RADWAG Scales User Manual

Page 1
User Manual
Single load cell platform scales
Manual number: ITKU-18-02-01-12-A
Scales of WPT series
Waterproof scales of WPT/H series
Waterproof scales of WPT/HR series
MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS
RADWAG Wagi Elektroniczne 26 – 600 Radom Bracka 28 Street - POLAND Phone +48 48 38 48 800, phone/fax. +48 48 385 00 10 Selling department +48 48 366 80 06
www.radwag.com
Page 2
JANUARY 2012
- 2 -
Page 3
TABLE OF CONTENTS
1. NTENDED USE .......................................................................................................................................5 I
2.
PRECAUTIONS........................................................................................................................................6
2.1. .....................................................................................................................................6 Maintenance
Accum lator / battery pack...............................................................................................................6u
2.2.
2.2.1. er supply of weighing indicators in plastic casings .........................................................7 Pow
2.2.2. Replacement of worn batteries..............................................................................................7
2.3. Operation in a strong electrostatic field ............................................................................................8
3. WARRANTY CONDITIONS ......................................................................................................................9
4. MAIN MENSIONS .................................................................................................................................9 DI
4.1. able scales WPT/F series...............................................................................................................9
T
4.2. aterproof scales of WPT/H series................................................................................................10 W
4.3. aterproof scales of WPT/HR series.............................................................................................12 W
4.4. Scales of WPT series.....................................................................................................................12
5. UNP CKING AND ASSEMBLY.............................................................................................................16 A
5.1.
PT/F series.........................................................................................................16 Table scales of W
PT series.....................................................................................................................17 Scales of W
5.2.
Waterproof scales of WPT/H, WPT/HR series................................................................................18
5.3.
6. GETTING STARTED..............................................................................................................................19
7. KEYPAD.................................................................................................................................................20
8. KEYS’ FUNCTIONS ...............................................................................................................................20
9. I SCRIPTIONS ON THE DISPLAY........................................................................................................21 N
10.
USER MENU ........................................................................................................................................22
10.1. ....................................................................................................................................22 Submenus
10.2. Browsing user menu.....................................................................................................................23
10.2.1. pad..............................................................................................................................23 Key
Return to the weighing mode............................................................................................23
10.2.2.
11. WEIGHING ...........................................................................................................................................24
11.1. ..........................................................................................................................................25 Tarring
11.2. alue.....................................................................................................................25 Inscribing tare v
11.3. .........................................................................................................................................26 Zeroing
11.4. eighings in two ranges ..............................................................................................................26 W
11.5. eight unit.......................................................................................................27 Selection of basic w
Temporarily selected unit .............................................................................................................28 11.6.
MAIN PARAMETERS...........................................................................................................................29
12.
12.1. el ..................................................................................................................29 Setting a filtering lev
12.2. ..................................................................................................................................30 Median filter
12.3. .........................................................................................................................31 Autozero function
Tare function ................................................................................................................................32 12.4.
RS 232 PARAMETERS........................................................................................................................33
13.
13.1. pe.................................................................................................................................33 Printout ty
13.2. s threshold................................................................................................................34 Minimal mas
13.3. ......................................................................................................................................35 Baud rate
Serial transmission parameters....................................................................................................36 13.4.
OTHER PARAMETERS........................................................................................................................37
14.
14.1. Backlight function.........................................................................................................................37
14.1.1. ing from mains ...................................................................................37 Backlight for supply
Backlight for supplying from batteries...............................................................................38 1
4.1.2.
“Beep” signal – after pressing a key
14.2. .............................................................................................39
14.3. atic switch-off.....................................................................................................................39 Autom
14.4. Batter oltage level check...........................................................................................................40y v
14.4.1. ......................................................................................................40 Checking the batteries
14.4.2. discharge pictogram.............................................................................................41 Battery
14.4.3. .............................................................................................41 Accumulator charging option
Formatting rechargeable battery packs.............................................................................42
14.4.4.
- 3 -
Page 4
15. WORK MODES ....................................................................................................................................43
15.1. of operation modes.......................................................................................43 Setting accessibility
15.2. of operation modes.......................................................................................... 44 Selecting quantity
15.3. ..............................................................................................45 Counting pieces of the same mass
15.4. .....................................................................47 +/- control referring to the inscribed standard mass
15.5. Control of % deviation referring to the inscribed standard mass...................................................49
15.5.1. its weighing.......................................................................49 Standard mass determined by
Mass of standard inscribed to scale memory....................................................................50 1
5.5.2.
Autom
15.6. atic tare.............................................................................................................................. 51
15.7. force on the pan – latch..................................................................................52 Measurement max
15.8. Totalizing ......................................................................................................................................52
15.8.1. ork mode....................................................................................................53 Enabling the w
15.8.2. .........................................................................................................53 Totalizing procedure
15.8.3. of the last value of sum of weighed goods ..........................................................54 Memory
5.8.4. Return to weighing............................................................................................................551
15.9. eighing animals.........................................................................................................................56 W
Tare memory...............................................................................................................................57
15.10.
15.10.1. Entering the tare value to the scale memory...................................................................57
15.10.2. Selecting a tare value from the memory..........................................................................59
16. USER CALIBRATION...........................................................................................................................60
16.1. ....................................................................................................................................60 Calibration
Start mass adjustment..................................................................................................................62 16.2.
COOPERATION WITH PRINTER.........................................................................................................63
17.
18. COOPERATION WITH COMPUTER.................................................................................................... 64
19. COM UNICATION PROTOCOL..........................................................................................................65 M
19.1. ......................................................................................................................65
General information
19.2. ...........................................................................................66 A set of commands for RS interfaces
19.3. ............................................................................................................66 Respond message format
Command’s description................................................................................................................67
19.4.
19.4.1. ..............................................................................................................................67 Zeroing
19.4.2. ..............................................................................................................................67 Tarring
19.4.3. alue ...................................................................................................................67 Get tare v
19.4.4. ...................................................................................68 Send the stable result in basic unit
19.4.5. in basic unit..........................................................................68 Send the result immediately
19.4.6. .................................................................................69 Send the stable result in current unit
19.4.7. in current unit.......................................................................70 Send the result immediately
19.4.8. itch on continuous transmission in basic unit ...............................................................70 Sw
19.4.9. Switch off continuous transmission in basic unit...............................................................70
19.4.10. Switch on continuous transmission in current unit..........................................................71
19.4.11. Switch off continuous transmission in current unit ..........................................................71
9.4.12. Send all implemented commands...................................................................................711
19.5. omatic printouts..........................................................................................72 Manual printouts / aut
19.6. ..............................................................................................................73 Continuous transmission
Configuring printouts....................................................................................................................73 19.7.
ERROR COMMANDS...........................................................................................................................74
20.
21. TECHNICAL PARAMETERS................................................................................................................75
21.1. PT series................................................................................................................... 75 Scales of W
21.2. PT/F series.......................................................................................................77 Table scales of W
21.3. aterproof scales of WPT/H series .............................................................................................78 W
Waterproof scales of WPT/HR series...........................................................................................81 21.4.
TROUBLE SHOOTING.........................................................................................................................83
22.
23. ADDITIONAL EQUIPMENT..................................................................................................................83
- 4 -
Page 5

1. INTENDED USE

Scales are designed for fast and precise measurements of weighed loads masses and direct commercial settlements. Tarring in full weighing range enables to determine net mass of weighed loads.
Functions:
backlight of display
level of filtration
autozero function
setting baud rate of transmission
continuous data transmission for RS 232
automatic operation for RS 232
designed printouts
designation minimum mass for function operating
counting pieces
+/- mass control
percentage deviation from standard mass
latch of maximum scale indication
automatic tare
memory of tare
Memory of 9 tare values
inscribing tare value
automatic scale switch-off
user calibration
Totalizing
Weighing animals
User functions may have attribute of accessibility. For this reason it is possible to adjust scale to individual needs to provide access to only these functions which are currently needed. Attribute determination accessible / inaccessible is possible in user menu and described in further part of manual.
- 5 -
Page 6

2. PRECAUTIONS

2.1. Maintenance

A. Please, read carefully this user manual before and use the device
according to its intended use.
B. Devices that are to be withdrawn from usage should be sent back to
the producer or in case of own utilization do it according to the law.

