RAE E-1400 Operations Manual

Page 1
E-1400
Four Channe l Menu Drive Programmable Inductive Loop Vehicle
Four Loop Inputs, Eight Vital Outputs, and Four Vital Inputs
Detector Designed for Railroad Applications
Built-in Loop Analyzer for Each Channel
Loop Detector
Firmware Version 4.02
Operations Manual
This manual contains technical in formation for the
E-1400 Loop Detector
pn 889-0505-03 Revision: April 2020
Page 2
This page intentionally left blank.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 2 of 43
Page 3
THE FOLLOWING PRODUCT WAS DESIGNED, INSPECTED,
TESTED AND MANUFACTURED IN THE USA BY
EBERLE DESIGN, INC. IN PHOENIX, ARIZONA.
INFORMATION CONTAINED HEREIN IS PROPRIETARY
TECHNICAL INFORMATION OF EBERLE DESIGN, INC.
PUBLICATION, REPRODUCTION OR USE IN WHOLE OR
PART IS NOT PERMITTED EXCEPT UNDER TERMS
AGREED UPON IN WRITING.
ALL REGISTERED TRADEMARKS O F EBERLE DESIGN INC.
ARE UNDER © COPYRIGHT
MAINTENANCE NOTE
IT IS AN EBERLE DESIGN, INC. RECOMMENDATION THAT
EACH UNIT BE TESTED AT LEAST ANNUALLY TO
ENSURE COMPLIANCE WITH FACTORY SPECIFICATIONS
AND MEETS PROPER OPERATIONAL STANDARDS. THE
RESULTS OF THIS TESTING WILL BE DOCUMENTED.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 3 of 43
Page 4
This page intentionally left blank.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 4 of 43
Page 5
MODEL E-1400 OPERATION MANUAL
TABLE OF CONTENTS
1.0 GENERAL DESCRIPTION ..................................................................................................................... 7
2.0 GENERAL CHARACTERISTICS ........................................................................................................... 8
2.1 Loop Frequency ........................................................................................................................ 8
2.2 Sensitivity ................................................................................................................................. 8
2.3 Option 1, Loop Inductance Display........................................................................................... 8
2.4 Option 2, Loop Inductance -ΔL/L Displa y ................................................................................ 8
2.5 Option 3, Noise Filter ............................................................................................................... 8
2.6 Option 4, Detection Zone Tracking ........................................................................................... 9
2.7 Option 5, Sensitivity Boost ....................................................................................................... 9
2.8 Option 6, Audible Detect Signal ............................................................................................... 9
2.9 Loop Check .............................................................................................................................. 9
2.10 Detector ID ............................................................................................................................. 10
2.11 Optio n 7, Sync h Feature.......................................................................................................... 10
2.12 Optio n 8, Induc tive Island Circuit Fea ture .............................................................................. 10
2.13 Option 9, Inductive Shunt Enhancement Circuit Feature ......................................................... 11
2.14 Prima ry Vita l Output Delay .................................................................................................... 12
2.15 Primary Vital Output Extension .............................................................................................. 12
2.16 Input / Output Mapping .......................................................................................................... 12
2.17 Full Restore To Factory Default Sett ings ................................................................................ 12
2.18 Diagnostic Features Activat ion ............................................................................................... 12
2.19 Password Protection ................................................................................................................ 12
3.0 SPECIFICATIONS ................................................................................................................................. 13
3.1 Physical .................................................................................................................................. 13
3.2 Electrical ................................................................................................................................. 13
3.3 Operational ............................................................................................................................. 14
3.4 TABLE: Sens itivity, -L/L, & Response Time ...................................................................... 16
3.5 TABLES: Default Settings ..................................................................................................... 16
3.6 TABLES: Pin Assignments.................................................................................................... 17
4.0 USER INTER FACE ............................................................................................................................... 19
5.0 INSTALLATION AND SET-UP ............................................................................................................ 20
5.1 Initial Installation Star t-up Mode Display Screens .................................................................. 21
5.2 Program Mode Display Screens .............................................................................................. 22
5.2.1 Primary Vital Output (Vital Out) Mapping Screens ................................................ 25
5.2.2 Auxiliary Vital Output (Vital Au x) Mapping Screens .............................................. 28
5.3 Normal Mode Displa y Screens ............................................................................................... 32
5.4 Error Message Display Screens ............................................................................................... 34
5.5 Loop Check Pass / Fail Indications ......................................................................................... 36
5.6 Vital Input (Vital In) Status / Fail Indicat ions ......................................................................... 37
5.7 Primary Vital Output (Vital Out) Status / Fail Indications ....................................................... 37
5.8 Auxiliary Vital Output (Vital Au x) Status / Fail Indications .................................................... 37
5.9 Synch Fail In dications............................................................................................................. 37
5.9.1 Series Synch Fail In dications .................................................................................. 37
5.9.2 Parallel Synch Fail Indications ................................................................................ 37
5.10 Loop Fail Indications .............................................................................................................. 38
5.11 Setti n g Sensitivity using the B argrap h .................................................................................... 38
5.12 Full Restore To Factory Default Sett in gs ................................................................................ 39
5.13 Diagnostic Features Activat ion ............................................................................................... 39
5.14 Display Test ............................................................................................................................
5.15 Changing or Relocating Detectors........................................................................................... 39
6.0 BLOCK D IAGRAM ............................................................................................................................... 40
7.0 THEORY OF OPERATION ................................................................................................................... 41
8.0 MAINTENANCE AND TROUBLESHOOTING ................................................................................... 42
8.1 Basic Troubleshooting Tips .................................................................................................... 42
8.2 Things To Know About Loops ................................................................................................ 43
39
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 5 of 43
Page 6
This page intentionally left blank.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 6 of 43
Page 7

1.0 GENER AL DESCRIPTION

This product manual was wr itten for people installing, operating, a nd maintaining the R eno A&E Model E-1400 Series inductive loop vehicle detector. The Model E-1400 is a rack mount, four-chan ne l i nductive loop vehicle detector designed to work in conjunction with Railroad Control Limited’s Exit Gate Management System (EGMS).
The Model E-1400 monitors and process es signals from four loop / lead-in circuits, fo ur check loop circuits, and four Vita l inputs. These inputs can be programmed to control eight Vital outputs, fo ur primary Vital outputs and four auxiliar y Vital outputs. Each of the four primary Vital outputs can be controlled by any co mbinat ion of t he four loop inputs and/or a ny co mbination of t he four V ital inputs . In addition to the four loop inputs and four Vital inputs (as descr ibed above), the four auxiliary Vita l outputs can be controlled by the four loop Health Status inputs that are co ntinuously monitored by the p rocessor. A loop failure will deactivate t he Vital out put that the loop status inp ut and/or loop Health Status input is programmed to control. The check loop circuits can be programmed to simulate a vehicle detection (every 1 to 255 minutes) in the loop area as a means of verifying proper loop operation. The Vital outp uts and Vital inputs are monitored to ensure they are in the correct state. If the state of a Vital input is not correct, the front panel LED indicato r correspond ing to that input will t urn red and a ll Vital o utputs will be deactivated. If the state of any Vital output is not correct, the front panel LED corresponding to that output will turn red and all Vital outputs will be dea ctivated.
A Liquid Crystal Display (LCD), s ixteen L EDs, and fo ur front panel p ushbuttons are used to displa y and program all detector functions. T he Model E-1400 also incorporates a fro nt mounted nine pin RS-232 connector to facilitate set-up and operation.
The use of a LCD is one feature that distinguishes this detector from that of other manufacturers. It allows for more information, never before available, to be displayed to the user dur ing normal operation of the detector. Several diagnostic modes are available to aid the technician in troubleshooting detection problems. It also allows easy viewing and setting of all programmab le va lues in the detector. The detector continues to operate normally while being programmed. An eight-seg ment bargraph at the top of the LCD provides a graphical representation o f the relative change of inductance as seen by the detector. This automatically takes into account loop size, loop inductance, number of loops, number of turns, geometry, lead-in length, etc. The first bargraph segment represents the minimum inductance change necessary for the detector to output a call. Larger inductance changes will be indicated by more segments. Each additional segment represents the next sensitivity level exceeded. Therefore, the bargraph indicates if the sensitivity is set too high or too low, e as ily fac ilita ting the ideal sett i ng of sensitivity level.
All programmed settings a re stored in non-volatile memory and ca n only be changed by programming new settings. Memory storage consists of an internal EEPROM and an external EEPROM memory module. Settings programmed on either component must match t hose p rogra mmed on the o ther f or the detector to operate. Loss of power or a detector reset will not change the programmed settings. If a loop failure occurs, the LCD w ill display the type of loop failure as L lo (for -25% change or shorted loop conditions) or L hi (for +25% change or open loop conditions). Each loop failure is counted (up to a maximum of 255 failures) and accumulated in t he Loop Failure Me mory. Information on the nu mber of failures since the last detecto r reset or power interruption is very useful during analysis of proble ms d ue to intermittent loop operation.
The Model E-1400 detector is a scanning detector. The scanning ope ration sequentially activates the ON and OFF cycle of each detector channel’s oscillator. Because only one channel’s input loop(s) is (are) active at a given time, crosstalk between adjacent loops connected to the same scanning detector is minimized. Several Mo d e l E-1400 detectors can be linked together via t he Synch In and Synch Out lines. This allows loops connected to different d etectors to be scanned in series or pa rallel. When in Program Mode, the Model E-1400 detector displays the real time loop freque ncy read ing for each cha nnel. T he eight frequency levels can be inc remented or decremented to provide prec ise frequency settings. This removes the guesswork when changing frequency sett ings to e liminate crossta lk. NO TE: Adjacent loops connected to different channels of a non-scanning detector or different scanning detectors should be set to diffe rent fre quencies with the maximum separation possible.
The Reno A&E Model E-1400 uses the first major innovation in inductive loop detectors since the introduction of digital detecto rs. The progra mming of all o f the detector’s parameters with fo ur normally open pushbutton switches not only simplifies setup by re moving b inary coded rotary or DI P s witches , b ut also increases the reliability of the detect or by removing the dependence on switch contac ts during normal operation. The detailed descriptions displayed on the LCD eliminate the interpretation of numerous LED flash rates to determine the detector status.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 7 of 43
Page 8

2.0 GENERAL CHARACTERISTICS

2.1 Loop Frequency

There are eight selectable loop frequency settings (normally in the range of 20 to 100 kilo hertz) per loop input. The actual loop operating frequency is a function of the loop / lead-in network and t he components of the loop oscillator circuit. The digital display of the actual loop opera ting frequenc y for each setting makes it easy to qu ickly identify a nd eliminate crossta lk in the most diff icult to configure instal latio ns. The freq uenc y disp lay is typically very stable when the loop is vacant and vehicles are not passing close to the loops. If the reading is varying by more than ±1 in the last digit, this is an indication of possible crosstalk between loops.

2.2 Sensitivity

There are ten selectable sensitivity levels pe r loop input, plus Continuous-Call and Loop-Off. The sensitivity levels are designed so that a one level increase actually doubles the sensitivity and a one level decrease halves the sensitivity. A unique bargraph displayed on the LCD make s it easy to quickly set sensitivity at the ideal level for any lo op / lead-in network situation. (See Section 3.4 for actual detection levels at each sensitivity level.)
C
ONTINUOUS-CALL: W hen se t to the Cont inuo us-Call state, the Lo op Sta tus is conti nuo usly in t he
Call state regardless of the presence or absence of vehicles over the loop. If the Loop Status Input is assigned to a Vital Output, the Cont inuous-Call state will deactivate the Vital Output. The loop oscillator is disab led w he n in t he Co nt inuo us-Ca ll State. This sta te is indicated b y Call flashing on the LCD. This option can be selected from the Sensitivity menu in Program Mode and is useful for checking controller response and other troubleshooting activities.
L
OOP-OFF: When set to the Loop-Off state, the Loop Status is continuously in the No-Call state
regardless of the presence or absence of vehicles over the loop. If the Loop Status Input is assigned to a Vital Output, t he Loop -O ff state w ill not a ffec t the Vita l O ut put. T he lo op osc il la t o r is disabled when in t he Loo p -O ff S tate . This state is indicated by OFF flashing o n the LCD. This option can be selected from the Sensitivity menu in Program Mode and is useful for checking controller response and other troubleshooting activit ies.

