Clear polycarbonate (PC) versionSmoky gray polysulphone (UDEL) version
ResMed HumidAire 2i – Disinfection Guide
This Disinfection Guide is intended for multipatient use of the HumidAire 2i in a sleep lab, clinic or hospital. If
you use the water chamber as a single user in the home, refer to the User Guide for cleaning instructions.
The following are the recommended and validated procedures for cleaning and disinfection of the water
chamber. We advise that each healthcare facility consult its own procedures before carrying out the following
instructions.
WARNINGS and CAUTIONS
• ResMed cannot give any assurance that deviations from the procedures listed in this guide, and their
effect on the performance of the product, will be acceptable.
• When using detergents or disinfectants always follow the manufacturer’s instructions.
Note: The HumidAire 2i is available in both a clear polycarbonate (PC) version as well as a smoky gray
polysulphone (UDEL) version. The following cleaning methods apply only to the clear polycarbonate (PC)
version.
2
Page 3
Clinical Use Only
ResMed Validated Procedures
Table 1: Option 1 (Automatic Cleaning and High-Level Thermal Disinfection)
Option 1 (Automatic Cleaning and High-Level Thermal Disinfection)
DisassemblyDisassemble the humidifier water chamber according to the “Disassembling and Reassembling”
instructions on page 5.
Pre-cleaningRinse all water chamber components 10 times and brush thoroughly with a soft bristle brush in a
warm (113°F or 45°C) solution of Neodisher MediZym (2%).
Note:Pre-cleaning must occur immediately after use.
Cleaning1.Support the components in the standard racking of the automatic washer-disinfector so that
each will drain water.
1
2. Add the detergent solution
of cleaning, followed by thorough rinsing.
DryingNot applicable.
Pre-disinfectionNot applicable.
DryingNot applicable.
Disinfection
(high-level
thermal
disinfection)
1.Activate the automatic washer-disinfector for a standard cycle of thermal disinfection
2. Ensure that the machine has thoroughly rinsed each component.
and activate the automatic washer-disinfector for a standard cycle
2
.
DryingAir dry out of direct sunlight.
InspectionPerform a visual inspection of the water chamber. If any visible deterioration is apparent (cracking,
crazing, tears etc), the water chamber should be discarded and replaced. Slight discoloration of the
silicone components may occur and is acceptable.
ReassemblyReassemble the humidifier water chamber according to the “Disassembling and Reassembling”
instructions on page 5.
Packaging &
Storage
1 Cleaning: ResMed has tested the detergent Neodisher MediZym™ (diluted at 0.5%) using warm water (approximately 113°F or 45°C)
according to the manufacturer’s instructions.
2 Disinfection: Due to specific regional requirements, ResMed water chambers have been tested for disinfection for 20 cycles using hot
water (approximately 199°F or 93°C) for 10 minutes.
Store in a dry, dust-free environment away from direct sunlight.
Storage temperature: -4°F to 140°F (-20°C to 60°C).
3
Page 4
Clinical Use Only
Table 2: Option 2 (Manual Cleaning and High-Level Chemical Disinfection)
Option 2 (Manual Cleaning and High-Level Chemical Disinfection)
DisassemblyDisassemble the humidifier water chamber according to the “Disassembling and Reassembling”
instructions on page 5.
Pre-cleaningRinse all water chamber components 10 times and brush thoroughly with a soft bristle brush in a
warm (113°F or 45°C) solution of Alconox (1%).
Note:Pre-cleaning must occur immediately after use.
Cleaning1.Clean all components of the water chamber with a soft bristle brush for one minute while
soaking them in the detergent
1
solution. Pay particular attention to all crevices and cavities.
2. Thoroughly rinse each component according to the manufacturer’s instructions.
DryingAllow the water chamber to air dry out of direct sunlight.
Pre-disinfectionNot applicable.
DryingNot applicable.
Disinfection
(high-level
chemical
disinfection)
1.Soak the disassembled water chamber components in a commercially available solution of
ortho-phthalaldehyde 0.55% (eg, CIDEX™ OPA)
2
. Take care that no air bubbles are trapped
against the components.
2. Thoroughly rinse
3
each component according to the manufacturer’s instructions.
DryingAir dry out of direct sunlight.
