Rae RAEGuard 2 PID User Manual

Page 1
User’s Guide
RAEGuard 2 PID
P/N D03-4001-000 Rev D April 2012
Page 2
RAEGuard 2 PID User’s Guide
Contents
Section 1: RAEGuard 2 PID User’s Guide
1 General Informat io n ............................................................ 2
2 General Specifications ......................................................... 4
2.1 RAEGuard 2 PID Specifications ........................... 4
2.2 Proper Product Disposal At End Of Life ............... 5
3 Operation ............................................................................. 6
4 Physical Description ............................................................ 7
5 Physical Dimensions ........................................................... 7
6 Installation And Access Instructions ................................... 8
6.1 Mounting ............................................................... 8
6.2 Instrument Disassembly ...................................... 10
6.3 Electrical Wiring ................................................. 11
6.4 Wiring Procedure ................................................. 12
6.5 Earth Grounding Instructions .............................. 13
6.5.1 External Earth Grounding ................................ 13
6.5.2 Internal Earth Grounding ................................. 13
6.5.3 Finished Grounding Wires .............................. 14
6.6 Alarm Contact Setup ........................................... 14
7 Display And User Interface ............................................... 16
7.1 User Interface ...................................................... 16
7.2 Magnet Key ......................................................... 16
7.3 Using The Magnet Key ....................................... 17
7.4 System Initialization ............................................ 17
7.5 Reading Display .................................................. 17
7.6 Instrument Status Display .................................... 18
8 Navigating Settings ........................................................... 19
9 Programming Menus ......................................................... 23
Zero/Fresh Calibration......................................................... 23
Meas. Unit* ...................................................................... 23
High Alarm ....................................................................... 23
Temp. Unit ........................................................................ 23
Span Calibration ................................................................. 23
Meas. Range* ................................................................... 23
Low Alarm ........................................................................ 23
Language .......................................................................... 23
Page 3
RAEGuard 2 PID User’s Guide
2
Set Span Value ................................................................. 23
Meas. Gas* ....................................................................... 23
Ext. Alarm Delay* ............................................................ 23
Pump Duty (%)* ............................................................... 23
Calibration Gas* ............................................................... 23
Pump Cycle(s)* ................................................................ 23
Pump Status* .................................................................... 23
Bus Baudrate .................................................................... 23
Analogout 4mA ................................................................ 23
Analogout 20mA ................................................................ 23
Unit ID .............................................................................. 23
LCD Contrast .................................................................... 23
LCD Backlight .................................................................. 23
Change Password* ............................................................ 23
9.1 Entering Programming Mode .............................. 24
9.2 Calibration ........................................................... 25
9.3 Measurement (Advanced Mode) ......................... 27
9.4 Alarm Setting ....................................................... 31
9.5 Monitor Setup ...................................................... 33
10 Calibration .................................................................. 39
10.1 Zero Calibration ................................................... 40
10.2 Span Calibration .................................................. 42
11 Alarm Signal Summary .............................................. 44
12 Maintenance ............................................................... 45
13 Troubleshooting .......................................................... 45
14 Appendix A: Range, Sensor and Related
Configuration .............................................................. 46
15 ModBus/RS-485 Information ..................................... 46
16 Appendix B: Controlled Section ................................ 48
16.1 Scope ................................................................... 48
16.2 Contents ............................................................... 48
16.3 Scope ................................................................... 50
16.4 Responsibility ...................................................... 50
16.5 Contents ............................................................... 50
Page 4
RAEGuard 2 PID User’s Guide
3
Section 2: DigiPID User’s Guide
Safety Instructions .................................................................... 57
17 Read Before Operating ............................................... 58
17.1 DigiPID Marking ................................................. 59
17.2 Operation Area and Conditions ........................... 60
17.2.1 Hazardous Areas classified by Zones ............ 60
Hazardous Areas classified by Divisions ................. 60
17.3 Instruction For Safe Use ...................................... 60
17.4 Use In Hazardous Areas ...................................... 61
17.5 Year of manufacture ............................................ 61
17.6 Specifications ...................................................... 62
18 General Informat io n ................................................... 63
19 Grounding (Earth Connection) ................................... 64
20 Physical Description ................................................... 65
21 Sensor Parts And Dimensions .................................... 66
22 Operating the Sensor Module ..................................... 67
22.1 Preparing The Sensor For Use ............................. 67
23 Using The Sensor Module .......................................... 68
23.1 Gas Flow Routing Connection ............................ 68
23.2 Sensor Module Calibration .................................. 69
23.3 Maintenance And Calibration .............................. 70
23.4 Replacing the Lamp And Filter ........................... 70
24 Replacing The Sensor’s Teflon UV Shield ................ 72
25 Sensor & Lamp Cleaning/Replacement ..................... 75
25.1 Cleaning The PID Sensor .................................... 75
25.2 Cleaning The Lamp Housing Or Changing
The Lamp ............................................................. 76
25.3 Cleaning The Instrument ..................................... 77
25.4 Ordering Replacement Parts ................................ 77
25.5 Special Servicing Note ........................................ 77
26 Electronic Waste Disposal .......................................... 78
27 Technical Support ....................................................... 79
28 RAE Systems Contacts ............................................... 79
Page 5
RAEGuard 2 PID User’s Guide
4
© Copyright 2012 RAE Systems by Honeywell
Page 6
RAEGuard 2 PID User’s Guide
1
Section 1: RAEGuard 2 PID User’s Guide
Page 7
RAEGuard 2 PID User’s Guide
2
1 General Information
The RAEGuard 2 PID is a fixed photoionization detector (PID) that measures a broad range of volatile organic compounds (VOCs). It operates on 10 to 28 VDC and provides an analog (4-20mA) three­wire signal output, and RS-485 Modbus digital signal output. RAEGuard 2 PID uses an intelligent processing platform and digital sensor technology, supports multiple ranges and resolutions, and can perform off-line calibration and environmental self-adaptation. RAEGuard 2 PID has a graphic display and local sound alarm and light status indicator. A magnetic key interface enables the detector to be calibrated and operational parameters adjusted with the explosion­proof enclosure in place. In addition, the PID module can be easily removed in hazardous locations for calibration or maintenance.
Key features:
Digital Smart Sensor Technology
Three wires, supports 4-20mA analog output
Can operate with diffusion and flow-through modes and can be
used in most environments
RS-485 digital communication in ModBus protocol
Explosion-proof stainless-steel enclosure for hazardous
environment applications
Magnetic-key interface eliminates the need to open the explosion-
proof housing when adjusting parameters.
Matrix LCD of 128x64 supports graphic display
Three-level switch signal outputs, normally closed/normally open
optional (30V load voltage, 2A)
LED alarms when High or Low alarm point is reached
Operation at 10 to 28 VDC
Three dry contacts (<30V, 2A) normally open (or normally closed),
one for High and Low alarm, another for Fault alarm
Page 8
RAEGuard 2 PID User’s Guide
3
Applications:
Refineries, petrochemical and natural gas plants
Metallurgical
Chemical, medication
Environmental protection
Electricity, communications
Fire protec ti on
Utilities
Pulp and paper, printing
Storage
Sewage disposal
Food, brewing
Scientific research, education, homeland security
Hazardous Location Classification:
ATEX
II 2G Ex d ia II CT4 Gb
UL/CSA
Class I, Div. 1, Groups A B C D T4 - pending
IECEx
Ex d ia IIC T4 Gb
Page 9
RAEGuard 2 PID User’s Guide
4
2 General Specifications
2.1 RAEGuard 2 PID Specifications
Basic Parameters
Principle
PID (photoionization detector)
Sensor
Digital Smart Sensor
Sampling
Internal diaphragm pump
Working Current
DC 10 to 28V, 210mA at 24V
Power
<5W
Output
4-20mA
Three-level programmable alarm relays
(30 VDC, 2A)
RS-485 (Supports Modbus)
IP Rating
IP-65
Mechanical Interface
3/4" NPT Male
Installation
2" pipe-holding or wall mounting
User Interface
Three-key magnetic bar adjustment
Calibration
Two points
Environmental Parameters
Temperature
-20° C to +55° C (-4° F to 131° F)
Humidity
0 to 95% relative humidity, non-condensing
Pressure
90 to 110kPa
Display
Display
128x64 matrix backlit LCD, supports graphic display
Page 10
RAEGuard 2 PID User’s Guide
5
RAEGuard 2 PID Specifications (continued)
Physical Parameters
Dimensions, L x W x H
257 x 201 x 107 mm (10.1" x 7.9" x 4.2")
Material
Stainless steel
Weight
3.5 kg (7.7 lbs)
Certification
ATEX
II 2G Ex d ia II CT4 Gb
UL/CSA
Class I, Div. 1, Groups A B C D T4 - pending
IECEx
Ex d ia IIC T4 Gb
Ex Entity Parameters
Um: 28VDC (input supply)
2.2 Proper Produc t Disposal At End Of Life
This symbol (crossed-out wheeled bin) indicates separate collection of waste electrical and electronic equipment in the EU countries. This product may contain one or more Nickel-metal hydride (NiMH), Lithium-ion, or Alkaline batteries. Specific battery information is given in this user guide.
