnnn<CR>
WLIN.C1<CR>
the sending device,
*CAL* while
6. To remove an offset value, clear all weight
from the sc
ale, including items used to hang
test weights, then send the REZERO EDP
command. The indicator displays
*CAL* while
the zero and span calibrations are adjusted.
7. Send the KUPARROW EDP command to exit
set
up mode.
4.3Revolution Calibration
To calibrate the indicator using Revolution, the
indicator EDP port must be connected to a PC running
Revolution configuration utility.
the
Use the following procedure to calibrate the indicator:
1. Select
tools menu.
2.
Revolution uploads calibration data from the
indicator then presents the information in the
calibration wizard, as shown in Figure 4-2
Figure 4-2. Revolution Calibration Display
3. Enter the Value of Test Weight to be used for span
calibration then click OK.
4. The Zero Calibration dialog box prompts you
to remove
scale and click
5. When zero calibration is complete, the Span
Calibration dialog box prompts you to place
test weights on the scale for span calibration.
Place tests weights on the scale then click
Calibration Wizard from the Revolution
all weight from the scale. Clear the
OK to begin zero calibration.
If your test weights require hooks or
chains, place them on the scale for zero
calibration
.
OK.
The WLIN.V1 command sets the test weight
value (nnnnn) for linearization point 1. The
WLIN.C1 command calibrates the point.
Repeat using the WLIN.Vx and WLIN.Cx
commands as required for additional
linearization points.
420 Plus Installation Manual - Calibration31
Page 36
6. When span calibration is complete, a dialog box
Note
asks whether you wish to perform linear
calibration. Click
Yes to perform linear
calibration for up to five linearization points, or
click No to complete calibration (see step 8).
7. On the Linear Calibration display, select the
po
int (1–5) to calibrate, then click
Place test weights on scale then click
Calibrate.
OK.
When prompted, enter the test weight value
then press Enter. Repeat for additional
linearization points, then click
Exit to return to
the Indicator Calibration display.
8. When calibration is complete, the
New Settings
fields of the Indicator Calibration display are
filled in. Click
return to the
the previous calibration values, click
Settings
.
Exit to save the new values and
Revolution main menu; to restore
Restore
4.4More About Calibration
The following topics provide additional information
about compensating for environmental factors
(Section 4.4.1) and diagnostic information for
determining expected zero and span coefficients.
4.4.1Adjusting Final Calibration
Calibration may be affected by environmental factors
including wind, vibration, and angular loading. For
example, if the scale is calibrated with 1000 lb, a strain
test may determine that at 2000 lb the calibration is 3 lb
high. In this case, final calibration can be adjusted by
tweaking WVAL to 998.5 lb. This adjustment provides
a linear correction of 1.5 lb per 1000 lb.
To adjust the final calibration, return to the
prompt and press to show the test weight value.
Press
or to adjust calibration up or down. Press
to save the value, then press to return to the CALIBR
menu.
4.4.2Zero Deadload A/D Counts
Table 4-1 lists the ideal A/D counts that result from
input signals of 0–3.0 mV with zero deadload. Actual
values will typical
ly be higher than the values shown in
Table 4-1 but the ideal values can be used when
calibrating the indicator with
no attached scale.
WVAL
Input Signal (mV)Raw A/D Counts
0.0 mV/V
0.5 mV/V
1.0 mV/V
1.5 mV/V
2.0 mV/V
2.5 mV/V
3.0 mV/V
Table 4-1. Ideal A/D Raw Counts
When Raw A/D Count is displayed, the six most
significant digits appear. Scroll left or right to
see the full value.
167,840
335,613
503,377
671,143
838,908
1,100,668
1,174,446
32420 Plus Installation Manual
Page 37
5.0EDP Commands
The 420 Plus indicator can be controlled by a personal
computer or remote keyboard connected to the
indicator EDP port. Control is provided by a set of EDP
commands that can simulate front panel key press
functions, display and change setup parameters, and
perform reporting functions. The EDP port provides
the capability to print configuration data or to save that
data to an attached personal computer. This section
describes the EDP command set and procedures for
saving and transferring data using the EDP port.
5.1The EDP Command Set
The EDP command set can be divided into five groups:
key press commands, reporting commands, the
RESETCONFIGURATION special function
command, parameter setting commands, and transmit
weight data commands.
When the indicator processes an EDP command, it
sponds with the message
re
verifies that the command was received and has been
executed. If the command is unrecognized or cannot be
executed, the indicator responds with
The following sections list the commands and
command syntax use
5.1.1Key Press Commands
d for each of these groups.
Key press EDP commands (see Table 5-1) simulate
pressing the keys on the front panel of the indicator.
st commands can be used in both setup and
Mo
weighing mode. Several of the commands serve as
“pseudo” keys, providing functions that are not
represented by a key on the front panel.
For example, to enter a 15-pound tare weight using
EDP commands:
1. Type K1 and press
2. Type K5 and press
3. Type KTARE and press
shifts to net mode when the tare is entered.
OK. The OK response
??.
ENTER (or RETURN).
ENTER.
