20. Technical Support ....................................................................................................... 60
21. RAE Systems Contacts ............................................................................................... 60
22. Appendix A: Regulatory Information ......................................................................... 63
2
ToxiRAE Pro User’s Guide
WARNINGS
Read Before Operating
This manual must be carefully read by all individuals who have or will have the
responsibility of using, maintaining, or servicing this product. The product will perform
as designed only if it is used, maintained, and serviced in accordance with the
manufacturer’s instructions.
CAUTION!
Never operate the monitor when the cover is removed. Remove the monitor top cap, rear
cover, or battery only in an area known to be non-hazardous.
Product Registration
Register your product online by visiting:
http://www.raesystems.com/support/product-registration
By registering your product, you can:
• Receive notification of product upgrades or enhancements
• Be alerted to Training classes in your area
• Take advantage of RAE Systems special offers and promotions
3
ToxiRAE Pro User’s Guide
Use only RAE Systems lithium-ion battery part number G02-3004-000 (3.7V,
WARNINGS
1,800mAh). This instrument has not been tested in an explosive gas/air atmosphere
having an oxygen concentration greater than 21%. Substitution of components may
impair suitability for intrinsic safety. Replace batteries only in non-hazardous locations.
STATIC HAZARD: Clean only with a damp cloth to avoid damage from electrostatic
discharge.
For safety reasons this equipment must be operated and serviced by qualified personnel
only. Read and understand this instruction manual completely before operating or
servicing.
The monitor should be calibrated every time it does not pass a bump test, but no less
frequently than every six months, depending on use and exposure to gas and
contamination, and its operational mode.
The ToxiRAE Pro monitor is factory-preconfigured to a 30-day calibration interval, so it
will go into calibration alarm if it is not calibrated for 30 days or more.
• Calibration intervals and bump test procedures may vary due to national legislation.
• RAE Systems recommends using RAE calibration gas cylinders containing the gas that
is appropriate to the sensor you are using, and in the correct concentration.
4
ToxiRAE Pro User’s Guide
Utiliser seulement la batterie Lithium-ion RAE Systems référence G02-3004-000 (3.7V,
AVERTISSEMENT
1,800mAh). Cet instrument n’a pas été testé dans une atmosphère de gaz/air explosive
ayant une concentration d’oxygène plus élevée que 21%. La substitution de composants
peut compromettre la sécurité intrinsèque. Ne changer les batteries que dans un
emplacement désigné non dangereux.
RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer uniquement avec un chiffon
humide.
Pour des raisons de sécurité, cet équipement doit être utilisé et entretenu uniquement par
un personnel qualifié. Étudier le manuel d’instructions en entier avant d’utiliser, ou
d’entretenir l’équipement.
Le ToxiRAE Pro est configuré en usine avec une période de calibration de 30 jours, si il
n’est pas calibré durant cette période , il passe en alarme.
• Une vérification d’étalonnage est définie par une exposition du détecteur au gaz
d’étalonnage qui doit déclencher le seuil d’alarmes bas.
• Les périodes de calibration peuvent varier suivant les législations locales.
• RAE Systems recommande l’utilisation de gaz étalons de RAE Systems qui
contiennent le gaz approprié au capteur que vous utilisez et avec une concentration
adéquate
5
ToxiRAE Pro User’s Guide
1. General Information
ToxiRAE Pro is a personal wireless monitor for toxic gases and oxygen
deficiency/enrichment. The ToxiRAE Pro can detect a broad range of gases—from
carbon monoxide, hydrogen sulfide, and oxygen to ammonia, chlorine, and phosgene.
The ToxiRAE Pro’s intelligent interchangeable sensors store calibration data, so if you
need to monitor one gas one day and another gas the next, swapping the sensors in your
ToxiRAE Pro is all it takes4.
Key Features
Wireless1. Personal. Proven
• Wireless access to real-time instrument readings and alarm status from any location
• Unmistakable five-way l ocal and remote wireless notification of alarm conditions
• Largest display in its class
• Reliable, rugged, and intrinsically safe
• Continuous datalogging capability
• Fully automated bump testing and calibration with AutoRAE 22
• Easy to maintain with replaceable sensor and Lithium-ion battery
• Wireless or non-wireless versions
3
are available
1
Wireless functionality is currently offered in North America and Europe. Contact RAE Systems for
availability in other regions.
2
Contact RAE Systems for availability.
3
Wireless versions have a functioning wireless modem and are ready for wireless deployments. “Nonwireless” versions cannot be upgraded to wireless in the future. Additional equipment and/or software
licenses may be required to enable r e mote wireless monitoring and alarm transmission.
6
2. Specifications
ToxiRAE Pro Specifications
Dimensions
118 mm x 60 mm x 30 mm (4.6" x 2.4" x 1.2")
Weight
220g (7.76 oz)
Sensor
Over 20 intelligent interchangeable field-replaceable electrochemical sensors
Battery
Rechargeable Li-ion battery
Display
Graphical LCD display with white LED b a c klighting (activated when monitor
Display Readout
Real-time reading of concentrations in parts per million; battery status;
Keypad
2 buttons for operation and programming
Sampling
Diffusion
Calibration
Automatic with AutoRAE 2 Test and Calib ration Station or manual
• Continuous Datalogging with a three-month capacity (at one-minute
Communication
• Data download and instrument set-up on PC via charging and PC
Wireless Network
RAE Systems Dedicated Wireless Network
Wireless
Frequency
ISM license-free bands
Wireless Range
(Typical)
300 feet (100 meters)
Operating
Temperature
-4° to 131° F (-20° to 55° C)
Humidity
0 to 95% relative humidity (non-condensing)
Dust and Water
IP-65 rating for dust and water ingress protection (validated by an independent
US and Canada
Class I, Zone 0 A/Exia IIC T4
Europe
IECEx: Ex ia IIC Ga T4
China
Ex ia IIC T4
CE Compliance
EMC directive: 2004/108/EC
Warranty
• Two years on non-consumable components and CO, H2S, and O2 sensors
Specifications are subject to change.
• Operating time > 30 hours (normal operation)
• Recharge time < 4 hours through charging cradle
is in alarm or with a button press)
datalogging on/off status; wireless on/off and reception quality; STEL, TWA,
and peak values (as applicable)
alarm (flashing bright red LEDs), and on-screen indication of alarm conditions
• Additional diagnostic alarm and d isplay message for low battery
ToxiRAE Pro User’s Guide
and Data
Download
Resistance
Hazardous
Location
Approvals
(European
Conformity)
intervals)
• User -configurable d atalogging interval (from 1 to 3,600 seconds)
communication cradle
• Data download via AutoRAE 2 Automated Test and Calibration Station
• Wireless data and status transmission via built-in RF modem (optional)
test laboratory)
Class I, Div. 1; Groups A,B,C,D T4
ATEX: II 1G, Ex ia IIC Ga T4
R&TTE directive: 1999/5 /E C
ATEX directive: 94/9/EC
• One year on all other sensors, battery, and consumable parts
7
ToxiRAE Pro User’s Guide
Sensor
Range
Resolution
Ammonia (NH3)
0 to 100 ppm
1 ppm
Carbon Monoxide (CO)
0 to 500 ppm
1 ppm
Carbon Monoxide (CO),
Extended Range
0 to 2,000 ppm
10 ppm
Carbon Monoxide (CO),
H2 Compensated*
0 to 2,000 ppm
10 ppm
Chlorine (Cl2)
0 to 50 ppm
0.1 ppm
Chlorine Dioxide (ClO2)
0 to 1 ppm
0.03 ppm
Ethylene Oxide (EtO-A)
0 to 100 ppm
1 ppm
Ethylene Oxide (EtO-B)
0 to 10 ppm
0.1 ppm
Ethylene Oxide (EtO-C),
Extended Range
0 to 500 ppm
10 ppm
Hydrogen (H2)*
0 to 1,000 ppm
2 ppm
Hydrogen Chloride (HCl)*
0 to 15 ppm
0.2 ppm
Hydrogen Cyanide (HCN)
0 to 50 ppm
0.5 ppm
Hydrogen Fluoride (HF)*
0 to 10 ppm
0.1 ppm
Hydrogen Sulfide (H2S)
0 to 100 ppm
0.1 ppm
Hydrogen Sulfide (H2S),
Extended Range
0 to 1,000 ppm
1 ppm
Methyl Mercaptan
(CH3-SH)
0 to 10 ppm
0.1 ppm
Nitric Oxide (NO)
0 to 250 ppm
0.5 ppm
Nitrogen Dioxide (NO2)
0 to 20 ppm
0.1 ppm
Oxygen (O2)
0 to 30% Vol.
0.1% Vol.
Phosgene (COCl2)*
0 to 1 ppm
0.02 ppm
Phosphine (PH3)
0 to 20 ppm
0.1 ppm
Phosphine (PH3),
Extended Range*
0 to 1,000 ppm
1 ppm
Sulfur Dioxide (SO2)
0 to 20 ppm
0.1 ppm
3. Available Electrochemical Sensors For ToxiRAE Pro
The following sensors are available for use in the ToxiRAE Pro.
Note: When replacing a sensor (even with the same type), the sensor may require
conditioning before first use, in addition to typical warm-up time prior to normal
operation. This can take from a few minutes to 24 hours, depending on the type of sensor.
Refer to RAE Systems Technical Note TN-151 for information on electrochemical sensor
replacement and maintenance.
