Rae RAEGuard EC, RAEGuard LEL, RAEGuard IR User Manual

Page 1
RAEGuardS EC
Toxic Gas and Oxygen Transmitter
RAEGuardS LEL
Combustible Gas Transmitter
RAEGuardS IR
Non-Dispersive Infrared (NDIR)
Transmitter
User’s Guide
P/N 033-4117-E00 Rev A August 2009
Page 2
RAEGuardS EC / LEL / IR User’s Guide
- READ BEFORE OPERATING -
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions. The user should understand how to set the correct parameters and interpret the obtained results.
CAUTION!
To reduce the risk of electric shock, turn the power off before removing the instrument cover. Disconnect the power before removing the sensor module for service. Never operate the instrument when the cover is removed. Remove instrument cover and sensor module only in an area known to be non­hazardous.
WARNING!
The calibration of all newly purchased RAE Systems instruments should be tested by exposing the sensors to known concentration calibration gas before the instrument is used or put into service. For maximum safety, the accuracy of the RAEGuard
S
should be checked by exposing the sensor(s) to
known concentration calibration gas every three (3) months.
1
Page 3
RAEGuardS EC / LEL / IR User’s Guide
Contents
1. General Information......................................................................... 3
1.1 General Specifications............................................................... 6
RAEGuard
S
EC Specifications ................................................. 6
RAEGuard
S
IR Specifications................................................... 7
RAEGuard
S
LEL Specifications................................................ 9
2. Operation....................................................................................... 10
2.1 Physical Description ................................................................ 11
2.2 Installation and Access Instructions ........................................ 11
Mounting.................................................................................. 12
Instrument Assembly Removal................................................ 13
Electrical Wiring....................................................................... 14
Instrument Assembly Installation............................................. 15
Earth Grounding Instructions................................................... 16
2.3 Display And User Interface...................................................... 18
User Interface .......................................................................... 18
Magnet Key ............................................................................. 18
Using The Magnet Key ............................................................ 19
System Initialization................................................................. 19
Reading Display ...................................................................... 20
Alarm Contacts ........................................................................ 20
2.4 Calibration ............................................................................... 21
Zero Calibration ....................................................................... 22
Span Calibration ...................................................................... 23
2.5 Advanced Menu (EC & LEL) ................................................... 25
2.6 Advanced Menu (IR)................................................................ 31
2.7 Restoring Default Values (EC & LEL) ................................... 31
3. Theory of Operation....................................................................... 33
4. Maintenance .................................................................................. 35
5. Troubleshooting............................................................................. 35
Table A: EC Sensor configuration ................................................ 37
6. ModBus/RS-485 Information ......................................................... 38
2
Page 4
RAEGuardS EC / LEL / IR User’s Guide
1. General Information
RAEGuard
S
EC The RAEGuard
S
EC is a fixed electrochemical
detector for oxygen and toxic gas (CO, H
2
S, SO2, NO, NO2, Cl2, O2,
ClO
2
, NH3, PH3, HCN and ETO). It operates with voltages from 9 to 36 VDC and provides an analog (4 to 20mA) output signal and digital (RS-485, ModBus) output signal within the corresponding ranges of gas detection. The RAEGuard
S
EC uses an interchangeable electrochemical smart sensor, which can be pre-calibrated independently offline. Housed in an explosion-proof enclosure, the RAEGuard
S
EC is equipped with a local digital display of the gas
concentration and unit of measurement, status LEDs, and function
keys for performing calibration.
RAEGuard
S
IR The RAEGuard
S
IR is fixed non-dispersive infrared (NDIR) detector for hydrocarbon combustible gases, carbon dioxide, and other gases, It operates with voltages from 9 to 36 VDC and provides an analog (4 to 20mA) output signal and digital (RS-485, ModBus) output signal within the corresponding ranges of gas detection of 0 to 100% LEL or 0 to 50,000 ppm CO
2
. The RAEGuardS IR uses a smart NDIR sensor, which can be pre-calibrated independently offline. Housed in an explosion-proof stainless steel enclosure, the RAEGuard
S
IR is equipped with a local digital display
of the gas concentration, status LEDs, and function keys for
performing calibration.
