red lion PAX Series Series Manual

1
! ACCEPTS STANDARD 3-WIRE 100
RTD SENSORS (ALPHA =
0.00385 or ALPHA = 0.00392)
! CONFORMS TO ITS-90 STANDARDS
! SELECTABLE °F OR °C WITH 0.1 OR 1 DEGREE DISPLAY
RESOLUTION
ACCURACY AND RELIABILITY
! FULL 4-DIGIT, HIGH VISIBILITY, 0.56" (14.2 mm) HIGH RED LED
DISPLAY
! PROGRAMMABLE TEMPERATURE OFFSET
! PROGRAMMABLE DIGITAL FILTERING
! PEAK/VALLEY (HI/LO READING) MEMORY
! NEMA 4X/IP65 SEALED FRONT BEZEL
! CUSTOM UNITS OVERLAY WITH BACKLIGHT
GENERAL DESCRIPTION
The Pax Lite RTD Meter accepts standard RTD inputs and precisely linearizes them into temperature readings. A full 4-digit display accommodates a wide range of temperature inputs. State-of-the-art digital circuitry virtually eliminates errors due to drift.
The meter features a readout choice of either Fahrenheit or Celsius with 0.1 or 1 degree resolution. English Style display prompts and front panel buttons aid the operator through set-up and operation. Programmable digital filtering enhances the stability of the reading. All set-up data is stored in EEPROM, which will hold data for a minimum of 10 years without power.
The meter provides a Peak (HI) and Valley (LO) reading memory with selectable capture delay time. The capture delay is used to prevent detection of false Peak or Valley readings that may occur during start-up or unusual process events. The Peak and Valley readings are stored at power-down to allow monitoring the process limits over any length of time (shifts, days, etc.).
The meter has several built-in diagnostic functions to alert operators of any malfunction. Extensive testing of noise interference mechanisms and full burn­in makes the meter extremely reliable in industrial environments. The front bezel meets NEMA 4X/IP65 requirements for wash down applications.
SAFETY SUMMARY
All safety related regulations, local codes and instructions that appear in the literature or on equipment must be observed to ensure personal safety and to prevent damage to either the instrument or equipment connected to it. If equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Do not use this unit to directly command motors, valves, or other actuators not equipped with safeguards. To do so, can be potentially harmful to persons or equipment in the event of a fault to the unit.
DEFINITION OF TERMS
INSTALLATION CATEGORY (overvoltage category) I:
Signal level, special equipment or parts of equipment, telecommunication, electronic, etc. with smaller transient overvoltages than Installation Category (overvoltage category) II.
INSTALLATION CATEGORY (overvoltage category) II:
Local level, appliances, portable equipment, etc. with smaller transient overvoltages than Installation Category (overvoltage category) III.
MODEL PAXLRT - PAX LITE RTD METER
CAUTION: Risk of Danger
Read complete instructions prior to
installation and operation of the unit.
CAUTION: Risk of electric shock.
Bulletin No. PAXLRT-B
Drawing No. LP0562
Released 11/08
Tel +1 (717) 767-6511
Fax +1 (717) 764-0839
www.redlion.net
DIMENSIONS In inches (mm)
Note: Recommended minimum clearance (behind the panel) for mounting clip installation is
2.1" (53.4) H x 5" (127) W.
C
US LISTEDUS
LISTED
U
L
R
IND. CONT. EQ.
51EB
2
ORDERING INFORMATION
Meter Part Numbers
Ordering Information . . . . . . . . . . . . . . . . . . . 2
General Meter Specifications. . . . . . . . . . . . . 3
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . 3
Installing the Meter . . . . . . . . . . . . . . . . . . . . 4
Wiring the Meter . . . . . . . . . . . . . . . . . . . . . . 4
Reviewing the Front Buttons and Display . . . 5
Programming the Meter . . . . . . . . . . . . . . . . 6
Calibrating the Meter. . . . . . . . . . . . . . . . . . . 7
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 7
TABLE OF CONTENTS
Accessories Part Numbers*
MODEL NO. DESCRIPTION PART NUMBERS
Accessories
TYPE
PAXLBK Units Label Kit Accessory PAXLBK30
*This meter is shipped with °F and °C overlay labels. The label kit is only
needed if another units label is desired.
