Renishaw RMP400 User Manual

Page 1
Installation guide
Draft 5 16/04/18
H-6570-8501-01-A
RMP400 high accuracy radio machine probe
Page 2
Draft 5 16/04/18
This document may not be copied or reproduced in whole or in part, or transferred to any other media or language, by any means, without the prior written permission of Renishaw plc.
The publication of material within this document does not imply freedom from the patent rights of Renishaw plc.
Renishaw part no: H-6570-8501-01-A
Issued: ??.????
Page 3
Contents
Draft 5 16/04/18
Before you begin .............................................................1.1
i
Before you begin
Disclaimer
Trade marks
Warranty
Changes to equipment
CNC machines
Care of the probe
Patents
EU declaration of conformity
FCC Information to user (USA only)
WEEE directive
Radio approval
Safety
....................................................................1.5
............................................................1.1
..............................................................1.1
.............................................................1.1
................................................................1.1
.....................................................1.1
...........................................................1.1
.........................................................1.1
.................................................................1.2
...................................................1.3
.............................................1.3
.............................................................1.3
.............................................................1.4
RMP400 basics ...............................................................2.1
Introduction
Getting started
System interface
Trigger Logic™
Probe modes
Configurable settings
Switch-on/switch-off methods
Enhanced trigger filter
Auto-reset function
Multiple probe mode
Acquisition mode
RMP400 dimensions
RMP400 specification
Recommended styli
................................................................2.1
...........................................................2.1
..........................................................2.2
...........................................................2.2
...............................................................2.2
.........................................................2.2
................................................2.2
.....................................................2.4
........................................................2.4
.......................................................2.4
.........................................................2.5
.........................................................2.6
........................................................2.7
..........................................................2.9
Page 4
RMP400 installation guide
Draft 5 16/04/18
System installation ...........................................................3.1
Installing the RMP400 with an RMI or RMI-Q ......................................3.1
Operating envelope .......................................................3.1
Positioning the RMP400 and RMI or RMI-Q ....................................3.2
Contents
ii
Performance envelope .....................................................3.2
Preparing the RMP400 for use .................................................3.3
Fitting the stylus ..........................................................3.3
Installing the batteries .....................................................3.4
Mounting the probe on a shank ..............................................3.5
Stylus on-centre adjustment .................................................3.6
Calibrating the RMP400 ......................................................3.7
Why calibrate a probe? ....................................................3.7
Calibrating in a bored hole or on a turned diameter ...............................3.7
Calibrating in a ring gauge or on a datum sphere ................................3.7
Calibrating the probe length .................................................3.7
Trigger Logic™ ...............................................................4.1
Reviewing the probe settings ..................................................4.1
Multiple probe mode settings ..................................................4.2
Probe settings record ........................................................4.3
Changing the probe settings ...................................................4.4
RMP400 – RMI partnership .................................................... 4.6
RMP400 – RMI-Q partnership .................................................. 4.7
Operating mode .............................................................4.8
Maintenance .................................................................5.1
Maintenance ...............................................................5.1
Cleaning the probe ..........................................................5.1
Changing the batteries .......................................................5.2
Fault-finding .................................................................6.1
Parts list .....................................................................7.1
Page 5
see Section 1, “Before you begin”
Draft 5 16/04/18

Before you begin

1.1

Before you begin

Disclaimer

RENISHAW HAS MADE CONSIDERABLE EFFORTS TO ENSURE THE CONTENT OF THIS DOCUMENT IS CORRECT AT THE DATE OF PUBLICATION BUT MAKES NO WARRANTIES OR REPRESENTATIONS REGARDING THE CONTENT. RENISHAW EXCLUDES LIABILITY, HOWSOEVER ARISING, FOR ANY INACCURACIES IN THIS DOCUMENT.

Trade marks

RENISHAW and the probe symbol used in the
RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other countries. apply innovation and names and designations of other Renishaw products and technologies are trade marks of Renishaw plc or its subsidiaries.

Warranty

Equipment requiring attention under warranty must be returned to your equipment supplier.
Unless otherwise specifically agreed in writing between you and Renishaw, if you purchased the equipment from a Renishaw company, the warranty provisions contained in Renishaw’s CONDITIONS OF SALE apply. You should consult these conditions in order to find out the details of your warranty but, in summary, the main exclusions from the warranty are if the equipment has been:
neglected, mishandled or inappropriately used;
or
modified or altered in any way except with the
prior written agreement of Renishaw.
If you purchased the equipment from any other supplier, you should contact them to find out what repairs are covered by their warranty.
Google Play and the Google Play logo are trademarks of Google LLC.
Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.,registered in the U.S. and other countries.
All other brand names and product names used in this document are trade names, trade marks, or registered trade marks of their respective owners.

Changes to equipment

Renishaw reserves the right to change equipment specifications without notice.

CNC machines

CNC machine tools must always be operated by fully trained personnel in accordance with the manufacturer’s instructions.

Care of the probe

Keep system components clean and treat the probe as a precision tool.
Page 6
Before you begin
Draft 5 16/04/18

Patents

Features of the RMP400, and other similar Renishaw products, are the subject of one or more of the following patents and/or patent applications:
1.2
Patents will be listed here when we have them.. Possibly all RMP40 patents and strain gauge patents (Steve to confirm)
Page 7

EU declaration of conformity

Draft 5 16/04/18

FCC Information to user (USA only)

C
Renishaw plc declares that the RMP400 complies with the applicable standards and regulations.
Contact Renishaw plc or visit www.renishaw.com/mtpdoc for the full EU declaration of conformity.

WEEE directive

47 CFR Section 15.19
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference that may cause undesired operation.
47 CFR Section 15.21
The user is cautioned that any changes or modifications not expressly approved by Renishaw plc or authorised representative could void the user’s authority to operate the equipment.
Before you begin
1.3
The use of this symbol on Renishaw products and/or accompanying documentation indicates that the product should not be mixed with general household waste upon disposal. It is the responsibility of the end user to dispose of this product at a designated collection point for waste electrical and electronic equipment (WEEE) to enable reuse or recycling. Correct disposal of this product will help to save valuable resources and prevent potential negative effects on the environment. For more information, please contact your local waste disposal service or Renishaw distributor.
Page 8
Before you begin
Draft 5 16/04/18
1.4