RAEGuard
S
LEL The RAEGuard
S
LEL is a fixed, highly poison­resistant catalytic bead lower explosion limit (LEL) sensor for combustible gases detection. It operates with voltages from 9 to 36 VDC and provides an analog (4 to 20mA) output signal and digital (RS-485, ModBus) output signal within the corresponding range of 0 to 100% LEL. The RAEGuard
S
LEL uses a smart LEL sensor, which can be pre-calibrated independently offline. Housed in an explosion­proof stainless-steel enclosure, the RAEGuard
S
LEL is equipped with
a local digital display of the gas concentration, status LEDs, and
function keys for performing calibration.
3
Page 5
RAEGuardS EC / LEL / IR User’s Guide
Key features:
Interchangeable smart electrochemical sensor that can be
calibrated offline
4 to 20mA analog output signal.
RS-485 digital communication in ModBus Protocol
Explosion-proof stainless-steel enclosure for hazardous
environment applications
Magnetic-key interface eliminates the need to open the explosion-
proof housing when adjusting parameters.
LCD and status/error LEDs
LED alarms when High or Low alarm point is reached
Operation at 9 to 36 VDC
Two dry contacts (<30V, 2A) normally open (or normally closed),
one for High and Low alarm, another for Fault alarm
Applications:
Waste water treatment plants
Petroleum and natural gas fields
Marine and offshore oil wells
Refineries and petrochemical plants
Solvent recovery systems
Chemical plants
Industrial safety
Sewers and pipelines
Pulp & paper plants
Heavy industry
Power plants
Steel mills
4
Page 6
RAEGuardS EC / LEL / IR User’s Guide
Hazardous Location Classification:
ATEX: II 2G Ex d IIC T6 -40° C < T
amb
< +60° C
Complies with EN60079-0 :2004, EN60079-1 :2004
Example Of Name Plate On RAEGuardS Models:
Notes:
Sensor Name is replaced by the actual sensor name. FGM-1XXXS is replaced by model number. * <
4
** Optional
5
Page 7
RAEGuardS EC / LEL / IR User’s Guide
1.1 General Specifications
RAEGuardS EC Specifications
Size
7.4" L x 5.9" W x 4.3" H
188mm x 150mm x 109mm
Weight
7.7 lbs
3.5 kg
Detector
Off-line pre-calibrated interchangeable smart
electrochemical sensor
Calibration
2-point calibration
Certification
ATEX: II 2G Ex d IIC T6 -40° C <
T
amb
< +60° C Complies with EN60079-0 :2004, EN60079­1 :2004
IP Rating
IP-65
Power
9 to 36VDC, max 40mA at 24V or 1W
Output
4-20mA RS-485, baud-rate 4.8, 9.6 or 19.2kb/sec
Sampling
Internal diffusion
Display
7-segment, 4-digit LCD and 4 color-coded alarm LEDs
User Interface
Magnetically accessed keys for non-intrusive calibration and adjustment
Temperature
-40°
C to 60° C
Humidity
0% to 95% relatively humidity (non-condensing)
Pressure
0.9 to 1.1 Atm
Dry contact
Max 30V, 2A
Dry Contacts
30V, 2A, normally open (or normally closed), one for High and Low alarm, another for Fault
alarm
Terminal
AWG 24-12
Mounting
2 holes, 5.25" (133mm) from center to center
6
Page 8
RAEGuardS EC / LEL / IR User’s Guide
RAEGuardS IR Specifications
Size
6.6” L x 5.9” W x 4.3” H
168mm x 150mm x 109mm
Weight
7.7 lbs or 3.5 kg
Detector
Protected, Off-line pre-calibrated,
interchangeable smart NDIR sensor
Calibration
2 point calibration
Range
CO
2
0 to 50000 ppm
CH
4
0 to 100% LEL
Resolution
CO
2
When CO
2
< 10000 ppm,resolution is 1 ppm;
When CO
2
>10000 ppm,resolution is 1000 ppm
CH
4
resolution is 1% LEL.
