19. Technical Support ....................................................................................................... 64
20. RAE Systems Contacts ............................................................................................... 64
21. Appendix A: Regulatory Information ......................................................................... 67
2
ToxiRAE Pro PID User’s Guide
WARNINGS
Read Before Operating
This manual must be carefully read by all individuals who have or will have the
responsibility of using, maintaining, or servicing this product. The product will perform
as designed only if it is used, maintained, and serviced in accordance with the
manufacturer’s instructions.
CAUTION!
Never operate the monitor when the cover is removed. Remove the monitor top cap, rear
cover, fan, or battery only in an area known to be non-hazardous.
Product Registration
Register your product online by visiting:
http://www.raesystems.com/support/product-registration
By registering your product, you can:
Receive notification of product upgrades or enhancements
Be alerted to Training classes in your area
Take advantage of RAE Systems special offers and promotions
3
ToxiRAE Pro PID User’s Guide
Use only RAE Systems lithium-ion battery part number G02-3004-000 (3.7V,
1,800mAh). This instrument has not been tested in an explosive gas/air atmosphere
having an oxygen concentration greater than 21%. Substitution of components may
impair suitability for intrinsic safety. Replace batteries only in non-hazardous locations.
STATIC HAZARD: Clean only with a damp cloth to avoid damage from electrostatic
discharge.
For safety reasons this equipment must be operated and serviced by qualified personnel
only. Read and understand this instruction manual completely before operating or
servicing.
The monitor should be calibrated every time it does not pass a bump test, but no less
frequently than every six months, depending on use and exposure to gas and
contamination, and its operational mode.
The ToxiRAE Pro PID monitor is factory-preconfigured to a 30-day calibration interval,
so it will go into calibration alarm if it is not calibrated for 30 days or more.
• Calibration intervals and bump test procedures may vary due to national legislation.
• RAE Systems recommends using RAE calibration gas cylinders with one of the
Constant-flow DGA-600 regulator with adapter, PN 007-3001-000
WARNINGS
4
ToxiRAE Pro PID User’s Guide
Utiliser seulement la batterie Lithium-ion RAE Systems référence G02-3004-000 (3.7V,
1,800mAh). Cet instrument n’a pas été testé dans une atmosphère de gaz/air explosive
ayant une concentration d’oxygène plus élevée que 21%. La substitution de composants
peut compromettre la sécurité intrinsèque. Ne changer les batteries que dans un
emplacement désigné non dangereux.
RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer uniquement avec un chiffon
humide.
Pour des raisons de sécurité, cet équipement doit être utilisé et entretenu uniquement par
un personnel qualifié. Étudier le manuel d’instructions en entier avant d’utiliser, ou d’entretenir l’équipement.
Le ToxiRAE Pro PID est configuré en usine avec une période de calibration de 30 jours,
si il n’est pas calibré durant cette période , il passe en alarme.
Une vérification d’étalonnage est définie par une exposition du détecteur au gaz
d’étalonnage qui doit déclencher le seuil d’alarmes bas.
Les périodes de calibration peuvent varier suivant les législations locales.
RAE Systems préconise d’utiliser les gaz de calibration suivants:
débit constant DGA-600 regulateur avec adaptateur, PN 007-3001-000
AVERTISSEMENT
5
ToxiRAE Pro PID User’s Guide
1
Wireless functionality is currently offered in 2011 in North America and Europe. Contact RAE Systems for
availability in other regions.
2
Available with Industrial Hygiene configurations only.
3
Contact RAE Systems for availability.
4
Wireless versions have a functioning wireless modem and are ready for wireless deployments. “Nonwireless” versions cannot be upgraded to wireless in the future. Additional equipment and/or software
licenses may be required to enable remote wireless monitoring and alarm transmission.
1. General Information
ToxiRAE Pro PID is a personal wireless monitor for volatile organic compounds
(VOCs), the world’s smallest.Featuring RAE Systems’ next-generation PID sensor, the
ToxiRAE Pro PID can promptly detect and accurately monitor over 300 VOCs. With an
onboard library of 190 correction factors, the ToxiRAE Pro PID can be programmed to
automatically read in concentrations of the specified compound.
