The ResMed Tango CPAP system is indicated for the treatment of obstructive sleep
apnoea (OSA) in patients weighing more than 30kg (66 lb). It is intended for home and
hospital use.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with the
following pre-existing conditions:
• severe bullous lung disease
• pneumothorax
• pathologically low blood pressure
• dehydration
• cerebrospinal fluid leak, recent cranial surgery or trauma.
English
Adverse Effects
Patients should report unusual chest pain, severe headache or increased
breathlessness to their prescribing physician. An acute upper respiratory tract infection
may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the Tango:
• Mirage Swift II Nasal Pillows System• Ultra Mirage™ Nasal Mask
• Papillon™ Nasal Mask • Ultra Mirage II Nasal Mask
• Silent Papillon™ Nasal Mask• Mirage Quattro™ Full Face Mask
• Meridian™ Nasal Mask• Mirage Liberty™ Full Face Mask
For information on using masks, see your mask manual.
Humidifiers
For information on using the ResMed C-Series Heated Humidifier or the ResMed
Passover, see your humidifier manual.
2
The Tango Interface
!
!
Heating
indicator
Power
indicator
Standby modeWhen the Tango is powered, the power indicator is lit and the run hours
are displayed on the LCD. All changes to altitude and ramp settings are
made in standby mode.
RampSetting a ramp time enables therapy to start at a low pressure and build
up to full pressure during the ramp time (see “Set ramp time”).
AltitudeChanges in altitude affect the pressure delivered by the Tango. The
altitude feature compensates for pressure changes related to changes in
altitude (see “Set altitude”).
Run hoursRun hours are displayed when the Tango is in standby mode.
HumidifierThe humidifier buttons are used when using the ResMed C-Series Heated
Humidifier and when setting the altitude. The heating indicator lights
when the humidifier is on.
English
Setting up the Tango
CAUTION
• Be careful to place the device where it cannot be bumped, and where no
one will trip over the power cord.
• If you put the device on the floor, make sure the area is free from dust
and clear of bedding, clothes or other objects that could block the air
inlet.
• Make sure the area around the device is dry and clean.
WARNING
• Make sure the power cord and plug are in good condition and the
equipment is not damaged.
3
• Only ResMed air tubing should be used with the device. A different type
of air tubing may alter the pressure you actually receive and reduce the
effectiveness of your treatment.
• Blocking the hose and/or air inlet of the device while in operation could
lead to overheating of the device.
2
1
3
4
1. Connect the power cord to the rear of the Tango.
2. Plug the other end into a power outlet.
3. Connect the air tubing to the air outlet of the Tango.
4. Connect the assembled mask system to free end of air tubing.
Using the Tango
Start therapyPress when the Tango is in standby mode. A circulating dash symbol
displays on the LCD.
If ramp is set, the ramp time displays on the LCD. When ramping is
finished, therapy continues at the prescribed pressure.
If no ramp is set, therapy starts immediately at the prescribed pressure.
Stop therapyPress .
Set ramp time1. Repeatedly press until the LCD displays the desired time.
Set altitude1. H o l d down for 3 seconds until the current altitude setting is
2. Press to accept changes and start therapy. (The Tango returns to
standby mode after 5 seconds if no button is pressed.)
The ramp settings are (in minutes): Off, 10, 20, 45.
displayed.
2. Press or to change the altitude setting.
3. Press to accept changes and return to standby mode. (The Tango
returns to standby mode after 10 seconds if no button is pressed.)
The altitude settings are:
• (0–609 m/0–2000’),
• (610–1219 m/2001–4000’),
• (1220–1828 m/4001–6000’),
• (1829–2591 m/6001–8500’).