Page 11
RAEGuard 2 PID User’s Guide
6
3 Operation
The calibration of all newly purchased RAE Systems instruments should be tested by exposing the sensor to a known concentration calibration gas before the instrument is used or put into service. Prior to factory shipment, the RAEGuard 2 PID is calibrated and tested. However, the user should calibrate the instrument before the first use.
Kit Accessories include: RAEGuard 2 Magnet Key, and User’s Guide.
Page 12
RAEGuard 2 PID User’s Guide
7
4 Physical D escrip tio n
The RAEGuard 2 PID can be easily installed and integrated with various control systems. It is designed with flexible pipe­holding/wall-mounting options and standard connection terminals.
5 Physical D imen sion s
The physical dimensions are as shown:
201mm (7.9")
290 mm (11.4")
257mm
(5.23")
107mm
(4.2")
Page 13
RAEGuard 2 PID User’s Guide
8
6 Installation And A ccess Instructions
6.1 Mounting
First, decide where the transmitter will be mounted. (Refer to installation drawing, below.) Drill two holes in mounting surface, with the center of the holes 5.25" (133mm) apart.
Besides directly mounting the RAEGuard 2 PID to a wall, it can be mounted on a pipe.
WARNING
1. To prevent ignition of hazardous atmospheres, area must
be free of flammable vapors and supply circuit must be disconnected before removing cover.
2. For European applicati o n, the installation must comply with the requirements of EN 60079-14.
Page 14
RAEGuard 2 PID User’s Guide
9
Note: When installing the RAEGuard 2 PID, make sure the sensor is vertically oriented (pointing straight down).
Wall
Pole
Page 15
RAEGuard 2 PID User’s Guide
10
6.2 Instrument Disassembly
Prior to service: Make sure power is OFF. Observe all Hazardous Location Safety procedures.
1. Loosen the fastening bolt
before unscrewing the housing lid. Un screw the housing lid from the housi ng body by rotating it counter­clockwise. (As shipped, one of the conduit holes is covered by the provided hex­head plug. The o ther condu it is for wire connections.)
3. Tilt the instrument assembly
90˚.
4.
Unlock the 24-pin connector on the ribbon cable.
5.
Lift the entire instrument assembly out of the housing.
6.
Unscrew the sensor by turning it counterclockwise.
Disassembled instrument.
2. Pull off the retaining
clip to release the instrument assembly.
Page 16
RAEGuard 2 PID User’s Guide
11
To reassemble the instrum ent: 1 Reconnect the 24-pin connector of the ribbon cable and lock it in
place.
2 Fasten the cable to the board. 3 Mate the board with the clip on the left side and tilt it over. 4 Click the board into place. 5 Screw on the sensor, making sure not to block the gas inlet hole
and connectors.
6 Screw on the housing top. 7 Screw down the fastening bolt.
6.3 Electrical Wiring
The detector connects to peripheral equipment via three terminal blocks. These blocks accept 12AWG to 24AWG wire (0.2 to 4.0 mm
2
).
Terminal blocks
Page 17
RAEGuard 2 PID User’s Guide
12
6.4 Wiring Procedure
1. Inside the housing bottom, unplug the two green terminal block
plugs from the terminal block on the PC boards.
Note: The terminal block plugs accept 12 AWG to 24 AWG wire.
2. Lace the wires through the RAEGuard 2 PID’s wire hole(s) and
connect wires to the corresponding pin numbers of the terminal blocks:
Terminal Terminal Definition
Terminal
description
No.
Block 1
Failure Alarm Contact
ERR
1
Low Level Alarm Contact
LOW
2
High Level Alarm Contact
HIGH
3
Switch signal Common Contact
COM
4
Block 2
4-20 mA Output
OUT
5
Power Supply Negative
GND
6
Power Supply Positive
+V
7
Block 3
RS485A terminal
485A
8
RS485B terminal
485B
9
Page 18
RAEGuard 2 PID User’s Guide
13
6.5 Earth Grounding Instructions
6.5.1 External Earth Grounding
Fasten the crimped ground wire with hardware as illustrated below. The wire should have a minimum cross-section area of 4mm
2
for its
conductor.
6.5.2 Internal Earth Grounding
Use the same hardware as shown in the illustration of external earth grounding. The wire should be no less than the size of the power lines. Signal grounding can connect to a cable’s shielding layer if shielded cable is used. If a separate wire is used for grounding, its cross section should be greater than that of the power line.
M4 flat washer
M4 Split washer
M4 Screw
Wire
(Min. 4 mm2 cross-section)
Crimped terminal
Internal earth
grounding
Page 19
RAEGuard 2 PID User’s Guide
14
6.5.3 Finished Grounding Wires
Internal and external grounding are shown here, as well as an alternate external grounding point. Always follow local electrical guidelines.
6.6 Alarm Contact Setup
The alarm contacts can drive external alarms such as a light or buzzer. By default, the three relays are set for normally open (NO) operation, meaning that contact is closed when in alarm.
You can separately change each relay from normally open to normally closed (NC) operation by changing the position of the jumpers on the internal printed circuit board.
Alternate
earth grounding point
Internal earth
grounding
External earth
grounding
Page 20
RAEGuard 2 PID User’s Guide
15
Three jumper blocks are located below the three green terminal blocks. They are labeled (from left to right): J4, J3, and J1. Here are their functions:
Jumper
Function
J1
High
J3
Low
J4
Error
Lift off the jumper for each one that you want to change and slide it back on, either connecting the middle and top points or middle and bottom points.
Important!
Once the system is operational, test the functionality of all three relays.
Jumper blocks
Normally
Open
Normally
Closed
Page 21
RAEGuard 2 PID User’s Guide
16
7 Display And User Interface
7.1 User Interface
The RAEGuard 2 PID’s user interface consists of a status LED, an LCD display, and three keys, [+], [MODE], and [-]. The three keys are operated by using the Magnet Key.
RAEGuard 2 PID user interface.
7.2 Magnet Key
The RAEGuard 2 PID has no external switches, but instead uses the magnetic end of the RAEGuard Magnet Key (p/n 033-3032-000) to activate switches built into the un it.
Magnet
SIDE VIEW
TOP VIEW
Page 22
RAEGuard 2 PID User’s Guide
17
7.3 Using The Magnet Key
Using the magnet end of the Magnet Key, briefly touch the glass above the MODE circle or the triangles labeled [+] and [-]. Then remove the key straight out and away from the RAEGuard 2 PID.
RAEGuard Magnet Key touching glass above the [+] triangle.
Important! Never drag the key sideways, or two functions may be activated.
7.4 System Initialization
When the RAEGuard 2 PID’s system power is turned on, it initializes and a RAE Systems logo appears on the display. As the transmitter is warming up, each component is checked. The countdown timer appears on the screen for the warm-up.
7.5 Reading Display
As the transmitter enters the Reading Display, it automatically starts testing for errors and goes through a cycle of checking each alarm
Page 23
RAEGuard 2 PID User’s Guide
18
condition. If there are no errors or alarm conditions, the green “OK” LED is lit and the gas concentration is displayed.
If there is an error, the “Fault” LED blinks and an error message blinks. Each alarm condition has a corresponding LED that blinks an amber color when the readings are outside a specified range or limit.
7.6 Instrument Status Display
After warm-up, the detector enters the Status Display. It automatically starts testing for errors and goes through a cycle of checking each alarm condition. If there is no error message or alarm condition, gas concentration is displayed and LED flashes green.
If there is an error or alarm, a corresponding message shows on the screen, and the LED indicates different statuses:
Status
External
Alarm
LED
LCD
Output
Current
Normal
-
Green
Reading
Reading
Low alarm
LOW
Red flashing, 2 Hz
Reading
Reading
High alarm
HIGH
Red flashing, 3 Hz
Reading
Reading
Over
Range
HIGH Red flashing, 3 Hz
22mA
Pump Error ERR Yellow flash ing, 1 Hz
2mA
Lamp Error ERR Yellow flashing, 1 Hz
2mA
Max Alarm -- Red flashing, 3 Hz
--
Sensor Drift
-- Yellow flash ing, 1 Hz
--
Calibration
Fault
ERR Yellow flash ing, 1 Hz
2mA
Humidity
Sensor
Fault
ERR Green, 1 Hz
Reading
Page 24
RAEGuard 2 PID User’s Guide
19
8 Navigating Settings
You can view the basic settings, clear the Peak and Minimum readings, and perform a test of the Analog system without entering Programming Mode.