Key Features
Wireless1. Personal. Proven
• Wireless access to real-time instrument readings and alarm status from any location
• Unmistakable five-way local and remote wireless notification of alarm conditions
• Cutting-edge PID with 190 built-in correction factors
• Largest display in its class
• Reliable, rugged, and intrinsically safe
• Continuous datalogging2 capability
• Fully automated bump testing and calibration with AutoRAE 23
• Easy to maintain with replaceable sensor, fan, and Lithium-ion battery
• Available in Industrial Hygiene (advanced) or Safety (basic)
• Wireless or non-wireless versions4 are available
3
configurations
6
ToxiRAE Pro PID User’s Guide
ToxiRAE Pro PID Specifications
Dimensions
118 mm x 60 mm x 30 mm (4.6" x 2.4" x 1.2")
Weight
235g (8.3 oz)
Battery
Internal rechargeable Li-ion
Charging Time
Less than 4 hours
Operating Time
More than 12 hours on full charge (normal operation)
Charger
Charging and PC communication cradle: AC / DC adapter with international
prong kit (100 to 240V, 50/60 Hz AC to 12V DC)
Multi-unit charging station for up to 5 units (data download from one):
AC/DC adapter for the multi-unit charging station
Alarms
Audio alarm – 95 dB @ 30cm
Visual alarm – 4 bright red LED lights
Vibration alarm
Activated display backlighting and on-screen indication of alarm conditions
Wireless alarm notification sent to remote observers (wireless versions only)
Display
Monochrome dot-matrix graphical LCD display
Display Backlight
Manually and on alarm
Keypad
2 programming/operation keys
Direct Readout
Real-time reading of VOC concentrations in parts per million or mg/m3;
battery status; data logging on/off status; wireless reception quality; STEL,
TWA, and peak values; calibration and measurement gases, correction factors
Sampling Method
Fan-assisted diffusion
Datalogging
Capacity: 3 months at 1-minute interval
Datalog interval: user-configurable from 1 to 3,600 seconds
Calibration
Two-point field calibration of zero and standard reference gas
Adjustable span value
Maintains calibration date
IP Rating
IP-54
Docking Station
AutoRAE 2 Automatic Test and Calibration Station
Protection:
Password-protected calibration settings, alarm limits, and data
Hazardous
Location
Approvals
Class I, Div. 1; Gr. A,B,C,D T4
Class I, Zone 0 A/Exia IIC T4
ATEX: II 1G,
Ex ia IIC Ga T4
IECEx: Ex ia IIC Ga T4
China: Ex ia IIC T4
Two years on non-consumable components (per RAE Systems Standard
Warranty); one year on fan, sensor, battery, and other consumable components
2. Specifications
7
ToxiRAE Pro PID User’s Guide
[Y/+] key
[MODE] key
Gas inlet
Display
LEDs
Alarm
Buzzer
Belt clip (on back)
LED
Top cap
3. Overview
The ToxiRAE Pro PID is a wireless, personal volatile organic compounds (VOC)
monitor, the world’s smallest. The ToxiRAE Pro PID takes worker protection to the next
level by providing safety professionals wireless access to real-time instrument readings
and alarm status from any location for better visibility and faster response. The ToxiRAE
Pro PID features RAE Systems’ next-generation PID sensor that delivers fast, reliable
detection and accurate monitoring of a wide variety of VOCs in a broad range of
temperature and humidity conditions.
4. User Interface
The ToxiRAE Pro PID’s user interface consists of the display, four LEDs, an alarm
buzzer, and two keys, [MODE] and [Y/+]. The LCD displays real-time reading in
specified measurement units, sensor type, alarm type (when in alarm, including cal.
overdue), battery status, datalog (if on), and radio and connection quality (if available).
8
ToxiRAE Pro PID User’s Guide
Icon
Function
Wireless status: the radio is on
Wireless status: the radio is off
The instrument is not equipped with a radio (no icon)
Wireless strength (0 to 5 bars), only when radio is turned on
Datalogging status (shown when datalogging is on, blank when off)
Battery status (three segments show battery charge level)
Man Down alarm enabled
Sensor calibration due
Bump test due
All sensors tested and calibrated tick mark (all sensors have been bump tested
and calibrated; no sensor is overdue for a bump test or calibration according to
the intervals configured on the instrument
Calibration and bump
test up to date
Sensor name and “Alarm” are
displayed alternately when in alarm
Wireless on/off status
Wireless signal strength
Unit of measure
Soft keys
(functions change by activity)
Alarm type
(when in alarm)
Bump test due
Sensor
calibration due
Reading
Man Down alarm on
Datalog on
Battery status
4.1. Display Overview
The LCD display provides visual feedback that includes the sensor types, readings, time,
battery condition, and other functions.
