Raytek MX6 Operating Manual

How to use this Manual

To get
the full functionality of this manual we recommend to use Adobe®Acrobat Reader®5 or higher. It is availabe at:
http://www.adobe.com
Once you have open this manual in Acrobat Reader choose the tab “Book­mark” at the left to see the Table of Contents. These headlines are linked with the corresponding chapters of the manual. Just click on a headline to see the chapter in the main window. Sometimes you’ll find phrases in the text marked red. These will guide you to additional information. To go back to
the manual use this arrow button in the upper bar of Acrobat Reader.
1
3
2

Basic Mode

This chapter is dedicated to explore the unit and the software and to get fami­lar with its basic functions.
It helps to get first results fast and easy.
It explains the software installation and how to use the unit with factory defaults. Please take a bit time and follow up carefully.
Once you feel trained enough go to the “Expert Mode” to find out what else this unit can do for you.
5

Before Starting

4

Requirements

To use the unit and the software you need a Windows compatible PC with a minimal config­uration:
• 200 Mhz processor (400 Mhz recommended)
• CD/DVD drive
• USB 1.1 - port
• 15 MByte free diskspace on the hard drive
The software works with Windows
®
98, ME, XP and 2000.
Note! With a fresh set of batteries, the unit can store photographs for up to 2,000 hours. However, we recommend that you download photographs soon after you have taken them.
To avoid the loss of photos you should change the batteries within two minutes after the unit has shut down completly.
6
7

Signs and Icons in the Manual

Introduction

We hope that you enjoy using your photographic infrared thermometer! It measures the amount of infrared energy emit­ted by a target object, and calculates the temperature of that object´s surface. In addition, it takes a photo of the measured spot highlight­ed by a bright laser circle. In a few easy steps you can create reports that incorporate the photo of the target and also the temperature, location, date/time, and a description - all this can be customized by using the software delivered with the unit.
Your unit includes:
-A digital camera synchronized with the temperature measurement
-True Spot laser Sighting
- Nameable locations
- Adjustable emissivity
- High/Low alarm
- Data logger
- ...and a lot more!

Features

Delivery Content

If you see this icon, use the unit to follow the steps described.
If you see this icon, you should have started your PC and the companion Software delivered with your unit.
If you see this icon, the unit is connected to the PC.
Software descriptions appear in a grey textbox
This icon shows that you can adjust values with these buttons on the unit.
To store these values press the ENTER but­ton on the unit.
This icon shows that there are additional options related to the DIP-switch settings.
The unit • Thermocouple Type K probe • Win­dows-based software on CD • USB cable • Two AA batteries • Manual on CD • Quick start
8
9

Configuration of the PC

Installing the software

Hardware Setup

Connect the unit to the PC via USB.
New hardware will be found. You have to install three different device drivers - two for USB ports and one for the USB camera. A device driver is software needed by your PC to communicate with the unit. You will be asked three times to install a driver.
Note! During driver installation, a message
window could appear, indicating that a particu­lar driver is not supported or authorized by Microsoft. Please ignore this message and con­tinue with the installation.
The Windows Hardware Assistant will guide you through the installation process. If you are asked where to look for the drivers, choose CD-ROM. In most cases Windows will find the drivers automatically. It could happen, that Windows asks for the file STV680u.dll. This file is located in the drivers folder on the CD.

Software Setup

1. Insert the CD into your CD drive.
2. Double click on setup.exe.
3. Follow the instructions, shown on the screen.
4. Launch the software.
The software is now ready to use.
Plese read the following information first and configure language, time, and date if necessary.
This configuration can be repeated at any time.
Software Language
To choose one of the available languages of the software, go to “Setup”.
Click on “Language” in the Pull Down menu.
Then make your selection.
11

Configure Date and Time

10
Configure Date and Time
Set Date and Time of the unit
To get the right date and time stamp in your picture, proceed with the following steps:
1. Choose “Setup” in the software’s menu bar.
2. Go to “Device Alarm and Display”
3. Look for “Date and Time” and verify that the correct values are shown (these values depend on your computer system settings)
4. Enable the checkbox to the right of the val­ues.
5. Click on “OK”.
6. Disconnect the unit from the PC.
13
Basic Mode
12
Simply turn the lens focus ring depending on the distance of your target.