2.2. Accumulator / battery pack

The device connected to mains inteligently monitors the battery state and charges it if possible. After sudden lack of power supply from the mains the device automatically switches to accumulator without breaking operation.
Scales aquipped with indicator PUE C/31 (plastic casing) are devices designed to be supplied from NiMH batteries (nickel-metal-hydrogen) with rated voltage of 1.2V, size R6 and capacities from 1800 to 2800mAh charged while connected to mains without stopping operation.
Scales equipped with PUE C/31H and PUE C/31H/Z weighing indicators (stainless steal housing) are devices designed to be supplied from SLA accumulators (Sealed lead acid type) 6V o and capacity 3 to 4Ah charged while connected to mains without stopping operation.
In case of an elongated storage period in low temperatures, it is not allowed the full discharge of the accompanied batteries.
The equipment including accumulators does not belong to your regular household waste. The European legislation requires that electric and electronic equipment be collected and disposed separately from other communal waste with the aim of being recycled.
- 6 -
Page 7
Notice: Some symbols on accumulators identify harmful elements/compounds: Pb = lead, Cd = cadmium, Hg = mercury.

2.2.1. Power supply of weighing indicators in plastic casings

Indicators in plastic casing are intended to be supplied from a power adapter or from NiMH rechargeable battery pack (standard equipment). New rechargeable batteries should be formatted according to the description in the chapter 14.4.4. of this manual.
Alternatively, you can use to power the device R6 size standard non­rechargible batteries. If you want to use normal batteries instead of rechargeable ones, proceed as follows:
Before installing non-rechargeable batteries turn on the device and set <5.5.CHr6> to <no>, to switch off charging.
Then install the batteries.
Installing batteries without changing <5.5.CHr6> to <no> may cause damage of batteries and the indicator.

2.2.2. Replacement of worn batteries

Users have the ability to replace worn out batteries to new ones in weighing indicators PUE C/31 (plastic casing).
Procedure:
Open the lid of the chamber for batteries placed in the bottom of the indicator casing:
- 7 -
Page 8
Remove discharged and then insert new batteries into the chamber, according to given polarity (+/-):
Close the lid of the chamber for batteries:
In PUE C/31H and PUE C/31H/Z weighing indicators (stainless steel housing) the worn out accumulator can be exchanged to a new one by the authorized service of the manufacturer.

2.3. Operation in a strong electrostatic field

If the device is about to operate in a strong electrostatic field (e.g. printing houses etc.) it should be connected to the earthing.
Connect it to the clamp terminal signed
.
- 8 -
Page 9

3. WARRANTY CONDITIONS

A. RADWAG is obliged to repair or change those elements that
appears to be faulty because of production and construction reason,
B. Defining defects of unclear origin and outlining methods
of elimination can be settled only in participation of a user and the manufacturer representatives,
C. RADWAG does not take any responsibility connected with
destructions or losses derives from non-authorized or inappropriate (not adequate to manuals) production or service procedures,
D. Warranty does not cover:
Mechanical failures caused by inappropriate maintenance of the device or failures of thermal or chemical origin or caused by atmospheric discharge, overvoltage in mains or other random event,
Inappropriate cleaning.
E. Loss of warranty appears after:
Access by an unauthorized service,
Intrusion into mechanical or electronic construction
of unauthorized people,
Removing or destroying protection stickers.
F. Warranty conditions outline the warranty period for rechargeable
batteries attached to the device for 12 months. G. The detailed warranty conditions one can find in warranty certificate. H. Contact with the central authorized service:
+48 48 384 88 00 ext. 106 or 107.

4. MAIN DIMENSIONS

4.1. Table scales WPT/F series

WPT/F…/C series – main dimensions
- 9 -
Page 10
WPT/F…/C/K series – main dimensions
WPT/F…/C/R series – main dimensions

4.2. Waterproof scales of WPT/H series

WPT…H1 series (pillar 24cm) –main dimensions
- 10 -
Page 11
WPT…H1series (pillar 7cm) –main dimensions
WPT…H2 - WPT…H6 series – main dimensionS
- 11 -
Page 12

4.3. Waterproof scales of WPT/HR series

WPT…HR2 - WPT…HR6 series – main dimensions

4.4. Scales of WPT series

WPT…C1/K series – main dimensions
- 12 -
Page 13
WPT…C1/R series - main dimensions
WPT…C1 series – main dimensions
- 13 -
Page 14
WPT…C2/K series – main dimensions
WPT…C2/R series – main dimensions
- 14 -
Page 15
WPT…C2 series – main dimensions
WPT…C3/K series – main dimensions
- 15 -
Page 16
WPT…C3 series – main dimensions

5. UNPACKING AND ASSEMBLY

5.1. Table scales of WPT/F series

Unpack and put the scale on a flat even stable surface far away from sources of heat.
maximally screw out the transport protection according to the
drawing below:
Install the pillar, be careful about the cable linking the indicator with
the load cell:
- 16 -
Page 17

5.2. Scales of WPT series

Unpack and put the scale on a flat even stable surface far away from sources of heat and then:
- Remove transport protection:
For versions with an indicator on the pillar:
- 17 -
Page 18

5.3. Waterproof scales of WPT/H, WPT/HR series

Unpack and put the scale on a flat even stable surface far away from sources of heat and then:
Unscrew the pillar and the transport handler from the platform:
- 18 -
Page 19
Turn the pillar and mount it to
the platform. The surplus cable
place inside the pillar.
Pick up the pan and remove
the transport protection.

6. GETTING STARTED

After unpacking and mounting the scale level it out.
Use levelling legs and the level condition
indicator installed in the basis of the scale.
Turn the device on using the
for about 0.5 sec,
Wait for the test completion,
Then you will see zero indication and pictograms:
key – keep pressing the key
- zero indication
- stable result
kg - weight unit
If the indication is not zero press zero key.
- 19 -
Page 20

7. KEYPAD

8. KEYS’ FUNCTIONS

Switching on/off
Function key (operation mode selection)
Sending a weighing result to RS232
Zeroing
Tarring
Notice: After pressing
in this mode is described in details further in this manual.
+ keys’ functions changes. The way of operation
- 20 -
Page 21

9. INSCRIPTIONS ON THE DISPLAY

No Text string Description
1. FIL
2. bAud
3. PCS
4. HiLo
5. rEPL
6. StAb
7. Auto
8. t1
9. toP
10. Add
11. AnLS
12. tArE
13.
14.
15. PCS Operation mode – counting pieces
16. kg (g) Operation mode – weighing
17.
18. Net
19. Min
20. OK
21. Max
Filter level Transmission baud rate Piece counting +/- control according to a standard mass Automatic printout The condition of printing data Autozero correction Power save – time to switch off while no operation Latch of the max measurement Totalizing Weighing animals Memory of 9 tare values Indication in autozero zone (indication = exact zero) Stable result (ready to read)
Rechargeable battery pack or battery discharged (BAT-LO) Tare function has been used +/- control with reference to the standard mass: setting
the lower threshold or mass below the first threshold +/- control with reference to the standard mass:
load mass between the thresholds +/- control with reference to the standard mass: setting
the upper threshold or mass over the second threshold
- 21 -
Page 22

10. USER MENU

10.1. Submenus

User’s menu is divided into 6 basic submenus. Each group has its own characteristic name preceded by the letter P and a number.
P1 rEAd
P 1.1 Fil | 2 P 1.2 Auto | YES P 1.3 tArA | no P 1.4 Fnnd | no P2 Prnt P2.1 Pr_n | StAb P2.2 S_Lo | P2.3 bAud | 9600 P2.4 S_rS | 8d1SnP
P3 Unit
P3.1 StUn | kg
P4 Func
P4.1 FFun | ALL P4.2 Funi | no P4.3 PcS | no P4.4 HiLo | no P4.5 PrcA | no P4.6 Prcb | no P4.7 AtAr | no P4.8 toP | no P4.9 Add | no P4.A AnLS | no P4.b tArE | no
P5 othr
P5.1 bL | Auto P5.2 bLbt | 70 P5.3 bEEP | YES P5.4 t1 | Auto P5.5 CHr6 | YES
P6 CAL
P6.1 St_u | * FUNCTION * P6.2 uCAL | * FUNCTION *
- 22 -
Page 23

10.2. Browsing user menu

Use scale’s keys to move inside the menu.

