RaySafe Xi User Manual

4.7 (3)

User Manual

RaySafe Xi

© 2013.02 Unfors RaySafe 5000071-L

All rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic, mechanical or otherwise, is prohibited without the prior written consent of the copyright owner.

2

RaySafe Xi User Manual – Contents

 

Contents

 

Introduction.....................................................................................................

5

Versions...........................................................................................................

5

The RaySafe Xi....................................................................................................

6

The RaySafe Xi Detectors................................................................................

6

The RaySafe Xi Base Unit.................................................................................

7

Getting Started..................................................................................................

8

Keys on the RaySafe Xi Base Unit....................................................................

8

Transfer data to a computer............................................................................

9

Battery and Charger........................................................................................

9

R/F Measurement...............................................................................................

11

Mammography Measurement.......................................................................

14

Special measurement cases..........................................................................

14

Transparent Measurement...........................................................................

18

CT Measurement..............................................................................................

20

How to measure CT dose.............................................................................

20

How to measure CT kVp................................................................................

21

Light Measurement........................................................................................

23

Survey Measurement.....................................................................................

26

mA/mAs Measurement.....................................................................................

29

Passive mAs...................................................................................................

32

Measurement definitions.............................................................................

33

Dose................................................................................................................

34

Dose Rate......................................................................................................

35

kV/kVp............................................................................................................

35

Time...............................................................................................................

36

Pulse...............................................................................................................

37

Frame rate......................................................................................................

37

Dose per Frame.............................................................................................

38

3

RaySafe Xi User Manual – Contents

 

HVL (Half Value Layer)...................................................................................

38

Total Filtration................................................................................................

38

mA and mAs...................................................................................................

38

RaySafe Xi View.................................................................................................

39

Installing RaySafe Xi View.............................................................................

39

Setup of Bluetooth communication..............................................................

40

Getting help...................................................................................................

40

Accessories.......................................................................................................

41

The RaySafe Xi Flexi stand.............................................................................

41

The RaySafe Xi Cassette holder.....................................................................

41

Tips and Tricks.................................................................................................

42

Arrangement of Measurement......................................................................

42

Sensor Position..............................................................................................

42

Reset Switch...................................................................................................

43

Vertical Detector Placement..........................................................................

43

kVp on AMX....................................................................................................

43

Display Codes.................................................................................................

43

Frequently Asked Questions..........................................................................

44

Troubleshooting.............................................................................................

46

Warranty, Service and Support.................................................................

48

Service...........................................................................................................

48

Return Procedure for Service and Warranty.................................................

48

Disposal............................................................................................................

49

4

RaySafe Xi User Manual – Introduction

Introduction

Versions

This user manual covers the following articles:

Base unit version:

8201011

Detector version:

8202011

 

8201021

 

8202021

 

8201013

 

8202022

 

8201023

 

8202031

 

 

 

8202040

 

 

 

8202041

 

 

 

8202050

 

 

 

8202060

 

 

 

8202062

 

 

 

8202070

This manual discribes the RaySafe Xi which is compatible with the Unfors Xi Platinum Plus version.

About this user manual

This user manual is intended to assist users in the safe and effective operation of the product described.

Before attempting to operate the product, you must read these instructions for use, noting and strictly observing all WARNINGS and CAUTION notices.

WARNING A WARNING alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the operator or patient.

CAUTION A CAUTION alerts you to where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury, and/or cause environmental pollution.

NOTE Notes highlight unusual points as an aid to an operator.

These Instructions for Use describe the most extensive configuration of the product, with the maximum number of options and accessories. Not every function described may be available on your product.

5

RaySafe Xi User Manual – The RaySafe Xi

The RaySafe Xi

The RaySafe Xi consists of a base unit and several different external detectors measuring: Radiography/ Fluoroscopy (R/F), mammography (MAM), on AEC or ABC systems (Transparent), Computed Tomography (CT), ambient and emitted light (Light) and scattered or low level radiation (Survey).

Communication between detector and base unit is purely digital, thereby minimizing sensitivity to mechanical or electrical stress. The base unit may also be equipped with an optional integrated tube current meter (mA/mAs).

The RaySafe Xi Detectors

The R/F detector has two sensors: R/F high is designed for conventional, high dose rate measurements normally generated without a phantom between the detector and the X-ray source. R/F low is designed for low dose rate measurements normally generated with a phantom between the detector and the X-ray source.

The mammography sensor is used for both low and high dose rates generated in mammography applications.

The transparent detector is a solid-state detector with a very small radiologically visible footprint. It is intended to be used for applications where the R/F detector would influence the automatic exposure control on X-ray equipment and serves as a complement to the R/F detector. The transparent detector is designed to mimic the response of a pancake ion chamber detector and therefore it has no backscatter protection.

The CT detector is an ionization chamber, designed to measure CT dose for applications such as Dose Length Product (DLP) and Computed Tomography Dose Index (CTDI).

The light detector is designed to measure luminance on LCD and CRT monitors, as well as light boxes, and illuminance for diagnostic X-ray applications.