Response Time
(T
90
)
T
90
30 seconds
Certification
ATEX: II 2G Ex d IIC T6 -40° C <
T
amb
< +60° C
Complies with EN60079-0 :2004, EN60079-1 :2004
IP Rating
IP-65
Power
9 to 36VDC, max 60mA at 24V or 1.5W
Output
4-20mA RS-485 Baud rate: 4.8, 9.6 or 19.2kb/sec
Sampling
Internal diffusion
Display
7-segment, 4-digit LCD 4 color-coded LEDs
User Interface
Magnetically accessed keys for non-intrusive calibration and adjustment
Operating
Temperature
CO
2
: -20° C to +60° C
CH
4
: -40° C to +60° C
Humidity
0 to 95% relative humidity (non-condensing)
Pressure
0.9 to 1.1 Atm
Dry Contacts
30V, 2A, normally open (or normally closed), one
for High and Low alarm, another for Fault alarm
Terminal
AWG 24 to 12 (diameter 0.5mm to 2mm)
Mounting
2 holes 5.25" (133mm), from center to center
7
Page 9
RAEGuardS EC / LEL / IR User’s Guide
RAEGuardS IR Default Sensor Settings
Gas Range Resolution Span Gas
Carbon Dioxide (CO2)
0
to 50,000
ppm
10 ppm or
1000 ppm
2000 ppm to
10000 ppm
Methane (CH4)
0
to 100%
LEL
1% LEL
20% LEL to
99% LEL
Note: When CO
2
< 10000 ppm, the LCD displays the reading. When
CO
2
>10000, the LCD displays X.XE4 (which represents X.X x 10
4
).
For example, when the monitor display 1.5E4, it means 15000 ppm; when the LCD displays 3.6E4, it means 36000 ppm. The LCD on the
RAEGuard
S
IR CH4 displays the %LEL (percentage of LEL) reading.
For example, when the LCD displays 25, it means 25% LEL; when the LCD displays 75, it means 75% LEL.
8
Page 10
RAEGuardS EC / LEL / IR User’s Guide
RAEGuard
S
LEL Specifications
Size
6.6” L x 5.9” W x 4.3” H
168mm x 150mm x 109mm
Weight
7.7 lbs or 3.5 kg
Detector
Protected, Off-line pre-calibrated, interchangeable catalytic bead smart LEL
sensor
Calibration
2 point calibration
Range
0 to 100% LEL
Resolution
1% LEL
Response Time
(T
90
)
<
15 seconds to 90% of reading to 50% LEL
methane
Certification
ATEX: II 2G Ex d IIC T6 -40° C <
T
amb
< +60° C Complies with EN60079-0 :2004, EN60079­1 :2004
IP Rating
IP-65
Power
9 to 36VDC, max 60mA at 24V or 1.5W
Output
4-20mA RS-485, Baud-rate 4.8, 9.6 or 19.2kb/sec
Sampling
Internal diffusion
Display
7-segment, 4-digit LCD 4 color-coded LEDs
User Interface
Magnetically accessed keys for non­intrusive calibration and adjustment
Operating
Temperature
-40° C to +60° C
Humidity
0 to 95% relative humidity (non-condensing)
Pressure
0.9 to 1.1 Atm
Dry Contacts
30V, 2A, normally open (or normally closed), one for High and Low alarm, another for
Fault alarm
Terminal
AWG 24 to AWG 12 (diameter 0.5mm to 2mm)
Mounting
2 holes, 5.25" (133mm) from center to center
9
Page 11
RAEGuardS EC / LEL / IR User’s Guide
2. Operation
The calibration of all newly purchased RAE Systems instruments should be tested by exposing the sensor(s) to a known concentration calibration gas before the instrument is used or put into service. For maximum safety, the accuracy of the RAEGuard
S
EC should be checked by exposing the sensor(s) to a known concentration calibration gas, after a period of time.
Calibration should be verified daily during the period of initial use in the intended atmosphere to ensure nothing is poisoning the sensor(s). The period of initial use must be of sufficient duration to ensure that the sensors are exposed to all conditions that might have an adverse effect on the sensors.
Verify the calibration with a known concentration test gas before use. This “bump” test is very simple. Recalibrate the unit if readings are off.
Prior to factory shipment, the RAEGuard
S
EC is calibrated and tested using span gas. However, the user should calibrate the instrument before the first use. After the unit is installed, run for 24 hours, and calibrated, it is ready for immediate operation. Calibration setting values for EC sensors are shown as Table A (on page 32).
Kit Accessories include: Calibration Adapter, RAEGuard Magnet
Key, and User’s Guide.
10
Page 12
RAEGuardS EC / LEL / IR User’s Guide
11
2.1 Physical Description
The design of RAEGuardS allows it to be easily mounted and interfaced to a fixed-point gas monitoring system. The RAEGuardS EC and LEL are housed in a 7.4" L x 5.9" W x 4.3" H (188mm x 150mm x 109mm) case with two holes 5.25" (133 mm) from center to center. The RAEGuard
S
IR’s dimensions vary slightly because of the
sensor size: 6.6" L x 5.9" W x 4.3" H (168mm x 150mm x 109mm).
