Rae MeshGuard, FTD-2000 User Manual

MeshGuard
User’s Guide
Rev. A
P/N D01-4002-000
1
FCC Part 15 statement
This device contains FCC ID: QNT PT2400GS20-A. This device complies with Part15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
© Copyright 2007 RAE Systems, Inc.
MeshGuard User’s Guide
2
Contents
1 Standard Kit ............................................................................. 5
2 General Information................................................................. 5
3 Physical Description ................................................................ 7
3.1 LCD Display .............................................................. 8
3.2 Specifications............................................................. 9
4 Operating The MeshGuard .................................................... 10
4.1 Turning The MeshGuard On.................................... 10
4.2 Turning The MeshGuard Off ................................... 11
4.3 Low Battery Indicator & Action.............................. 12
4.4 Wireless Communication Indicator ......................... 12
5 Operation Modes.................................................................... 13
5.1 Detection Mode........................................................ 14
5.2 Detection Mode (When Portable Mode Is Enabled) .... 16
5.3 Programming Mode ................................................. 18
5.3.1 Entering Programming Mode .............................. 19
5.3.2 Exiting Programming Mode ................................ 20
5.3.3 Changing Values.................................................. 21
5.3.4 Zero Calibration................................................... 23
5.3.5 Span Calibration................................................... 24
5.3.6 Change High Alarm ............................................. 27
5.3.7 Change Low Alarm.............................................. 28
5.3.8 Change STEL setting (Portable Mode Only)....... 29
5.3.9 Change TWA setting (Portable Mode Only) ....... 30
5.3.10 Change SPAN value .......................................... 31
5.3.11 Fixed or Portable Operation Selection............... 32
5.3.12 Radio Enable/Disable......................................... 32
5.3.13 Change Communication Interval ....................... 33
6 Diagnostic Mode.................................................................... 34
6.1 Entering Diagnostic Mode ....................................... 34
6.2 Exiting Diagnostic Mode ......................................... 35
6.3 Diagnostic Mode Readings ...................................... 35
6.3.1 Sensor Raw Count................................................ 35
6.3.2 Battery Raw Count............................................... 35
6.3.3 Temperature Raw Count ...................................... 35
6.3.4 Sudden Power Off Count..................................... 35
MeshGuard User’s Guide
3
6.3.5 Zero Raw Count................................................... 36
6.3.6 Calibration Delta Counts...................................... 36
6.4 Diagnostic Mode Programming............................... 36
6.4.1 Change Pan ID ..................................................... 38
6.4.2 Enable RFD or FFD ............................................. 38
6.4.3 Join Mesh Network.............................................. 38
6.4.4 Factory Setting..................................................... 38
6.4.5 Audible & Visible Alarm Enable/Disable ........... 38
6.4.6 Sensor Type ......................................................... 38
7 Sensor And Battery Replacement .......................................... 39
7.1 Battery replacement ................................................. 39
7.2 Sensor Filter Replacement ....................................... 40
7.3 Sensor Replacement................................................. 41
8 Troubleshooting ..................................................................... 42
9 Alarm Signal Summary.......................................................... 43
10 Warnings................................................................................ 45
11 Installation.............................................................................. 46
11.1 Magnet-Mount Installation ...................................... 46
11.2 Fixed Installation ..................................................... 47
11.3 Magnetic Mount Alternative Installation................. 50
MeshGuard User’s Guide
4
WARNINGS
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions.
Warning:
Use only the Lithium battery or external rechargeable battery provided by RAE Systems. This instrument has not been tested in an explosive gas/air atmosphere having an oxygen concentration greater than 21%. Substitution of components may impair suitability for intrinsic safety. Replacement of batteries only in non-hazardous locations.
STATIC HAZARD: Clean only with a damp cloth.
For safety reasons this equipment must be operated and serviced by qualified personnel only. Read and understand instruction manual completely before operating or servicing. Any rapid up-scale reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit, which may be hazardous.
Intrinsically safe marking:
IECEx EX ia II C IECEx number: TSA 07.0032X UL/CSA C1D1, group A,B,C, D T4 ATEX II 1G EEx ia IIC T4
MeshGuard User’s Guide
1 Standard Kit
Monitor with antenna User’s Guide CD with resources Calibration adapter Maintenance tool Calibration certificate
2 General Information
MeshGuard (FTD-2000) is a single toxic gas detector integrated with a wireless mesh network-enabled transmission radio module. It can work as a fixed device or as a portable device. Its built-in radio board operates on a frequency of 2.4GHz and complies with IEEE 802.15.4 standard. The FTD-2000 works with the FMC-2000 wireless controller on a flexible, robust wireless network to provide reliable, low-cost operation. It also works in a ProRAE.net network with a PC, and it supports point­to-point and point-to-multi-point networks.
