ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
Distributed by:
ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA
ResMed (UK) Ltd 96 Milton Park Abingdon Oxfordshire OX14 4RY UK
See www.resmed.com for other ResMed locations worldwide.
For patent information, see www.resmed.com/ip
C-Series, Tango, Mirage, Mirage Activa, Mirage Vista, Ultra Mirage, Mirage Swift, Mirage Quattro, Mirage
Liberty and Meridian are trademarks of ResMed Ltd. Mirage, Mirage Activa, Mirage Vista, Ultra Mirage,
Mirage Swift and Meridian are registered in US Patent and Trademark Office. Papillon and Silent Papillon
are trademarks of MAP Medizin-Technologie GmbH.
The ResMed Tango CPAP system is indicated for the treatment of obstructive sleep
apnoea (OSA) in patients weighing more than 30kg (66 lb). It is intended for home and
hospital use.
Contraindications
Positive airway pressure therapy may be contraindicated in some patients with the
following pre-existing conditions:
• severe bullous lung disease
• pneumothorax
• pathologically low blood pressure
• dehydration
• cerebrospinal fluid leak, recent cranial surgery or trauma.
English
Adverse Effects
Patients should report unusual chest pain, severe headache or increased
breathlessness to their prescribing physician. An acute upper respiratory tract infection
may require temporary discontinuation of treatment.
The following side effects may arise during the course of therapy with the Tango:
• Mirage Swift II Nasal Pillows System• Ultra Mirage™ Nasal Mask
• Papillon™ Nasal Mask • Ultra Mirage II Nasal Mask
• Silent Papillon™ Nasal Mask• Mirage Quattro™ Full Face Mask
• Meridian™ Nasal Mask• Mirage Liberty™ Full Face Mask
For information on using masks, see your mask manual.
Humidifiers
For information on using the ResMed C-Series Heated Humidifier or the ResMed
Passover, see your humidifier manual.
2
Page 5
The Tango Interface
!
!
Heating
indicator
Power
indicator
Standby modeWhen the Tango is powered, the power indicator is lit and the run hours
are displayed on the LCD. All changes to altitude and ramp settings are
made in standby mode.
RampSetting a ramp time enables therapy to start at a low pressure and build
up to full pressure during the ramp time (see “Set ramp time”).
AltitudeChanges in altitude affect the pressure delivered by the Tango. The
altitude feature compensates for pressure changes related to changes in
altitude (see “Set altitude”).
Run hoursRun hours are displayed when the Tango is in standby mode.
HumidifierThe humidifier buttons are used when using the ResMed C-Series Heated
Humidifier and when setting the altitude. The heating indicator lights
when the humidifier is on.
English
Setting up the Tango
CAUTION
• Be careful to place the device where it cannot be bumped, and where no
one will trip over the power cord.
• If you put the device on the floor, make sure the area is free from dust
and clear of bedding, clothes or other objects that could block the air
inlet.
• Make sure the area around the device is dry and clean.
WARNING
• Make sure the power cord and plug are in good condition and the
equipment is not damaged.
3
Page 6
• Only ResMed air tubing should be used with the device. A different type
of air tubing may alter the pressure you actually receive and reduce the
effectiveness of your treatment.
• Blocking the hose and/or air inlet of the device while in operation could
lead to overheating of the device.
2
1
3
4
1. Connect the power cord to the rear of the Tango.
2. Plug the other end into a power outlet.
3. Connect the air tubing to the air outlet of the Tango.
4. Connect the assembled mask system to free end of air tubing.
Using the Tango
Start therapyPress when the Tango is in standby mode. A circulating dash symbol
displays on the LCD.
If ramp is set, the ramp time displays on the LCD. When ramping is
finished, therapy continues at the prescribed pressure.
If no ramp is set, therapy starts immediately at the prescribed pressure.
Stop therapyPress .
Set ramp time1. Repeatedly press until the LCD displays the desired time.
Set altitude1. H o l d down for 3 seconds until the current altitude setting is
2. Press to accept changes and start therapy. (The Tango returns to
standby mode after 5 seconds if no button is pressed.)
The ramp settings are (in minutes): Off, 10, 20, 45.
displayed.
2. Press or to change the altitude setting.
3. Press to accept changes and return to standby mode. (The Tango
returns to standby mode after 10 seconds if no button is pressed.)
