4. HOME SCREEN............................................................................................................................................. 8
4.1. Top Bar....................................................................................................................................................9
4.2. Weighing Result Window.........................................................................................................................9
5. OPERATING THE MENU............................................................................................................................. 10
5.1. Entering the Menu..................................................................................................................................10
5.2. Menu Keys............................................................................................................................................. 10
5.3. Entering Numeric and Text Characters and Signs..................................................................................11
5.3.2. Text Box...................................................................................................................................... 13
5.3.4. Special Signs Table.....................................................................................................................15
5.4. Return to Weighing................................................................................................................................ 15
7. LOG IN OPERATION................................................................................................................................... 16
7.1. Log in Procedure....................................................................................................................................17
7.2. Log out Procedure..................................................................................................................................17
8.3. Dual Range Devices.............................................................................................................................. 19
8.4. Weighing Unit Change...........................................................................................................................19
8.5. Setting MIN, MAX Thresholds................................................................................................................ 20
8.5.1. Setting MIN, MAX Threshold Using Programmable Button..........................................................20
8.5.2. Setting MIN, MAX Thresholds by Product Selection....................................................................21
8.5.3. Setting MIN, MAX Threshold Using Proximity Sensor................................................................. 21
8.5.4. Setting MIN, MAX Threshold Using Digital Input......................................................................... 21
9. COMMUNICATION ...................................................................................................................................... 22
9.4. USB ....................................................................................................................................................... 23
13.1.2. Text Box.................................................................................................................................... 43
13.1.3. Bar Graph .................................................................................................................................43
14.2. Date and Time..................................................................................................................................... 48
15. UNITS ........................................................................................................................................................ 49
15.1. Units Availability................................................................................................................................... 49
16.3. Date and Time..................................................................................................................................... 51
16.4. Backlight Standby Time .......................................................................................................................52
17.2. User Adjustment .................................................................................................................................. 53
17.3. Start Mass Determination.....................................................................................................................54
20.2. Save Mode........................................................................................................................................... 57
20.3. Auto Threshold ....................................................................................................................................57
20.4. Result Control......................................................................................................................................57
21. WORKING MODE – WEIGHING................................................................................................................60
21.1. Home Screen.......................................................................................................................................60
21.2. Local Settings ...................................................................................................................................... 61
22. WORKING MODE – PARTS COUNTING .................................................................................................. 61
22.1. Home Screen.......................................................................................................................................61
22.2. Local Settings ...................................................................................................................................... 61
22.2.1. Function of Automatic Correction of Reference Sample Mass...................................................62
22.3. Setting Reference Sample Mass by Entering Mass of a Single Part.................................................... 63
22.4. Setting Reference Sample Mass by Determining Mass of a Single Part .............................................. 63
22.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part from Database...........................64
22.6. Entering Reference Sample Mass to Weighing Device Memory.......................................................... 64
23. WORKING MODE - PERCENT WEIGHING............................................................................................... 65
23.1. Home Screen.......................................................................................................................................65
23.2. Local Settings ...................................................................................................................................... 65
23.3. Reference Sample Mass Determined by Weighing..............................................................................66
23.4. Reference Sample Mass Entered to the Weighing Device Memory.....................................................66
23.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part from Database...........................66
24. WORKING MODE – DOSING.................................................................................................................... 66
24.1. Home Screen.......................................................................................................................................67
24.2. Local Settings ...................................................................................................................................... 67
24.5. Dosing Process Report........................................................................................................................ 73
25. WORKING MODE – FORMULATIONS ..................................................................................................... 73
25.1. Home Screen.......................................................................................................................................73
25.2. Local Settings ...................................................................................................................................... 74
25.3. New Formulation.................................................................................................................................. 74
28.1. Data Export.......................................................................................................................................... 86
28.2. Data Import..........................................................................................................................................87
30.1. Variables List ....................................................................................................................................... 88
PUE HX5.EX is a versatile weighing indicator designed to make industrial
scales based on load cells. It is constructed in accordance with Directive
2014/34/EU, and can be used in environment filled with explosive gases and
dust, classified as zones: 1, 2, 21, 22. The PUE HX5.EX indicator's functionality
makes it a perfect solution for vast range of various industry applications.
