Renishaw PSU3 Installation And User Manual

Installation and user’s guide
H-2000-5057-03-A
PSU3 power supply unit
English Français Deutsch Italiano
Français
Italiano
English
Manuel d’installation et d’utilisation
PSU3 Netzteil
Guida d’installazione ed uso
Alimentatore PSU3
Installations- und Benutzerhandbuch
2-0
4-0
3-0
Deutsch
Installation and user’s guide
PSU3 power supply unit
1-0
Bloc d’alimenation PSU3
This page is intentionally left blank
1
English
Installation and user’s guide
PSU3 power supply unit
© 2001 - 2006 Renishaw. All rights reserved.
This document may not be copied or reproduced in whole or in part, or transferred to any other media or language, by any means, without the prior written permission of Renishaw.
The publication of material within this document does not imply freedom from the patent rights of Renishaw plc.
Disclaimer
Considerable effort has been made to ensure that the contents of this document are free from inaccuracies and omissions. However, Renishaw makes no warranties with respect to the contents of this document and specifically disclaims any implied warranties. Renishaw reserves the right to make changes to this document and to the product described herein without obligation to notify any person of such changes.
Trademarks
RENISHAW® and the probe emblem used in the
RENISHAW logo are registered trademarks of Renishaw plc in the UK and other countries.
apply innovation is a trademark of Renishaw plc.
All other brand names and product names used in this document are trade names, service marks, trademarks, or registered trademarks of their respective owners.
Renishaw part no: H-2000-5057-03-A Issued: 10. 2006
1-1
PRELIMINARY
Safety ...............................................................1-3
Warranty ........................................................... 1-3
Changes to equipment .....................................1-3
Caring for the PSU3 ......................................... 1-3
Probe IP rating ................................................. 1-3
Environment .....................................................1-3
GENERAL
Introduction .......................................................1-4
PSU3 connections ............................................ 1-5
PSU3 dimensions ............................................. 1-6
FAULT FINDING ...............................................
1-6
PARTS LIST .....................................................
1-6
Contents
1-2
EC DECLARATION OF CONFORMITY
Renishaw plc declares that the product:
Name Description PSU3 24 Vdc power supply unit
has been manufactured in conformity with the following standards:
BS EN 61326: 1998/ Electrical equipment for measurement, A1:1998/A2:2001 control and laboratory use - EMC
requirements. Immunity to annex A ­industrial locations. Emissions to class A (non-domestic) limits.
BS EN 61010-1: 2001 Safety requirements for electrical
equipment for measurement, control and laboratory use. Part 1: General requirements.
and that it complies with the requirements of the following directives (as amended):
89/336/EEC Electromagnetic compatibility 73/23/EEC Low voltage
The above information is summarised from the full EC Declaration of Conformity. A copy is available from Renishaw on request.
FCC
Information to user (FCC Section 15.105)
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with this installation guide, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case you will be required to correct the interference at your own expense.
Information to user (FCC Section 15.21)
The user is cautioned that any changes or modifications not ex­pressly approved by Renishaw plc or authorised representative could void the user’s authority to operate the equipment.
Preliminary information
1-3
Safety
The PSU3 has been designed with all connections and information indicators on the outside of the enclosure.
There are no serviceable parts inside. In case of failure the unit should be returned to your supplier for repair.
Warranty
Equipment requiring attention under warranty must be returned to your supplier. No claims will be considered where Renishaw equipment has been misused, or repairs or adjustments have been attempted by unauthorised persons.
Changes to equipment
Renishaw reserves the right to change specifications without notice.
Care of the PSU3
No routine maintenance is required.
Probe IP rating
IP30
Environment
The PSU3 is specified for the following temperature ranges:
Storage: –10 °C to 70 °C (14 °F to 158 °F)
Operation:
5 °C to 50 °C (41 °F to 122 °F)
Preliminary information
!
1-4
General
Introduction
The PSU3 provides a +24 Vdc supply for Renishaw interface units.
The PSU3 power supply unit may be used with several types of Renishaw interface. Please see individual product handbooks for connection details. More than one interface may be driven from one power supply as long as the maximum output current is not exceeded (24 Vdc ±2 Vdc 0.5 A maximum).
Front
Rear
When then green LED is lit, the power supply is operating.
Output
24 Vdc ±2 Vdc 0.5 A maximum.
The PSU3 is over voltage and over current protected.
Power is provided via an IEC mains connector.
This is the mains isolation device and must
remain accessible at all times.
Input
100 Vac to 240 Vac (-15% to +10%)
50 Hz to 60 Hz, 25 W maximum.
Mains switch
1-5
PSU3 connections
Optional cable tie
A cable tie and an extra long screw are provided
if the low voltage wires
need anchoring.
To use, remove the
existing left hand
rear panel screw and
replace with longer
screw and tie.
Non-rewirable
machine socket with cable
(Not supplied)
Fit when required by
local regulations.
OR
Rewirable
mains socket
(Supplied with kit)
Use three core insulated
and sheathed cable.
Each core
Ø0.5 mm to Ø1.0 mm.
(Ø0.019 in to Ø0.039 in).
Output terminal block
4 BA screw (Ø3.6 mm dia.).
Maximum fork width 7.5 mm (0.30 in).
+24 V
0 V
Mains switch
Mains plug
Green Red Black
Three low voltage wires
16/0.2 mm insulated.
GND
+ 24 V
0 V
Interface unit
Cable colour code
Europe USA N Blue Black E Green/yellow Green L Brown White
Warning
This equipment must be supplied from a mains circuit in which overcurrent protection devices of 10 A or less are fitted.
1-6
140 mm (5.5 in)
3.5 mm (0.14 in)
183 mm (7.20 in)
Self adhesive foot
12,7 mm (0.5 in) diameter
44 mm (1.73 in)
PSU3 dimensions, fault finding and parts list
Fault finding
Power LED not illuminated
Check mains switch at rear of unit is switched on.
Check that mains supply cable is plugged in securely and that the mains source is functioning.
Power LED dim or flickering
Check all connections are tight.
Check that the load on the power supply does not exceed 0.5 A.
In the event of continuing malfunction please return the PSU3 to your supplier for repair.
PARTS LIST - Please quote the Part no. when ordering equipment.
PSU3 power supply unit Part no. A-2019-0018