2.3 Option 1, Loop Inductance Display

The Loop Inductance Display sett ing can be toggled ON or OFF b y momentarily pressing either the (UP) or  (DOWN ) pus hbu tton. Whe n th is opt ion is ena bled (ON), the LCD displays the total loop inductance (actual loop inductance plus actual lead-in inducta nce) in micr ohenr ies f or loo p ind ucta nce values in the range of 20 to 2500 microhenries. By recording the inductance of the loop / lead-in circuit when it is first installed, the actual inductance can be compared to the expected inductance to help identify defective loop / lead-in circuits. Loop / lead-in inductance can be easily estimated using the simple formulas included in Section 8.2 of this manual.
NOTE: Enabling this option activates it for all loop inputs. This option is automatically disabled 15 minutes after activation or on loss of power.

2.4 Option 2, Loop Ind uc t a nc e -Δ L/L Display

The Loop Inductance -ΔL/L Display setting can be toggled ON or OFF b y momentarily pressing eithe r the  (UP) or  (DOWN) p ushbutto n. When this option is enabled (ON), the LCD disp lays the percentage of inductance change (-ΔL/L value) during the Call state. To facilitate the viewing of the maximum amount of change in the -ΔL/L value while traffic is in motion over the detection zone, the channel holds the peak -ΔL/L value for a period of two seconds.
NOTE: Enabling this option activates it for all loop inputs. This option is auto matically disabled 15 minutes after activation or on loss of power.

2.5 Option 3, Noise Filter

Each loop input’s Option 3 has four settings ranging from 3.00 to 3.03. When Option 3 is set to 3.00, the opt io n is dis ab le d ( OFF), and normal noise filter ing is used. W hen Opt ion 3 is set to 3.01, 3. 02, or
3.03 the option is ON, and increased filtering of the loop signal is used to eliminate problems associated with particular types of noise. Turning this option ON when not required can degrade the performance of the detector. The default setting of this option is 3.00 (OFF). This option should only
be turned ON at the advice of Reno A&E.
3.00 = 250 milliseconds filter.
3.01 = 1 second filter.
3.02 = 2 seconds filter.
3.03 = 10 seconds filter.
NOTE: Each loop can be enabled indepe ndently.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 8 of 43
Page 9

2.6 Option 4, Detection Zone Tracking

Each loop input’s O pt io n 4 has nine settings ranging from 4.00 to 4.08. This option allows the user to select a detection zone tracking setting that will maintain detection zone occupancy for an extended period of time. When Option 4 is set to 4.00, the detector will track out the vehicle (i.e. drop the detection of the vehicle) in four minutes to several hours depending on the amount of inductance change (-ΔL/L) caused by the vehicle when it is detected. The greater the change caused by the vehicle, the longer the time needed to track out the vehicle. When Option 4 is set to 4.01, 4.02, 4.03,
4.04, 4.05, 4.06, 4.07, or 4. 08 the option is ON, and the detector will provide extended detection times. The default setting of Option 4 is 4.02.
There are many factors that influence t he proper setting of this option. A full explanation of how to determine the correct setting is beyond the scope of this manual. Contact Technical Support at Re no A & E regarding proper setting of this option.
NOTE: Using this option in a long loop application is not reco mmended. The detector will drop t he call prior to the vehicle totally exitin g the loop.
4.00 = Tracking Rate 0.
4.01 = Tracking Rate 1.
4.02 = Tracking Rate 2.
4.03 = Tracking Rate 3.
4.04 = Tracking Rate 4.
4.05 = Tracking Rate 5.
4.06 = Tracking Rate 6.
4.07 = Tracking Rate 7.
4.08 = Tracking Rate 8.

2.7 Option 5, Sensitivity Boost

Each loop input’s Option 5 has five settings ranging from 5.00 to 5.04. When Optio n 5 is set to 5.00, the opt ion is dis abled (OFF), and no sensitivity boost is added. When Option 5 is set to 5.01, 5. 02,
5.03, or 5.04 the option is ON, and sensitivity boost is enabled. After detection, the sensitivity is boosted to the currently selected sensitivity level plus the number of leve l(s) selec ted . The max imum sensitivity level which can be obtained with or without sensitivity boost is level 10.
5.00 = Off.
5.01 = 1 level.
5.02 = 2 levels.
5.03 = 3 levels.
5.04 = 4 levels.

2.8 Option 6, Audible Detect Signal

Each loop input’s Audible Detect Signal setting can be toggled ON or OFF by momentarily pressing either the  (UP) or  (DOWN) button. Only one loop inp ut can be turned ON at a time. Turning this option ON for one loop input automatically turns it OFF for the other loop inputs. When th is option is enabled (ON), an audible signa l will be activated whenever the detection zone for the selected loop input is occupied. The a udible s ignal indicates act ual occ upancy of t he loop de tection zo ne. This feature allows a technician to watch the detection zone and confirm correct detector operation without having to look at the detector display as well. NOT E : This opt ion is automatically disabled 15 minutes after activation or on loss of power.

2.9 Loop Check

The Loop Check feature provides a means of automatically verifying the proper operatio n of eac h loop circuit. The Loop Check Timer ca n be set for a n inter val ranging from 0 to 255 minutes in o ne-minute increments. A setting of 000 indicates that the Loop Check feature is off.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 9 of 43
Page 10

2.10 Detector ID

This feature is used in conj unction with the Synch feature (Opt ion 7) and is also use d as a means of providing each detector with a unique identification signature fo r communication purposes. Detector ID settings ca n range from 000 to 255. If, however, the detector is set to an ID value of 255, it is considered to be in an uninitialized state a nd the de tector will not save any powe r down data. W hen a detector with an ID setting of 255 is powered up, t he LCD will alternate between id and 255 until the ID setting is changed to something other than 255 and the new setting is saved in detector memory. When any detector is first initialized, its ID setting is, by default, 255.

2.11 Option 7, S ynch Feature

This option provides a means of linking together up to four (4) E-1400 detectors. Each detector’s Option 7 has three settings ranging from 7.00 to 7.02. There are two modes of Synch, Series Synch and Parallel Synch. When set to operate in Series Synch mode, one detector channel in the entire detector system is active at any given time. T his has the advantage of minimizing crosstalk between adjacent loops connected to different E-1400 detectors. When set to operate in Parallel Synch mode, all detectors scan t heir respective channel 1, 2, 3, or 4 inputs at the same time. Parallel Synch offers a faster response time than Series Synch, however the potential for crosstalk is greater.
7.00 = Off.
7.01 = Series Synch.
7.02 = Parallel Synch.
NOTE: When operating in eit her Series or Parallel S ynch mode, a ll linked detect ors must have Op tion 7 set to the same setting.

2.12 Option 8, Inductive Island Circuit Feature

The detector’s Inductive Is land Circuit option can be toggled ON or OFF by momentarily pressing either the  (UP) or  (DOWN) pushbutt on. Opt io n 8 is a detecto r wide o ption. Turning t his opt ion ON for one loop input automatically turns it ON for the ot her loop input s. NOTE: Op tion 8 is mutually exclusive with Option 9. Only one option can be enabled at a given time . If an attempt is made to turn one option ON while the ot her is enabled, a warning buzze r will sound t o provide a n alert that the other mutually exclus ive option is ON.
For proper operation, the arming loops must be connected to input channels 1 and 3 and the main channel loops must be connected to input channels 2 and 4.
The sequence of operation starts with a detection on one of the two arming loops. Once this presence detection has occurred, the remaining two or three input channels arm and Auxiliary Vital Output 1 deactivates. As the vehicle proceeds from the first ar ming loop onto the main channe l loop(s) and onto the other arming loop, pres ence detection continues and de activation of Auxiliary Vita l Output 1 is maintained. Once the vehicle has cleared the second arming loop, presence detection ends and Auxiliary Vital Output 1 reactivate s .
This feature is used in applications where it is necessary to monitor the presence of vehicles within at­grade crossings. The expected installation is two 2 0-foot arming loops located on either side of the crossing, and one or two main c ha nnel loops spanning the crossing. Contact a Technical Support representative at Reno A&E regarding proper loop configurations and spacing for specific applications.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 10 of 43
Page 11

2.13 Option 9, Inductive Shunt Enhancement Circuit Feature

The detector’s Inductive Shunt Enhancement Circuit optio n can be toggled ON or OFF b y momentar ily pressing either the  (U P) or  (DOWN) pushbutton. Option 9 is a pa ired input option. T his means that it takes two inputs to implement the feature. When this option is to ggled ON or OFF for one input, its paired input is also set to the same sta te. In the Model U-1400, input Channel 1 is paired with input Channel 2 and input Channel 3 is paired with input Channel 4. NOTE: Opt ion 9 is mutually exclus i ve w it h Option 8 . O nl y o ne o p t io n c a n be e na b le d a t a g iven time. I f a n a t tempt is ma d e to t ur n one option ON while the other is e nabled, a warning buzzer will sound to provide an alert that the other mutua lly exclusive option is ON.
When Op tio n 9 is t urne d ON, a vehicle e ntering the first loop w ill cause that input channel to e nter a pending state. As the vehicle enters the second loop while still occupying the first loop, the second input channel will enter the Ca ll state w hile the first input c hannel re mains in the pend ing state . A Cal l is neve r o ut p ut on the fi rs t i np u t c hannel w it h a d e t ec t ion. Under normal conditio ns it is not pos sible to have both outputs on at the same time. However, if one of the loops fails, both outputs will come on and stay on until the failure is corrected.
The sequence of operation starts with a detectio n on one input channel. When both of the paired input channels have detection, t he last input channel to have dete c tion will output a Call and its Primary Vital Output(s) (if mapped) a nd Auxiliary V ital Output(s) (if mapped) deac tivate until the detection for the last input ends, even if the detectio n ends for the fir st input c hannel. Pr imary Vita l Output ti ming funct io ns ( Primary V it a l Output D e la y o r P r i ma r y V i tal Output E x tension) a r e u na f fe c t e d b y t he s e tting of this feature. When detection end s for both input channels, the Call o utput will cease and the Primary Vital Output(s) (if mapped) and A uxilia r y Vital Output(s) (if mapped) will reactivate.
This feature is used in applications where it is necessary to determine t he travel direction of vehicles operating on a two-way section of railway trackage. The expected installation is two 20-foot long loops spaced approximately 25 to 28 feet apart. C ontact a Technical Support representative at Reno A&E regarding proper loop configurations and spacing for specific applications.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 11 of 43
Page 12

2.14 Primary Vital Output Delay

Each Primary Vital Output (Vita l Out) c a n be c o nd itioned w it h a Dela y t i me that is adjustable from 0.0 to 25.5 seconds in 0.1-second steps. Primary Vital Output Delay time starts counting down w he n a Call state is received from a Loop Sta tus Inp ut or V ital Inp ut (Vital In ) that is mapped to the primary Vital Output (Vital Out). When the timer reaches zero, the primary V ital Output is deactivate d. Primary Vital Output Delay does not condition the auxiliary Vital Outputs (Vital Aux).

2.15 Primary Vital Output Extension

Each Pr imary Vital O utput ( Vital Out) can be conditioned with an E xtension time that is adjustable from 0.0 to 25.5 seconds in 0.1-second steps. Pr imary Vital Output Extension ti me starts counting down when a Call state is cleared from a Loop Status I nput or Vital Input (Vital In) that is mapped to the primary Vital Outp ut (Vital Out). The primary Vital Output w ill remain OFF until the Exte nsion time reaches zero. Any vehicle entering the loop detection zone during the Exte ns ion time period causes the primary Vital Output to remain in the OFF state, and later, when the last vehicle clears the loop detection zone, the full Primary Vital Output E xt e nsion time sta rt s c o u nt in g d o w n a gain. Primary Vital Output Extension does not cond ition the auxiliary Vital Outputs (Vital Aux).