Radio approval

Will be listed here, when there are some (testing to take place first).
Page 9

Safety

Draft 5 16/04/18
Information to the user
Do not get batteries wet.
If a battery is damaged, exercise caution when
handling it.
The RMP400 is supplied with two non­rechargeable ½AA lithium-thionyl chloride batteries (approved to BS EN 62133:2013 [IEC 62133:2012]). Once the charge in these batteries is depleted, do not attempt to recharge them.
The use of this symbol on the batteries, packaging or accompanying documents indicates that used batteries should not be mixed with general household waste. Please dispose of the used batteries at a designated collection point. This will prevent potential negative effects on the environment and human health which could otherwise arise from inappropriate waste handling. Please contact your local authority or waste disposal service concerning the separate collection and disposal of batteries. All lithium and rechargeable batteries must be fully discharged or protected from short circuiting prior to disposal.
Please ensure replacement batteries are of the correct type and are fitted in accordance with the instructions in this manual (see page 5.2, “Changing the batteries”), and as indicated on the product. For specific battery operating, safety and disposal guidelines, please refer to the battery manufacturer’s literature.
Ensure that all batteries are inserted with the
correct polarity.
Do not store batteries in direct sunlight or rain.
Do not heat or dispose of batteries in a fire.
Avoid forced discharge of the batteries.
Please ensure that you comply with international and national battery transport regulations when transporting batteries or the products.
Lithium batteries are classified as dangerous goods and strict controls apply to their shipment by air. To reduce the risk of shipment delays, if you need to return the products to Renishaw for any reason, do not return any batteries.
In all applications involving the use of machine tools or CMMs, eye protection is recommended.
The RMP400 has a glass window. Handle with care if broken to avoid injury.
Information to the machine supplier/ installer
It is the machine supplier’s responsibility to ensure that the user is made aware of any hazards involved in operation, including those mentioned in Renishaw product literature, and to ensure that adequate guards and safety interlocks are provided.
Under certain circumstances, the probe signal may falsely indicate a probe seated condition. Do not rely on probe signals to halt the movement of the machine.
Information to the equipment installer
All Renishaw equipment is designed to comply with the relevant EC and FCC regulatory requirements. It is the responsibility of the equipment installer to ensure that the following guidelines are adhered to, in order for the product to function in accordance with these regulations:
any interface MUST be installed in a position
away from any potential sources of electrical noise, i.e. power transformers, servo drives etc;
Before you begin
1.5
Do not short-circuit the batteries.
Do not disassemble, pierce, deform or apply
excessive pressure to the batteries.
Do not swallow the batteries.
Keep the batteries out of the reach of children.
all 0 V/ground connections should be
connected to the machine “star point” (the “star point” is a single point return for all equipment ground and screen cables). This is very important and failure to adhere to this can cause a potential difference between grounds;
Page 10
all screens must be connected as outlined in
Draft 5 16/04/18
the user instructions;
cables must not be routed alongside high
current sources, i.e. motor power supply cables etc, or be near high-speed data lines;
Before you begin
1.6
cable lengths should always be kept to a
minimum.
Equipment operation
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Page 11
see Section 2, “RMP400 basics”
Draft 5 16/04/18

RMP400 basics

2.1

Introduction

The RMP400 offers an unrivalled combination of size, accuracy, reliability and robustness and, allows high-accuracy probing on small to medium machining centres or other machines where line-of-sight problems affect optical signal transmission.
Successfully combining patented RENGAGE™ strain gauge technology with the patented frequency hopping radio transmission system of the RMP40, the RMP400 provides existing probe users with a simple upgrade to solid-state strain gauge technology and all the associated benefits this brings:
excellent 3D performance to allow probing of
contoured surfaces;
improved repeatability in all probing directions;
a low triggering force combined with low pre-
travel variation to provide high accuracy, even when used with long styli;
Performing this calibration for every 3D direction can be time-consuming.
The RMP400 has almost no pre-travel variation, so a single calibration value may be used for any probing angle in 2D or 3D. This results in a vastly reduced calibration time. An additional benefit is a corresponding reduction in errors introduced by environmental changes within the machine during a long calibration cycle.
The ability to be used in applications
where axial and radial reorientations are used, enabled by the use of solid state accelerometers:
The auto-reset function is required and recommendations should be followed for optimum metrology performance.

Getting started

Three multicolour probe LEDs provide visual indication of selected probe settings.
a proven ten-fold improvement in life (10 million
triggers); ????
the elimination of reseat failures;
high resistance to machine tool vibration;
resistance to shock and false triggering
through the use of solid state accelerometers.
In addition to providing high-accuracy measurement on your machine tool, the RMP400 also offers:
Faster calibration:
On complex 3D parts, it is common to measure in several different directions. Each direction of a standard mechanical probe must be calibrated, to ensure that the pre-travel variation is compensated in the measurement.
For example:
Switch-on and switch-off methods
Probe status – triggered or seated
Battery condition
Batteries are inserted or removed as shown (see page 3.5, “Installing the batteries”) for further information).
On insertion of batteries, the LEDs will begin to flash (see page 4.1, “Reviewing the probe settings”).
Page 12

System interface

Draft 5 16/04/18

Configurable settings

The RMI and RMI-Q are integrated interfaces/ receivers used to communicate between the RMP400 probe and the machine control.
RMP400 basics
2.2

Trigger Logic™

Trigger Logic (see Section 4, “Trigger Logic™”) is a method that allows the user to view and select all available mode settings in order to customise a probe to suit a specific application. Trigger Logic is activated by battery insertion and uses a sequence of stylus deflections (triggering) to systematically lead the user through the available choices to allow selection of the required mode options.
A Trigger Logic app is available that simplifies this process with clear, interactive instructions and informative videos and is available for download on the following app stores.

Switch-on/switch-off methods

The following switch-on/switch-off options are user-configurable.
Radio on/Radio off
Radio on/Timer off
Spin on/Spin off
Spin on/Timer off
or
Current probe settings can be reviewed by simply removing the batteries for a minimum of
seconds, and then replacing them to activate
five
the Trigger Logic review sequence.
Pr
obe modes
The RMP400 probe can be in one of three modes:
Standb
signal.
NOTE: The RMP400 will enter hibernation mode should the system interface be powered off or out of range for a period of 30 seconds (only applicable to “radio on” mode).
y mode – Probe is waiting for a switch-on
Operational mode – When activated by one of
the switch-on methods, the probe is switched on and ready for use.
Configuration mode – Ready to change the probe settings using Trigger Logic.
Page 13
RMP400 switch-on method
Draft 5 16/04/18
Switch-on options are configurable
RMP400 switch-off method
Switch-off options are configurable
Probe ready time
Radio on
Radio switch on is commanded by machine input.
Spin on
Spin at 500 rev/min for one second minimum.
Radio off
Radio switch off is commanded by machine input. A timer automatically switches the probe off 90 minutes after the last trigger if it is not turned off by machine input.
Timer off (timeout)
Timeout will occur 12, 33 or 134 seconds (user configurable) after the last probe trigger or reseat.
Spin off
Spin at 500 rev/min for one second minimum. A timer automatically switches the probe off 90 minutes after the last trigger if it is not spun.
Timer off (timeout)
Timeout will occur 12, 33 or 134 seconds (user configurable) after the last probe trigger or reseat.
1.7 seconds maximum.
RMP400 basics
2.3
2.5 seconds. (The probe must be stationary for 2.5 seconds minimum after it has stopped spinning.)
NOTES:
In “radio on” mode, the switch-on time is user selectable “fast” or “standard” when using RMI-Q (selection is made in RMI-Q). Otherwise
1.7 seconds.
For more information on the user selectable switch-on time when operating with RMI-Q, please refer to the installation guide RMI-Q radio machine interface (Renishaw part no. H-5687-8504).
In “radio on” mode, the switch-on time assumes a good radio communication link. In a poor RF environment this may rise to a maximum of
3.0 seconds.
In “spin on” mode, the one second starts from the moment the spindle reaches 500 rev/min.
The RMP400 must be on for a minimum of one second before being switched off.
Page 14
RMP400 installation guide
Draft 5 16/04/18

Enhanced trigger filter

Probes subjected to high levels of vibration or shock loads may output signals without having contacted any surface. The enhanced trigger filter improves the probe’s resistance to these effects.
RMP400 basics
2.4
When the filter is enabled, a constant 8 ms or 16 ms delay is introduced to the probe’s output. The factory setting is 8 ms. If false triggering is noticed, then consider increasing the filter delay to 16 ms.