The altitude settings are:
• (0–609 m/0–2000’),
• (610–1219 m/2001–4000’),
• (1220–1828 m/4001–6000’),
• (1829–2591 m/6001–8500’).
4
Page 7
Cleaning and Maintenance
!
!
WARNING
Beware of electric shock. Do not immerse the flow generator or power cord
in water. Always unplug the flow generator before cleaning and make sure
it is dry before reconnecting.
CAUTION
• Do not hang the air tubing in direct sunlight as it may harden over time
and eventually crack.
• Do not use bleach, chlorine, alcohol, or aromatic-based solutions,
moisturising or antibacterial soaps or scented oils to clean the air tubing
or the Tango. These solutions may cause hardening and reduce the life
of the product.
Daily
Disconnect the air tubing and hang it in a clean, dry place until next use.
Weekly
1. Remove the air tubing and wash it in warm water and mild detergent. Rinse
thoroughly, hang and allow to dry.
2. Before next use, reassemble the mask and headgear.
3. Reconnect the air tubing.
Monthly
• Clean the exterior of the Tango with a damp cloth.
• Check the air filter for holes and blockage by dirt. Replace it every six months, or
more often in a dusty environment.
For information on cleaning your mask or humidifier, refer to the manual supplied with
your mask or humidifier.
English
5
Page 8
Changing the Air Filter
!
Air filter
CAUTION
Do not wash the air filter.
1. With the power cord and air tubing disconnected, turn the Tango upside down so
the air inlet is facing you. The air filter is inside the air inlet.
2. Pull the old air filter out of the air inlet.
3. Put the new air filter into the air inlet in the direction shown by the arrow, with the
blue side facing towards you.
Troubleshooting
ProblemPossible CauseSolution
6
Tango won’t startPower cord not connected
properly.
Faulty power outlet.
No/low air flowAir tubing not connected properly.
Air tubing is blocked, pinched or
punctured.
Treatment pressure seems
either low or too high
LCD displays an error
message and treatment
stops or can’t be started
Unblock or free air
tubing. Check air tubing
for punctures.
Page 9
Technical Specifications
Operating pressure range4 to 20 cm H2O
Dimensions (L x W x H)215 mm x 189 mm x 114 mm (8.5” x 7.4” x 4.5”)
Weight1.1 kg (2.4 lb)
Power SupplyInput range 100–240V, 50/60Hz, 20VA/10W (typical power
consumption), 36VA/15W (maximum power consumption)
Operating Temperature+5°C to +40°C (+41°F to +104°F)
Operating Humidity10%–95% non-condensing
Storage and Transport
Temperature
Storage and Transport
Humidity
-20°C to +60°C (-4°F to +140°F)
10%–95% non-condensing
English
Atmospheric Pressure
Range
Electromagnetic
Compatibility
Air FilterTwo-layered, powder-bonded, polyester non-woven fiber
Air TubingFlexible plastic, 2 m (6’6”)
Air OutletThe 22 mm (0.87”) conical air outlet complies with
IEC 60601-1 ClassificationsClass II (double insulation), Type CF
Noise levelRadiated sound pressure is measured at <32dBA at
1060 hPa (sea level) to 680 hPa (2591 m/8500’)
Product complies with all applicable electromagnetic
compatibility requirements (EMC) according to IEC60601-1-2,
for residential, commercial and light industry environments
EN 1281-1
10 c m H
O and 1 m according to ISO17510-1.
2
The manufacturer reserves the right to change these specifications without notice.
7
Page 10
Symbols which appear on the product
Attention, consult accompanying documents
Type CF equipment
Class II equipment
Drip proof
European Authorised Representative
Canadian Standards Association
Manufacturer
Environmental information
WEEE 2002/96/EC is a European Directive that
requires the proper disposal of electrical and
electronic equipment. This device should be
disposed of separately, not as unsorted municipal
waste. To dispose of your device, you should use
appropriate collection, reuse and recycling systems
available in your region. The use of these collection,
reuse and recycling systems is designed to reduce
pressure on natural resources and prevent hazardous
substances from damaging the environment.
If you need information on these disposal systems,
please contact your local waste administration. The
crossed-bin symbol invites you to use these disposal
systems. If you require information on collection and
disposal of your ResMed device please contact your
ResMed office, local distributor or go to
www.resmed.com/environment.
8
Page 11
!
General Warnings and Cautions
A warning alerts you to possible injury.
