RAE Systems ToxiRAE Pro PID Operating Manual

P/N G02-4008-000 Rev. C
March 2013
ToxiRAE Pro PID
User’s Guide
© Copyright 2013 RAE Systems, Inc.
ToxiRAE Pro PID User’s Guide
Contents
Read Before Operating ................................................................................................... 3
1. General Information ........................................................................................................ 6
2. Specifications .................................................................................................................. 7
3. Overview ......................................................................................................................... 8
4. User Interface .................................................................................................................. 8
4.1. Display Overview .......................................................................................... 9
4.2. Status Indicator Icons ..................................................................................... 9
5. Charging The ToxiRAE Pro PID .................................................................................. 10
5.1. Removing The ToxiRAE Pro PID From The Charging Cradle .................. 11
5.2. Low Voltage Shutoff.................................................................................... 11
6. Turning The ToxiRAE Pro PID On .............................................................................. 12
7. Turning The ToxiRAE Pro PID Off ............................................................................. 13
8. Testing The Alarms (Manually).................................................................................... 13
9. Low Battery Warnings .................................................................................................. 14
10. Dead Battery Warning & Shutoff ............................................................................... 14
11. Operating The ToxiRAE Pro PID ............................................................................... 15
11.1. Overview ...................................................................................................... 15
11.1.1. Normal Mode ................................................................................................ 15
11.1.2. Basic User Mode ........................................................................................... 16
11.1.3. Advanced User Mode ................................................................................... 17
11.2. Navigation In Normal Mode ........................................................................ 17
11.2.1. TWA, STEL, Peak ........................................................................................ 18
11.2.2. Date, Time, And Temperature ...................................................................... 18
11.2.3. Wireless Settings ........................................................................................... 18
11.2.4. Calibration Gas, Measurement Gas, Correction Factor ................................ 18
11.2.5. Enter Communications and stop measurement? ........................................... 19
11.3. Programming................................................................................................ 19
11.4. Entering Programming Mode ...................................................................... 19
11.4.1. Navigating Programming Menus .................................................................. 21
11.5. Programming Menus .................................................................................... 21
11.5.1. Calibration..................................................................................................... 22
11.5.2. Measurement ................................................................................................. 23
11.5.3. Alarm Setting ................................................................................................ 24
11.5.4. Datalog .......................................................................................................... 28
11.5.5. Monitor Setup ............................................................................................... 30
11.6. Exit (Exit Programming Mode) ................................................................... 35
12. Policy Enforcement ..................................................................................................... 36
12.1. Setting Policy Enforcement ......................................................................... 36
12.1.1. AutoRAE 2 Cradle ........................................................................................ 36
12.1.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger ................................ 40
12.2. Deactivating Policy Enforcement ................................................................ 42
12.2.1. AutoRAE 2 Cradle ........................................................................................ 42
12.2.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger ................................ 42
13. Bump Testing & Calibrating The ToxiRAE Pro PID ................................................. 44
13.1. Bump Testing & Calibrating With An AutoRAE 2™ ................................. 44
13.1.1. Bump (Functional) Testing ........................................................................... 44
1
ToxiRAE Pro PID User’s Guide
13.2. Calibration Adapter ...................................................................................... 44
13.2.1. Connecting The Calibration Adapter ............................................................ 44
13.2.2. Disconnecting The Calibration Adapter ....................................................... 45
13.3. Zero Calibration ........................................................................................... 45
13.4. Span Calibration........................................................................................... 46
13.4.1. Changing Span Gas Concentration ............................................................... 47
14. Diagnostic Mode ......................................................................................................... 48
14.1. Entering Diagnostic Mode ........................................................................... 48
14.2. Navigating Diagnostic Mode ....................................................................... 48
14.3. Exiting Diagnostic Mode ............................................................................. 48
15. Alarm Signal Summary............................................................................................... 49
16. Downloading Datalogs And Performing PC-Based
Instrument Configuration and Firmware Upgrades .................................................... 50
17. Maintenance ................................................................................................................ 51
17.1. Replacing the ToxiRAE Pro PID Li-Ion Battery ......................................... 51
17.2. Proper Battery Disposal ............................................................................... 54
17.3. Removing And Installing The Rubber Boot ................................................ 55
17.4. Filter Replacement ....................................................................................... 56
17.5. Cleaning Or Replacing The PID .................................................................. 58
17.6. Replacing The Fan ....................................................................................... 60
17.7. Cleaning The Exterior .................................................................................. 62
17.7.1. Ordering Replacement Parts ......................................................................... 62
18. Troubleshooting .......................................................................................................... 63
19. Technical Support ....................................................................................................... 64
20. RAE Systems Contacts ............................................................................................... 64
21. Appendix A: Regulatory Information ......................................................................... 67
2
ToxiRAE Pro PID User’s Guide
WARNINGS
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions.