4.2. Status Indicator Icons
Along the top of most screens are status indicators that tell you whether a function is
operating and/or its strength or level.
9
ToxiRAE Pro PID User’s Guide
Rear view of
ToxiRAE Pro PID
ToxiRAE Pro
Charging and PC
Communications
Cradle
Raised lip on
rear of ToxiRAE Pro PID
locks into cradle’s latch
Rear view of
charging cradle
Charging status
LED
Connect AC
adapter to the
charging cradle
Latch release
5. Charging The ToxiRAE Pro PID
Always fully charge the battery before using the ToxiRAE Pro PID. The ToxiRAE Pro
PID’s Li-ion battery is charged by placing the ToxiRAE Pro PID in its cradle. Contacts
on the bottom of the ToxiRAE Pro PID meet the cradle’s contact pins, transferring
power.
Note: Before setting the ToxiRAE Pro PID into its charging cradle, visually inspect the
contacts to make sure they are clean. If they are not, wipe them with a soft, dry cloth. Do
not use solvents or cleaners.
10
ToxiRAE Pro PID User’s Guide
Press on latch
Lift ToxiRAE Pro PID out of
charging cradle
Follow this procedure to charge the ToxiRAE Pro PID (the procedure is applicable to the
instrument with or without the rubber boot installed):
1. Plug the AC/DC adapter into the ToxiRAE Pro PID’s cradle.
2. Plug the AC/DC adapter into the wall outlet.
3. Place the ToxiRAE Pro PID into the cradle and press down until it is locked in
place.
The ToxiRAE Pro PID begins charging automatically. The LED in the cradle should
glow red to indicate charging. When charging is complete, the LED in the cradle glows
green.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge the instrument
only in areas known to be non-hazardous. Remove and replace batteries only in
areas known to be non-hazardous.
5.1. Removing The ToxiRAE Pro PID From The Charging
Cradle
To remove the ToxiRAE Pro PID from its charging cradle, press down on the latch
release on the back of the charging cradle while slightly tilting the ToxiRAE Pro PID
toward the front of the cradle, and lift it out.
5.2. Low Voltage Shutoff
When the battery’s charge falls below the shutoff voltage, the ToxiRAE Pro PID then
turns itself off. Recharge the battery by placing the ToxiRAE Pro PID in its cradle (the
cradle should be connected to the AC adapter).
11
ToxiRAE Pro PID User’s Guide
6. Turning The ToxiRAE Pro PID On
To turn the ToxiRAE Pro PID on, press and hold [MODE] for 3 seconds.
Caution: The alarm buzzer is very loud. During startup, you can mute most of the sound
by holding a finger over the alarm buzzer port.
Note: Never put tape over the alarm buzzer port to permanently mute it or otherwise limit
the loudness of the buzzer.
1. When the ToxiRAE Pro PID turns on, the buzzer, vibration alarm, and LEDs are
tested. The firmware version is shown in the display (for example, “V1.00” indicates
firmware version 1.00).
2. A battery test is performed. If the battery has sufficient charge, then the next step is a
self test of the system.
Note: If the battery is almost fully discharged, the message “Battery too low! Needs
charging. Powering off!” shows in the display and the ToxiRAE Pro PID shuts off.
You must charge the battery before you can use the instrument.
3. Self-testing is performed by the ToxiRAE Pro PID.
Note: If one of the following error messages is displayed, shut off the ToxiRAE Pro
PID by pressing [MODE]:
Sensor error!
Wrong sensor!
If any of the following error messages occurs, you can either shut off the instrument
by pressing [MODE], or continue to the next startup test by pressing [Y/+]:
No fan or fan error!
Real time clock error!
Memory error!
Sensor changed!
Failed to turn on lamp!
For any of these error types, try restarting the instrument. If the error message appears
again after five more attempts to restart the instrument, consult the distributor from
whom the instrument was purchased for technical support.
12
ToxiRAE Pro PID User’s Guide
4. After self-testing, two start-up options are available for the ToxiRAE Pro PID:
Normal and Fast Start-up. The following information is displayed in sequence under
Normal Start-up and skipped under Fast Start-up:
Sensor type
Sensor serial number
Date, time, and temperature
Calibration gas
Last calibration date
Measurement gas
Correction factor
User mode
Battery type, voltage, and shutoff voltage
Alarm mode
Buzzer enabled/disabled status
Light enabled/disabled status
Vibration alarm enabled/disabled status
Alarm settings for High, Low, STEL, and TWA
Datalog type and interval
5. Once the startup process is complete, a sensor warm-up screen is shown (with
countdown, if necessary). Then the main measurement screen is shown. The type of
monitored hazard is displayed, with current reading. The ToxiRAE Pro PID is now
ready for use.