* Check with RAE Systems for availability.
8
ToxiRAE Pro User’s Guide
[Y/+] key
[MODE] key
Gas inlet
Display
LEDs
Alarm
Buzzer
Belt clip (on back)
LED
Top cap
4. Overview
The ToxiRAE Pro is the world’s first wireless personal monitor for toxic gases and
oxygen. The ToxiRAE Pro takes worker protection to the next level by providing safety
professionals wireless access to real-time instrument readings and alarm status from any
location for better visibility and faster response. The ToxiRAE Pro can detect a broad
range of gases—from carbon monoxide, hydrogen sulfide, and oxygen to ammonia,
chlorine, and phosgene. The ToxiRAE Pro’s intelligent interchangeable sensors store
calibration data, so if you need to monitor one gas one day and another gas the next,
swapping the sensors in your ToxiRAE Pro is all it takes.
5. User Interface
The Toxi RAE Pro’s user interface consists of the display, four LEDs, an alarm buzzer,
and two keys, [MODE] and [Y/+]. The LCD displays real-time readings in specified
measurement units, sensor type, alarm type (when in alarm, including cal. overdue),
battery status, datalog (if on), and radio and connection quality (if available).
9
ToxiRAE Pro User’s Guide
Icon
Function
Wireless status: the radio is on
Wireless status: the radio is off
The instrument is not equipped with a radio (no icon)
Wireless strength (0 to 5 bars), only when radio is turned on
Datalogging status (shown when datalogging is on, blank when off)
Battery status (three segments show battery charge level)
Man Down alarm enabled
Sensor calibration due
Bump test due
All sensors tested and calibrated tick mark (all sensors have been bump tested
the intervals configured on the instrument
Calibration and bump
test up to date
Sensor name and “Alarm” are
displayed alternately when in alarm
Wireless on/off status
Wireless signal strength
Unit of measure
Soft keys
(functions change by activity)
Alarm type
(when in alarm)
Bump test due
Sensor
calibration due
Reading
Man Down alarm on
Datalog on
Battery status
5.1. Display Overview
The LCD display provides visual feedback that includes the sensor types, readings, time,
battery condition, and other functions.
5.2. Status Indicator Icons
Along the top of most screens are status indicators that tell you whether a function is
operating and/or its strength or level.
and calibrated; no sensor is overdue for a bump test or calibration according to
10
ToxiRAE Pro User’s Guide
Rear vi ew of
ToxiRAE Pro
ToxiRAE Pro
Charging and PC
Communications
Cradle
Raised lip on
rear of ToxiRAE Pro
locks into cradle’s latch
Rear vi ew of
charging cradle
Charging status
LED
Connect AC
adapter to the
charging cradle
Latch release
6. Charging The ToxiRAE Pro
Always fully charge the battery before using the ToxiRAE Pro. The ToxiRAE Pro’s Liion battery is charged by placing the ToxiRAE Pro in its cradle. Contacts on the bottom
of the ToxiRAE Pro meet the cradle’s contact pins, transferring power.
Note: Before setting the ToxiRAE Pro into its charging cradle, visually inspect the
contacts to make sure they are clean. If they are not, wipe them with a soft, dry cloth. Do
not use solvents or cleaners.
11
ToxiRAE Pro User’s Guide
Press on latch
Lift ToxiRAE Pro out of
charging cradle
Follow this procedure to charge the ToxiRAE Pro (the procedure is applicable to the
instrument with or without the rubber boot installed):
1. Plug the AC/DC adapter into the ToxiRAE Pro’s cradle.
2. Plug the AC/DC adapter into the wall outlet.
3. Place the ToxiRAE Pro into the cradle and press down until it is locked in place.
The ToxiRAE Pro begins charging automatically. The LED in the cradle should glow red
to indicate charging. When charging is complete, the LED in the cradle glows green.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, recharge the instrument
only in areas known to be non-hazardous. Remove and replace batteries only in
areas known to be non-hazardous.
6.1. Removing The ToxiRAE Pro From The Charging
Cradle
To remove the ToxiRAE Pro from its charging cradle, press down on the latch release on
the back of the charging cradle while slightly tilting the ToxiRAE Pro toward the front of
the cradle, and lift it out.
6.2. Low Voltage Shutoff
When the battery’s charge falls below the shutoff voltage, the ToxiRAE Pro then turns
itself off. Recharge the battery by placing the ToxiRAE Pro in its cradle (the cradle
should be connected to the AC adapter).
12
ToxiRAE Pro User’s Guide
7. Turning The ToxiRAE Pro On
To turn the ToxiRAE Pro on, press and hold [MODE] for 3 seconds.
Caution: The alarm buzzer is very loud. During startup, you can mute most of the sound
by holding a finger over the alarm buzzer port.
Note: Never put tape over the alarm buzzer port to permanently mute it or otherwise limit
the loudness of the buzzer.
1. When the ToxiRAE Pro turns on, the buzzer, vibration alarm, and LEDs are tested.
The firmware version is shown in the display (for example, “V1.00” indicates
firmware version 1.00).
2. A battery test is performed. If the battery has sufficient charge, then the next step is a
self test of the system.
Note: If the battery is almost fully discharged, the message “Battery too low! Needs
charging. Powering off!” shows in the display and the ToxiRAE Pro shuts off. You
must charge the battery before you can use the instrument.
3. Self-testing is performed by the ToxiRAE Pro.
Note: If one of the following error messages is displayed, shut off the ToxiRAE Pro
by pressing [MODE]:
• Sensor error!
• Wrong sensor!
If any of the following error messages occurs, you can either shut off the instrument
by pressing [MODE], or continue to the next startup test by pressing [Y/+]:
• Real time clock error!
• Memory error!
• Sensor changed!
For any of these error types, try restarting the instrument. If the error message appears
again after five more attempts to restart the instrument, consult the distributor from
whom the instrument was purchased for technical support.
13
ToxiRAE Pro User’s Guide
4. After self-testing, two start-up options are available for the ToxiRAE Pro: Normal
and Fast Start-up. The following information is displayed in sequence under Normal
Start-up and skipped under Fast Start-up:
• Sensor type
• Sensor serial number
• Date, time, and temperature
• Calibration gas
• Last calibration date
• Measurement gas
• User mode
• Battery type, voltage, and shutoff voltage
• Alarm mode
• Buzzer enabled/disabled status
• Light enabled/disabled status
• Vibration alarm enabled/disabled status
• Alarm settings for High, Low, STEL, and TWA (as applicable)
• Datalog type and interval
5. Once the startup process is complete, a sensor warm-up screen is shown (with
countdown, if necessary; the countdown time may vary for different sensors). Then
the main measurement screen is shown. The type of monitored hazard is displayed,
with the current reading. The ToxiRAE Pro is now ready for use.
8. Turning The ToxiRAE Pro Off
Press and hold [MODE]. In 2 seconds, a 5-second countdown to shutoff begins,
accompanied by one buzzer beep per second. You must hold your finger on the key for
the entire shutoff process. If you remove your finger from the key during the countdown,
the shutoff operation is canceled and the ToxiRAE Pro continues normal operation.
When you see “Unit Off,” release your finger from the [MODE] key. The ToxiRAE Pro
is now off.
9. Testing The Alarms (Manually)
Under normal non-alarm conditions, the buzzer, vibration alarm, and LEDs can be tested
at any time by pressing and holding [Y/+].
Important! If any of the alarms does not respond to this test, check the Alarm Settings. It is
possible that any or all of the alarms have been turned off. If all the alar ms are tu rne d on, but
one or more of them (buzzer, LED lights, or vibration alarm) does not respond to this test, do
not use the instrument; contact your RAE Systems distributor for technical support.
14
ToxiRAE Pro User’s Guide
10. Low Battery Warnings
If, during operation, the battery’s charge is low, the battery icon in the display screen
turns empty and blinks once every second. At the same time, the following notifications
occur every minute: the buzzer beeps and LEDs blink once, followed by a vibration and a
screen message “Battery low! Shut off in X minutes.” The countdown to shutoff starts 10
minutes prior to automatic shutoff.
11. Dead Battery Warning & Shutoff
If you turn on the ToxiRAE Pro and the battery’s charge is below its lowest threshold, the
display shows this message:
Battery too low!
Needs charging.
Powering off!
It then automatically shuts off. You must charge the ToxiRAE Pro’s battery before you
can continue to use this instrument.
Important! Do not operate the ToxiRAE Pro again until its battery has been charged.
WARNING!
To reduce the risk of ignition of hazardous atmospheres, charge the battery only in
areas known to be non-hazardous.
15
ToxiRAE Pro User’s Guide
12. Operating The ToxiRAE Pro
12.1. Overview
The ToxiRAE Pro has two operational modes:
• Normal — see page 16 for detailed instructions.
• Diagnostic — see page 48 for detailed instructions.
The following is an overview of the two modes:
Normal Mode is the default mode. It is accessed when you turn on the ToxiRAE Pro.
Two sub-modes are available: Basic User and Advanced User. In Basic User, some
restrictions are applied, including password protection that guards against entering
Programming Mode by unauthorized personnel. In Advanced, there are no such access
restrictions (you do not need a password), and it provides the indications and data you
need most for typical monitoring applications.
