RAE Systems ToxiRAE Pro Quick Start Guide

ToxiRAE Pro
(Electrochemical Sensors)
QuickStart Guide
User Interface
The ToxiRAE Pro’s user interface consists of the display, four LED lights, an alarm buzzer, and two keys, [MODE] and [Y/+]. The LCD displays real-time reading in specified measurement units, sensor type, alarm type (when in alarm, including cal. or bump overdue), Man Down status, battery status, datalog (if on), and radio and connection quality (if available).
Charging The ToxiRAE Pro
Always fully charge the battery before use. Contacts on the bottom of the ToxiRAE Pro meet the cradle’s contact pins, transferring power. While charging, the LED on the cradle glows red. When the battery is fully charged, the LED glows green.
Turning The ToxiRAE Pro Off
Press and hold [MODE]. A 5-second countdown to shutoff begins. You must continue pressing on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the ToxiRAE Pro continues normal operation.
When the countdown beeps stop and you see “Unit Off,” release your finger from the [MODE] key.
The ToxiRAE Pro is now off.
Belt clip (on back)
Alarm Buzzer
LEDs Display
[MODE] key
[Y/+] key
2.
3.
5.
6.
4.
Gas inlet LED
Place ToxiRAE Pro in charging cradle
Connect AC adapter to charging cradle and plug it into AC power source
Latch release
Raised lip locks into cradle’s latch
A
B
Charging status LED
Phone: 408-952-8200 Fax: 408-952-8480 Email: customerserv@raesystems.com
www.raesystems.com
RAE Systems, Inc. 3775 N. First St. San Jose, CA 95134-1708 USA
Calibration
The ToxiRAE Pro can be automatically bump tested and calibrated using the AutoRAE 2 Test and Calibration Station (refer to its User’s Guide for instructions). Manually calibrate using a fixed-flow regulator (flow rate between 0.5 and 1.0 liters per minute) and the supplied special calibration adapter that covers the gas inlet:
1. To begin calibration, connect the zero air or
calibration gas cylinder, flow regulator, and calibration adapter to the ToxiRAE Pro.
2. Enter Programming Mode by pressing and holding
[MODE] and [Y/+] simultaneously until the password screen appears.
3. Input the 4-digit password. (The default password is
“0000.” If you do not know the password, highlight
and select “OK.”) Then follow the instructions for
zero calibration and span calibration.
Important! After a bump test or calibration, remove the calibration adapter to ensure correct readings.
Manual Span Calibration
In Programming Mode, and with “Span Calib” highlighted:
1. Press [Y/+]. The screen displays the current (or
default) calibration gas and its concentration.
· Press [MODE] if you do not want to change it.
· Press [Y/+] if you do want to change it.
2. Attach the calibration adapter, and connect the
calibration gas cylinder and flow regulator to the ToxiRAE Pro, and start the gas flow.
3. Press [Y/+] to start calibration.
4. Upon completion, "Span is done!" and the reading
are shown. Make sure it is within ±10% of the span gas value.
5. Remove the calibration adapter.
Bump (Functional) Testing
RAE Systems recommends periodic bump testing to confirm that the sensor is working and that the alarms are functional.
With the ToxiRAE Pro in Normal Mode:
1. Connect the calibration gas cylinder, flow
regulator, and calibration adapter to the ToxiRAE Pro and start the gas flow.
2. Make sure the unit goes into alarm and that the
buzzer produces at least two beeps per second, the LED lights flash on and off, and the vibration alarm functions. The display backlight should illuminate and an alarm message should be shown in the display.
3. Turn off the gas flow.
4. Remove the calibration adapter.
8.
9.
10.
P/N: G02-4012-000 Rev C 201303
Menu Navigation In Normal Mode
Pressing [MODE] repeatedly allows you to step through the screens as shown here.
Note: The gray boxes indicate the datalog functions when datalogging is in Manual mode. When datalogging is in Automatic mode, these screens do not appear. Refer to the User’s Guide for information on selecting Automatic or Manual datalogging.
11.
7.
* If your ToxiRAE Pro is configured with an oxygen (O2) sensor, TWA and STEL are not included in the menus because neither TWA nor STEL are calculated for this sensor. Min (for minimum) is included only when an oxygen sensor is used.