10.2.1. Keypad

Entering main menu
+
Inscribing tare value
+
Increasing a digit value by „1” moving down in the menu
Battery / accumulator state monitoring
+
Toggling between gross / net values
+
Selecting the parameter or changing the value of a selected parameter
Entering the selected submenu or activating a parameter for changes
Confirmation (enter)
Leaving without changes or reaching a higher level of the menu

10.2.2. Return to the weighing mode

The changes that have been introduced should be saved in order to keep them in the memory for good.
While leaving parameters press <SAuE?> appears on the display. Then press: – to save changes or – to leave without changes.
- 23 -
key until the text
Page 24

11. WEIGHING

Put a load you want to weigh on the weighing pan. When the pictogram appears it means that the result is stable and ready to read. In order to assure long-term operation and appropriate measurements of weighted loads following precautions should be taken into consideration:
Loads should be placed on the pan delicately and carefully
in order to avoid mechanical shocks:
Loads should be placed centrally on the pan (errors caused
by eccentric weighing are outlined by standard PN-EN 45501
ch. 3.5 and 3.6.2):
Do not load the pan with concentrated force:
Avoid side loads, particularly side shocks should be avoided:
- 24 -
Page 25

11.1. Tarring

In order to determine the net mass put the packaging on the pan. After stabilising press -
(Net pictogram will be displayed in the left
upper corner and zero will be indicated).
After placing a load on the weight pan net mass will be shown. Tarring is possible within the whole range of the scale. After unloading the pan the display shows the tarred value with minus sign.
Notice:
Tarring cannot be performer when a negative or zero value is being displayed. In such case <Err3> appears on the display and short audible signal will be emitted.

11.2. Inscribing tare value

You can also inscribe a tare value. While in weighings mode press:
Press simultaneously and ,
You will see :
- 25 -
Page 26
Using and set the tare value,
Press
,
Program returns to weighings mode. The inscribed tare value can be seen on the display with „–” sign,
Tare can be inscribed anytime in weighings mode.
Notice:
1. You cannot inscribe a new tare value when the tare value in memory is greater than zero. In the case of trying this the <Err3> message will be displayed and short audible signal will be emitted.
2. Users can also enter up to 9 tare values to the scale memory (see 15.10 of his manual).

11.3. Zeroing

To ZERO the scale press: . The scale will display zero and following pictograms:
and . Zeroing is only possible within the scope of ±2% of full scale. While zeroing outside the scope of ±2% you will see <Err2>. Zeroing is possible only in stable state.
Notice:
Zeroing is possible only within the ±2% interval of the maximal range. If zeroing is performed beyond this range the <Err2> message and short audible signal will be emitted.

11.4. Weighings in two ranges

Switching between the I range and the II range happens automatically (exceeding Max of the I range). Weighings in the second range is signalled by a pictogram in the top left corner of the display.
Then weighings is done with the accuracy of the II range to the moment of returning to zero (autozero range
) where the scale switches back
to the I range.
- 26 -
Page 27

11.5. Selection of basic weight unit

This function is used to set weight unit the scale will start with.
Procedure:
Enter the submenu <P3.Unit> and then:
press
, until the expected unit appears on the display:
Options:
A. When the basic unit is [kg], users can toggle between:
[kg, lb, N], for verified scales [lb] is not accessible,
B. If the basic unit is [g], users can toggle between:
[g, ct, lb], for verified scales [lb] is not accessible,
- 27 -
Page 28
After you select the unit press , the scale returns to:
Return to weighing according to chapter - 10.2.2.
Notice:
After turning on the scale always sets the basic unit.

11.6. Temporarily selected unit

This function is used to set weight unit the scale will use temporarily until the next power off or next selection.
Procedure:
Press and then:
After you select the unit you want come back to weighing procedure.
- 28 -
Page 29
Options:
A. When [kg] is a basic unit, users can select following units:
[kg, lb, N], [lb] is not accessible for verified scales.
B. When [g] is a basic unit, users can select following units:
[g, ct, lb], [lb] is not accessible for verified scales.

12. MAIN PARAMETERS

Users can adjust the scale to external ambient conditions (filtering level) or particular needs (autozero operation, tare memory). This parameters are placed in <P1.rEAd> submenu.

12.1. Setting a filtering level Procedure:

Enter the submenu <P1.rEAd> and then:
1 - 4 - level of filtering
By pressing
select the filtering level you need.
- 29 -
Page 30
Notice:
Filtering level influences the time of stabilization. The higher the filtering level is the longer stabilization time is needed.
Return to weighing:
See - 10.2.2.

12.2. Median filter

This filter eliminates short changes (impulses) of measure signal (e.g. shocks).
Procedure:
Enter the submenu <P1.rEAd> and then:
Fnnd no - filter disabled Fnnd YES - filter enabled
Return to weighing:
See - 10.2.2.
- 30 -
Page 31

12.3. Autozero function

The autozero function has been implemented in order to assure precise indications. This function controls and corrects „0” indication. While the function is active it compares the results continuously with constant frequency. If two sequentional results differ less than the declared value of autozero range, so the scale will be automatically zeroed and the pictograms
and will be displayed.
When AUTOZERO is disabled zero is not corrected automatically. However, in particular cases, this function can disrupt the measurement process e.g. slow pouring of liquid or powder on the weighing pan. In this case, it is advisable to disable the autozero function.
Procedure:
Enter the submenu <P1.rEAd> and then:
Fnnd no - filter disabled Fnnd YES - filter enabled
Return to weighing:
See - 10.2.2.
- 31 -
Page 32

12.4. Tare function

This parameters enables users to configure a tare function.
Procedure:
Enter the submenu <P1.rEAd> and then:
tArA AtAr
-
automatic tare function on and is stored in balance memory after unplugging it from mains (Description of function operating point 15.6 automatic tare)
tArA
no
automatic tare function off (user can turn on
operating of automatic tare F6 AtAr – till unplugging
­the balance from mains)
tArA tArF
-
tare memory function – stores last value of tare in balance memory. It is automatically displayed after starting the balance. Value of tare is displayed with minus sign, and there is Net symbol indicated on the display. (user can turn on operating of automatic tare
F6 AtAr – till unplugging the balance from mains)
Return to weighing:
See - 10.2.2.
- 32 -
Page 33

13. RS 232 PARAMETERS

External devices connected to RS 232C have to be supplied from the same mains and common electric shock protection. It prevents from appearing a potential difference between zero leads of the two devices. This notice does not apply to the devices that do not use zero leads.
Transmission parameters:
Baud rate - 2400 – 38400 bit / s
Data bits - 7,8
Stop bits - 1,2
Parity control - no, even, odd.
There are four ways of sending data via RS232 interface:
Manually – after pressing
,
Automatically – after stabilizing the indication over LO threshold
Continuously – after it is activated in parameter or by a command
sent via RS232
On external request - see - „List of scale - computer commands”.
The indication can be sent as:
• stable – the indication is sent after the scale stabilizes.
any – the indication is sent immediately after pressing the
key,
this state is assign with <?> in the printout.

13.1. Printout type

This parameter is to select the type of printout.
Procedure:
Enter the submenu <P2.Prnt> and then:
- 33 -
Page 34
Pr_n noStAb
- immediate printout (not accessible in verified scales)
Pr_n StAb Pr_n rEPL Pr_n CntA Pr_n Cntb
- sending stable results
- automatic operation
- continuous transmission in basic unit
- continuous transmission in present unit
Return to weighing:
see 10.2.2.

13.2. Minimal mass threshold This function is necessary while working with automatic tare or

automatic operation or weighing animals. Automatic tarring will not be applied until the indication (gross)
is lower than the value inscribed in S_Lo parameter. In automatic operation measurements (net) are sent via RS232
when the indication is equal or greater than the value inscribed in S_Lo parameter.
Weighings animals is performer when the indication is equal or greater than the value inscribed in S_Lo parameter.
Procedure:
Enter the submenu <P2.Prnt> and then:
- 34 -
Page 35
Return to weighing:
see 10.2.2.

13.3. Baud rate Procedure:

Enter the submenu <P2.Prnt> and then:
- 35 -
Page 36
Return to weighing:
see 10.2.2.

13.4. Serial transmission parameters Procedure:

Enter the submenu <P2.Prnt> and then:
7d2SnP - 7 data bits; 2 stop bits, no parity control 7d1SEP - 7 data bits; 1 stop bit, EVEN parity control 7d1SoP - 7 data bits; 1 stop bit, ODD parity control 8d1SnP - 8 data bits; 1 stop bit, no parity control 8d2SnP - 8 data bits; 2 stop bits, no parity control 8d1SEP - 8 data bits; 1 stop bit, EVEN parity control 8d1SoP - 8 data bits; 1 stop bit, ODD parity control
Return to weighing:
See 10.2.2.
- 36 -
Page 37

14. OTHER PARAMETERS

The user can set parameters which influence the scale operation. They are gathered in the submenu <P5.othr> e.g. backlight and beep signal. Enter this submenu <P5.othr> according to chapter 11.2.

14.1. Backlight function

Program recognises the way the scale is supplied (mains, battery) and automatically selects the way of operating on the backlight:
bl – for mains, blbt – for batteries or rechargeable battery pack.