Diagnostic Mode is primarily intended for technicians to use during troubleshooting, alt hough it
also offers access to a few changeable parameters that you may rarely (if ever) need to change. A
password is required to enter Diagnostic Mode (the password is the same as (and
automatically kept in sync with) the one used to enter Programming Mode). The default
password is 0000. To enter Diagnostic Mode, power the instrument off and then turn it
back on by pressing and holding both the buttons. In Diagnost ic Mode, ToxiRAE Pro
displays readings in raw co unts instea d of uni ts such as parts per million ( ppm).
It is not possible to switch to Programming from Diagnostic Mode without turning off the
instrument first and then restarting it in the desired mode.
12.1.1. Normal Mode
Normal Mode is the default mode of the ToxiRAE Pro when it is turned on. The selected
sub-mode (Basic or Advanced) determines the number of screens that are accessible.
16
ToxiRAE Pro User’s Guide
12.1.2. Basic User Mode
When the ToxiRAE Pro is in Basic User Mode, pressing [MODE] allows you to step
through the screens in the following fashion:
* If your ToxiRAE Pro is configured with an oxygen (O
) sensor, TWA and STEL are
2
not included in the menus because neither TWA nor STEL are calculated for this sensor.
Min (for minimum) is included only when an oxygen sensor is used.
Note: The gray boxes indicate the datalog functions when datalogging is in Manual
mode. When datalogging is in Automatic mode, these screens do not appear. Refer to
page 27 for information on selecting Automatic or Manual datalogging.
17
ToxiRAE Pro User’s Guide
12.1.3. Advanced User Mode
Note: To enter Advanced User mode from Basic User mode, you must enter
Programming mode on an instrument that is running in Basic User mode by
simultaneously pressing and holding both buttons for 3 seconds and then supplying a
password.
To exit Programming Menu and return to the measurement screen (in Advanced User
mode), press the [MODE] button repeatedly to highlight the Exit icon in Programming
Menu and press Y to exit to the measurement screen.
When the ToxiRAE Pro is in Advanced User mode, you may enter Programming without
using a password.
* If your ToxiRAE Pro is configured with an oxygen (O2) sensor, TWA and STEL are
not included in the menus because neither TWA nor STEL are calculated for this sensor.
Min (for minimum) is included only when an oxygen sensor is used.
18
ToxiRAE Pro User’s Guide
12.2. Navigation In Normal Mode
Step through the screens in Normal Mode by pressing [MODE]. Some screens allow
actions on your part, such as clearing the data. In these instances, pressing [Y/+] performs
the action.
12.2.1. TWA, STEL, Peak
The first screen includes a readout for TWA, STEL, and Peak readings.
TWA*The TWA (time-weighted average) reading is the average reading of the gas
concentration times that the monitor has been on, prorated to an 8-hour period.
STEL*Short Term Exposure Limit (STEL) data based on 15-minute STEL values,
expressed in ppm or mg/m3. A STEL value is only displayed after the
ToxiRAE Pro has been on for 15 minutes or longer. After exiting PC
Communications mode, the STEL value is reset.
PeakThe Peak value tells you the highest reading for the sensor since the ToxiRAE
Pro was turned on. Press [Y/+] twice to clear the Peak value.
Note: If you do not press a key within 1 minute, the display reverts to the
normal reading.
* If your ToxiRAE Pro is configured with an oxygen (O2) sensor, TWA and STEL are
not included in the menus because neither TWA nor STEL are calculated for this sensor.
Press [MODE] to advance from screen to screen.
12.2.2. Date, Time, And Temperature
The current Date is expressed as Month/Day/Year, with two digits for month and day,
and four digits for the year. Below the date is the time, which is expressed as
Hours/Minutes/Seconds, with two digits for each. The time is in 24-hour (military)
format. Below the time is the internal temperature, which can be in degrees Fahrenheit or
Celsius (settable in Programming Mode).
12.2.3. Wireless Settings
The current wireless settings are shown here. These include the PAN ID, Channel, and
Unit ID, which are helpful when creating or troubleshooting networks. Parent ID data
does not apply.
19
ToxiRAE Pro User’s Guide
12.2.4. Enter Communications and stop measurement?
Before communicating with a PC running ProRAE Studio II Instrument Configuration
and Data Management software (for downloading datalogs, configuring the instrument,
or uploading new firmware), the ToxiRAE Pro must be set to communicate with the
computer. Press [MODE] repeatedly until you reach the “Enter Communications and stop
measurement?” screen. Then press [Y/+]. Measurement and datalogging are stopped, and
the instrument is now ready to communicate with the PC.
12.3. Programming
You must provide a 4-digit password to enter Programming Mode, unless the ToxiRAE
Pro is already running in Advanced User mode. The default password is 0000.
12.4. Enter i ng Pr ogr ammi ng Mode
1. Press and hold [MODE] and [Y/+] simultaneously until the Password screen appears.
2. Input the 4-digit password:
• Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again "wraps" around back to 0.
• Step from digit to digit using [MODE].
• When you are done, continue pressing [MODE] until OK or Cancel is highlighted.
• Press [Y/+] to register the password (if OK is highlighted) or to exit without entering
the password (if Cancel is selected).
If you make a mistake, you can cycle through the digits by pressing [MODE] repeatedly
and then using [Y/+] to change the number in each position.
Note: The default password is 0000.
20
ToxiRAE Pro User’s Guide
When you have successfully entered Programming Mode, the following screen appears:
Note: The password can only be changed by connecting the instrument to a PC running
ProRAE Studio II software. Follow the instructions in ProRAE Studio II to change it.
The Calibration icon is highlighted first by default, but you can press [MODE] to step
from one programming menu item to the next, with the name of the menu shown at the
top of the display and the corresponding icon highlighted. As you repeatedly press
[MODE], the selection moves from left to right, and the following screens appear in
sequence:
Note: When you reach Exit and press [MODE], the menu cycles back to Calibration.
To exit the Programming Menu, highlight Exit by pressing the [MODE] button, and press
the [Y/+] key.
21
ToxiRAE Pro User’s Guide
Calibration
Alarm Setting
Datalog
Monitor Setup
Zero Calibration
High Alarm
Clear Datalog
Wireless
Span Calibration
Low Alarm
Interval
Site ID
STEL Alarm*
Data Selection
User ID
TWA Alarm*
Datalog Type
User Mode
Alarm Mode
Date
Buzzer & Light
Time
Man Down
Alarm
Temp. Unit
Language
Zero At Start
LCD Contrast
12.4.1. Navigating Programming Menus
Navigating through the Programming menus is easy and consistent, using a single
interface format of a “forward arrow” and “Select,” controlled by the [MODE] key and
[Y/+] key, respectively.
In the submenus, the functions of a “down arrow” and “Select” are controlled by the
[MODE] key and [Y/+] key, respectively.
Also in submenus, the functions of the keys may correspond to the instruction directly
above each key. Therefore, if it says “OK” above the [MODE] key, pressing [MODE]
selects “OK.” When setting numerical data, the [MODE] key typically selects which digit
is highlighted, while each press of [Y/+] increases the digit by 1.
12.5. Programming Menus
Programming allows anyone with the password to change the instrument’s settings, calibrate
the instrument, enter user information, etc. Programming has four menus. Each menu
includes several submenus to perform additional programming functions.
This table shows the menus (and their icons) and submenus:
* STEL and TWA are not included in the menus on a ToxiRAE Pro equipped with an O2
sensor.
Note: There is also an “Exit” icon. It has only one function. When you select it, the
instrument exits Programming Mode, updates the settings, and returns to monitoring.
22
ToxiRAE Pro User’s Guide
Once you enter Programming, the LCD displays the first menu, Calibration. Each
subsequent menu is accessed by pressing [MODE] repeatedly until the desired menu is
displayed. To enter a submenu of a menu, press [Y/+].
12.5.1. Calibration
Two types of calibration are available: Zero (fresh air) and Span.
In the Calibration submenu, select Zero or Span Calibration by pressing [MODE]. Once
your choice is highlighted, press [Y/+].
Zero Calib
The procedure for performing a zero calibration is covered on page 43.
Span Calib
The procedure for performing a basic span calibration is covered on page 46.
Exit
Press [Y/+] to exit. Calibration is exited, and Measurement is highlighted in the
Programming Menu screen.
12.5.2. Alarm Setting
High Alarm
Set the High Alarm as follows:
1. In the High Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] until “OK” is
highlighted.
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
Low Alarm.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the High Alarm submenu and advance to the Low Alarm
submenu.
23
ToxiRAE Pro User’s Guide
Low Al arm
Set the Low Alarm as follows:
1. In the Low Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] until “OK” is highlighted.
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
STEL Alarm.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the Low Alarm submenu and advance to STEL Alarm
submenu.
STEL Alarm
Set the STEL Alarm as follows (except if the ToxiRAE Pro has an O2 sensor; then there
is no STEL Alarm):
1. In the STEL Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] several times until “OK”
is highlighted.
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
TWA Alarm.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the STEL Alarm submenu and advance to the TWA Alarm
submenu.
TWA
Set the TWA Alarm as follows (except if the ToxiRAE Pro has an O2 sensor; then there
is no TWA Alarm):
1. In the TWA Alarm submenu, press [Y/+] repeatedly to select a desired number.
Numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
2. Press [MODE] to move to the next digit.
3. When you are satisfied with the number, press [MODE] until “OK” is
highlighted.
24
ToxiRAE Pro User’s Guide
4. To save the new value, press [Y/+]. The value is saved, and the menu advances to
Alarm Mode.