ENTER. The display
CommandFunction
KZEROIn weighing mode, press the ZERO key
KGROSSNETIn weighing mode, press the GROSS/
NET key
KGROSSGo to gross mode (pseudo key)
KNETGo to net mode (pseudo key)
KTAREPress the TARE key
KUNITSIn weighing mode, pr
KPRIMGo to primary units (pseudo key)
KSECGo to secondary units (pseudo key)
KPRINTIn weighing mode, pr
KLEFTARROWIn setup mode, move left in the menu
KRIGHTARROWIn setup mode, move right in the menu
KUPARROWIn setup mode, move up in the menu
KDOWNARROWIn setup mode, move down in the menu
KEXITIn setup mode, exits to normal mode
KCOUNTGo to piece count mode (pseudo key)
KTIMEPress the TIME/DATE key once
KDATEPress the TIME/DATE key twice
KDISPTAREPress the DISPLAY TARE key
KDISPACCUMDisplays accumulator
KCLRACCUMClears accumulator
KCLRPress the CLEAR key
K0Press number 0 (zero)
K1Press number 1
K2Press number 2
K3Press number 3
K4Press number 4
K5Press number 5
K6Press number 6
K7Press number 7
K8Press number 8
K9Press number 9
KDOTPress the decimal point (.)
KENTERPress the ENTER key
ess the UNITS key
ess the PRINT key
Table 5-1. EDP Key Press Commands
420 Plus Installation Manual - EDP Commands33
Page 38
5.1.2Reporting Commands
Note
Reporting commands (see Table 5-2) send specific
information to the EDP port. These commands can be
sed in both setup mode and normal mode.
u
CommandFunction
DUMPALLList all parameter values
VERSION
PWrite current displayed weight with units
ZZWrite current weight and annunciator
SWrite one frame of stream format. See
Table 5-2. EDP Reporting Commands
5.1.3The RESETCONFIGURATION Command
420 Plus software version
Write
fier. See Section 7.2 on page 43 for
identi
more information.
status. See Section 7.2 on page 43 for
more information.
Ta bl e 5-12 on page 37 for information
about the SX and EX commands.
The RESETCONFIGURATION command can be used
to restore all configuration parameters to their default
values. Before issuing this command, the indicator
must be placed in test mode (press and hold setup
switch for approximately three seconds to show TEST
menu).
This command is equivalent to
using the DEFLT
function on the TEST menu. See Section 7.9 on
page 51 for more information about test mode and
using the TEST menu.
All load cell calibration settings are lost when
the RESETCONFIGURATION command is run.
5.1.4Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 5-3 through 5-11).
Current configuration para
meter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command<ENTER>
Most parameter values can be changed in setup mode
only. Use the following command syntax when
changing parameter values:
command=value<ENTER>
where value is a number or a parameter value. Use no
spaces before or after the equal (=) sign. If you type an
incorrect command or value, the display reads
??.
Changes to the parameters are saved as they are entered
but typically do not take effect until you exit setup
mode.
For example, to set the motion band parameter
to 5,
type the following:
MOTBAND=5D<ENTER>
To view choices, type the following:
MOTBAND=?<ENTER>
5.1.5Soft Reset
The command RS will reset the indicator without
losing configuration and calibration. It does reset
consecutive number and count mode.
EDP.BAUDEDP port baud rate300, 600, 1200, 2400, 4800, 96
EDP.BITSEDP port data bits/parity8NONE, 7EVEN, 7ODD
EDP.TERMINEDP port termination characterCR/LF, CR
EDP.EOLDLYEDP port end-of-line delay0–255 (0.1-second
EDP.ECHOEDP port echoON, OFF
PRN.BAUDPrinter port baud rate300, 600, 1200, 2400, 4800, 96
PRN.BITSPrinter port data bits/parity8NONE, 7EVEN, 7ODD
PRN.TERMINPrinter port termination characterCR/LF, CR
PRN.EOLDLYPrinter port end-of-line delay0–255 (0.1-second
PRN.ECHOPrinter port echoON, OFF
STREAMStreaming portOFF, EDP, PRN
STRRTEStream rateINDUST, LFT
PRNDESTPrint destinationEDP, PRN
PRNMSGPrint messageOFF, ON
ing
NFMTNet demand print fo
CFMTCount demand print format
ing
str
rmat string
See Section 6.0 on page 39 for detailed information
Table 5-8. PFORMT EDP Commands
CommandDescriptionVal ues
SP1SP2
SP1.ENABLESP2.ENABLESetpoint enableOFF, ON
SP1.KINDSP2.KINDSetpoint kindGROSS, NET
SP1.VALUESP2.VALUESetpoint valuenumber
SP1.TRIPSP2.TRIPTr ipHigher, Lower, Inband, Outbnd
SP1.BNDVALSP2.BNDVALBand valuenumber
SP1.HYSTERSP2.HYSTERHysteresisnumber
SP1.ACCESSSP2.ACCESSSetpoint accessOFF, ON
Table 5-9. SETPNTS EDP Commands
CommandDescriptionValues
DIGIN1
DIGIN2
Digital input functionOFF, ZERO, TARE, NT/GRS, UNITS,
KBDLOC, HOLD, CLRTAR, ACCUM, CLRACC, NEWID
DSPTAR, PRINT, CLRCN,
36420 Plus Installation Manual
Table 5-10. DIG IN EDP Commands
Page 41
CommandDescriptionValues
SOURCEAnalog output source GROSS, NET
OFFSETZero offset0%, 20%
ERRACTError actionFULLSC, HOLD, ZEROSC
MINMinimum value tracked0–999 999
MINNEGMinimum negativeOFF, ON
MAXMaximum value tracked0–999 999
MAXNEGMaximum negativeOFF, ON
TWZEROZero calibration0–16 383
TWSPANSpan calibration0–16 383
Table 5-11. ALGOUT EDP Commands
5.1.6Normal Mode Commands
The serial transmit weight data commands (see Tab le 5-12) transmit data to the EDP port on demand. The SX and
EX commands are valid only in normal operating mode; all oth
er commands are valid in either setup or normal
mode.
