Rae TN-204 User Manual

Page 1
Technical Note TN-204 04/17/vk
RAE Systems
TVOC Mode
Benzene Mode
MultiRAE Benzene with RAE Cartridge
RAE-Sep Cartridge
INTRODUCTION
Benzene is a very well-known human carcinogen that is widely distributed throughout nature (coal, crude oil, gas and gas con­densate). In most fuel products (gasoline, diesel, jet fuel), it ex­ists in concentrations of up t o several percent, by volume (vol%). Benzene is a very important industrial chemical, as well. It belongs to the BTEX aromatic family and is a liquid with a ra­ther high vapor pressure. Due to its ha z­ardous properties, Cal/OSHA set its PEL (Personal Exposure Limit) as 1 ppm of TWA and 5 ppm of STEL (Short Term
1
Exposure Limit).
ACGIH recommends
even more strict limits of 0.5 and 2.5
2
ppm, respectively.
There are several portable products on the market (from RAE Systems, IonScience, Dräger) that are able to measure such low concentra­tions of benzene, but only the MultiRAE Benzene combines the ability to make quick benzene-specific assessment (“snapshot”) measurements with the ad-
-Sep Tube
vantages of a versatile multi-gas monitor (up to six gas sensors) with real-time wireless capability. It can support more than 20 intelligent interchangeable sensor options.
Note: This Technical Note applies only to the MultiRAE Ben­zene monitor. For information regarding the UltraRAE 3000, re­fer to RAE Systems Technical Note TN-127: “Benzene-Specific Measurements In Petroleum Hydrocarbons Using The RAE Systems UltraRAE 3000.”
INSTRUMENT AND CARTRIDGE
The MultiRAE Benzene is designed specifically for use with the replaceable benzene-specific RAE-Sep Tube Cartridge, which comprises six individual RAE-Sep tubes.
The MultiRAE Benzene has two modes: TVOC (Total Volatile Organic Compounds) mode and Benzene mode. When used in TVOC mode, its display shows all sensor readings continuously.
When the instrument i s i n benzene mode, t he dis play p rompts for snapshot sampling. (Refer to the MultiRAE User’s Guide for details on managing, calibrating, and measuring in both TVOC and Ben-
3
zene modes
.) To perform snapshot compound-specific (ben­zene) measurement in addition to general VOC measurement requires a RAE-Sep Tube Cartridge (available in packs of six: P/N M01-0312-000).
Note: Each tube should be used for one measurement only.
The RAE-Sep Tube Cartridge is designed to provide six one­time specific measurements of benzene. This cartridge has unique properties and performance features not found in other monitors’ separation tube configurations. It has been designed to provide accurate, benzene-only measurement and has only been evaluated and tested on the MultiRAE Benzene monitor.
Failure to use the MultiRAE Benzene monitor with the
Note:
RAE-Sep Tube Cartridge will result in inaccurate detection or measurement, which may include the failure to detect benzene and possible serious injury or death.
For proper cartridge operation and replacement, refer to the doc­ument “RAE-Sep Tube Cartridge For MultiRAE Benzene.”
4
by Honeywell 877-723-2878 raesystems.com
Honeywell Internal
1
Page 2
Technical Note TN-204 04/17/vk
RAE Systems
Sensor/Mode
Range
Reso-
Some measurable parameters of the instrument are presented be­low. Refer to the MultiRAE Benzene datasheet for comprehensive sensor specification data.
5
lution
9.8eV for use in TVOC mode 10 ppb to 2,000 ppm 10 ppb
9.8eV for use in Benzene mode (with RAE-Sep Tube Cartridge)
0.1 to 200 ppm 0.1 ppm
MULTIRAE BENZENE-SPECIFIC MEASUREMENT PRINCIPLE
In Benzene mode, the monitor is designed to make specif ic ben­zene measurements. This is achieved via these two approaches:
1. The monitor uses a lower-energy, 9.8eV PID lamp. This al­lows excluding from cross-contamination all the VOCs with ionization energy (IE) above 9.8eV.
2. RAE-Sep tubes eliminate most VOCs. Two major effects are present in them:
Gas absorption in the tube. This is usually a fast pro-
cess, and it is reversible.
Gas reaction (chemical) in the tube. This is a slower
process, and it is not reversible.
Note: Some VOCs can be eluted (removed) from the RAE-Sep Tube and cause false positive benzene readings.
In addition, if the instruments are insensitive to some gases (for example, their IE is above 9.8eV or if they require high correction factors), the products of their chemical reactions in RAE-Sep Tubes may become sensitive and may distort the benzene read­ing. A separate case is when gases don’t absorb in the RAE­Sep Tube and are insensitive to the monitor equipped with a
9.8eV lamp (for example, lower hydrocarbons such as methane,
ethane, etc.).
Example of Elution When the instrument in Benzene mode was exposed to 400 ppm of acetone for 60 seconds, the reading didn’t exceed 0.2 ppm. After 5 minutes running in air, the reading increased and reached almost 30 ppm.
by Honeywell 877-723-2878 raesystems.com
Acetone eluate from Sep-Tube
35000 30000 25000 20000 15000 10000
5000
Response, ppb
0
0 10 20 30
Time, min
ppb
Elution in Benzene mode with exposure to 400 ppm of acetone.
Acetone didn’t convert in the RAE-Sep Tube within a short pe­riod of time (slow chemical reaction) and further eluted with the air stream.
Important: For the correct measurement an d to avoid a false positive response, RAE-Sep Tubes should be used just one time.