14.1.1. Backlight for supplying from mains Procedure:

Enter the submenu <P5.othr> and then:
bL no bL YES bL Auto
- backlight switched off
- backlight switched on
- backlight switched off automatically if indication becomes stable for about 10s
- 37 -
Page 38
Return to weighing:
See 10.2.2.
Notice:
When bl=Auto, and the indication has not changed for 10s, the backlight is automatically switched off. The backlight is switched on again automatically after the result changes.

14.1.2. Backlight for supplying from batteries

The user can change the intensity of backlight from 0% to 100%. The lower the intensity is the longer the scale operates without recharging or exchanging batteries. When the intensity is set this function works as AUTO (described above).
Procedure:
Enter the submenu <P5.othr> and then:
Return to weighing:
See 10.2.2.
Notice:
The more intense the backlight is the shorter the scale operates on batteries.
- 38 -
Page 39

14.2. “Beep” signal – after pressing a key Procedure:

Enter the submenu <P5.othr> and then:
bEEP no - switched off bEEP YES - switched on
Return to weighing:
See 10.2.2.

14.3. Automatic switch-off

This function is essential to save the battery power. The scale is switched off automatically when (function t1 = YES) no weighing appears in 5 minutes. (no changes on the display). In case when this function disrupts the operation (e.g. long time weighing procedures) or while working with connection to mains, switch off this function.
Operation according to the power supply:
Setting
t1 = 0 disabled disabled t1 = YES enabled enabled t1 = Auto * disabled enabled
Mains Batteries/accumulator
- 39 -
Operation
Page 40
* automatic enabling/disabling according to the source of power.
Procedure:
Enter the submenu <P5.othr> and then:
Returnto weighing:
See 10.2.2.

14.4. Battery voltage level check

While supplying from batteries too low level of voltage is measured by software the pictogram
is displayed. It means that charging
or exchanging batteries is required.

14.4.1. Checking the batteries

This function is to check the level of battery supply. It works only if:
Weighing mode is set,
Battery supply is set in parameters.
- 40 -
Page 41
Procedure:
After displaying the level of batteries (in per cents) the program returns to weighing.

14.4.2. Battery discharge pictogram

The symbol (bat low) switches on when the voltage level drops to 18% of the accepted level of voltage. It means that charging or exchanging batteries is required.
Low level of batteries:
Pictogram on the display,
After one time the device will automatically switch off
to protect the batteries from distructable discharging,
Charging is signalled by (blinking period about
2 seconds) on the display.

14.4.3. Accumulator charging option

This function allows to switch on charging algorithm for a NiMH battery pack (for indicators in plastic casings) or a gel cell SLA accumulator (for indicators in metal housings):
a) Parameter <CHr6> set to <no>:
Pictogram does not appear, charging disabled,
During software initializing, after turning on <bAtt>.
b) Parameter <CHr6> set to <YES>:
Pictogram blinks slowly (period about 2 seconds), charging is enabled,
- 41 -
Page 42
Message <nImh> appears on the display (for indicators in plastic casings) or <SLA> (for indicators in metal housings).
In case of damaging accumulators or lack of it the pictogram
blinks quickly (period about 0.5 sec).
Notice:
Indicators in plastic casings are equipped with the set of rechargeable batteries NiMH R6 (AA) and power adapter.
Procedure:
Enter the submenu <P5.othr> and then:
CHr6 YES - enabled CHr6 no - disabled
Return to weighing:
See 10.2.2.

14.4.4. Formatting rechargeable battery packs Every plastic indicator is equipped with a brand new NiMH R6 (AA)

battery pack and a power adapter. They need formatting after first powering up. It is crucial for batteries lifetime to undertake this process. Formatting consist in charging and total discharging (without meantime charging).
- 42 -
Page 43
Procedure:
1. Supply the indicator from mains.
2. Charge batteries for 12 hours (time of charging 2200mAh batteries).
3. After 12 hours unplug from mains.
4. Use the device up to the moment of self powering down.
5. Repeat the process of charging starting from point 1.
Notice:
They reach their optima capacity after three cycles of full charging and discharging.

15. WORK MODES

15.1. Setting accessibility of operation modes

In this parameter group users can disable/enable accessibility of functions after pressing
key.
Procedure:
Enter the submenu <P4.Func> and then:
- 43 -
Page 44
no – mode is disabled YES – mode is enabled
Return to weighing:
See 10.2.2.

15.2. Selecting quantity of operation modes

This function enables user to set if ,after pressing key, all operating modes will be accessible (ALL) or only one from the list chosen and used by operator.
Procedure:
Enter the submenu <P4.Func> and then:
After choosing setting press key. The program will return to displaying name of submenu <P4.1.FFun>.
Return to weighing:
See 10.2.2.
- 44 -
Page 45

15.3. Counting pieces of the same mass

Standard solution is equipped with option of counting small pieces of the same mass. It is possible to execute a tare function in this operating mode in order to tare a container value.
Notice:
1. Counting pieces does not work together with other scale functions,
2. The counting pieces function is not saved as a default start function so it is not remembered after restarting.
Procedure:
Enter to <PcS> function:
You will see a blinking value of sample quantity.
Press
key to start setting quantity of sample, you have a few
options to chose from:
- 45 -
Page 46
If option <LASt> is choosen in the scale program displays estimated unit mass of the last piece (about 3 seconds) and then goes to Counting pieces automatically setting the previously displayed value as valid for the procedure.
If the <FrEE> option is selected you will see:
Using and enter the required sample quantity,
where:
Confirm the value by pressing
- selection of digit position, - setting the digit,
,
You will see <LoAd> on the display and then:
If weighing is performed in a container put the container on the pan
first and then tare it. Then put the declared quantity of pieces on the pan and confirm it when stable (signalled by
):
The program will automatically calculate the mass of a single piece and go on to the Piece Counting mode (pcs). You will see the following display:
- 46 -
Page 47
Notice:
1. If a user presses the
key when load is not present on the pan, the message -Lo- will be indicated for a few seconds and the scale will automatically return to weighing.
2. In order to comply with the rules of appropriate counting pieces put as many pieces as possible during unit mass adjustment. Single piece mass should not be less than 5 divisions.
3. If a single piece mass is lower than a reading interval d the display will show the <Err5> message (see ch. 20. Error messages) and short audible signal will be emitted than the scale returns to weighing.
Return to weighing:
Press the
key twice.

15.4. +/- control referring to the inscribed standard mass Procedure:

Enter to <HiLo> function:
The program enters the window of setting the lower threshold
of weighing (Min):
- 47 -
Page 48
The inscribed value confirm by pressing , the program will automatically go to the higher threshold of weighing (Max):
The inscribed value confirm by pressing , the program will automatically go to the main window.
During setting threshold values following cases take place:
Notice:
If a user erroneously enters a value of the lower threshold higher than the upper one, the scale will indicate an error message and will return to weighing.
- 48 -
Page 49
Return to weighing:
Press the
key twice.

15.5. Control of % deviation referring to the inscribed standard mass

Scale software enables control of deviation (in %) of weighed loads mass referring to the inscribed standard mass. Mass of standard can be determined by its weighing (PrcA function) or entered to the scale memory by an user (PrcB function).

15.5.1. Standard mass determined by its weighing Procedure:

Enter to <PrcA> function:
You will see <LoAd> on the display and then:
place an load on the pan which mass will be accepted as standard
press
to confirm this operating mode
after few seconds the indication 100,00% will be displayed
From this moment display will not indicate mass of weighed load
but deviation of load mass placed on the pan referring to the mass of standard (in %).
- 49 -
Page 50
Return to weighing:
Press the
key twice.

15.5.2. Mass of standard inscribed to scale memory Procedure:

Enter to <PrcB> function:
The program goes to the weight display window:
Using and set standard mass,
where:
Confirm the entered value by pressing
- digit selection, - digit setting.
,
You will see the indication equal to 0,000%,
From this moment display will not indicate the mass of weighed
load but deviation of the load mass placed on the pan referring mass of standard (in
%).
- 50 -
Page 51
Return to weighing:
Press the
key twice.

15.6. Automatic tare

This function is useful for fast net mass determination of weighed load in case when tare value of is different for each load. In case when the function is active the cycle of scales operating looks as follows:
press zeroing key when the pan is empty,
place the container for pieces,
when indication is stable automatic tarring of the container mass will
be performed (Net marker will appear in the upper part of the display),
place a sample into the package,
display will indicate net mass of sample,
remove the sample together with the container,
display will indicate tare mass with minus sign,
place a container for the next sample. When indication is stable
automatic tarring will take place (Net marker will appear in the upper part of the display),
place next sample into the package.
Procedure:
Return to weighing:
Press the
key twice.
- 51 -
Page 52

15.7. Measurement max force on the pan – latch Procedure:

Enter to <toP> function:
Confirmation of choice of <toP> function is indication of the Max
pictogram:
Apply a force to the weighing pan.
The display of scale will latch the maximum value of the force
Remove loads from the pan
Before the next measurement press the
key.
Return to weighing:
Press the
key twice.

