ResMed AirMini User Manual

User guide

English | Français | Español | Português| | Bahasa Melayu

ResMed.com/AirMiniApp

ENGLISH

Welcome

The ResMed AirMini™ system combines ResMed’s AirMini selfadjusting pressure machine, masks and the AirMini by ResMed™ App.

WARNING

Read this entire guide before using the machine.

CAUTION

In the US, Federal law restricts this device to sale by or on the order of a physician.

Indications for use

The AirMini self-adjusting system is indicated for the treatment of obstructive sleep apnea (OSA) in patients (female patients with mild to moderate OSA when using AutoSet™ for Her treatment mode) weighing more than 66 lb (30 kg).

It is intended for home and hospital use.

Contraindications

Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions:

severe bullous lung disease

pneumothorax

pathologically low blood pressure

dehydration

cerebrospinal fluid leak, recent cranial surgery, or trauma.

Adverse effects

You should report unusual chest pain, severe headache, or increased breathlessness to your prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment.

English 1

The following side effects may arise during the course of therapy with the machine:

drying of the nose, mouth, or throat

nosebleed

bloating

ear or sinus discomfort

eye irritation

skin rashes.

At a glance

The AirMini system includes the following:

AirMini: the AirMini machine, power supply unit, drawstring bag and one of the following:

AirMini setup pack - F20: Air tubing, F20 connector OR

AirMini setup pack - N20: Air tubing, N20 connector, HumidX™, HumidX Plus OR

AirMini mask pack for P10: Air tubing, AirFit P10 for AirMini mask (M), P10 cushion (S), P10 cushion (L), HumidX, HumidX Plus, AirFit™ P10 headgear clips.

Note: The F20 set up pack can also be used to connect the AirFit F30 mask with AirMini.

2

Your AirMini system

Your AirMini system is comprised of the AirMini machine and one mask pack or one setup pack.

A

AirMini machine

6

20W power supply unit

B

F20 connector for AirMini and air

7

F20 connector for AirMini

 

tubing

 

 

C

N20 connector for AirMini and air

8

Vent

 

tubing

 

 

D

AirFit P10 for AirMini and tubing

9

Vent module

1

Start/Stop button

10

Anti Asphyxia Valve (AAV)

2

Power inlet

11

N20 connector for AirMini

3

Air outlet

12

AirFit P10 for AirMini

4

Air filter cover

13

Heat moisture exchanger - HumidX

 

 

 

(blue) and HumidX Plus (gray)

5

Bluetooth® button

14

AirMini tubing

Your machine is compatible with ResMed masks. For a full list of compatible masks for this device, see the Mask device compatibility list on ResMed.com/downloads/masks.

English 3

Note: Not all masks are available in all regions.

Make sure all parts and accessories used with the machine are compatible. For compatibility information, refer to ResMed.com for more information.

Your AirMini machine is designed to work with ResMed’s AirMini app. However, the AirMini app is not essential to operate the machine. For more information on the AirMini app, refer to Using your machine with the AirMini app or the AirMini app Patient eHelp.

HumidX and HumidX Plus

The HumidX and HumidX Plus are waterless humidifiers. These are designed to provide more comfort by improving the moisture level in the air you breathe and offering relief from possible dryness.

The HumidX is a good starting point for finding the right level of humidification. The HumidX Plus will improve the humidity level further and is useful when the ambient humidity level is very low (eg, high altitudes, on an aircraft).

The HumidX and HumidX Plus can be used with the N20 connector for AirMini and AirFit P10 for AirMini only.

The HumidX and HumidX Plus must be replaced within 30 days of opening the HumidX/HumidX Plus pack.

4

Setting up and starting therapy
Using your mask

Use your AirMini machine only as directed by your physician or healthcare provider.

For information on fitting your mask, refer to the mask user guide.

Set up your machine

1.Plug the power supply unit into the machine and the power outlet. The connector has an angled edge so that the power cord can only be inserted into the machine one way.