Meshguard network with FMC-2000 controller
ProRAE.net network with PC controller
5
MeshGuard User’s Guide
6
Key Features
z Up to 6 months continuous operation z IEEE 802.15.4 Mesh network functionality with 64-bit encryption z Robust wireless mesh network with auto network forming and
configuration
z Operating distance: up to 300 m, line of sight z Very low-cost installation no hardwiring involved z Large area coverage with multi-hop mesh network z Sensors supported: H
2
S and CO (for others, consult RAE Systems)
z Field-replaceable battery and sensor. z Loud audio alarm, 90dB @ 30cm (12″) z Large, easy-to-read continuous display of gas concentration in ppm z User-adjustable high and low alarms z User-adjustable STEL and TWA alarms (in portable mode) z Bright red flashing alarm z Simple calibration z Highly resistant to RFI interference z IP-65 rated for outdoor use in harsh environments
MeshGuard User’s Guide
3 Physical Description
1
2
3
4
7
5
6
1 LED alarm 2 LCD 3 Buzzer alarm 4 Sensor gas inlet 5 Battery cover 6 Y/+, MODE, and N/- keys 7 Antenna Not visible Optional magnetic mount on rear
7
MeshGuard User’s Guide
3.1 LCD Display
1 2
3
4 5
6 7
8
9
10
11
12
13
14
11 22
3
4 5
6 7
8
9
10
11
12
13
14
1 Zero Calibration
2 Low Battery Indicator
3
Wireless Communication (
if on, the monitor is
in RDF; if blinking, the monitor is FFD)
4 Short Term Exposure Limit (STEL)
5 Time Weighted Average (TWA)
6, 8 High Alarm
7, 8 Low Alarm
9 Gas Concentration unit, ppm
10 Gas Concentration unit, %
11 Save Setting
12,13 Span Calibration
14 Reading Value
Remark: Concentration unit is shown as either ppm (9)
or % (10).
8
MeshGuard User’s Guide
9
3.2 Specifications
RF
Certifications
FCC Part15 CE EN 300328 SRRC(Pending)
Display
Customized LCD (1 x 1.5) with backlight
Audible alarm
90dB @ 30cm
Visual alarm
2 super-bright red LEDs
Calibration
Two-point field calibration
RF
IEEE 802.15.4/Zigbee with mesh stack
Operating
Range
Up to 300 meters, line of sight
Transmission
Power
Up to 18dBm (63mw EIRP)
Receiver
Sensitivity
Minimum -95dBm at 2.4GHz
User Interface
Three keys (Y/+, MODE, N/-)
Power Supply
D-size Lithium primary battery, +3.6V, 19 Ah
Max Current
Consumption
200mA@3.6V during transmission <0.5mA@3.6V during standby
Operation
Time
Internal Battery: For RFD* Mode: up to 6 Months For FFD** Mode: up to 10 days *RFD is reduced-function device **FFD is full-function device
Operating
Temperature
-20
°
to +50° C
(-30
°
to +50° C for intermittent operation)
Humidity
10% to 90% relative humidity, non-condensing
Dimensions
26.5cm x 9.5cm x 5.5cm (10.5 L x 3.7 W x 2.1 H)
Weight
0.6 kg (1.3 lbs)
Package
IP-65
Mounting
Optional stainless-steel bracket mount or magnetic mount
MeshGuard User’s Guide
4 Operating The MeshGuard
Make sure the battery is installed before operating the MeshGuard. Refer to page 39 for information on battery installation and replacement.
4.1 Turning The MeshGuard On
Hold down the [MODE] key and release it when the MeshGuard beeps. The monitor is now on, as indicated by the display:
The MeshGuard performs a self-test, followed by warm-up and zero calibration. The display briefly shows the firmware version (for example, “F1.0” means firmware version 1.0, “F2.0” means firmware version 2.0, etc.):
Then the MeshGuard initializes the wireless network and displays this information:
Note: If MeshGuard is unable to find a radio network to connect with, it searches, and the display alternates between “rdo” and “Srh” (for “radio search”).