At the main reading screen, hold the magnetic key over [-] for two seconds. The RAEGuard 2 screen changes to show the current peak value in memory:
If you do not want to clear the peak value, press [-].
To clear the peak value (reset it), press [+].
This screen appears:
If you do not want to clear the peak value, press [-]. The screen
advances to Min. Clear.
To confirm that you want to clear the peak value (reset it), press
[+]. The value is cleared, and this screen is shown:
Advance to the next screen, Min value, by pressing [-].
Page 25
RAEGuard 2 PID User’s Guide
20
If you do not want to clear the minimum value, press [-].
To clear the minimum value (reset it), press [+].
If you do not want to clear the minimum value, press [-]. The
screen advances to Min. Clear.
To confirm that you want to clear the minimum value (reset it),
press [+]. The value is cleared, and this screen is shown:
Press [-] to advance to the screen showing the measurement gas, calibration gas, correction factor (CF), and molecular weight (MW) of the gas:
Page 26
RAEGuard 2 PID User’s Guide
21
Press [-] to advance to the screen showing the measurement range (M.Range) and high alarm (H.Alarm) and low alarm (L.Alarm) values:
Press [-] to advance to the screen showing the firmware version and build time and date, and the serial number of the instrument:
Press [-] to advance to the Analog Output Test screen.
The analog output test is primarily used when commissioning a system. It ensures that the mA output at the sensor location is the same as the input received on the system controller. Run the analog test to make sure that the readings on the detector match the readings on the controller.
The analog test uses a timed step function to output specific mA values in sequence. The test starts at 4 mA, wh ich typically corresponds to 0 ppm output. The analog test then progresses in increments of 2 mA (4, 6, 8, etc., all the way up to 20mA). The corresponding sensor reading in ppm is shown on the unit’s LCD display.
Page 27
RAEGuard 2 PID User’s Guide
22
For example, if you have a sensor range of 0 to 1000 ppm, the detector LCD displays in sequence 0, 125, 250, 375…all the way up to 1000 ppm. These are the ppm equivalent values of 4,6, 8, 10, etc., all the way up to 20mA.
Note: You can stop the test and exit at any time by pressing [-].
Page 28
RAEGuard 2 PID User’s Guide
23
9 Programming Menus
Programming allows anyone with the password to change the instrument’s settings, calibrate the instrument, etc. Depending on whether you use the Basic or Advanced Mode, both of which have different passwords, you can access different menus and submenus.
This table shows the menus (and their icons) and submenus. Items marked with an asterisk (*) are only available in Advanced Mode. They are not available in Basic Mode.
Calibration
Measurement*
Alarm Setting
Monitor Setup
Zero/Fresh Calibration
Meas. Unit*
High Alarm
Temp. Unit Span Calibration
Meas. Range*
Low Alarm
Language
Set Span Value
Meas. Gas*
Ext. Alarm Delay*
Pump Duty (%)* Calibration Gas*
Pump Cycle(s)*
Pump Status*
Bus Baudrate
Analogout 4mA
Analog out 20m A
Unit ID
LCD Contrast
LCD Backlight
Change Password*
Page 29
RAEGuard 2 PID User’s Guide
24
9.1 Entering Programming Mode
Enter Programming Mode by using the magnetic key to press these buttons in sequence:
1. Press [+].
2. Press [-].
3. Press [MODE].
The password screen appears, with the cursor on the first digit:
Input the 4-digit password:
Increase the number from 0 through 9 by pressing [+].
Step from digit to dig i t using [-].
Press [MODE] when you are done.
If you make a mistake, you can cycle through the digits by pressing [-] and then using [+] to change the number in each position.
Note: The default password for Basic Mode is 1111. The default password for Advanced Mode is 1250.
Once you enter Programming Mode, the LCD displays the first menu, Calibration. Each subsequent menu is accessed by pressing [-] repeatedly until the desired menu is displayed. To enter a submenu of a menu, press [+].
The Calibration label is shown and its icon is highlighted, but you can press [-] to step from one programming menu to the next, with the name of the menu shown at the top of the display and the corresponding icon highlighted.
Page 30
RAEGuard 2 PID User’s Guide
25
As you repeatedly press [-], the selection moves from left to right, and you see these screens in sequence:
9.2 Calibration
Two types of calibration are availabl e: Zero (f resh ai r) and Span.
Note: Complete calibration procedures are detailed on page 39.
Press [+] to select Calibration. Otherwise, press [MODE] to go back, or press [-] to advance to the next menu.
The Calibration screen is shown:
Select Zero or Span Calibration by pressing [-]. Once your choice is highlighted, press [+].
Page 31
RAEGuard 2 PID User’s Guide
26
Zero/Fresh Calibration
Your options are indicated along the bottom of the screen:
Press [+] to start.
Press [MODE] to quit and return to the menu.
If you choose to start zero/fresh air calibration, you see this message:
Please apply zero gas…
Connect the zero gas, start its flow, and then press [+]. There is a countdown as calibration is performed.
Span Calibration
Your options are indicated along the bottom of the screen:
Press [+] to start.
Press [MODE] to quit and return to the menu.
If you choose to start span calibration, you see this message:
Please apply gas…
Connect the calibration gas, start its flow, and then press [+]. There is a countdown as calibration is performed.
Calibration Gas (Advanced Mode)
You can set the gas that you want as your calibration gas. Calibration gases are organized in two lists:
Last Ten is a list of the last ten gases used by your instrument.
The list is built automatically and is only updated if the gas selected from the Gas Library is not already in the Last Ten. This ensures that there is no repetition.
Page 32
RAEGuard 2 PID User’s Guide
27
Gas Library is a library that consists of all the gases found in
RAE Systems’ Technical Note TN-106 (available online at www.raesystems.com).
Press [-] to step between the two options.
Press [MODE] to go back to the menu level above it.
Press [+] to select your choice and advance to its submenu.
Then follow the steps for the type of selection group you have chosen:
Last Ten
1. Step through the list’s gases by pressing [-].
2. Register your selection by pressing [+].
3. Save your choice by pressing [MODE].
Gas Library
1. Step through the list’s gases by pressing [-].
2. Register your selection by pressing [+].
3. Save your choice by pressing [MODE].
Note:
Whenever you change the calibration gas type, you must set the
span value.
9.3 Measurement (Advanced Mode )
Measurement settings allow you to set measurement units, range, gas, and other parameters.
Meas. Unit (Advanced Mode)
Standard available measurement units include:
Abbreviation
Unit
ppm
parts per million
mg/m3
milligrams per cubic meter
Page 33
RAEGuard 2 PID User’s Guide
28
1. Press [-] to move the cursor from “ppm” to “mg/m3” or cycle
through them.
2. Press [+] to register your choice.
3. Press [MODE] to save the selection.
4. Press [+] to save the Measurement Unit, [MODE] to return to
editing it, and [-] to cancel the change.
The new Measurement Unit is saved, and the RAEGuard 2 PID exits back to the “Monitor Setup” menu.
Meas. Range (Advanced Mode)
Each sensor head has a maximum range for its sensor (for instance, 100 ppm). However, if you want to increase sensitivity in a lower range, it is possible to change the measurement range (from the preset 100 ppm to 60 ppm, for example).
1. Press [+] to increase the number from 0 to 9 (it wraps around
to 0 again after 9).
2. Press [-] to step through digits.
3. Press [MODE] to save the new value.
If you have set a value that is within range, then you see this screen:
Press [+] to save the new value.
Press [MODE] to go back and edit it again.
Press [-] to exit without making a change.
If the value is within the sensor’s range when you press [+], the value is saved and “Saved!” is displayed.
Page 34
RAEGuard 2 PID User’s Guide
29
If you have set a value that is outside the sensor’s range, you see this, along with a message that says, “Invalid Input!”:
You have two options:
Press [-] to exit without making a change.
Press [+] to go back to change the value.
Meas. Gas (Advanced Mode)
PIDs can be calibrated to a reference gas, yet display the concentration readings in equivalent units of the target gas. Isobutylene is typically used as the calibration gas because it is inexpensive, readily available, has immediate sensitivity, and has very low toxicity. The RAEGuard PID can display concentrations of over 200 gases using the built-in library of correction factors. Refer to RAE Systems Technical Note TN-106, available online at www.raesystems.com, for further details.
Measurement gases are organized in two lists:
Last Ten is a list of the last ten gases used by your instrument.
The list is built automatically and is only updated if the gas selected from the Gas Library is not already in the Last Ten. This ensures that there is no repetition.
Gas Library is a library that consists of all the gases found in
RAE Systems’ Technical Note TN-106 (available online at www.raesystems.com).
Press [-] to step between the two options.
Press [MODE] to go back to the menu level above it.
Press [+] to select your choice and advance to its submenu.