2.Connect the gray end of the AirMini tubing firmly to the air outlet.

Add optional humidification (for N20 or AirFit P10 for AirMini masks only)

To insert the HumidX or HumidX Plus:

1.Open the connector by gently twisting it.

2.Hold the HumidX/HumidX Plus at the sides, with the colored side pointing down and insert it.

3.Gently push together the connector and twist until it locks in place.

English 5

ResMed AirMini User Manual
Connect your mask

1.Fit your mask as instructed in the mask user guide.

2.Attach the connector to your mask.

F20 Series

F30 Series

N20 Series

AirFit P10 for AirMini

Start therapy

Start therapy by pressing the Start/Stop button or breathe normally if SmartStart™ is enabled.

Stop therapy

1.Remove your mask.

2.Press the Start/Stop button or wait until the machine stops if SmartStop is activated.

6

Comfort features

Ramp Time, Pressure Relief and SmartStart are enabled on your machine.

Ramp Time

Designed to make the beginning of therapy more comfortable, Ramp Time is the period during which pressure gradually increases from a lower start pressure to the prescribed treatment pressure.

SmartStart™

When SmartStart is turned on, therapy starts automatically when you breathe into your mask.

SmartStop

When SmartStop is turned on, therapy stops automatically after a few seconds when you remove your mask.

Expiratory Pressure Relief

Designed to make therapy more comfortable, Expiratory Pressure Relief (EPR) maintains optimal treatment during inhalation and reduces the delivered mask pressure during exhalation.

Changing comfort features

Your AirMini machine has been set up for your needs by your provider, but you may find that you want to make small adjustments to comfort settings to make your therapy more comfortable.

For more information about comfort settings, or for information on changing these settings, refer to the AirMini app Patient eHelp.

English 7

Using your machine with the AirMini app

Your AirMini machine can be used with ResMed’s AirMini app. Once you have paired and connected your machine to your smart device via Bluetooth, you can start and stop therapy, change comfort settings and view your therapy data via the AirMini app. Data sent via Bluetooth is encrypted to protect your privacy.

Connecting your machine and smart device

Before connecting the AirMini machine to a smart device, ensure that the latest version of the AirMini app is installed on the smart device. If not, download the app from the App StoreSM or the Google Play store.

1.Ensure your machine is set up correctly and plugged into a power source.

2.On your smart device, enable Bluetooth.

3.Open the AirMini app.

The first time you open the AirMini app, you will be asked to:

provide consent for the use of analytics

accept ResMed’s terms of use and privacy notice

provide consent for the AirMini app to upload data to the cloud

Select the necessary check boxes to provide consent and accept policies.

Notes:

The check box for analytics consent is selected by default.

From time to time, the terms of use may be updated and you may be required to provide consent again.

If you agree to allow the AirMini app to upload your data to the cloud, it will be uploaded from the AirMini app to a secure server.

There are two distinct data upload features:

8

Background data upload – allows the app to upload all data that has not already been uploaded to the cloud without any user interaction (ie, app automatically uploads data whenever new data is available and active internet connection is present). You must provide consent to enable this feature as it is disabled by default. This feature can be enabled or disabled anytime on the More screen.

On-demand data upload – clinician or patient-initiated feature that uploads all data that has not already been sent to the cloud. This feature can be used by clinician or patient regardless if background data upload consent has been provided. To use this feature, the app must have active Bluetooth connection with the AirMini.

For additional information, links are also provided on the Welcome page:

Learn More – explanation of the way analytics are used

Terms – ResMed’s terms of use

Privacy – ResMed’s privacy notice

my data – information about data privacy.

4.Once selections are complete on the Welcome page, tap Continue.

5.On your machine, press the Bluetooth button when prompted by the AirMini app.

When the light starts to flash blue , Bluetooth is enabled on your AirMini machine and ready to connect.

6.On your smart device, tap Connect.