10
MeshGuard User’s Guide
Next, the display tells you that MeshGuard is initiating network communication:
Next, if a network is located, an antenna icon appears (if no network is found, then the icon is not shown; press [Y/+] to search for a network). The current gas concentration reading is also displayed:
The MeshGuard is now operational.
4.2 Turning The MeshGuard Off
Hold down the [MODE] key through the “5…4…3… 2… 1… oFF” sequence. The monitor is off when the display is blank. Release the [MODE] key.
The MeshGuard is now off.
11
MeshGuard User’s Guide
4.3 Low Battery Indicator & Action
The battery is designed for up to six months’ life in RFD mode. When the battery gets low, the MeshGuard beeps once per minute. It is recommended that the battery be changed immediately, to minimize disruption.
When battery is completely depleted, the LCD displays “OFF,” and the LED, buzzer and vibration alarm activate once per second. The battery icon also blinks on and off. The MeshGuard shuts down after you press any key, or shuts down automatically if you do not press a key for 60 seconds.
4.4 Wireless Communication Indicator
When wireless communication is turned on, the LCD displays the wireless link status in the upper left corner:
If the MeshGuard finds and joins a wireless network, an antenna icon is shown in the display. If no link or a weak link is established, no antenna icon is shown. Press [Y/+] to search for a network.
A flashing antenna means the MeshGuard is in Full Function Mode (FFD), and the radio is always on.
12
MeshGuard User’s Guide
13
5 Operation Modes
The MeshGuard can operate as a Reduced Function Device (RFD) or a Full Function Device (FFD). In RFD mode, the MeshGuard transmits data to the host at a set interval (the default is 30 seconds) or anytime an alarm occurs. In FFD mode, the MeshGuard receives data in real time, and it can also work as a router as needed to relay data from RFD devices back to the host.
Note: Operating the MeshGuard in FFD mode reduces battery life.
The interval can be changed in Programming Mode. See page 18 for details.
In addition, there is a Detection Mode for standard operation, Programming Mode for making changes to values (such as the High Alarm, etc.), Diagnostic Mode (for servicing and checking the sensor, etc.), and Diagnostic Programming Mode, which is for selecting the sensor type, etc.
MeshGuard User’s Guide
5.1 Detection Mode
The MeshGuard can operate in fixed or portable mode. Whenever you start MeshGuard, it is automatically in fixed Detection Mode, which is the default.
The MeshGuard displays the current reading:
Pressing [MODE] steps through the Detection Mode screens:
Press [MODE], and it displays the sensor type:
14
MeshGuard User’s Guide
Press [MODE], and it alternates between EUI (Extended Unique Identifier) and its value:
Press [MODE], and it alternates between Pan and ID (Personal Area Network Identifier) and its value:
Press [MODE] to return to the detector reading:
Manually Sending Data
While the MeshGuard typically sends reading data to the network on a fixed interval, you can send the data anytime.
Press the [Y/+] key. The screen alternates between “Ini” and “nEt” one time, sends the current sensor data, and returns to the detector reading.
15
MeshGuard User’s Guide
5.2 Detection Mode (When Portable Mode Is Enabled)
When Portable Mode is enabled (see page 32), pressing [MODE] steps through three more screens, STEL, TWA, and Peak:
16
MeshGuard User’s Guide
The initial screen is the reading display:
Press [MODE], and it displays STEL. The STEL (short-term exposure
limit) is the average reading of the gas concentration for the last 15 minutes.
Note: The STEL reading does not appear until 15 minutes have elapsed.
If the MeshGuard has not been on for 15 minutes, it displays three dashes:
Press [MODE], and it displays TWA. The TWA (time-weighted average)
is the accumulated reading of the gas concentration, divided by 8 hours, since the monitor was turned on.
Press [MODE], and the display indicates the Peak reading by alternating between “P” (for “Peak”) and a numerical value:
17
MeshGuard User’s Guide
Press [MODE] to return to the current reading:
5.3 Programming Mode
Programming Mode allows you to perform any of the following
actions (listed in order of appearance):
Calibrate the MeshGuard
Changing Preset Limits or Span Gas Values
The menus accessed in Programming Mode are:
Zero Calibration
Span Calibration
High Alarm
Low Alarm
STEL (Portable Mode only)
TWA (Portable Mode only)
Fixed or portable operation
Radio enable/disable
Communication interval
18
MeshGuard User’s Guide
5.3.1 Entering Programming Mode
To enter the Programming Mode, press [MODE] and [N/-] for 3 seconds. “Pro” appears in the display:
Pressing [N/-] steps you through all the screens and then returns to the first programming display:
Each display alternates between its name and a status message or value.