How to focus the camera

Between
0.2 m (8 in) and
0.3 m (12 in.) adjust the lens to the flower symbol.
Between
0.5 m (19 in) and infinity adjust the lens to the mountain symbol.
This is a simple aid to understanding the camera’s field of view. The opening angle of the lens is approximately 40°. Rotate the cap up to a horizontal position and look over it, as shown below. The embossed triangle will show you the approximate width of the photo.

The view finder guide

Laser ON and OFF

laser ON symbol
The laser circle shows the spot size that includes the measured target. To turn the laser On or Off, press the LASER button when the trigger is pulled. A laser symbol appears when the laser is on. The laser automatically turns off if you release the trigger.
CAUTION!
Do not stare into beam!
Avoid indirect exposure via reflective materials!
Laser Warning Label
Basic Mode
40°
Between 0.25 m (10 in.) and 0.6 m (23 in.) adjust the lens in the middle of both symbols
15
Basic Mode
14
Basic Mode
1. To switch on the unit, pull the trigger.
2. Press the “Enter” button to activate the
camera while the display is active.
3. First the word LOG” flashes and then the
camera icon appears.
The unit is ready to use now! It is preset to take 26 pictures!
4. Pull the trigger and hold it. The laser circle
shows where you are measuring.
5. Aim at the target. Be sure that the laser
circle is inside the target.
6. Gently release the trigger to record the
photo and the temperature. Successful recording of picture and temperature is indi­cated by two short beeps and a green
shining LED above the display. The next
location will be shown on the display.
Caution:
If you hear a longer beep and the LED above the display is shining red, look at the display. If you see “Use Flash!” and a flash symbol, repeat your last measurement. A small green LED above the camera icon signals: “Flash has charged”. The flash will now fire automatically.
7. For the next measurement repeat points
5 and 6.
8. Once you have taken all your photos,
connect the unit to the PC via USB.
9. Start the software on your PC. After a few seconds, the data download begins. Click on “OK” if a message about logger file and unit
configuration incompatibility appears. Click
on “Yes” if you are asked to download all images
10. See your pictures and data as a thumbnail
collection.
11. If this doesn’t happen automatically, toggle to this button.
16
17
Basic Mode
Basic Mode
Once you have downloaded all data from the unit to the PC, save the data as library.lgg file.
To edit a picture, click on it once with the left mouse button. It will be enlarged.
Now you are able to add discriptions to the photo or to insert a drawing of the laser circle. This may be necessary, if the ambient light was too bright while taking the photo and the laser circle is hard to see.
Right clicking on the photo will open a menu, where you can choose what you want to do.

Additional software features

If you choose “Paint circle” the following window appears.
With the distance slider you could adjust the size of the laser circle drawing corresponding to the original size of the real laser circle.
NOTE: This feature will only work properly, if you have adjusted once the circle drawing to the camera with the following procedure:
Ta ke a photo - enlarge it in the software - draw a laser circle - adjust the size with the distance slider as shown above - adjust the place of the
circle by using “Adj.” and then these sliders.
The drawing will be stored within the photo. The same happens with the description function.
With the “Copy image” function you can create a report easily by pasting the photo into a pro­gramme of your choice.
“Print Image” starts the printers dialog of your standard printer. Now you can print out a single photo.
19
18

Expert Mode

Expert Mode
This chapter is dedicated to the advanced user and based on the
“Basic Mode” part of this
manual.
We recommend to going through this part step by step to get the most benefit out of the unit.
For additional information contact the manufacturer or refer to its website. (See back page of the cover of the printed manual.)
In the “Basic Mode” chapter you have seen how easy it is to use this unit and to store photos and data. A special file has been created automatically by the unit, based on the factory defaults. This file is called “Logger” and has the extension “lgg”.
Under special circumstances it may happen, that the factory default settings don’t match your needs. In that case, it is possible to customize the factory default logger file.
You may wish, for instance, to name the loca­tions where you want to measure or to add dif­ferent emissivity values or special materials. Additionally it may necessary to add more than just temperature, location, date and time infor­mation to the photos.
All this is possible and explained in the next few pages
21
Expert Mode
20
Expert Mode

How to create a logger file?