CommandDescriptionResponse Format
UIDSet unit IDnnnnnnn
SDSet dateMMDDYY, DDMMYY, YYMMDD (enter using DATEFMT
specified)
STSet timehhmm (enter using 24-hour format)
SXStart EDP streamingOK or ??
EXStop EDP streamingOK or ??
RSReset system—
XATransmit accumulator valuennnnnn UU
XGTransmit gross weight in
XNTransmit net weight in displayed units
XTTransmit tare weight in displayed units
XG2Transmit gross weight in non-displayed units
XN2Transmit net weight in non-displayed units
XT2Transmit tare weight in non-displayed units
XW Transmit current piece weight nnnnnn UU
XCTransmit current part countnnnnnn PC
XEQuery system error conditionsnnnnn nnnnn
displayed units
where nn
See Section 7.1.2 on page 43 for detailed information about
the XE command response format.
nnnn is the weight value, UU is the units.
Table 5-12. Normal Mode Commands
420 Plus Installation Manual - EDP Commands37
Page 42
5.2Saving and Transferring Data
Note
Connecting a personal computer to the 420 Plus EDP
port allows you to save indicator configuration data to
the PC or to download configuration data from the PC
to an indicator. The following sections describe the
procedures for these save and transfer operations.
5.2.1Saving Indicator Data to a Personal Computer
Configuration data can be saved to a personal computer
connected to the EDP port. The PC must be running a
communications program such as PROCOMMPLUS
See Section 2.3.5 on page 10 for information about
serial communications wiring and EDP port pin
ignments.
ass
When configuring the indicator
, ensure that the values
set for the BAUD and BITS parameters on the SERIAL
menu match the baud rate, bits, and parity settings
configured for the serial port on the PC. Set the
PRNDEST parameter to EDP.
To save all configuration data, place the indicator in
setup mode and
the indicator. The
send the DUMPALL EDP command to
420 Plus responds by sending all
configuration parameters to the PC as ASCII-formatted
text.
®
5.2.2Downloading Configuration Data from PC to
Indicator
Configuration data saved on a PC or floppy disk can be
downloaded from the PC to an indicator. This
procedure is useful when a number of indicators with
similar configurations are set up or when an indicator is
replaced.
To download configuration data, connect the PC to the
port as described in Section 5.2.1. Place the
EDP
indicator in setup mode and use the PC
.
communications s
oftware to send the saved
configuration data to the indicator. When transfer is
complete, calibrate the indicator as described in
Section 4.0.
Calibration settings are included in the
configuration data downloaded to the indicator.
the receiving indicator is a direct
If
replacement for another
is not changed, recalibration is not required.
When downloading configurations that include changed
ial communications settings, edit the data file to place
ser
the serial communications changes at the end of the file.
Communication between the PC and indicator will be lost
once the indicator receives settings for baud rate (BAUD
parameter) or data bits and parity (BITS parameter) that do
not match those configured for the PC.
420 Plus and the attached scale
38420 Plus Installation Manual
Page 43
6.0Print Formatting
Note
Note
The 420 Plus provides three print formats, GFMT,
NFMT, and CFMT that determine the format of the
printed output when the
PRINT key is pressed or when a
KPRINT EDP command is received. If a tare has been
entered or acquired, NFMT is used; otherwise, GFMT
is used.
Each print format can be customized to include up to
0 characters of information, such as company name
30
and address, on printed tickets. You can use the
indicator front panel (PFORMT menu), EDP
commands, or the
Revolution
®
configuration utility to
customize the print formats.
Display will only show capital letters, press the
down arrow to view the ASCII character
decimal value on
to Tab le 7-4 on page 45 and Tab le 7-5 on
page 46 to view the ASCII character charts.
the second menu level. Refer
6.1Print Formatting Commands
Table 6-1 lists commands you can use to format the
gross, net and count print formats
in the format strings must be enclosed between < and >
delimiters. Any characters outside of the delimiters are
printed as text on the ticket. Text characters can include
any ASCII character that can be printed by the output
device.