7. Turning The ToxiRAE Pro PID Off
Press and hold [MODE]. In 2 seconds, a 5-second countdown to shutoff begins,
accompanied by one buzzer beep per second. You must hold your finger on the key for
the entire shutoff process. If you remove your finger from the key during the countdown,
the shutoff operation is canceled and the ToxiRAE Pro PID continues normal operation.
When you see “Unit Off,” release your finger from the [MODE] key. The ToxiRAE Pro
PID is now off.
8. Testing The Alarms (Manually)
Under normal non-alarm conditions, the buzzer, vibration alarm, and LEDs can be tested
at any time by pressing and holding [Y/+].
Important! If any of the alarms does not respond to this test, check the Alarm Settings. It is
possible that any or all of the alarms have been turned off. If all the alarms are turned on, but
one or more of them (buzzer, LED lights, or vibration alarm) does not respond to this test, do
not use the instrument; contact your RAE Systems distributor for technical support.
13
ToxiRAE Pro PID User’s Guide
9. Low Battery Warnings
If, during operation, the battery’s charge is low, the battery icon in the display screen
turns empty and blinks once every second. At the same time, the following notifications
occur every minute: the buzzer beeps and LEDs blink once, followed by a vibration and a
screen message “Battery low! Shut off in X minutes.” The countdown to shutoff starts 10
minutes prior to automatic shutoff.
10. Dead Battery Warning & Shutoff
If you turn on the ToxiRAE Pro PID and the battery’s charge is below its lowest
threshold, the display shows this message:
Battery too low!
Needs charging.
Powering off!
It then automatically shuts off. You must charge the ToxiRAE Pro PID’s battery before
you can continue to use this instrument.
Important! Do not operate the ToxiRAE Pro PID again until its battery has been
charged.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, charge the battery only in
areas known to be non-hazardous.
14
ToxiRAE Pro PID User’s Guide
11. Operating The ToxiRAE Pro PID
11.1. Overview
The ToxiRAE Pro PID has two operational modes:
Normal — see page 15 for detailed instructions.
Diagnostic — see page 48 for detailed instructions.
The following is an overview of the two modes:
Normal Mode is the default mode. It is accessed when you turn on the ToxiRAE Pro
PID. Two sub-modes are available: Basic User and Advanced User. In Basic User, some
restrictions are applied, including password protection that guards against entering
Programming Mode by unauthorized personnel. In Advanced, there are no such access
restrictions (you do not need a password), and it provides the indications and data you
need most for typical monitoring applications.
Diagnostic Mode is primarily intended for technicians to use during troubleshooting, although it
also offers access to a few changeable parameters that you may rarely (if ever) need to change. A
password is required to enter Diagnostic Mode (the password is the same as (and
automatically kept in sync with) the one used to enter Programming Mode). The default
password is 0000. To enter Diagnostic Mode, power the instrument off and then turn it
back on by pressing and holding both the buttons. In Diagnostic Mode, ToxiRAE Pro PID
displays readings in raw counts instead of units such as parts per million (ppm).
It is not possible to switch to Programming from Diagnostic Mode without turning off the
instrument first and then restarting it in the desired mode.
11.1.1. Normal Mode
Normal Mode is the default mode of the ToxiRAE Pro PID when it is turned on. The
selected sub-mode (Basic or Advanced) determines the number of screens that are
accessible.
15
ToxiRAE Pro PID User’s Guide
11.1.2. Basic User Mode
When the ToxiRAE Pro PID is in Basic User Mode, pressing [MODE] allows you to step
through the screens in the following fashion:
Note: The gray boxes indicate the datalog functions when datalogging is in Manual
mode. When datalogging is in Automatic mode, these screens do not appear. Refer to
page 28 for information on selecting Automatic or Manual datalogging.
16
ToxiRAE Pro PID User’s Guide
11.1.3. Advanced User Mode
Note: To enter Advanced User mode from Basic User mode, you must enter
Programming mode on an instrument that is running in Basic User mode by
simultaneously pressing and holding both the buttons for 3 seconds and supply a
password.