CAUTION!
Never operate the monitor when the cover is removed. Remove the monitor top cap, rear cover, fan, or battery only in an area known to be non-hazardous.
Product Registration
Register your product online by visiting: http://www.raesystems.com/support/product-registration By registering your product, you can:
Receive notification of product upgrades or enhancements Be alerted to Training classes in your area Take advantage of RAE Systems special offers and promotions
3
ToxiRAE Pro PID User’s Guide
Use only RAE Systems lithium-ion battery part number G02-3004-000 (3.7V, 1,800mAh). This instrument has not been tested in an explosive gas/air atmosphere having an oxygen concentration greater than 21%. Substitution of components may impair suitability for intrinsic safety. Replace batteries only in non-hazardous locations.
STATIC HAZARD: Clean only with a damp cloth to avoid damage from electrostatic discharge.
For safety reasons this equipment must be operated and serviced by qualified personnel only. Read and understand this instruction manual completely before operating or servicing.
The monitor should be calibrated every time it does not pass a bump test, but no less frequently than every six months, depending on use and exposure to gas and contamination, and its operational mode.
The ToxiRAE Pro PID monitor is factory-preconfigured to a 30-day calibration interval, so it will go into calibration alarm if it is not calibrated for 30 days or more.
• Calibration intervals and bump test procedures may vary due to national legislation.
• RAE Systems recommends using RAE calibration gas cylinders with one of the
following:
Isobutylene, 34L, 100 ppm (balance air), PN 600-0002-000 Isobutylene, 100 ppm 103L steel cylinder, PN 600-0002-001 Isobutylene Calibration Gas, 100 ppm, 34L cylinder; constant-flow CGA-600
regulator with adapter, PN 002-3025-000
Isobutylene, 100 ppm (balance air), 34L cylinder, Dry Air, 0 ppm, 34L cylinder;
Constant-flow DGA-600 regulator with adapter, PN 007-3001-000
WARNINGS
4
ToxiRAE Pro PID User’s Guide
Utiliser seulement la batterie Lithium-ion RAE Systems référence G02-3004-000 (3.7V, 1,800mAh). Cet instrument n’a pas été testé dans une atmosphère de gaz/air explosive
ayant une concentration d’oxygène plus élevée que 21%. La substitution de composants
peut compromettre la sécurité intrinsèque. Ne changer les batteries que dans un emplacement désigné non dangereux.
RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer uniquement avec un chiffon humide.
Pour des raisons de sécurité, cet équipement doit être utilisé et entretenu uniquement par un personnel qualifié. Étudier le manuel d’instructions en entier avant d’utiliser, ou d’entretenir l’équipement.
Le ToxiRAE Pro PID est configuré en usine avec une période de calibration de 30 jours, si il n’est pas calibré durant cette période , il passe en alarme.
Une vérification d’étalonnage est définie par une exposition du détecteur au gaz
d’étalonnage qui doit déclencher le seuil d’alarmes bas.