5.0" (127)
5.24" (133)
Ø3.15"(8)
3/4" -14NPT
7.40" (188)
4.92" (125)
Note: This dimension
is 6.6
" (168mm) on the
RAEGuardS IR and LEL.
Ø 4 1/2" (114.30)
3/4" -14NPT
1.04" (26.4)
4.29" (109.0)
Page 13
RAEGuardS EC / LEL / IR User’s Guide
2.2 Installation and Access Instructions
WARNING
1. A minimum of 18" (457mm) of explosion-proof conduit must be used at cable entry in group A and B atmospheres. CSA requires seals in conduit exceeding 5' (1.524 m) in group C atmospheres.
2. To prevent ignition of hazardous atmospheres, area must be free of flammable vapors and supply circuit must be disconnected before removing cover.
3. For European application, the installation must comply with the requirements of EN 60079-14.
Mounting
First, decide where the transmitter will be mounted. (Refer to
installation drawing, below.) Drill two holes in mounting surface, with
the center of the holes 5.25" (133mm) apart.
12
Page 14
RAEGuardS EC / LEL / IR User’s Guide
Instrument Assembly Removal (EC and LEL)
Prior to service: Make sure power is OFF. Observe all Hazardous Location Safety procedures.
1. Unscrew the housing lid from the housing body by rotating it counter­clockwise.
A
s shipped, one of the conduit holes is covered by the provided hex-head plug. The other conduit is shipped with connected wires.
2. Pull off the retaining clip to release the
instrument assembly.
3. Tilt the instrument assembly
90˚.
4. Unlock the black 16-pin connector.
5. Lift the entire instrument assembly out of the housing.
6. Unlock the white 8-pin connector.
7. Remove the sensor.
Picture shows disassembled instrument
13
Page 15
RAEGuardS EC / LEL / IR User’s Guide
Instrument Assembly Removal (IR)
Prior to service: Make sure power is OFF. Observe all Hazardous Location Safety procedures.
1. Unscrew the housing lid from the housing body by rotating it counter­clockwise.
A
s shipped, one of the conduit holes is covered by the provided hex-head plug. The other conduit is shipped with connected wires.
6. Unlock the 12-pin connector.
2. Pull off the retaining clip
to release the
instrument assembly.
7. Remove the sensor.
Picture shows disassembled instrument
6. Tilt the instrument assembly
90˚.
7. Unlock the 20-pin connector.
8. Lift the entire instrument assembly out of the housing.
14
Page 16
RAEGuardS EC / LEL / IR User’s Guide
Electrical Wiring
1. Inside the housing bottom, unplug the two green terminal block
plugs from the terminal block on the PC boards. Note: The
terminal block plugs accept 12 AWG to 24 AWG wire.
2. Lace the wires through the RAEGuardS EC’s wire hole(s) and
connect wires to the corresponding pin numbers of the terminal blocks:
Terminal Wire Pin#
Alarm Common (COM) 1 High/Low Alarm (ALM1) 2
Fault Alarm (ALM2) 3 RS485A (485A) 4
Block 1
RS485B (485B) 5 4-20 mA Output (4-20mA) 6
Power Supply - Output Common (P-) 7
Block 2
Power Supply + (9 to 36VDC) (P+) 8
Instrument Assembly Installation
1. Plug all terminal block plugs into the correct terminal block headers.
Keep the extra wires as close to the inside enclosure wall as possible.
2. Screw the sensor module to the housing and plug the 8-pin (EC) or 10-
pin (LEL model) or 12-pin (IR model) connector back into its socket.
3. Plug the 16-pin (EC or LEL model) or 20-pin (IR model) connector back
into its socket.
4. Pull off the clip to place the instrument assembly back into the housing
bottom.
5. Tightly screw the housing top to the housing bottom.
15
Page 17
RAEGuardS EC / LEL / IR User’s Guide
Earth Grounding Instructions
External Earth Grounding
Fasten the crimped ground wire with hardware as illustrated below. The wire should have a minimum cross-section area of 4mm
2
for its
conductor.
Wire (Min. 4 mm
2
cross-section)
Crimped terminal
M4 flat washer
M4 Split washer
M4 Screw
Internal Earth Grounding
Use the same hardware as shown in the illustration of external earth grounding. The wire should be no less than the size of the power lines.