CommandDescription
<G>Gross weight in displayed units
<G2>Gross weight in non-displayed units
<N>Net weight in displayed units
<N2>Net weight in non-displayed units
<T>Tare weight in displayed units
<T2>Tare weight in non-displayed units
<A>Accumulated weight in displayed units
<AC>Number of accumulator events (5-digit counter)
<AD>Date of last accumulator event
<AT>Time of last accumulator event
<UID>Unit ID number
<CN>Consecutive number
>New line (nn = number of termination (<CR/LF>
<NLnn
or <CR>) characters)*
<C>Current piece count
<W>Current piece weight
>Space (nn = number of spaces)*
<SPnn
<SU>Toggle weight data format (formatted/
atted)**
unform
<TI>Time
. Commands included
CommandDescription
<DA>Date
<TD>
Gross, net, and tare weights are 9 digits in length, including
sign (10 digits with decimal point), followed by a space and a
two-digit units identifier. Total field length with units identifier
is 12 (or 13) characters.
ID and consecutive number (CN) fields are 1–6 characters in
length, as required.
* If nn is not specified, 1 is assumed. Value must be in the
range 1–99.
** After receiving an SU command, the indicator sends
unformatted data until the next SU command is received.
Unformatted data omits decimal points, leading and trailing
characters.
Time & Date
Table 6-1. Print Format Commands
The default GFMT, NFMT and CFMT print formats
use only the new line (<NL>) command and the
commands for gross, net, and tare weights in displayed
units (<G>, <N>, and <T>). The default
420 Plus print
formats are shown in Table 6-2:
FormatDefault Format StringSample Output
GFMT
NFMT
CFMT
ID numbers included in the print format string (<UID>
comma
The 300-character limit of each print format string includes
the output field length o
not the command length. For example, if the indicator is
configured to show a decimal point, the <G> command
generates an output field of 13 characters: the
10-character weight value (including decimal point), one
space, and a two-digit units identifier.
PT (preset tare) is added to the
keyed in.
<G> GROSS<NL>2046.81 lb GROSS
<G> GROSS<NL>
<T> TARE<NL>
<N> NET<NL>
<C><NL>512 PC
4053.1 lb GROSS
15.6 lb TARE
4037.5 lb NET
Table 6-2. GFMT and NFMT Formats
The <G2>, <N2>, and <T2> commands listed in
Tabl e 6-1 print the gross, net, and tare weights
in non-displayed units—that is, in the units not
currently displayed
Scroll right in format stringScroll left in format string
Increment ASCII value of active characterDecrement ASCII value of active character
Pressto insert a space
befor
e the active character
VERS
Same as GFMT
CFMT
Note
Note
The following sections describe procedures for
customizing the GFMT, NFMT and CFMT formats
using the EDP port, the front panel (PFORMT menu),
and the
Revolution configuration utility.
A ticket printed using this format might look like the
following:
FINE TRANSFER CO
32400 WEST HIGHWAY ROAD
SMALLTOWN
6.2.1Using the EDP Port
With a personal computer, terminal, or remote
keyboard attached to the
420 Plus EDP port, you can
use the EDP command set to customize the print
format strings.
To view the current setting of a
format string, type the
name of the string (GFMT, NFMT or CFMT) and press
ENTER. For example, to check the current configuration
of the GFMT format, type GFMT and press
ENTER.
The indicator responds by sending the current
configuration for the gross format:
GFMT=<G> GROSS<NL>
To change the format, use the GFMT, NFMT or CFMT
EDP command followed by an equals sign (=) and the
modified print format string. For example, to add the
name and address of a company to the gross format,
you could send the following EDP command:
Indicator must be placed in the setup mode,
CONFIG is
GFMT=FINE TRANSFER CO<NL>32400 WEST
HIGHWAY ROAD<NL>SMALLTOWN<NL2><G>
GROSS<NL>
shown on the display.
1345 lb GROSS
6.2.2Using the Front Panel
If you have no access to equipment for communication
through the EDP port or are working at a site where
such equipment cannot be used, you can use the
PFORMT menu (see Figure 6-1) to customize the print
formats.
Using the PFORMT menu, you can edit the print
format strings by
changing the decimal values of the
ASCII characters in the format string.
Lowercase letters and some special characters
cannot be displayed on the 420 Plus front
panel (see the ASCII character charts on
page 45 and page 46) and are shown as blanks. The
Plus
can send or receive any ASCII character; the
character printed depends on the particular ASCII
character set implemented for the receiving device.
420
Figure 6-1. PFORMT Menu, Showing Alphanumeric Character Entry Procedure
40420 Plus Installation Manual
Page 45
6.2.3Using Revolution
The Revolution configuration utility provides a print formatting grid with a tool bar. The grid allows you to
construct the print format without the formatting commands (<NL> and <SP>) required by the front panel or EDP
command methods. Using
Revolution, you can type text directly into the grid, then select weight value fields from
the tool bar and place them where you want them to appear on the printed ticket.
Figure 6-2 shows an example of the
Revolution print formatting grid.
Figure 6-2. Revolution Print Format Grid
420 Plus Installation Manual - Print Formatting41
Page 46
7.0Appendix
7.1Error Messages
The 420 Plus indicator provides a number of error messages. When an error occurs, the message is shown on the
indicator LED display. Error conditions can also be checked remotely by using the XE EDP command as described
in Section 7.1.2 on page 43.
7.1.1Displayed Error Messages
The 420 Plus provides a number of front panel error messages to assist in problem diagnosis. Tab l e 7-1 lists these
messages and their meanings.