15.8. Totalizing

Scale software is equipped in a totalizing function of single weighings. The totalizing procedure can be documented on the printer connected to the indicator.
- 52 -
Page 53

15.8.1. Enabling the work mode Procedure:

Enter to <Add> function:
A letter „P” in the left side of the display is a confirmation that
<Add> function have been selected:

15.8.2. Totalizing procedure

Enter <Add> function according to ch. 16.8.1,
Put the first load on the pan. If the weighing procedure is performed
in a container put the container on the pan first and tare it. Then put the first load on the pan and confirm it by pressing
(signalled by
),
when stable
You will see a sum of weighings on the display, the „▲” pictogram in the upper right corner will be displayed and the weighing result will be printed on the printer connected to the indicator.
- 53 -
Page 54
Take off the load from the pan, indication returns to ZERO and the letter „P” in the left part of the display appears,
Put the next load on the pan,
After stabilizing press
, the sum of first and second weighing will appear on the display, the „” pictogram in the upper right corner will be displayed and the second weighing result will be printed on the printer connected to the indicator:
Press
to complete the procedure (with the loaded or unloaded
pan), a sum of all weighings will be printed:
(1) 1.912 kg (2) 1.912 kg
------------------------ TOTAL: 3.824 kg
In case of pressing
one more time with loaded pan, you will see the <unLoAd> message. Unload the pan, the scale will return to ZERO and the letter „P” in the left part of the display will appear. The scale is ready for the next procedure.
In case of pressing
one more time with loaded pan, you will see the letter „P” in the left part of the display will appear. The scale is ready for the next procedure.

15.8.3. Memory of the last value of sum of weighed goods

After interrupting (e.g. switching off) the totalizing procedure, it is possible to restart the procedure without loosing data. In order to do it just enter the totalizing procedure:
Enter <Add> function again according to the ch. 16.8.1 of the manual,
You will see the last memorized sum of weighings on He display.
- 54 -
Page 55
- In order to continue the procedure press , the indication returns to ZERO and the letter „P” appears in the left part of the display. The scale is ready for weighing;
- In order to terminate the previous totalizing procedure press key,
, or . You will see the letter „P” in the left part of
the display. The scale is ready for weighing.

15.8.4. Return to weighing

Press
key, you will see:
Before leaving the <Add> function it is possible to print out subsequent weighings and the sum of weighings on the
connected printer (press
to print, press to cancel),
The following message will appear on the display:
Press key to return to weighing,
Press
to return to totalizing.
Notice:
In case of overflow of the range of the display in totalizing you will see <5-FULL> message in the display. In that case unload the pan and
press
to complete the procedure with a printout of sum of all weighings or put a lower mass on the pan which does not cause the overflow error.
- 55 -
Page 56

15.9. Weighing animals Procedure:

Enter to <AnLS> function:
The <tinnE> message appears on the display for 1s, and then the
program goes to the window of setting the duration time (in seconds) of the animal weighing process:
Confirm the selected value by pressing ,
You will see the following window:
- 56 -
Page 57
Load an animal to the platform,
After exceeding the -LO- value (see 13.2), program starts the
weighings process. The appearance of subsequent hyphens < - - - - - - - > showing the progress,
After completing the process of weighings the result is latched on the display and additionally the OK pictogram is shown in the upper part of the display:
You can start the procedure of weighing animals again by pressing
,
After removing the animal from the platform program returns to the window:
Return to weighing:
Press .

15.10. Tare memory

Users are allowed to Enter Up to 9 tare values to the memory.

15.10.1. Entering the tare value to the scale memory Procedure:

Enter to <tArE> function:
- 57 -
Page 58
The program goes to displaying the first value from the selection of tare values <tArE 0> (press
to chose different values):
After selecting the right position press and you will see an editing field:
Enter the selected tare value to the scale memory ,
The program returns to the following window:
- 58 -
Page 59
Return to weighing:
Press .

15.10.2. Selecting a tare value from the memory

Enter <tArE> function according to the ch. 16.10.1 of the manual,
The program goes to displaying the first value from the selection
of tare values <tArE 0> (press
to chose different values):
To use an entered tare value press , you will see the tare value on the display preceded by the „-„ sign and the Net pictogram:
Caution:
A tare value from the memory is not remembered after powering off and on the scale.
- 59 -
Page 60

16. USER CALIBRATION

Only for non-verified scales
Confirmation of high accuracy of weighing requires periodical correcting of calibration factors in the scale memory – this is adjustment of the scale. Calibration should be performed when we start weighing or dynamic change of temperature occurs. Before starting calibration remove loads from the pan.

16.1. Calibration Procedure:

Enter the submenu <P6.CAL> and then:
Following inscriptions will appear
A new start mass is adjusted during this period of time.
After that a mass of calibration weight is shown (e.g. 3 000kg).
- 60 -
Page 61
Put a weight of the displayed mass value on the pan and press . The calibration process will start which is signalled by the message:
After completion of the process of calibration the following screen will appear
Take off the weight , then the following sequence of screens will appear
Calibration process can be terminated anytime by pressing which is signalled by the following message on the display:
Return to weighing with saving changes that have been made.
Caution:
If the calibration process (span adjustment) lasts longer than 15 the <Err8> message will be displayed and short audible signal will be
emitted. Press
to perform calibration again with more stable
ambient conditions!
- 61 -
Page 62

16.2. Start mass adjustment

If the scale does not require the full calibration process sit is possible to adjust only a new start mass.
Procedure:
Enter the submenu <P6.CAL> and then:
The display will show the following information
After the completion of the start mass adjustment the following
screen will appear:
The process of start mass adjustment can be terminated by pressing
, which is signalled on the display:
- 62 -
Page 63
Return to weighing performing the procedure of saving parameters.
Caution:
If the start mass adjustment lasts longer than 15 the <Err8> message will be displayed and short audible signal will be emitted. Press
to perform calibration again with more stable ambient conditions!

17. COOPERATION WITH PRINTER

Each time the mass units is sent to RS 232 interface.
Depending on setting of STAB parameter it can be printed out with temporary or stable value. Depending on setting of REPL parameter, printout will be automatic or manual.
One of thermal printer in KAFKA series can cooperate with each platform scales:
a) KAFKA
Only result of weighing with mass unit can be printed.
b) KAFKA 1/Z
This printer is equipped with an internal real time clock. Both date and time can be printed.
c) KAFKA SQ S
This printer is equipped with an internal real time clock and possibility of running statistics from measurements. Statistic contents: quantity of samples, sum of masses of all samples, average value, standard deviation, variation factor, min value, max value, difference max - min.
key is pressed a current mass value together with
- 63 -
Page 64
Cable diagrams:
Scale – Kafka printer cable diagram for plastic casing
Scale – Kafka printer cable diagram for steel housing

18. COOPERATION WITH COMPUTER

Sending weighing results to the computer can be done:
- manually
- in continuous way
- after pressing
- after function activating or sending an appropriate command,
- automatically
- After stabilizing the indication
- on the request from the computer
- After sending a control command
These scales can cooperate with „EDYTOR WAG” program. The indicator window comprises the most important information from the scale display. The program allows to configure easily, e.g. design printouts, edit parameters. A precise description is issued in the „Help” file that accompanies the program.
key,
- 64 -
Page 65
Cable diagrams:
Scale – computer cable diagram for plastic casing
Scale – computer cable diagram for metal housing

19. COMMUNICATION PROTOCOL

19.1. General information

A. A character protocol scale-terminal has been designed for
communication between RADWAG scales and external devices via RS-232 interface.
B. It consists of commands sent from an external device to
the scale and a responses from a scale.
C. Responses are sent every time after receiving a command
(reaction for any command).
D. Using commands allows users to receive some information
about the state of scale and/or influence the operation e.g.:
Requesting weighing results,
Display control.
- 65 -
Page 66

19.2. A set of commands for RS interfaces

Commands
Z T TO S SI SU SUI C1 C0 CU1 CU0 PC
Description of commands
Zeroing Tarring Get tare Send the stable result in basic unit Send the result immediately in basic unit Send the stable result in current unit Send the result immediately in current unit Switch on continuous transmission in basic unit Switch off continuous transmission in basic unit Switch on continuous transmission in current unit Switch off continuous transmission in current unit Send all implemented commands
Notice:
1. Each command have to be terminated in CR LF;
2. The best Policy for communication is not sending another command until the former answer has been received.