Hazardous area
PUE HX5.EX
Safe area
IM01.EX
OR
PM01.EX-1
AC 100÷240V
50/60Hz
PM01.EX-2
AC 100÷240V
50/60Hz
The standard version of the indicator is equipped with the following
communication interfaces enabling integration with devices placed in the
hazardous area: 2 x RS232, RS485, and 4I/4O (optionally). The indicator
is powered using intrinsically safe power supplies: PM01.EX-1 or PM01.EX-2.
It is possible to connect the PUE HX5.EX indicator to IM01.EX communication
module, placed outside the hazardous area. Due to its intrinsically safe
interface, the IM01.EX module can expand the indicator’s interfaces range:
Ethernet, 2 x RS232, USB, RS485 (option), 12I/12O (option), analog outputs
(option), PROFIBUS (option).
6
Page 7
2. OPERATION PANEL
Keys:
Press to switch the indicator on / off.
Press to enter the main menu.
Press to log in.
Press to zero the scale.
Press to tare the scale.
Press to change the weighing unit.
Press to cancel the message.
Press to confirm the message.
Press to change the working mode.
Press to confirm the weighing result (PRINT).
Press to confirm the messages (ENTER).
Press to cancel the messages.
7
Page 8
Programmable key assigned to a pictogram displayed in the bottom screen
area.
Programmable key assigned to a pictogram displayed in the bottom screen
area.
Programmable key assigned to a pictogram displayed in the bottom screen
area.
Programmable key assigned to a pictogram displayed in the bottom screen
area.
Programmable key assigned to a pictogram displayed in the bottom screen
area.
3. START-UP
• Press key, it is to be found on the top of the operation panel.
• Upon completed start-up the home screen is displayed automatically.
• The scale is ready for operation.
4. HOME SCREEN
The home screen features 4 sections:
• top bar,
• weighing result window,
• workspace,
• pictograms.
Home screen view:
8
Page 9
4.1. Top Bar
The top bar displays the following information:
Working mode name and symbol.
Weighing device name.
4.2. Weighing Result Window
Weighing result window provides all weighing related data.
4.3. Workspace
The workspace is to be found underneath the weighing result window.
Symbol informing that printer is connected.
Symbol informing that communication with a PC computer is on.
Symbol informing that communication with E2R SYSTEM is on.
The workspace comprises 4 programmable widgets. Each working mode
features default home screen widgets set. You can customize the screen.
For detailed information concerning the workspace read section 'Display'.
4.4. Pictograms
The pictograms assigned to operation panel keys are to be found underneath
the workspace.
9
Page 10
You can define on-screen pictograms individually for each working mode.
For detailed procedure informing you how to define on-screen pictograms read
section 'Display'.
5. OPERATING THE MENU
In order to navigate the program menu use the operation panel.
5.1. Entering the Menu
In order to enter the menu press
key. Background colour of the first menu
entry differs from the remaining ones. To navigate the program menu use the
keys that operate as arrow keys.
Menu view
5.2. Menu Keys
Press to enter the main menu.
Press to go to the home screen.
Press to move one menu level up, or to discard entering parameter
modifications.
Press to move one menu level up.
Press to delete a character when editing numeric and text values.
10
Page 11
Press to change keyboard mode when editing numeric and text values.
5.3. Entering Numeric and Text Characters and Signs
Press to select/change working mode.
Press to confirm/accept modifications.
Press to move one menu level up, or to discard entering parameter
modifications.
Press to select higher-level parameters group, or to edit parameter value and
change it by one digit up.
Press to select parameters group that you want to operate. The first parameter
of the selected parameters group is displayed.
Press to select lower-level parameters group, or to edit parameter value and
change it by one digit down.
Depending on a type of data entered to indicator's memory the software offers
two different edit boxes:
• numerical box (for entering part mass values, tare values, etc.).
In order to activate special signs table while editing text box it is necessary
to press and hold
key.
Where:
Function inactive.
ąëñ
5.4. Return to Weighing
Press to switch to diacritical signs keyboard.