2.16 Input / Output Mapping

The Loo p Sta tus I nputs and/o r Vita l Inp uts (Vital In) can be ass igned to one or more of the primary Vital (Vital Out) and a uxiliary Vital (Vital Aux ) Outputs. Health Status can o nly be mapped to the auxiliary Vital Outputs (Vital Aux).

2.17 Full Restore To Factory Default Settings

Pressi ng a ll fo ur fro nt pa ne l p us hb utt on sw it che s si mu lta ne o us ly a nd co nt in uo us ly fo r five (5 ) s e co nds restores all factory default settings. The countdow n o f the five sec ond period is displa yed o n the LCD. Releasing any or all of the switches before the countdown ends aborts the full restore operation. (See Section 3.5 for default settings). Refer to Sect ion 5.0 for details. Caution: Restoring the factory default settings will clear all previously programmed settings including any m appe d outpu t(s).

2.18 Diagnostic F eat ur e s Acti vat io n

When the detector is operating in normal mode, pressing either the  (UP) or  (DOWN ) button will temporarily activate Option 1 (Loop Inductance Display) and Option 2 (Loop Inductance -ΔL/L Display) for all loop inputs and Optio n 6 (Audible Detect Signal) for the loop input ind icated in the lower portion of the LCD. All three of these diagnostic options will remain active for 15 minutes.

2.19 Password Protection

This detector has a password protection feature to gua rd against the accidental change of detector settings. In order to change any detector setting, a valid password must be e ntered to disable the password protection feature. Refer to Section 5.0 for details.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 12 of 43
Page 13

3.0 SPECIFICATIONS

3.1 Physical

W
EIGHT: 1.0 lb (16.0 oz) (453.6 gm).
S
IZE: 9.50 inches (24.13 cm) high x 1.50 inches (3.81 cm) wide x 7.75 inches (19.69 cm) deep
including connectors and card ejectors. O
PERATING TEMPERATURE: -40°F to +180°F (-40°C to +82°C).
C
IRCUIT BOARD: Printed circuit boards a re 0.062 inch thick FR4 material with 2 oz. copper on both
sides and plated through holes. C
ONNECTORS:
P
RIMARY INTERFACE CONNECTORS (CON1 and CON2): 34 Pin, right angle, PC Board mount.
C
OMMUNICATIONS PORT CONNECTOR (CON4): DB-9, nine pin, right angle, PC Board mount, metal
shell, D subminiature receptacle with gold plated female contacts and nuts for retaining screws.

3.2 Electrical

P
OWER: 9.5 to 18 VDC, 3.0 Amps maximum.
F
USE: 3 amp, 2AG type (Slo-Blow) type located in a fuse holder mounted on the PC Board.
P
OWER DOWN: When t he DC volt age drop s belo w the low powe r det ectio n t hresho ld (7.9 VDC), t he
detector will save all needed information to hold calls through the power down sequence. Upon power resto ration (a minim um of 10.2 V DC), the detector will use this saved data to determine the correct occupancy state of all loops.
L
OOP FEEDER LENGTH: Up to 1500 feet (450 m) maximum with proper feeder cable and appropria te
loops. L
OOP INDUCTANCE RANGE: 20 to 2500 micro Henries w ith a Q factor of 5 or greater.
L
OOP INPUTS: Transformer isolated. The minimum capacitance added by the detector is 0.068
microFarad. L
IGHTNING PROTECTION: The detector can tolerate, without damage, a 10 microFarad capacitor charged
to 2,000 volts being discharged directly i nto the loop input terminals, or a 10 microFarad capacitor charged to 2,000 volts being discharged between either loop terminal and earth ground.
R
ESET: App lication of a 30 millisecond low state (relative to +12 VDC ) to Pin DD o f the lower
primary interface connector (CON2) resets all loops. Each loop input channel ca n be manually reset by selecting the loop input channel then pressing and holding the front pa nel CHAN button for three seconds, or by changing the sensitivity or loop frequency of the loop input channel.
S
YNCHRONIZE SCAN CONTROL: The SYNC H IN input and SYNCH OUT outpu t allow the loops from
different detectors to be scanned sequentially, either in series or parallel. This feature can be used to eliminate crosstalk between adjacent loops connected to different E-1400 detectors. NOTE: Response times will vary depending on the number of detectors (up to a maximum of four) on the SYNCH Line. (See Section 3.4.)
V
ITAL OUTPUTS: The primary and auxiliary Vital Outputs are transformer isolated and are capable of
sourcing 200 mA at 12 VDC. Opto-isola ted circuitry verifies t hat the Vital Output is in the desired state. If any Vital Output fails its self-chec k, all of the detector’s Vital Outputs will be dea ctivated.
V
ITAL INPUTS: 12 volt inputs isolated from the detector. Opto -isolated circ uitry verifies that t he Vital
Inputs are operational. C
HECK LOOP RELAYS: Co ntacts rated for 6.0 Amps maximum, 150 VDC, 300 VAC, and 500 V A
maximum switched power.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 13 of 43
Page 14

3.3 Operational

L
OOP STATUS INDICATORS: Each loop input has a super high intensity red light emitting diode (LED) to
indicate a CALL (vehicle detected over loop) or failed loop. I
NPUT / OUTPUT INDICATORS: The detector has 12 dual color (green / red) LEDs to indicate the
following:
Vital In LED Status Meaning
OFF Vital Input is deactivated (low voltage level). GREEN ON Vital Input is activated (high voltage level). RED ON Vital Input has failed its self-check (Vital Inputs, Primary Vital Outputs, and Auxiliary Vital Outputs revert to OFF state).
Vital Out LED Status Meaning
OFF Primary Vital Output is deactivated (low voltage level). GREEN ON Primary Vital Output is activated (high voltage level). GREEN 2Hz FLASHING Timing Delay activated and Primary Vital Output is working (high voltage
GREEN 5Hz FLASHING Timing Extension activated and Primary Vital Output is working (low
RED ON Primary Vital Output has failed its self-check (Vital Inputs, Primary Vital
Vital Aux LED Status Meaning
OFF Auxiliary Vital Output is deactivated (low voltage level). GREEN ON Auxiliary Vital Output is activated (high voltage level). RED ON Auxiliary Vital Output has failed its self-check (Vital Inputs, Primary Vital
level until delay times out).
voltage level until extension times out).
Outputs, and Auxiliary Vital Outputs revert to OFF state).
Outputs, and Auxiliary Vital Outputs revert to OFF state).
RESPONSE TIME: See Section 3.4 for actual response times. S
ELF-TUNING: Each loop circuit tunes and is operational within two seco nds after app licatio n of power
or after being reset. Full sensitivity and hold time require 30 seconds of operation. E
NVIRONMENTAL & TRACKING: The detector is fully se lf-compensating for enviro nmental c hanges and
loop drift over the full temperature range and the entire loop inductance range. G
ROUNDED LOOP OPERATION: The loop isolation transformer allows ope ration with poor qua lity loops
(which may include one short to ground at a single point). L
OOP (FAIL) MONITOR: If the to ta l in d ucta nc e o f a c ha nne l’ s loop input network goes out of the range
specified for the detector, or rapidly changes by more than ±25%, and the Loop and/or Health status outputs are mapped, the Loop a nd/or Health status o utputs will immediately enter Fail-Safe mode and the LCD wil l disp lay LOOP FAI L. The type of loop failure will also be displayed as L lo (for -25% change or shorted loop conditions) or L hi (for +25% change o r open loop conditions). This will continue as long as the loop fault exists. At t he time of a loop failure, the Loop stat us LED will begin to flash at a rate of three flashes per second. The LED will continue this display pattern until the loop input channel is manually reset or power is removed. If the loop self-hea ls, t he LO OP FAIL message on the LCD will extinguish and the loop input cha nnel will resume operation in a normal manner; except the Loop status LED will continue the three flashes per second display pattern, thus providing an alert that a prior Loop Fail condition has occurred. Eac h loop failure for the loop input cha nnel is counted and accumulated into the Loop Fail Memory. The total number of loop failures written into t he Loop Fail Memory (since the last powe r interruption or manual reset) ca n be seen by stepping through the loop input channel’s detector functions in Program Mode to t he LOOP F AIL mes sage . W hen t he status of one or more loops is ass igned to a Vita l Outp ut, a loop failu re will de act ivate the Vita l Outp ut and the c orrespo nding Vital Out LED will turn from green to OFF to indicate the failure. Whe n a Health Status is assigned to a n auxiliary Vital O utput, a loop failure will deact ivate the auxiliar y Vital Output and the corresponding Vital Aux LE D will turn off.
L
OOP CHECK: This feature is used as a means of verifying proper operation of each inductive loop
connected to a detector. In a typical installation, a separate loop of wire called a check loop will be wound around the perimeter of each inductive loop. T he Loop C heck T imer can be programmed with a value ranging from 1 to 255 minutes. Once this time interval has elapsed, the detector will automatically short the check loop simulating vehicle detection in the inductive loop. If this initial loop check test fails, the detector will enter a verification mode to make certain that the loop failure is indeed due to a loop related problem. While operating in this verification mode, the detector will perfor m a series of ten verification tests and the LCD will flash Pcl. If the ten loop check ver ification tests all fail,
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 14 of 43
Page 15
the Vital Output(s) to which the Loop Status Input(s) and/or Health Status Input(s) is (are) mapped will be deactivated. The loop check test will be repeated at regular intervals depending on the value programmed into the Loop Check Timer. The Vital Output(s) to which the Loop S ta t us I nput(s) a nd / o r Health Status Input(s) is (are) mapped will not be activated until the problem with the check loop is corrected and the channel reset or until the Loop Check function is turned off. During a check loop failure the LCD will flash Lcl. NOTE: If the Loop Check Timer is set to 0 minutes, the Loop C heck function is off.
S
YNCH MODE: The Synch feature provides a means of linking up to four E-1400 detectors together.
There are two modes in which the detectors can be configured to operate; Series Synch or Parallel Synch.
Series Synch mode can be used to eliminate the possibility of cross talk between adjacent loops. Series Synch will activate one loop at a time, cycling through all loops configured to de tect in the Series Synch mod e. W he n the detector system is configured to operate in Series Synch mo de, it is necessary for one detector to be identified as the master unit, i.e. the detector that starts the Synch pulse. This master unit must be assigned a Detector ID value of 000. The scanning sequence of detector loop input channels is Detector ID 000 c hanne l 1, D etector ID 000 channel 2, Detector ID 000 channel 3, Detector ID 000 channel 4, second detector channe l 1, second detector channel 2, second detector channe l 3, second detector channel 4, and so on until the synch pulse returns to the first detector. The Synch-out of the first (master) unit is connected to the Synch-in of the second de tector; the Synch-out of the second detector is connected to the Synch-in of the third detector. This connection scheme is continued until the Synch-out of the last detector is connected to the Synch in of the first (master) detector.
When set to operate in Parallel Synch mode, all detectors configured to operate in Parallel Synch mode start scanning their respective channel 1 loop input s at the sa me t ime a nd the n pr oce ed to seq ue ntia ll y scan their respective channel 2, 3, and 4 loop inputs. When the detector system is configured to operate in Parallel Synch mode, it is necessar y for one detector to be identified as the master unit. This master unit mus t be assigned a Detector ID value of 000. Connections between detectors differ from those in Series Synch mo de in tha t the Sync h-out of first detector is connected to the Synch-in of the other detectors and the Synch-in of the first detector is connected to the Synch-out of the other detectors.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 15 of 43
Page 16
Operation in either Synch mode require s that the Rese t inp uts ( pin DD of the lower primary interface connector) of all detectors be tied together, the DC Common and S ync h C o mmo n inputs (pins BB and Z of the upper primary interface connector) of all detectors be tied together, a nd the Synch 12 VDC inputs (pin V of the upper primary interface connector) be tied to 12 VDC +. To sta rt operation in either Synch mode, the detectors must be reset using the RES ET pin ( pin DD o f the lower primary interface connector).
When ope ra tin g in Ser ies S ync h mo de, if a ny de tec tor fai ls to Sy nch, al l of t he V ita l Out put s o f ever y detector configured to operate in Synch mode will be deactivated. If t he detector that has failed to Synch has the Synch feature turned off, the other detectors configured to operate in Synch mode will continue to operate in Series Synch mode.
When operating in Parallel Synch mode, if the master unit fails, all of the Vital Outp uts of every detector configured to operate in Synch mode will be deactivated. If one or more of t he detectors connected to the master unit Fails, only those Vita l Outputs of the detecto r(s) which has (have ) failed will be deactivated.
When ope rating in either Series or Parall el mode, FSY will be d isplayed on the LCD screen o f the detector(s) associated with the failure. This display state will continue until the problem is corrected and the detectors are powered down and repowered, reset, or until the Synch function is turned off.
In either mode of operation, the S ynch feature will continue to function if one or more detectors has (have) been disabled.
3.4 TABLE: Sensitivity, -∆L/L, & Response Time

