008-4022-000, Revision B, November 2003

with a MultiRAE Plus monitor that is datalog-enabled.

Attention! For European Applications
with the manufacturer’s instructions.

Any rapid up-scale reading followed by a declining or erratic

Adjust Lamp Failure Threshold • 27

. The monitor will beep once and go through a
(customizable using ProRAE Suite software)
(As the MultiRAE Plus checks each sensor, it displays
(for High, Low, STEL, TWA)
(The last calibration date must be within the last 30
after the warm-up is complete, press and hold

(volatile organic compound) – ppm
(lower explosive limit) – percentage

will be 4.8 V or greater. When
- the runtime is how long the
Correction factor may only
Correction factor may only

While all instruments are calibrated prior to leaving the factory,
When should you calibrate the monitor?
will be prompted for a password. Use [Y/+] to

Attach the filter to the MultiRAE Plus during

Apply the RAE Systems 4-gas span gas
If calibration is accidentally started and gas has run out or has
repeatedly. Calibration will stop and revert

A time stamp is created and stored each time a sensor calibration is
twice to exit calibration mode and return to the main display.
Residual gas may linger on the sensors for up to 60 seconds.
to move from sensor to sensor.
Apply span gas mixture and start gas flow.

Always keep the MultiRAE Plus on charge in a dry indoor area when
The clock will also need to be reset.
at any level in the confined space it is

Auto Reset Alarm (default):
The alarm signals automatically reset
when the alarm condition is no longer present.

The backlight can shorten the operating time by 20-30%.

A small “L” at the center of the screen means the monitor is datalogging.
STEL and TWA values will only be logged if average is selected.
All data is stored in non-volatile memory that can be downloaded to a
Datalogging pauses upon entering Program Mode or
Datalogging automatically starts/stops when the monitor is
Set a monthly start/stop time to datalog using ProRAE

with particles, or draws in liquid.
Do not use the external filter when calibrating reactive gases.
Although this shortens battery life, it provides a faster response.

Alternatively, the calibration gas can be sampled from:

with calibration gas flow

when utilizing an RS-232 cable and/or transformer.

Monitoring gas concentrations continues during Programming
and [N/-] together until the first

Calibrate Monitor?
Change Alarm Limits?
Change Monitor Setup?
Change Sensor Configuration?
Change Datalog?
Fresh Air Calibration?
Multiple Sensor Calibration?
Single Sensor Calibration?
Modify Span Gas Value?
Change LEL / VOC Span Gas?
Change High Alarm Limit?
Change Low Alarm Limit?
Change STEL Alarm Limit?
Change Average Alarm Limit?
Clear All Data?
Change Datalog Period?
Select Data Type?
Enable / Disable Datalog?
Select Memory Full Type?
Change Site ID?
Change ID Mode?
Change User ID?
Change Alarm Mode?
Change User Mode?
Change Real Time Clock?
Change Backlight Mode?
Change Password?
Change Pump Speed?
Change Averaging Method?
Change Display Language?
Set Temperature Unit?
Change LEL / VOC Gas Selection?
Enable / Disable Sensor?
Change Dilution Ratio?
Change PID Lamp Type?
Sensor Name / Instantaneous Readings
Y/+ or N/-
Y/+ or N/-
Y/+ or N/-
Y/+ or N/-
Y/+ or N/-
MODE
MODE
MODE
MODE & N/-
Y/+
MODE
Y/+
Y/+
Y/+
Y/+
MODE
MODE
MODE
N/-
N/-
N/-
N/-
MODE
N/-MODE
MODE

For security purposes “0000” always

with pure nitrogen gas, by

halts datalogging when the memory is full, and the “MEM”

– the monitor turns the back light on/off when ambient

– (default) use when operating conditions are slow to change;
– use for long lengths of tubing or when rapid changes in input
(default) – an eight-hour Time Weighted Average or
The correction factor allows the unit to
to move from sensor to sensor. An asterisk
until “Save?” appears. Then press [Y/+] to accept.
to escape and cancel changes.

Attach an optional dilution fitting on the gas inlet port to dilute the gas
VOC and LEL sensors respond to a broad range of gases and show
Correction factors provide an estimate of the measurement gas

which are very critical to the operation of the monitor. Extra care
together; release both keys when the unit
until the desired diagnostic parameter appears:
Adjust Lamp Failure Threshold

Adjust the threshold for the LCD backlight to automatically turn on/off;

will slowly drain even if the monitor is turned off. If the battery packs

The alkaline battery adapter supplied by RAE Systems is intrinsically
safe.
The adapter is intended to be used in emergency situations
when there is no time to recharge the Li-Ion battery pack. The adapter
the adapter, remove the monitor cover. Remove the Li-Ion

All other toxic sensors have a 1-year warranty.
The Oxygen and the LEL sensors have assigned sockets.
Any toxic sensor can be installed in socket 2/B.
sensor(s), turn the monitor off
Calibrate all sensors prior to use!

sensors require a special high bias. The bias pins are

Avoid contact with the flat window surface.

email: Tech@raesystems.com

Adjust threshold. Clean or

oxygen and toxic gases (2)
with internal smart charging, 120V AC/DC wall adapter,

Auto shut-off pump at low ow condition
for pump-equipped instruments)