Note! The following operation will
change the current settings in the unit!
Configure an inspection table using the menu SETUP/DEVICE LOGGER. The following screen display will appear.
Continue by setting the desired number of the inspection points to be measured in the table at
the lower right. In our example it is 5. Choose the desired image resolution. “High” means a resolution of 640x480 pixels, “Low” is 320x240 pixels. Please note! In both modes you could confi­guere up tp 100 positions. If you have chosen “High” the maximum number of photos is 26.
In the “Low” resolution mode, you can take more than 100 photos. The camera memory­can store up to 106 photos. This could be hel­pful, if you want to repeat a photo at a previous position.
Next, enter the name of the location and insert the related material from the material table by clicking here. For a known emissivity enter the value in the column EMS. In the column “Image” you could enable or disable taking a photo by checking the appropriate box
The material names are stored in a material table file. This can be customized. See chapter “How to create a material table?”
Save this logger file with a new name by clicking on the “Save as” button. With a click on “OK” transfer the table to the unit.
Now go to the Logger view by clicking here and open the file you have just saved
23
Expert Mode
22
Expert Mode
Examples
Disconnect the unit from the PC and begin your inspection.
Once the inspection has been done, transfer data and photos to the PC. Connect the unit via USB to the computer, while the software is run­ning. The data transfer begins automatically when the message “Connected” appears.
Now you can set up which data fields should be inserted in the photos after you have taken them. All columns you choose here will be inclu­ded in the final photo.You can’t delete any infor­mation from the photo itself. To open the sub-
menu simply right click into the table
With the following dialog choose and order the columns.
Note! The more data you insert in the photo the more these data will hide the picture. Especially if your working in the low resolution mode, we recommend to choose only a few columns.
See the examples!
Many columns
chosen
Fewer columns
chosen
Ta ke your photos and data as described above and save them by pressing disc buttoninto the same file as before. The number of inspections is shown in the bot-
tom line of the window To see changes of the temperature over time, you can create a routing graph.
This feature is very useful to monitor measure­ment points over time and identify problematic trends.
For instance, in electrical panels, it is difficult to find out that something is outside the normal temperature range, especially if the panel has live voltage. If you monitor the panel over time, you find out, where the problems are, before they become dangerous.
Go to
and then open the logger file you have just created.
25
Expert Mode
24
Expert Mode
The photos and data appear in a thumbnail pre­view. To see the table click here.
With a right click into the table you open a sub menu.
By choosing “Columns” you can add or delete columns to show all the data you have saved.
Save the data. This will delete the unit’s memory and the unit is now ready for the next inspection.
27
Expert mode
26
The result.will look as follows:
The routing graph
Expert Mode
The following dialog appears:
Now select in the curves menu up to five objects measured and set up their appearance in the graph.
Then choose the time frame of the measure­ment.
29
28
Expert Mode

How to create or edit a material table?

This function enables you to easily create customized material / emissivity tables that meet your requirements.You can then transfer the table, online, to your thermometer. Proceed as follows:
Choose the item DEVICE MATERIAL TABLE from the menu SETUP. The following screen display appears.

Creating a New Material Table

The cursor is located in the first line under the entry MATERIAL. Enter the name of the first material here. In the same line enter the asso­ciated emissivity under EMS. To access the other fields, use the keyboard or the mouse and fill in the complete table. Now save the table by clicking the button SAVE TO FILE. In the subse­quent window (see next page) enter the file name; the file suffix "mat" is automatically appended.
Save the file in a directory on the drive of your choice. We recommend the directory name "Material."

Default Material Table

For your convenience we have provided a selec­tion of 30 common materials. To access these, simply click FACTORY DEFAULT with the pri­mary mouse button or choose a material table in the language you need by selecting LOAD FROM FILE.
A complete table will appear as shown above. By clicking OK you transfer this table to the unit.