Les périodes de calibration peuvent varier suivant les législations locales.
RAE Systems préconise d’utiliser les gaz de calibration suivants:
Isobutylene, 34L, 100 ppm (balance air), PN 600-0002-000 Isobutylene, 100 ppm 103L Cylindre acier, PN 600-0002-001 Isobutylene Calibration, 100 ppm, 34L cylindre; débit constant CGA-600
regulateur avec adaptateur, PN 002-3025-000
Isobutylene, 100 ppm (balance air), 34L cylindre, Air sec, 0 ppm, 34L cylindre;
débit constant DGA-600 regulateur avec adaptateur, PN 007-3001-000
AVERTISSEMENT
5
ToxiRAE Pro PID User’s Guide
1
Wireless functionality is currently offered in 2011 in North America and Europe. Contact RAE Systems for availability in other regions.
2
Available with Industrial Hygiene configurations only.
3
Contact RAE Systems for availability.
4
Wireless versions have a functioning wireless modem and are ready for wireless deployments. “Non­wireless” versions cannot be upgraded to wireless in the future. Additional equipment and/or software
licenses may be required to enable remote wireless monitoring and alarm transmission.
1. General Information
ToxiRAE Pro PID is a personal wireless monitor for volatile organic compounds (VOCs), the world’s smallest. Featuring RAE Systems’ next-generation PID sensor, the ToxiRAE Pro PID can promptly detect and accurately monitor over 300 VOCs. With an onboard library of 190 correction factors, the ToxiRAE Pro PID can be programmed to automatically read in concentrations of the specified compound.
Key Features
Wireless1. Personal. Proven
• Wireless access to real-time instrument readings and alarm status from any location
• Unmistakable five-way local and remote wireless notification of alarm conditions
• Cutting-edge PID with 190 built-in correction factors
• Largest display in its class
• Reliable, rugged, and intrinsically safe
• Continuous datalogging2 capability
• Fully automated bump testing and calibration with AutoRAE 2 3
• Easy to maintain with replaceable sensor, fan, and Lithium-ion battery
Available in Industrial Hygiene (advanced) or Safety (basic)
• Wireless or non-wireless versions4 are available
3
configurations
6
ToxiRAE Pro PID User’s Guide
ToxiRAE Pro PID Specifications
Dimensions
118 mm x 60 mm x 30 mm (4.6" x 2.4" x 1.2")
Weight
235g (8.3 oz)
Battery
Internal rechargeable Li-ion
Charging Time
Less than 4 hours
Operating Time
More than 12 hours on full charge (normal operation)
Charger
Charging and PC communication cradle: AC / DC adapter with international prong kit (100 to 240V, 50/60 Hz AC to 12V DC)
Multi-unit charging station for up to 5 units (data download from one): AC/DC adapter for the multi-unit charging station
Alarms
Audio alarm – 95 dB @ 30cm Visual alarm – 4 bright red LED lights Vibration alarm Activated display backlighting and on-screen indication of alarm conditions Wireless alarm notification sent to remote observers (wireless versions only)
Display
Monochrome dot-matrix graphical LCD display
Display Backlight
Manually and on alarm
Keypad
2 programming/operation keys
Direct Readout
Real-time reading of VOC concentrations in parts per million or mg/m3; battery status; data logging on/off status; wireless reception quality; STEL, TWA, and peak values; calibration and measurement gases, correction factors
Sampling Method
Fan-assisted diffusion
Datalogging
Capacity: 3 months at 1-minute interval Datalog interval: user-configurable from 1 to 3,600 seconds
Calibration
Two-point field calibration of zero and standard reference gas Adjustable span value Maintains calibration date
IP Rating
IP-54
Docking Station
AutoRAE 2 Automatic Test and Calibration Station
Protection:
Password-protected calibration settings, alarm limits, and data
Hazardous
Location
Approvals
Class I, Div. 1; Gr. A,B,C,D T4
Class I, Zone 0 A/Exia IIC T4 ATEX: II 1G,
Ex ia IIC Ga T4 IECEx: Ex ia IIC Ga T4 China: Ex ia IIC T4
CE Compliance
(European
Conformity)
EMC directive: 2004/108/EC R&TTE directive: 1999/5/EC ATEX directive: 94/9/EC
Operating
Temperature
-4° to 131° F (-20° to 55° C) Humidity
0 to 95% relative humidity (non-condensing)
Warranty
Two years on non-consumable components (per RAE Systems Standard Warranty); one year on fan, sensor, battery, and other consumable components
2. Specifications
7
ToxiRAE Pro PID User’s Guide
[Y/+] key
[MODE] key
Gas inlet Display
LEDs
Alarm Buzzer
Belt clip (on back)
LED
Top cap
3. Overview
The ToxiRAE Pro PID is a wireless, personal volatile organic compounds (VOC) monitor, the worlds smallest. The ToxiRAE Pro PID takes worker protection to the next level by providing safety professionals wireless access to real-time instrument readings and alarm status from any location for better visibility and faster response. The ToxiRAE Pro PID features RAE Systems next-generation PID sensor that delivers fast, reliable detection and accurate monitoring of a wide variety of VOCs in a broad range of temperature and humidity conditions.
4. User Interface
The ToxiRAE Pro PID’s user interface consists of the display, four LEDs, an alarm buzzer, and two keys, [MODE] and [Y/+]. The LCD displays real-time reading in specified measurement units, sensor type, alarm type (when in alarm, including cal. overdue), battery status, datalog (if on), and radio and connection quality (if available).
8
ToxiRAE Pro PID User’s Guide
Icon
Function
Wireless status: the radio is on
Wireless status: the radio is off
The instrument is not equipped with a radio (no icon)
Wireless strength (0 to 5 bars), only when radio is turned on
Datalogging status (shown when datalogging is on, blank when off)
Battery status (three segments show battery charge level)
Man Down alarm enabled
Sensor calibration due
Bump test due
All sensors tested and calibrated tick mark (all sensors have been bump tested and calibrated; no sensor is overdue for a bump test or calibration according to the intervals configured on the instrument
Calibration and bump test up to date
Sensor name and “Alarm” are
displayed alternately when in alarm
Wireless on/off status
Wireless signal strength
Unit of measure
Soft keys (functions change by activity)
Alarm type (when in alarm)
Bump test due
Sensor calibration due
Reading
Man Down alarm on
Datalog on
Battery status
4.1. Display Overview
The LCD display provides visual feedback that includes the sensor types, readings, time, battery condition, and other functions.
4.2. Status Indicator Icons
Along the top of most screens are status indicators that tell you whether a function is operating and/or its strength or level.
9
ToxiRAE Pro PID User’s Guide
Rear view of ToxiRAE Pro PID
ToxiRAE Pro Charging and PC Communications Cradle
Raised lip on rear of ToxiRAE Pro PID locks into cradle’s latch
Rear view of charging cradle
Charging status LED
Connect AC adapter to the charging cradle
Latch release
5. Charging The ToxiRAE Pro PID
Always fully charge the battery before using the ToxiRAE Pro PID. The ToxiRAE Pro PID’s Li-ion battery is charged by placing the ToxiRAE Pro PID in its cradle. Contacts on the bottom of the ToxiRAE Pro PID meet the cradle’s contact pins, transferring power.
Note: Before setting the ToxiRAE Pro PID into its charging cradle, visually inspect the contacts to make sure they are clean. If they are not, wipe them with a soft, dry cloth. Do not use solvents or cleaners.