Auto-reset function

In previous strain gauge products, the probe was required to be turned off during reorientation moves. The auto-reset function in the RMP400 can compensate for stylus forces, resulting from changes in probe orientation, that can cause the probe to trigger.
This feature is controlled by solid state accelerometers and is suitable for applications where axial and radial reorientation of the probe is applied.
To achieve optimum metrology performance when the auto-reset function is turned on, a dwell is recommended before making a programmed move that follows any reorientation of the probe.
When using a stylus of up to 150 mm long, a 0.2 second dwell is necessary. In most applications the machine response time will adequately provide this.
Up to four RMP400s can be used with a single RMI-Q in “radio on/radio off” mode. For further details of this functionality, please refer to the installation guide RMI-Q radio machine interface (Renishaw part no. H-5687-8504).
NOTES:
Multiple probe mode is a function of the RMP400, as such, the option will not appear when the “radio on” option has been selected.
RMP400 probes which are set to “multiple probe mode on” can coexist alongside any number of RMP400 probes set to “multiple probe mode off”.
To allow multiple radio probes to work in close proximity, and with a single RMI or RMI-Q, 16 choices of “mode on” colours are available, each representing a different machine tool installation. (see page 4.2, “Multiple probe settings”).
All probes operating with a single RMI or RMI-Q must be set to the same “mode on” colour choice; any multiple probes located on adjacent machines must all be set to an alternative “mode on” colour choice.
NOTE: Each probe per “mode on” colour choice needs to be partnered with the RMI or RMI-Q. By configuring multiple probes to a single “mode on” colour choice, all probes using this “mode on” colour choice will have the same identity.
When using a heavy stylus configuration, or a stylus longer than 150 mm, it is recommended that the RMP400 should be turned off during reorientation moves.

Multiple probe mode

The RMP400 can be configured, using Trigger Logic, to allow multiple radio probes in “spin on/ spin off” or “shank on/shank off” to be used with a single RMI or RMI-Q.
The probe to be partnered is partnered after selecting the multiple probe mode setting and choosing the “mode on” option. (see page 4.4, “Changing the probe settings”).
There is no limit to the number of probes that can be used with a single RMI or RMI-Q so long as they all have the same “mode on” colour choice. All RMP400 probes are factory set to “mode off”.
The addition of any further probe(s) into a single probe installation will require that all probes are reconfigured to the same “mode on” colour choice and that one of the probes is then repartnered with the RMI or RMI-Q.
Page 15
The addition of any further probe(s), or
Draft 5 16/04/18
replacements, into a multi-probe installation can be achieved simply through the reconfiguration of the probe to the same “mode on” colour choice.

Acquisition mode

System set-up is achieved using Trigger Logic and powering-on the RMI or RMI-Q.
Partnering is only required during initial system set-up. Further partnering is only required if either the RMP400 or RMI/RMI-Q is changed.
NOTES:
Systems using the RMI-Q can be partnered with up to four RMP400s manually. Alternatively this can be achieved by using ReniKey; a Renishaw machine macro cycle which does not require the RMI-Q to be power cycled.
RMP400 basics
2.5
For more information or to download ReniKey free of charge visit:
www.renishaw.com/mtpsupport/renikey
Partnering by ReniKey is not available for RMI.
Partnering will not be lost by reconfiguration of probe settings or when changing batteries, except where multiple probe mode is selected.
Partnering can take place anywhere within the operating envelope.
Page 16

RMP400 dimensions

Draft 5 16/04/18
RMP400 basics
2.6
50 (1.97)
M4 stylus
11°
Ø40 (Ø1.57)
Dimensions given in mm (in)
Stylus length ±X/±Y Z
50 (1.97) 12 (0.47) 6 (0.24)
100 (3.94) 22 (0.87) 6 (0.24)
11°
RMP400 window
Stylus overtravel limits
19 (0.75)
50.5 (1.99)
Battery cassette
A range of probe-ready shanks is available from Renishaw
Probe status LED
Page 17

RMP400 specification

Draft 5 16/04/18
Principal application
Dimensions
Weight (without shank)
Transmission type
Radio frequency
Switch-on methods
Switch-off methods
Probe feedrate (minimum)
Spindle speed (maximum)
Operating range
Receiver/interface
Sense directions
Unidirectional repeatability
Workpiece inspection and job set-up on multi-tasking machines, machining centres and gantry machining centres.
Length Diameter
With batteries Without batteries
Frequency hopping spread spectrum (FHSS) radio
2400 MHz to 2483.5 MHz
Radio M-code or spin on
Radio M-code, timer or spin off
3 mm/min (0.12 in/min) (see note 6)
1000 rev/min
Up to 15 m (49.2 ft)
RMI or RMI-Q combined interface and receiver unit
±X, ±Y, +Z
0.25 µm (10 µin) 2s – 50 mm stylus length (see note 1)
0.35 µm (14 µin) 2s – 100 mm stylus length
50.5 mm (1.99 in) 40 mm (1.57 in)
262 g (9.24 oz) 242 g (8.54 oz)
RMP400 basics
2.7
X, Y (2D) form measurement deviation
X, Y, Z (3D) form measurement deviation
Stylus trigger force
(see notes 2 and 5)
XY plane (typical minimum) +Z direction (typical minimum)
Stylus overtravel force
XY plane (typical minimum) +Z direction (typical minimum)
Stylus overtravel
Note 1 Performance specification is tested at a standard test velocity of 240 mm/min (9.45 in/min) with a 50 mm (1.97 in)
carbon fibre stylus. Significantly higher velocity is possible depending on application requirements.
Note 2 Trigger force, which is critical in some applications, is the force exerted on the component by the stylus when the
probe triggers. The maximum force applied will occur after the trigger point (overtravel). The force value depends on related variables including measuring speed and machine deceleration. RENGAGE™ equipped probes offer ultra­low trigger forces.
Note 3
Note 4
Note 5 These are the factory settings, manual adjustment is not possible.
Note 6 Speeds below 3 mm/min commonly occur when manually moving the probe using the handwheel with a very fine
Stylus overtravel force in the XY plane typically occurs 70 µm (2755.91 µin) after the trigger point and rises by
0.1 N mm 10 gf/mm (9.1 oz/in) until the machine tool stops (in the high force direction and using a 50 mm (1.97 in) carbon fibre stylus).
Stylus overtravel force in the +Z direction occurs 10 µm (393.70 µin) to 11 µm (433.07 µin) after the trigger point and rises by 1.2 N/mm, 122 gf/mm (109.60 oz/in) until the machine tool stops.
feedrate.
±0.25 µm (10 µin) – 50 mm stylus length (see note 1) ±0.25 µm (10 µin) – 100 mm stylus length
±1.00 µm (40 µin) – 50 mm stylus length (see note 1) ±1.75 µm (70 µin) – 100 mm stylus length
0.06 N, 6 gf (0.22 ozf)
2.55 N, 260 gf (9.17 ozf)
1.04 N, 106 gf (3.74 ozf) (see note 3)
5.50 N, 561 gf (19.78 ozf) (see note 4)
XY plane +Z plane
TBC by end of March
±11° 6 mm (0.23 in)
TBC
Page 18
RMP400 installation guide
Draft 5 16/04/18
RMP400 basics
2.8
Typical battery life
Environment
Battery types
Battery reserve life
Typical battery life
Battery type
Standby life 5% usage
IP rating IPX8, BS EN 60529:1992+A2:2013
(IEC 60529:1989+A1:1999+A2:2013)
IK rating IK01 (BS EN IEC 62262: 2002) [for glass
window]
Storage temperature
Operating temperature +5 °C to +50 °C (+41 °F to +122 °F)
2 × ½AA 3.6 V lithium-thionyl chloride (LTC)
Approximately one week after a low battery warning is first given (based on 5% usage).
See the table below.
Spin switch on Radio switch on
(72 minutes/day)
–10 °C to +70 °C (+14 °F to +158 °F)
Standby life 5% usage
(72 minutes/day)
Continuous use
Lithium-thionyl
chloride
NOTE: Using RMP400 with “fast radio on” mode will result in a 20% reduction in standby battery life
and a 10% reduction in 5% usage battery life.
230 days 90 days 230 days 90 days 165 hours
Page 19