Editing an Existing Material Table

Just as easily you can edit a material table. To call up the table, choose the function LOAD FROM FILE in the editing window.
By selecting and overwriting, you can modify the list as you prefer and then save it again.
31
Expert Mode
30
Expert Mode

Temperature Monitoring mode

Using the software, the unit can monitor temperature and take photos automatically. This happens when the temperature exceeds the preset values for high or low alarm.
Proceed as follows:
Install the unit on a tripod.
Adjust the measurement direction with the help of the live video window of the software. Then go to “Setup” menu in the software and choose “Device Alarm and Display...”
Choose the alarm values for which you want the unit to take a photo when exceeded. If necessary, choose the material of the target or emissivity value.
Then click “OK”. Now go to “Recording setup” in the setup menu.Mark when the snapshot should be taken
- at high alarm or low alarm.
The “min. interval(s)” setting means that the unit only checks after this interval for the alarm val­ues exceeded.
A photo will only be taken if an alarm is newly exceeded. The snapshots will be taken in the measure­ment and recording mode. But you can only save the data for later docu­mentation in the recording mode. The “Max. number of images” is limited only by hard disk space. The file size of a low resolution photo is approxomate 30 kB and for high reso­lution 60 kB.
The live video window appears in both modes. Please do not close it. Otherwise the snapshot mode will not work.
To resize the live video window, simple click the right mouse button and choose another size.
When you press the recording or measuring button again to stop the process, you will be prompted for a file in which to save the temperature data.
33
32
Expert Mode
After taking the snapshot, you can see the photos by clicking on the square markers in the graph.
34
35

Appendix A - Unit functions

Appendix A - Unit functions

Selecting a function

1. To switch on the unit pull and release
the trigger.
2. While the display is active, push the button
of the desired function.
3. Many of the functions are adjustable by
using the Up and Down keys. For storing, press ENTER.
1
2
3

The DIP switches

Some functions of the unit be changed by using the DIP switches. These switches are located in the Battery compartement of the unit.
ON
Sets the unit into con­tinuous measuring mode
Switches between degrees Celsius and Fahrenheit
Switches acoustic signals On or Off
Switches the backlight of the display On or Off
Sets the unit’s values to default, if On
If On, the function buttons are blocked
If On, the laser will flash when temperature val­ues exceed alarm
If On, photoflash is activated
Auto = automatic flash use All = every photo is taken with flash
Time or Date will be displayed
Thermocouple settings ­refer to the corresponding chapter
ON
Lock
°C/°F
Buzzer
Backlight
Set Default
Ltd. Access
Laserflash
Photoflash
Auto/All
Time/Date
NTC/TC
TC-J/TC-K
37
Appendix A - Unit functions
36
Appendix A - Unit functions

The Data button

Press Data once to activate the log mode of the thermometer and the camera. With the right logger file configuration, you can store temperature values and photos.
The configuration of the logger file is performed by the companion software.
By pressing the Data button twice, one of the
following four displays appears.
“RCL” is displayed for the recall mode.
A photo and a tempera­ture value have been stored at this Logger posi­tion.
Only a temperature value has been stored at this logger position.
Camera is active at this logger posi­tion. Nothing has been stored.
Camera is inactive at this logger position. Nothing has been stored.

The Display button

The last ten measurements are shown in the bar graph on the unit’s display. Auto Range of the bar graph is automatically defined by the measured maximum and minimum value. Manual Range (Man Range) is user defined.
CYCLE allows the adjustment of the display interval of the bar graph.
Press Display once to see how many photos are already taken.
This sets up the BEGIN value for the graphic display of the bar graph. The graphic display shows the temperature as a picture.
This sets up the END value for the graphic display of the bar graph.
39
Appendix A - Unit functions
38
Appendix A - Unit functions

The Setup button - HiAL, LoAL, Time...