When your AirMini machine is discovered, the machine name appears in the machine selection list.

7.Select the machine name from the list to connect.

8.The first time you pair your AirMini machine with your smart device, you will need to perform the authentication procedure.

When prompted by the AirMini app, authenticate by entering the four-digit key located on the back of your machine.

English 9

Alternatively, you can tap Scan code. If permission has not been granted already, you will be prompted by the AirMini app to allow the use of your smart device's camera and then to position the machine’s QR code within the camera's viewing frame.

Note: The QR code is located on the back of your AirMini machine.

9. Tap Done.

When successfully connected, the Bluetooth connection icon is displayed in the top right corner of the AirMini app.

Bluetooth status indicators

Bluetooth indicator light is flashing blue.

Bluetooth indicator light is constantly blue.

Bluetooth indicator light is constantly white (airplane mode).

Your machine is in discoverable mode and ready for pairing.

Your machine is connected to your smartphone.

Bluetooth on your machine is disabled.

Bluetooth indicator light is off.

Your machine is not connected to your smartphone.

Note: Once therapy starts, the indicator light will start to fade.

10

Cleaning and maintenance

It is important that you regularly clean your AirMini machine to make sure you receive optimal therapy. The following sections will help you with disassembly, cleaning and checking your machine.

WARNING

As part of good hygiene, always follow cleaning instructions. Some cleaning products may damage the air tubing and affect its function, or leave harmful residual vapours that could be inhaled if not rinsed thoroughly.

The AirMini connectors and AirFit P10 for AirMini contain a vent system to prevent carbon dioxide build-up in the mask. The vent system must be kept clean and clear of contamination in order to function correctly. Blocking or modifying the vent system may result in excessive carbon dioxide rebreathing.

Regularly clean your AirMini tubing, connectors and mask components to receive optimal therapy and to prevent the growth of germs that can adversely affect your health.

Regularly inspect the HumidX/HumidX Plus and follow the cleaning and maintenance instructions to prevent the growth of germs that can adversely affect your health.

CAUTION

If any visible deterioration of a system component is apparent (cracking, discoloration, tears etc.), the component should be discarded and replaced.

English 11

Disassembling

Remove the connector (F30, F20 and N20 masks only)

Remove the connector from the mask by pressing the side buttons and pulling it away.

Disconnect the AirMini tubing (F30 and F20 masks only)

Disconnect the F20 connector from the AirMini tubing by gently twisting and pulling it away.

Disconnect the AirMini tubing (N20 and AirFit P10 for AirMini masks only)

Disconnect the vent module from the AirMini tubing by gently twisting it and pulling the two pieces apart.

12

Remove the HumidX/HumidX Plus (N20 and AirFit P10 for AirMini masks only)

Disconnect the mask tube from the vent module by gently twisting and pulling the two pieces apart. If you are using the HumidX or HumidX Plus, remove it from the vent module.

Note: The HumidX and HumidX Plus cannot be washed as this will affect the performance of therapy.

Unplug the machine

1.Disconnect the AirMini tubing from the machine by pressing the side buttons on the cuff and pulling it away.

2.Unplug the power supply unit from the power outlet and the machine.

For mask disassembly instructions, refer to your mask user guide.

Cleaning and replacing parts

Detach all components according to the disassembly instructions and make sure that the HumidX/HumidX Plus is removed and stored in a clean, dry location.

Inspect the HumidX/HumidX Plus

1.Check daily for any signs of damage or blockages caused by dirt or dust.

2.It must be replaced within 30 days after opening.

3.When the HumidX/HumidX Plus is not being used, store it in a clean, dry location.

Note: HumidX/HumidX Plus cannot be washed.

English 13

Clean the AirMini tubing

1.Clean and check the air tubing each week. AirMini tubing should be replaced at least every six months.

2.Wash the AirMini tubing in warm water using a mild liquid detergent.

