RAE Systems AutoRAE Lite Quick Start Guide

AutoRAE Lite
AutoRAE Lite
Quick Reference
CALIBRATION LED REFERENCE
• Make sure calibration gas cylinder is attached and pressure gauge reads ≥50 psi.
• Make sure the power LED is glowing.
inlet is not blocked.
1. Turn on the ToxiRAE 3: Press and hold the
MODE ( ) button.
2. After the ToxiRAE 3 has started and displays
a reading, press the Y/+ button until its display shows “Ir.”
3. Firmly press the ToxiRAE 3
face-down into the dock on the AutoRAE Lite.
4. When both the Bump and Cal. LEDs glow green, press
“Bump” to perform a bump test, or press “Cal.” to perform a calibration.
Cal.Bump Cal.Bump
or
Quick Reference
Status LED explanation
LED Red Green Solid Blinking Orange Orange
Power Battery low, External - - - - - ­ no external power or power
Bump Fail Pass - - - Processing Cal Fail Pass - - - Processing Service No SD card, - - - - - - SD card is
Request or commu- locked or nication needs to be failure, or formatted, or SD card SD card space is space is very low very low
SET THE TIME
Set the AutoRAE Lite’s time before its first use, and whenever the AutoRAE Lite has not been used for more than 10 days. Refer to the AutoRAE LiteUser’s Guide for instructions.
WARNING
Only use this instrument in non-hazardous locations.
battery OK
During a bump test or calibration, the status LED blinks and is orange.
PASS: A green LED indicates the bump test or calibration was successful.
FAIL: A red LED indicates the bump test or calibration failed.
SIDE A SIDE B
Finished size should be 85mm x 170mm.
Colors: Black and red.
For further information on use and maintenance, refer to the AutoRAE Lite User’s Guide.
Tech Support Email: tech@raesystems.com Tech Support Phone: +1 408-952-8461
RAE Systems by Honeywell 3775 N. First St., San Jose, CA 95134 USA www.raesystems.com
Rev. B Feb 2008 P/N T01-0901-000D