AutoRAE Lite
AutoRAE Lite
Quick Reference
CALIBRATION LED REFERENCE
• Make sure calibration gas cylinder is attached
and pressure gauge reads ≥50 psi.
• Make sure the power LED is glowing.
• Make sure a filter is attached and that the air
inlet is not blocked.
1. Turn on the ToxiRAE 3: Press and hold the
MODE ( ) button.
2. After the ToxiRAE 3 has started and displays
a reading, press the Y/+ button until its
display shows “Ir.”
3. Firmly press the ToxiRAE 3
face-down into the dock on
the AutoRAE Lite.
4. When both the Bump and
Cal. LEDs glow green, press
“Bump” to perform a bump
test, or press “Cal.” to perform
a calibration.
Cal.Bump Cal.Bump
or
Quick Reference
Status LED explanation
LED Red Green Solid Blinking
Orange Orange
Power Battery low, External - - - - - no external power or
power
Bump Fail Pass - - - Processing
Cal Fail Pass - - - Processing
Service No SD card, - - - - - - SD card is
Request or commu- locked or
nication needs to be
failure, or formatted, or
SD card SD card
space is space is
very low very low
SET THE TIME
Set the AutoRAE Lite’s time before its first use,
and whenever the AutoRAE Lite has not been
used for more than 10 days. Refer to the
AutoRAE LiteUser’s Guide for instructions.
WARNING
Only use this instrument in non-hazardous locations.
battery OK
During a bump test or calibration, the status LED
blinks and is orange.
PASS: A green LED indicates the bump test or
calibration was successful.
FAIL: A red LED indicates the bump test or
calibration failed.
SIDE A SIDE B
Finished size should be 85mm x 170mm.
Colors: Black and red.
For further information on use and maintenance,
refer to the AutoRAE Lite User’s Guide.