Recommended styli

Draft 5 16/04/18
High modulus carbon fibre styli are designed to minimise pre-travel and improve accuracy, as the stem material is extremely stiff. This inherent stiffness makes the following styli most suitable for strain gauge applications.
B
RMP400 basics
2.9
D
A
C
Part number A-5003-7306
Carbon fibre
A
Ball diameter mm (inch) 6 (0.24) 6 (0.24) 6 (0.24) 6 (0.24)
B
Length mm (inch) 50 (1.97) 100 (3.94) 150 (5.91) 200 (7.88)
C
Stem diameter mm (inch) 4.5 (0.18) 4.5 (0.18) 4.5 (0.18) 4.5 (0.18)
D
EWL mm (inch) 38.5 (1.52) 88.5 (3.48) 138.5 (5.45) 188.5 (7.42)
Mass in g (oz) 4.1 (0.14) 6.2 (0.22) 7.5 (0.26) 8.7 (0.31)
The featured range of solid carbon fibre styli ensure the best possible performance of the RMP400.
It is possible that the featured range of solid carbon fibre styli may not be suitable for every RMP400 application and that it may be necessary to select specialised styli configurations to meet specific application requirements.
In applications where specialised styli are to be used, it may be beneficial to reduce the speed of probing moves. It has been seen in some cases that specialist styli configurations do not exhibit the probing characteristics and performance that would have otherwise been expected and achieved when using standard styli. Reducing the speed of the probing move may, in some cases, improve the performance of the probe.
A-5003-6510 Carbon fibre
When selecting components for an application specific stylus, it is recommended that a configuration with the least number of components is chosen. The stylus diameter should always be as large as possible and the overall stylus length kept to a minimum. If a stem with a reduced diameter is required, then it is recommended that an M4 stem with a short length and reduced diameter is selected.
A-5003-6511 Carbon fibre
A-5003-6512 Carbon fibre
Page 20
RMP400 basics
Draft 5 16/04/18
2.10
This page is intentionally left blank.
Page 21
see Section 3, “System installation”
Draft 5 16/04/18

System installation

Installing the RMP400 with an RMI or RMI-Q
3.1
CNC machining centre spindle
RMP400 inspection probe
Stylus
Workpiece
Typical tool setting probe

Operating envelope

Radio transmission does not require line-of-sight between the probe and interface as it works via reflected paths, and will pass through gaps and machine tool windows. This allows easy installation, either inside or outside the machine enclosure, as long as the probe and RMI or RMI-Q are kept within the performance envelope shown overleaf.
RMI or RMI-Q interface
When operating, do not cover the probe glass window, RMI or RMI-Q with your hands, as this will affect the performance.
Mounting bracket
CNC
machine
control
Cable
Interface unit
Coolant and swarf residue accumulating on the RMP400 and RMI or RMI-Q may have a detrimental effect on transmission performance. Wipe clean as often as is necessary to maintain unrestricted transmission.
Page 22
Positioning the RMP400 and RMI or
Draft 5 16/04/18
RMI-Q
NOTE: Installing the RMP400 and RMI or RMI-Q with the RMP400 in radio-on configuration
System
installation
3.2
The probe system should be positioned so that the optimum range can be achieved over the full travel of the machine’s axes. Always face the front cover of the RMI or RMI-Q in the general direction of the machining area and the tool magazine, ensuring both are within the performance envelope shown below. To assist in finding the optimum position of the RMI or RMI-Q, the signal quality is displayed on an RMI or RMI-Q signal LED.
The RMP400 has a built-in hibernation mode (battery-saving mode) that saves battery life when the RMI or RMI-Q is unpowered in radio-on (radio-off or timer-off) configurations. The RMP400 goes into hibernation mode 30 seconds after the RMI or RMI-Q is unpowered (or the RMP400 is out of range). When in hibernation mode, the RMP400 checks for a powered RMI or RMI-Q every 30 seconds. If found, the RMP400 goes from hibernation mode to standby mode, ready for radio-on.

Performance envelope

The RMP400 and RMI or RMI-Q must be within each other’s performance envelope, as shown below. The performance envelope shows line-of­sight performance, however, this is not necessary for the RMP400 radio transmission as it will operate with any reflected radio path provided that the reflected path length does not exceed the 15 m (49.2 ft) operating range.
Performance envelope when using the RMP400 with the RMI or RMI-Q
75°
60°
45°
75°
15 (49)
60°
45°
30°
15°
15°
30°
45°
60°
Typical plot at +20 °C (+68 °F) Transmission range in m (ft)
75°
10 (33)
5 (16)
90°
75°
75°
60°
45°
30°
15°
15°
30°
45°
60°
Operating and switch-on / switch-off
30°
15°
15°
30°
45°
60°
5 (16)
10 (33)
15 (49)
75°
Page 23

Preparing the RMP400 for use

Draft 5 16/04/18

Fitting the stylus

1
System
3.3
installation
2
M-5000-3707
1.8 Nm – 2.2 Nm
(1.3 lbf.ft – 1.6 lbf.ft)
Page 24
System
Draft 5 16/04/18
installation
3.4