Internal earth grounding
16
Page 18
RAEGuardS EC / LEL / IR User’s Guide
Finished Grounding Wires
Internal and external grounding are shown here, as well as an alternate external grounding point. Always follow local electrical guidelines.
External earth grounding
Internal earth grounding
Alternate grounding point
17
Page 19
RAEGuardS EC / LEL / IR User’s Guide
2.3 Display And User Interface
User Interface
The RAEGuardS EC’s user interface consists of four status LEDs, a 4­digit LCD display, and three keys, [+], [MODE], and [-]. The three keys are operated by using the Magnet Key.
RAEGuardS EC user interface.
Magnet Key
The RAEGuardS EC has no external switches, but instead uses the magnetic end of the RAEGuard Magnet Key (p/n 033-2032-000) to activate switches built into the unit.
Magnet
SIDE VIEW
TOP VIEW
RAEGuard Magnet Key.
18
Page 20
RAEGuardS EC / LEL / IR User’s Guide
Using The Magnet Key
Using the magnet end of the Magnet Key, briefly touch the glass above the MODE circle or the triangles labeled [+] and [-]. Then remove the key straight out and away from the RAEGuard
S
EC.
RAEGuard Magnet Key touching glass above the [+] triangle.
Important! Never drag the key sideways, or two functions may be
activated.
System Initialization
When the RAEGuardS EC system power is turned on, it initializes and an “InIt” message appears on the display. As the transmitter is warming up, each component is checked, and the LEDs blink all at once. The countdown timer appears on the screen for the warm-up.
19
Page 21
RAEGuardS EC / LEL / IR User’s Guide
Reading Display
As the transmitter enters the Reading Display, it automatically starts testing for errors and goes through a cycle of checking each alarm condition. If there are no errors or alarm conditions, the green “OK” LED is lit and the gas concentration is displayed.
If there is an error, the “Fault” LED blinks and an error message blinks. Each alarm condition has a corresponding LED that blinks an amber color when the readings are outside a specified range or limit.
Alarm Contacts
The alarm contacts or alarm relay can be used to drive user-supplied external alarms such as a light or buzzer. The external alarms open in normal conditions and close when an alarm occurs.
Default Alarm Relay Logic
External
Alarm
LED LCD Analog Output
Exceeds Low alarm limit
ALM 1 Alarm Low reading Based on
reading
Exceeds High alarm limit
ALM 1 Alarm High reading Based on
reading
Over Range ALM 2 Alarm High 8888 22mA
Calibration fail ALM 2 Alarm Fault Flashing
E003
2mA
Sensor drift* ALM 2 Alarm Fault Flashing
E004
2mA
ADC saturated *(max)
ALM 2 Alarm Fault Flashing
E005
2mA
* RAEGuardS EC and LEL only.
20
Page 22
RAEGuardS EC / LEL / IR User’s Guide
2.4 Calibration
WARNING
The calibration of all newly purchased RAE Systems instruments should be tested by exposing the sensor(s) to a known concentration calibration gas before the instrument is used or put into service. For maximum safety, the accuracy of the RAEGuard
S
should be checked by exposing the sensor to a
known concentration calibration gas, after a period of time.
T
he RAEGuard
S
is calibrated using a two-point calibration process. First, use a “zero gas.” Then use a “span gas” containing a known concentration of a standard reference gas, to set the second point of reference.
Note:
The zero
calibration must be performed before the span calibration.
RAEGuard
S
connected to gas cylinder with a calibration adapter.
Gas calibration requires a bottle of zero gas, a bottle of span gas, and a calibration adapter.
21
Page 23
RAEGuardS EC / LEL / IR User’s Guide
Zero Calibration
1. Connect the zero gas cylinder to the metal filter gas adapter on
the RAEGuard
S
transmitter using the provided calibration
adapter.
2. To access the Calibration menu, press [MODE] from the Reading
Display. Zero calibration appears first, as indicated by the “Zero” message on the screen.
Optional: To advance to span calibration, press [MODE] a
second time. “Span” appears on the screen. Skip to Span Calibration instructions on the next page.
Optional: To exit the Calibration menu, press [-] to return to the
Reading Display.
3. Turn on the gas flow. Allow gas to flow into the sensor for 30
seconds before zero calibration. Then press [+] to start calibration. The “Zero” LED starts blinking, and the “Zero” message alternates with a countdown timer.