The survey detector is designed for measuring leakage or scattered radiation from X-ray tubes or in examination rooms and leakage radiation from γ-emitting isotopes.

6

RaySafe Xi User Manual

RaySafe Xi User Manual – The RaySafe Xi

The RaySafe Xi Base Unit

There are two versions of the RaySafe Xi base unit: one with and one without mA/mAs capabilities. Both versions are fully compatible and interchangeable with the RaySafe Xi detectors as long as the firmware supports the detector (see Specifications). In this manual all information refers to both versions unless otherwise specified.

The RaySafe Xi base unit automatically identifies the connected detector and displays the settings and parameters available for that detector.

The built-in active compensation automatically applies corrections for different beam qualities, filtrations and temperatures. During fluoroscopy, survey or light measurements the displayed values are continuously

updated. Data is sent on the serial port, and RaySafe Xi view, the program included with your RaySafe Xi, shows numerical data as well as waveforms on a PC.

7

RaySafe Xi User Manual – Getting Started

Getting Started

1.Connect a RaySafe Xi detector of your choice to the RaySafe Xi base unit with one of the two (2 and 10 m) RaySafe Xi cables.

2.Position as required for the selected sensor.

3.Turn on the RaySafe Xi (ON/OFF key, see below), and the instrument specific setup information is displayed. In the SENSOR MENU a detector or sensor field for your application can be selected.

4.The RaySafe Xi is now in MEASURE MODE and ready to measure. Press STEP to scroll through measured parameters (also possible during fluoroscopy). The last three displayed parameters will automatically show up after the next exposure. To change setup values (SETUP MENU) such as various delays, displayed units and other choices, press SELECT.

Keys on the RaySafe Xi Base Unit

ON/OFF: Turns on the RaySafe Xi and off when in SENSOR MENU.

EXIT: Exit to the previous menu.

STEP: A short press steps through available options.

SELECT: A long press selects an option.

8

RaySafe Xi User Manual – Getting Started

Transfer data to a computer

To view the waveform in the RaySafe Xi view and/or transfer measured data to a PC, connect a serial cable or Bluetooth module (option) to the communication port on the RaySafe Xi base unit.

Battery and Charger

Caution! Only authorized personnel may remove and replace the battery.

Fully charge the battery with the provided battery charger before first use. Battery status in percentage is displayed, together with setup information, when the instrument is switched on.

RaySafe Xi/Unfors Xi Platinum Plus

Unfors Xi Platinum

For RaySafe Xi

The provided battery charger may be connected during all measurements.

For Unfors Xi Platinum

The provided battery charger may be connected during measurement except when measuring mAs, as it may create ground currents affecting the measurement. However, the RaySafe Xi battery charger may be connected during all measurements.

Xi version

Battery type

Hours of operation

Charge time

RaySafe Xi

7.4 V Li-ion

20–40 hours

4 hours

Unfors Xi Platinum

9 V NiMH

10–20 hours

15 hours

9

RaySafe Xi User Manual – Getting Started

There are two battery warning levels:

First warning level, “Battery low”: Finish your measurements and then charge the battery.

Second warning level, “Battery down”: No further measurements are possible.

When the provided battery charger is connected, the external power indicator is lit and Charging battery will be briefly displayed. The RaySafe Xi battery will be charged even if the RaySafe Xi is turned off.

There is a battery charger LED indicator that is indicating red while charging and green when the battery is fully charged.

10

RaySafe Xi User Manual – R/F Measurement

R/F Measurement

For best accuracy, center the selected sensor field (R/F low or R/F high) and position the long axis of the sensor field perpendicular to the anode-cathode axis of the tube.

1. SENSOR MENU

R/F low

Sensor for conventional low dose rate measurements lower than 1 mGy/s (7 R/min),

 

normally after a phantom.

R/F high

Sensor for conventional high dose rate measurements higher than 1 mGy/s (7 R/min),

 

normally before a phantom.

 

 

 

 

 

 

Press STEP to step between sensors and SELECT to select a sensor.

2. MEASURE MODE

The displayed values are updated after each exposure, or continuously after 4 seconds of fluoroscopy.

kVp

kVp (for the R/F high sensor at adequate signal levels) or kV

Dose

Gy (Air kerma, free in air) or R (Exposure)

Dose rate

Gy or R per s, m or h

Time

ms, s

HVL

mm Al

Total Filtration, TF

mm Al

11

RaySafe Xi User Manual – R/F Measurement

Pulses

pulses

Frame rate

frames/s

Dose per frame

Gy or R

mAs

mAs (requires RaySafe Xi base unit with mAs)

mA

mA (requires RaySafe Xi base unit with mAs)

3. SETUP MENU

Press SELECT to enter the SETUP MENU (from MEASURE MODE) and STEP to step between setup parameters.

All values are stored in a non-volatile memory and are valid until manually changed. At start up (after the battery status information), valid Trig delay, kVp delay and Calc. delay are displayed in sequence.

(Factory settings in bold.)