Error MessageDescriptionSolution
E A/DA/D physical errorCall Rice Lake Weighing Systems (RL
VIREEVirgin EEPROMUse TEST menu to perform DEFLT (restore defaults) procedure, then
OVERFLOverflow errorWeight value too large to be displayed.
------Gross > overload limitGross value exceeds overload limit. Check conf
______A/D underrangeA/D reading < –4 mV. Check scale for binding or damage.
RNGERRGRADS > 100,000
WVAL > 100,000
CNTERRCount error, insufficient sample sizeReconfigure, parts sample too sm
EEPERREEPROM errorCall Rice Lake Weighing Systems (RL
r
n XE command (see Section 7.1.2 on
page 43) to determine error type.
Call RLWS Service if unable to clear err
recurs.
input level. Overload can be caused by input signal > 45 mV or
common mode voltage > 950 mV.
Only shows up in Config mode.
WS) Service.
or by cycling power or if error
iguration or signal
all for current configuration
WS) for service
42420 Plus Installation Manual
Table 7-1. 420 Plus Error Messages
Page 47
7.1.2Using the XE EDP Command
The XE EDP command can be used to remotely query
420 Plus for the error conditions shown on the front
the
panel. The XE command returns two 5-digit numbers
in the format:
xxxxx yyyyy
where xxxxx contains a decimal representation of any
existing error conditions as described in Tab le 7-2.
If more than one error condition exists, the number
returned is the
sum of the values representing the error
conditions. For example, if the XE command returns
the number 1040, this value represents the sum of an A/
D reference error (1024) and an A/D calibration
checksum error (16).
The second number returned (
yyyyy) uses the same bit
assignments as shown in Table 7-2 to indicate whether
the test for the error condition was run. For example,
the value
yyyyy = 50815 represents the decimal
equivalent of the binary value 1100 0110 0111 1111.
Using the bit assignments in Tabl e 7-2, this value
indicates all tests were run.
Two EDP commands, P and ZZ, can be used to provide
status about the indicator. These commands are
described in the following sections.
7.2.1Using the P EDP Command
The P EDP command returns the current displayed
weight value to the EDP port, along with the units
identifier. If the indicator is in an underrange or
overload condition, the weight value is replaced with
&&&&&& (overload) or :::::: (underrange).
7.2.2Using the ZZ EDP Command
The ZZ EDP command can be used to remotely query
which annunciators are currently displayed on the
indicator front panel. The ZZ command returns the
currently displayed weight and a decimal number
representing the LED annunciators currently lit. The
format of the returned data is:
wwwwww uu zzz
where wwwwww uu is the current displayed weight and
zzz is the annunciator status value (see Tabl e 7-3).
units,
If more than one annunciator is lit, the second number
returne
annunciators.
Example: If the annunciator status value returned on
the ZZ command is 145, the gross, standstill, and lb
annunciators are lit. 145 represents the sum of the
values for the standstill annunciator (128), gross mode
annunciator (16), and the lb/primary units annunciator
(1).
d is the sum of the values representing the active
Decimal
Value
1lb/primary units
2kg/secondary units
4Count
8Tare entered
16Gross
32Net
64Center of zero
128Standstill
Table 7-3. Status Codes Returned on the ZZ Command
Annunciator
420 Plus Installation Manual - Appendix43
Page 48
7.3Continuous Output (Stream) Format
<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>
ASCII 02
(decimal)
Weight data: 7 digits, right-justified, with
decimal point, l eading zer o s upr e ssion.
Overload = ^^^^^^^
Underrange = ] ] ] ] ] ] ]
Display overflow = OVERFL
L = pounds
K = kilogram
T = ton
G = grams
O = ounces
<space> = none
Figure 7-1 shows the continuous output format sent to the 420 Plus EDP or printer port when the STREAM
parameter (SERIAL menu) is set to either EDP or PRN.
Figure 7-1. Continuous Output Data Format
44420 Plus Installation Manual
Page 49
7.4ASCII Character Chart
Use the decimal values for ASCII characters listed in Tables 7-4 and 7-5 when specifying print format strings on
420 Plus PFORMT menu. The actual character printed depends on the character mapping used by the output
the
device.
420 Plus can send or receive any ASCII character value (decimal 0–255), but the indicator display is limited to
The
numbers, upper-case, unaccented letters, and a few special characters. See Section 7.5 on page 47 for information
about the
Figure 7-2 shows the 7-segment LED character set used to display alphanumeric characters on the 420 Plus front
panel.
Figure 7-2. 420 Plus Display Characters
420 Plus Installation Manual - Appendix47
Page 52
7.6Conversion Factors for Secondary
Note
Note
Units
The 420 Plus has the capability to mathematically
convert a weight into many different types of units and
instantly display those results with a press of the
key.
Secondary units can be specified on the FORMAT
men
u using the SECNDR parameter, or by using EDP
commands.
Multipliers are preconfigured within the
indicator. Manual entry is only necessary when
E is selected under UNITS. Long tons and
grain units listed in Tab le 7-6 cannot be directly specified
as primary or secondary units. For these or other unlisted
units
Ensure that the secondary decimal point position is set
appropriately for the scale capacity in the secondary units.