Installing the batteries

NOTES:
See (Section 5,“Maintenance”)for a list of suitable battery types.
If dead batteries are inadvertently inserted, the LEDs will remain a constant red.
Do not allow coolant or debris to enter the battery compartment. When inserting batteries, check that the battery polarity is correct.
After the batteries have been inserted, the LEDs will display the current probe settings, for details, (see Section 4, “Trigger Logic™”).
1
3
2
Please remove the battery isolation device from the battery compartment before use.
4
Page 25

Mounting the probe on a shank

Draft 5 16/04/18
1
2 mm A/F
× 4
× 2
2 mm A/F
× 2
× 4
System
3.5
installation
2
2 mm A/F
× 2
0.5 Nm – 1.5 Nm
(0.4 lbf.ft – 1.1 lbf.ft)
Page 26

Stylus on-centre adjustment

Draft 5 16/04/18
System
installation
3.6
NOTES:
If a probe and shank assembly is dropped, it must be rechecked for correct on-centre adjustment.
Do not hit or tap the probe to achieve on-centre adjustment.
1
× 4
360°
±10 µm
2
1.5 Nm – 2.2 Nm
(1.1 lbf.ft – 1.6 lbf.ft)
360°
±10 µm
× 2
3
× 4
1.5 Nm – 2.2 Nm (1.1 lbf.ft – 1.6 lbf.ft)
360°
±2.5 µm
Page 27

Calibrating the RMP400

Draft 5 16/04/18

Why calibrate a probe?

A spindle probe is just one component of the measurement system which communicates with the machine tool. Each part of the system can introduce a constant difference between the position that the stylus touches and the position that is reported to the machine. If the probe is not calibrated, this difference will appear as an inaccuracy in the measurement. Calibration of the probe allows the probing software to compensate for this difference.

Calibrating in a bored hole or on a turned diameter

Calibrating a probe, either in a bored hole or on a turned diameter of known size, automatically stores values for the offset of the stylus ball to the spindle centre line. The stored values are then used automatically in the measuring cycles. Measured values are compensated by these values so that they are relative to the true spindle centre line.

Calibrating in a ring gauge or on a datum sphere

System
3.7
installation
During normal use, the difference between the touch position and the reported position does not change, but it is important that the probe is calibrated in the following circumstances:
when a probe system is to be used for the first
time;
when the enhanced trigger filter delay is
changed;
when a new stylus is fitted to the probe;
when it is suspected that the stylus has
become distorted or that the probe has been crashed;
at regular intervals to compensate for
mechanical changes of your machine tool;
if repeatability of relocation of the probe shank
is poor. In this case, the probe may need to be recalibrated each time it is selected.
It is good practice to set the tip of the stylus on­centre, because this reduces the effect of any variation in spindle and tool orientation (see page
3.6, “Stylus on-centre adjustment”). A small amount of run-out is acceptable, and can be compensated for as part of the normal calibration process.
Calibrating a probe either in a ring gauge or on a datum sphere with a known diameter automatically stores one or more value for the radius of the stylus ball. The stored values are then used automatically by the measuring cycles to give the true size of the feature. The values are also used to give true positions of single surface features.
NOTE: The stored radius values are based on the true electronic trigger points. These values are different from the physical sizes.

Calibrating the probe length

Calibrating a probe on a known reference surface determines the length of the probe, based on the electronic trigger point. The stored value for length is different from the physical length of the probe assembly. Additionally, the operation can automatically compensate for machine and fixture height errors by adjusting the probe length value that is stored.
Three different operations are to be used when calibrating a probe. They are:
calibrating either in a bored hole or on a turned
diameter of known position;
calibrating either in a ring gauge or on a datum
sphere;
calibrating the probe length.
Page 28
System
Draft 5 16/04/18
installation
3.8
This page is intentionally left blank.
Page 29
see Section 4, “Trigger Logic™”
Draft 5 16/04/18

Trigger Logic™

Reviewing the probe settings

4.1
1
Key to the symbols
LED short flash
LED long flash
2
> 5 s
3
Switch-on method
Radio on
(omitted if “multiple probe mode” is selected)
Switch-off method
Radio off or
Spin off
Auto-reset off
Trigger filter
on 8 ms
Auto-reset off
Trigger filter
or
Short timeout
or
Enhanced trigger filter and auto-reset facility
on 16 ms
12 s
or
Auto-reset on
Trigger filter
Medium timeout
or
on 8 ms
33 s
Auto-reset on
Trigger filter
or
Spin on
or
or
on 16 ms
Long timeout
134 s
Auto-reset off
Trigger filter
or
LED check
off
Hibernation mode (only for “radio on”)
On
30 s
Multiple probe mode (omitted for “radio on”)
see “Multiple probe mode settings” to view all 16 choices
Mode off Mode on
Machine 1
or
Battery good
or
Machine 2
or
Battery status
or
Probe in standby mode (after 5 seconds)
On 5 s
or
or
Battery low
Off
or
Machine 16
Page 30
Trigger Logic
Draft 5 16/04/18
4.2
RMP400 installation guide
Multiple probe mode settings
Deflect the stylus for less than 4 seconds to cycle to the next setting.
Multiple probe mode
Mode off Mode on
Machine 1
or
or
Machine 2
or
Machine 3
or
Machine 4
Machine 5
Machine 9
Machine 13
Machine 6
or
Machine 10
or
Machine 14
or
Machine 7
or
Machine 11
or
Machine 15
or
or
Machine 12
or
Machine 16
or
Return to
“Mode off”
Machine 8
Page 31

Probe settings record

Draft 5 16/04/18
This page is provided to note your probe’s settings.
Switch-on method
Radio on
Spin on
tick
Factory settings
tick
New settings
Trigger Logic
4.3
Switch-off method
Auto-reset and enhanced trigger filter
Radio or spin
Short timeout (12 s)
Medium timeout (33 s)
Long timeout (134 s)
Auto reset off / Trigger filter on (8 ms)
Auto reset off / Trigger filter on (16 ms)
Auto reset on / Trigger filter on (8 ms)
Auto reset on / Trigger filter on (16 ms)
Auto reset off / Trigger filter off
Hibernation mode
Multiple probe mode
Factory settings are for kit (A-6570-0001) only.
On (30 s)
On (5 s)
Off
Off (factory set)
On (machine number) See “Multiple
probe settings”
RMP400 serial no ........................................
Page 32
RMP400 installation guide
Draft 5 16/04/18
Trigger Logic
4.4
Probe partnering function
The probe partnering function enables the RMP400 to be partnered with the RMI or RMI-Q independently of the configuration process for other probe settings. To partner RMP400 with RMI or RMI-Q, insert the batteries or, if they have already been installed, remove them for five seconds and then refit them.
Following an LED check, the RMP400 will proceed to show the probe settings, this will end with “Probe status” being displayed. If the battery power is good, probe status will be eight green flashes. If battery power is low, each green flash will be followed by a blue flash.
Key to the symbols
LED short flash
LED long flash
Deflect the stylus for less than 4 seconds to move to the next menu option.
To exit, leave the stylus untouched for more than 120 seconds.
Acquisition successful. Probe is now in standby.
Whilst the “Probe status” is being displayed, deflect and immediately release the stylus to enter “Acquisition mode”.
Acquisition mode off” will be displayed as a sequence of light blue flashes, at this point the RMI or RMI-Q must be turned on.
On the RMP400 select “Acquisition mode on” by deflecting the stylus for less than 4 seconds. After a successful acquisition, the RMP400 will timeout after 8 seconds and then go into standby. If “Acquisition mode on” is not selected, the RMP400 will timeout after 120 seconds and then go into standby (see page 4.10, “RMP400 – RMI partnership”) or (see page 4.11, “RMP400 – RMI-Q partnership”).
1
2
> 5 s
3
LED check
All probe settings will be shown, ending with “Probe status” being displayed.
Probe status (seated)
Battery good
or
Whilst the “Probe status” is being displayed, deflect and release the stylus to enter “Acquisition mode off”. Probe status will flash red to acknowledge this.
Battery good
Battery low
Probe status
or
Battery low
At this point turn on either the RMI or RMI-Q.
Page 33
After 8
Draft 5 16/04/18
seconds
Acquisition mode
Acquisition mode off
120 seconds
probe in standby
If acquisition is unsuccessful “Acquisition mode off” will be displayed again after 8 seconds. Deflect the stylus for less than 4 seconds to select “Acquisition
mode on” again.
Acquisition mode on
8 seconds
probe in standby
Acquisition
successful
Trigger Logic
4.5
Page 34
RMP400 installation guide
Draft 5 16/04/18