10
ToxiRAE Pro PID User’s Guide
Press on latch
Lift ToxiRAE Pro PID out of charging cradle
Follow this procedure to charge the ToxiRAE Pro PID (the procedure is applicable to the instrument with or without the rubber boot installed):
1. Plug the AC/DC adapter into the ToxiRAE Pro PID’s cradle.
2. Plug the AC/DC adapter into the wall outlet.
3. Place the ToxiRAE Pro PID into the cradle and press down until it is locked in place.
The ToxiRAE Pro PID begins charging automatically. The LED in the cradle should glow red to indicate charging. When charging is complete, the LED in the cradle glows green.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge the instrument only in areas known to be non-hazardous. Remove and replace batteries only in areas known to be non-hazardous.
5.1. Removing The ToxiRAE Pro PID From The Charging
Cradle
To remove the ToxiRAE Pro PID from its charging cradle, press down on the latch release on the back of the charging cradle while slightly tilting the ToxiRAE Pro PID toward the front of the cradle, and lift it out.
5.2. Low Voltage Shutoff
When the battery’s charge falls below the shutoff voltage, the ToxiRAE Pro PID then turns itself off. Recharge the battery by placing the ToxiRAE Pro PID in its cradle (the cradle should be connected to the AC adapter).
11
ToxiRAE Pro PID User’s Guide
6. Turning The ToxiRAE Pro PID On
To turn the ToxiRAE Pro PID on, press and hold [MODE] for 3 seconds.
Caution: The alarm buzzer is very loud. During startup, you can mute most of the sound by holding a finger over the alarm buzzer port.
Note: Never put tape over the alarm buzzer port to permanently mute it or otherwise limit the loudness of the buzzer.
1. When the ToxiRAE Pro PID turns on, the buzzer, vibration alarm, and LEDs are
tested. The firmware version is shown in the display (for example, “V1.00” indicates firmware version 1.00).
2. A battery test is performed. If the battery has sufficient charge, then the next step is a
self test of the system.
Note: If the battery is almost fully discharged, the message “Battery too low! Needs charging. Powering off!” shows in the display and the ToxiRAE Pro PID shuts off.
You must charge the battery before you can use the instrument.
3. Self-testing is performed by the ToxiRAE Pro PID.
Note: If one of the following error messages is displayed, shut off the ToxiRAE Pro PID by pressing [MODE]:
Sensor error! Wrong sensor!
If any of the following error messages occurs, you can either shut off the instrument by pressing [MODE], or continue to the next startup test by pressing [Y/+]:
No fan or fan error! Real time clock error! Memory error! Sensor changed! Failed to turn on lamp!
For any of these error types, try restarting the instrument. If the error message appears again after five more attempts to restart the instrument, consult the distributor from whom the instrument was purchased for technical support.
12
ToxiRAE Pro PID User’s Guide
4. After self-testing, two start-up options are available for the ToxiRAE Pro PID:
Normal and Fast Start-up. The following information is displayed in sequence under Normal Start-up and skipped under Fast Start-up:
Sensor type Sensor serial number Date, time, and temperature Calibration gas Last calibration date Measurement gas Correction factor User mode Battery type, voltage, and shutoff voltage Alarm mode Buzzer enabled/disabled status Light enabled/disabled status Vibration alarm enabled/disabled status Alarm settings for High, Low, STEL, and TWA Datalog type and interval
5. Once the startup process is complete, a sensor warm-up screen is shown (with
countdown, if necessary). Then the main measurement screen is shown. The type of monitored hazard is displayed, with current reading. The ToxiRAE Pro PID is now ready for use.
7. Turning The ToxiRAE Pro PID Off
Press and hold [MODE]. In 2 seconds, a 5-second countdown to shutoff begins, accompanied by one buzzer beep per second. You must hold your finger on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the ToxiRAE Pro PID continues normal operation.
When you see “Unit Off,” release your finger from the [MODE] key. The ToxiRAE Pro PID is now off.
8. Testing The Alarms (Manually)
Under normal non-alarm conditions, the buzzer, vibration alarm, and LEDs can be tested at any time by pressing and holding [Y/+].