Introduced modifications are automatically recorded upon return to the home
screen. To return to the home screen:
15
Page 16
• press key repeatedly, keep pressing the key until you see the home
screen,
• press
6. PROGRAM
key, the home screen is displayed immediately.
Program menu is divided into function groups. Function group is a group
of interrelated parameters.
Function Groups:
• Working modes,
• Databases,
• Reports,
• Communication,
• Peripherals,
• Printouts,
• Inputs / Outputs,
• Display,
• Permissions,
• Units,
• Adjustment,
• Misc,
• Scale data.
7. LOG IN OPERATION
In order to access operator-related parameters and to edit databases, you
need to log in as an operator with <Administrator> permission level.
<Admin> is a default operator set on each brand new
indicator. The default operator is assigned with
<Administrator> permission level. <Admin> account is not
protected by password. Logging of default operator is carried
out automatically upon indicator start-up. In case
of modification of default operator data or upon adding new
operators, it is necessary to log in manually.
16
Page 17
7.1. Log in Procedure
• Go to the home screen, press key, wait for the operators database
to open.
• Select particular operator, wait for the on-screen keyboard with
a password box to open.
• Enter the password and press
key to confirm, the home screen
is displayed.
7.2. Log out Procedure
• Go to the home screen, press
key, wait for the operators database
to open.
• Select <Log out> option, the home screen is displayed.
7.3. Permission Levels
There are 4 permission levels: Administrator, Advanced Operator, Operator,
None.
Access to operator-related parameters and program functions
is conditioned by a permission level:
Permission level Available parameters and functions
Permission to edit operator-related parameters not granted. Operator
can neither accept the weighing result nor start the following
None
Operator
Advanced operator
Administrator
operations: entering reference sample mass and determining
reference sample quantity in <Parts counting> mode, entering
reference sample mass and determining reference sample in <Percent
weighing> mode, dosing, making formulations.
Operator can edit <Readout>, <Misc> parameter groups (operation
of <Date and time> parameter and <Restore operator's default
settings> function excluded). Operator can run and carry out all
weighing operations.
Operator can edit all operator-related parameters excluding <Date and
time> parameter. Operator can run and carry out all weighing
operations.
Operator can edit all operator-related parameters and all databases,
and use all functions. Operator can run and carry out all weighing
operations.
17
Page 18
8. WEIGHING
Load the weighing pan. Read the result when stability marker is
displayed.
Only stable weighing results can be recorded (stability marker
).
8.1. Zeroing
To zero mass indication press
pictograms are displayed:
key. Zero indication and the following
and .
Zeroing operation means determining new zero point, recognized by the
weighing device as precise z ero. The instrumen t can be zeroed only when the
indication is stable.
Indication can be zeroed only within ±2% range of maximum
capacity. If the zeroed value is greater than ±2% of the
maximum capacity, then the software displays a respective
error message:
8.2. Taring
To determine net weight value, load the weighing pan with a packaging, wait
for a stable indication and press
pictograms are displayed:
tared.
key. Zero indication and the following
and . The weighing device has been
Remember not to exceed scale's maximum capacity, i.e. sum of tare weight
value and load weight value must be lower than the maximum capacity value.
Upon unloading the weighing pan, the sum of tared masses with minus sign
is displayed.
You can assign tare value to a particular product stored in the database.
In such case the assigned tare value is automatically acquired upon selection
of the given product.
18
Page 19
It is impossible to tare zero or negative values. When you tare
zero or negative values the indicator responds with the
following message:
8.3. Dual Range Devices
N/A in case of single range scales
Switching from weighing with the accuracy of the I weighing range to weighing
with the accuracy of the II weighing range takes place automatically upon
exceeding Max of the I weighing range.
In case of dual range scales:
• upon switching to weighing with the accuracy of the I weighing range,
pictogram/marker is displayed on the left,
• upon switching to weighing with the accuracy of the II weighing range,
pictogram/marker is displayed on the left.
Switching from weighing with the accuracy of the II weighing range
to weighing with the accuracy of the I weighing range takes place
automatically upon unloading the weighing pan and returning to AUTOZERO
(
8.4. Weighing Unit Change
pictogram/marker is displayed).
–
You can change the weighing unit, to do it press key.