High alarm (HiAl) generates an audible and visual (flashing LED and laser) alarm if the temperature is above the setpoint.
Low alarm (LoAl) generates an audible and visual (flashing LED and laser) alarm if the temperature is below the setpoint.
This function is used with a selected emissivity to add or subtract an offset value (±10°C/±18°F) to the temperature value.
The OFFSET feature allows the temperature values for several units to be matched, correcting for the allowed temperature tolerance difference between units. The OFFSET function can also be used to increase the accuracy for a narrow temperature range.
Change the time using the Up and Down keys.Then press ENTER for each time segment to activate this time setting. The segment flashes whilst being set. The time is stored within the data logger.
Change the date using the Up and Down keys.Then press ENTER for each date seg­ment to activate this date setting. The seg­ment flashes while being set. The date is stored within the data logger.

...Date, Offset, Clear Images

This feature deletes all photos from the unit’s memory. “Clear Image Press Enter” is shown as a run-
ning string in the bottom line of the display BE CAREFUL if you use this function. The pictures cannot be restored.
Back to the
Setup Button
The configuration of the alarm values can also be performed by the companion software.
41
Appendix A - Unit functions
40
Appendix A - Unit functions

The Mode button - MINMAX, Material, MAX...

...MIN, DIF, AVG,

To activate the MAX mode, press MODE until the MAX symbol appears . The measured maximum temperature is displayed as long as the trigger is pulled or locked on. The real time temperature is shown in the lower part of the dis­play (NORM).
To show which specific material has been choosen for the specific location, press MODE until the name appears in the bottom line. If “Free” appears, you can change the emissivity settings by using the Emiss. button, when not in LOG mode. See chapter The
Emiss. button!
To show the minimum and maxi­mum temperature values during a measurement at the bottom of the display, press MODE until the two values appear.
To activate the MIN mode, press MODE until the MIN symbol appears. The measured minimum temperature is displayed as long as the trigger is pulled or locked on.The real time temperature is shown in the lower part of the dis­play (NORM)
To activate the DIF mode, press MODE until the DIF symbol appears. The difference between the measured max and min tem­peratures is displayed as long as the trigger is pulled or locked on. The real time temperature is shown in the lower part of the dis­play (NORM).
To activate the AVG mode, press MODE until the AVG symbol appears. The average value of measured temperatures is diplayed as long as the trigger is pulled or locked on. The real time temperature is shown in the lower part of the display (NORM).
Next page
43
Appendix A - Unit functions
42
Appendix A - Unit functions
s
ON
Open the battery compart­ment and set the switches ON or Off according to the desired probe type.
NTC - thermistor TC - thermocouple Thermocouple type J Thermocouple type K
Connect the probe to the input. The input is located at the bot­tom of the handle behind the rubber cover.
Press MODE, until the desired probe symbol (TC-K shown here) appears. The probe temperature is shown in the lower part of the display. The real time infrared temperature is shown in the main display.

The Mode button - Thermocouple Settings

back to
Mode
button

The Emiss. button

To choose the emissivity of a mate­rial, press EMISS.
The display shows a material name, an emissivity value, and the calculated temperature value. To choose another material, use the Up and Down keys. Press ENTER to activate this set­ting.
To adjust the emissivity value, press EMISS., when not in LOG mode. Use the Up and Down keys to select “Free“ (“F
ree” will have a flashing underline). Press EMISS again. “Free” is not underlined, and the emissivity icon flashes. Use the Up and Down keys to adjust. Press ENTER (D) to acti­vate this setting.
ON
Lock
°C/°F
Buzzer
Backlight
Set Default
Ltd. Acces
Laserflash
Photoflash
Auto/All
Time/Date
NTC/TC
TC-J/TC-K
45

Appendix B

44
Appendix B
To adjust the unit’s emissivity value for a material with unknown emissivity, plug in the probe.
Pull the unit’s trigger. Place the measuring tip of the probe on the area to be measured. Wait for the reading to stabilize. Note the indicated probe temperature reading. Release the trigger. Pull the trigger again. Measure the same area using infrared measurement. Press the emissivity button. Use the Up and Down keys to select the material name “Free” which will be shown in the display. Press the emissivity button again until the emissivity sign flashes. Use the arrow keys to change the emissivity value until the temperature matches the probe’s reading.
The amount of infrared energy radiated by an object depends on its emissivity and its temperature.
The emissivity depends on the material and its surface characteristics.
Note: For more accurate readings, adjust the emissivity value for the type of material being measured.