Changing the probe settings

Insert the batteries or, if they have already been installed, remove them for five seconds and then refit them.
1
2
> 5 s
3
Trigger Logic
4.6
Following the LED check, immediately deflect the stylus and hold it deflected until eight red flashes have been observed (if the battery power is low, each red flash will be followed by a blue flash).
Keep the stylus deflected until the “Switch-on
method” setting is displayed, then release it.
CAUTION: Do not remove the batteries whilst
in configuration mode. To exit, leave the stylus untouched for more than 20 seconds.
Key to the symbols
LED short flash
LED long flash
Deflect the stylus for less than 4 seconds to move to the next menu option.
Deflect the stylus for more than 4 seconds to move to the next menu.
To exit, leave the stylus untouched for more than 120 seconds.
To exit, leave the stylus untouched for more than 20 seconds.
LED check
Probe status (triggered)
Battery good
or
Battery low
Switch-on method
(omitted if “multiple probe mode” was selected)
Radio on
Switch-off method
Radio off or
Spin off
Enhanced trigger filter setting and auto-reset facility
Auto reset off
Trigger filter
on 8 ms
Auto reset off
Trigger filter
on 16 ms
Short
timeout
12 s
Auto reset on
Trigger filter
Medium
timeout
on 8 ms
33 s
Spin on
Long
timeout
134 s
Auto reset on
Trigger filter
on 16 ms
Auto reset off
Trigger filter
off
Page 35
Acquisition mode
Draft 5 16/04/18
Acquisition mode off
Acquisition mode on
120 seconds
probe in standby
Trigger Logic
Hibernation mode (only for “radio on”)
On
30 s
Stop triggering here, unless “Multiple probe
mode” is required, in which case deflect the
stylus for more than 4 seconds.
Multiple probe mode (omitted for “radio on”)
(see “Multiple probe mode settings” to view all 16 choices)
Mode off Mode on
Machine 1 Machine 2
If no changes are made in “Multiple probe mode”, then deflecting the
stylus for more than 4 seconds will return the probe settings to beginning
of the Trigger Logic menu. If “Multiple probe mode” is selected, proceed to
Acquisition mode” to repartner one probe with the RMI-Q.
On 5 s
Off
Machine 16
4.7
Acquisition mode
Acquisition mode off
Return to the beginning of the Trigger Logic menu
NOTE: If using “Multiple probe mode”, refer to the installation guide RMI radio machine interface (Renishaw part no. H-4113-8554) or the installation guide RMI-Q radio machine interface (Renishaw part no. H-5687-8504).
NOTE: To partner an RMP400 with an RMI please see “RMP400 – RMI partnership”. Once acquisition has been successful, the RMP400 will revert to “Acquisition mode off”.
Acquisition mode on
120 seconds
probe in standby
New settings
complete,
probe in standby
NOTE: Further probes used require the same
Multiple probe mode” setting, but do not need to be partnered with the RMI or RMI-Q.
NOTE: To partner an RMP400 with an RMI-Q please see “RMP400 – RMI-Q partnership”. Once acquisition has been successful, the RMP400 will revert to “Acquisition mode off”.
Page 36
RMP400 installation guide
Draft 5 16/04/18
Trigger Logic
4.8
Master reset function
RMP400 features a master reset function to assist user’s who have mistakenly changed the probe settings into an unintended state.
The application of the master reset function will clear all current probe settings and return the probe to default settings.
The default settings are as follows:
Radio switch-on
Radio switch-off
Auto-reset on, Enhanced trigger filter on 8 ms
Hibernation mode on 30 s
Multiple probe mode off
The default settings may not be representative of the required probe settings. Further configuration of RMP400 may subsequently be necessary to achieve the required probe settings.
To reset the probe
1. First enter into the Trigger Logic™ menu and ensure that the stylus is no longer deflected.
2. From within the Trigger Logic menu, hold the stylus deflected for 20 seconds. After this the status LEDs will proceed to flash yellow eight times. A confirmation for master reset is required, if nothing is done the probe will timeout.
To confirm that a master reset is required, release the stylus and then hold the stylus deflected again until the eight yellow flash sequence has ended. This action will clear all probe settings and return the probe to default settings. Following an LED check the RMP400 will then go back into Trigger Logic and will display “Switch-on method”.
3. Further configuration using Trigger Logic may be necessary to achieve the required probe settings.
1. Probe is in the Trigger Logic menu. Ensure that the stylus is no longer deflected.
2.
Whilst the status LEDs are flashing yellow to confirm that a master reset is required, release the stylus and then hold the stylus deflected again until the eight yellow flash sequence has ended.
System status LED
Deflect the stylus for 20 seconds until the status LEDs start to flash yellow eight times.
Previous settings have been cleared. The probe now has default settings.
Page 37
LED check
Draft 5 16/04/18
Switch-on method
3.
NOTE: RMP400 will continue to be partnered with either the RMI or RMI-Q following the activation of the master reset function, unless “Multiple probe mode” has been used.
Configure probe settings as required using Trigger Logic
Probe is now back in the Trigger Logic menu and will display “Switch-on method”.
Trigger Logic
4.9
Page 38
Trigger Logic
Draft 5 16/04/18
4.10