Optional: Before the countdown reaches zero, you may press
any key to interrupt zero calibration and advance to span calibration.
4. Once the countdown reaches zero, the “Zero” LED stops
blinking, and the zero calibration data is saved.
Note: The transmitter returns to the Reading Display after 60 seconds of idle time.
5. Turn off the zero calibration gas and remove the cylinder.
22
Page 24
RAEGuardS EC / LEL / IR User’s Guide
Zero calibration automatically advances to span calibration when complete.
Span Calibration
1. Connect the span gas cylinder to the metal filter gas adapter on
the RAEGuard
S
transmitter using the provided calibration
adapter.
23
Page 25
RAEGuardS EC / LEL / IR User’s Guide
Optional: To access span calibration from the Reading Display,
press [MODE]. After “Zero” appears on the screen, press
[MODE] a second time to advance to span calibration.
Optional: To access span calibration after zero calibration has
already started, press any key to advance to span calibration.
Optional: To exit the calibration menu, press [-] to return to the
Reading Display.
Note: It is not necessary to exit manually. After 60 seconds of idle time,
the RAEGuard
S
automatically returns to the Reading Display.
2. Turn on the gas flow. First, let gas flow into the sensor for 30
seconds before span calibration.Then press [+] to start
calibration. The “Span” LED starts blinking. The “Span” message alternates with a countdown timer.
Note: Wait for the entire countdown for a complete calibration.
Optional: Before the countdown reaches zero, you may press
any key to interrupt span calibration and return to the Reading Display.
3. Once the countdown reaches zero, the “Span” LED stops
blinking.
4. If the sensor sensitivity is unacceptable, the screen alternately
flashes a “Span” and “FAIL” message until any key is pressed.
Note: If span calibration fails, it may be necessary to replace the
sensor.
5. You may choose to press [-] or [MODE] to return to zero
calibration and restart the entire calibration process, or press [+] to go to the Reading Display.
24
Page 26
RAEGuardS EC / LEL / IR User’s Guide
6. If the sensor’s sensitivity is acceptable, the span data is
calculated and saved.
7. The calibration procedure is complete. After a few seconds, the
transmitter returns to the Reading Display.
8. Turn off span calibration gas and remove the cylinder.
2.5 Advanced Menu (EC & LEL)
The Advanced menu on the RAEGuardS EC and LEL allows you to change the values for each reading and setup option. To access the
Advanced menu from the Reading Display, first press [+], [-],and then
[MODE] in sequence. Then scroll through the parameters shown below by pressing [MODE].
Once you have entered the Advanced Menu, you can edit the parameters for each item in the list.
25
Page 27
RAEGuardS EC / LEL / IR User’s Guide
Note: Anytime a submenu screen is idle for more than 60 seconds,
the transmitter returns to the Reading Display.
Advanced Menu
Display Explanation
CALu Calibration Concentration Value CF Correction Factor, only for RAEGuardS LEL and IR LO Low Alarm HI High Alarm Id Client ID bAUd Baud rate. Supports 19200, 9600, or 4800 LItE Backlight Aout Analog output 4mA and 20mA adjustment tESt Test Function of LED, External Alarm, and 4-20mA
To change a value, press [MODE] until the desired parameter appears.
Press [+] to enter the Process Menu.
To increase a value, press [+].
To decrease a value, press [-].
When finished changing the value, press [MODE].
If a value has changed, the new value blinks on the LCD.
Press [-] or [MODE] to discard changes and advance to the next submenu item.
Press [+] to save changes.
The “SAVE” message appears on the screen to confirm changes have been saved.
26
Page 28
RAEGuardS EC / LEL / IR User’s Guide
After entering the Advanced Menu, step through the parameters by pressing [MODE]. To edit a parameter, press [+]. The screen shows the current reading value of the parameter.
You can exit Advanced Menu at any screen by pressing [-]. In addition, Advanced Menu is automatically exited if you do not make any changes within 60 seconds.
27
Page 29
RAEGuardS EC / LEL / IR User’s Guide
This diagram shows how to edit parameter values and save changes for most of the parameters:
28
Page 30
RAEGuardS EC / LEL / IR User’s Guide
This diagram shows the submenu of parameters for the 4-20mA Analog Menu:
29
Page 31
RAEGuardS EC / LEL / IR User’s Guide
This diagram shows the submenu of parameters for the Test Menu:
* The analog output steps from 4mA to 20mA in 1mA steps and then steps back down to 4mA before returning to the upper menu. Each step is held for 3 seconds. Press any key to interrupt the process.