Page 35
RAEGuard 2 PID User’s Guide
30
Then follow the steps for the type of selection group you have chosen:
Last Ten
1. Step through the list’s gases by pressing [-].
2. Register your selection by pressing [+].
3. Save your choice by pressing [MODE].
Gas Library
1. Step through the list’s gases by pressing [-].
2. Register your selection by pressing [+].
3. Save your choice by pressing [MODE].
Page 36
RAEGuard 2 PID User’s Guide
31
9.4 Alarm Setting
During each measurement period, the gas concentration is compared with the programmed high and low gas concentration alarm limits. If the concentration exceeds any of the preset limits, the alarms (and relays) are activated immediately to warn of the alarm condition.
An alarm signal summary is shown in the Instrument Status Display table on page 44.
In this menu, you can change the High and Low alarm limits. Press [Y/+] to enter the Alarm Setting menu.
Note: All settings are shown in ppm (parts per million).
1. Select High or Low Alarm by pressing [-].
2. Press [+] to select edit that alarm’s setting. The display shows
a flashing cursor on the left-most digit of the previously stored alarm setting.
3. Press [+] to increase each digit’s value.
4. Press [-] to advance to the next digit.
5. Again, use [+] to increase the number.
Repeat this process until all numbers are entered. Press [MODE] when you are done.
Note: If you set an invalid input value, then it says “Invalid input!” and this set of buttons is shown along the bottom of the display:
Press [MODE] to go back and input a valid value.
Press [-] to exit back to the main menu without making a change.
Page 37
RAEGuard 2 PID User’s Guide
32
Ext. Alarm Delay (Advanced Mode)
In order to prevent false alarms, you can set a delay time between when an alarm event occurs and when the RAEGuard 2 sends out the alarm signal, up to 60 seconds. If the alarm duration is less than the alarm delay setting, then no relays will be triggered.
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until all digits of the new value are entered.
3. Press [MODE] to register the Ext Alarm Delay value.
4. Press [+] to save the value, [MODE] to go back and edit the
value, or [-] to cancel and return to the menu.
Note: If you set an invalid input value, then it says “Invalid input!” and this set of buttons is shown along the bottom of the display:
Press [MODE] to go back and input a valid value.
Press [-] to exit back to the main menu without making a change.
Page 38
RAEGuard 2 PID User’s Guide
33
9.5 Monitor Setup
This menu has sub-menus for setting the date, time, and other parameters.
Temp. unit
Set the temperature units to either Fahrenheit or Celsius.
1. Press [-] to place the cursor on the temperature unit you want.
2. Press [+] to make the selection.
3. Press [+] to save your selection, exit the Temp. unit submenu
and return to the Monitor Setup menu.
Note: If you do not want to save the change and want to make further changes, press [MODE]. Press [-] to exit without saving the changes.
Language
Choose from English or Chinese.
1. Press [-] to place the cursor on the language you want.
2. Press [+] to make the selection.
3. Press [MODE] to save your selection, exit the Language
submenu and return to the Monitor Setup menu.
Pump Duty (%) (Advanced Mode)
The pump’s duty cycle is the ratio of its on time to off time. The duty cycle ranges from 50% to 100% (always on), and the period is 10 seconds. Therefore, a duty cycle of 60% means that the pump is on for 6 seconds and off for four seconds. RAE Systems’ patented duty cycling is employed by the instrument to clean the PID. A lower duty cycle has a greater effect on keeping the PID clean than a higher duty cycle.
Important! Pump duty cycling is interrupted when the instrument senses a gas. The pump’s duty cycle is disabled when the
Page 39
RAEGuard 2 PID User’s Guide
34
measurement is greater than the 2ppm threshold and is re-enabled when the reading falls below 90% of the threshold (1.8 ppm).
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until all digits of the new value are entered.
3. Press [MODE] to register the Pump Duty value.
4. Press [+] to save the value, [MODE] to go back and edit the
value, or [-] to cancel and return to the menu.
Note: If you set an invalid input value, then it says “Invalid input!” and this set of buttons is shown along the bottom of the display:
Press [MODE] to go back and input a valid value.
Press [-] to exit back to the main menu without making a change.
Page 40
RAEGuard 2 PID User’s Guide
35
Pump Cycle(s) (Advanced Mode)
This sets the length of time for one cycle of the pump running.
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until all digits of the new value are entered.
3. Press [MODE] to register the Pump Cycle(s) value.
4. Press [+] to save the value, [MODE] to go back and edit the
value, or [-] to cancel and return to the menu.
Note: If you set an invalid input value, then it says “Invalid input!” and this set of buttons is shown along the bottom of the display:
Press [MODE] to go back and input a valid value.
Press [-] to exit back to the main menu without making a change.
Pump Status (Advanced Mode)
This tells you whether the pump should be on or off, and allows you to set whether the pump is on or off.
Press [-] to select On or Off (the black dot in the circle indicates the selection).
Press [+] to make the change to your selection.
Press [MODE] if you want to exit without making a change.
Page 41
RAEGuard 2 PID User’s Guide
36
Bus Baudrate
Choose from three different baud rates for Modbus data transmission: 4800, 9600, or 19200.
1. Press [-] to place the cursor on the baud rate you want.
2. Press [+] to make the selection.
3. Press [MODE] to save your selection.
4. Press [+] to store the selection and exit the submenu and
return to the Monitor Setup menu.
Note: If you do not want to save the change and want to make further changes, press [MODE]. Press [-] to exit without saving the changes.
Analogout 4mA
This feature sets the analog output to 4mA for system setup and troubleshooting.
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until all three digits of the new value are entered.
3. Press [MODE] to save the 4mA analog output value.
Analogout 20mA
This feature sets the analog output to 20mA for system setup and troubleshooting.
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until all four digits of the new value are entered.
3. Press [MODE] to save the 20mA analog output value.
Page 42
RAEGuard 2 PID User’s Guide
37
Unit ID
When using Modbus communication, you can give each RAEGuard 2 PID on a network a unique Unit ID that helps identify it.
1. Press [+] repeatedly to select a desired number. Numbers
increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to 0.
2. Press [-] to advance to the next digit.
Repeat this process until both digits of the new value are entered.
3. Press [MODE] to save the 20mA analog output value.
LCD Contrast
The display’s contrast can be increased or decreased from its default setting. You may never need to change the default setting, but sometimes you can optimize the display to suit extreme temperature and ambient brightness/darkness conditions.
1. Press [+] to step through the contrast levels. The range is 1 to
99. Once 99 is reached, pressing [+] again "wraps" around back to 10.
2. When you have completed your selection, press [MODE].
3. Press [+] to save your selection, and exit LCD Contrast.
Note: If you want to make further changes, press [MODE] to return to the adjustment screen. Pr ess [-] to exit without saving the changes.
LCD Backlight
The backlight glows whenever you press a button with the magnetic key. In addition, you can set the backlight to be either always on or only on when an event (alarm) occurs.
1. Press [-] to place the cursor on the setting you want.
2. Press [+] to make the selection.
3. Press [MODE] to save your selection, exi t the LCD Ba cklight
submenu and return to the Monitor Setup menu.
Page 43
RAEGuard 2 PID User’s Guide
38
Change Password (Advanced Mode)
You can change the four-dig it passwo rd for acc ess ing Basic Mode.
The default basic password is: 1111 The advanced password is: 1250
Note: You cannot change the password for accessing Advanced Mode.
1. Press [+] to step through all 10 numerals (0 to 9).
2. Press [-] to advance to the next digit. The next digit to the
right flashes. Repeat this process until all four digits of the new password
are entered.
3. Press [MODE] to register the selection.
4. Press [+] to save the new password, [MODE] to return to
editing the password, and [-] to cancel the change.
The new password is saved, and the RAEGuard 2 PID exits back to the “Monitor Setup” menu.
Page 44
RAEGuard 2 PID User’s Guide
39
10 Calibration
The RAEGuard 2 PID is calibrated using a two-point calibration process. First, use a “zero gas.” Then use a “span gas” containing a known concentration of a standard reference gas, to set the second point of reference.
Note:
Zero calibration must be performed before span calibration.
RAEGuard2 PID connected to gas bottle with calibration adapter.
To perform a calibration, you need a cylinder of zero gas, a cylinder of span gas, and a calibration adapter.
WARNING
The cali brat ion of al l newl y purcha sed RA E Syste ms ins tru ments should b e teste d by exp osing the sens or(s ) to a kn own concen tra tion cal ibr at io n gas be for e t he i ns tru men t is us ed or put into service. F or maximu m safety, the accuracy of the RAEGuard 2 PID should be checked by ex posing t he sensor to a known c onc entr a tion cal ibra ti on g as.
Page 45
RAEGuard 2 PID User’s Guide
40
10.1 Zero Calibration
1. Connect the zero gas cylinder to the gas inlet port on the
RAEGuard2 PID using the provided calibration adapter.
2. In measurement mode, use the magnet key to press [+],
[-], and then [MODE] in sequence.