Important! If any of the alarms does not respond to this test, check the Alarm Settings. It is possible that any or all of the alarms have been turned off. If all the alarms are turned on, but one or more of them (buzzer, LED lights, or vibration alarm) does not respond to this test, do not use the instrument; contact your RAE Systems distributor for technical support.
13
ToxiRAE Pro PID User’s Guide
9. Low Battery Warnings
If, during operation, the battery’s charge is low, the battery icon in the display screen turns empty and blinks once every second. At the same time, the following notifications occur every minute: the buzzer beeps and LEDs blink once, followed by a vibration and a screen message Battery low! Shut off in X minutes.” The countdown to shutoff starts 10 minutes prior to automatic shutoff.
10. Dead Battery Warning & Shutoff
If you turn on the ToxiRAE Pro PID and the battery’s charge is below its lowest threshold, the display shows this message:
Battery too low! Needs charging. Powering off!
It then automatically shuts off. You must charge the ToxiRAE Pro PID’s battery before
you can continue to use this instrument.
Important! Do not operate the ToxiRAE Pro PID again until its battery has been charged.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, charge the battery only in areas known to be non-hazardous.
14
ToxiRAE Pro PID User’s Guide
11. Operating The ToxiRAE Pro PID
11.1. Overview
The ToxiRAE Pro PID has two operational modes:
Normal see page 15 for detailed instructions. Diagnostic see page 48 for detailed instructions.
The following is an overview of the two modes: Normal Mode is the default mode. It is accessed when you turn on the ToxiRAE Pro
PID. Two sub-modes are available: Basic User and Advanced User. In Basic User, some restrictions are applied, including password protection that guards against entering Programming Mode by unauthorized personnel. In Advanced, there are no such access restrictions (you do not need a password), and it provides the indications and data you need most for typical monitoring applications.
Diagnostic Mode is primarily intended for technicians to use during troubleshooting, although it also offers access to a few changeable parameters that you may rarely (if ever) need to change. A password is required to enter Diagnostic Mode (the password is the same as (and automatically kept in sync with) the one used to enter Programming Mode). The default password is 0000. To enter Diagnostic Mode, power the instrument off and then turn it back on by pressing and holding both the buttons. In Diagnostic Mode, ToxiRAE Pro PID displays readings in raw counts instead of units such as parts per million (ppm).
It is not possible to switch to Programming from Diagnostic Mode without turning off the instrument first and then restarting it in the desired mode.
11.1.1. Normal Mode
Normal Mode is the default mode of the ToxiRAE Pro PID when it is turned on. The selected sub-mode (Basic or Advanced) determines the number of screens that are accessible.
15
ToxiRAE Pro PID User’s Guide
11.1.2. Basic User Mode
When the ToxiRAE Pro PID is in Basic User Mode, pressing [MODE] allows you to step through the screens in the following fashion:
Note: The gray boxes indicate the datalog functions when datalogging is in Manual mode. When datalogging is in Automatic mode, these screens do not appear. Refer to page 28 for information on selecting Automatic or Manual datalogging.
16
ToxiRAE Pro PID User’s Guide
11.1.3. Advanced User Mode
Note: To enter Advanced User mode from Basic User mode, you must enter Programming mode on an instrument that is running in Basic User mode by simultaneously pressing and holding both the buttons for 3 seconds and supply a password.
To exit Programming Menu and return to the measurement screen (in Advanced User mode), press the [MODE] button repeatedly to highlight the Exit icon in Programming Menu and press Y to exit to the measurement screen.
When the ToxiRAE Pro PID is in Advanced User mode, pressing [MODE] steps through the same screens, but with the addition of a screen that allows for selection of different measurement gases from the internal library. Also, once in Advanced User mode, you may enter Programming without using a password.
Note: In Advanced User Mode, you may select different measurement gases from the internal library.