3.Rinse thoroughly and allow to dry out of direct sunlight and/or heat.

4.Check the AirMini tubing and replace it if there are any holes, tears or cracks.

Note: Do not wash the AirMini tubing in a dishwasher or washing machine.

Clean the F20 connector for AirMini

1.Clean and check the F20 connector daily. The connector should be replaced at least every six months.

2.Wash in warm water using a mild liquid detergent.

3.Clean with a soft bristle brush paying particular attention to the vent holes.

4.Rinse well under running water. Shake to remove excess water and allow to dry out of direct sunlight.

5.Check that there is no dirt or dust in the vent holes.

Note: Do not wash in a dishwasher or washing machine.

14

Clean the N20 connector for AirMini

1.Clean and check the N20 connector daily. The connector should be replaced at least every six months.

2.Wash the connector in warm water using a mild liquid detergent.

3.Clean with a soft bristle brush paying particular attention to the vent holes inside the connector.

4.Rinse well under running water. Shake to remove excess water and allow to dry out of direct sunlight.

5.Check that there is no dirt or dust in the vent holes.

Note: Do not wash in a dishwasher or washing machine.

Clean the AirFit P10 for AirMini

Refer to your mask guide for cleaning instructions.

Change the air filter

Replace at least every 6 months. It cannot be washed.

Check and replace it more often if there are any holes or blockages caused by dirt or dust.

To change the air filter:

1.Slide the air filter cover off and remove the old air filter.

2.Insert a new air filter and replace the air filter cover.

English 15

Notes:

Make sure the air filter cover is fitted at all times to prevent water and dust from entering the machine.

The use of ResMed approved hypoallergenic filter will result in a small reduction in the accuracy of the delivered pressure at high leaks.

Cleaning the machine

Wipe the exterior with a dry cloth each week.

Reprocessing

The AirMini tubing, AirMini connectors, vent module and HumidX/HumidX Plus must be discarded and replaced, as they cannot be disinfected and cannot be used between other patients.

For mask cleaning instructions, see your mask user guide.

16

Traveling

You can take your AirMini machine with you wherever you go. Make sure you have the appropriate power supply for the region you are traveling to. For information on purchasing, contact your care provider.

Traveling by airplane

For some airlines, medical devices do not count toward carry-on luggage limits. Please check with your airline for their policy regarding medical equipment.

You can use your AirMini machine on a plane as it meets the Federal Aviation Administration (FAA) requirements. Air travel compliance letters can be downloaded and printed from ResMed.com.

When using the machine on an airplane:

When connected to power, disable Bluetooth (enter airplane mode) by pressing and holding the Bluetooth button for at least ten seconds until the indicator light is white.

Do not use the AirMini app.

Use the Start/Stop button on your machine to start therapy,

To reconnect Bluetooth (exit airplane mode), press the Bluetooth button.

English 17

Troubleshooting

If you have any problems, have a look at the following troubleshooting topics. If you are not able to fix the problem, contact your care provider or ResMed. Do not try to open the machine.

Problem/possible cause

Solution

 

 

Air is leaking from around my mask / Mask is too noisy

Mask may be fitted incorrectly.

The AirMini system may be assembled incorrectly.

Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or run the Mask Fit function using the AirMini app to check your mask fit and seal.

Make sure your AirMini system is correctly assembled. Refer to Setting up and starting therapy for more information.

Air pressure in my mask seems too low / It feels like I am not getting enough air

Ramp may be in progress.

Using the AirMini app, confirm the Ramp function has been enabled. If so, wait for air pressure to build up or turn Ramp Time off using the AirMini app.

HumidX or vent might be blocked.

Inspect the HumidX and vent for blockages or damage. Refer to Cleaning and maintenance for more information.

HumidX may be wet.

The HumidX should be dry when starting

 

therapy. Check if HumidX is wet and

 

replace as necessary.

 

 

18

Problem/possible cause

Solution

I cannot start therapy

Power may not be connected.