3. Input your password, followed by [-].
The instrument enters the calibration menu.
4. Press [+] to enter the Calibration menu.
5. Press [-] until “Zero/Fresh calibration” is highlighted.
Press [+], and this screen appears:
6. Start the gas flow.
Page 46
RAEGuard 2 PID User’s Guide
41
7. Press [+]. A countdown screen is shown as the calibration is
performed:
Note: You may abort the calibration during the countdown by
pressing [-]. If you stop the calibration, the process stops and this screen is displayed befo re the men u screen is s how n:
8. Once the calibration is complete, this screen is shown:
9. If the calibration is not successful, and you see the “Zero
Calibration Failed” message, check that the inlet is not blocked and that there are no other problems before retrying.
Page 47
RAEGuard 2 PID User’s Guide
42
10.2 Span Calibration
1. Connect the calibration gas cylinder to the gas inlet port on
the RAEGuard2 PID using the provided calibration adapter.
Note: Make sure the labeled concentration of calibration gas matches the value that is set in the RAEGuard 2 PID.
2. If the RAEGuard 2 PID is not already in Programming Mode,
use the magnet key to press [+], [-], and then [MODE] in sequence.
3. Input your password, followed by [-].
The instrument enters the calibration menu.
4. Press [+] to enter the Calibration menu.
5. Press [-] until “Span calibration” is highlighted.
Page 48
RAEGuard 2 PID User’s Guide
43
6. Press [+], and this screen appears:
7. Start the gas flow.
8. Press [+]. A countdown screen is shown as the calibration is
performed:
Note: You may abort the calibration during the countdown by
pressing [-]. If you stop the calibration, the process stops and this screen is displayed before the menu screen is shown:
9. Once the calibration is complete, this screen is shown:
10. If the calibration is not successful, and you see the “Span
Low Failed!” message, check that the inlet is not blocked, the sensor or lamp are not fouled or damaged, and there are no other problems before retrying.
Page 49
RAEGuard 2 PID User’s Guide
44
11 A larm Signal Summary
The following are reading-related alarms.
Status
LCD
LED Indication
Relay Status
Analog Output
Normal
Green
Failure: Disable
Low alarm: Disable
High alarm: Disable
According
to Readi ng
Value
Low Ala r m
Red
(2 flashes
per second)
Failure: Disable
Low alarm: Enable
High alarm: Enable
According
to Readi ng
Value
High Alarm
Red
(3 flashes
per second)
Failure: Disable
Low alarm: Disable
High alarm: Enable
According
to Readi ng
Value
Over Range
Red
(3 flashes
per second)
Failure: Disable
Low alarm: Disable
High alarm: Enable
22mA
Pump Error
Yellow
(1 flash
per second)
Failure: Enable
Low alarm: Disable
High alarm: Disable
2mA
Lamp Error
Yellow
(1 flash
per second)
Failure: Enable
Low alarm: Disable
High alarm: Disable
2mA
Max Alarm
Red
(3 flashes
per second)
Failure: Disable
Low alarm: Disable
High alarm: Enable
2mA
Sensor
Negative
Drift
Yellow
(1 flash
per second)
Failure: Enable
Low alarm: Disable
High alarm: Disable
2mA
Calibration
Failure Error
Yellow
(1 flash
per second)
Failure: Enable
Low alarm: Disable
High alarm: Disable
2mA
Humidity
Sensor Error
Green
Failure: Disable
Low alarm: Disable
High alarm: Disable
According
to Readi ng
Value
Page 50
RAEGuard 2 PID User’s Guide
45
12 Maintenance
As a guide, it is recommended to regularly “bump test” a RAEGuard 2 PID unit with a known percentage gas. Also, periodically check the sensor inlet and outlet to make sure they are clean and unobstructed.
13 Troubleshooting
Note: Before attempting to diagnose measurement problems, perform zero and span calibrations.
Message
Reason and Solution
Reason: Instrument over range.
Solution: Confirm the level of gas concentration.
Check sensor.
Reason: Pump failure.
Solution: Make sure that the air path is not blocked.
Check pump.
Reason: Lamp failure. Solution: Check Sensor.
Reason: Sensor drift.
Solution: Check sensor and/or perform zero and
span calibration.
Reason: Calibration fails.
Solution: Check calibration operation.
Change sensor or contact RAE Systems
Customer Service.
Reason: Humidity sensor failu re.
Solution: Check the humidity in the area. Make sure that the air path is not blocked.
Contact RAE Systems Customer Service.
Page 51
RAEGuard 2 PID User’s Guide
46
14 Appendix A: Range, Sensor and
Related Configuration
The following table shows the measurement range and related configurations for four different DigiPID sensor modules:
Sensor
(ppm)
Resolution
(ppm)
Span Gas
Setup
Default
Value
Default Alarm Setup (ppm)
Low Alarm High Alarm
0-100
0.01
10
10
20
0-1000
1
100
10
100
15 ModBus/RS-485 Information
The RAEGuard s communicates by means of MODBUS RTU. All monitors provide a 4-byte register value. Note: Gas concentration is the only value that can be retrieved.
For example, 34 hex = 52 decimal
Overview
This document describes the modbus protocol used in RAEGuard 2 PID products.
1. Communication Setting
Communication Mode: RTU Controller: PC or Controller Client: RAEGuard 2 PID Baud Rate: 4800, 9600, 19200 Client ID: 1 to 99 (0x0001 to 0x0064)
Page 52
RAEGuard 2 PID User’s Guide
47
2. Data Structure/C ommunication Steps
RAEGuard 2 PID only supports Function Code 0x03 (read-only registers). The detector only supports value reading.
0x03: Read-Only Register Requesting Message:
Device
Address
Func-
tion Code
Register
Address High
Byte
Register
Address Low
Byte
Quantity
of Registers
High Byte
Quantity
of Registers
Low Byte
CRC
Low Byte
CRC
High Byte
Client ID
03
00
08
00
02
CRC
CRC
Answering Message:
Device
Address
Function
Code
Byte
Count
Register Value
CRC
Low
Byte
CRC
High
Byte
Client ID
03
04
Reading
Highest
Byte
Reading
Higher
Byte
Reading
High
Byte
Reading
Low Byte
CRC
CRC
Note: The length of data from the detector is 4 bytes.
Example:
Request: 01 03 00 08 00 02 45 C9 Answer: 01 03 04 00 00 00 D1 3A 6F
Note: The maximum communication distance should be less than 1 km when using a 1.5 mm
2
cable.
Page 53
RAEGuard 2 PID User’s Guide
48
16 Appendix B: Co ntrolled Section
RAEGuard 2 And DigiPID
16.1 Scope
The scope of this document is to describe the system aspects for use of the RAEGuard 2 PID by connection of the RAEGuard2 and DigiPID according to EN60079-25; UL913 and CAN/CSA-C22.2 No. 157-92 in hazardous locations.
16.2 Contents
RAEGuard2
certification
DIGI PID
certification
Comment
to
integration
ATEX/IECEx certification for zones
ITS 11ATEX17231X
KEMA10ATEX0059
-
IECEx ITS
11.0005X
IECEx KEM
10.0005
-
Ex d [ ia Ga ]IIC T4 Gb
Ex ia IIC T4 Ga
­Gb
Ga
Gb < Ga
UL/CSA certification for Divisions
RAEGuard 2
Associated [ia] output
Intrinsically safe, ia, PID
DIGI PID
Page 54
RAEGuard 2 PID User’s Guide
49
CID1
CID1
same
-40º C < Tamb < +55º C
­Um: 28VDC
-
-
Uo: 5.91V
Ui: 6.13V
Uo < Ui; OK
Io: 0.8A
Ii: 1.5A
Io < Ii; OK
Po: 1.16W
Pi: 1.225W
Po < Pi; OK
Co: 30µF
Ci: 20.2µF
Ci < Co; OK
Lo: 55µH
Li: 1µH
Li < Lo; OK
As demonstrated in the table above, RAEGuard 2 and DigiPID have matching entity parameters, which allows the two devices to be connected to form a single piece of equipment: RAEGuard 2 PID, with the resulting Ex specification given below, suitable for installation in hazardous areas defined by zones as Ex d ia IIC T4 Gb.
The marking is found on each of the two connected devices.
Resulting RAEGuard2 PID Ex specification
Ex d ia IIC T4 Gb
-20º C < Tamb < +55º C
Page 55
RAEGuard 2 PID User’s Guide
50
RAEGuard 2
16.3 Scope
The scope of this document is to identify the section of the RAEGuard2 controlled part of manual.
16.4 Responsibility
The included sections cannot be changed without prior approval from the Notified Body.
16.5 Contents
Below are the sections controlled by the Notified Body, including all safety-related information in the manual.