Note: You can exit Programming Mode at any time by pressing [MODE].
Also, if you do not make a change within one minute, the MeshGuard exits Programming Mode and returns to Detection Mode.
19
MeshGuard User’s Guide
Settings can be changed as follows:
1. When a menu is selected, it flashes between two screens. For
example:
2. Press [MODE] to exit Program Mode and return to Detection Mode,
or press [N/-] to advance to the next menu.
3. Press [Y/+] to enter. The LCD displays “go.”
5.3.2 Exiting Programming Mode
You may exit Programming Mode anytime and return to Detection Mode in either of these two ways:
1. Press [MODE]. MeshGuard exits Program Mode and shows the
current reading in Detection Mode.
2. Do not press any buttons for 1 minute. MeshGuard automatically
exits Programming Mode and returns to Detection Mode, showing the current reading.
20
MeshGuard User’s Guide
5.3.3 Changing Values
When an editable numerical value is shown, the most significant digit flashes:
Change the value by pressing [Y/+] to increase it, and [N/-] to decrease it.
Press [MODE] to advance to the next digit:
After editing the last digit, press [MODE]. A question mark (?) is shown in the display, asking if you want to save the change:
21
MeshGuard User’s Guide
Press [Y/+] for “yes.” The message “dn” indicates the change is done.
Press [N/-] for “no.” A “no” message means that the change was abandoned.
Press [MODE] to return to the first digit and continue editing values for that menu.
22
MeshGuard User’s Guide
5.3.4 Zero Calibration
When “CAL” and “go” are displayed in alternation, and “ZERO” is shown, the MeshGuard is ready to perform a zero calibration.
Press [Y/+]. The LCD displays “go.”
The display counts down from 10 to 0.
After the countdown reaches 0, the LCD displays “dn,” for “done.” The reading should show 0 (zero). Otherwise, repeat the zero calibration.
23
MeshGuard User’s Guide
Note: To stop zero calibration before the countdown reaches 0, press any
key. The LCD displays “no” and advances to the next programming menu, Span calibration.
5.3.5 Span Calibration
“CAL” and “go” flash in alternation, and “SPAn” is shown. The MeshGuard is now ready to perform a span calibration.
To start calibration, press [Y/+]. The LCD displays “go.”
The MeshGuard waits for 10 seconds so that you have time to connect the span gas. Connect the calibration gas adapter to the MeshGuard, and connect the gas cylinder to the adapter. Start the flow of gas.
24
MeshGuard User’s Guide
When the gas flow starts, the LCD displays “gAS” and the span concentration value.
The MeshGuard now counts down to 0.
Note: The countdown time varies according to the type of sensor used in
the MeshGuard.
After counting down and reaching 0, the LCD displays “dn.” The reading should be the span concentration value. Otherwise, the span calibration should be repeated.
25
MeshGuard User’s Guide
If the MeshGuard does not detect gas after counting down to 0, the LCD displays “Err” (for “error”). The LED glows red and the buzzer sounds to provide extra warning. The MeshGuard automatically returns to the span calibration display.
Note: To stop span calibration before the countdown reaches 0, press any
key. The LCD displays “no” and advances to the next programming menu, Change High Alarm.
26
MeshGuard User’s Guide
5.3.6 Change High Alarm
At the menu for changing the High Alarm setting, “ Set” and “ go” flash in alternation, and both “HIGH“ and “ALARM” are shown.
Press [Y/+] to enter and change the setting. Press [MODE] to go back to Detection Mode or [N] to advance to the next menu.
The LCD displays the current value. Change the value if necessary.
To change the value:
1. Press [Y/+] to increase the number and [N/-] to decrease it.
2. Press [MODE] to advance to the next digit.
3. After moving to the last digit and making changes, press
[MODE]. A question mark (?) is shown in the display, asking if you want to save the change.
Press [Y/+] for yes. The message “dn” means the change is
done.
Press [N/-] for no. A “no” message means that the change
was abandoned.