30
Page 32
RAEGuardS EC / LEL / IR User’s Guide
2.6 Advanced Menu (IR)
The Advanced menu allows the user to change the values for each reading and setup option. To access the Advanced menu from the
Reading Display, first press [+], [-],and then [MODE] in sequence.
Then scroll through the parameters shown below by pressing
[MODE].
Anytime a submenu screen is idle for more than 60 seconds, the transmitter will return to the Reading Display.
Advanced Menu
Submenu LCD display Default
Standard gas concentration
CO2: C5E3
(represents 5000 ppm)
CH
4
: C050
(represents 50% LEL)
CO
2
: C5E3
CH
4
: C050
Low alarm CO2: L4E3 (4000 ppm)
CH
4
: L020 (20% LEL)
CO
2
: L4E3
CH
4
: L020
High alarm CO2: H8E3 (8000 ppm)
CH
4
: H050 (50% LEL)
CO
2
: H8E3
CH
4
: H050
User ID IdXX Id00
Baud-rate (kBaud)
B19.2, B09.6, B04.8 (19200, 9600, and 4800)
B09.6
4mA output adjust
A04
20mA output adjust
A20
31
Page 33
RAEGuardS EC / LEL / IR User’s Guide
To change a value, press [MODE] until the desired parameter appears. Then press [+] to enter the Process Menu.
To increase a value, press [+].
To decrease a value, press [-].
Press [MODE] when finished changing the value. If a value has
changed, the new value blinks on the LCD.
Press [-] or [MODE] to discard changes and advance to the next
submenu item.
Press [+] to save changes. The “SAVE” message appears on the
screen to confirm changes have been saved.
2.7 Restoring Default Values (EC & LEL)
You can reset all values in the RAEGuardS EC and LEL to factory default values. Follow this procedure:
1. Press [MODE] to Zero.
2. Press [MODE] to SPAN.
3. Press [MODE] key to show DEFT in the display.
4. Press [+]. The display flashes DEFT and awaits confimation.
5. Press [+] to Save the change. All data is cleared and the default
values are now restored.
32
Page 34
RAEGuardS EC / LEL / IR User’s Guide
3. Theory of Operation
The RAEGuard
S
EC uses electrochemical sensors. The electrons
generated at the working electrode travel via the contact pins to be measured at the external circuit and are returned to the counter electrode to complete the circuit, The current of the sensor response is proportional to the related EC sensors’ gas concentration.
The RAEGuard
S
IR is based on the patented technology of non-
dispersive infrared (NDIR) for hydrocarbon, carbon dioxide and other gas detection. After sample gas diffuses into the infrared optical gas sensor, the electro-magnetic radiation cause oscillation on gas molecules. If this happens the oscillation reduces the amount of radiation, it absorbs a certain degree of radiation. The absorbance at specified wavelength is a measure of the present gas concentration.
The RAEGuard
S
LEL uses a highly poison-resistant catalytic bead
lower explosion limit (LEL) sensor. After sample gas diffuses through the metal sinter into the combustion chamber, a pair of combustion elements inside the combustion chamber burns the combustible gas and generates an electrical signal.
The RAEGuard
S
is a microcontroller-based instrument. After the electrical signal is conditioned and converted to digital, the microcontroller processes the data, which displays the results locally and reconstructs the digital data into a standard 4-20 mA analog current output signal and RS-485 digital output signal.
The instrument has three magnetic sensing keys. The user may calibrate the transmitter and change the alarm level setting via the magnetic front-panel key pad.
33
Page 35
RAEGuardS EC / LEL / IR User’s Guide
The instrument is powered by a DC power supply located in a safe area. The power supply on the PC board contains a switching and linear regulator that converts the main input voltage to 3.3V and 5V DC to power the entire circuit.
34
Page 36
RAEGuardS EC / LEL / IR User’s Guide
4. Maintenance
As a guide, it is recommended to regularly “bump test” a RAEGuardS unit with a known percentage gas.
LCD
Main PCB
Enclosure Cover
Cable
Socket PCB
Enclosure Body
7R EC, LEL, or NDIR Sensor
Holder
Exploded View of RAEGuardS EC and LEL Components
Periodically examine the sensor’s opening to make sure it is not dirty or covered in dust or debris.