Emissivity explained

Unknown Emissivity Value

Reflected energy
Emitted energy
Tr ansmitted energy
Target
EMISSIVITY
47
Appendix B
46
Appendix B
The measured spot size depends on the distance between the object you are measuring and the infrared thermometer. The relationship between distance and spot size is 60:1 at the focus point. The D:S in the far field (>33ft/10m) is 35:1.
The measured spot size depends on the distance between the object you are measuring and the infrared thermometer. The relationship between dis­tance and spot size is 50:1 at the focus point. The D:S in the far field is 12:1. The best distance between sensor and object is from 10 cm (4“) up to 50 cm (20“).
Optical Resolution (Infrared Themometer) Standard Focus Model
Optical Resolution (Infrared Thermometer)
Close Focus Model
STANDARD MODEL Optical Chart
DISTANCE: SENSOR TO OBJECT (IN)
0
30
20
36
46
48
0.76 IN @ 46 IN
SPOT DIA. (IN)
1.0
0.9
0.87
23
24
24
250
100
500
0
SPOT DIA. (mm)
DISTANCE: SENSOR TO OBJECT (mm)
FOCUS POINT D:S = 60:1 FAR FIELD D:S = 35:1
1.2
0.82
0.81
19.8
29.0
19 mm @ 1150 mm
1000
60
1500
1.51
84
72
1.85
44.0
2000
58.0
108
100
2.3
2500
120
2.9
2.5
72.0
3000
CLOSE FOCUS MODEL Optical Chart
DISTANCE: SENSOR TO OBJECT (IN)
0
6
7.9
2
SPOT DIA. (IN)
1
0.9
0.75
18,7
22
25
0
50
SPOT DIA. (mm)
DISTANCE: SENSOR TO OBJECT (mm)
FOCUS POINT D:S = 50:1 FAR FIELD D:S = 12:1
9.8
4
0.24 IN @ 11.8 IN
0.62
0.49
0.37
9,2
12,3
15,5
6 mm @ 300 mm
150
100
200
250
20
0.93
Close Focus
23
500
40
2.81
70
1000
49
Appendix B
48
Appendix B
Aluminum* 0.30 Asbestos 0.95 Asphalt 0.95 Basalt 0.70 Brass* 0.50 Brick 0.90 Carbon 0.85 Ceramic 0.95 Concrete 0.95 Copper* 0.95 Dirt 0.94 Frozen food 0.90 Hot food 0.93 Glass (plate) 0.85 Ice 0.98 Iron* 0.70 Lead* 0.50 Limestone 0.98 Oil 0.94 Paint 0.93 Paper 0.95 Plastic** 0.95 Rubber 0.95 Sand 0.90 Skin 0.98 Snow 0.90 Steel* 0.80 Te xtiles 0.94 Water 0.93 Wood*** 0.94
*oxidized ** opaque, over 20 mils ***natural

Table of emissivity values of common materials

The DIP-switches are located in the battery compartment (8). In the beginning we recom­mend you use the default DIP-switch settings.

DIP switch settings - Factory Defaults

Factory Defaults DIP-Switch Settings
ON
Lock
°C/°F
Buzzer
Backlight
Set Default
Ltd. Access
Laserflash
Photoflash
Auto/All
Time/Date
NTC/TC
TC-J/TC-K
Celsius setting shown
ON
Lock
°C/°F
Buzzer
Backlight
Set Default
Ltd. Access
Laserflash
Photoflash
Auto/All
Time/Date
NTC/TC
TC-J/TC-K
Fahrenheit setting shown
51