3.5 TABLES: Default Settings CHANNEL & DETECTOR SETTINGS:

889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 16 of 43
Page 17
VITAL OUTPUT SETTINGS:

3.6 TABLES: Pin Assignments

PRIMARY INTERFACE CONNECTORS S
EQUENTIAL PIN ASSIGNMENTS - UPPER CONNECTOR (CON1)
UNCTIONAL PIN ASSIGNMENTS - UPPER CONNECTOR (CON1)
F
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 17 of 43
Page 18
S
EQUENTIAL PIN ASSIGNMENTS - LOWER CONNECTOR (CON2)
F
UNCTIONAL PIN ASSIGNMENTS - LOWER CONNECTOR (CON2)
COMM PORT (DB-9 Communic a ti o n Interface Connec tor)
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 18 of 43
Page 19
E-1400
-L/L =
FREQ
LOOP FAIL
OPTION
MAX PRESENCE
OF
EOG
SCANNIN
EXTENSION
SENSITIVITY
DELAY
PULSE
SECONDS
3 2 1
4
ON
1 2 3 4
CHAN
FUNC

4.0 USER INTERFACE

Loop Input LED Indicators
Seven Segment Display
Numbered Loop Symbols
UP Pushbutton
Channel Select Pushbutton
Function Select Pushbutton
VITAL AUX LED Indicators VITAL IN LED Indicators
VITAL OUT LED Indicators Parameter Description
Bargraph
DOWN Pushbutton
DB-9 RS-232 Connector
Vehicle Signal Strength
Frequency Setting
Output Mapping
Parameter Values
Frequency Reading
Inductance Value
• -∆L/L Value
Output Mapping
Type of Loop Failures:
L hi = Open Loop
L lo = Shorted Loop
Change Displayed Loop Input
Channel - Press and Release
Exit Program Mode - Press and
Hold for 1 second
RESET Channel - Press and
Hold for 3 seconds
Change Output Mapping
Parameter - Press and Release
Increments Values
Toggles ON and OFF
Activates Detector Diagnostic
Features
Input Status - On / Off
Input Failed Self-check
Name of Parameter
Timer in Operation
Loop Input Channel Displayed
Flashing = Program Mode
Call
Loop Failure
Loop Fault has Occurred
Press momentarily to enter
Program Mode and to step through parameters
Decrements Values
Toggles ON and OFF
Activates Detector Diagnostic
Features
Output Status - On / Off
Delay Time Activated
Extend Time Activated
Output Failed Self-check
NOTES: The LCD backlighting illuminates whenever any pushbutton is pressed. Backlighting will extinguish 15 minutes after the last pushbutt on press .
Pressing any two or three pushbuttons illuminates the LCD backlighting, displays all LCD segments, and illuminates the Vital In, Vital Out, and Vital A ux Output LEDs. The Output LEDs are illuminated in their red phase as a means of confirming proper function of the detector’s Vital Output failure feature.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 19 of 43
Output Status - On / Off
Output Failed Self-check
Communications Port
Page 20

5.0 INSTALLATION AND SET-UP

The Model E-1400 detector has no DIP switches or jumpers to co nfigure. In order for the detector to operate, an external EEPROM memory module MUST be connected to the EGMS rack. Plug the detector into an appropriately wired rack and app ly power. If the detecto r is not new fro m the factory, it may be advantageous to restore a ll factory default se ttings to avoid having to check every setting for every loop input channel. To restore all factory default settings, disable the password protection feature as explained below. Press and hold all fou r pushbutt on switches simultaneo usly for five seconds . When all four pushbuttons are depressed and held, the display will start counting down from five (5). When the countdown reaches zero (0), releasing the pushbuttons restores the factory default settings. Releas ing any or all of the switches before the countdown ends aborts the full restore ope rat ion. Caution: Restoring the factory default settings will clear all previously programm ed settings including any m appe d output(s ).
All operating parameters can be adjusted from the front panel. The detector continues to operate normally while it is in the Program Mode. The va lue currently displayed is always the actual va lue being used. Example: If you are changing the delay time, the time displayed at the instant that a vehicle entered the detection zone for that loop input channel would be the value used for the delay timer.
Pressing the FUNC pushbutton enters the Program Mode. The FUNC pushbutton has an auto repeat function. This allows quick navigatio n to the desired parameter. The FUNC p ushbutto n only moves forward through all of the parameters. There is no way to move backwards through the parameters.
While viewing any parameter, pressing the CHAN pushbutt o n w i ll displa y the same paramet e r i n t he next loop input channel. The currently selected channel is indicated at the bottom of the LCD display. Pressi ng a nd ho ldi ng the CHAN pushbutton for one second w ill exit the Progra m Mode and retur n to the Normal Mode.
Pressing and holding either the  (UP) or  (DOWN) pushbutton will caus e the va lue to c ha nge rap id l y until the pushbutton is releas ed.
Disabling Password Protection
To change any detector setting, the password protection feature must be disabled. When viewing a setting in the Program Mode, press the  (UP) or  (DOWN) pushbutton to enter password entry mode. The LCD will display PSd to provide an indication that a valid pass word must be e ntered be fore any detect or setting can be changed. NOTE: Pressing the CHAN button for two seconds at any time will exit password entry mode. Press the (UP) or (DO WN) p ushbut ton. The LC D dis play w ill c hange to 000 wit h the left most 0 flashing. Press the (DOW N) pus hbutto n to change the flas hing d igit to 9. Press the CHAN pushb utto n a sec ond t ime to get the cente r 0 to flas h. Press the  (U P) pus hbut ton t o change the flas hing digit to 1. Press the CHAN pushbutton a third time to get the rightmost 0 to flash. Press the  (UP) pushb utton to change the flashing digit to 1. Press t he FUNC pushbutton to enter the password. Detector settings can now be changed. Press the FUNC pushbu tton to e nter the passwor d. Detector settings can now be changed. Password protection can be re-enabled by pres sing and holding the CHAN p ushbutton for two seconds followe d by pressing the (UP) pushbutton twice. If no act ion is taken to re-enable the password protection feature, it will be automatically re-enabled 15 minutes after changing any detector setting.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 20 of 43
Page 21

5.1 Initial Installation Start-up Mode Display Screens

1
©
1
©
LOOP FAIL
1
LOOP FAIL
1
Upon application of power, the LCD scre en will show the detector model letter and firmware version, in this case, Model E-1400 fir mw a re ve rsion 3. This screen will be displayed for two seconds.
The ne xt LCD scree ns displayed w ill show the Detector ID setting. The display will toggle between the two screens shown with a 50% duty cycle. When shipped from the factory, the Detector ID is set to 255 and the detector is considered to be uninitialized. This will be the case on initial start-up. Before proceeding, the detector ID must be changed to something other than 255. To do s o, use the  (UP) or  (DOWN) button to change the sett ing to t he desired ID number. Whe n set to the correct value, press the FUNC button to accept and store the setting. If the detector in question is being used as the master unit when ut iliz ing t he Sy nch fe atur e, t he Det ect or must be as signed an ID value of 000.
The detector will ne xt co nfirm t he state o f the check loops c onnecte d to it. I f all four check loops are connected and are intact, the LCD screen will display an indication of the successful completion of the chec k loop verification. This screen will flash twice
.
If all four chec k loop s are not connec ted o r one (or more) is (are) d amaged in so me way, the detector will initiate its check loop verification process. The LCD screen will d ispla y a n ind ica tio n of a pe ndi ng st at e dur ing t he c hec k lo op ver ific atio n. T his screen will flash for 55 seconds until one of three things happen; the check loop verification fails, the check loop(s) causing the check loop verification is (are) connected or restored to a functional state, or the check loop(s) causing the check loop verification is (are) turned off, i.e. the Loop Check Timer of the loop(s) is set to 000.
If the check loop verification process continues for the entire 55 second period and no action has been take n to connect or restore the check loops causin g the check loop verificat ion, the LCD screen will toggle between the two screens shown with a 50% duty cycle.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 21 of 43
Page 22
If the detector passes its check loop verification process or a pe nding loop check failure state is resolved in one of the ways mentioned above, the LCD screen will show a normal (idle) condition.