Controlled sections are:
1.Warnings and directive information
2. Marking of RAEGuard2
3. Hazardous Location Classifications
4. Instructions for Safe Use
5. Connections and Ratings
6. Maintenance
7. Physical dimensions
Page 56
RAEGuard 2 PID User’s Guide
51
1. Warnings And Directive Information
- READ BEFORE OPERATING -
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions.
CAUTION!
To reduce the risk of electric shock, turn the power off before removing the instrument cover. Disconnect the power before removing the sensor module for service. Never operate the instrument when the cover is removed. Remove instrument cover and sensor module only in an area known to be non- hazardous.
Page 57
RAEGuard 2 PID User’s Guide
52
2. Marking Of RAEGuard 2
The RAEGuard 2 is certified according to ATEX and the IECEx schemes as protected by a flameproof enclosure provided one or two associated equipment output port protected as intrinsically safe [ia], one of which can include a set of flame arrestors for gas sampling. The product is marked with the following information:
RAE Systems Inc. 3775 N. 1st. St. San Jose, CA95134 USA Serial number: XXXXXXXXXX RAEGuard 2, FGM-2XXX
IECEx
ATEX
IECEx ITS 11.0005X
ITS 11ATEX17231X
Ex d [ia Ga] IIC T4, Gb
0575 II 2(1)G, Ex d [ia Ga]
IIC T4 Gb
-20ºC < Tamb < +55ºC Um: 28VDC; Uo: 5.91V; Co: 30µF; Lo: 55uH; Io: 0.8A; Po:
1.16W
Year of manufacture: XXXX
WARNING: DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT. REFER TO USER MANUAL FOR ENTRY TYPE AND SIZE
RATING: < 5W. 10 ~28VDC
4. Hazardous Location Classification
Hazardous Areas Classified By Zones
RAEGuard 2 are intended to be used in hazardous areas classified for zone 1 or zone 2, within the temperature range of -20
o
C to +55oC, Where gases of explosion groups IIA, IIB or IIC and T4 may be present.
Page 58
RAEGuard 2 PID User’s Guide
53
5. Instructions For Safe Use
It must be ensured that the supply voltage for RAEGuard 2 comply with the Um input entity parameter. The 8-pin output interface connector facilitates connection to intrinsically safe equipment, e.g. sensors or modems. The connected devices must be protected by intrinsically safety, ia, ib or ic for gases in group IIA, IIB or IIC.
The connected device must have entity parameters that comply with general fitting rules for entity parameters with the output entity parameters of the RAEGuard 2 interface. The DigiPID sensor can be mounted on the on the RAEGuard 2 interface inside hazardous are in powered condition. This constellation is called RAEGuard 2 PID.
6. Connections And Ratings Outline drawing of the interface:
Figure1
Output interface port definition:
The pin definition of the 8-pin output connector is given in the table below:
Pin No.
1 2 3 4 5
6 7 8
Definition
Power
Out
CS
Analog
Single
RXD
Open
Drain
GND Cn TXD
7
8
1
2
In Out
Page 59
RAEGuard 2 PID User’s Guide
54
Input/output entity parameters
The rated RAEGuard2 input/output parameters are as follows: Input: Um: 28VDC Output: Uo: 5.91V Co: 30µF Lo: 55uH Io: 0.8A
Po: 1.16W
Cable Entry type and Size
RAEGuard2 is provided with three female 3/4" - 14 NPT cable entry holes tapped into its side walls. One or two of these three cable entries are mounted with safety adapter interface type ISA-100/ ISA-
101.
7. Maintenance
Calibration
The monitor should be calibrated for no less than every 6 months depending on use and exposure to gas and contamination.
- Calibration interval may very due to national legislation
- RAE Systems recommend the use of RAE Systems
calibration gas
Installation and Access Instructions
When used in North America, conduit runs must have a
sealing fitting connected within 18" (457mm) of the enclosure. CSA requires seals in conduit exceeding 5'
(1.524 m) in Group C atmospheres.
The appropriate regulations for installation, service and repair
must be properly observed during such activities. Installation in ATEX/IECEx area must be done as outlined in EN/IEC 60079-14.
To prevent ignition of hazardous atmospheres, area must be
free of flammable vapors and supply circuit must be disconnected before removing cover.
Page 60
RAEGuard 2 PID User’s Guide
55
8. EARTH GROUNDING INSTRUCTIONS
External Earth Grounding
Fasten the crimped ground wire with hardware as illustrated below. The wire should have a minimum cross-section area of 4mm
2
for its
conductor.
Internal Earth Grounding
Use the same hardware as shown in the illustration of external earth grounding. The wire should be no less than the size of the power lines.
Finished Grounding Wires
Internal and external grounding are shown here, as well as an alternate external grounding point. Always follow local electrical guidelines.
Sensors are not interchangeable; use only RAE System s
sensors, and use only the sensor type specified for your RAEGuard2 monitor. Use of non-RAE Systems components will void the warranty and can compromise the safe performance of this product.
Page 61
RAEGuard 2 PID User’s Guide
56
9. Physical Dimensions
RAEGuard 2 can be easily installed and integrated with various control systems with its flexible pipe-holding/wall-mounting options and standard connection terminals. The physical dimensions are as follows:
130mm
187mm
112mm 118mm
Page 62
RAEGuard 2 PID User’s Guide
57
SECTION 2: DigiPID User’s Guide
Page 63
RAEGuard 2 PID User’s Guide
58
SAFETY INSTRUCTIONS
17 Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions. The user should understand how to set the correct parameters and interpret the obtained results.
CAUTION!
Sensor interface connector pins cannot be interchanged for multiple applications. Apply the sensor in accordance with connector definitions described in the datasheet and in accordance with the general fitting rules for the applied intri nsi ca lly safe entity parameters. Use only RAE Systems lamps and the sensor detector type specified for your module. Maintain by using only RAE Systems accessories, including a Teflon filter or UV shield. Use of non-RAE Systems components will void the warranty and can compromise the safe performance of this product.
Page 64
RAEGuard 2 PID User’s Guide
59
17.1 Dig iPID Markin g
The DigiPID is certified according to the IECEx scheme, ATEX and CSA for US and Canada as protected by intrinsic safety.
The product is marked with the following information: RAE SYSTEMS 3775 N. 1
st
. St., San Jose CA 95134, USA DIGI PID Type DS100/DS101/DS102 Serial No/barcode: XX X-XXXX-000
IECEx KEM
10.0005 Ex ia IIC T4 Ga Ex ia I
0575
IM1/II 1G Ex ia IIC T4 Ex ia I KEMA10ATEX0059
Class I, Gr. A,B,C,D T4 Only as to intrinsic safety for use in hazardous loc. Intrinsically safe/ Securite
Intrinseque/Exia
Pi: 1.225W; Vi: 6.13V; Ci: 20.2µF; Ii: 1.5A; Li: 1µH.
-40º C < Tamb < +55º C
Warning: Read User’s Manual for Intrinsic Safety Precautions
Page 65
RAEGuard 2 PID User’s Guide
60
17.2 Operation Area and Conditions
17.2.1 Hazardous Areas classified by Zones
DIGI PID is intended to be used in hazardous areas or mines susceptible to fredamp classified zone 0, zone 1 or zone 2, within the temperature range of -40º C to +55º C, where gases of explosion groups IIA, IIB or IIC and T4 may be present.
Hazardous Areas classified by Divisions
DIGI PID is intended to be used in hazardous areas classified for Class I Div. 1 or 2, within the temperature range of -40º C to +55º C, where gases of explosion groups A, B, C or D and temperature class T4 may be present.
17.3 Instruction For Safe U se
Strictly follow the instructions for safe use. Application of the DigiPID requires full understanding and strict observation of the instructions.
The DigiPID can be connected to Ex-certified equipment for use in hazardous areas, when it is secured tha t the connec tio n is com ply ing with the intrinsically safe input entity parameters for the two units, and the intended function area corresponds to the certified area.
The DigiPID sensor can be connected and disconnected inside the hazardous area.
Page 66
RAEGuard 2 PID User’s Guide
61
17.4 Use In Hazardous Areas
Equipment which is intended for use in explosive atmospheres and which has been assessed and certified according to international regulations may be used only under s pecif ied conditions. The components may not be modified in any way.
The appropriate regulations for service and repair must be properly observed during such activities.
17.5 Year of manufacture
To identify the year of manufacture, refer to the serial number of the instrument.
The second to last digit in the serial number indicates the year of manufacture. For example, “H” indicates the manufacturing year is
2008.