Troubleshooting

50
Troubleshooting
Symptom
Display Code Problem Action
Camera cannot Driver installation Open send photos failed Harware Manager on to the PC the PC to fix it
-O- Target temperature is over Select target
-U- or under range within unit’s specs
EEPROM-Err EEPROM error Contact Factory
CalAreaErr calibration errors Contact Factory ProbCalEr
Flash needs Battery is low Replace Batteries too long to charge or does not charge at all
Blank display Battery is dead Replace Batteries
Laser won’t Low or dead battery Replace Batteries work
Ambient above 45°C (113°F) Operate unit in
45°C (113°F) ambient or below
Display “ON” Display locked “ON” Disconnect the unit
from the PC
Photos lost Batteries changed before Wait for shutdown
shutdown or too long to end. Change after removing old ones batteries within two
minutes of removing
old ones PC displays “logger The setup in the unit and Click “OK”. Either save file does not match in the PC software do not the current data to a device logger setup” match new file name or find
the correct *.lgg file,
open it in “Setup
Device Logger”, save,
and redownload the
data with the “Data
Logger/Load Device
Data Logger”
Sometimes a long beep and the red LED above the display shines to indicate a problem, check your display to find out which problem has occured.
Symptom
Display Code Problem Action
Photo quality Light is sufficient for photo Use Photoflash diminished with without flash always on (Dip switch available light. “Auto/All”set to “All” (Abrupt transitions or provide more between color shades) ambient light. Connection of unit Wrong com port selected Select correct com to PC software or correct com port in use. or disconnect other does not happen device from correct when software started com port No picture in More than one video Right mouse click “Live” mode source is installed on the “Live” mode window window. Choose
correct “Video source”
in the menu. Unit indicates The light is too bright for Go to next location “Use flash” in a photo or reduce brightness. bright light Unit leaves LOG The last position has been used If for the data and after a photo and to record photo and data. photos required, then data recorded. push Enter button and
use arrow keys to
re-record a position. Photo Mem! All memory for storing photos 1. Download existing
has been used. photos and data and
clear memory.
- or -
2. If you don’t mind
losing photos
in the unit, you can go
to Setup in the unit,
push the button until
“Clear photo” appears
in the display and
press Enter.
52
53

Maintenance

Lens Cleaning: Blow off loose parti­cles using clean compressed air. Brush remaining debris away with a camel’s hair brush. Wipe the surface with a moist cotton swab. The swab
may be moistened with water or a water based glass cleaner. NOTE: DO NOT use solvents to clean the plas­tic lens.
Cleaning the Hous­ing: To clean the exterior housing, use soap and water or a mild commer­cial cleaner. Wipe with a damp sponge or soft rag.
55

CE Confirmity

54

Important Hints

Avoid static electricity, arc welders, and induction heaters. Keep away from very strong EMF (electromagnetic fields).
WARNING: DO NOT touch live voltage with contact probe.
Manufacturer, Address
Made in Germany: Month, Year
Model: XXXXXXX Serial: 000000-0000-0000
Power Requirements 3 V
SERIAL NUMBER LABEL
Avoid abrupt changes in temperature. If this occurs, allow 40 minutes for thermal stabiliza­tion before use to prevent the possibility of inaccurate temperature readings.
Thermal Shock
This instrument conforms to the following standards:
EMC: - EN 61326-1
Safety: - EN 61010-1:1993 / A2:1995
- EN 60825-1:1994
This product herewith complies with the requirements of the EMC Directive 89/336/EEC and the Low Voltage Directive 73/23/EEC.
This instrument conforms to the Standards of the European Community.
Certification
The temperature sources used to calibrate this instrument are traceable to the U.S. National Institute of Standards and Technology (NIST) and the Deutscher Kalibrierdienst (DKD). Calibra­tion certificates are available as an option from the manufacturer.
Don’t leave the unit on or near objects of high temperature.
57
56

Secifications (Camera)

Secifications (Thermometer)