To exit without saving the new value:
1. Press [MODE] to advance to Cancel.
2. Press [Y/+] to exit the TWA Alarm submenu and advance to the Alarm Mode
submenu.
Alarm Mode
There are two selectable alarm types:
Auto ResetWhen an alarm is triggered and after the alarm condition is no
longer present, the alarm stops and resets by itself.
LatchWhen the alarm is triggered, it will not cease until you manually
turn it off.
The latched setting only controls alarms for High Alarm and Low
Alarm, not the STEL or TWA alarms.
Note: To clear an alarm when the instrument is set to “Latched,”
press [Y/+] when the main (measurement) display is shown.
1. Press [MODE] to step from one alarm type to the other.
2. Press [Y/+] to select an alarm type.
3. When you have completed your selection, press [MODE] until “Exit” is
highlighted. Then press [Y/+] to save the change and exit to “Buzzer and Light.”
Buzzer & Light
The buzzer, light, and vibration alarms can be programmed so that any combination, all,
or none of them are activated when the ToxiRAE Pro goes into alarm.
Important! If you program the ToxiRAE Pro so that no alarms are activated in an alarm
condition, you may put yourself or your workers at risk of being exposed to a hazardous
atmosphere and being unaware of it, as the indication of a dangerous condition will be
limited to the concentration shown in the display and the name of the type of alarm
flashing on the display.
1. Choose an alarm combination setting by pressing [MODE] until your choice is
highlighted.
2. Press [Y/+] to select it.
3. Press [MODE] repeatedly until “Exit” is highlighted.
4. Press [Y/+] to register your selection and exit.
25
ToxiRAE Pro User’s Guide
To exit without saving the new value:
1. Make sure your previous setting is selected.
2. Press [MODE] repeatedly to advance to Exit.
3. Press [Y/+] to exit.
Man Down Alarm
The Man Down Alarm is a critical and potentially lifesaving safety feature of every
ToxiRAE Pro. The Man Down Alarm is based on the premise that if the instrument is
motionless when it is not supposed to be, something wrong may be happening to its user.
If that is the case, the ToxiRAE Pro not only goes into alarm locally on the instrument,
but also remotely, over the RAE Systems Dedicated Wireless Network, to notify people
in the vicinity, as well as remote safety officers at a command center, that a person is
down, so that help can be dispatched quickly.
Whenever the Ma n Down fea ture is on , the main scree n display s a Man Dow n icon along the
top to indic ate it is ac tive :
When the Man Down feature is on and there is no gas alarm, the ToxiRAE Pro senses that it is
motionless for the amount of time set in the “M otion less Time” parame ter . If the instr umen t is
not moved during that ti me, then a pr e-a larm i s activated to al ert the u ser, and shows the “Are
You OK?” screen. Pressing [Y/+] clears the alarm and returns the ToxiRAE Pro to its normal
operation. Pressing [MODE] sets it i nto Man D own A lar m (a nd if w irele ss c onn ecti vity is
enabled, a Man Down message is sent in rea l time to remote obse rvers). If neither key is
pressed, then after the countdown, it goes into Man Down Alarm (again sending a message to
remote observers if wirelessly enabled).
26
ToxiRAE Pro User’s Guide
Settings are available for :
• Off/On
• Motionless T ime (t ime th e ins trumen t is mo tionle ss be fore i nitia ting a pre-alarm)
• Sensitivity (set to low, medium , or h igh t o compe nsate for a mbient vibrat ion or m otion)
• Warning Time (coun tdown, in se conds, fr om pre-alarm to Man Down alarm)
When the Man Down a larm is ac tivated , the bu zzer sound s and L EDs flash co ntinuous ly, and a
countdown begins.
• If the ToxiRAE Pro’s user presses [Y/+] for “Yes” in response to the “Are You OK?”
question on the screen be fore t he countdo wn reac hes zero, t he Man Down alarm s tops
and the main reading screen is displayed.
• If the pers on d oes not p ress [Y/+ ] f or “Ye s” in resp onse to the “Are Yo u OK ?”
question on the screen be fore the countdown reaches zero, the Man Down alarm is
sounded and LEDs flash continuously.
• If the pers on pr es ses [MOD E] du ring the co untd own, answ eri ng the “A re Y ou OK ?”
question w ith “ No,” the Ma n Dow n a larm star ts.
If the ToxiRAE Pro’s wireless connectivity is enabled, a Man D own mes sage is a lso sent to
remote observers.
Exit (the Alarm Settings Submenu)
Press [Y/+] to exit the Alarm Settings submenu. Alarm Settings is exited, and Datalog is
highlighted.
If you want to cycle through the Alarm Setting submenus, press [MODE] instead of [Y/+].
Exit (the Alarm Settings Submenu)
Press [Y/+] to exit the Alarm Settings submenu. Alarm Settings is exited, and Datalog is
highlighted.
If you want to cycle through the Alarm Setting submenus, press [MODE] instead of [Y/+].
12.5.3. Datalog
Clear Datalog
This operation erases all data stored in the datalog.
Note: Once the datalog is cleared, the data cannot be recovered.
Press [Y/+] to clear the datalog. The display asks, “Are you sure?”
• Press [Y/+] if you want to clear the datalog. When it has been cleared, the display
shows “Datalog Cleared!”
• Press [MODE] if you do not want to clear the datalog.
27
ToxiRAE Pro User’s Guide
You are taken to the next submenu, Interval.
Interval
Intervals are shown in seconds. The default value is 60 seconds. The maximum interval is
3600 seconds, and the minimum is 1 second.
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
2. Press [MODE] to advance to the next digit.
3. Repeat this process until the desired value is entered.
When you have completed your selections, press [MODE] repeatedly until “OK” is
highlighted.
• To save the new value, press [Y/+].
• To cancel the changes and keep the settings as they were, press [MODE] to
highlight “Cancel.” Then press [Y/+].
You are taken to the next submenu, Data Selection.
Data Selection
Data Selection allows you to select which types of data are stored and made available
when you download your datalog to a computer via ProRAE Studio II software.
You can choose any or all of four types of data (you must choose at least one):
• Minimum
• Average
• Maximum
• Real Time
1. Press [MODE] to step from one option to the next. The highlighter indicates your
choice.
2. Press [Y/+] to toggle your selection on or off (the check box indicates “on” with
an “X”).
3. When you have completed your selections, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save the changes and exit.
28
ToxiRAE Pro User’s Guide
Datalog Type
The instrument has two datalog types:
AutoCollects datalog information when the instrument is sampling until the
datalog memory is full.
ManualDatalogging occurs only when you manually initiate datalogging (see
below for details).
1. Press [MODE] to step from one option to the next.
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
that an option has been selected).
3. When you have completed your selection, press [MODE] until “Exit” is
highlighted.
4. Press [Y/+] to save your selection and exit.
About Manual Datalog
When the instrument is set to Manual Datalog, you can turn datalogging on and off from
the main measurement display in the Normal Mode by pressing the [MODE] key three
times.
• When you reach the screen that says “Start Datalog?” press [Y/+] to start it. You
see “Datalog Started,” confirming that datalogging is now on.
• When you reach the screen that says “Stop Datalog?” press [Y/+] to stop it. You
see “Datalog Stopped,” confirming that datalogging is now off.
Exit (the Datalog Submenu)
Press [Y/+] to exit. Datalog is exited, and Monitor Setup is highlighted.
29
ToxiRAE Pro User’s Guide
12.5.4. Monitor Setup
Wireless
There are seven Wireless parameters:
• Radio On/Off
• PAN ID
• Channel
• Join Network
• Interval
• Off Network Alarm
• Factory Reset
Radio On/OffThis turns the wireless modem on or off.
Note: Turning the radio off does not change any of the settings.
They are stored and ready for use the next time the wireless modem
is turned on.
1. Choose an “On” or “Off” by pressing [MODE] until your choice is highlighted.
(The dark circle in the “radio button” indicates that an option has been selected).
2. Press [Y/+] to select it.
3. Press [MODE] repeatedly until “Exit” is highlighted.
4. Press [Y/+] to register your selection and exit.
To exit without saving the new value:
1. Make sure your previous setting is selected.
2. Press [MODE] repeatedly to advance to Exit.
3. Press [Y/+] to exit.
PAN IDAll units on a network must be programmed with the same Pan ID
(Personal Network Identifier) to ensure communication compatibility.
Important! The allowed range for Pan ID numbers is 001 through
999.
Choose and enter a 3-digit PAN ID:
1. Press [Y/+] and the display shows the current PAN ID. Example: “866.” The left-
most digit is highlighted.
2. Press [Y/+] repeatedly to select a desired number. Once 9 is reached, pressing
[Y/+] again “wraps” around back to 0.
3. Press [MODE] to advance to the next digit.
Repeat this process until all eight characters of the new PAN ID are entered.
30
ToxiRAE Pro User’s Guide
To save your new PAN ID and exit, press [MODE] repeatedly until “OK” is
highlighted. Then press [Y/+]. It says, “Applying settings…” and then “Settings
applied successfully!” to indicate they have been registered. The menu now advances
automatically to “Channel.”
Channel Every network operates on its own channel, between 01 and 10 (do not
use 00). If you need to know the network’s channel, check the
EchoView Host’s setting. Note: In Europe, only Channel 11 can be
used, and it is preset (it cannot be changed).
Choose and enter a 2-digit Channel:
1. Press [Y/+] and the display shows the current Channel. Example: “10.” The left-
most digit is highlighted.