First digit
Year
J
2008
K
2009
M
2010
N
2011
P
2012
Q
2013
R
2014
S
2015
T
2016
U
2017
V
2018
W
2019
Page 67
RAEGuard 2 PID User’s Guide
62
17.6 Specifications
Power Supply
5V ±0.25V DC
Current
110 mA max
Power Consumption
< 0.6W
Measuring Range
0 to 100 ppm, 1 to 1000 ppm
Resolution
10 ppb, 100 ppb, 1 ppm (depends on model)
Response Time
Diffusion (T90): 30 s Pumped (T90): 5 s
Calibration
Two-point off-line and field calibrat ion
Accuracy
±2% for calibration point
Zero Drift
±10% FSS/Month
Span Drift
±10% FSS/Month
Analog Output
0.5 - 2.5V (ro=1.0k)
Digital Interface
Serial Interface (UART)
Transmit (Tx): 3.3V TTL
Receive (Rx): 3.3V TTL
Sensor Lifetime
2 years
Operating Temperature
-20º C to +55º C (-4º F to +131º F) Humidity
0 to 95% RH non-condensing
EMI/RFI
Highly resistant to EMI/RFI Compliant with EMC Directive 2004/108/EC
Package
Rugged case for 1.2m drop test
Spray watertight for IP65 rating
Dust membrane for sensor front protection
Size (DxL)
49mm x 150.8mm (1.92″ x 5.94″)
Weight
< 550g (19.4 oz)
Ex entity parameters
Pi: 1.225W; Vi: 6.13V; Ci: 20.2µF; Ii:
1.5A; Li: 1µH
Page 68
RAEGuard 2 PID User’s Guide
63
18 General Information
The DigiPID Sensor is a self-contained intelligent sensor module with built-in photoionization detector (PID) sensor processor, lamp driver, and analog and digital interface circuits. It is designed to detect volatile organic compounds (VOC) and has a standard external interface. This sensor module can be easily integrated into a wired or wireless communication system for remote, wide-area, and pervasive monitoring applications. The sensor module can be powered by a 5V ±0.25V DC power supply. It is housed in a weatherproof enclosure with a standard 8-pin interface connector. The chart below details the functions of the eight pins:
Pin
Number
1 2 3 4 5 6 7
8
Definition
Power
In
CS
Analog
Signal Out
RXD
Open
Drain
GND
Control
Signal (optional)
TXD
Notes:
Power in: Power supply input connector for the sensor module. The input voltage range is DC 5V ±0.25V at 200mA.
CS: Communication Select. Select (falling edge) to start communication, deselect (rising edge) to stop communication.
Analog Signal Out: Analog signal output with 1kΩ output resistance for the sensor module. The output signal range is between 0.5V and
2.5V. RXD & TXD: Serial interface poles (receive and transmit) for UART
mode with 3.3V TTL level.
Open drain: Switching signal output pole. It can deliver low or high level to external equipment according to customer requirements.
GND: Power and signal ground.
Page 69
RAEGuard 2 PID User’s Guide
64
Control signal: This optional pole only delivers a high (3.3V) or low (0V) signal le v el.
Caution: 6.2V is the maximum input voltage for all of the sensor module’s input poles.
19 Grounding (Earth Connection)
DigiPID includes a grounding connection via pin 6 of the 8-pin connector. If the complete installation must be grounded, ensure that the thread on which DigiPID is mounted is correctly grounded. Check local ordinances regarding further requirements.
Page 70
RAEGuard 2 PID User’s Guide
65
20 Physical Description
The DigiPID is a pumped sensor. The pump is in this version intended to pump the sample gas in though one of the DigiPID gas tubes and return it through the other.
Page 71
RAEGuard 2 PID User’s Guide
66
21 Sensor Parts And Dimensions
DigiPID sensor shown with optional mounting adaptor
8-pin
interface connector
Sample outlet
Pump connections
Sensor module
PID Lamp
Sensor detector
Teflon O-ring
Sensor cover with
filter installed
Optional
mounting adapter
Fastening nut
Optional mounting
adapter
Sample inlet
Page 72
RAEGuard 2 PID User’s Guide
67
22 Operating the Sensor Module
22.1 Preparing The Sen sor For Use
Make sure the 8-pole input of the external intrinsically safe equipment matches the interface connector pinout configuration of the DigiPID. Make sure the input power supply range is between
4.75V and 5.25V DC, and the input current is about 200mA when the sensor module is operated. The following mechanical drawing includes the sensor module interface connector’s pinout. The Ex entity parameter of the equipment connected to the DigiPID must fit the entity parameters of DigiPID according to intrinsically safe fitting rules.
8-pin connector pinout configuration:
1: Power in 2: CS (Communication Select). Select (falling edge) to start
communication, deselect (rising edge) to stop communication.
3: Analog Signal Output 4: RXD 5: Open Drain 6: GND 7: Control Signal (Optional) 8: TXD
Note: Some external equipment requires no connection to pins 5 and 7.
Page 73
RAEGuard 2 PID User’s Guide
68
23 Using The Sensor Module
23.1 Gas Flow Routing Connection
The DigiPID has an “IN” and “OUT” port for sampling gas. Make sure that the in and out sample tubes are properly oriented.
After making all gas connections and ensuring connection to the 8-pin connector, turn on the power supply and warm up the sensor module.
Note: A two-hour warm-up time is recommended before making measurements.
Firmly affix the sensor module to external equipment and make sure the 8-pole interface connector makes a reliable, tight connection with other equipment.
When you begin measurement, the sensor provides data to external equipment, showing the current concentration of the specified gas. Data is output digitally via the UART port or as an analog signal (0.5V to 2.5V) from pin 3 of the interface connector. The sensor module can be calibrated for different operations and gases.
If the DigiPID sensor is connected to RAE Systems equipment, you can affix the sensor module directly to the equipment and then turn on the instrument’s power supply. The sensor module runs normally. If the other external equipment is not a RAE Systems product, additional firmw are and software suppo rt is necess ary. RAE Systems can provide essential information such as the handshaking protocol, port setting, flow rating, etc.
Page 74
RAEGuard 2 PID User’s Guide
69
23.2 Sensor Module Calibration
The sensor module is calibrated before it leaves the factory. However, you can also calibrate the sensor module. Allow it to warm up for two hours before calibrating it. Calibration should be performed using external equipment, using a two-point calibration for the 100 and 1000 ppm VOC sensor.
100 ppm and 1000 ppm range sensor module calibration. First perform a zero gas calibration by applying fresh air to the sensor module. If the module is configured for use with a pump, the fresh air’s flow rate must be greater than the pump’s draw rate. A calibration time of 1 minute is recommended. The other calibration point requires exposure of the sensor to span gas. The span gas concentration should be 10 ppm (for 100 ppm range) or 100 ppm (for 1000 ppm range) isobutylene, and calibration time is also 1 minute.
IMPORTANT!
Whenever you change the sensor head while the instrument is in operation, follow this procedure:
Press and hold [+] until the screen says:
Change Sensor
Press [+]. The RAEGuard 2 performs a full startup.
If you leave the sensor head detached from the RAEGuard 2 for more than 60 seconds, it goes into alarm and displays this message:
Sensor Fail:0
Replace the sensor head and follow the procedure above for restarting it.
Page 75
RAEGuard 2 PID User’s Guide
70
23.3 Maintenance A nd Calibration
The sensor module should be calibrated if it does not pass a bump test, but no less than every 6 months, depending on use and exposure to gas and contamination.
23.4 Replacing the Lamp And Filter
The sensor module is shipped with the lamp, sensor, and stainless­filter installed. Periodically check the filter for dirt and contamination, which can impact the reading and response time.
Note: Always turn off power to the DigiPID before checking and servicing any parts. After inspection and/or replacement of parts, recalibra te th e instrument.
The lamp and sensor are not interchangeable with parts from other manufacturers; use only RAE Systems replacements. Use of non­RAE Systems components will void the warranty and can compromise the safe performance of this product.
Loosening the
fastening nut allows the DigiPID to be easily detached from a base.
Page 76
RAEGuard 2 PID User’s Guide
71
Housing
Teflon O-ring
PID Lamp
Sensor
Module
Attachment
ring
IMPORTANT!
Only handle lamp
by holding sides of
end. Do not touch
flat surface.
Filter
IMPORTANT!
Align hole in filter with hole in top of
PID sensor to
ensure
unobstructed
sample flow
Inlet/Outlet
Page 77
RAEGuard 2 PID User’s Guide
72
24 Replacing The S ensor’s Teflon UV Shield
On the underside of the sensor module is a Teflon UV Shield that should be replaced every 90 days, in order to ensure sensor accuracy.
1. Dism antle the DigiPID as shown on page 70, and remove the
Sensor Module. Note: B e careful to not touch the PID lamp’s surface.
2. Remove the Sensor Module, and replace it with a new one (a
spare Sensor Module is included with the unit).
3. Take the old Sensor Module to your shop/service center to
perform the rest of the activities, which will refresh the Sensor Module and make it ready as a replacement spare.
Note: Teflon UV Shields are available from RAE Systems in packages of 10 (P/N: C01-2017-000).
Page 78
RAEGuard 2 PID User’s Guide
73
4. Unscrew and remove all four of the gold-plated contact “legs”
from the Sensor Module.
