Rae MultiRAE Quick Start Manual

Page 1
MultiRAE
(Pumped Models)
QuickStart Guide
User Interface
The MultiRAE’s user interface consists of the display and three keys, [Y/+], [MODE] and [N/-]. The flippable LCD displays information such as monitored threats, real-time readings, measurement units, alarm type (when in alarm, including cal. overdue), battery and pump status, datalog (if on), and radio and connection quality (if available).
Charging The MultiRAE
Always fully charge the battery before use. Contacts on the bottom of the MultiRAE meet the Travel Charger’s or Charging Cradle’s contact pins, transferring power. Make sure the charger and MultiRAE are firmly attached. Then connect the AC Adapter’s plug to the charger, and plug its transformer into an AC outlet. While charging, the LED on the cradle glows red. When the battery is fully charged, the LED glows green.
Turning The MultiRAE Off
Press and hold [MODE]. A 5-second countdown to shut­off begins. You must continue pressing on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the MultiRAE continues normal operation.
When the countdown ends and the screen displays “Unit Off,” release your finger from the [MODE] key. The MultiRAE is now off.
Testing The Alarms
Under normal-operation mode and non-alarm condi­tions, the buzzer, vibration alarm, LED, and backlight can be tested anytime by pressing [Y/+] once. If any alarm does not respond, check the Alarm Settings in Programming Mode to make sure all alarms are enabled. If any alarms are enabled but are not functional, do not use the instrument.
2.
3.
5.
6.
4.
Charging status LED
[N/-] key
[MODE] key
Gas inlet
Display
LEDs
Alarm Buzzer
LEDs
[Y/+] key
Charging and Communication Contacts
External filter
AC adapter plug
Cable from AC adapter
MultiRAE
Charger
Phone: 408-952-8200 Fax: 408-952-8480 Email: customerserv@raesystems.com
www.raesystems.com
RAE Systems by Honeywell 3775 N. First St. San Jose, CA 95134-1708 USA
Page 2
Bump Testing and Calibration Setup
The MultiRAE can be automatically bump tested and calibrated using the AutoRAE 2 Test and Calibration System (refer to its User’s Guide for instructions). Manually calibrate using a fixed-flow regulator (flow rate between 0.5 and 1.0 liters per minute) and the supplied special calibration adapter that covers the gas inlet:
1. Connect the gas cylinder, flow regulator, tubing
(must use Teflon tubing for PID sensor), and calibration adapter to the MultiRAE.
2. With the MultiRAE in Normal Mode, enter
Programming Mode by pressing and holding both [MODE] and [N/-] until the password screen appears.
3. Input the 4-digit password. (The default password is
“0000.” If you do not know the password, select “Done.”) Then follow the menus to select single- or multi-sensor bump test, zero, or span calibration.
Important! After a bump test or calibration, remove the calibration adapter to ensure correct readings.
Span Calibration
In Programming Mode, and with “Multi Sensor Span” or “Single Sensor Span” highlighted:
1. Press [Y/+]. The screen displays the sensor(s) to be
calibrated.
· Multi: The list is shown.
· Single: Select a sensor and press [Y/+]
2. Attach the calibration adapter, and connect the
calibration gas cylinder’s flow regulator to the MultiRAE, and start the gas flow.
3. Press [Y/+] to start calibration.
4. Upon completion, a pass/fail calibration result
appears and the readings are shown (they should be within ±10% of the span gas value). Note: If a VOC sensor is installed, a second calibration can be performed to enhance linearity, requiring different calibration gas.
5. Turn off the gas and remove the calibration adapter.
Bump (Functional) Testing
Bump test to confirm that the sensors and alarms are functional.
Important! Test the alarms, as described in panel 6 (above), prior to performing a bump test.
Important! Make sure all sensors have warmed up before performing the bump test. You can tell a sensor has warmed up if you see a reading next to it name on the display. If it has not warmed up, you see three dashes next to it.
With the MultiRAE in Normal Mode:
1. Enter the Bump Test menu. Follow instructions
in panel 7 (above) or use the easy shortcut: Press both the [Y/+] and [N/-] buttons at the same time and hold them for 5 seconds. The Multi-Bump menu then appears.
BATTERY PACKS
A Li-Ion battery pack (PN: M01-3051-000 or M01-
3053-000) and an alkaline battery adapter (PN: M01-3052-000 or M01-3054-000) are supplied with each MultiRAE.
There are two types of output power for battery packs or adapters. The battery pack (PN: M01­3051-000) and adapter (PN: M01-3052-000) are used for MultiRAE model number PGM-62x0. Battery pack (PN: M01-3053-000) and adapter (PN: M01-3054-000) are used for model number PGM-62x6/PGM-62x8.