2. Press [Y/+] repeatedly to select a desired number. Once 9 is reached, pressing
[Y/+] again “wraps” around back to 0.
3. Press [MODE] to advance to the next digit.
Repeat this process until all eight characters of the new Channel are entered.
To save your new Channel number and exit, press [MODE] repeatedly until “OK” is
highlighted. Then press [Y/+]. It says, “Applying settings…” and then “Settings
applied successfully!” to indicate they have been registered. The menu now advances
automatically to “Join Network.”
Join Network This screen is for joining an existing network. When you see this
screen, press [Y/+] to join, or [MODE] to exit without joining.
While searching for the network, the message “Joining network…” is displayed.
If the network is found and successfully joined, then “Join network successful!” is
displayed. Press [MODE] to acknowledge the connection and return to the Wireless
menu.
If the network is not joined, the display shows: “Join network failed!” You have two
options: Exit and Retry. If you want to retry, check that your network setting is
correct (matches the EchoView Host) and that the EchoView Host is turned on and
operating. Then press [Y/+] to retry.
Interval This sets the amount of time between data transmissions to the
EchoView Host.
To change the Interval:
1. Press [MODE] to step from one option to the other. The highlighted option
changes each time you press [MODE].
31
ToxiRAE Pro User’s Guide
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
the option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Interval menu, and advance to “Off
Netwk Alarm.”
Off Netwk Alarm When this is enabled, the unit goes into alarm whenever an active
network connection is lost. This informs you that connectivity has
been broken between the ToxiRAE Pro and the EchoView Host.
1. Choose “On” or “Off” by pressing [MODE] until your choice is highlighted. (the
dark circle in the “radio button” indicates that an option has been selected).
2. Press [Y/+] to select it.
3. Press [MODE] repeatedly until “Exit” is highlighted.
4. Press [Y/+] to register your selection and exit.
To exit without saving the new value:
1. Make sure your previous setting is selected.
2. Press [MODE] repeatedly to advance to Exit.
3. Press [Y/+] to exit.
Factory Reset Selecting this resets the wireless settings to their original factory
configurations.
CAUTION! A factory reset of the wireless settings deletes any customized
settings and cannot be undone.
1. Press [Y/+] to select “Factory Reset.” You will see the message “Reset wireless
settings?”
2. Press [Y/+] for “Yes” or [MODE] for “Exit.” If you select “Exit,” the ToxiRAE
Pro exits without performing a factory reset.
Exit (the Wireless Submenu)
Press [Y/+] to exit. Wireless is exited, and Monitor Setup is highlighted.
Site ID
Choose and enter an 8-digit alphanumeric Site ID to uniquely identify the particular site
where the instrument is to be used. This Site ID is included in the datalog report.
1. Press [Y/+] and the display shows the current Site ID. Example: “SITE0001.” The
left-most digit is highlighted.
2. Press [Y/+] repeatedly to select a desired character. Letters increase from current
value to Z followed by digits from 0 to 9. Once 9 is reached, pressing [Y/+] again
“wraps” around to the letter A. For the last four characters that must be numerical,
32
ToxiRAE Pro User’s Guide
numbers increase from 0 to 9. Once 9 is reached, pressing [Y/+] again “wraps”
around back to 0.
Note: The last four characters must be numerals.
3. Press [MODE] to advance to the next character.
Repeat this process until all eight characters of the new Site ID are entered.
To save your new Site ID and exit, press [MODE] repeatedly until “OK” is highlighted.
Then press [Y/+].
User ID
Enter an 8-digit alphanumeric User ID to uniquely identify a user. This User ID is
included in the datalog report.
1. Press [Y/+] and the display shows the current User ID. Example: “USER0000.”
The left-most character is highlighted.
2. Press [Y/+] repeatedly to select a desired character. Letters increase from current
value to Z followed by digits from 0 to 9. Once 9 is reached, pressing [Y/+] again
“wraps” around to the letter A.
3. Press [MODE] to advance to the next character.
Repeat this process until all eight characters of the new User ID are entered.
To save your new User ID and exit, press [MODE] repeatedly until “OK” is highlighted.
Then press [Y/+].
User Mode
The ToxiRAE Pro has two user modes:
Basic Basic users can only see and use a basic set of functions.
Advanced Advanced users can see all screens and take advantage of all available
functions.
Note: The default value for User Mode is Basic.
To change the User Mode:
1. Press [MODE] to step from one option to the other. The highlighting changes
each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
the option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to register your selection, exit the User Mode submenu and advance
to Date.
33
ToxiRAE Pro User’s Guide
Date
The Date is expressed as Month/Day/Year, with two digits for each (MM/DD/YY).
The left-most digit is highlighted to indicate it is selected when you enter the Date
submenu.
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
2. Press [MODE] to advance to the next digit.
Repeat this process until all six digits of the new date are entered.
3. Press [MODE] repeatedly until “OK” is highlighted.
4. Press [Y/+] to save your selection and advance to Time.
Note: If you do not want to save the change, press [MODE] repeatedly until
“Cancel” is highlighted. Press [Y/+] to exit to Time without saving the changes.
Time
The Time is expressed as Hours/Minutes/Seconds, with two digits for each. The time is in
24-hour (military) format of HH:MM:SS.
The left-most digit is highlighted to indicate it is selected.
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
2. Press [MODE] to advance to the next digit.
Repeat this process until all six digits of the new time are entered.
3. Press [MODE] repeatedly until “OK” is highlighted.
4. Press [Y/+] to save your selection, exit the Time submenu and advance to Temp.
Unit.
Note: If you do not want to save the change, press [MODE] until “Cancel” is
highlighted. Press [Y/+] to exit to Temp. Unit without saving the changes.
34
ToxiRAE Pro User’s Guide
Temp. Unit
The display unit of the internal temperature sensor can be switched between Fahrenheit
and Celsius.
To change the Temp. Unit:
1. Press [MODE] to step from one option to the other. The highlighted option
changes each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
the option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Temp. Unit menu, and advance to
Language.
Language
English is the default language, but other languages can also be selected for the
instrument.
1. Press [MODE] to step from one option to the next. The highlighted option
changes each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
this option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Language submenu, and advance to
Zero At Start.
Zero At Start
When Zero At Start is turned on, the instrument starts with zero calibration every time it
is turned on.
1. Press [MODE] to choose between “Off” and “On.”
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
your selection).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Zero At Start submenu, and advance to
LCD Contrast.
LCD Contrast
The display’s contrast can be increased or decreased from its default setting. You may
not need to ever change the default setting, but sometimes you can optimize the display to
suit extreme temperature and ambient brightness/darkness conditions.
1. Press and hold [Y/+] to select a desired level of contrast. The range is 1 to 100.
Once 100 is reached, pressing [Y/+] again "wraps" around back to 1.
35
ToxiRAE Pro User’s Guide
2. When you have completed your selection, press [MODE] repeatedly until “OK” is
highlighted.
3. Press [Y/+] to save your selection, and exit the LCD Contrast.
Note: If you do not want to save the change, press [MODE] repeatedly until
“Cancel” is highlighted. Press [Y/+] to exit without saving the changes.
Exit (Monitor Setup Submenu)
Press [Y/+] to exit. Monitor Setup is exited. Exit (to exit Programming Mode) is
highlighted.
12.6. Exit (Exit Programming Mode)
To exit Programming Mode and return to normal operation, press [MODE] until the Exit
icon is highlighted. Press [Y/+] to exit. You will see “Updating Settings…” as changes
are registered and the mode changes to normal operation.
13. Policy Enforcement
The ToxiRAE Pro can be configured to enforce a facility/company’s requirements that
calibration and/or bump testing be performed at specified intervals, and to warn the user
that calibration/bump testing is required. In addition, it can require calibration or bump
testing and not allow normal operation of the instrument unless calibration or bump
testing is performed.
If Policy Enforcement is enabled and if the instrument has been bump tested and
calibrated in compliance with the policy settings, a check-mark icon is included along the
top of the ToxiRAE Pro screen:
If Policy Enforcement is enabled, then after startup the ToxiRAE Pro displays a screen
that informs the user that the instrument requires either a bump test or a calibration. If
both are required, then they are shown in sequence.
13.1. Setting Policy Enforcement
You must use ProRAE Studio II to make changes to Policy Enforcement settings. The
procedure differs, depending on whether you are using an AutoRAE 2 Cradle, a
ToxiRAE Pro Travel Charger, or a ToxiRAE Pro Desktop Cradle.
13.1.1. AutoRAE 2 Cradle
Note: This procedure only works with a single AutoRAE 2 Cradle. It will not work with
an AutoRAE 2 Controller.
36
ToxiRAE Pro User’s Guide
To program a ToxiRAE Pro via an AutoRAE 2 Cradle, you need ProRAE Studio II
Instrument Configuration and Data Management Software, the AutoRAE 2 Cradle
connected to a power source, and a USB PC communications cable.
1. Connect a USB cable between a PC with ProRAE Studio II and the AutoRAE 2
Cradle.
2. Apply power to the AutoRAE 2 Cradle.
3. Turn off the ToxiRAE Pro (or put the ToxiRAE Pro into PC Communications Mode)
and set it in the cradle.