Connect the power supply unit and make sure the plug is fully inserted. The green indicator light on the power supply unit should be illuminated.

Check that the green indicator light above the Start/Stop button on the machine is also illuminated.

My therapy has stopped

The air tubing may be disconnected.

You may have high leak and SmartStop enabled.

Ensure that the air tubing is properly connected. Press the Start/Stop button to continue therapy.

Make sure your mask is fitted correctly. Alternatively, disable SmartStop.

The Bluetooth indicator light is constantly white. I cannot pair to my machine

Bluetooth is disabled.

Press the Bluetooth button for 3 seconds.

 

When the Bluetooth indicator light flashes

 

blue, the machine is ready for pairing.

 

 

I cannot scan the QR code to pair my machine

The camera is not focused or broken, or the QR code label is damaged.

The QR code label is unreadable.

You can pair your machine manually by entering into your smart device the fourdigit key found on the back of your machine.

Contact your care provider.

English 19

Problem/possible cause

Solution

The light above the Start/Stop button is flashing green

An error has occurred on the machine.

Unplug your machine. Wait a few seconds and then plug it back in.

If the problem persists, contact your care provider. Do not open the machine.

I am experiencing dryness (dry or blocked nose)

Mask may be fitted incorrectly

Adjust and refit your mask to improve seal.

 

Check that there is no air leaking from the

 

mask.

 

If mask seal is good, try the HumidX Plus.

Humidification may be required.

Try therapy using the HumidX.

 

 

I am using the HumidX/HumidX Plus and I am getting uncomfortable droplets of water on my nose, mask or air tubing

Humidity level is too high

If using the HumidX: Remove the HumidX

 

and try therapy without humidification.

 

If using the HumidX Plus: Try the HumidX

 

instead. This will help to reduce the water

 

droplets.

 

If you are still experiencing problems,

 

contact your care provider.

 

Note: You may need to use the

 

HumidX/HumidX Plus when ambient

 

humidity conditions change.

 

 

20

General warnings and cautions

WARNING

Make sure that you arrange the air tubing so that it will not twist around the head or neck.

Keep the power cord away from hot surfaces.

Make sure the power cord and plug are in good condition and the equipment is not damaged.

If you notice any unexplained changes in the performance of the machine, if it is making unusual sounds, if the machine or the power supply are dropped or mishandled, or if the enclosure is broken, discontinue use and contact your care provider or your ResMed Service Center.

Do not open or modify the machine. There are no user serviceable parts inside. Repairs and servicing should only be performed by an authorized ResMed service agent.

Beware of electrocution. Do not immerse the machine, power supply or power cord in water. If liquids are spilled into or onto the machine, unplug the machine and let the parts dry. Always unplug the machine before cleaning and make sure that all parts are dry before plugging it back in.

Do not use with oxygen. Any sources of oxygen must be located more than 1 m away from the machine to avoid risk of fire and burns.

Do not perform any maintenance tasks while the machine is in operation.

Use only AirMini masks and accessories that allow for normal breathing to avoid asphyxiation.

The use of accessories other than those specified for the machine is not recommended. This may result in increased electromagnetic emissions or decreased immunity of the machine and result in improper operation.

English 21

The machine should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the machine should be observed to verify normal operation in the configuration in which it will be used.

The F20, N20 and P10 connectors for AirMini are fitted with vents. The F20 connector for AirMini is also fitted with an AAV. The AAV and vents have specific safety functions to prevent carbon dioxide build-up in the mask. The AAV and vents should be kept clean and not covered or blocked. The connectors should not be used if the vent or AAV are damaged as they will not be able to perform their safety functions. The connectors should be replaced if the valves of the vent or AAV are damaged, distorted or torn.

The machine is not intended to be operated by persons (including children) with reduced physical, sensory or mental capabilities without adequate supervision by a person responsible for the patient’s safety.