5.2 Program Mode Display Screens

PARAMETER .................... Password Protection Screen.
S
ETTING .......................... 911 disables the Password Protection.
ETTING DISPLAYED ....... PSd.
S 7-
SEGMENT DISPLAY ....... Press the (UP) or (DOWN) pushbutto n to display 000 to
EFAULT SETTINGS ........ Password Protection feature is active.
D
XAMPLE ........................ Password entry mode is available for access.
E N
OTES ............................ Pressing the CHAN button for two seconds at any time will exit
ARAMETER .................... Frequency.
P S
ETTINGS ........................ Eight selections - 1 to 8.
ETTING DISPLAYED ....... Bargraph indicates settings from 1 (left) to 8 (right).
S 7-
SEGMENT DISPLAY ....... Actual frequency of the loop circuit. Typically 20.0 to 99.9
EFAULT SETTINGS ........ Loop Input 1 = 2, Loop Input 2 = 4, Loop Input 3 = 6, and Loop
D
XAMPLE ........................ Frequency setting 4 is selected for loop input. The loop frequency
E
OTES ............................ Changing the frequency will reset the Loop Input. An unstable
N
P
ARAMETER .................... Sensitivity.
ETTINGS ........................ 12 selections - 1 to 10, OFF, or CALL.
S S
ETTING DISPLAYED ....... 7-segment display will show currently selected setting.
SEGMENT DISPLAY ....... Currently selected Sensitivity.
7­D
EFAULT SETTING .......... 6 for all loop inputs.
XAMPLE ........................ Sensitivity 6 is selected for loop input 1.
E N
OTES ............................ Changing the sensiti vity will reset the loop input. The bar graph
ARAMETER .................... Option 1 (Loop/Lead-In Inductance Display).
P
ETTINGS ........................ ON or OFF.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7­D
EFAULT SETTING .......... OFF for all loop inputs.
XAMPLE ........................ Option 1 is turned ON for all loop inputs.
E
NOTES ............................ This option is a detector-wide setting. Changing it for one loop
ARAMETER .................... Option 2 (Percentage of Inductance change, -ΔL/L).
P
ETTINGS ........................ ON or OFF.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7­D
EFAULT SETTING .......... OFF for all loop inputs.
XAMPLE ........................ Option 2 is turned OFF for all loop inputs.
E N
OTES ............................ This option is a detector-wide setting. Changing it for one loop
access password entry mode.
password entry mode.
kiloHertz.
Input 4 = 8.
is 34.9 kHz.
frequency display varying more than ±0.2 kiloHertz may indicate loop crosstalk or other interference.
will show strength of vehicle calls so that correct sensitivity can be verified from this screen.
input changes it for all loop inputs. This option will automatically turn off in 15 minutes.
input changes it for all loop inputs. This option will automatically turn off in 15 minutes.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 22 of 43
Page 23
PARAMETER .................... Option 3 (Noise Filter).
ETTINGS ........................ Four selections - 3.00 to 3.03.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7­D
EFAULT SETTING .......... 3.00 (OFF) for all loop inputs.
XAMPLE ........................ Option 3.02 is turned ON for loop input 1.
E N
OTES ............................ This option is a channel specific setting. Changing it for one loop
ARAMETER .................... Option 4 (Detection Zone Tracking).
P
ETTINGS ........................ Nine selections - 4.00 to 4.08.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7­D
EFAULT SETTING .......... 4.02 (ON) for all loop inputs.
XAMPLE ........................ Option 4.03 is turned ON for loop input 1.
E N
OTES ............................ This option is a loo p input specific setting. Changing it for one
ARAMETER .................... Option 5 (Sensitivity Boost).
P S
ETTINGS ........................ Five selections - 5.00 to 5.04.
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
S 7-
SEGMENT DISPLAY ....... The number of this option.
EFAULT SETTING .......... 5.00 (OFF) for all loop inputs.
D E
XAMPLE ........................ Option 5.04 is turned ON for loop input 1
OTES ............................ This option is a loo p input specific setting. Changing it for one
N
P
ARAMETER .................... Option 6 (Audible Detect).
ETTINGS ........................ ON or OFF.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7-
EFAULT SETTING .......... OFF for all loop inputs.
D E
XAMPLE ........................ Option 6 is turned ON for loop input 1.
OTES ............................ This option is mutually exclusive with the same option in any
N
P
ARAMETER .................... Loop Check Timer setting.
ETTINGS ........................ 256 selections - 000 to 255 minutes in 1 minute steps.
S S
ETTING DISPLAYED ....... 7-segment display will show currently selected setting.
7-SEGMENT DISPLAY ....... Currently selected time in minutes.
D
EFAULT SETTING .......... Five minutes (005) for all loop inputs.
XAMPLE ........................ The Loop Check Timer for Loop 1 is set for 25 minutes.
E N
OTES ............................ This option is a loo p input specific setting. Changing it for one
P S S
7­D
E N
input changes it for that loop input only. It is strongly recommended that this option be set to 3.00 (OFF) for normal operation.
loop input changes it for that loop input only.
loop input changes it for that loop input only.
other loop input. Turning it on for one loop input turns it off for all other loop inputs. This option will automatically turn off after 15 minutes.
loop input changes it for that loop input only.
ARAMETER ................... Detector ID setting.
ETTINGS ....................... 256 selections - 000 to 255.
ETTING DISPLAYED ....... 7-segment display will show currently selected
SEGMENT DISPLAY ...... Current Detector ID setting. The displa y will toggle
EFAULT SETTING .......... 255.
XAMPLE ........................ Detector ID is set as 158.
OTES ............................ When set to Detector ID number 255, the dete ctor is
setting.
between the two screens with a 50% duty cycle.
considered uninitialized.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 23 of 43
Page 24
PARAMETER .................... Option 7 (Synch Feature).
ETTINGS ........................ Three selections - 7.00 to 7.02.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7­D
EFAULT SETTING .......... 7.00 (OFF) for all loop inputs.
XAMPLE ........................ Option 7 is turned OFF.
E N
OTES ............................ This option is a detector-wide setting. Setting Option 7 to 7.01
ARAMETER .................... Option 8 (Inductive Island Circuit).
P
ETTINGS ........................ ON or OFF.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7­D
EFAULT SETTING .......... OFF for all loop inputs.
XAMPLE ........................ Option 8 is turned OFF for all channels.
E N
OTES ............................ This option is a detector wide setting. Changing it for one channel
ARAMETER .................... Option 9 (Inductive Shunt Enhancement Circuit).
P
ETTINGS ........................ ON or OFF.
S S
ETTING DISPLAYED ....... The word ON or OFF will be displayed.
SEGMENT DISPLAY ....... The number of this option.
7­D
EFAULT SETTING .......... OFF for all loop inputs.
XAMPLE ........................ Option 9 is turned OFF for channels 1 and 2.
E N
OTES ............................ This is a paired channel option. Channel 1 is paired with channel
ARAMETER .................... Primary Vital Output Delay.
P
ETTINGS ........................ 256 selections - 00.0 to 25.5 Seconds in 0.1 second steps.
S S
ETTING DISPLAYED ....... 7-segment display will show currently selected setting. Loop
SEGMENT DISPLAY ....... Currently selected Delay time in seconds.
7-
EFAULT SETTING .......... 00.0 seconds.
D E
XAMPLE ........................ Delay of 10 seconds selected for Primary Vital Output 1.
OTES ............................ If the loop input’s detection zo ne is occupied when this parameter
N
.....................................When viewing and/or setting the Primary Vital Output Delay time
P
ARAMETER .................... Primary Vital Output Extension.
ETTINGS ........................ 256 selections - 00.0 to 25.5 Seconds in 0.1 second steps.
S
SETTING DISPLAYED ....... 7-segment display will show currently selected setting. . Loop
SEGMENT DISPLAY ....... Currently selected Extension time in seconds.
7-
EFAULT SETTING .......... 00.0 seconds.
D E
XAMPLE ........................ Extension of 2.5 seconds selected for Primary Vital Output 1.
OTES ............................ If the loop input’s detection zo ne is occupied when this parameter
N
....................................... When viewing and/or setting the Primary Vital Output Extension
activates Series Synch (SYS), setting Option 7 to 7.02 activates Parallel Synch (SYP).
changes it for all channels. Turning ON Option 8 automatically turns OFF Option 9.
2 and Channel 3 is paired with channel 4. Changing the setting for one channel also changes the setting for the paired channel. Turning ON Option 9 automatically turns OFF Option 8.
symbol indicates which Primary Vital Output the Delay time applies to.
is changed, it will not take effect until the detection zone is empty or the loop input is reset.
settings, use the CHAN button to advance to the Delay time settings of the next Primary Vital Output.
symbol indicates which Primary Vital Output the Extension time applies to.
is changed, it will not take effect until the detection zone is empty or the loop input is reset.
time settings, use the CHAN button to advan ce to the Extension time settings of the next Primary Vital Output.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 24 of 43
Page 25

5.2.1. Primary Vital Output (Vital Out) Mapping Screens P

ARAMETER .................Primary Vital Output (Vital Out) Mapping.
S
ETTINGS .....................256 selections - any and all combinations of the four
Loop Status Inputs and the four Vital Inputs (Vital In).
S
ETTING DISPLAYED ....Bargraph indicates Loop Sta tus Inputs and/or Vital
Inputs mapped to the Primary Vital Outputs. In Loop Status Input mode, the leftmost segment = Loop St atus Inp ut 4, next s egment = Lo op Status Input 3 , next segment = Loop Status Input 2 , next segment = Loop Status Input 1. In Vital Input mode, the center four segments are used. The leftmost center segment = Vital Input 4, next segme nt = Vita l Input 3, next segment = Vita l Inp ut 2, next segme nt = Vital Input 1. T he loop symbol indicates which Primary Vital Output the Loop Status Inputs and/or Vital I nputs a re bein g mapped to.
7-
SEGMENT DISPLAY ....7-segment display will show currently selected
setting. The first character cor responds to the Loop Status Inputs mappe d to the Primary Vital Outputs , the second character corresponds to the Vital I nputs mapped to the Primary Vital Out puts.
D
EFAULT SETTINGS .....Primary Vital Output 1 = 00. No Loop Status
Inputs mapped to Pr imary Vita l Output 1. No V ital Inputs mapped to Primary Vital O utput 1.
.....................................Primary Vital Output 2 = 00. No Loop Status
Inputs mapped to Pr imary Vita l Output 2. No V ital Inputs mapped to Primary Vital O utput 2.
.....................................Primary Vital Output 3 = 00. No Loop Status
Inputs mapped to Pr imary Vita l Output 3. No V ital Inputs mapped to Primary Vital O utput 3.
.....................................Primary Vital Output 4 = 00. No Loop Status
Inputs mapped to Primary Vital Outp ut 4. No Vital Inputs mapped to Primary Vital O utput 4.
E
XAMPLE ......................Loop Status Inputs 1, 2, & 3 ma pped to Primary
Vital Output 1. No Vital Inputs mapped to Pr imary Vital Output 1.
N
OTES ..........................The characters are displayed in Hexadecimal
notation (Base 16).
.....................................When viewing and/or setting the Primary Vital
Output mapping settings, use the FUNC button to advance to the mapping settings of the next Primary Vital Output.
.....................................A Primary Vital Output is deactivated if no Loop
Status Inputs or Vita l Inputs are mapped to it.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 25 of 43
Page 26
To set the mapping status of the Loop Status Inputs, repeatedly press the FUNC button to enter Program Mode and toggle through the set-up paramete rs until the display screen shows the bargraph and two of the seve n-segment display characters. The left character and a loop symbol will flash indicating the current setting and t he four leftmost bargraph segments will show the c orresponding Loop Status Input mapping. Use t he  (UP) or  (DOWN) button to change the setting to the desired config uration. Us e the FUNC b utton to ad vance to the next Primary Vital Output (indicated by a flashing numbered loop symbol at the bottom of the LCD screen). The various combinations o f displa y characters and bargraph display are shown below.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 26 of 43
Page 27
To set the mapping status of the V ital Inputs when in t he Prima ry Vital Output Mapping Program Mode, press the CHAN butto n once. The right charac ter and a loop symbol w ill flash indicating the current setting and the center four bargraph segments will show the corresponding Vital Inputs mapping. Use the  (UP) or  (D OWN) b utton to c hange the setting to the desired configuration. Use the FUNC butt on to adva nce to the next Primary Vita l Outp ut (indicated by a flashing numbered loop symbol at the bottom of the LCD screen). The various combinations of display characters and bargra ph disp lay are s hown belo w.
When in Normal Mode, repeatedly press the FUNC b utton to enter Program Mode and toggle thro ugh the set-up parameters until the display screen shows the bargraph and two of the seven-segment d is play characters. The left c haracter and a loop symbol will flash indicating t he current se tting and the four leftmost bargraph segments will show t he corresponding Loop Stat us I nput mapping. Press the C HAN button once. The right characte r and a loop symbol will flash indicating the current setting and the center four ba rgra ph s eg ment s w ill s how t he c orr esp ond ing Vit al I np uts mapping. U se the  (UP ) or (DOWN) button to change the setting to the desired configuration. Use the FUNC button to advance to the ne xt Pr imar y Vita l Output (indicated by a flashing numbered loop symbol at the bottom of the LCD screen). The various combinations of display characters and bargraph display are shown below.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 27 of 43
Page 28