Press [MODE] to return to the first digit.
27
MeshGuard User’s Guide
5.3.7 Change Low Alarm
At the Change Low Alarm menu, “Set” and “go” flash in alternation, and “LOW“ and “ALARM” are visible in the display.
Press [Y/+] to enter and change the setting. Press [MODE] to exit and return to Detection Mode or [N] to advance to the next menu.
The LCD displays current value. Change the value if necessary.
To change the value:
1. Press [Y/+] to increase the number and [N/-] to decrease it.
2. Press [MODE] to advance to the next digit.
3. After moving to the last digit and making changes, press
[MODE]. A question mark (?) is shown in the display, asking if you want to save the change.
Press [Y/+] for yes. The message “dn” means the change is
done.
Press [N/-] for no. A “no” message means that the change
was abandoned.
Press [MODE] to return to the first digit.
28
MeshGuard User’s Guide
5.3.8 Change STEL setting (Portable Mode Only)
Note: This menu is only available when the MeshGuard is in Portable
Mode. If it is in Fixed Mode, you will not see this menu and cannot change its settings.
“Set” and “go” flash in alternation, and “STEL” is shown.
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to advance to the next menu.
The LCD displays current value. Change the value if necessary.
To change the value:
1. Press [Y/+] to increase the number and [N/-] to decrease it.
2. Press [MODE] to advance to the next digit.
3. After moving to the last digit and making changes, press
[MODE]. A question mark (?) is shown in the display, asking if you want to save the change.
Press [Y/+] for yes. The message “dn” means the change is
done.
Press [N/-] for no. A “no” message means that the change
was abandoned.
Press [MODE] to return to the first digit.
29
MeshGuard User’s Guide
5.3.9 Change TWA setting (Portable Mode Only)
Note: This menu is only available when the MeshGuard is in Portable
Mode. If it is in Fixed Mode, you will not see this menu and cannot change its settings.
“Set” and “go” flash in alternation, and “TWA” is shown.
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to advance to the next menu.
The LCD displays the current value. Change the value if necessary.
To change the value:
1. Press [Y/+] to increase the number and [N/-] to decrease it.
2. Press [MODE] to advance to the next digit.
3. After moving to the last digit and making changes, press
[MODE]. A question mark (?) is shown in the display, asking if you want to save the change.
Press [Y/+] for yes. The message “dn” means the change is
done.
Press [N/-] for no. A “no” message means that the change
was abandoned.
Press [MODE] to return to the first digit.
30
MeshGuard User’s Guide
5.3.10 Change SPAN value
“Set” and “go” flash in alternation, and “SPAN” and a gas cylinder icon are shown.
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to advance to the next menu.
The LCD displays the current value. Change the value if necessary.
To change the value:
1. Press [Y/+] to increase the number and [N/-] to decrease it.
2. Press [MODE] to advance to the next digit.
3. After moving to the last digit and making changes, press
[MODE]. A question mark (?) is shown in the display, asking if you want to save the change.
Press [Y/+] for yes. The message “dn” means the change is
done.
Press [N/-] for no. A “no” message means that the change
was abandoned.
Press [MODE] to return to the first digit.
31
MeshGuard User’s Guide
5.3.11 Fixed or Portable Operation Selection
The next menu is for setting the MeshGuard for fixed or portable operation. In Portable Mode, STEL, TWA, and Peak readings are included.
“SEt” and “ Ptb” flash in alternation, to indicate that the MeshGuard can
now be set as fixed or portable. The default value is “fixed.”
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to go to the next menu.
The LCD displays the current value (enabled or disabled). Change the value if necessary.
Note: “dIS” means fixed (Portable Mode disabled) and “En” means
portable (Portable Mode enabled).
5.3.12 Radio Enable/Disable
The next menu is for setting turning the MeshGuard’s radio on or off.
“SEt” and “ rdo” flash in alternation, to indicate that the MeshGuard radio can be turned on (enabled) or off (disabled). The default value is “on.”
32
MeshGuard User’s Guide
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to go to the next menu.
The LCD displays the current value (enabled or disabled). Change the value if necessary.
Note: “dIS” means disabled (radio off) and “En” means on (radio on).
5.3.13 Change Communication Interval
This menu allows you to change the interval between wireless transmissions. “SET” and “ItUL” flash, to indicate that you can change the interval. The interval can be set to 10, 30, 60, 300, or 600 seconds.