If the converted value requires more digits than are
available, the indicator will display an overflow message
(OVERFL).
For example, if the primary units are short tons, secondary
units are
to 8888.88, the indicator will overflow if 5 tons or more are
applied to the scale. With 5 tons applied, and a conversion
factor of 2000, the secondary units display needs five
digits to the left of the decimal point to display the 10000 lb
secondary units value.
NON
of weight, specify NONE under UNITS.
pounds, and the secondary decimal point is set
UNITS
Primary Unitx MultiplierSecondary Unit
grains0.064799grams
0.002286ounces
0.000143pounds
0.000065kilograms
ounces437.500grains
28.3495grams
0.06250pounds
0.02835kilograms
pounds7000.00grains
453.592grams
16.0000ounces
0.453592kilograms
0.000500short tons
0.000446long tons
0.000453metric tons
short tons2000.00pounds
907.185kilograms
0.892857long tons
0.907185metric tons
grams15.4324grains
0.035274ounces
0.002205pounds
0.001000kilograms
kilograms15432.4grains
35.2740ounces
1000.00grams
2.20462pounds
0.001102short tons
0.000984long tons
0.001000metric tons
metric tons2204.62pounds
1000.00kilograms
1.10231short tons
0.984207long tons
long tons2240.00pounds
1016.05kilograms
1.12000short tons
1.01605metric tons
Multipliers in italics are preconfigured
Table 7-6. Conversion Factors
48420 Plus Installation Manual
Page 53
7.7Digital Filtering
1
4
2
8
DIGFL1
32
16
64
1
4
2
8
DIGFL2
32
16
64
1
4
2
8
DIGFL3
32
16
64
8OUT
32OUT
16OUT
64OUT
DFSENS DFTHRH
2OUT
128OUT
4OUT
2DD
NONE
5DD
20DD
10DD
50DD
200DD
100DD
250DD
DIGFL1= n1 DIGFL2= n2 DIGFL3= n3
3rd Stage
Filter A verage
2nd Stage
Filter A verage
1st Stage
Filter A verage
Displayed
Value
n3
2nd Stage
Filter A verages
n2
1st Stage
Filter A verages
n1
A/D Readings
The 420 Plus uses averaged digital filtering to reduce the effect of vibration on weight readings. Adjustable
threshold and sensitivity functions allow quick settling by suspending filter averaging, allowing the weight reading
to jump to the new value. Figure 7-3 shows the digital filter parameters on
Figure 7-3. Digital Filtering Parameters on the Configuration (CONFIG) Menu
the CONFIG menu.
Figure 7-4. Flow Diagram for 420 Plus Digital Filters
7. 7. 1D I G F L x P a r a m e t e r s
The first three digital filtering parameters, DIGFL1,
DIGFL2, and DIGFL3, are configurable filter stages
that control the effect of a single A/D reading on the
displayed weight. The value assigned to each
parameter sets the number of readings received from
the preceding filter stage before averaging (see
Figure 7-4).
7.7.2DFSENS and DFTHRH Parameters
The three digital filters can be used by themselves to
eliminate vibration effects, but heavy filtering also
increases settling time. The DFSENS (digital filter
sensitivity) and DFTHRH (digital filter threshold)
parameters can be used to temporarily override filter
averaging and improve settling time:
•DFSENS specifies the number of consecutive
The overall filtering effect can be expressed by adding
values assigned to the three filter stages:
the
DIGFL1 + DIGFL2 + DIGFL3
For example, if the filters are configured as
•DFTHRH sets a threshold value, in display
DIGFL1=4, DIGFL2=8, DIGFL3=8, the overall
filtering effect is 20 (4 + 8 + 8). With this
configuration, each A/D reading has a 1-in-20 effect on
the displayed weight value. Setting the filters to 1
effectively disables digital filtering.
scale readings that must fall
outside the filter
threshold (DFTHRH) before digital filtering is
suspended.
divisions. When
a specified number of
consecutive scale readings (DFSENS) fall
outside of this threshold, digital filtering is
suspended. Set DFTHRH to NONE to turn off
the filter override.
420 Plus Installation Manual - Appendix49
Page 54
7.7.3Setting the Digital Filter Parameters
Note
XXXXXXX XXXXXXX ALGOUT DIGIN SETPNT XXXXXXX PROGRM PFORMT SERIAL CALIBR XXXXXXX CONFIG FORMAT
SOURCE
GROSS
0%
20%
OFFSET
ERRACT
FULLSC
HOLD
000000
number
MIN
10000
number
MAX
TWZERO
NET
ZEROSC
VERS
MINNEG
OFF
ON
MAXNEG
OFF
ON
Fine-tuning the digital filter parameters greatly
improves indicator performance in heavy-vibration
environments. Use the following procedure to
determine vibration effects on the scale and optimize
the digital filtering configuration.
1. In setup mode, set all three digital filters
(DIGF
L1, DIGFL2, DIGFL3) to 1. Set
DFTHRH to NONE. Return indicator to
normal mode.
2. Remove all weight from the scale, then watch
indicator display to determine the
the
magnitude of vibration effects on the scale.
Record the weight below which all but a few
readings fall. This value is used to calculate the
DFTHRH parameter value in Step 4.