4. Start ProRAE Studio II software on the PC.
5. Select “Administrator” and input the password (the default is “rae”).
6. Click “Detect the instruments automatically” (the magnifying glass icon with the
letter “A” in it). After a few seconds, the AutoRAE 2 Cradle is found and it is shown,
along with its serial number:
7. Click on the icon to highlight it, and then click “Select.”
37
ToxiRAE Pro User’s Guide
8. In ProRAE Studio II, the AutoRAE 2 Cradle is shown, including its Serial Number,
under “Online”:
9. Expand the view to show the ToxiRAE Pro in the AutoRAE 2 Cradle by clicking the
“+” to the left of the image of the AutoRAE 2 Cradle:
10. Double-click on the icon representing the ToxiRAE Pro.
11. Click “Setup.”
38
ToxiRAE Pro User’s Guide
12. In the menu that now appears on the left side, click “Policy Enforcement.” It is
highlighted, and the Policy Enforcement pane is shown:
For “Must Calibrate” and “Must Bump,” you have the options of no enforcement or
enforcement (including “Can’t Bypass,” and “Can Bypass”).
Must Calibrate. The user is prompted to calibrate the instrument when calibration is due
(as set by the calibration interval). There are two programmable options:
•Can’t Bypass. Unless calibration is performed, the instrument cannot be used,
and the only option is to turn off the instrument.
•Can Bypass. If calibration is due but the user does not want to perform a
calibration, the instrument can still be used.
Must Bump
•Can’t Bypass. Unless a bump test is performed, the instrument cannot be used,
and the only option is to turn off the instrument.
•Can Bypass. If a bump test is due but the user does not want to perform one, the
instrument can still be used.
39
ToxiRAE Pro User’s Guide
These are the screens that are shown on a ToxiRAE Pro after startup if “Can Bypass” is
selected:
If “Can’t Bypass” is selected, the display looks like this, and only allows the options of
performing the test or shutting down:
13.1.2. ToxiRAE Pro Desktop Cradl e Or Travel Cha rger
Make sure the AC adapter is connected and that a USB cable is connected to a computer
running ProRAE Studio II.
1. Turn on the ToxiRAE Pro.
2. Hold down [MODE] and [Y/+] to enter Programming Mode.
3. Provide the password.
4. Press [MODE] until “Enter Communications Mode?” is displayed.
5. Press [Y/+]. The screen shows that it is ready by saying “Communications Mode
ready…” It is now ready to communicate with the computer running ProRAE
Studio II. If you want to exit, press [MODE].
11. Click on the instrument’s icon when it appears to highlight it.
12. Click “Select.”
13. Click “Setup.”
14. Click “Policy Enforcement.”
40
ToxiRAE Pro User’s Guide
The Policy Enforcement pane is shown:
You can select “Must Calibrate” and/or “Must Bump” and then set whether the user
must perform the selected operation in order to use the instrument.
15. Once you have made your selections in ProRAE Studio II, you must upload the
changes to the instrument. Click the icon labeled “Upload all settings to the
instrument.”
16. A confirmation screen is shown. Click “Yes” to perform the upload, or “No” to abort.
Uploading takes a few seconds, and this progress bar is shown. You can abort the
upload by clicking “Cancel.”
41
ToxiRAE Pro User’s Guide
17. Exit ProRAE Studio II.
18. Press [Y/+] on the ToxiRAE Pro to exit Communication Mode.
13.2. Deactivating Policy Enforcement
13.2.1. AutoRAE 2 Cradle
To deactivate Policy Enforcement when using an AutoRAE 2 Cradle, follow the
procedure for changing settings. See page 36 for details.
13.2.2. ToxiRAE Pro Desktop Cradle Or Travel Charger
If the ToxiRAE Pro screen displays the message that it must be bump tested or calibrated,
and if the option to bypass bump testing or calibration is not available, you should shut
off the instrument and follow the procedure outlined here if you want to change the
Policy Enforcement settings:
1. Use a USB cable to connect the ToxiRAE Pro in its Travel Charger or Desktop
Cradle to a computer running ProRAE Studio II.
2. Enter Diagnostic Mode on the ToxiRAE Pro (with the instrument turned off, press
and hold [Y/+] and [MODE] until it starts up.
3. After startup, enter the password when prompted (default is “0000”) and press
[MODE].
4. Press [MODE] repeatedly until you see the “Enter Communications Mode?” screen.
5. Press [Y/+] to enter Communications Mode.
6. Start ProRAE Studio II.
7. Select “Administrator.”
8. Input the password (the default is “rae”).
9. Click “OK.”
10. Click “A” (detect instruments automatically).
11. Click on the instrument’s icon when it appears.
12. Click “Select.”
13. Click “Setup.”
14. Click “Policy Enforcement.” The Policy Enforcement pane is shown.
15. Deselect Policy Enforcement features you do not wish to use.
16. Click “Upload all settings to the instrument.”
42
ToxiRAE Pro User’s Guide
17. When you see this confirmation. Click “Yes.”
Uploading will take a few seconds, and this progress bar is shown:
18. When the upload is done, exit ProRAE Studio II.
19. Press [Y/+] on the ToxiRAE Pro to exit Communication Mode.
14. Bump Testing & Calibrating The ToxiRAE Pro
Note: Calibration intervals and bump test procedures may vary due to national
legislation.
14.1. Bump Testing & Calibrating With An AutoRAE 2™
Bump testing and calibrating a ToxiRAE Pro can be done with an AutoRAE 2 Automatic
Test And Calibration Station. Follow all instructions in the AutoRAE 2 User’s Guide for
successful bump testing or calibration.
14.1.1. Bump (Functional) Testing
With the ToxiRAE Pro in Normal Mode:
1. Connect the calibration gas cylinder, flow regulator, and calibration adapter to the
ToxiRAE Pro and start the gas flow.
2. Make sure the unit goes into alarm and that the buzzer produces at least two beeps per
second, the LED lights flash on and off, and the vibration alarm functions. The
display backlight should illuminate and an alarm message should be shown in the
display.
3. Turn off the gas flow.
4. Remove the calibration adapter.
43
ToxiRAE Pro User’s Guide
Note: Bump testing and calibration can also be performed by using the RAE Systems
AutoRAE 2 Automatic Test And Calibration Station. Refer to the AutoRAE 2 User’s
Guide for instructions.
14.2. Calibration Adapter
The ToxiRAE Pro is a diffusion monitor and must be calibrated using a fixed-flow
regulator with a flow rate between 0.5 and 1.0 liters per minute. The ToxiRAE Pro is
supplied with a special calibration adapter that covers the gas inlet. If the local air has
impurities that might interfere with zero (fresh air) calibration, the calibration adapter and
a cylinder of clean zero air should be used.
14.2.1. Connecting The Calibration Adapter
The supplied Calibration Adapter has a removable insert to allow bump testing and
calibration to be performed on the instrument regardless of whether it has a rubber boot
installed. Use the Calibration Adapter without the insert to calibrate the instrument in the
rubber boot, and with the insert to calibrate the instrument without the rubber boot on it.
1. Place the Calibration Adapter on the top cap of the ToxiRAE Pro by pressing it
into place.
2. Attach the hose from the regulator/calibration gas cylinder to the Calibration
Adapter’s inlet.
Caution! After calibration is complete, remove the Calibration Adapter. When
monitoring, never operate the ToxiRAE Pro with the Calibration Adapter attached. If the
Calibration Adapter is attached during normal operation, inconsistent and lower-thannormal readings will occur because of decreased concentration of the gas being
monitored.
14.2.2. Disconnecting The Calibration Adapter
After calibrating the ToxiRAE Pro, you must remove the Calibration Adapter by lifting it
off the ToxiRAE Pro.
44
ToxiRAE Pro User’s Guide
14.3. Zero Calibration
This operation sets the zero point of the sensor calibration curve for clean air. Expose the
ToxiRAE Pro to a clean air source with 20.9% oxygen and without any organic, toxic, or
combustible gas impurities. This “Zero Air” can come from a cylinder or clean ambient
air.
Note: If you use a zero air cylinder, you must use the ToxiRAE Pro Calibration Adapter.
Using a calibration adapter is not necessary for calibration in fresh air.
Turn on the ToxiRAE Pro. Once it ha s been thr oug h it s st art up, ente r Pr ogra mmi ng Mode :
1. Press and hold [MODE] and [Y/+] until you see the Password screen.
2. Input the 4-digit password:
• Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to
9. Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
• Step from digit to digit using [MODE].
• When you are done, press [MODE] until OK or Cancel is highlighted.
• Press [Y/+] to register the password (if OK is highlighted) or to exit without
entering the password (if Cancel is selected).
If you make a mistake, you can cycle through the digits by pressing [MODE] and
then using [Y/+] to change the number in each position.
Note: The default password is 0000.
If you do not know the password, do not enter anything. Just press [MODE]
repeatedly to highlight OK, and then press [Y/+].
Note: If the ToxiRAE Pro is in Programming Mode and you do not press a key within 60
seconds, it exits Programming Mode and re-enters Normal Mode.
45
ToxiRAE Pro User’s Guide
1. At the Calibration screen, press [Y/+].
2. “Zero Calib” is highlighted. Make sure the ToxiRAE Pro is in clean air.
Otherwise attach a cylinder of zero gas to the Calibration Adapter installed on the
instrument.
3. Press [Y/+]. The message “Apply zero gas…” is displayed.
4. Start the flow of the zero gas (if used) and press [Y/+] to start the zero calibration.
5. There is a 60-second countdown, during which time the screen shows the word
“Zeroing…” and the countdown.