If the sensor requires replacement, refer this unit to qualified service personnel.
35
Page 37
RAEGuardS EC / LEL / IR User’s Guide
5. Troubleshooting
Note: Before diagnosing measurement problems, perform zero and
span calibration.
Symptom Reason & Solution
E001
Reason: Sensor not installed Solution: Install the sensor
E002
Reason: Wrong sensor or Sensor
EEPROM error
Solution: Replace sensor
E003
Reason: Calibration failure Solution: Make sure of standard gas flow
and perform recalibration
Replace Sensor
E004
Reason: EC sensor zero drift Solution: Recalibration
E005
Reason: EC sensor exceeds max raw
count
Solution: Call RAE Systems technical
support
E006
Reason: Wire connection error Solution: Check and reconnect wires
E007
Reason: Main PCB EEPROM error Solution: Check and replace main PCB
Couldn’t turn
on the unit
Reason: Wrong position of one switch
(S3)
Solution: Check the position of switch S3
Reading
abnormally
High
Reason: Calibration failure Solution: Recalibration Eliminate/control source of
temperature difference
“Span failure”
signal
Reason: Sensor broken Solution: Replace Sensor
Low 4-20mA
output
Reason: Power supply voltage is lower
than specified.
Solution: Check power supply voltage
and connection.
“Err”
Reason: Sensor ID is creating an error. Solution: Check that the correct sensor is
installed. Replace the sensor or update the firmware.
36
Page 38
RAEGuardS EC / LEL / IR User’s Guide
Table A: EC Sensor configuration
The following table contains specifications for sensors in the RAEGuard
S
EC Series.
Sensors Range
(ppm)
Resolution
(ppm)
Response
Time (T
90
)
Span
(ppm)
Low
(ppm)
High
(ppm)
CO 0-1000 1 <30s 50 35 200 H2S 0-300 0.1 <20s 25 10 20 SO2 0-150 0.1 <30s 5 2 10
NO 0-1000 0.5 <20s 25 25 50
NO2 0-50 0.1 <25s 5 1 10
Cl2 0-30 0.1 <30s 10 0.5 5
O2 0-30% 0.1% <20s 20.9% 19.5% 23.5%
ClO2 0-1 0.01 <120s 0.5 0.2 0.5
NH
3
0-100 0.5 <60s 50 25 50
PH3 0-20 0.05 <30s 5 1 2
HCN 0-100 0.5 <200s 10 4.7 50
ETO 0-100 0.1
<80s
37
Page 39
RAEGuardS EC / LEL / IR User’s Guide
6. ModBus/RS-485 Information
Retrieving Gas Concentration Data From A RAEGuard
S
Via RS-485
The RAEGuard s communicates by means of MODBUS RTU. All
monitors provide a 4-byte register value. Note: Gas concentration is
the only value that can be retrieved.
As example be low 34 hex = 52 decimal
Overview
This document describes the modbus protocol used in RAEGuardS products.
1. Communication Setting
Transmission MODE: RTU Controller: PC or Controller Client: RAEGuard
S
EC, LEL, IR Baud Rate: 4800, 9600, 19200 Client ID: 1 to 99 (0x0001 to 0x0063), except for 58 (0x3A)
2. Message Frame/Communications Procedure
RAEGuardS EC/LEL/IR only support Function Code 0x03 (read holding registers), which only supports the “Get Reading Value” from the detector.
38
Page 40
RAEGuardS EC / LEL / IR User’s Guide
39
0x03: Read Holding Registers Requesting Message:
Device Address
Func­tion Code
Register Address High Byte
Register Address Low Byte
Quantity of Registers High Byte
Quantity of Registers Low Byte
CRC Low Byte
CRC High Byte
Client ID 03 00 08 00 02 CRC CRC
Answering Message:
Device Address
Function Code
Byte Count
Register Value CRC
Low Byte
CRC High Byte
Client ID 03 04 Reading
Highest Byte
Reading Higher Byte
Reading High Byte
Reading Low Byte
CRC CRC
Note: The length of data from the detector is 4 bytes.
Example:
Request: 01 03 00 09 00 02 14 09 Answer: 01 03 04 00 00 00 34 FB E4
Note: The maximum distance should be less than 1 km when using a
1.5 mm
2
cable.
Page 41
P/N 033-4117-E00 Rev A August 2009
Loading...