PAXL
RT
0
0
RT - RTD Temperature Meter
3
GENERAL METER SPECIFICATIONS
1. DISPLAY: 4-digit, 0.56" (14.2 mm) high LED, minus sign displayed for negative temperatures.
Overrange/Underrange Input: Flashing “!"!"” or “#"#"
Overrange/Underrange Display: “$ $ $ $” or “&$ $ $ $
2. POWER: 85 to 250 VAC, 50/60 Hz, 6 VA Isolation: 2300 Vrms for 1 min. between input and supply (300 V working
voltage)
3. CONTROLS: Three front panel push buttons for meter set-up. Rear terminal input for disabling the front panel.
4. RESOLUTION: 0.1 or 1 degree
5. RANGE: Decimal Point Dependent
0.1° res: -199.9° to 850.0 °C (-199.9° to 999.9 °F); 1° res: -200° to 850 °C (-328° to 1562 °F)
6. OPEN/SHORTED RTD DETECTION: Display flashes: “!'()” or “*+,-
7. LEAD RESISTANCE EFFECT: 20 max., 2.5 °C/error for V exc. and common lead unbalance
8. ACCURACY: 0.3 °C, @ 23 °C and 30 min. warm-up
9. READING RATE: 2.5 readings/second
10. RESPONSE TIME: 2 seconds to settle for step input (increases with programmable digital filtering)
11. LOW FREQUENCY NOISE REJECTION: Normal Mode Rejection: 40 dB @ 50/60 Hz (may be improved by programmable digital filtering) Common Mode Rejection: 120 dB, DC to 50/60 Hz
12. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL Recognized Component, File # E179259, UL61010A-1, CSA C22.2 No. 61010-1
Recognized to U.S. and Canadian requirements under the Component Recognition Program of Underwriters Laboratories, Inc.
UL Listed, File # E137808, UL508, CSA C22.2 No. 14-M95
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards Type 4X Enclosure rating (Face only), UL50
IECEE CB Scheme Test Certificate # US/8843A/UL
CB Scheme Test Report # 04ME11209-20041018
Issued by Underwriters Laboratories, Inc.
IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment
for measurement, control, and laboratory use, Part 1.
IP65 Enclosure rating (Face only), IEC 529
ELECTROMAGNETIC COMPATIBILITY
Emissions and Immunity to EN 61326: Electrical Equipment for
Measurement, Control and Laboratory use.
Note:
1. Criterion A: Normal operation within specified limits.
13. ENVIRONMENTAL CONDITIONS: Operating Temperature Range: 0 to 50 °C Storage Temperature Range: -40 to 80 °C Operating and Storage Humidity: 85% max (non-condensing) from 0 to
50 °C
Span Drift: 50 ppm/ °C Zero Drift: 0.001 °C/°C Vibration According to IEC 68-2-6: Operational 5 to 150 Hz, in X, Y, Z
direction for 1.5 hours, 2g’s.
Shock According to IEC 68-2-27: Operational 30g’s (10g relay), 11 msec in
3 directions.
Altitude: Up to 2000 meters.
14. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 indoor use. One piece bezel/case. Flame resistant. Panel gasket and mounting clip included.
15. CONNECTIONS: High compression cage-clamp terminal block Wire Strip Length: 0.3" (7.5 mm) Wire Gage: 30-14 AWG copper wire Torque: 4.5 inch-lbs (0.51 N-m) max.
16. WEIGHT: 0.65 lbs. (0.24 Kg)
Class BEN 55011Emissions
Emissions:
0.5 cycle
Criterion AEN 61000-4-11Voltage dip/interruptions
3 V/rms
Criterion AEN 61000-4-6RF conducted interference
1 kV signal
1 kV L-L,
Criterion AEN 61000-4-5Surge
2 kV signal
2 kV power
Criterion AEN 61000-4-4Fast transients (burst)
2 kV L&N-E power
10 V/m
Criterion AEN 61000-4-3Electromagnetic RF fields
8 kV air discharge
4 kV contact discharge
Criterion AEN 61000-4-2Electrostatic discharge
Immunity:
ACCESSORIES
UNITS LABEL KIT (PAXLBK)
Each meter has a units indicator with backlighting that can be customized using
the Units Label Kit. The backlight is controlled in the programming.
Each meter is shipped with °F and °C overlay labels which can be installed into
the meter’s bezel display assembly.
Loading...
+ 5 hidden pages