6. Upon completion, it says “Zeroing is done! Reading = 0.0ppm” along with the
reading (it should be 0.0 ppm or very close to it).
7. “Span Calib” is highlighted.
Note: You can a bort the zero cal ibration during its countdown by pressing the [M ODE] key .
Important! If you used the Calibration Adapter and a zero air cylinder, you must remove
the Calibration Adapter from the ToxiRAE Pro. If the Calibration Adapter is attached
during normal operation, inconsistent and lower-than-normal readings will occur because
of decreased concentration of the gas being monitored.
14.4. Span Calibration
This procedure determines the second point of the sensor calibration curve for the sensor.
To begin calibration, connect the calibration gas cylinder, flow regulator, and calibration
adapter to the ToxiRAE Pro.
With the ToxiRAE Pro in Programming Mode, and with Span Calib highlighted, follow
this procedure:
1. Press [Y/+]. The screen displays the current (or default) calibration gas and its
concentration. For example:
Cal. Gas: CO
Span=50ppm
Change?
• Press [MODE] if you do not want to change it.
• Press [Y/+] if you want to change it.
If you entered the Calibration Menu without supplying the correct password, or
supplied the correct password but chose not to change the Cal.Gas, you will see a
message like this:
Cal.Gas: CO
Span=50ppm
Apply gas…
2. Press [Y/+] to start calibration.
46
ToxiRAE Pro User’s Guide
3. Press [MODE] if you want to quit without calibrating.
During calibration, the calibration gas name, its concentration, and the countdown are
displayed. Depending upon the device alarm settings, the monitor may go into low and
then high alarm once span calibration is complete, which is expected in response to the
test gas.
Note: You can abort span calibration during its countdown by pressing [MODE].
Upon completion, the screen “Span is done!” appears along with the reading (it should be
within 10% of the span gas concentration).
Important! After calibration is complete, you must remove the Calibration Adapter from
the ToxiRAE Pro. If the Calibration Adapter is attached during normal operation,
inconsistent and lower-than-normal readings will occur because of decreased
concentration of the gas being monitored.
14.4.1. Changing Span Gas Concentration
If you selected “Change” for span gas concentration, the following screen appears:
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
2. Step from one digit to the next by pressing [MODE].
3. When you are done, press [MODE] until either “OK” or “Cancel” is selected.
Press [Y/+] to register your selection (if “OK” is highlighted) or to exit without
registering it (if “Cancel” is selected).
47
ToxiRAE Pro User’s Guide
15. Diagnostic Mode
In Diagnostic Mode, the ToxiRAE Pro provides raw counts for sensor, battery, and other
readings. Most of these screens are useful only to service technicians. Therefore, if you
require information on Diagnostic Mode, refer to the ToxiRAE Pro Service Guide.
The ToxiRAE Pro’s Diagnostic Mode can only be accessed at startup time. In Diagnostic
Mode, Toxi RAE Pro d isplay s reading s in raw cou nts instea d of un its such a s parts per million
(ppm).
15.1. Entering Diagnostic Mode
1. With the ToxiRAE Pro turned off, press and hold both [MODE] and [Y/+].
2. When the display turns on and the password screen appears, release the keys.
3. Enter the 4-digit password (the password is the same as the one for the
Programming Mode):
• Step from one position in the four-character string to the other by pressing
[MODE].
• Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
• Once 9 is reached, pressing [Y/+] again “wraps” around back to 0.
4. When you are done, press [MODE] until either OK or Cancel is selected. Press
[Y/+] to register the password (if “OK” is highlighted) or to exit without entering
the password (if “Cancel” is selected).
15.2. Navigating Diagnostic Mode
Step through Diagnostic Mode by pressing [MODE].
15.3. Exiting Diagnostic Mode
1. Turn off the ToxiRAE Pro by pressing and holding [MODE]. There will be a
standard shutoff countdown.
2. When the instrument shuts off, you will be alerted. Release your finger.
Note: The next time you start ToxiRAE Pro, hold only [MODE], and it will
automatically start in Normal Mode.
48
ToxiRAE Pro User’s Guide
Alarm
Condition
Alarm Signal
Super Alarm
4 beeps/flashes per second
(Man
Down)
Man Down Alarm
3 beeps/flashes per second
(Man Down
Warning)
Man Down Warning
2 beeps/flashes per second
LAMP
Lamp failure
3 beeps/flashes per second
MAX
Gas exceeds electronics’
maximum range
3 beeps/flashes per second
OVER
Gas exceeds measurement
3 beeps/flashes per second
HIGH
Gas exceeds “High Alarm”
3 beeps/flashes per second
LOW
Gas exceeds “Low Alarm”
2 beeps/flashes per second
NEG
Gas reading measures less
than zero
1 beep/flash per second
STEL*
Gas exceeds “STEL” limit
1 Beep/flash per second
TWA*
Gas exceeds “TWA” limit
1 Beep/flash per second
CAL
Calibration failed, or needs
calibration
1 beep/flash per second
Offline
Offline from network
1 beep/flash per second
Bat
Low battery
1 flash, 1 beep per minute plus
16. Alarm Signal Summary
Alarm types include:
Message/
Priority
range
limit
limit
communcations
battery icon flashes on display
The alarm priority is as follows (in the event of multiple simultaneous alarms):
Super Alarm > Man Down Alarm > Man Down Warning > Max > Over > High > Low >
Neg > STEL > TWA > Cal Failed > Bump Failed > Cal Required > Bump Required >
Offline > Bat
* If your ToxiRAE Pro is configured with an oxygen (O2) sensor, alarms for TWA and
STEL are not included because neither TWA nor STEL are calculated for this sensor.
49
ToxiRAE Pro User’s Guide
17. Downloading Datalogs And Performing PC-Based
Instrument Configuration and Firmware Upgrades
The ToxiRAE Pro communicates with a PC running ProRAE Studio II Instrument
Configuration and Data Management software (version 1.03 or higher) to download
datalogs, configure the instrument, or upload new firmware.
The ToxiRAE Pro must be connected to a PC through the supplied Charging / PC
Communications Cradle and must be in the PC communications mode.
1. Make sure the Charging/PC Communications Cradle has power from its AC
adapter (the LED on the front of the cradle should be illuminated).
2. Use the supplied PC Communications Cable (USB to mini-USB cable) to connect
the Cradle to a PC.
3. Turn on the ToxiRAE Pro. Make sure it is running in Normal mode (with the
main measurement screen with its reading showing).
4. Insert the ToxiRAE Pro in the cradle.
5. Activate the PC communications mode on the ToxiRAE Pro by pressing [MODE]
repeatedly, starting from the main measurement screen until you reach the “Enter
PC Comm and stop measurement?” screen.
6. Press [Y/+]. Measurement and datalogging stop, and the instrument is now ready
to communicate with the PC. The display now says “PC comm ready…”
7. Start up the ProRAE Studio II software, enter a password, and detect the
instrument following the directions provided in the ProRAE Studio II User’s
Manual.
8. Follow the instructions in the ProRAE Studio II User’s Manual to download the
datalog, configure the instrument settings, or update the ToxiRAE Pro’s firmware.
9. When you are done, press [MODE] to exit the PC communications mode on the
ToxiRAE Pro. The screen displays “Sensor initializing,” after which the
instrument returns to operating in Normal mode.
50
ToxiRAE Pro User’s Guide
18. Maintenance
WARNING!
Maintenance should be performed only by a qualified person who has proper
training and fully understands the contents of this manual.
The following guidelines should be followed when changing the battery:
1. Turn off the unit before changing the battery.
2. When replacing the battery, take note of its orientation. The connectors on the circuit
board should go into the matching slots in the battery.
WARNING
To reduce the risk of ignition of hazardous atmospheres, replace the battery only in
areas known to be non-hazardous.
18.1. Replacing the ToxiRAE Pro Li-Ion Battery
Caution: Turn off the ToxiRAE Pro before re mov ing or re plac ing the ba ttery .
1. If the rubber boot is on the ToxiRAE Pro, remove it (ref er to page 55 for details).
2. Remove the cap by unscrewing it.
3. Remove the sensor module by gripping it with your fingers and pulling it out. Do not
use tools, which could damage the sensor module.
4. Place the ToxiRAE Pro face-down on a soft surface.
51
ToxiRAE Pro User’s Guide
Ribbon cable
5. Remove the four screws on the back of the instrument by turning them counterclockwise.
5. Hold the ToxiRAE Pro down a nd separate the rear cover from the front by pulling
from the top and tilting it open without disconnecting the ribbon cable connecting the
two halves.
6. Carefully disconnect the battery from the battery connectors on the circuit board
by pulling it toward the bottom of the instrument. Be careful not to damage the
ribbon cable.
7. Before installing a new battery, visually inspect the contact pins on the ToxiRAE
Pro circuit board to make sure they are not bent. If they are, carefully straighten
them before installing the battery. Install the new battery (part number G02-3004-
52
ToxiRAE Pro User’s Guide
000), making sure it is inserted fully. Make sure the battery connectors on the
circuit board go all the way into the connector slots in the battery.
8. Inspect to make sure the ribbon connector is still connected, and plug it back in if
it has come unplugged.
9. Place the cover over the compartment by tilting it up and then fitting it in place.
10. Tighten all four screws by turning them clockwise with a torque-limiting
screwdriver (preferably an electric one). Tighten them in the numerical order
shown here, to a torque rating of 3.54 to 4.43 lbf-in (4.08 to 5.10 kgf-cm).