The alkaline battery adapter accepts four AA alkaline batteries (use only Duracell MN1500). Do not mix old and new batteries or batteries from different manufacturers.
8.
9.
10.
7.
11.
P/N: M01-4016-000 Rev C 201403
Page 3
Wireless Operation
If your MultiRAE is equipped with a wireless modem, its settings are controlled via the menu items under “Wireless.” In order to save time while operating the MultiRAE in a network, it is best to configure the settings before taking the MultiRAE into the field. Consult the User’s Guide for more detailed instructions.
1. Enter Programming Mode by pressing and holding
[MODE] and [N/-] simultaneously until the password screen appears.
2. Input the 4-digit password. (The default password
is “0000.” If you do not know the password, select “Done.”) Then follow the instructions for individual or multiple zero and span calibration.
3. Press [N/-] repeatedly until “Wireless” is
highlighted.
4. Press [Y/+] to select Wireless Settings.
WARNING
To reduce the risk of ignition of hazardous atmospheres, recharge, remove, or replace the battery only in an area known to be non-hazardous!
WARNING
Do not replace sensors in hazardous locations.
Basic Menu Navigation In Hygiene Mode
Pressing [N/-] repeatedly allows you to step through the screens as shown here.
Note: The first gray box only applies if a gamma radiation sensor is installed. The second gray box indicates the datalog functions when datalogging is in Manual mode. When datalogging is in Automatic mode, this screen does not appear. Refer to the User’s Guide for information on selecting Automatic, Manual, or Snapshot datalogging.
Note: Dashed line indicates automatic progression.
14.
13.
12.
Page 4
SPECIAL CONDITIONS FOR SAFE USE
1. The PGM-62xx shall only be fitted with
RAE Systems Battery Pack type M01­3051-000 or M01-3053-000 or Battery Adapter M01-3052-000 or M01-3054-000 fitted with Duracell MN1500 batteries.
2. The PGM-62xx shall only be charged
outside hazardous areas. No precautions against electrostatic discharge are necessary for portable equipment that has an enclosure made of plastic, metal, or a combination of the two, except where a significant static-generating mechanism has been identified. Activities such as placing the item in a pocket or on a belt, operating a keypad or cleaning with a damp cloth, do not present a significant electrostatic risk.
HAZARDOUS LOCATION APPROVALS
Exia Class I, Division 1, Groups A, B, C, D, T4
SIRA 11ATEX2152X, 0575 II 1G Ex ia IIC T4 Ga (for PGM62x0/PGM62x6)
SIRA 11ATEX2152X, 0575 II 2G Ex ia d IIC T4 Gb (for PGM62x8)
UM = 20V
IECEx SIR 11.0069X, Ex ia IIC T4 Ga (for PGM62x0/PGM62x6)
IECEx SIR 11.0069X, Ex ia d IIC T4 Gb (for PGM62x8)
WIRELESS CERTIFICATION
Complies with the following: FCC Part 15 R&TTE Directive (1999/5/EC)
WARNINGS
ANY RAPID UP-SCALE READING FOL­LOWED BY A DECLINING OR ERRATIC READING MAY INDICATE A GAS CONCEN­TRATION BEYOND UPPER SCALE LIMIT, WHICH MAY BE HAZARDOUS.
TOUTE LECTURE RAPIDE ET POSITIVE, SUIVE D’UNE BAISSE SUBITE AU ERRATIQUE DE LA VALEUR, PEUT INDIQUER UNE CONCENTRATION DE GAZ HORS GAMME DE DÉTECTION QUI PEUT ÊTRE DANGEREUSE.
CAUTION: BEFORE EACH DAY’S USAGE, SENSITIVITY OF THE LEL SENSOR MUST BE TESTED ON A KNOWN CONCENTRATION OF METHANE GAS EQUIVALENT TO 20 TO 50% OF FULL-SCALE CONCENTRATION. ACCURACY MUST BE WITHIN 0 AND +20% OF ACTUAL. ACCURACY MAY BE CORRECTED BY CALIBRATION PROCEDURE.
ATTENTION: AVANT CHAQUE UTILISATION JOURNALIERE, VERIFIER LA SENSIBILITE DU CAPTEUR DE LIE AVEC UNE CONCENTRATION CONNUE DE METHANE EQUIVALENTE DE 20 A 50% DE LA PLEINE ECHELLE. LA PRECISION DOIT ETRE COMPRISE ENTRE 0 ET 20% DE LA VALEUR VRAIE ET PEUT ETRE CORRIGEE PAR UNE PROCEDURE D’ETALONNAGE.
15.
16.
17.
Loading...