RAE Systems AutoRAE 2 Quick Start Guide

by Honeywell
Setup
Exhaust
Calibration Cylinder 1
Calibration
Cylinder 2
Press filter onto end of Cradle
Plug barrel from AC
Install left and right end caps
Connect gas cylinders with gas inlets
The ToxiRAE Pro Cradle
mechanism. ToxiRAE Pro
and ToxiRAE Pro PID (R). MultiRAE
ToxiRAE Pro
QRAE 3
Before using the AutoRAE 2 Cradle to bump test or calibrate ToxiRAE Pro-family and/or QRAE 3 and/or MultiRAE-family (pumped versions) instruments, make sure the AutoRAE 2 Cradle and (if one is used) the AutoRAE 2 Controller is running latest firmware. Also, follow the instructions in the AutoRAE 2 User’s Guide for programming the gas configurations. Note: Each calibration gas cylinder must have a demand-flow regulator.
Connections
Gas
Gas
adapter into jack, then AC Adapter to AC power source
IMPORTANT!
Always check that the gas configuration and the type/concentration of the calibration gases connected to the AutoRAE 2 Cradle match before you begin any bump test or calibration. Make sure there is enough gas in the cylinder(s). Set the RTC (real-time clock) date and time on the Cradle via ProRAE Studio II before first use.
receptacle at
Placing An Instrument In The Cradle
Important! If the Cradle is for a ToxiRAE Pro-family instrument, make sure the correct calibration adapter is in place.
1. Make sure the filter on the instrument is not dirty or clogged.
2. Make sure the instrument is turned off or in AutoRAE 2 Mode.
3. Place the instrument into the cradle face-down, making check that it is aligned correctly with the contacts on the
AutoRAE 2 Cradle’s charging port.
4. Press in on the capture mechanism to lock the instrument in place.
requires an adapter to be inserted in the capture
and ToxiRAE Pro LEL (L)
by Honeywell
Automatic Warm-Up Before Bump Testing Or Calibrating
When you place an instrument in the cradle and lock the capture mechanism, the instrument begins charging and is automatically warmed up. Warm-up time depends on the sensors installed in the instrument and their individual warm-up requirements.
Performing A Bump Test
1. Insert an instrument in the AutoRAE 2 Cradle.
2. Press “Bump.” If the instrument is warmed up (both LEDs are glowing green), the bump test begins.
Consult the Bump & Cal Error & Status chart for an explanation of the LED indications that tell you which operations are underway and which LED indications tell you of errors during calibration.
Note: If the instrument does not successfully pass a bump test, the AutoRAE 2 Cradle automatically begins a full calibration.
Performing A Calibration
Always perform a full calibration after replacing a sensor, using a new ToxiRAE Pro, QRAE 3, or MultiRAE, or if the instrument has been unused for a long period of time. In addition, if the instrument has failed a bump test, perform a full calibration.
1. Make sure the span value in the monitor(s) and the AutoRAE 2 Controller match.
2. Make sure the correct Gas Configuration has been selected (use G8; G1 through G7 are not supported).
3. Insert a ToxiRAE Pro, QRAE 3, or MultiRAE monitor in the AutoRAE 2 Cradle.
4. Press “Cal.”
When the instrument is warmed up (both LEDs are glowing green), calibration begins.
Note: For MultiRAE and QRAE 3 instruments, calibration order is important. Refer to cross-interference information in section 21.2 (Connecting Calibration Gas) in the AutoRAE 2 User’s Guide and RAE Systems Technical Note TN-114: Sensor Specifications And Cross-Sensitivities (available for free download from www.raesystems.com).
by Honeywell
Bump And Cal Errors And Status Messages
Status
Bump LED
Cal LED
User Action
Warm-up in progress
Orange (blinking in alternation)
Do not remove the instrument from the cradle, or the warm-up process will be interrupted. Pressing keys has no effect.
Warm-up error
Red (blinking in alternation)
Remove the instrument from the cradle and take the action indicated on the instrument’s display.
Warm-up completed successfully
Green
Green
Press Bump or Cal key to perform a bump test or calibration.
Bump test in progress
Green (blinking)
Off
Do not remove the instrument, or the process will be interrupted. Pressing keys has no effect.
Bump test completed successfully
Green
Off
The bump test result (pass) has been logged. You may remove the instrument from the Cradle for use or leave it on the Cradle for the next AutoRAE 2 operation or to charge its battery.
Bump test failed
Red (blinking slowly)
Off
The bump test result (fail) has been logged. The AutoRAE 2 will perform calibration on failed sensor(s) automatically.
Calibration in progress
Off
Green (blinking)
Do not remove the instrument, or the process will be interrupted. Pressing keys has no effect.
Calibration completed successfully
Off
Green
The result has been logged in the instrument. You may remove the instrument from the Cradle.
Calibration failed
Off
Red (blinking slowly)
The result has been logged in the instrument; you may remove it from the Cradle and read its display for an error code.
Sleep mode
Orange
Orange
Charging continues when in sleep mode. Press either key to wake up the instrument.
Monitor not connected / system idle
Off
Off
Check to make sure the monitor is properly installed in the cradle and the capture mechanism is fully engaged. Check to make sure the monitor is in AutoRAE 2 Communications Mode or turned off.
Cradle error
Red (blinking slowly)
Take the action indicated in the Controller’s display.
Otherwise, contact RAE Systems service.
CORPORATE HEADQUARTERS
RAE Systems by Honeywell 3775 N. First St. San Jose, CA 95134 USA Phone: 408.952.8200 Fax: 408.952.8480 customerserv@raesystems.com
WORLDWIDE SALES OFFICES
USA/Canada: 1.877.723.2878 Europe: +45.86.52.51.55 Middle East: +00971.4.440.5949 China: +86.10.5885.8788-3000 Asia Pacific: +852.2669.0828
by Honeywell
Removing An instrument From The Cradle
Printer Port
Serial port on
Serial
Rev C April 2014 P/N T02-4014-000
MultiRAE
ToxiRAE Pro
QRAE 3
To remove an instrument from the cradle, press the button to release the capture mechanism. Then lift the instrument from its inlet end.
Charging An Instrument’s Battery
The AutoRAE 2 Cradle automatically charges an instrument’s battery when it is docked. The Charge Status LED
glows red to indicate that charging is taking place. When the instrument’s battery is fully charged, the LED glows
green.
Printing Bump/Calibration Reports
The AutoRAE Cradle has a port for connecting to a serial printer (or you can use a serial-to-USB adapter to drive a compatible USB printer).
Connect the serial port of the AutoRAE Cradle to the serial port on a printer.
Cable
printer
When you perform a bump test or calibration with a printer attached to an AutoRAE 2 Cradle, a full report is sent to the printer after the bump test or calibration concludes. If a bump test fails, the AutoRAE 2 Cradle sends the report to the printer and it initiates a full calibration. When the calibration is complete, the results are then also sent automatically to the printer.
Loading...