Note: The default interval is 30 seconds.
Press [Y/+] to enter and change the setting, [MODE] to exit and return to Detection Mode, or [N/-] to advance to the next MENU.
33
MeshGuard User’s Guide
6 Diagnostic Mode
Diagnostic Mode provides raw data from sensors and about settings.
6.1 Entering Diagnostic Mode
Note: To enter Diagnostic Mode, you must begin with the MeshGuard
turned off.
Press and hold [Y/+] and [MODE] until the MeshGuard starts.
The instrument goes through a brief startup, and then displays “dIA” to indicate it is in Diagnostic Mode. It then switches to showing raw data for the sensor (the display shows “rAU” followed by a number).
The following chart shows how to navigate Diagnostic Mode (pressing N/- repeatedly steps through the screens):
34
MeshGuard User’s Guide
35
6.2 Exiting Diagnostic Mode
Note: You can exit Diagnostic Mode and enter Programming Mode and
calibrate the MeshGuard as usual by pressing both [MODE] and [N/-] for three seconds.
Note: You can exit Diagnostic Mode and enter Detection Mode by
pressing [MODE] and [Y/+] together for three seconds, or by turning it off and on again.
6.3 Diagnostic Mode Readings
In Diagnostic mode, you can step through readings by pressing [N/-].
6.3.1 Sensor Raw Count
Sensor Raw Count is indicated by “rAU” followed by a number.
Press [N/-] to advance to the next reading.
Press [MODE] and [Y/+] together for three seconds to exit
Diagnostic Mode and enter Detection Mode.
6.3.2 Battery Raw Count
Battery Raw Count is indicated by “bAt” followed by a number.
Press [N/-] to advance to the next reading.
Press [MODE] and [Y/+] together for three seconds to exit
Diagnostic Mode and enter Detection Mode.
6.3.3 Temperature Raw Count
Temperature Raw Count is indicated by “tNp” followed by a number.
Press [N/-] to advance to the next reading.
Press [MODE] and [Y/+] together for three seconds to exit
Diagnostic Mode and enter Detection Mode.
6.3.4 Sudden Power Off Count
Sudden Power-Off Count is indicated by “rSt” followed by a number.
Press [N/-] to advance to the next reading.
MeshGuard User’s Guide
36
Press [MODE] and [Y/+] together for three seconds to exit
Diagnostic Mode and enter Detection Mode.
6.3.5 Zero Raw Count
Zero Raw Count is indicated by “0rC” followed by a number.
Press [N/-] to advance to the next reading.
Press [MODE] and [Y/+] together for three seconds to exit
Diagnostic Mode and enter Detection Mode.
6.3.6 Calibration Delta Counts
Calibration Delta CTS is indicated by “CdC” followed by a number.
Press [N/-] to return to the first raw count.
Press [MODE] and [Y/+] together for three seconds to exit
Diagnostic Mode and enter Detection Mode.
6.4 Diagnostic Mode Programming
You can enter a special programming mode from Diagnostic Mode in order to perform advanced programming functions. These include:
Change Pan ID
Join a Mesh Network
Enable RFD or FFD
Return MeshGuard to original factory settings
Enable/disable audible and visible alarms
Select the sensor type
Note: When the MeshGuard is in this programming mode, if you do not
make a change or press a key for 60 seconds, it reverts to the standard Diagnostic Mode.
Important! After you make changes in Diagnostic Mode, it is
recommended that you turn off the MeshGuard and turn it on again before using it.
MeshGuard User’s Guide
Enter this programming mode by first entering Diagnostic Mode. This requires starting the MeshGuard while holding Y/+ and [MODE]. When you see the Sensor Raw Count screen, hold [MODE] and N/- until you see “Pro” in the display, indicating that you are in Programming Mode.
Step through the menus by pressing N/-. Exit by pressing [MODE], and then shutting off the MeshGuard and restarting it.
37
MeshGuard User’s Guide
38
6.4.1 Change Pan ID
Press Y/+ to enter the menu to make changes to the value.
1. Press [Y/+] to increase the number and [N/-] to decrease it.
2. Press [MODE] to advance to the next digit.
3. After moving to the last digit and making changes, press
[MODE]. A question mark (?) is shown in the display, asking if you want to save the change.
Press [Y/+] for yes. The message “dn” means the change is
done.
Press [N/-] for no. A “no” message means that the change
was abandoned.