Note: Incorrectly tightening the screws may compromise the instrument’s
IP rating.
11. Inspect the filter in the top cap and replace if necessary. Screw on the top cap.
Important! Charge the battery completely before using the ToxiRAE Pro, and check the
calibration before placing the instrument in service.
53
ToxiRAE Pro User’s Guide
18.2. Proper Battery Disposal
The Waste Electrical and Electronic Equipment (WEEE) directive
(2002/96/EC) is intended to promote recycling of electrical and electronic
equipment and their components at end of life. This symbol (crossed-out
wheeled bin) indicates separate collection of waste electrical and electronic
equipment in the EU countries. This product may contain one or more
Nickel-metal hydride (NiMH), Lithium-ion, or Alkaline batteries. Specific
battery information is given in this user guide. Batteries must be recycled or
disposed of properly.
At the end of its life, this product must undergo separate collection and recycling from
general or household waste. Please use the return and collection system available in your
country for the disposal of this product.
Sensor Specifications, Cross-Sensitivities, And Calibration
Information
For information on sensor specifications, cross-sensitivities, and calibration information,
refer to RAE Systems Technical Note TN-114: Sensor Specifications And CrossSensitivities (available for free download from www.raesystems.com). All specifications
presented in this Technical Note reflect the performance of standalone sensors. Actual
sensor characteristics may differ when the sensor is installed in different instruments. As
sensor performance may change over time, specifications provided are for brand-new
sensors.
54
ToxiRAE Pro User’s Guide
18.3. Removing And Installing The Rubber Boot
The protective rubber boot on the ToxiRAE Pro must be removed before the sensor top
cap or back of the instrument can be removed. When removing the boot, follow these
steps outlined here.
1. Flip up the tab on the belt clip on the rear of the instrument.
2. Remove the Philips screw that holds the clip in place.
3. Stretch the bottom corners of the boot and gently slide the bottom of the boot over
the rear of the ToxiRAE Pro.
4. Slide the boot up until it comes off of the instrument. Make sure that you do not
force it to move if it sticks against the top cap. Gently nudge it.
5. Inspect the rubber boot for damage, and replace it if it is cracked or torn.
When replacing the boot, simply reverse the procedure, sliding the boot onto the
instrument and then carefully stretching it over the bottom portion of the instrument. Do
not forget to replace the belt clip.
55
ToxiRAE Pro User’s Guide
18.4. Filter Replacement
A filter is permanently attached inside the ToxiRAE Pro’s top cap, which helps to keep
debris and moisture out. This filter should be inspected periodically, and the cap and
rubber insert should be inspected for damage. If the filter is dirty or the cap shows signs
of damage, replace the entire instrument cap assembly, including the cap and filter (part
number G02-0307-003, package of three).
To replace the cap assembly:
1. Remove the top cap by unscrewing it:
2. Discard the cap/filter and replace it with a new one by screwing it on:
56
ToxiRAE Pro User’s Guide
Sensor module
Match index keys
on sensor module
with ToxiRAE Pro
enclosure
Sensor
indexing
key
18.5. Re placing The Sensor
Follow the instructions for removing the cap in the previous section.
1. Gently lift out the sensor module with your fingers:
2. The sensor can only go into its slot one way. The connector inside the ToxiRAE
Pro and the indexing guides are good visual indicators of how to set the sensor
into position.
3. Place the sensor module into the ToxiRAE Pro. Make sure the indexing keys are
aligned (the sensor can only go in one way) and that it is seated firmly.
4. Carefully place the cap back onto the ToxiRAE Pro by screwing it on.
Note: Always calibrate the ToxiRAE Pro after replacing the sensor module.
57
ToxiRAE Pro User’s Guide
18.6. Cleaning The Exterior
Occasional external cleaning of the instrument with a soft cloth is recommended. Do not
use detergents or chemicals. If necessary, you can use a damp cloth (water only). It is a
good idea to install the Calibration Adapter before cleaning the ToxiRAE Pro’s housing,
to keep dirt, dust, or moisture away from the sensor openings and to keep the filter clean.
Visually inspect the contacts at the base of the ToxiRAE Pro and on the charging cradle
to make sure they are clean. If they are not, wipe them with a soft, dry cloth. Never use
solvents or cleaners.
18.6.1. Ordering Replacement P art s
If you need replacement parts, contact your local RAE Systems distributor. A list of
authorized distributors is available online:
http://www.raesystems.com
58
ToxiRAE Pro User’s Guide
Problem
Possible Reasons & Solutions
Cannot turn on power
Reasons: Depleted battery.
Defective battery.
Solutions: Replace battery.
Lost password
Solutions:Use ProRAE Studio II
+1 888-723-4800
Reading abnormally
Reasons: Calibration Adapter is
Pro.
Buzzer, LEDs, or
Reasons: Alarms may be turned
center.
19. Troubleshooting
software to change the
password.
Call Technical Support
at +1 408-752-0723 or
toll-free at
Low
vibration alarm
Inoperative
attached.
Incorrect calibration.
Solutions:Remove Calibration
Adapter.
Calibrate the ToxiRAE
off.
Blocked alarm port.
Bad buzzer, LEDs, or
vibration alarm.
Solutions:Enable alarms in the
Buzzer & Light section
of the Alarm Settings
sub-menu in the
Programming menu.
Unblock alarm port.
Call authorized service
59
ToxiRAE Pro User’s Guide
20. Technical Support
To contact RAE Systems Technical Support Team:
Monday through Friday, 7:00AM to 5:00PM Pacific (US) Time
Phone (toll-free): +1 888-723-4800
Phone: +1 408-952-8461
Email: tech@raesystems.com
21. RAE Systems Contacts
RAE Systems by Honeywell
World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: +1
Fax: +1
E-mail: customerserv@raesystems.com
Web Site: www.raesystems.com
RAE Systems Technical Support
Monday through Friday, 7:00AM to 5:00PM Pacific Time
Phone: +1.408.952.8461
Email: tech@raesystems.com
This manual must be carefully read by all individuals who have or will have the
responsibility of using, maintaining, or servicing this product. The product will perform
as designed only if it is used, maintained, and serviced in accordance with the
manufacturer’s instructions. The user should understand how to set the correct parameters
and interpret the obtained results.
CAUTION!
Only use RAE Systems battery pack P/N: G02-3004-000
Do only charge the battery pack using RAE Systems charger outside hazardous area.
Use of non-RAE Systems components will void the warranty and can compromise the
safe performance of this product.
Warning: Substitution of components may impair intrinsic safety.
ToxiRAE Pro Ma rking
The ToxiRAE Pro is certified according to the IECEx scheme, ATEX and CSA for US
and Canada as protected by intrinsic safety.
The ToxiRAE Pro is marked with the following information:
RAE SYSTEMS
3775 N. 1st. St., San Jose
CA 95134, USA
ToxiRAE Pro
Type PGM-1860.
Serial No/barcode: XXXX-XXXX-XX
IECEx SIR
10.0096X
Ex ia IIC T4 Ga
SIRA 10 ATEX 2189X
Ex ia IIC T4 Ga
-20º C < Tamb < +55º C;
Um: 20V
Battery pack: G02-3004-000
Warning: Read User’s Manual for Intrinsic Safety Precautions
Warning: Substitution of components may impact intrinsically safety
63
0575 II 1G
2010 2360054
I.S. for Cl.I Dv 1, Grps A,B,C,D TCode T4.
Class I, Zone 0 A/Exia IIC T4
Only as to intrinsic safety for use in
hazardous loc. Intrinsically safe/
ToxiRAE Pro User’s Guide
Operation Area and Conditions
Hazardous Areas classified by Zones
ToxiRAE Pro (PGM1860) are intended to be used in hazardous areas classified zone 0,
zone 1 or zone 2, where gases of explosion groups IIA, IIB or IIC and T4 may be present.
For North America and Canada the equipment can further be used in Class I, Zone 0.
Hazardous Areas classified by Divisions
ToxiRAE Pro is intended to be used in hazardous areas classified for Class I Div. 1 or 2,
within the temperature range of -20º C to +55º C, where gases of explosion groups A, B,
C or D and temperature code T4 may be present.
Instruction For Safe Use
Strictly follow the instructions for safe use. Application of the ToxiRAE Pro requires full
understanding and strict observation of the instructions.
Only use RAE Systems battery pack P/N: G02-3004-000
Do only charge the battery pack using RAE Systems charger outside hazardous area
(Um: 20V).
Use In Hazardous Areas
Equipment which is intended for use in explosive atmospheres and which has been
assessed and certified according to international regulations may be used only under
specified conditions. The components may not be modified in any way.
The appropriate regulations for service and repair must be properly observed during such
activities.
64
ToxiRAE Pro User’s Guide
First digit
Year
J
2008
K
2009
M
2010
N
2011
P
2012
Q
2013
R
2014
S
2015
T
2016
U
2017
V
2018
W
2019
Year of manufacture
To identify the year of manufacture, refer to the serial number of the instrument.
The second to last digit in the serial number indicates the year of manufacture. For
example, “M” indicates the manufacturing year is 2010.
65
P/N G02-4009-000 Rev. C
RAE Systems by Honeywell
World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: 408.952.8200
Fax: 408.952.8480
E-mail: customerserv@raesystems.com
Web Site: www.raesystems.com
March 2013
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.