Trig delay

A delay in ms after the normal trig of the RaySafe Xi. Utilize when an unwanted part

 

of an exposure, such as a pre-pulse, should be excluded from the measurement. No

 

measurements are performed during the Trig delay. (0, 5, 10, 50, 100, 200, 500, 1000,

 

2000 ms)

Trig level

Use Trig level to measure correct exposure time on waveforms with slowly increasing

 

output, such as with single phase dental. The default setting is a low value, that depends

 

on the selected sensor; see Specifications. Can be set to 25, 50 or 75 % of the peak

 

value of the previous exposure. Use it with the same generator settings and FDD. Not

 

selectable before the first exposure or at low dose rates.

kVp delay

Defined as the delay in ms, after the trig delay, but before the kVp measurement (and

 

waveform) window begins. Use a kVp delay on machines with slow rising output, such

 

as single phase intra oral machines and fluoroscopy systems. (0, 2, 5, 10, 50, 150, 300,

 

1000, 1200, 1500, 1700, 2000 ms)

Calc. delay

Defined as the dead time after end trig, before data is calculated. Default is 0.5 s but it is

 

recommended to use a longer delay when measuring on pulsed fluoroscopy where the

 

time between pulses may exceed 0.5 s. (0.5, 2, 4, 6, 7 s)

12

RaySafe Xi User Manual – R/F Measurement

Dose unit

Change the default units and/or use this feature to recalculate a value to a different unit.

 

(Gy or R per s, m or h, where 1 Gy = 114.1 R.)

Backlight

Display lighting option. (off, 2, 5, 10 s, on)

Auto Power Off Never, or after 5, 20 or 60 minutes of inactivity.

Info

Displays the serial number (S/N) and firmware version of the connected detector and

 

the RaySafe Xi base unit and the calibration date for the selected sensor. There is also a

 

possibility to reset the instrument settings to factory defaults.

kVp Mode

“kV only” prevents RaySafe Xi from calculating kVp, which otherwise occurs

 

automatically when R/F high is selected and the signal level is high enough (“kV/kVp”).

 

Useful for example when measuring near that signal level.

13

RaySafe Xi User Manual – Mammography Measurement

Mammography Measurement

Place the detector along the X-ray tube axis (with the cable end pointing towards the X-ray machine). For kV measurements, make sure that the center of the MAM sensor square is placed 6 cm from the examination table front edge. Dose and HVL can be measured from other positions as well.

The following concerns Unfors Xi Platinum, not RaySafe Xi:

NOtE! When measuring on mammography machines with carbon fibre table, we recommend not to use the power adapter to charge Xi, or use something that creates at least a 5 mm distance between the table and the detector, such as a plastic plate.

Special measurement cases

The table below indicates when to use filter for kV measurements.

Beam quality

Filter

Comment

Mo/Mo

No filter

 

Mo/Rh

2 mm Al

 

W/Rh

No filter

 

W/Al 20–39 kV

No filter

 

W/Al 40–49kV

2 mm Al

Use the R/F high detector

W/AlScanning

2 mm Al

 

Mo/Rh

When measuring kV with Mo/Rh, use the included 2 mm Al filter. Follow the instructions printed on the filter.

14

RaySafe Xi User Manual – Mammography Measurement

W/Al Scanning (separate detector)

When measuring kV on W/Al Scanning, use the included 2 mm Al filter. Follow the instructions printed on the filter. Please use the dedicated detector holder. Be sure to push it tight to the examination table front. For more information, see the separate mammography W/Al Scanning Quick Guide (enclosed with the holder).

SCANNING

Note! Some of the scanning W/Al mammography machines have inhomogenous radiation fields. A consequence of this is that the position of the detector is very important. Also, detectors of different viewing angles will result in different dose readings.

W/Al (Hologic)

To measure kV in the range of 40-49 kV (W/Al), the R/F MAM detector and a 2 mm Al filter is needed. Set the detector to R/F-high (see R/F Measurement). When measuring in Combo mode on Hologic Selenia Dimensions 3D, use the sensor menu choice “Combo” and 7 s Calc. delay (located in the Setup Menu).

1. SENSOR MENU

Mammography

Sensor for mammography measurements with submenu for

Mo/Mo No Paddle

 

selection of beam qualities, see list to the right (depending

Mo/Mo Paddle

 

on configuration). The choice of “paddle” or “no paddle” only

Mo/Rh

 

applies if the kV measurement is going to be made with or

Mo/Al

 

without a paddle in the beam and does not affect dose or HVL

Rh/Rh

 

measurements.

Rh/Al

 

 

W/Rh No Paddle

 

 

W/Rh Paddle

 

 

W/Rh* No Paddle

 

 

W/Rh* Paddle

 

 

W/Ag

 

 

W/Al S (scanning)

 

 

W/Al No Paddle

 

 

W/Al Paddle

 

 

Combo

Note! Select W/Rh* for measurements on Hologic Selenia and Planmed Nuance. W/Al (scanning) is on a separate detector.

15

Loading...
+ 34 hidden pages