19.3. Respond message format

After sending a request message you can receive:
XX_A CR LF XX_D CR LF XX_I CR LF XX _ ^ CR LF XX _ v CR LF XX _ OK CR LF ES_CR LF
XX _ E CR LF
command accepted and in progress command completed (appears only after XX_A) command comprehended but cannot be executed command comprehended but time overflow error appeared command comprehended but the indication below the Command done Command not comprehended error while executing command – time limit for stable result
exceeded (limit time is a descriptive parameter of the scale)
- 66 -
Page 67
XX - command name _ - substitutes spaces

19.4. Command’s description

19.4.1. Zeroing

Syntax Z CR LF Possible answers:
Z_A CR LF Z_D CR LF
Z_A CR LF Z_^ CR LF
Z_A CR LF Z_E CR LF
Z_I CR LF
- command accepted and in progress
- command completed
- command accepted and in progress
- command comprehended but zero range overflow appeared
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed

19.4.2. Tarring

Syntax: T CR LF Possible answers:
T_A CR LF T_D CR LF
- command accepted and in progress
- command completed
T_A CR LF T_v CR LF
T_A CR LF T_E CR LF
T_I CR LF
- command accepted and in progress
- command comprehended but tare range overflow appeared
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed

19.4.3. Get tare value

Syntax: TO CR LF Possible answers:
- 67 -
Page 68
TO_TARA CR LF
- command executed
Frame format:
1 2 3 4 5-6 7-15 16 17 18 19 20 21 T O space stability space tare space unit CR LF
Tare - 9 characters with decimal point justified to the right Unit - 3 characters justified to the left

19.4.4. Send the stable result in basic unit

Syntax: S CR LF Possible answers:
S_A CR LF S_E CR LF
S_I CR LF S_A CR LF
MASS FRAME
- command accepted and in progress
- time limit for stable result exceeded
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in basic unit is returned
Frame format:
1 2-3 4 5 6 7-15 16 17 18 19 20 21
S space stability space sign mass space unit CR LF
Example: S CR LF – computer command
S _ A CR LF - command accepted and in progress S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.

19.4.5. Send the result immediately in basic unit

Syntax: SI CR LF Possible answers:
- 68 -
Page 69
SI_I CR LF
MASS FRAME
- command comprehended but cannot be executed at the moment
- mass value in basic unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space stability space sign mass space unit CR LF
Example: S I CR LF – computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass value
in basic unit is returned immediately.

19.4.6. Send the stable result in current unit

Syntax: SU CR LF Possible answers:
SU_A CR LF SU_E CR LF
- command accepted and in progress
- timeout while waiting for stable results
SU_I CR LF SU_A CR LF
MASS FRAME
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in current unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U space stability space sign mass space unit CR LF
Example: S U CR LF – computer command
S U _ A CR LF - command accepted and in progress S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass value
in current unit is returned.
- 69 -
Page 70

19.4.7. Send the result immediately in current unit

Syntax: SUI CR LF Possible answers:
SUI_I CR LF MASS FRAME
- command comprehended but cannot be executed
- mass value in current unit is returned immediately
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF
Example: S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed
and mass returned

19.4.8. Switch on continuous transmission in basic unit

Syntax: C1 CR LF Possible answers:
C1_I CR LF C1_A CR LF
MASS FRAME
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21 S I space stability space sign mass space unit CR LF

19.4.9. Switch off continuous transmission in basic unit

Syntax: C0 CR LF
- 70 -
Page 71
Possible answers:
C0_I CR LF C0_A CR LF
- command comprehended but cannot be executed
- command comprehended and executed

19.4.10. Switch on continuous transmission in current unit

Syntax: CU1 CR LF Possible answers:
CU1_I CR LF CU1_A CR LF
MASS FRAME
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF

19.4.11. Switch off continuous transmission in current unit

Syntax: CU0 CR LF Possible answers:
CU0_I CR LF CU0_A CR LF
- command comprehended but cannot be executed
- command comprehended and executed

19.4.12. Send all implemented commands

Syntax: PC CR LF Possible answers: PC_- >_Z,T,TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC – command executed,
the indicator have sent all the implemented commands.
- 71 -
Page 72

19.5. Manual printouts / automatic printouts

Users can general manual or automatic printouts from the scale.
Manual printouts can be performed after loading the pan and stabilizing indication by pressing
.
Automatic printouts can be performed only after loading the pan and stabilizing indication.
Notice:
If a scale is verified printouts of immediate values are blocked.
Format frame:
1 2 3 4 -12 13 14 15 16 17 18
stability space sign mass space unit CR LF
Stability character
[space] if stable [?] if not stable [^] if an indication over the range [v] if fan indication below the range
sign
[space] for positive values or [-] for negative values
mass unit command
9 characters justified to the right 3 characters justified to the left 3 characters justified to the left
Example 1: _ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF – the printout generated from
the scale after pressing ENTER/PRINT.
Example 2: ? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.
Example 3: ^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - the printout generated from
the scale after pressing ENTER/PRINT.
- 72 -
Page 73

19.6. Continuous transmission

The indicator can work in a continuous transmission mode. It can be switched on or off in parameters or using RS232 commands.
The frame format sent by the indicator in case of setting <P2.Prnt> to CntA:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space stability space sign mass space Unit CR LF
Stability character
[space] if stable [?] if not stable [^] if an indication over the range [v] if fan indication below the range
sign
[space] for positive values or [-] for negative values
mass unit command
9 characters justified to the right 3 characters justified to the left 3 characters justified to the left
The frame format sent by the indicator in case of setting <P2.Prnt> to Cntb:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF

19.7. Configuring printouts General information

If some information included are redundant or not sufficient and there is a necessity of changes one can design their own protocol format in EDYTOR WAG computer program. This piece of software is accessible in: http://www.radwag.com
- 73 -
Page 74

20. ERROR COMMANDS

Err2 Err3 Err4
Err5 Err8
NULL FULL2 LH
5-FULL
Notice:
1. Errors: Err2, Err3, Err4, Err5, Err8, null, that appear on the display are also signalled by a short beep sound (about 1 sec.);
2. Error FULL2 that appears on the display is also signalled by a continuous sound until the cause of error disappears.
- Value beyond the zero range
- Value beyond the tare range
- Calibration mass or start mass beyond the acceptable range (±1% for weight, ±10 for start mass)
- Mass of a single piece lower than the scale division
- Exceeded the time for tarring, zeroing, start mass adjustment or span adjustment
- Zero value from the AD converter
- Measurement range overflow
- Start mass error, the mass on the weighing platform is beyond the acceptable range (-5% to +15% of start mass)
- Display range overflow in totalizing
- 74 -
Page 75