• Read the entire guide before using the Tango.
• The advice in this guide should not supersede instructions given by the prescribing
physician.
• The Tango should only be used with masks (and connectors*) recommended by
ResMed, or by a physician or respiratory therapist. A mask should not be used
unless the Tango is turned on and operating properly. The vent hole or holes
associated with the mask should never be blocked.
Explanation: The Tango is intended for use with special masks (or connectors*)
with vent holes allowing a continuous flow of air out of the mask.
• When the device is turned on and functioning properly, new air from the device
flushes the exhaled air out through the mask vent holes. If the device is not
operating, insufficient fresh air is provided through the mask, and the exhaled air
may be rebreathed. Rebreathing exhaled air for longer than several minutes can
sometimes lead to suffocation. If there is a power failure or the device machine
malfunctions, remove the mask.
• The Tango can be set to deliver pressures up to 20 cm H
certain fault conditions, pressures up to 30 cm H
• If oxygen is used with this device, the oxygen flow must be turned off when the
2
device is not operating. If oxygen has been left on while the device is not operating,
disconnect the oxygen and wait 30 minutes before turning on the flow generator
again.
Explanation: When the CPAP device is not in operation and the oxygen flow is left
on, oxygen delivered into the air delivery tubing may accumulate within the CPAP
machine enclosure and create a risk of fire.
• Oxygen supports combustion. Do not use oxygen while you are smoking or in the
presence of an open flame.
• Always ensure air is flowing from the device before turning on the oxygen supply.
• Always turn the oxygen supply off before stopping the airflow from the device.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen
concentration will vary, depending on where the oxygen is introduced, pressure
settings, patient breathing pattern, mask selection and leak rate.
• Explosion hazard—do not use in the vicinity of flammable anaesthetics.
• Do not use the Tango if there are obvious external defects or unexplained changes
in performance.
• Do not open the Tango case. There are no user serviceable parts inside. Repairs
and internal servicing should only be performed by an authorised service agent.
O. In the unlikely event of
2
O are possible.
English
*Ports may be incorporated into the mask or in connectors that are near the mask.
9
Page 12
A caution explains special measures for safe and effective use of the device.
• At low pressures, the flow through the exhalation ports of your mask may not clear
all exhaled gas from the tubing. Some rebreathing may occur.
• The airflow for breathing produced by this device can be as much as 6°C (11°F)
higher than the temperature of the room. Be careful if the room temperature is
warmer than 32°C (90°F).
Servicing
The Tango should be inspected by an authorised ResMed service centre 5 years from
the date of manufacture. Before this, the device is intended to provide safe and reliable
operation if it is operated and maintained according to the instructions provided by
ResMed. Warranty details are provided with the device at the time of original supply.
As with all electrical devices, if any irregularity becomes apparent, you should have the
device inspected by an authorised ResMed service centre.
10
Page 13
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions and Immunity
Guidance and manufacturer’s declaration—electromagnetic emissions
The C-Series Tango system is intended for use in the electromagnetic environment specified
below. The customer or the user of the C-Series Tango system should assure that the device is
used in such an environment.
Emissions testCompliance
RF emissions CISPR11 Group 1The C-Series Tango system uses RF
RF emissions CISPR 11 Class BThe C-Series Tango system is suitable
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/Flicker
Emissions IEC 61000-3-3
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed
and put into service according to EMC information provided in this document.
Warnings: The C-Series Tango system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the C-Series Tango system should be observed
to verify normal operation in the configuration in which it will be used.
The use of accessories (eg humidifiers) other than those specified in this manual is not
recommended. They may result in increased emissions or decreased immunity of the C-Series
Tango system.
Class A
Complies
Electromagnetic
environment—guidance
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equ pment.
for use in all establishments, including
domestic establishments and those
directly connected to the public lowvoltage network that supplies
buildings used for domestic purposes.
English
11
Page 14
Guidance and manufacturer’s declaration—electromagnetic immunity
The C-Series Tango system is intended for use in the electromagnetic environment spec fied
below. The customer or the user of the C-Series Tango system should assure that the device is
used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: Ut is the AC mains voltage prior to application of the test level.