5. Use a small flat-bladed screwdriver or similar tool to get between
the Sensor Module and the Teflon UV Shield at the point nearest the shield’s large “notch,” and gently pry it up.
6. Work the blade or tool around the bottom of the Sensor Module,
lifting the Teflon UV Shield as you go around.
7. Once the Teflon UV Shield has been separated far enough from
the Sensor Module to remove it, use your fingers to pull it over the four contact “legs” and discard it.
Sensor
Discard the used
Teflon Sheld
Page 79
RAEGuard 2 PID User’s Guide
74
8. Slide a new Teflon UV Shield over the contact “legs,” and gently
press it down until it is snug against the surface of the Sensor Module.
9. Use your fingers to screw on the four contact “legs.”
10. Use a soft, clean cloth to wipe any finger oils from the surface of
the contact “legs,” and set the Sensor Module aside.
11. Inspect the surface of the PID lamp. Clean it with GC-Grade
methanol on a swab, as outlined in the next section.
12. After the surface of the PID lamp dries, reassemble the unit.
Page 80
RAEGuard 2 PID User’s Guide
75
25 Sensor & Lamp Cleaning/Replacement
Clean the PID sensor module, the lamp and housing only if:
1. The reading is inaccurate even after calibration.
2. The reading is very sensitive to air moisture.
3. A liquid has been sucked into the unit and damaged the unit.
Use of the external filter helps to prevent contamination of the sensor. To access the sensor components and lamp, gently unscrew the sensor
cover. Then hold the PID sensor and pull it straight out. A slight, gentle rocking motion helps release the sensor.
25.1 Cleaning The P ID Sensor
Place the entire PID Sensor Module into GC-grade methanol. It is highly recommended that an ultrasound bath to be used to clean the sensor for at least 15 minutes. Then dry the sensor thoroughly. Never touch the electrodes of the sensor by hand.
Also use a methanol-soaked cotton swab to wipe off the lamp housing where it contacts the sensor when the sensor is installed.
Turn over the sensor so that the pins poin t up and the sensor cavity is visible. Examine the sensor electrodes for any corrosion, damage, or bending out of alignment. The metal sensor electrode “fingers” should be flat and straight. If necessary, carefully bend the sensor fingers to ensure that they do not touch the Teflon portions and that they are parallel to each other. Make sure that the nuts on the sensor pins are snug but not overtight. If the sensor is corroded or otherwise damaged, it should be replaced.
Page 81
RAEGuard 2 PID User’s Guide
76
25.2 Cleaning The Lamp Housing O r Changing The Lamp
If the lamp does not turn on, the instrument will display an error message to indicate replacement of the lamp may be required.
1. If the lamp is operational, clean the lamp window surface and the lamp housing by wiping it with GC grade methanol using a cotton swab using moderate pressure. After cleaning, hold the lamp up to the light at an angle to detect any remaining film. Repeat the process until the lamp window is clean. Never use water solutions to clean the lamp. Dry the lamp and the lamp housing thoroughly after cleaning.
CAUT ION: Never touch the windo w sur face with th e finger s
or anything else that may leave a film. Never use acetone or aqueous solutions.
2. If the lamp does not turn on, remove the lamp from the lamp housing. Place the lamp O-ring onto the new lamp. Insert the new lamp, avoiding contact with the flat window surface.
3. Reinstall the PID sensor module.
4. Tighten the sensor cover.
Page 82
RAEGuard 2 PID User’s Guide
77
25.3 Cleaning The Instrument
Occasional cleaning with a soft cloth is recommended. Do not use detergents or chemicals.
25.4 Ordering Replacement Parts
If you need replacement parts, contact your local RAE Systems distribut or . A lis t is available online:
http://www.raesystems.com
In the U.S., you can order sensors, replacement batteries, and other accessories online at:
http://istore.raesystems.com/
25.5 Special Servicing Note
If the instrument needs to be serviced, contact either:
1. The RAE Systems distributor from whom the instrument was purchased; they will return the instrument on your behalf.
or
2. The RAE Systems Technical Service Department. Before returning the instrument for service or repair, obtain a Returned Material Authorization (RMA) number for proper tracking of your equipment. This number needs to be on all documentation and posted on the outside of the box in which the instrument is returned for service or upgrade. Packages without RMA Numbers will be refused at the factory .
Page 83
RAEGuard 2 PID User’s Guide
78
26 Electronic Waste Disposal
At the end of its life, this product must undergo separate collection and recycling from general or household waste.
Page 84
RAEGuard 2 PID User’s Guide
79
27 Technical Support
To contact RAE Systems Technical Support:
Monday through Friday, 7:00AM to 5:00PM Pacific (US) Time
Phone (toll-free): +1 888-723-4800 Phone: +1 408-952-8461 Email: tech@raesystems.com
28 RAE Sy stems Co ntac ts
RAE Systems by Honeywell World Headquarters
3775 N. First St. San Jose, CA 95134-1708 USA
Phone: +1 408.952.8200 Fax: +1 408.952.8480 E-mail: customerserv@raesystems.com Web Site: www.raesystems.com
RAE Systems Technical Support
Monday through Friday, 7:00AM to 5:00PM Pacific Time
Phone: +1.408.952.8461 Email: tech@raesystems.com
Page 85
RAEGuard 2 PID User’s Guide
80
RAE Systems Europe ApS Kirstinehøj 23 A DK-2770 Kastrup Denmark
Phone: +45 86 52 51 55 Fax: +45 86 52 51 77
orders@raeeurope.com sales@raeeurope.com service@raeeurope.com
Web: www.raesystems.eu
RAE Systems UK Ltd
D5 Culham Innovation Centre Culham Science Centre Abingdon, Oxon OX14 3DB United Kingdom
Phone: +44 1865408368 Fax: +44 1235531119 Mobile: +44 7841362693 Email: raeuk@raeeurope.com
RAE Systems France
336, rue de la fée des eaux 69390 Vernaison France
Phone: +33 4 78 46 16 65 Fax: +33 4 78 46 25 98 Email: info-france@raeeurope.com Web: www.raesystems.fr
Page 86
RAEGuard 2 PID User’s Guide
81
RAE BeNeLux BV Rijndal 20 2904 DC Capelle a/d IJssel
Phone: +31 10 4426149 Fax: +31 10 4426148 Email: info@rae.nl Web: www.rae.nl
RAE Systems Spain, s.l.
Av. Remolar, 31 08820 El Prat de Llobregat Spain
Phone: +34 933 788 352 Fax: +34 933 788 353 Mobile: +34 687 491 106 Email: mdelgado@raespain.com Web: www.raespain.com
RAE Systems Middle East
LOB 7, Ground Floor, Office 19, Jebel Ali Free Zone Dubai, United Arab Emirates
Phone: +971.4.887.5562 Email: mesales@raesystems.com
RAE Systems (Hong Kong) Ltd.
Room 8, 6/F, Hong Leong Plaza 33 Lok Yip Road Fanling, N.T, Hong Kong
Phone: +852.2669.0828 Fax: +852.2669.0803 Email:
hksales@raesystems.com
Page 87
RAEGuard 2 PID User’s Guide
82
RAE Systems Japan 403 Plaza Ochanomizu Bldg. 2-1 Surugadai Kanda Chiyoda-Ku Tokyo, Japan
Phone: 81-3-5283-3268 Fax: 81-3-5283-3275 Email: jpsales@raesystems.com
RAE Systems Korea
#1010, DaeMyungAnsVill First, Sang-Dong 412-2, Wonmi-Gu, Bucheon, Kyungki-Do, Korea
Phone: 82-32-328-7123 Fax: 82-32-328-7127 Email: krsales@raesystems.com
Page 88
RAE Systems by Honeywell
World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: 408.952.8200
Toll-Free: 888.723.4800
Fax: 408.952.8480
E-mail (sales support): raesales@raesystems.com
Web Site: www.raesystems.com
Technical Support
Phone: 408.952.8461 (Mon-Fri, 7AM-5PM, PST)
Email: tech@raesystems.com
Product Registration
www.raesystems.com/support/product-registration
RAE Systems Europe
Kirstinehøj 23A, DK-2770 Kastrup • Denmark
Tel: +45.8652.5155 • Fax: +45.8652.5177
Web: www.raesystems.eu • Email: sales@raeeurope.com
RAE Systems (Hong Kong) Ltd.
Units 1516-18, 15/F, Delt a H ouse, 3 On Yiu Street
Shatin, N.T. Hong Kong
Web: www.raesystems.hk • Email: asiasales@raesystems.com
Phone: +852.2669.08 28
RAE Systems Middle East
LOB 7, Ground Floor, Office 19, Jebel Ali Free Zone
Dubai, United Arab Em irates
Email: mesales@raesystems.com • Phone: +971.4.887.5562
P/N D03-4001-000 Rev D April 2012
Loading...