The device has not been tested or certified for use in the vicinity of X-ray, CT or MRI equipment. Do not bring the device within 13 ft (4 m) of X-ray or CT equipment. Never bring the device into an MR environment.

22

CAUTION

Use only ResMed AirMini parts, masks and accessories with the machine. Non-ResMed AirMini parts may reduce the effectiveness of the treatment, may result in excess carbon dioxide rebreathing and/or damage the machine. For compatibility information, refer to www.resmed.com for more information.

Blocking the air tubing and/or air inlet of the machine while in operation could lead to overheating of the machine.

Keep the area around the machine dry, clean and clear of anything (eg, clothes, pillows or bedding) that could block the air inlet or cover the power supply unit.

Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or antibacterial soaps or scented oils to clean the machine or air tubing. These solutions may cause damage and reduce the life of the products.

Do not insert any USB cable into the machine or attempt to plug the power supply unit into a USB device. This may cause damage to the machine or USB device.

English 23

Technical specifications

Units are expressed in cm H2O and hPa. 1 cm H2O is equal to 0.98 hPa.

Power supply unit

 

AC input range:

100–240V, 50–60Hz, 0.5-0.3A

 

115V, 400Hz for aircraft use, 0.4A

DC output:

24V 0.83A

Typical power consumption:

6.3W

Peak power consumption:

27W

If power is interrupted during therapy, the machine will automatically restart therapy when power is restored.

AirMini power

 

Standby power consumption:

1 W

Environmental conditions

 

Operating temperature:

+41°F to +95°F (+5°C to +35°C)

 

Note: Under extreme ambient temperature

 

conditions (104°F/40°C) the air and parts

 

of the tubing near the mask may reach

 

temperatures up to 109°F/43°C. These

 

parts of the tubing are not in contact with

 

the user and the machine remains safe

 

under these extreme conditions.

Operating humidity:

10 to 95% relative humidity, non-

 

condensing

Operating altitude:

Sea level to 8,500' (2,591 m); air pressure

 

range 1013 hPa to 738 hPa

Storage and transport temperature:

-13°F to +158°F (-25°C to +70°C)

Storage and transport humidity:

5 to 95% relative humidity, non-

 

condensing

Air filter

 

Standard:

Material: Polyester non woven fiber

 

Average arrestance: >75% when tested to

 

EN779

24

 

Hypoallergenic:

Material: Blended synthetic fibers in a

 

polypropylene carrier

 

Efficiency: >80% (average) when tested to

 

EN 13274-7

AirMini machine

 

Dimensions:

5.4" (W) x 3.3" (D) x 2.1"(H)

 

136 mm (W) x 84 mm (D) x 52 mm (H)

Weight:

10.6 oz (300 g)

Housing construction:

Flame retardant engineering thermoplastic

Air outlet:

Proprietary connector, 16mm inner

 

diameter. Not compatible with EN ISO

 

5356-1 connectors.

AirMini air tubing

 

Material:

Flexible plastic

Length:

6'2" (1.9 m)

Inner diameter:

0.6" (15 mm)

Electromagnetic compatibility

The AirMini complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2:2014, for residential, commercial and light industry environments. Portable and mobile RF communications equipment should be used no closer to any part of the machine, including cables, than the recommended 10 cm separation distance. The AirMini has been designed to meet EMC standards. However, should you suspect that the machine performance (eg, pressure or flow) is affected by other equipment, move the machine away from the possible cause of interference.

The AirMini complies with Part 15 of the FCC Rules and Industry Canada license-exempt RSS standards. Operation is subject to the following two conditions: This machine may not cause harmful interference, and this machine must accept any interference received, including interference that may cause undesired operation.

FCC ID: QOQBT121, IC: 5123A-BGTBT121

Additional information regarding the FCC Rules and IC compliance for this machine can be found on www.resmed.com/downloads/devices.

Aircraft use

ResMed confirms that machine meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel.

