ResMed AirSense 11 Disinfection Guide

Disinfection Guide

Preparing the device for use between patients

When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a health care provider, the outlet connector, air tubing , cleanable humidifier tub and device enclosure should be reprocessed between each patient.

Described here are ResMed's recommended and validated procedures for cleaning and disinfecting the outlet connector, air tubing, cleanable humidifier tub and device enclosure. The person executing reprocessing activities is responsible for ensuring that reprocessing is completed in line with ResMed's validated procedures. Components not identified in the reprocessing instructions do not require reprocessing or are intended for single patient use.

Note: The standard HumidAir 11 tub cannot be reprocessed.

WARNING

Always follow cleaning and disinfection instructions. Some cleaning products may damage the components and their function or leave harmful residual vapors.

Any deviations from the procedures or claimed maximum number of cycles in this guide can have an adverse effect on the components and consequently the safety or the quality of therapy.

When using detergents, disinfectants or equipment, always follow the instructions provided by the manufacturer of those products. In the event of conflict, this guide takes precedence.

Always follow safe operating practices, including the use of appropriate Personal Protective Equipment (PPE), as required. Refer to the instructions provided by the manufacturer of those products for more details.

Beware of electrocution:

Do not immerse the device, AC Adaptor or power cord in water.

Do not connect to power while the device is wet. Make sure that all parts are dry before plugging it in.

1

ResMed AirSense 11 Disinfection Guide

General summary

ResMed has validated the following number of cycles for cleaning and disinfection using a manual cleaning method.

Components

 

Cleaning - Mild alkaline, anionic detergent (eg, Alconox™)

 

 

 

System component

Chemical high level disinfection

Thermal high level disinfection

 

eg, CIDEX®-OPA

167ºF (75ºC) for 30 minutes

Outlet connector

120

80

 

 

 

HumidAir 11 Cleanable tub

120

80

 

 

 

ClimateLineAir 11

30

30

 

 

 

Standard tubing

30

30

 

 

 

SlimLine

30

30

 

 

 

Device

 

 

 

 

 

 

Cleaning and low level disinfection

 

 

 

 

Device Enclosure

CaviWipes1™

 

 

 

 

Disassembling

Before disassembly, turn off the device and ensure the AC adaptor has been removed.

Cleanable Humidifier tub

1.Hold the humidifier tub at the top and bottom, press it gently and pull it away from the device. Note: take care when handling the humidifier tub as the humidifier tub may be hot. Allow 10 minutes for the heater plate and any excess water to cool.

2.Hold the base of the humidifier tub and fully open the humidifier tub lid and pull it away so that it easily detaches from the base.