21. TECHNICAL PARAMETERS

21.1. Scales of WPT series

WPT 6 C1
Scale type:
Max capacity 6kg 15kg 30kg 30kg Min capacity 40g 100g 200g 200g Readability 2g 5g 10g 10g Verification division 2g 5g 10g 10g Tare range -6kg -15kg -30kg -30kg Platform size 290 × 360mm 400 x 500mm OIML class III Operation temperature -10°C to +40°C Interfaces RS 232 Power supply 230 V 50Hz/11V AC, and 6×AA NiMH Supplied from batteries Average operation when supplied from batteries 35h Ingress protection rating IP 43 - plastic Display LCD with backlight Net / Gross weight 6,5 / 7,8kg 15,5 / 17,8kg Package dimensions 550 x 420 x 220mm 720 x 580 x 220mm
WPT
6 C1/K
WPT
6 C1/R
WPT
15 C1
WPT
15 C1/K
WPT
15 C1/R
WPT
30 C1
WPT
30 C1/K
WPT
30 C1/R
WPT 30
C2
WPT 30
C2/K
WPT 30
C2/R
WPT
60 C2
Scale type:
Max capacity 60kg 150kg 300kg 150kg 300kg Min capacity 400g 1000g 2000g 1000g 2000g Readability 20g 50g 100g 50g 100g Verification division 20g 50g 100g 50g 100g Tare range -60kg -150kg -300kg -150kg -300kg Platform size 400 × 500mm 500 × 700mm OIML class III Operation temperature -10°C to +40°C Interfaces RS 232 Power supply 230 V 50Hz/11V AC, and 6×AA NiMH Supplied from batteries Average operation when supplied from batteries 35h Ingress protection rating IP 43 - plastic Display LCD with backlight Net / Gross weight 15,5 / 17,8kg 20,5 / 26,8kg Package dimensions 720 x 580 x 220mm 820 x 640 x 260mm
WPT 60
C2/K
WPT 60
C2/R
WPT
150 C2
WPT 150
C2/K
WPT 150
C2/R
WPT
300 C2
WPT 300
C2/K
WPT 300
C2/R
WPT
150 C3
WPT 150
C3/K
WPT
300 C3
WPT 300
C3/K
- 75 -
Page 76
Notice:
/K designation describes a scale with an indicator with the cable, /R signifies a scale with an indicator integrated with the scale platform.
WPT 3/6 C1 WPT 6/15 C1 WPT 15/30 C1
Scale type:
Max capacity 3/6kg 6/15kg 15/30kg Min capacity 20/40g 40/100g 100/200g Readability 1/2g 2/5g 5/10g Verification division 1/2g 2/5g 5/10g Tare range -6kg -15kg -30kg Platform size 290 × 360mm OIML class III Operation temperature -10°C to +40°C Interfaces RS 232 Power supply 230 V 50Hz/11V AC, and 6×AA NiMH Supplied from batteries Average operation when supplied from batteries 35h Ingress protection rating IP 43 - plastic Display LCD with backlight Net / Gross weight 6,5 / 7,8kg Package dimensions 550 x 420 x 220mm
Scale type:
Max capacity 15/30kg 30/60kg 60/150kg 150/300kg Min capacity 100/200g 200/400g 400/1000g 1000/2000g Readability 5/10g 10/20g 20/50g 50/100g Verification division 5/10g 10/20g 20/50g 50/100g Tare range -30kg -60kg -150kg -300kg Platform size 400 × 500mm OIML class III Operation temperature -10°C to +40°C Interfaces RS 232 Power supply 230 V 50Hz/11V AC, and 6×AA NiMH Supplied from batteries Average operation when supplied from batteries 35h Ingress protection rating IP 43 - plastic Display LCD with backlight Net / Gross weight 15,5 / 17,8kg Package dimensions 720 x 580 x 220mm
WPT 3/6 C1/K WPT 6/15 C1/K WPT 15/30 C1/K WPT 3/6 C1/R WPT 6/15 C1/R WPT 15/30 C1/R
WPT 15/30
C2
WPT 15/30
C2/K
WPT 15/30
C2/R
WPT 30/60
C2
WPT 30/60
C2/K
WPT 30/60
C2/R
WPT 60/150
C2
WPT 60/150
C2/K
WPT 60/150
C2/R
WPT 150/300
WPT 150/300
C2/K
C2
- 76 -
Page 77

21.2. Table scales of WPT/F series

WPT/F 3C WPT/F 6C WPT/F 15C WPT/F 30C
Scale type:
Max capacity 3kg 6kg 15kg 30kg Min capacity 20g 40g 100g 200g Readability 1g 2g 5g 10g Verification division 1g 2g 5g 10g Range of tare -3kg -6kg -15kg -30kg Platform size 300 x 300mm OIML class III Operation temperature -10°C to +40°C Interfaces RS 232 Ingress protection rating IP 43 - plastic Power supply 230 V 50Hz/11V AC, and 6×AA NiMH Supplied from batteries Average operation when supplied from batteries 35h Display LCD with backlight Net / Gross weight 5,5 / 6,5kg Package dimensions 410 x 380 x 210mm
WPT/F 3C/K WPT/F 6C/K WPT/F 15C/K WPT/F 30C/K WPT/F 3C/R WPT/F 6C/R WPT/F 15C/R WPT/F 30C/R
WPT/F 3/6C WPT/F 6/15C WPT/F 15/30C/K
Scale type:
Max capacity 3/6kg 6/15kg 15/30kg Min capacity 20/40g 40/100g 100/200g Readability 1/2g 2/5g 5/10g Verification division 1/2g 2/5g 5/10g Range of tare -6kg -15kg -30kg Platform size 300 x 300mm OIML class III Operation temperature -10°C to +40°C Interfaces RS 232 Power supply 230 V 50Hz/11V AC, and 6×AA NiMH Supplied from batteries Average operation when supplied from batteries 35h Ingress protection rating IP 43 - plastic Display LCD with backlight Net / Gross weight 5,5 / 6,5kg Package dimensions 410 x 380 x 210mm
WPT/F 3/6C/K WPT/F 6/15C/K WPT/F 15/30C WPT/F 3/6C/R WPT/F 6/15C/R WPT/F 15/30R
Notice:
/K designation describes a scale with an indicator with the cable, /R signifies a scale with an indicator integrated with the scale platform.
- 77 -
Page 78

21.3. Waterproof scales of WPT/H series

WPT
Scale type:
Max capacity 3kg 6kg 15kg 15kg 30kg Min capacity 20g 40g 100g 100g 200g Readability 1g 2g 5g 5g 10g Verification division 1g 2g 5g 5g 10g Tare range -3kg -6kg -15kg -15kg -30kg Platform size 200×150mm 250 × 300mm 410 × 410mm OIML class III Work temperature -10°C to +40°C Interfaces RS 232 Ingress protection rating IP 67 - platform, IP 66/67 - indicator Power supply 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 7 / 8,3kg 9 / 10,3kg 15,5 / 17,3kg Package dimensions 520x260x290mm 580x320x360mm 670x510x330mm
3H1
WPT
3H1/K
WPT
6H2
WPT
6H2/K
WPT 15H2
WPT
15H2/K
WPT
15H3
WPT
15H3/K
WPT
30H3 WPT
30H3/K
WPT
Scale type:
Max capacity 60kg 150kg 60kg 150kg Min capacity 400g 1000g 400g 1000g Readability 20g 50g 20g 50g Verification division 20g 50g 20g 50g Tare range -60kg -150kg -60kg -150kg Platform size 410 × 410mm 500 × 500mm OIML class III Work temperature -10°C to +40°C Interfaces RS 232 Ingress protection rating IP 67 - platform, IP 66/67 – indicator Power supply 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 15,5 / 17,3kg 23,5 / 25,8kg Package dimensions 670x510x330mm 520x260x290mm
60H3 WPT
60H3/K
WPT
150H3
WPT
150H3/K
WPT
60H4
WPT
60H4/K
WPT
150H4
WPT
150H4/K
- 78 -
Page 79
Scale type:
Max capacity 60kg 150kg 300kg 150kg 300kg Min capacity 400g 1000g 2000g 1000g 2000g Readability 20g 50g 100g 50g 100g Verification division 20g 50g 100g 50g 100g Tare range -60kg -150kg -300kg -150kg -300kg Platform size 600 × 600mm 800 × 800mm OIML class III Work temperature -10°C to +40°C Interfaces RS 232 Ingress protection rating IP 67 - platform, IP 66/67 - indicator Power supply 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 29,5 / 31,8kg 42,5 / 45,8kg Package dimensions 840x700x400mm 1160x820x340mm
WPT 60H5
WPT
60H5/K
WPT
150H5
WPT
150H5/K
WPT
300H5
WPT
300H5/K
WPT
150H6
WPT
150H6/K
WPT
300H6
WPT
300H6/K
WPT
Scale type:
Max capacity 1,5/ 3kg 3/6kg 6/15kg 6/15kg 15/30kg Min capacity 10/20g 20/40g 40/100g 40/100g 100/200g Readability 0,5/1g 1/2g 2/5g 2/5g 5/10g Verification division 0,5/1g 1/2g 2/5g 2/5g 5/10g Tare range -3kg -6kg -15kg -15kg -30kg Platform size 200×150mm 250 × 300mm 410 × 410mm OIML class III Work temperature -10°C to +40°C Interfaces RS 232 Ingress protection rating IP 67 - platform, IP 66/67 - indicator Power supply 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 7 / 8,3kg 9 / 10,3kg 15,5 / 17,3kg Package dimensions 520x260x290 580x320x360mm 670x510x330mm
1,5/3H1
WPT
1,5/3H1/K
WPT
3/6H2
WPT
3/6H2/K
WPT
6/15H2
WPT
6/15H2/K
WPT
6/15H3
WPT
6/15H3/K
WPT
15/30H3
WPT
15/30H3/K
- 79 -
Page 80
Scale type:
Max capacity 30/60kg 60/150kg 30/60kg 60/150kg Min capacity 200/400g 400/1000g 200/400g 400/1000g Readability 10/20g 20/50g 10/20g 20/50g Verification division 10/20g 20/50g 10/20g 20/50g Tare range -60kg -150kg -60kg -150kg Platform size 410 × 410mm 500 × 500mm OIML class III Work temperature -10°C to +40°C Interfaces RS 232 Ingress protection rating IP 67 - platform, IP 66/67 - indicator Power supply 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 15,5 / 17,3kg 23,5 / 25,8kg Package dimensions 670x510x330mm 840x600x400mm
WPT
30/60H3
WPT
30/60H3/K
WPT
60/150H3
WPT
60/150H3/K
WPT
30/60H4
WPT
30/60H4/K
WPT
60/150H4
WPT
60/150H4/K
WPT
30/60H5
Scale type:
WPT
30/60H5/K
Max capacity 30/60kg 60/150kg 150300kg 150300kg Min capacity 200/400g 400/1000g 1000/2000g 1000/2000g Readability 10/20g 20/50g 100g 100g Verification division 10/20g 20/50g 100g 100g Tare range -60kg -150kg -300kg -300kg Platform size 600 × 600mm 800 × 800mm OIML class III Work temperature -10°C to +40°C Interfaces RS 232 Ingress protection rating IP 67 - platform, IP 66/67 - indicator Power supply 230V AC, 50Hz / 11V AC and internal gell cell SLA 6V accumulator Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 29,5 / 31,8kg 42,5 / 45,8kg Package dimensions 840x700x400mm 1160x820x340mm
WPT
60/150H5
WPT
60/150H5/K
WPT
150/300H5
WPT
150/300H5/K
WPT
150/300H6
WPT
150/300H6/K
- 80 -
Page 81