Humidity Effect
Humidity has no effect on the response to benzene. However, high humidity affects the capacitance of the tubes to remove in­terfering hydrocarbons, especially when they are in high con­centrations.
The figure below shows the response of the instrument to the mix of benzene (2 ppm) and octane (180 ppm). HPS-2 is the case when the tube was opened and the instrument run with a high pump speed (HPS) for about 5 minutes in ambient air with relative humidity (RH) of 75 to 80% before a benzene measure­ment in Benzene mode was done. HPS-1 i s the case when ben­zene measurements were done right after opening the tube.
Benzene+Octane, response vs. numbers
of measurements
4.5 4
3.5 3
2.5 2
1.5 1
Reading, ppm
0.5 0
0 2 4 6 8
Numbers of measurements
HPS-1
HPS-2
Response of Benzene and Octane with different times of measurement after opening the tubes, HPS-1 and HPS-2. Each line represents a single tube, and each triangle repre­sents a single measurement with that tube.
2
Honeywell Internal
Page 3
Technical Note TN-204 04/17/vk
RAE Systems
Gas
Gas concentration,
Capacitance,
Toluene
100
≥10
i-Butylene
100
≥10
i-Butylene
1000
5
Methane
25000
No response
n-Hexane
2400
10 sec
n-Octane
500
1*
n-Pentane
3500
7 sec
n-Propane
11000
No response
Gas Cross-
Gas concentration,
Benzene
Pentane
1700
0.1
Propane
5500
0
i-Butane
4500
0.5
i-Butylene
100
0.6
i-Butylene
1000
0.8
Methane
12500
0
Octane
250
0.3
Hexane
1200
0.6
Hexane
100
0
Butadiene 1,3
2.5
0.5
Toluene
50
1
Acetone
200
0.1
Tube capacitance can be extended significantly while using the instrument’s low pump speed (LPS) instead of its high pump speed as default setting. Test conditions are the same as above.
Benzene+Octane, response vs. numbers
of measurements
4.5 4
3.5 3
2.5 2
1.5 1
Reading, ppm
0.5 0
0 2 4 6 8
Numbers of measurements
HPS-1
LPS-1
HPS-2
LPS-2
Note: Previous results reinforce our strong recommendation to only open RAE-Sep Tubes right before performing a benzene­specific measurement. Other approaches could lead to false positive results.
Temperature Effects
Temperature variation affects RAE-Sep Tube performance. De­creasing te mper ature lea ds to the highe r tube ab sorpt ion, but sl ows down chemical transformation. As a result, the tube may not show a color change, but the tube fully or partially loses its capacitance. To avoid confusion, follow the recommendation of “One tube, one measurement.” Note: Calibration should be performed at the same temperature as the measurement because time necessary for sam­pling changes wit h tempe ratur e.
Tube Capacitance
Tube capacitance in this case is defined as: MultiRAE Benzene with a RAE-Sep Tube exposed to the gases in the table for a limited time until the reading reached 0.5 ppm.
Concentration vs. Exposure Time
Potential Cross-Sensitivity
sensitivity
ppm
Response, ppm
Benzene Absorption by RAE-Sep Tubes
There is a small effect of the benzene absorbtion by the RAE­Sep Tube that potentially could slightly decrease a benzene reading, but it doesn’t exceed 50 to 60 ppb, which is below the instrument’s resolution of 0.1 ppm (100ppb).
Effect of tube on Benzene absorption
300 250 200 150 100
50
Reading, ppb
0
110 149 225 328
Benzene concentration, ppb
ppm
* Reading 1 ppm after 1 minute exposure
by Honeywell 877-723-2878 raesystems.com
Honeywell Internal
minutes
Without tube With tube
3
Page 4
Technical Note TN-204 04/17/vk
RAE Systems
Examples of RAE-Sep Tubes’ appearance after exposure to different VOCs and environment
Unused RAE-Sep tube. Note the uniform orange color of the fresh absorbant.
The following examples show various degrees of capacitance. Do not use opened tubes that look like these.
1. Freshly opened tube exposed to a mixture of benzene and octane. The exposure was over the tube’s capacitance.
2. Open tube with instrument run for 5 minutes at 75 to 80%
relative humidity and then exposed to a mixture of benzene and octane. The tube lost capacitance. The front part didn’t absorb VOC.
3. Open tube with instrument run for 5 minutes at 75 to 80%
relative humidity and then exposed to benzene and pro­pane. The tube lost capacitance. The front part didn’t ab­sorb VOC, and the color is mostly brownish instead of green.
REFERENCES
1 California Code of Regulations, Section 5218:
https://www.dir.ca.gov/title8/5218.html
2 American Conference of Governmental Industrial Hygien-
ists (ACGIH) Threshold Limit Value: https://www.osha.gov/ dts/chemicalsampling/data/CH_220100.html
3 MultiRAE User’s Guide:
http://www.raesystems.com/sites/default/files/content/ resources/
4 MultiRAE Benzene Tube Cartridge Instructions:
http://www.raesystems.com/sites/default/files/content/ resources/MultiRAE%20Benzene%20Tube%20Car­tridge%20QSG%20-%20A5_Page%20RevA2.pdf
5 MultiRAE Benzene Datasheet:
http://www.raesystems.com/sites/default/files/content/ resources/Datasheet_MultiRAE-%20Benzene_DS-1098-
01.pdf
4. Freshly opened tube exposed to acetone. The tube’s color
almost didn’t change, though tube capacitance suffered.
by Honeywell 877-723-2878 raesystems.com
4
Honeywell Internal
Loading...