InspectionPerform a visual inspection of the water chamber. If any visible deterioration is apparent (cracking,
crazing, tears etc), the water chamber should be discarded and replaced. Slight discoloration of the
silicone components may occur and is acceptable.
ReassemblyReassemble the humidifier water chamber according to the “Disassembling and Reassembling”
instructions on page 5.
Packaging &
Storage
1 Cleaning: ResMed has tested the detergent Alconox (diluted at 1%) using hot water (approximately 140°F or 60°C) or warm water
(approximately 120°F or 50°C) according to the manufacturer’s instructions.
2 Disinfection: ResMed water chambers have been validated for 20 cycles using solutions of ortho-phthalaldehyde (0.55% CIDEX OPA
for 12 minutes) according to the manufacturer’s instructions.
3 Rinsing: Thoroughly rinse each component in drinking quality water (five litres per assembly) by immersing it completely for a
minimum of one minute in duration. Repeat the rinse procedure two additional times using fresh water for a total of three rinses.
Store in a dry, dust-free environment away from direct sunlight.
Storage temperature: -4°F to 140°F (-20°C to 60°C).
4
Page 5
Clinical Use Only
Side catch
Front tab
Seal
Lid
Base
Cross-member
Disassembling and Reassembling the HumidAire 2i
WARNING
Beware of electrocution. Do not immerse the device or power cord in water. Always unplug the device before
cleaning and be sure that it is dry before plugging back in.
1Turn the main power switch at the back of the device to OFF, and unplug the power cord from the power outlet.
2Allow 10 minutes for the tub plate and any excess water to cool.
3Remove the water chamber. To do this, lift the latch at the bottom of the water chamber and pull the water chamber
away from the docking station.
4Discard any excess water by turning the water chamber upside down, allowing the water to drain out of the air outlet.
5Disassembling the Water Chamber
1. To open the side catches pull out from the lower edge.
2. Separate the water chamber by pulling the two halves apart. Use your thumb to pull down on the front tab.
3. Inside the chamber lid is the seal. Remove the seal by peeling it away from the inside of the lid.
4. Remove the cross-member from the water chamber base.
5
Page 6
Clinical Use Only
Tub plate
O-ring
Underside view
61. Push down on the tub plate from inside the chamber to remove it from the chamber.
2. Remove the O-ring from the tub plate.
3. Turn the water chamber base upside down. Firmly pull the latch up and away from the chamber until separated
from the base.
4. Disengage the catches from the sides of the water chamber base. Then, separate the inner and outer moulding of
the catch, pull back the inner moulding and squeeze. The inner moulding will pop out.
7Reassembling the Water Chamber
When all the water chamber parts are clean and dry, reassemble the water chamber:
1. Reassemble the catches by inserting one side of the inner moulding to the outer moulding and pushing
downwards.
6
Page 7
Clinical Use Only
Tu b p la t e
2. Take the assembled catches and fit them to the water chamber base.
3. Turn the water chamber base upside down. Insert the latch’s feet into the locking tabs. Push firmly and click into
place.
4. Refit the O-ring in the channel of the tub plate.
CAUTION
Ensure that you replace the O-ring if it is nicked or punctured.
5. Push the tub plate into the water chamber with the open side facing down. Then firmly press the tub plate down
into the water chamber.
6. Place the cross-member on its supports inside the water chamber base.
7
Page 8
Clinical Use Only
Seal
Rim
Lid
ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia, ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA, ResMed
(UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK, ResMed Offices Australia, Austria, Brazil, China, Finland, France, Germany, Hong Kong, India, Japan,
Malaysia, Netherlands, New Zealand, Norway, Singapore, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
Patents pending. Protected by design registrations: AU 147335, CH 128.711, CH 128.712, DE 40202008, DE 40201723, ES 153516, ES 153518, FR 021407, GB 3001820, GB 3001821, JP 1164266, JP 1164267,
SE 75598, SE 75715, US D476077, US D477868. Other designs pending.
HumidAire is a trademark of ResMed Ltd and is registered in the U.S. Patent and Trademark Office.
CIDEX is a trademark of Johnson & Johnson. Alconox is a trademark of Alconox Inc. Neodisher MediZym is a trademark of Chemische Fabrik Dr Weigert.