5.2.2 Auxiliary Vital Output (Vital Aux) Mapping Screens P

ARAMETER .................Auxiliary Vital Output (Vital Aux ) Mapping.
S
ETTINGS .....................4096 selections - any and all combinations of the four Loop
Status I np u t s , fo ur V ita l I np ut s (Vital In), and the four Hea lt h Status Inputs.
S
ETTING DISPLAYED ....Bargraph indicates Loop Status Inputs, Vital Inputs, and/or
Health Status Inputs mapped to the Auxiliary Vital Outputs. In Loop Status Input mode, the leftmost segment = Loop Status Input 4, next segment = Loop Status Input 3, next segme nt = Lo op Sta tus I nput 2, next segment = Loop Status Input 1. In Vital Input mode, the center four segments are used. The leftmost center segment = Vital Input 4, next segme nt = V ital Inpu t 3, next segment = Vital Inp ut 2, next segme nt = Vital I nput 1. In Health Sta tus Input mode, the rightmost segment = Health Stat us Input 1, next segment = Loop Status Input 2, next seg ment = Loop Status Input 3, next segment = Loop Status Input 4. T he loop symbol indicates which Auxiliary Vital Output the Loop St atus Inputs, Vital Inputs, a nd/or Hea lth Status Inp uts are bein g mapped to.
7-
SEGMENT DISPLAY ....7-segment disp lay will show currently sele cted setting. The
first character corresponds to the Loop Status I nputs mapped to the Auxiliary Vital Outputs, the second character corresponds to the Vital Inputs mapped to Auxiliary Vital Outputs, and the third character corresponds to the Health Status Inputs mapped to the Auxiliary Vita l Outputs.
D
EFAULT SETTINGS .....Auxiliary Vital Output 1 = 000. No Loop Status Inputs
mapped to Auxiliary Vital O utput 1 . No V ital I nputs mapped to Auxiliary Vital Output 1. Health Status Input 1 not mapped to Auxiliary Vital Output 1.
.....................................Auxiliary Vital Output 2= 000. No Loop Status Inputs
mapped to Auxiliary Vital Output 2. No V ital I nputs mapped to Auxiliary Vital Output 2. Health Status Input 2 not mapped to Auxiliary Vital Output 2.
.....................................Auxiliary Vital Output 3 = 000. No Loop Status Inputs
mapped to Auxiliary Vital O utpu t 3, No V ital I nputs mapped to Auxiliary Vital Output 3. Health Status Input 3 not mapped to Auxiliary Vital Output 3.
.....................................Auxiliary Vital Output 4 = 000. No Loop Status Inputs
mapped to Auxiliary Vital O utpu t 4. No V ital I nputs mapped to Auxiliary Vital Output 4. Health Status Input 4 not mapped to Auxiliary Vital Output 4.
E
XAMPLE ......................Loop Status Inputs 1, 2, & 3 mapped to Auxiliary Vital
Output 1. No Vital Inputs mapped to Auxiliary Vital Output
1. Health Status Inputs 1, 2, 3, & 4 mapped to Auxiliary Vital Output 1.
N
OTES ..........................The characters are displayed in Hexadecimal notation (Base
16).
.....................................When viewing and/or setting the Auxiliary Vital Output
mapping settings, use the FUNC button to advance to the mappi ng setti ngs of the next Auxiliary Vita l Output.
.....................................An Auxiliary Vital Output is deactivated if no Loop Status
Inputs, Vita l Inputs, or Health Status Inp uts are mapped to it.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 28 of 43
Page 29
To set the mapping status of the Loop Status Inputs, repeatedly press the FUNC button to enter Program Mode and toggle through the set-up pa rameters until the display scree n shows the bargraph and thr ee of the seven-segme nt display characters. The left charac ter and a loop symbol will flash indicating the current setting and t he four leftmost bargraph segments will show the c orresponding Loop Status Input mapping. Use the  (UP) o r  (DOWN) button to change the setting to the desired config uration. Use the FUN C button t o advance to the next Auxiliary Vital Output (indicated by a flashing numbered loop symbol at the bottom of the LCD screen). The various combinations o f displa y characters and bargraph display are shown below.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 29 of 43
Page 30
To set the mapping status of t he Vital Inputs whe n in the Auxiliar y Vital Output Mapping Program Mode, press the CHAN button once. The center character and a loop symbol will flash indicating the current setting and the center four bargraph segments will show the corresponding Vital Inputs mapping. Us e t he  (UP) or  (DOWN ) b utton to change the setting to the des ired configuration. Use the FUNC button to advance to the next Auxiliary Vital Output (indicated by a flashing numbered loop symbol at the bottom of the LCD screen). The various combinations of display characters and bargraph display are shown below.
When in Normal Mode, repeatedly press the FUNC button to enter Program Mode and toggle through the set-up parameters until the display screen shows the bar graph and three of the seven­segment display characters. The left char acter and a loop symbol will flash indicating the current setting and the four leftmost bargraph seg ments will show the corresponding Loop Status Input mapping. Press the CHAN button once. The center character and a loop symbol will flash indicating the current setting and the center four bargrap h segments will show the corresponding Vital Inpu ts mapping. Us e t he  (UP) or  (D OW N ) b utt o n to cha nge the setting to the desired configurat ion. Use the FUNC button to advance to the next Auxiliary Vital Output (indicated by a flashing numbered loop symbol at the bottom of the LCD screen). The various combinations of display characters and bargraph display are shown below.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 30 of 43
Page 31
To set t he mapp ing sta tus o f the He alth St atus I nputs when i n the Auxiliar y Vital Output Mapping Program Mode, press the CHAN button once again. The right character and a loop symbol will flash indicating the current setting and the four right most bargraph se gments will s how the correspo nding Health Status Input mapping. Use the  (UP) or  (DOWN) button to change t he sett ing to t he desired configuration. Use the FUNC button to advance to the next Auxiliary Vital Output (indicated by a flashing numbered loop symbol at the bottom of the LCD screen). The various combinations of display characters and bargraph display are shown below.
When in Normal Mode, repeatedly press the FUNC button to enter Program Mode and toggle through the set-up parameters until the display screen shows the bar graph and three of the seven­segme nt dis play c haracters. The left character and a loop symbol will flash indicating the current setting and the four leftmost bargraph segments will show the correspo nding Vital I nput mapping. Press the CHAN button twice. The right character and a loop symbol will flash indicating t he curre nt se tti ng a nd t he fo ur r igh tmos t ba rg rap h se gme nts w ill s how the cor res pond in g He alt h Sta tus Input mapping. Use the  (UP) or  (DOWN) button to change the setting to the desired config urat ion. Use t he FUN C but ton t o ad vance to the next Auxiliary Vita l O utput (indicated by a flashing numbered loop symbol at t he bottom of the LCD screen). The various combinatio ns of display charac ters and bargraph disp lay are s hown below.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 31 of 43
Page 32
5.2 Program Mode Display Screens (cont.)
LOOP FAIL
L =
1
PARAMETER .................... Loop Fail.
ETTINGS ........................ Pressing the (UP) or (DOWN) button will clear the Loop
S S
ETTING DISPLAYED ....... View only.
SEGMENT DISPLAY ....... Loop Failures since the last time it was cleared manually or due to
7-
EFAULT SETTING .......... 000.
D E
XAMPLE ........................ There are eight Loop Failures in the accumulator for channel 1.
OTES ............................ Count will be reset to 0 after loss of power, pressing the (UP)
N

5.3 Normal Mode Display Screens

TATE Idle.
S B
ARGRAPH DISPLAY Off.
SEGMENT DISPLAY Three Dashes.
7­L
OOP INPUT LED OFF. XAMPLE Loop Input Channel 1 is idle.
E
OTES This is the normal state for a loop input channel when the loop
N
S
TATE Call. ARGRAPH DISPLAY Number of sensitivity levels inductance change from vehicle
B
SEGMENT DISPLAY Call.
7-
OOP INPUT LED Solid ON.
L E
XAMPLE Loop Input Channel 1 detection zone is occupied by a vehicle that
TATE Loop Inductance Display (Option 1 ON).
S B
ARGRAPH DISPLAY OFF if no vehicle is detected. Number of sensitivity levels that the
SEGMENT DISPLAY Loop / Lead-In circuit inductance in microHenries. If the value
7-
EXT L=.
T L
OOP INPUT LED The loop input LED operates as normal, indicating call as
EXAMPLE Channel 1 Loop / Lead-In circuit inductance is 98 microHenries
OTES If Option 2 (-ΔL/L Display) is ON, this display is only shown
N
Fail memory.
power failure.
or (DOWN) button, or a reset input to the RESET terminal on the motherboard.
P
ARAMETER .................... Firmware Version and Revision.
ETTINGS.......................View Only.
S
ETTING DISPLAYED ....... View Only.
S 7-
SEGMENT DISPLAY ....... Model letter and firmware version on one scr een
D
EFAULT SETTING........... Not Applicable.
XAMPLE ........................ Model E-1400 firmware version 3, revision .00.
E
detection zone is unoccupied.
exceeds (first dot = current sensitivity level, second dot = next lower sensitivity level, etc.).
exceeds the detection threshold by four sensitivity levels.
inductance change caused by the vehicle exceeds the detection threshold (first dot = current sensitivity level, second dot = next lower sensitivity level, etc.) if a vehicle is detected.
exceeds 999, the display will alternate between the thousands place (1 or 2) and the lower three digits of the inductance value.
expected.
and it is not detecting a vehicle.
when the channel is not detecting a vehicle.
and revision on the other. The display will toggle between the two screens with a 50% duty cycle.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 32 of 43
Page 33
-L/L
-L/L = FREQ KHZ
SECONDS
©
SENSITIVITY
DELAY PU LSE
EXTENSION MAX
OPTION ON OFF
PRESENCE
EOG SCANNING
LOOP FAIL
1 2 3 4
S
TATE ..........................Loop Inductance -ΔL/L Display (% Change) (Option 2 ON).
B
ARGRAPH DISPLAY ....OFF i f no vehic le is detec ted. Numb er of sens itivity levels
that t he induc ta nce cha nge ca used by the vehicle exceeds the detection threshold (first dot = current sensitivity level, second dot = next lower sensitivity level, etc.) if a vehicle is detected.
7-
SEGMENT DISPLAY ....Percentage o f change in inductance of the Loop / Lead-In
circuit.
T
EXT .............................-L/L.
L
OOP INPUT LED .........The loop input LED operates as normal, indicating call as
expected.
E
XAMPLE ......................Loop Input Channel 1 detection zone is occupied by a vehicle
that exceeds the detection threshold by t hree (3) sensitivity levels and Channel 1 percentage change of inductance is
0.087%.
N
OTES ..........................This display is only visible while the channel is detecting a
vehicle.
S
TATE ..........................LCD Display Te st.
B
ARGRAPH DISPLAY ....All segments on.
7-
SEGMENT DISPLAY ....All segments on.
T
EXT .............................All segments on.
L
OOP INPUT LED .........The loop input LED operates as normal, indicating call as
E
XAMPLE ......................All segments on.
N
OTES ..........................This dis pla y is v isi ble whe ne ver an y two or three pushbutton
expected.
switches are pressed at the same time.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 33 of 43
Page 34