Factory defaults

Default Range
Emissivity/Gain 0.95 0.10 to 1.50
in steps of 0.01
Emissivity Table Free 30 materials
Mode normal
Hi Alarm Standard model 900°C (1600°F) -30 to 900°C
(-25 to 1600°F)
Sub Zero model 500°C (932°F) -50 to 500°C
(-58 to 932°F) Lo Alarm Standard model -30°C (-25°F) -30 to 900°C
(-25 to 1600°F) Sub Zero model -50°C (-58°F) -50 to 500°C
(-58 to 932°F) Offset Adjust 0°C (0°F) -10 to 10°C
(-18 to 18°F)
Graphic Display Auto Range Auto Range
/ Man Range
Cycle Time 0.2 sec 0.1 sec to 300 sec Data logger 26 points 100 points max.
pre-set with emissivity 0.95 adjustable only via Software
Temp. Range (Standard Model) - 30 to 900°C (- 25 to 1600°F)
Temp. Range (Sub Zero Model) - 50 to 500°C (-58 to 932°F)
Display Resolution 0.1°C (0.2°F)
Accuracy (Infrared, Standard Model) ± 0.75% of reading or at 25°C (77°F) ambient temperature ± 0.75°C (± 1,5°F), whichever is greater
±2°C (± 4°F) for targets below -5°C (23°F) Accuracy (Infrared, Sub Zero Model) ± 1% of reading or ± 1°C (2°F), at 25°C (77°F) ambient temperature whichever is greater for targets between
- 5 to 500°C (23 to 932°F), ± 1.5°C (3°F)
for targets between -30 to -5°C (-22 to 23°F)
and ± 2°C (4°F) for targets beween
-50 to -30°C (-58 to - 22°F)
Ambient derating < 0.05K/K or < 0.05%/K,
whichever is greater at
+ 25°C (77°F) ± 25°C (± 45°F)
Optical Resolution 60:1 (19mm spot size at 1.15m.)
(Standard Focus) (0.75in. spot size at 3.8 feet)
Optical Resolution 50:1( 6mm spot size at 0.3m.)
(Close Focus) (0.24in. spot size at 0.98 feet)
Accuracy ± 2°C or ± 0.75%,
(Thermocouple K & J) whichever is greater
Accuracy
(Thermistor)
-30 to 0°C (-22 to 32°F) ± 0.6°C 0to 70°C (32 to 158°F) ± 0.4°C
70 to 100°C (158 to 212°F) ± 1°C
100 to 120°C (212 to 248°F) ± 1.5°C
Repeatability ± 0.5% of reading or ± 0.5°C (1°F),
(Infrared) whichever is greater,
±1°C (± 2°F) for targets below -5°C (23°F)
Response Time (95%) 250 mSec
Hot Spot Detection (30%) 85 mSec
Spectral Range 8 to 14 µm
Ambient Operating Range 0 to 50°C (32 to 122°F)
Storage Temperature -20 to 50°C (-4 to 122°F) without batteries Relative Humidity 10 to 95% at 30°C (86°F), non condensing
Analog output 1 mV/°C(°F)
Digital Output USB 1.1 Power 2 x 1.5 V Alkaline Type AA
Battery Life (continuous use) 8 hrs (13 hrs with photo mode off)
Dimensions 240 x 170 x 50 mm (7.9 x 6.7 x 2 inches) Tripod Mount 1/4”-20 UNC
Maximum Picture Number 26 640x480 Pixels (VGA)
Maximum Picture Number 100 320x240 Pixels (1/4 VGA)
Recharge Time for Flash app. 5 sec
Useful Flash Range Standard Focus 0.5 to 2m (19 to 79 in.) Close Focus 0.2 to 1m ( 8 to 40 in)
Camera Lens 6 mm (approx. equal to
42 mm on a 35 mm camera)
Focal Points 200 mm (8 in.)(Close-up)
Infinity (Far Distance)
Light Sensitivity 6 lux
Shutter Speed variabel, max. 1/15 sec
Data Interface USB 1.1
Image File Format JPEG
To open the battery compartment, press gently on the top part of the handle to release the catch and pivot the grip as shown in the figure on the cover of the printed manual. Position the batteries [two alkaline R6 (AA, UM3)], as shown on the housing. (See coverpage of the printed manual)
Caution:
Wait for unit Auto Shut down before battery change! You will see the message “Shut down” for a few seconds (max. 30 sec) on the display after the unit is used in LOG mode. Download stored photos and data and remove batteries when you are not using the unit for a longer time!
If the battery life icon shows two or less illumi-
nated segments, be ready to replace the batteries since you are setting close to the end of battery life.
58

Batteries

Loading...