English 25

Wireless technology

 

Technology used:

Bluetooth

Connection types:

SPP, iAP2, GATT

Frequency:

2402 to 2480 MHz

Max RF power output:

+4 dBm

Operating range:

10 m (Class 2)

It is recommended that the machine is at a minimum distance of 0.43" (1.1cm) from the body during operation. Not applicable to masks, air tubing or accessories.

IEC 60601-1 (Edition 3.1) classification

Class II (double insulation), Type BF, Ingress protection IP22.

Operating pressure range

 

AutoSet, AutoSet For Her, CPAP:

4 to 20 cm H2O (4 to 20 hPa)

Maximum single fault steady pressure

Machine will shut down in the presence of a single fault if the steady state pressure exceeds 30 cm H2O (30 hPa) for more than 6 seconds or 40 cm H2O (40 hPa) for more than 1 second.

Pneumatic flow path

1. Flow sensor

2. Blower

3. Pressure sensor

4. AAV (F20 connector only)

5. Vent

6. HumidX (N20, P10 connectors only)

7. Mask

8. Air tubing

9. Machine

10. Inlet filter

Pressure accuracy

Maximum static pressure variation at 10 cm H2O (10 hPa) according to ISO 80601-2-70:2015:

Tested with full face mask: ±0.5 cm H20 (0.5 hPa)

26

Maximum dynamic pressure variation according to ISO 80601-2-70:2015

Machine with full face mask

 

 

Pressure

10 BPM

15 BPM

20 BPM

[cm H2O (hPa)]

 

 

 

4

0.5

0.7

1.0

8

0.5

0.7

1.0

12

0.5

0.7

1.0

16

0.5

0.7

1.0

20

0.5

0.7

1.0

Flow (maximum) at set pressures

The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing:

Pressure

AirMini machine and AirMini air tubing

cm H2O (hPa)

L/min (including uncertainty)

4

119

8

116

12

112

16

108

20

105

Measurement system uncertainties

In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the manufacturer's test equipment is:

For measures of flow

For measures of static pressure For measures of dynamic pressure For measures of volume (< 100 mL) For measures of volume (≥ 100 mL) For measures of time

±1.5 L/min or ± 2.7% of reading (whichever is greater)

±0.15 cm H2O (hPa)

±0.27 cm H2O (hPa)

±5 mL or 6% of reading (whichever is greater)

±20 mL or 3% of reading (whichever is greater)

±10 ms

Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this manual for these items already include the relevant measurement uncertainty from the table above.

English 27

Sound

Declared dual-number noise emission values in accordance with ISO 4871:1996

Pressure level measured according to ISO

29 dBA with uncertainty of 2 dBA

80601-2-70:2015 (CPAP mode)

 

Power level measured according to ISO 80601-

37 dBA with uncertainty of 2 dBA

2-70:2015 (CPAP mode)

 

The A-weighted sound pressure level measured according to ISO 17510:2015:

Air tubing connected to AirMini F20 connector

19 dBA with uncertainty of 3 dBA

with mask

 

 

Air tubing connected to AirMini N20 or P10

15 dBA with uncertainty of 3 dBA

connectors with mask

 

 

The A-weighted sound power level measured according to ISO 17510:2015:

 

Air tubing connected to AirMini F20 connector

27 dBA with uncertainty of 3 dBA

with mask

 

 

Air tubing connected to the AirMini N20 or P10

23 dBA with uncertainty of 3 dBA

connectors with mask

 

 

Pressure flow curve (ISO 17510:2015)

 

 

 

Pressure

Flow

 

[cm H2O

[L/min]

 

(hPa)]

 

 

4

21

 

8

26

 

12

29

 

16

31

 

20

33

Resistance

Inspiratory and expiratory resistance

F20 connector

with AAV open to atmosphere (ISO

 

17510:2015)

 

Inspiration at 50L/min

0.7 cm H2O (hPa)

Expiration at 50L/min

0.5 cm H2O (hPa)

28

 