For example, if a heavy-capacity scale
duces vibration-related readings of up to 50
pro
lb, with occasional spikes to 75 lb, record 50 lb
as the threshold weight value.
3. Place the indicator in setup mode and set the
digital fi
lters (DIGFLx) to eliminate the
vibration effects on the scale. (Leave
DFTHRH set to NONE.) Reconfigure as
necessary to find the lowest effective values
for the DIGFLx parameters.
4. With optimum values assigned to the DIGFLx
parameters, calculate the DFTHRH parameter
by converting the weight value recorded
value
in Step 2 to display divisions:
threshold_weight_value / DSPDIV
In the example in Step 2, with a threshold
weight value of 50 lb and a display division
value of 5lb:
50 / 5lb = 10DD. DFTHRH should
be set to 10DD for this example.
5. Finally, set the DFSENS parameter high
enough to
ignore transient peaks. Longer
transients (typically caused by lower vibration
frequencies) will cause more consecutive
out-of-band readings, so DFSENS should be
set higher to counter low frequency transients.
Reconfigure as necessary to find the lowest
fective value for the DFSENS parameter.
ef
7.8Analog Output Calibration
The following calibration procedure requires a multimeter to measure voltage or current output from the analog
output module. If the option is not already installed, see Section 2.3.7 on page 11.
The analog output must be calibrated after the indicator itself has been configured (Section 3.0) and calibrated
(Section 4.0).
Figure 7-5. Analog Output Menu
1. Enter setup mode and go to the ALGOUT
menu (see Figure 7-5).
•Set OFFSET to 0% for 0–10 V output,
% for 4–20 mA output
20
•Set MIN to lowest weight value to be
tracked by
the analog output
•Set MAX to highest weight value to be
tracked by
the analog output
2. Connect multimeter to analog output:
•For voltage output, connect voltmeter
leads to pins three and four
•For current output, connect ammeter leads
pins one and two
to
3. Adjust zero calibration: Scroll to the
TWZ
ERO parameter. Check voltage or current
50420 Plus Installation Manual
reading on multimeter. Press and hold
to adjust the zero value up or down.
4. Adjust span calibration: Scroll to the TWSPAN
parameter
multimeter. Press and hold
. Check voltage or current reading on
or to adjust the
span value up or down.
5. Final zero calibration: Return to the TWZERO
parameter a
nd verify that the zero calibration
has not drifted. Press and hold or to
re-adjust the zero value as required.
6. Return to normal mode. Analog output
fu
nction can be verified using test weights.
or
Page 55
7.9Test Mode
Note
A/DTST
TEST
ADOFFS
ADGAIN
DEFLT
EXIT
Not used
ENTER
Move LEFT
Move RIGHT
In addition to normal and setup modes, test mode
provides a number of diagnostic functions for the
, including:
Plus
•Display raw A/D count
•Reset configuration parameters to default
lues
va
•A/D internal calibration
Test Mode is intended for factory use only.
Figure 7-6. Test Mode Menu
420
To enter test mode, press and hold the setup switch
until the front panel display shows the word
TEST. After
about three seconds, the test mode display
automatically shifts to the first test menu function, A/
DTST.
Figure 7-6 shows the Test Menu structure; Figure 7-7
shows the front panel key functions
in test mode. Note
that, because the Test Menu functions are all on a
single menu level, the
function. Press the
GROSS/NET ( ) key has no
ZERO ( ) key to exit test mode.
Table 7-7 summarizes the test menu functions.
Figure 7-7. Front Panel Key Functions in Test Mode
TEST Menu
FunctionDescription
A/DTSTDisplay A/D test
Press and hold Enter key to display raw count from A/D converter.
DEFLTDefault parameters
Press setup switch and Enter key
default values. Load cells must be recalibrated before using the indicator (see Section 4.0 on page 30).
at the same time to reset configuration and calibration parameters to factory
Table 7-7. Test Mode Menu Functions
420 Plus Installation Manual - Appendix51
Page 56
7.10Regulatory Mode Functions
REGULAT
Parameter ValueWeight on ScaleTa r e in S yst e m
NTEPzero or negativenono actionZERO
yesCLEAR TARE
positivenoTARE
yesTARE
CANADAzero or negativenono actionTARE
yesCLEAR TARE
positivenoTARE
yesno action
OIMLzero or negativenono actionZERO
yesCLEAR TAREZERO and CLEAR TARE
positivenoTAREZERO
yesTAREZERO and CLEAR TARE
NONEzero or negativenoTAREZERO
yesCLEAR TARE
positivenoTARE
yesCLEAR TARE
Front Panel Key Function
TAREZERO
Table 7-8. TARE and ZERO Key Functions for REGULAT Parameter Settings
7.11LED Functions
The following is a list of diagnostic LEDs for the 420
. see Figure 2-4 8 for locations on the 420 Plus
Plus
CPU board.
