RAE Systems QRAE II Quick Start Guide

QRAE II Quick Reference
Covers Diffusion & Pump Models
with Firmware Version 3.12 and higher
WARNINGS
STATIC HAZARD: Clean only with a damp cloth. For safety reasons this equipment must be operated and serviced by qualified personnel only. Read and
understand instruction manual completely before operating or servicing. Only the combustible gas detection portion of this instrument has been assessed for performance.
All newly purchased RAE Systems instruments should be bump tested by exposing the sensor(s) to known concentration calibration gas before the instrument is put into service, and later, prior to each use. A bump test is defined as a brief exposure of the monitor to the calibration gas and the sensors to show response and trigger the lowest alarm set point for each sensor.
The QRAEII Multi Gas detector must be calibrated if it does not pass a Bump Test, or at least once
every 180 days, depending on use and sensor exposure to poisons and contaminants.
Calibration intervals and bump test procedures may vary due to national legislation.
When using an H2S sensor in a QRAE II, RAE Systems recommends using RAE calibration gas
cylinders with a 4 gas mix containing 10 ppm H2S, 50 ppm CO, 50% LEL Methane, and 18% Oxygen.
Any rapid up-scale reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit which may be hazardous.
Important
This Quick Reference provides operational information for the QRAE II. It is
not intended as a substitute for the User’s Guide. The User’s Guide must be
carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the
manufacturer’s instructions.
Standard Contents
QRAE II Calibration Adapter (diffusion model only) Water Trap Filter (pump model only) Quick Start Guide Alkaline Battery Adapter CD-ROM with User's Guide and related materials, ProRAE Studio Software Package for Windows 2000, NT and XP (Rechargeable version only) Charging/download cradle with computer interface cable (Rechargeable version only) AC/DC Adapter with international plugs (Rechargeable version only) Lithium-Ion battery pack
1
Setting Up The Charging Cradle
1. Remove the plastic spacer piece from adapter’s receptacle.
2. Select the correct plug for your region.
3. Tilt the plug into the adapter’s receptacle,
aligning the small ledge with the cutout inside the receptacle.
4. Connect the adapter’s plug to the receptacle on the back of the cradle.
5. If you need to remove the plug, slide the capture plate to release the plug, and tilt out the plug from the receptacle.
Before charging the rechargeable Li-ion battery, you must set up the charging cradle. Its AC/DC adapter is designed for use anywhere in the world and is able to handle voltages between 100 VAC and 240 VAC at either 50 Hz or 60 Hz. It comes with several plugs that snap into place on the base of the AC/DC adapter.
2
Cradle
Charging LED
Charging The Battery
Always charge the QRAE II before use. To charge the QRAE II:
1. Plug the AC/DC adapter into the QRAE II’s cradle.
2. Plug the AC/DC adapter into the wall outlet.
3. Place the QRAE II into the cradle and press down until the charging LED glows.
Red charging LED: Unit charging Green charging LED: Unit charged
Note: A full recharge of the Li-ion battery typically requires six hours.
Turning The QRAE II On
Hold down [MODE] for 2 seconds. When starting up, the QRAE II simultaneously turns the backlight on and off, beeps once, blinks once, and vibrates. It performs diagnostics and tells you current settings.
Turning The QRAE II Off
Press and hold [MODE]. In 2 seconds, a 5-second countdown to shutoff begins. You must hold your finger on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the QRAE II continues normal operation.