5.4 Error Message Display Screens

1
1
The Model E-1400 stores operating para meters and detection reference settings in two places. The detector hardware includes an internally mounted EEPROM and an externally mounted EEPROM memory module. In order for the detector to operate proper ly, an external EEPROM memory module MUST be connected to the EGMS rack.
Logic programmed into the detector operating firmware continually checks the data stored on the two components listed above. This check function pro vides a means of verifying several key factors t hat ensure that the detector is functioning properly. These incl ude:
Detector power up status - Has the detector powered up correctly and is it operating in the correct state?
Detector Failure status - Are failure cond itions being identified and is the detector responding to these failures correctly?
Stored data status - Is the data stored on each of the two data storage components consistent with one another? This phase of the check function allows the user to program the detector with data stored on a master external EEPROM module co nnected to the EGMS rack.
If the information obtained via the check function indicates that there is a problem with some phase of the detector operation, the LCD screen will display one or more of the erro r messages listed below.
Uninitialized E xternal EEPR OM. The display will toggle between the two screens s how n w it h a 5 0% du t y c ycle . Th is is a n ind ic at io n tha t the external EEPR OM memory mod ule is missing, that it is not installed properly, or that it is uninitialized. To correct this problem, verify that the module is installed correctly and press the FUNC b ut t o n. I f the erro r message does not clear, the external EEPROM memory module is defective. Replace it with a functional module and press the FUNC button to clear the error message. If the error message still does not clear, return the detector to the factory for servicing.
Warning: If the error message clears after pressing the FUNC button, the user must confirm that the stored detector settings are correct for the installation in question.
Uninitialized Internal E EPROM. The display will toggle between t he two sc ree ns sho w n w ith a 5 0% d u ty c yc le . Th is is a n i ndi ca t ion t ha t t he internal EEPROM is uninitialized. To correct this problem, press the FUNC button. If the error message does not clear, the internal EEPROM is defective. Return the detector to the factory for servicing.
Warning: If the error message clears after pressing the FUNC button, the user must confirm that the stored detector settings are correct for the installation in question.
Uninitialized Internal a nd External EEPROMs. The dis play will togg le between the two screens shown with a 50% duty cycle. This is an indication t hat the internal EEPRO M and the external EEPROM memory module are uninitialized. To correct this problem, press the FUNC button. If the error message does not cle ar, the internal EEPROM and the external EEPROM memory module are defective. Return the detector to the factory for servic ing. NOTE: Clearing this error message will reset the detector to factory default settings and will clear any previously mapped output(s).
Warning: If the error message clears after pressing the FUNC button, the user must confirm that the stored detector settings are correct for the installation in question.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 34 of 43
Page 35
External EEPROM Error. The display will toggle between the two
1
1
screens shown with a 50% duty cycle. This is an indication that the data stored on the external EEPROM memory module is co rrupt. Pres s the FUNC button to copy the reference data stored on the interna l EEP RO M to the external EEPROM memory module. If the error message does not clear, the external EEPROM memory module is defective. Replace it with a functional module and press the F UNC button to clear the error message. If the error message still does not clear, return the detector to the factory for servicing.
Warning: If the error message clears after pressing the FUNC button, the user must confirm that the stored detector settings are correct for the installation in question.
Internal EEPROM Error. The display will toggle between the two screens shown with a 50% duty cycle. This is an indication that the data stored on the i nternal EEPROM is corr upt. Pres s the FUN C butto n to copy the reference data stored on the external EEPROM memory module to the internal EEPROM. If the error message does not clear, the internal EEPROM is defective. Return the detector to the factory for servicing.
Warning: If the error message clears after pressing the FUNC button, the user must confirm that the stored detector settings are correct for the installation in question.
Internal and External EEPROM Error. The display w ill toggle between the two screens shown with a 50% duty cycle. This is an indication that the data stored on the internal EEPROM and the external EEPROM memory module is corrupt. Press the FUNC button to copy the reference data stored on the microprocessor to the internal EEPROM and the exter na l EEP RO M memory module. If the error message does not clear, the internal EEPROM and the external EEPROM memory mo dule are defective. Return the detector to the factory for servicing. NOTE: Clearing this error message will reset the detector to factory default settings and will clear any previously mapped output(s).
Warning: If the error message clears after pressing the FUNC button, the user must confirm that the stored detector settings are corre ct for the installation in question.
Detection Reference Error. This is an in dication that the in ternal EEPRO M and the exter nal EEPR OM memory module detection reference values do not match. Press the FUNC button to clear the error message.
Warning: If the error message clears after pressing the FUNC button, the user must confirm that the stored detector settings are correct for the installation in question.
Warning: Clearing this error will reset all internal and external detection reference values. If a vehicle (or vehicles) is (are) present over any or all loops at this time, the inductance change caused by the vehicle(s) will be incorporated into the new baseline detection reference and the vehicle(s) will no longer be detected.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 35 of 43
Page 36

5.5 Loop Check Pass / Fail Indications

The Loop Check feature is enabled whenever the Loop Check Timer for a specific loop input is set to a value between 1 and 255 minutes. W hen the Loop Check Timer has timed out, the detector will automatically short the check loop to simulate vehicle detection. If t he percenta ge of inducta nce change is greater tha n or equal to -0.04% ΔL/L, a nd t he c ha nne l o n w hic h t he c ur re nt loo p check test is selected and being viewed
Disabling the Loop Check feature (setting the Loop Check Timer to zero), powering down the detector, or rese ttin g the c han nel tha t fa iled t he loop check test will clear the Loop Check failure. However, if the problem that caused the Loop Check failure is not corrected, the c hannel will fail the next loop check test.
The Loop Check feature is suppressed if the loop has an open, a shorted, or a relative (±25% ΔL/L) failure.
The detector will perform a loop check tes t on all programmed channels two sec onds after start-up or two seconds after a reset.
NOTE: If the Loop Check Timer is set to 000, the Loop Check function is off.
on the LC D, the display will show PAS.
If the percentage of inductance change caused by activating the check loop is less than -
0.04% Δ L/L, the detector w ill initiate its check loop verification process. The LCD screen will display an indication of a pending state (Pcl) during the check loop verification. This screen will flash for 55 seconds until one of three things happen; the check loop verification fails (indication shown be low - Lcl), the check loop(s) ca using the check loop verification is (are) connected or restored to a functional state (indication shown above - PAS), or t he check loop(s) causing the check loop verification is (are) turned off, i.e. the Loop Check Timer of the loop(s) is set to 000.
If the check loop verification fails, the LED corres ponding to that loop channel will begin flashing at a three Hz rate. If the channel corresponding to the check loop failure is selected a nd being viewed on the LC D, the display will toggle betwee n a screen showing L and LOOP FAIL and a s cree n sho win g Lcl and LOOP FA IL with a 5 0% d uty cyc le to indicate a that Loop Check failure has occurred. In addition, the loop status will enter the Call state and the He alth Status will enter the Fail condition. A Loop Check failure will de ac tivate a ny Pr imar y Vital Outp ut to which any loop input or Health Status is mapped. When the Health Status of one or more loop inputs is (are) mapped to an Auxiliary Vital Output, a loop check failure will deactivate the Auxiliary Vital output.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 36 of 43
Page 37

5.6 Vital Input (Vital In) Status / Fail Indications

Each of the four Vital Inputs has a dedicated, dual color (green / red) LED which illumina tes in var ious ways to indicate the state of the Vital Input. When the LED is off, the Vital Input is deactivated (low volta ge leve l). W he n t he LED d isp la ys a s s ol id g re en, t he V ita l I np ut is a ct iva te d (h igh vo lta ge le ve l). When the LED dis plays as s o lid red, the Vital Input has failed its self-check.

5.7 Primary Vital Output (Vital Out) Status / Fail Indications

Each o f t he f o ur Primary Vital Outputs has a dedicated, dual color (green / red) LED which illuminates in var ious wa ys to indica te the state of the Primary Vital O utput. Whe n the LE D is o ff, the Primary Vital Output is deactivated (low voltage level). When the LED displays as solid green, the Primary Vital Output is activated (high volta ge level). When the LED displa ys as a tw o Hz flashing green, timing delay is activated and the Primary Vital Output is functioning (low voltage level until delay times out). When t he LED disp lays as a five Hz flas hing green, tim ing e xtens ion is act ivate d and t he Primary Vital Output is funct ioning (low voltage leve l until exte nsion times out. When the LED displays as solid red, the Primary Vital Output has failed its self-check.

5.8 Auxiliary Vital Output (Vital Aux) Status / Fail Indications

Each o f th e fo ur Auxilia ry Vita l O utputs has a dedicated, dual color (green / re d) LED that illuminates in various ways to indicate the state of the Auxiliary Vital Out p u t. When the LE D is o ff, t he Auxiliary Vital O utpu t is d eact ivat ed ( low vo lta ge leve l). Whe n the LED d ispla ys as soli d gre en, t he Auxiliary Vital Out put is ac ti vat ed (h igh vo lta ge le ve l). Whe n t he LED d isp la ys a s s oli d re d, th e Auxiliar y Vital Output has failed its self-check.

5.9 Synch Fail Indications

5.9.1 Series Synch Fail Indications

When the Synch feature is set to ope rate in Series Synch mode (Option 7 is set to 7.01), the LCD will displa y SYS. If, after completing t he sampling routine, the detector determines there is a failure on o ne or more loop inputs, the LC D will dis p la y FSY and co nt in ue s a mpl ing. I f the Sy nc h fa il ure is cleared, the fail indication (FSY) will no longer be displayed on the LCD.

5.9.2 Parallel Synch Fail Indications

When the Synch feature is set to operate in Parallel Synch mode (Opt ion 7 is set to 7.02), the LCD will display SYP If, after completing the sampling ro utine, the detector determines there is a failure on o ne or more loop inputs, the LC D will dis p la y FSY and continue sampling. If the Synch fa ilure is cleared, the fail indication (FSY) will no longer be displayed on the LCD .
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 37 of 43
Page 38

5.10 Loop Fail Indication s

If the total inductance of a channel’s loop input network goes out of the range specified for the detector, or rapidly changes by more than ±25%, the loop input channel will enter the Fail-Safe mode and disp lay LOOP F AIL on the LCD. The type of loop failure will also be displayed as L lo (for -25% change or shorted loop conditions) or L hi (for +25% change or open loop conditions). This will continue as long as the loop fault exists. Fail-Safe mode generates a continuous call. At the time of a loop failure, the loop input channel’s LED will begin to flash at a rate of three flashes per second. The LED will continue this display pattern until the channel is manually reset or power is removed.
If the loop self-heals, the LOOP FAIL message on the LCD will extinguish and the loop input channel will resume operation in a normal manner; except, the LED will continue the three flashes per second display pattern, thus, pro viding an alert that a prior Loop Fail condition has occurred. Each loop failure is counted and accumulated into the Loop Fail Memory. The total number of loop failures for the loop input channel (since the last power interruption or manual reset) is written into the Loop Fail Memory and can be seen by stepping throu gh the channel’s functions in Program Mode to the LOOP FAIL display.
This is a useful tool that can be used to identify inter mittent loop problems. If the count is extremely high for the period of time observed, the problem is very likely a loose connecti on (check for loose connections at the terminal strip and bad splices in the field). The Loop Fail Count is reset when power is removed from the detector. This prevents the Loop Failure Count from moving to another loop if the detector is moved to a new location.
To view the Loop Fail Count, select the desired loop input channel using the CHAN button then repeatedly press the FUNC button until the LOOP FAIL display is shown. The Loop Fail Count display is after the OPTION displays. Pressing the (UP) or (DOWN) button while the Loop Fail Cou nt is displayed, will reset the count to zero.

5.11 Setting Sensitivity using the Bargraph

The bargraph is a graphical representation of the relative chan ge of inductance as seen by the detector loop input channel; automatically taking into account sensitivity setting, loop geometry, configuration, lead-in length, etc. The first bargraph segment repres ents the minimum inductance change (set by the sensitivity level) necessary for the loop input channel to output a call. Each additional segment represents the inductance change in excess of the next sensitivity threshold. Usually, the lar ger the vehicle detected, the more bargraph segments displayed, i.e. the greater the -L/L. The bargraph can be used to select the proper sensitivity level.
The bargraph below shows the deflection (3 segments) of a vehicle with Sensitivity set to Level 5. The vehicle in the loop zone is causing a change of inductance greater than 1.28% -ΔL/Lor Sensitivity Level 3.
The bargraph below has the same vehicle in the loop zone causing the same inductance change as above. Because the sensitivity setting was increased to Level 8, six segments are now displayed. If the bargraph displays 5 or 6 segments for a vehicle in the loop, the sensitivity has been set to the proper range.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 38 of 43
Page 39

5.12 Full Restore To Factory Default Settings

1 © 1 © 1 © 1
©
Pressi ng a ll fo ur fro nt pa ne l p us hb utt on sw it che s si mu lta ne o us ly a nd co nt in uo us ly fo r five (5 ) s e co nds restores all factory default settings. The countdow n of the five second pe riod is displa yed o n the LCD. Releasing any or all of the switches before the countdown ends aborts the full restore operation. (See Section 3.5 for default settings). Refer to Sect ion 5.0 for details. Caution: Restoring the factory default settings will clear all previously programmed settings including any m appe d outpu t(s).