RMP400 – RMI partnership

System set-up is achieved using Trigger Logic and powering the RMI. Partnering is only required during initial system set-up. Further partnering will be required if either the RMP400 or RMI is changed, or if a system is reconfigured for multiple probes (multiple probe mode).
Acquisition mode
Acquisition mode off
Switch the RMI on
SIGNAL LED
Partnering will not be lost by reconfiguring the probe settings or when changing batteries, except where multiple probe mode is selected. Partnering can take place anywhere within the operating envelope.
In configuration mode, configure the probe settings as required until you reach the “Acquisition mode” menu, which defaults to “Acquisition mode off”.
NOTE: Please refer to the installation guide RMI radio machine interface (Renishaw part no. H-4113-8554) when partnering the RMP400.
RMI in acquisition mode
Deflect the stylus to select “Acquisition mode on”. Ensure this is done within 8 seconds of the RMI signal LED flashing green.
> 20 s
SIGNAL LED
New partner RMP acquired
The probe is now in standby and the system is ready for use.
Page 39

RMP400 – RMI-Q partnership

Draft 5 16/04/18
System set-up is achieved by using Trigger Logic and powering on the RMI-Q or applying ReniKey. Partnering is required during initial system set-up. Further partnering will be required if either the RMP400 or RMI-Q is changed.
Partnering will not be lost by reconfiguring the probe settings or changing the batteries. Partnering can take place anywhere within the operating envelope.
Acquisition mode
Acquisition mode off
Switch the RMI-Q on
An RMP400 that is partnered with the RMI-Q but then used with another system will need to be repartnered before being used again with the RMI-Q.
In configuration mode, configure the probe settings as required until you reach the “Acquisition mode” menu, which defaults to “Acquisition mode off”.
Trigger Logic
4.11
SYSTEM STATUS LED
RMI-Q in acquisition mode
with empty probe location
NOTE: Please refer to the installation guide RMI-Q radio machine interface (Renishaw part no. H-5687-8504) when partnering up to four RMPs.
or
Deflect the stylus to select “Acquisition mode on”. Ensure this is done within 60 seconds of the RMI-Q system status LED flashing green.
SYSTEM STATUS LED
New partner RMP acquired
Displayed for 5 seconds
SYSTEM STATUS LED
RMI-Q in acquisition mode
with full probe location
SYSTEM STATUS LED
Acquisition pending
SYSTEM STATUS LED
or
RMP cleared from location
> 20 s
The probe is now in standby and the system is ready for use.
Page 40

Operating mode

Draft 5 16/04/18
Trigger Logic
4.12
LEDs flashing
green
X/Y
Probe status LEDs
LED colour Probe status Graphic hint
Flashing green Probe seated in operating mode
Flashing red Probe triggered in operating mode
Flashing green and blue Probe seated in operating mode – low battery
Flashing red and blue Probe triggered in operating mode – low battery
LEDs flashing
red
LEDs flashing
red
Z
Constant red Battery dead
Flashing red or Flashing red and green or Sequence when batteries are inserted
Constant blue Probe damaged beyond use
NOTE: Due to the nature of lithium-thionyl chloride batteries, if a “low battery” LED warning is ignored, it is possible for the following sequence of events to occur:
1. When the probe is active, the batteries discharge until battery power becomes too low for the probe to operate correctly.
2. The probe stops functioning, but then reactivates as the batteries recover sufficiently to provide the probe with power.
Unsuitable battery
4. Again, the batteries discharge and the probe ceases to function.
5. Again, the batteries recover sufficiently to provide the probe with power, and the sequence repeats itself.
3. The probe begins to run through the LED review sequence (see page 4.1, “Reviewing the probe settings”).
Page 41
see Section 5,“Maintenance”
Draft 5 16/04/18

Maintenance

5.1

Maintenance

You may undertake the maintenance routines described in these instructions.
Further dismantling and repair of Renishaw equipment is a highly specialised operation, which must be carried out at an authorised Renishaw Service Centre.
Equipment requiring repair, overhaul or attention under warranty should be returned to your supplier.

Cleaning the probe

Wipe the window of the probe with a clean cloth to remove machining residue. This should be done on a regular basis to maintain optimum transmission.
CAUTION: The RMP400 has a glass window. Handle with care if broken to avoid injury.
Page 42

Changing the batteries

Draft 5 16/04/18
Maintenance
5.2
1
CAUTIONS:
Do not leave dead batteries in the probe.
When changing batteries, do not allow coolant or debris to enter the battery compartment.
When changing batteries, check that the battery polarity is correct.
Take care to avoid damaging the battery cassette gasket.
Only use specified batteries.
2
CAUTION: Please dispose of dead batteries in
accordance with local regulations. Never dispose of batteries in a fire.
Page 43
3
Draft 5 16/04/18
Ecocel Saft:
Tadiran: Xeno:
NOTES:
After removing the old batteries, wait more than 5 seconds before inserting the new batteries.
Do not mix new and used batteries or battery types, as this will result in reduced life and damage to the batteries.
Always ensure that the cassette gasket and mating surfaces are clean and free from dirt before reassembly.
If dead batteries are inadvertently inserted, the LEDs will remain a constant red.
Battery type
½ AA lithium-thionyl chloride (3.6 V) × 2
EB1426 LS 14250C LS 14250 SL-750 XL-050F
Dubilier: Maxell: Sanyo Tadiran:
Varta:
Maintenance
5.3
SB-AA02 ER3S CR 14250SE SL-350, SL-550, TL-4902, TL-5902, TL-2150, TL-5101 CR ½AA
4
5
Page 44
Maintenance
Draft 5 16/04/18
5.4
This page is intentionally left blank.
Page 45
see Section 6,“Fault-finding”
Draft 5 16/04/18