21.4. Waterproof scales of WPT/HR series

WPT
Scale type:
Max capacity 3kg 6kg 15kg 15kg 30kg Min capacity 20g 40g 100g 100g 200g Readability 1g 2g 5g 5g 10g Verification division 1g 2g 5g 5g 10g Tare range -3kg -6kg -15kg -15kg -30kg Platform size 250 × 300mm 410 × 410mm OIML class III Work temperature -10°C to +40°C Storage temperature -25° to +70°C Interfaces RS 232 Ingress protection rating IP 68 - construction, IP 68 – tensometer, IP 66/67 - indicator Power supply 220÷240VAC 50Hz (optional 110÷120VAC 60Hz) and SLA 6V/3,4Ah Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 9/10,3kg 15,5/17,3kg Package dimensions 580x320x360mm 670x510x330mm
3HR2
WPT
3HR2/K
WPT
6HR2
WPT
6HR2/K
WPT
15HR2
WPT
15HR2/K
WPT
15HR3
WPT
15HR3/K
WPT
30HR3
WPT
30HR3/K
WPT
Scale type:
Max capacity 60kg 150kg 60kg 150kg 60kg 150kg Min capacity 400g 1000g 400g 1000g 400g 1000g Readability 20g 50g 20g 50g 20g 50g Verification division 20g 50g 20g 50g 20g 50g Tare range -60kg -150kg -60kg -150kg -60kg -150kg Platform size 410 × 410mm 500 × 500mm 600 x 600mm OIML class III Work temperature -10°C to +40°C Storage temperature -25° to +70°C Interfaces RS 232 Ingress protection
rating Power supply 220÷240VAC 50Hz (optional 110÷120VAC 60Hz) and SLA 6V/3,4Ah Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 15,5/17,3kg 15,5/17,3kg 23,5 / 25,8kg Package dimensions 670x510x330mm 670x510x330mm 840x600x400mm
60HR3
WPT
60HR3/K
WPT
150HR3
WPT
150HR3/K
IP 68 - construction, IP 68 – tensometer, IP 66/67 - indicator
WPT
60HR4
WPT
60HR4/K
WPT
150HR4
WPT
150HR4/K
WPT
60HR5
WPT
60HR5/K
WPT
150HR5
WPT
150HR5/K
- 81 -
Page 82
Scale type:
Max capacity 3/6kg 6/15kg 15/30kg 30/60kg 60/150kg Min capacity 20/40g 40/100g 100/200g 200/400g 400/1000g Readability 1/2g 2/5g 5/10g 10/20g 20/50g Verification division 1/2g 2/5g 5/10g 10/20g 20/50g Tare range -6kg -15kg -30kg -60kg -150kg Platform size 250 × 300mm 410 × 410mm OIML class III Work temperature -10°C to +40°C Storage temperature -25° to +70°C Interfaces RS 232 Ingress protection rating IP 68 - construction, IP 68 – tensometer, IP 66/67 - indicator Power supply 220÷240VAC 50Hz (optional 110÷120VAC 60Hz) and SLA 6V/3,4Ah Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 9/10,3kg 15,5/17,3kg Package dimensions 580x320x360mm 670x510x330mm
WPT
3/6HR2
WPT
3/6HR2/K
WPT
6/15HR2
WPT
6/15HR2/K
WPT
15/30HR3
WPT
15/30HR3/K
WPT
30/60HR3
WPT
30/60HR3/K
WPT
60/150HR3
WPT
60/150HR3/K
WPT
Scale type:
Max capacity 30/60kg 60/150kg 30/60kg 60/150kg Min capacity 200/400g 400/1000g 200/400g 400/1000g Readability 10/20g 20/50g 10/20g 20/50g Verification division 10/20g 20/50g 10/20g 20/50g Tare range -60kg -150kg -60kg -150kg Platform size 500 x 500mm 600 x 600mm OIML class III Work temperature -10°C to +40°C Storage temperature -25° to +70°C Interfaces RS 232 Ingress protection rating IP 68 - construction, IP 68 – tensometer, IP 66/67 - indicator Power supply 220÷240VAC 50Hz (optional 110÷120VAC 60Hz) and SLA 6V/3,4Ah Supplied from batteries Average operation when supplied from batteries 45h Display LCD with backlight Net / Gross weight 23,5 / 25,8kg 29,5 / 31,8kg Package dimensions 840x600x400mm 840x700x400mm
30/60HR4
WPT
30/60HR4/K
WPT
60/150HR4
WPT
60/150HR4/K
WPT
30/60HR5
WPT
30/60HR5/K
WPT
60/150HR5
WPT
60/150HR5/K
- 82 -
Page 83

22. TROUBLE SHOOTING

Problem Cause Solution
Connect to mains or change batteries
Check the correctness of installation (polarization)
In „othr” submenu change „5.4 t1” to „no”
Unload the pan. Then the scale will indicator zero.
Turning on does not work
The scale turns off automatically
After turning on „LH” message on the display
Discharged batteries.
No batteries (not installed or improperly installed)
„t1” set to „YES” (Power save)
Loaded weight pan during powering up

23. ADDITIONAL EQUIPMENT

Accessories:
KAFKA printer cable for PUE C/31 indicators - P0136,
KAFKA printer cable for PUE C/31H, PUE C/31H/Z - P0253,
Computer cable for PUE C/31 - P0108,
Computer cable for PUE C/31H, PUE C/31H/Z - P0259,
EPSON printer cable for PUE C/31 - P0151,
EPSON printer cable for PUE C/31H, PUE C/31H/Z - P0261,
Power cord for car lighter 12V DC for PUE C/31 - K0047 ,
• Power cord for car lighter 12V DC for PUE C/31H/Z - K0042,
Thermal printer - KAFKA,
Dot matrix printer - EPSON,
Additional display in plastic casing for PUE C/31- WD- 4/1
(accessible with balance as complete set only),
Additional display in stainless metal housing for PUE C/31H, PUE C/31H/Z - WD- 4/3 (accessible with balance as complete set only),
Large size display (2”) for PUE C/31H, PUE C/31H/Z - WWG-2,
• Current loop in plastic casing for PUE C/31 - AP2-1,
• Current loop in metal housing PUE C/31H, PUE C/31H/Z - AP2-3
(accessible with balance as complete set only),
- 83 -
Page 84
RS232 / RS485 converter for PUE C/31 – KR-01,
RS232 / Ethernet converter for PUE C/31 - KR-04,
Stainless steel anti-vibration table - SAP/N,
A case for save carring/transporting a scale of WPT/C1/K series – W2,
Anti-dust case for Epson printer,
A rack for PUE C/31, PUE C/31H or PUE C/31H/Z indicator,
Handle for measuring indicator in plastic version,
A table for a scale (3 versions, for WPT/H3, WPT/H4 and WPT/H5),
A frame for weighings loads under a scale of WPT/F series,
Roller table.
Computer programs:
“EDYTOR WAG” computer program,
"RAD-KEY" computer program,
"PW-WIN" computer program.
- 84 -
Page 85
MANUFACTURER
OF ELECTRONIC WEIGHING
INSTRUMENTS
RADWAG WAGI ELEKTRONICZNE
26 – 600 Radom, Bracka 28 Street
POLAND
Tel. +48 48 38 48 800, tel./fax. + 48 48 385 00 10
Selling departament + 48 48 366 80 06
www.radwag.com
- 85 -
Loading...