IEC60601-1-2
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV d fferential
mode
±2 kV common
mode
<5% Ut
(>95% d p in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95%dip in Ut)
for 5 sec
3 A/m 3 A/m Power frequency magnetic
Compliance
level
±6 kV contact
±8 kV air
±2 kV
Not Applicable
±1 kV d fferential
mode
±2 kV common
mode
< 12V
(>95% d p in 240V)
for 0.5 cycle
96V
(60% dip in 240V)
for 5 cycles
168V
(30% dip in 240V)
for 25 cycles
<12V
(>95%d p in 240V)
for 5 sec
Electromagnetic
environment—
guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered w th
synthetic material, the
relative humid ty should be
at least 30%.
Mains power qual ty should
be that of a typical
commercial or hosp tal
environment.
Mains power quality should
be that of a typical
commercial or hosp tal
environment.
Mains power quality should
be that of a typical
commercial or hosp tal
environment.
If the user of the C-Series
Tango system requires
continued operation during
power mains interruptions,
it is recommended that the
C-Series Tango system be
powered from an
uninterruptible power
source.
fields should be at levels
characteristic of a typical
location in a typical
commercial or hosp tal
environment.
12
Page 15
Guidance and manufacturer’s declaration—electromagnetic immunity
The C-Series Tango system is intended for use in the electromagnetic environment specified below. The
customer or the user of the C-Series Tango system should assure that the device is used in such an
environment.
Immunity
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC60601-1-2
test level
3 Vrms
150 kHz to 80 MHz
10 V /m
80 MHz to 2.5 GHz
Complianc
e level
3 Vrms d = 1.17 √P
10 V/m d = 0.35 √P 80 MHz to 800 MHz
Electromagnetic environment—
guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the C-Series Tango system, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 0.70 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transm tter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transm tters, as
determined by an electromagnetic site
a
should be less than the compliance
survey,
level in each frequency range.
Interference may occur in the vicin ty of
equipment marked with the following symbol:
b
English
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all s tuations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transm tters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the C-Series Tango system is used exceeds the applicable RF compliance level above, the C-Series Tango
system should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the C-Series Tango system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
13
Page 16
Recommended separation distances between portable and mobile RF communications
equipment and the C-Series Tango system.
The C-Series Tango system is intended for use in an environment in which radiated RF disturbances are
controlled. The customer or the user of the C-Series Tango system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transm tters) and the C-Series Tango system as recommended below, according to the
maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output
power of transmitter
W
0.010.170.040.07
0.10.370.110.22
11.170.350.7
103.691.112.21
10011.703.507.0
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transm tter in watts (W) according to the transm tter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all s tuations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
150 kHz to 80 MHz
d = 1.17 √P
80 MHz to 800 MHz
d = 0.35 √P
800 MHz to 2.5 GHz
d = 0.7 √P
14
Page 17
Limited Warranty
ResMed warrants that your ResMed product shall be free from defects in material and
workmanship for the period specified below from the date of purchase by the initial
consumer. This warranty is not transferable.
ProductWarranty Period
ResMed humidifiers, ResControl™, ResLink™, ResTraxx™1 Year
ResMed flow generators2 Years
English
Accessories, water chambers and mask systems (including
mask frame, cushion, headgear and tubing).
Excludes single-use devices.
90 Days
Note: Some models are not available in all regions.
If the product fails under conditions of normal use, ResMed will repair or replace, at its
option, the defective product or any of its components. This Limited Warranty does not
cover:
a) any damage caused as a result of improper use, abuse, modification or alteration of
the product;
b) repairs carried out by any service organisation that has not been expressly authorised
by ResMed to perform such repairs;
c) any damage or contamination due to cigarette, pipe, cigar or other smoke;
d) any damage caused by water being spilled on or into a flow generator.
Warranty is void on product sold, or resold, outside the region of original purchase.
Warranty claims on defective product must be made by the initial consumer at the point
of purchase.
This warranty is in lieu of all other express or implied warranties, including any implied
warranty of merchantability or fitness for a particular purpose. Some regions or states
do not allow limitations on how long an implied warranty lasts, so the above limitation
may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed
to have occurred as a result of the sale, installation or use of any ResMed product.
Some regions or states do not allow the exclusion or limitation of incidental or
consequential damages, so the above limitation may not apply to you. This warranty
gives you specific legal rights, and you may also have other rights which vary from
region to region.
For further information on your warranty rights, contact your local ResMed dealer or
ResMed office.
15
Page 18
34840/2
2012-04
C-Series Tango
User
ROW ENG
34840
Global leaders in sleep and respiratory medicine www.resmed.com