Fault-finding

Symptom Cause Action
6.1
Probe fails to power up (no LEDs illuminated or fails to indicate current probe settings).
Probe fails to switch on.
Dead batteries. Change batteries.
Unsuitable batteries. Fit suitable batteries.
Batteries inserted incorrectly. Check battery insertion/polarity.
Batteries removed for too short a time and probe has not reset.
Poor connection between battery cassette mating surfaces and contacts.
Dead batteries. Change batteries.
Batteries inserted incorrectly. Check battery insertion/polarity.
Probe out of range. Check position of RMI or RMI-Q,
No RMI or RMI-Q “start/stop” signal (“radio on” method only).
Remove batteries for a minimum of 5 seconds.
Remove any dirt and clean the contacts before reassembly.
(see operating envelope).
Check RMI or RMI-Q for green start LED.
Incorrect spin speed (“spin on” method only).
Incorrect switch on method configured.
Incorrect multiple probe mode setting configured.
RMP400 in hibernation mode (radio on method only).
Spin on is within 1 second of spin off. Check for 1 second dwell following
Check spin speed and duration.
Check configuration and alter as required.
Check configuration and alter as required.
Ensure probe is in range and wait up to 30 seconds, then resend switch-on signal.
Check position of RMI or RMI-Q, see operating envelope.
spin off.
Page 46
Symptom Cause Action
Draft 5 16/04/18
Fault-finding
6.2
Machine stops unexpectedly during a probing cycle.
Radio link failure/RMP400 out of range.
RMI or RMI-Q receiver/machine fault.
Dead batteries. Change batteries.
Excessive machine vibration causing false probe trigger.
Probe unable to find target surface.
Adjacent probe. Reconfigure adjacent probe to low
Stylus not given sufficient time to settle from a rapid deceleration.
Check interface/receiver and remove obstruction.
Refer to receiver/machine user’s guide.
Enable enhanced trigger filter.
Check that part is correctly positioned and that stylus has not broken.
power mode and reduce range of receiver.
Add a short dwell before the probing move (length of dwell will depend on stylus length and rate of deceleration). Maximum dwell is one second.
Probe crashes.
Probe permanently triggered.
Workpiece obstructing probe path. Review probing software.
Probe length offset missing Review probing software.
In cases where there is more than one probe on a machine, incorrect probe activated.
Probe orientation has changed – i.e. from horizontal to vertical.
New stylus has been fitted. Turn probe off and on again.
Probe was switched on when stylus was deflected.
Probe has not settled before a trigger move occurs following a rotation or rapid move (“Auto­reset” mode only).
Probe has collided with an object during a rotation or rapid move (“Auto-reset” mode only).
Review interface wiring or part program.
Select probe “Auto-reset” function.
Turn probe off and on again. Ensure stylus is seated during switch on.
Turn probe off and on again, and increase the dwell from 0.2 to
0.5 second dwell before probing move.
Turn probe off and on again.
Page 47
Symptom Cause Action
Draft 5 16/04/18
Poor probe repeatability and/or accuracy.
Debris on part or stylus. Clean part and stylus.
Poor tool change repeatability. Redatum probe after each tool
change.
Loose probe mounting on shank or loose stylus.
Excessive machine vibration. Enable enhanced trigger filter.
Calibration out of date and/or incorrect offsets.
Calibration and probing speeds not the same.
Calibration feature has moved. Correct the position.
Measurement occurs as stylus leaves surface.
Check and tighten as appropriate.
Eliminate vibrations.
Review probing software.
Review probing software.
Review probing software.
Fault-finding
6.3
RMP400 status LEDs do not correspond to RMI or RMI-Q status LEDs.
Measurement occurs within the machine’s acceleration and deceleration zone.
Probing speed too high or too slow.
Temperature variation causes machine and workpiece movement.
Machine tool faulty. Perform health checks on
Radio link failure – RMP400 out of RMI or RMI-Q range.
RMP400 has been enclosed/ shielded by metal.
RMP400 and RMI or RMI-Q are not partnered.
Review probing software and probe filter settings.
Perform simple repeatability trials at various speeds.
Minimise temperature changes.
machine tool.
Check position of RMI or RMI-Q, see operating envelope.
Remove from obstruction.
Partner RMP400 and RMI or RMI-Q.
Page 48
Symptom Cause Action
Draft 5 16/04/18
Fault-finding
6.4
RMI or RMI-Q error LED lit during probing cycle.
RMI or RMI-Q low battery LED lit.
Reduced range.
Probe not switched on or probe timed out.
Probe out of range. Check position of RMI or RMI-Q,
Dead batteries. Change batteries.
RMP400 and RMI or RMI-Q are not partnered.
Probe selection error. Verify that one RMP is working
“Fast” turn-on error. Ensure that all RMPs are ‘Q’
Low batteries. Change batteries soon.
Local radio interference. Identify and remove.
Change setting. Review switch-off method.
see operating envelope.
Partner RMP400 with RMI or RMI-Q.
and is correctly selected.
marked probes, or change the RMI-Q turn-on time to “standard”.
Probe fails to switch off.
Probe goes into Trigger Logic™ configuration mode and cannot be reset.
Probe status LED shows a constant blue
Incorrect “switch-off” method configured.
No RMI or RMI-Q “start/stop” signal (“radio on” method only).
Probe in timeout mode and placed in tool magazine and being triggered by movement.
Incorrect spin speed (spin switch on only).
Spin off is within 1 second of a spin on
Probe was triggered when batteries were inserted.
Probe damaged beyond use. Return the probe to your nearest
Check configuration and alter as required.
Check RMI or RMI-Q for green start LED.
Use shorter timeout setting or use different switch-off method.
Check spin speed.
Check for a 1 second dwell following a spin on.
Do not touch the stylus or stylus mounting face during battery insertion.
Renishaw supplier for repair/ replacement.
Page 49
see Section 7,“Parts list”
Draft 5 16/04/18

Parts list

Item Part number Description
RMP400 A-6570-0001 RMP400 probe with batteries, tool kit and quick-start guide
Battery P-BT03-0007 ½AA battery – lithium-thionyl chloride (pack of two).
7.1
(factory-set to radio on/radio off).
Stylus A-5003-7306
Stylus A-5003-6510 100.0 mm (3.94 in) long carbon fibre stylus with Ø6.0 mm
Stylus A-5003-6511 150.0 mm (5.91 in) long carbon fibre stylus with Ø6.0 mm
Stylus A-5003-6512 200.0 mm (7.88 in) long carbon fibre stylus with Ø6.0 mm
Tool kit A-4071-0060 Probe tool kit comprising Ø1.98 mm stylus tool,
Battery cassette A-4071-0031 Battery cassette assembly.
Battery gasket A-4038-0301 Battery cap gasket kit.
RMI A-4113-0050
RMI-Q A-5687-0050 RMI-Q (side exit) with 15 m (49.2 ft) cable, tool kit and
Mounting bracket A-2033-0830 Mounting bracket with fixing screws, washers and nuts.
Styli tool M-5000-3707 Tool for tightening/releasing styli.
50.0 mm (1.97 in) long carbon fibre stylus with Ø6.0 mm (0.24 in) ball.
(0.24 in) ball.
(0.24 in) ball.
(0.24 in) ball.
2.00 mm AF hexagon key and shank grub screw (× 6).
RMI (side exit) with 15 m (49.2 ft) cable, tool kit and user’s guide.
quick-start guide.
Publications. These can be downloaded from our web site at www.renishaw.com.
RMP400 H-6570-8500 Quick-start guide: for rapid set-up of the RMP400 probe.
RMI QSG A-4113-8550 Quick-start guide: for rapid set-up of the RMI.
RMI IG H-4113-8554 Installation guide: for set-up of the RMI.
RMI-Q QSG H-5687-8500 Quick-start guide: for rapid set-up of the RMI-Q.
RMI-Q IG H-5687-8504 Installation guide: for set-up of the RMI-Q.
Styli H-1000-3200 Technical specifications guide: Styli and accessories – or visit
our Web shop at www.renishaw.com/shop.
Probe software H-2000-2298 Data sheet: Probe software for machine tools – programs and
features.
Taper shanks H-2000-2011 Data sheet: taper shanks for machine tool probes.
Page 50
Renishaw plc
Draft 5 16/04/18
New Mills, Wotton-under-Edge Gloucestershire, GL12 8JR United Kingdom
T +44 (0)1453 524524 F +44 (0)1453 524901 E uk@renishaw.com
www.renishaw.com
For worldwide contact details, visit
www.renishaw.com/contact
© 2018 Renishaw plc
Issued: ??.????
*H-6570-8501-01*
Part no. H-6570-8501-01-A
Loading...