5.13 Diagnostic F eat ur e s Acti vation

When the detector is operating in normal mode, pressing either the  (UP) or  (DOWN ) button will temporarily activate Option 1 (Loop Inductance Display) and Option 2 (Loop Inductance -ΔL/L Display) for all loop inputs and Optio n 6 (Audible Detect Signal) for the loop input ind icated in the lower portion of the LCD. All t hree of these diagnostic options w ill remain active for 15 minutes.

5.14 Display Test

Pressing any two or t hree of the front panel s witches simulta neously w ill turn on a ll possible s ymbols and mes sages o n the LCD screen.

5.15 Changing or Relocating Detectors

The Model E-1400 detector stores user-programmed parameters on an internal EEPROM and an exter nal EEPR OM memo ry mo dule conne cted to the EGMS rack. This a llows t he use r to c hange or relocate detectors and reset the parameters that were programmed for a particular site or detector. When a new detector is installed or an existing detector is relocated, an Error message will ap pear on the LCD. The LCD will alternate between Cpy and In. This message means that the parameters stored on the internal EEPROM do not match the parameters stored on the external EEPROM memory module. At this point, the user has two choices. Pressing the FUNC button will accept the parameters stored on the inter nal EEPR OM a nd store them o n the e xterna l EEPR OM memory module. Pressing either the  (UP) or  (DOW N) b utto n will change the alte rnat ing me ssa ge on t he LCD to Cpy and Out. Pressing the FUNC b utton will acce pt the para meters stored o n the externa l EEPROM memory module and st ore them on the int ernal EEPROM.
Parameters stored on the external EEPROM memory module do not match those stored on the inte rna l E EPR O M. Pres si ng t he FU NC but to n will accept the parameters stored on the internal EEPROM a nd store them on the exte rna l EEPR OM module, i.e. copy the internal EE PROM parameters to the external EEPROM module.
Parameters stored on the external EEPROM memory module do not match those stored on the inte rna l E EPR O M. Pres si ng t he FU NC but to n will accept the parameters stored on the external EEPROM memory module and sto re them o n the int ernal E EPROM , i.e. c opy the e xternal EEPROM memory module parameters to the internal EEPROM.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 39 of 43
Page 40

5.16 BLOCK DIAGRAM

Loop
Isolation
Loop
Oscillator
Loop
Capacitors
Non-
Loop
Isolation
Loop
Oscillator
Loop
Isolation
Loop
Oscillator
Loop
Isolation
Loop
Oscillator
Squaring Circuit
Vital Input 1
Vital Input 2
Vital Input 3
Vital Input 4
+9 to
Power
DC Com
Vs
+12 VDC
GND
Front Panel Pushbutton
32 MHz
Oscillator
Primary
Audible
Buzzer
Loop
Capacitors
Loop
Capacitors
Loop
Capacitors
Auxiliary
RS-232
Input
RS-232
Output
Check Inputs
Check
Relay
Check
Relay
Check
Relay
Check
Relay
+10 VDC
+5 VDC
Synch In
Synch Out
Loop 1
Check Inputs
Loop 2
Inputs
Check
Loop 3
Check Inputs
Loop 4
Internal
EEPROM
External
EEPROM
Inputs
Loop 1
Loop 2
Inputs
Loop 3
Inputs
Loop 4
+18 VDC
Loop
Activation
Loop
Activation
Loop
Activation
Loop
Activation
Suppl
y
Micro
Controller
volatile
DETECT
LEDs
LCD
Switches
Vital
Outputs
Vital
Outputs
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 40 of 43
Communication
Communication
Page 41

6.0 THEORY OF OPERATION

The Reno A & E Model E-1400 detector digitally measures changes in the resonant frequency of four independent loop circuits to determine if a vehicle has entered the detection zone. The Model E-1400 detector applies an excitation voltage to each loop circuit res ulting in the loops oscillating at their resonant frequencies. The current flow in the loop wire creates magnetic fields around the loop wire. When a vehicle passes over the loop area, the conductive metal of the vehicle causes a loading of the loop’s magnetic fie lds. The loading decreases the loop inductance, which causes the res onant frequency to increa se . B y co nt in uo us ly s a mpl in g t he lo op’ s res ona nt fr eq ue nc y, t he ma gni t ude a n d ra te o f cha n ge c an be determined. If the frequency change exceeds a selectable thres hold (set by the sen s it i vi t y s e t t in gs), the detector will deactivate a vital output if one is mapped. If the rate of change is slow, typical of environmental drift, the detector will continuously track and compensate for the change. T he Model E­1400 detector also monitors the loop frequency for out of range conditions such as an open o r shorted loop circuit.
The Model E-1400 detector is a scanning detector. The scanning method sequentially turns on and off each channel’s loop oscillator. Each channel’s oscillator c ircuit supplies the excitation voltage that is coupled to the loop circuit by a loop isolation tra nsformer. T he trans former p rovides high co mmon mode isolation between the loop and detector electronics, which allows t he detector to operate on poor quality loops including a single short to ground. The transformer also limits the amount of static energy (lightning) that can transfer to the detector electronics. A spark gap transient suppression device is connected across the loop inputs connected to the isolation t ransforme r. This de vice dissipa tes stat ic charges prior to the transformer. A network of four capacitors is connected to the detector side of the isolation transformer. Three of the capacitors can be switched in or out of the oscillator circuit to shift the frequency of the loop oscillator circuit thus provid ing frequency separation between a djacent loops. The three switchable capacitors are electronically switched using FETs and are selected when programming parameter values wit h the fro nt panel pushbutton switches.
The outputs from the four loop oscillators are tied together a nd fed into a common sq uaring circuit. T his is possible since the detector is a scanning detector allowing only a single loop osc illator to be operating at any give n time. The sine wave from each loop oscilla tor circuit is squared to provide a precise zero crossing signal for the input to the microcontro ller. This signal is called the loop sample. The loop sample is an integr al number of complet e osci llatio ns from the loop oscillator circuit. The number of loop oscillations counted is a function of the s elected sensitivity setting for the detector channel. The required number of loop oscillations needed for a loop sample increases as the se nsitivity setting is increased. The microcontroller uses the period of the loop sa mple for accumulating high-speed (32 MHz) crystal clock pulses generated by the microcontroller’ s internal high-speed crystal oscillator. The number of crystal clock pulses accumulated during consecutive loop samples is compared to the internal reference number of crystal clock pulses stored in the mic r o contro l le r ’ s me mo ry.
When a vehicle enters the loop zone the loop inductance decreases. This decrease in loop inductance causes an increase in the loop oscillator frequency. In turn, an increase in loop oscillator frequency results in a decrease of the time period for the loop sample. Hence, when a vehicle enters the loop zone, the number of crystal clock pulses accu mulated during a loop sample period decreases . By comparing the new count with the reference count, a percentage change can be calculated that indirectly relates to the inductance change. If the magnitude of the change exceeds a selectable threshold (sensitivity setting), the detector deactivates a vital output if one is map ped . The rate of c hange is also monitored. S low rates o f change caused by environmental fluctuations are tracked and automatically compensated for. This process is conducted independently for each of the four loop oscillator circuits.
The microcontroller uses the high-speed crystal clock count to calculate the loop inductance, frequency and percentage of change. If selected, the values are displayed on the seven-segment LCD. The microcontroller also processes the pushbutton switch selections for the LCD display and stores the operating parameters in non-volatile memory. Stored parameters are changed with the front panel switches or vi a the front panel RS-23 2 communications port and are unaffected by loss o f power or detector reset. The microcontroller co nt inuously proce sses the loop s amples a nd detec tor operat ion is not effected during the operation of the switches or the LCD. NOTE: When either the sensitivity or frequency of a loop input channel is changed, the loop inp ut channel is reset.
In addition, the microcontroller conditions the outputs based on Loop Status Inputs, Vital Inputs, Health Status Inputs , and the programmed settings of the various timers (Primary Vital Output Delay, Primary Vital Out put E xte nsion, and Loop Check) and options (Optio n 1, Option 2, Option 3, Optio n 4, Option 5, Option 6 and Option 7).
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 41 of 43
Page 42

7.0 MAINTENANCE AND TROUBLESHOOTING

The Reno A & E Model E-1400 Detector requires no maintenance. If you are having problems with a Model E-1400 detector, use the troubleshooting tips be low to help determine the cause of the problem.

7.1 Basic Troubleshooting Tips

Detector Enclosure Checks:
Verify correct input voltage to the detector. Check the detector diagnostics. If there is a current loop failure or t he loop fai l memory s hows a
loop failure has occurred, the problem is probably not related to the detector module. Check the loop / lead-in cable resistance of the detect loops and the check loops. A value greater
than fi v e Ohms is an indic ation of a splice or connection problem or a broken loop wire. Disconnect the detect loop or check loop from the detector. Use a 500 VDC Megger to measure t he
resistance from one of the loop wires to earth ground. It should be greater than 50 megOhms. If lower, separate the splice at the pull box and isolate the problem to the loop or lead-in cable.
Verify all terminal screws are tight. Inspect all loop connections in the detector enclosure, especially crimped lug connections. If
suspect, solder the lugs to the wire. If the lightening or surge suppression devices on the loop inputs in the detecto r enclos ure are s uspect,
remove or replace them. Check for places in the detector enclosure where the harness wires or lead-in cable may be pinched
or chaffed. Make sure that each pair of interconnect (lead-in) wires for the loop(s) are individually twisted
together.
Field Checks:
Inspect the loop. Look for exposed wires, debris pressed into the loo p wires, p ave me nt s hifts , e tc . Check s plices i n the pull box. The y must be so ldered and waterproofed. DO NOT USE WIRE
NUTS. Check that each set of loop lead-in wires in each pull box is twisted toget her [six (6) twists per foot,
minimum] and that lead-in lengths are not excessive.
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 42 of 43
Page 43

7.2 Things To Know About Loops

Always use a wire with cr oss-linked Polyethylene insulat ion (insulation type XLPE) for loop wire.
Typical sensing height is ⅔ of the shortest leg of a loop. Therefore, a 6΄ x 6΄ loop will have a detection height of 4΄.
The inductance of a con ventional four -sided loop can be estimated using the formula:
L = P x (T
2
+ T) / 4 Where: L = Loop Induct a nc e in mi c r o he n r ie s
P = Loop Perimeter in feet
T = Number of Turns of Wire. Therefore, a 6΄ by 6΄ loop with 3 turns would have a n inductance o f:
L = (6 + 6 + 6 + 6) x (3
2
+ 3) / 4 L = 24 x (9 + 3) / 4 L = 24 x 12 / 4 L = 24 x 3 L = 72 microhenries.
The inductance of a Qua drupole
L = [P x (T
2
+ T) / 4] + [CL x (T2 + T) / 4] Where: L = Loop Inductanc e in microhenries
TM
loop can be estimated using the formula:
P = Loop Perimeter in feet T = Number of Turns of Wire CL = Len gth of Center Leg in feet.
Therefore, a 6΄ by 50΄ loop with a 2-4-2 confi guration would have an inducta nce of:
L = [(6 + 50 + 6 + 50) x (2
2
+ 2) / 4] + [50 x (42 + 4) / 4] L = [112 x (4 + 2) / 4] + [50 x (16 + 4) / 4] L = (112 x 6 / 4) + (50 x 20 / 4) L = (112 x 1.5) + (50 x 5) L = 168 + 250 L = 418 microhenries.
Loop Feeder cable typically adds 0.22 microhenries of inductance per foot of cable. Total inductance of loops connected in ser ie s : L Total inductance of loops connected in pa rallel: L
= L1 + L2 + L3 + . . . + LN.
TOTAL
= 1 / [ (1 / L1) + (1 / L2) + (1 / L3) + . . . + (1 / LN) ].
TOTAL
889-0505-03 E-1400 Operation Manual Rev: Apr 2020 Page 43 of 43
Loading...