Press [MODE] to return to the first digit.
6.4.2 Enable RFD or FFD
Press [Y/+] to toggle the setting from RFD to FFD, and vice versa.
6.4.3 Join Mesh Network
Press [Y/-] to initiate joining a network. Three bars flash in sequence while it searches. When it is done, it alarms once and displays “dn” for “done.”
6.4.4 Factory Setting
Press [Y/+] to return the MeshGuard to its original factory settings.
6.4.5 Audible & Visible Alarm Enable/Disable
Press [Y/+] to toggle between the MeshGuard’s audible and visible alarms turned on and off.
6.4.6 Sensor Type
Select the type of sensor that is in the MeshGuard.
1. Press [Y/+] to step through the sensor types.
2. Press [MODE] to save your change. A question mark (?) is
shown in the display, asking if you want to save the change.
Press [Y/+] for yes. The message “dn” means the change is done.
Press [N/-] for no. A “no” message means that the change
was abandoned.
MeshGuard User’s Guide
7 Sensor And Battery Replacement
Sensor compartment
Battery compartment
Sensor and battery removal tool (P/N 019-2044-000)
3-pin end
Hexagonal end
7.1 Battery replacement
1. Use the 3-pin end of the tool to unscrew and open the battery cover
by turning it counterclockwise.
2. Remove the battery.
3. Insert the new battery with its positive (“+”) pole towards inside of
the unit.
4. Replace the battery cover by turning it clockwise with the 3-pin end
of the tool.
Note: Only Change battery in a safe location and use the battery RAE
Systems provided, model ER34615 or XL-205F. After changing the battery, wait at least 60 seconds before turning the MeshGuard on.
39
MeshGuard User’s Guide
Battery
Battery cover
7.2 Sensor Filter Replacement
1. Use the 3-pin end of the tool to unscrew and open the filter holder by
turning it counterclockwise.
Filter
Filter holder
2. Remove and discard the filter.
3. Place a new filter inside the monitor.
4. Replace the filter holder by turning it clockwise with the 3-pin end of
the tool.
40
MeshGuard User’s Guide
7.3 Sensor Replacement
1. Use the 3-pin end of the tool to unscrew and open the filter holder at
the bottom of the monitor.
Sensor
Sensor cover
Filter
2. Use the hexagonal end of the tool to open the remove the sensor
cover, turning counterclockwise.
Filter holder
3. Pull the old sensor out.
4. Gently push a new sensor into the compartment.
Important! Ensure that the RAE Systems part number matches the
sensor that was removed.
WARNING! Use only the same sensor model as the one installed
when the monitor was purchased.
5. Replace the sensor compartment cover using by turning it
clockwise, using the hexagonal end of the tool.
6. Replace the filter holder by turning it clockwise, using the 3-pin
end of the tool.
Note: Always recalibrate the MeshGuard after service to ensure
functionality.
41
MeshGuard User’s Guide
42
8 Troubleshooting
Failure Symptom Cause Solution
Cannot turn on Battery charge too low
Battery has been changed
New battery needs to be discharged before use
Replace battery
Wait at least 60 seconds to turn on MeshGuard
Check RAE Systems web site for inform­ation on batteries
Abnormally high reading
Incorrect gas calibration Sensor low sensitivity to calibration gas
Recalibrate Replace the sensor
0” Alarm
Sensor zero drift Perform zero
calibration
Controller cannot receive the MeshGuard’s signal
Too much distance between the MeshGuard and the controller.
There is an obstruction between the MeshGuard and the controller.
Controller does not receive completed data packet
Battery is low
MeshGuard and controller have different Pan ID numbers
The distance should be 300 m, line of sight. Deploy FFD MeshGuard or repeaters.
Relocate the MeshGuard or deploy FFD MeshGuard or repeaters. .
Press [Y/+] on the detector to force it to send data packets
Replace battery
Set both units to have
MeshGuard User’s Guide
the same Pan ID number
No Antenna Icon There is no reader or
controller nearby. The controller or reader’s network has changed.
The MeshGuard is out of its RF range.
Battery is low
Move the MeshGuard closer to a working controller or reader. Perform the network searching function in diagnostic mode.
Move the MeshGuard close to a working controller or reader and Press [Y/+]
Replace battery
Others Turn MeshGuard off
and on again.
Consult RAE Systems Customer Service.