Serial Port Status
LED1 (EDP Port)Flashing red = TXD activity
Flashing green = RXD activity
On steady or off = No activity
LED2 (PRN Port)Flashing red = TXD activity
Flashing green = RXD activity
On steady or off = No activity
Power Supply
LED (+5AV)Excitation/ADC supply on
LED (+5V & +3.3V)Digital logic supply on
Operation Status
LEDHB (Microprocessor)Flashing green = Active
Red = Fault
LED AD (ADC)Green = Active
Red = Fault
LED MEM (Memory Test)Green = Pass
Red = Fail
LED ANA Red = Not Implemented
(Analog Option Card)
Digital Input/Output Status
DO 1 (Digital Out 1)On when lit (active low)
DO2 (Digital Out 2)On when lit
DI 1 (Digital In 1)On when lit (active low)
DI 2 (Digital In 2)On when lit (active low)
(active low)
52420 Plus Installation Manual
Page 57
7.12Specifications
t
N
A
T
I
O
N
A
L
C
O
N
F
E
R
E
N
C
E
t
O
N
W
E
I
G
H
T
S
A
N
D
M
E
A
S
U
R
E
S
Model Numbers
United States420-2A (AC) (NEMA 5-15)
420-2C (DC)
International420-2B (AC) (CEE-7)
420-2D (DC)
Power - AC
Line Voltages115 or 230 VAC
Frequency50 or 60 Hz
Power Consumption1.5 A @ 115 VAC (8W)
0.75 A @ 230 VAC (8W)
Fusing2.5 A 5 x 20 mm fuse
Power - DC
Line Voltages9-36 VDC DC Input
Power Consumption1.5 A Max
FusingInternal short circuit protection
Analog Specifications
Full Scale Input Signal Up to 22.5 mV
Excitation Voltage+5V single sided
8 x 350 or 16 x 700 l
Sense AmplifierDifferential amplifier with
4- and 6-wire sensing
Analog Signal
Input Range–0.5 mV/V to +4.5 mV/V
Analog Signal
Sensitivity0.3 V/gra
1.5 V/gra
Input Impedance200 M,
duation minimum,
d recommended
typical
Noise (ref to input)0.3 V p-p with digital
Internal Resolution8,000,000 counts
Display Resolution100 000 dd
Measurement RateUp to 60 measurements/sec
Input Sensitivity10 nV per internal count
System LinearityWithin 0.01% of full scale
Zero Stability150 nV/°C, maximum
Span Stability3.5 ppm/°C, maximum
Calibration MethodSoftware, constants stored in EEPROM
Common Mode
Voltage–2.35 to +3.45 V, referred to ground
Rejection130 dB minimum @ 50 or 60 Hz
Normal Mode
Rejection90 dB minimum @ 50 or 60 Hz
Input Overload± 12 V conti
nuous, static discharge
protected
RFI ProtectionSignal, excitation
protected by capacitor bypass
Analog OutputOptional: fully isolated, voltage or
current output,14-bit resolution.
Voltage output: 0 –10 VDC
Load resistance:1K minimum
Current output: 4–20 mA
External loop resistance: 500
ma
ximum
oad cells
filters at 4-4-4
, and sense lines
Digital Specifications
MicrocomputerPhillips PXAG30K main processor @
Digital Inputs2 inputs, TTL or switch closure,
Digital Outputs2 outputs, 250 mA Max Sink
Digital Filters3 filters, softwar
19.6608 MHz
active-low
e selectable
Serial Communications
EDP PortFull duplex RS-232
Printer Portfully capable RS-232 or active 20 mA
curr
ent loop (transmit only)
Both Ports38400, 19200, 9600, 4800, 2400,
1200, 600, 300 bps; 7 or 8 data bits;
even, odd, or no parity
Operator Interface
Display6-digit LED display. 7-segment, 0.8 in
(20 mm) digits
LED annunciatorsGross, net, center of zer
o, standstill, lb/
primary units, kg/secondary units,
count, tare
Keypad21-key flat membrane panel
Environmental
Operating Temperature –10 to +40°C (legal);
–10 to +50°C (industrial)
Storage Temperature –25 to +70°C
Humidity0–95% relative humidity
Enclosure
Enclosure Dimensions 9.5 in x 6 in x 3.12 in
24 cm x 15 cm x 8 cm
Weight2.8 Kg (6.1 lb)
Rating/MaterialUL Type 4X
Certifications and Approvals
NTEP
CoC Number04-076
Accuracy ClassIII/III L
n
max : 10 000
Measurement Canada (Pending)
Accuracy ClassIII/III HD
n
max
: 10 000
420 Plus Installation Manual - Appendix53
Page 58
420 Plus Limited Warranty
Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a
Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the
Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials
and workmanship for two years.
RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized by
RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment fails
to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the warranty
period subject to the following conditions:
•Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with a
detailed explanation of the alleged deficiencies.
•Individual electronic components returned to RLWS for warranty purposes must be packaged to
prevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in a
publication, Protecting Your Components From Static Damage in Shipment, available from RLWS
Equipment Return Department.
•Examination of such equipment by RLWS confirms that the nonconformity actually exists, and was
not caused by accident, misuse, neglect, alteration, improper installation, improper repair or improper
testing; RLWS shall be the sole judge of all alleged non-conformities.
•Such equipment has not been modified, altered, or changed by any person other than RLWS or its duly
authorized repair agents.
•RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsible
for shipping charges both ways.
•In no event will RLWS be responsible for travel time or on-location repairs, including assembly or
disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.