Note: The gray box indicates the datalog functions when datalogging is in Manual mode. When datalogging is in Automatic mode, these screens do not appear. Refer to page 28 for information on selecting Automatic or Manual datalogging.
11.2. Navigation In Normal Mode
Step through the screens in Normal Mode by pressing [MODE]. Some screens allow actions on your part, such as clearing the data. In these instances, pressing [Y/+] performs the action.
17
ToxiRAE Pro PID User’s Guide
11.2.1. TWA, STEL, Peak
The first screen includes a readout for TWA, STEL, and Peak readings.
TWA The TWA (time-weighted average) reading is the average reading of the gas
concentration times that the monitor has been on, prorated to an 8-hour period.
STEL Short Term Exposure Limit (STEL) data based on 15-minute STEL values,
expressed in ppm or mg/m3. A STEL value is only displayed after the ToxiRAE Pro PID has been on for 15 minutes or longer. After exiting PC Communications mode, the STEL is reset.
Peak The Peak value tells you the highest reading for the sensor since the ToxiRAE
Pro PID was turned on. Press [Y/+] twice to clear the Peak value.
Note: If you do not press a key within 1 minute, the display reverts to the
normal reading.
Press [MODE] to advance from screen to screen.
11.2.2. Date, Time, And Temperature
The current Date is expressed as Month/Day/Year, with two digits for month and day, and four digits for the year. Below the date is the time, which is expressed as Hours/Minutes/Seconds, with two digits for each. The time is in 24-hour (military) format. Below the time is the internal temperature, which can be in degrees Fahrenheit or Celsius (settable in Programming Mode).
11.2.3. Wireless Settings
The current wireless settings are shown here. These include the PAN ID, Channel, and Unit ID, which are helpful when creating or troubleshooting networks. Parent ID data does not apply.
11.2.4. Calibration Gas, Measurement Gas, Correction Factor
This screen shows calibration gas (Cal. Gas), measurement gas (Meas. Gas), and correction factor (CF).
18
ToxiRAE Pro PID User’s Guide
11.2.5. Enter Communications and stop measurement?
Before communicating with a PC running ProRAE Studio II Instrument Configuration and Data Management software (for downloading datalogs, configuring the instrument, or uploading new firmware), the ToxiRAE Pro PID must be set to communicate with the computer. Press [MODE] repeatedly until you reach the “Enter Communications and stop measurement?” screen. Then press [Y/+]. Measurement and datalogging are stopped, and the instrument is now ready to communicate with the PC.
11.3. Programming
You must provide a 4-digit password to enter Programming Mode, unless the ToxiRAE Pro PID is already running in Advanced User mode. The default password is 0000.
11.4. Entering Programming Mode
1. Press and hold [MODE] and [Y/+] simultaneously until the Password screen appears.
2. Input the 4-digit password:
Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again "wraps" around back to 0.
Step from digit to digit using [MODE]. When you are done, continue pressing [MODE] until OK or Cancel is highlighted. Press [Y/+] to register the password (if OK is highlighted) or to exit without entering
the password (if Cancel is selected).
If you make a mistake, you can cycle through the digits by pressing [MODE] repeatedly and then using [Y/+] to change the number in each position.
Note: The default password is 0000.
19
ToxiRAE Pro PID User’s Guide
When you have successfully entered Programming Mode, the following screen appears:
Note: The password can only be changed by connecting the instrument to a PC running ProRAE Studio II software. Follow the instructions in ProRAE Studio II to change it.
The Calibration icon is highlighted first by default, but you can press [MODE] to step from one programming menu item to the next, with the name of the menu shown at the top of the display and the corresponding icon highlighted. As you repeatedly press [MODE], the selection moves from left to right, and the following screens appear in sequence:
Note: When you reach Exit and press [MODE], the menu cycles back to Calibration.
To exit the Programming Menu, highlight Exit by pressing the [MODE] button, and press the [Y/+] key.
20
Loading...
+ 50 hidden pages