To exit Programming Menu and return to the measurement screen (in Advanced User
mode), press the [MODE] button repeatedly to highlight the Exit icon in Programming
Menu and press Y to exit to the measurement screen.
When the ToxiRAE Pro PID is in Advanced User mode, pressing [MODE] steps through
the same screens, but with the addition of a screen that allows for selection of different
measurement gases from the internal library. Also, once in Advanced User mode, you
may enter Programming without using a password.
Note: In Advanced User Mode, you may select different measurement gases from the
internal library.
Note: The gray box indicates the datalog functions when datalogging is in Manual mode.
When datalogging is in Automatic mode, these screens do not appear. Refer to page 28
for information on selecting Automatic or Manual datalogging.
11.2. Navigation In Normal Mode
Step through the screens in Normal Mode by pressing [MODE]. Some screens allow
actions on your part, such as clearing the data. In these instances, pressing [Y/+] performs
the action.
17
ToxiRAE Pro PID User’s Guide
11.2.1. TWA, STEL, Peak
The first screen includes a readout for TWA, STEL, and Peak readings.
TWAThe TWA (time-weighted average) reading is the average reading of the gas
concentration times that the monitor has been on, prorated to an 8-hour period.
STELShort Term Exposure Limit (STEL) data based on 15-minute STEL values,
expressed in ppm or mg/m3. A STEL value is only displayed after the
ToxiRAE Pro PID has been on for 15 minutes or longer. After exiting PC
Communications mode, the STEL is reset.
PeakThe Peak value tells you the highest reading for the sensor since the ToxiRAE
Pro PID was turned on. Press [Y/+] twice to clear the Peak value.
Note: If you do not press a key within 1 minute, the display reverts to the
normal reading.
Press [MODE] to advance from screen to screen.
11.2.2. Date, Time, And Temperature
The current Date is expressed as Month/Day/Year, with two digits for month and day,
and four digits for the year. Below the date is the time, which is expressed as
Hours/Minutes/Seconds, with two digits for each. The time is in 24-hour (military)
format. Below the time is the internal temperature, which can be in degrees Fahrenheit or
Celsius (settable in Programming Mode).
11.2.3. Wireless Settings
The current wireless settings are shown here. These include the PAN ID, Channel, and
Unit ID, which are helpful when creating or troubleshooting networks. Parent ID data
does not apply.
This screen shows calibration gas (Cal. Gas), measurement gas (Meas. Gas), and
correction factor (CF).
18
ToxiRAE Pro PID User’s Guide
11.2.5. Enter Communications and stop measurement?
Before communicating with a PC running ProRAE Studio II Instrument Configuration
and Data Management software (for downloading datalogs, configuring the instrument,
or uploading new firmware), the ToxiRAE Pro PID must be set to communicate with the
computer. Press [MODE] repeatedly until you reach the “Enter Communications and stop
measurement?” screen. Then press [Y/+]. Measurement and datalogging are stopped, and
the instrument is now ready to communicate with the PC.
11.3. Programming
You must provide a 4-digit password to enter Programming Mode, unless the ToxiRAE
Pro PID is already running in Advanced User mode. The default password is 0000.
11.4. Entering Programming Mode
1. Press and hold [MODE] and [Y/+] simultaneously until the Password screen appears.
2. Input the 4-digit password:
Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again "wraps" around back to 0.
Step from digit to digit using [MODE].
When you are done, continue pressing [MODE] until OK or Cancel is highlighted.
Press [Y/+] to register the password (if OK is highlighted) or to exit without entering
the password (if Cancel is selected).
If you make a mistake, you can cycle through the digits by pressing [MODE] repeatedly
and then using [Y/+] to change the number in each position.
Note: The default password is 0000.
19
ToxiRAE Pro PID User’s Guide
When you have successfully entered Programming Mode, the following screen appears:
Note: The password can only be changed by connecting the instrument to a PC running
ProRAE Studio II software. Follow the instructions in ProRAE Studio II to change it.
The Calibration icon is highlighted first by default, but you can press [MODE] to step
from one programming menu item to the next, with the name of the menu shown at the
top of the display and the corresponding icon highlighted. As you repeatedly press
[MODE], the selection moves from left to right, and the following screens appear in
sequence:
Note: When you reach Exit and press [MODE], the menu cycles back to Calibration.
To exit the Programming Menu, highlight Exit by pressing the [MODE] button, and press
the [Y/+] key.
20
Loading...
+ 50 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.