9 Alarm Signal Summary
Alarm
Mode
When LCD
Buzzer
& LED
Over Range
H
2
S > 100 ppm
CO > 500 ppm
or 2000 ppm
3 beeps
per
second
High Alarm
> high alarm
setting
3 beeps
per
second
Low Alarm
> low alarm
setting
2 beeps
per
second
43
MeshGuard User’s Guide
TWA
> TWA
setting
1 beep
per
second
STEL
> STEL
Setting
1beep
per
second
Zero Drift < 0 ppm
1 beep
per
second
Battery Low < 3.2V
1 beep
per
minute
Battery
Exhausted
< 3.1V
1 beep
per
second
44
45
Appendix A 10 Warnings
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions.
Warning:
Use only the Lithium battery provided by RAE Systems. This instrument has not been tested in an explosive gas/air atmosphere having an oxygen concentration greater than 21%. Substitution of components may impair suitability for intrinsic safety. Replacement of batteries only in non-hazardous locations.
STATIC HAZARD: Clean only with a damp cloth.
For safety reasons this equipment must be operated and serviced by qualified personnel only. Read and understand instruction manual completely before operating or servicing. Any rapid up-scale reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit, which may be hazardous.
FCC Part 15 statement and CE
This device contains FCC ID: QNT PT2400GS20-A. This device complies with Part15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Intrinsically Safe Marking
IECEx EX ia II C IECEx number: TSA 07.0032X UL/CSA C1D1, group A,B,C, D T4 ATEX II 1G EEx ia IIC T4
MeshGuard User’s Guide
Appendix B
11 Installation
Two methods for mounting MeshGuard make it easy to install. The first method uses a magnet that screw onto the rear of the Meshguard, making ideal for moving from one location to another. The second method uses a specially designed stainless-steel enclosure that is permanently mounted. It protects the MeshGuard from damage in industrial settings.
11.1 Magnet-Mount Installation
Magnet For Mounting MeshGuard
Rear of MeshGuard
This magnet is powerful enough to support the MeshGuard when it is placed against a flat steel or iron surface.
Important! Keep the magnet away from computer hard drives. The
strong magnet can corrupt or erase data on these.
46
MeshGuard User’s Guide
11.2 Fixed Installation
Four reinforced holes in the rear of the enclosure allow for a screw to pass through to the mounting brackets.
Mounting Holes
Side View
Front View
Sensor Cover
The enclosure can be mounted to a vertical or horizontal pole.
V
ertical pole Horizontal pole
47
MeshGuard User’s Guide
Slip the screws through the two holes that are side by side in order to mount the enclosure to a vertical pole. Otherwise slip the screws through the two vertically aligned holes to attach the enclosure to a horizontal pole.
Loosely assemble the clamp parts around the pole. Note that the screws have nuts that fit into the clamp parts. The clamp parts are designed to hole the nut so that you do not need to use a wrench. Hand-tighten the parts until snug.
Tighten the hex screws from the front of the enclosure:
Once the clamp parts and the enclosure are securely held against the pole, stop tightening.
Note: The pole must be between 25mm (1″) and 63mm (2.5) in
diameter.
48
MeshGuard User’s Guide
Next, place the MeshGuard into the enclosure:
1. Lift up the hinged cover of the enclosure.
2. Slide the MeshGuard into the enclosure from the top.
3. Close the cover of the enclosure.
4. Insert the hex screw into the cover’s locking portion, and tighten
it.
5. The MeshGuard is now ready to use.
1 2 3 4 5
Note: The sensor cover on the bottom of the enclosure can be removed
so that the filter and sensor can be inspected without removing the MeshGuard from the enclosure. Simply pull off the cover and follow the maintenance procedures in this guide.
49
MeshGuard User’s Guide
11.3 Magnetic Mount Alternative Installation
The magnet-mount disc can be attached to the steel enclosure instead of the clamps. This approach provides the protection of the enclosure with the ease of installation afforded by the magnetic mounting.
1. Insert screw through magnetic disc.
2. Place the magnetic disc over the bottom hole on the rear of the
enclosure.
3. Tighten the screw until the disc is snug.
Side View
Rear View
50
RAE Systems
World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: 408.952.8200
Fax: 408.952.8480
E-mail: customerserv@raesystems.com
Web Site: www.raesystems.com
Rev. A
July 2008
P/N D01-4002-000
Loading...