Options in case when the main unit is [g]:
• g (gram),
• kg (kilogram),
• ct (carat),
• lb (pound)*,
• oz (ounce)*,
• N (newton)*,
• u1 (custom unit 1)*,
• u2 (custom unit 2)*.
*) – unit disabled for verified weighing devices.
19
Page 20
Options in case when the main unit is [kg]:
• kg (kilogram),
• lb (pound)*,
• N (newton)*,
• u1 (custom unit 1)*,
• u2 (custom unit 2)*.
*) – unit disabled for verified weighing devices.
You can declare start unit and two custom units – for detailed
information read section 15, „UNITS”.
8.5. Setting MIN, MAX Thresholds
MIN, MAX thresholds are used:
• to control mass of the weighed loads (read section 20.4 of this user
manual),
• for graphic visualization informing you how much of the weighing device
capacity is used (read section 13.1.3 of this user manual),
• to control the external automation systems using digital outputs of the
indicator (read section 12.2 of this user manual).
The mass is considered to be correct if it is comprised within MIN and MAX
threshold values. You can set MIN, MAX thresholds as follows:
• using programmable <
Set MIN, MAX> button,
• selecting product with set thresholds,
• using a proximity sensor,
• using a digital input.
MAX threshold value has to be greater than MIN threshold
value.
8.5.1. Setting MIN, MAX Threshold Using Programmable Button
• Enter <Display / Buttons functions> submenu.
• Edit a given button.
• Select <Set MIN and MAX> parameter from the list.
• Exit to the home screen.
• Press previously programmed button, <Min> numerical box is opened.
20
Page 21
• Enter respective value and press key to confirm changes, <Max>
numerical box is opened.
• Enter respective value and press
8.5.2. Setting MIN, MAX Thresholds by Product Selection
key to confirm changes.
• Enter <Databases / Products> submenu.
• Edit a given product and enter respective threshold values.
• Exit to the home screen and press
button to select product with set
thresholds.
8.5.3. Setting MIN, MAX Threshold Using Proximity Sensor
• Enter <Working modes> menu and select respective working mode.
• Edit a given proximity sensor (left or right).
• Select <Set MIN and MAX> parameter from the list.
• Exit to the home screen.
• Move your hand near the previously programmed proximity sensor, <Min>
numerical box is opened.
• Enter respective value and press
key to confirm changes, <Max>
numerical box is opened.
• Enter respective value and press
8.5.4. Setting MIN, MAX Threshold Using Digital Input
key to confirm changes.
• Enter <Inputs / Outputs> menu, next enter <Inputs> submenu.
• Edit a given input.
• Select <Set MIN and MAX> parameter from the list.
• Exit to the home screen.
• Activate the previously set input signal, <Min> numerical box is opened.
• Enter respective value and press
key to confirm changes, <Max>
numerical box is opened.
• Enter respective value and press
key to confirm changes.
21
Page 22
9. COMMUNICATION
The scale operating in hazardous area can communicate with peripheral
devices, the communication is established via the following ports: RS232 (1),
RS232 (2), RS485.
Since the scale is intended for operation in hazardous area, its
communication interfaces have been equipped with hermetic
intrinsically safe connectors.
It is possible to expand communication interfaces range outside hazardous
area, this is done using IM01.EX communication module, connected to the
scale via intrinsically safe interface. Standard IM01.EX communication module
is equipped with the following interfaces: 2 x RS232, 4I/O, Ethernet, USB.
To set the ports go to < / Communication> submenu.
Applicable in case of an active IM01.EX communication module
• Select <Ethernet> port.
• Set transmission parameters:
DHCP
IP Address 0.0.0.0
Subnet mask 0.0.0.0
Default gateway 0.0.0.0
MAC address ---
Set the transmission parameters in accordance with your local
network.
<MAC address> parameter is automatically assigned to the
weighing device with <Read-only> attribute.
Yes, No
If you select
transmission parameters are given <Read-only> attribute.
option for <DHCP> parameter, the remaining
9.4. USB
Applicable in case of an active IM01.EX communication module
USB port is used for:
• connecting a USB flash drive in order to enable:
- printout of measurement data (set <Printer/Port> parameter
to <Pendrive> value),
- databases export/import,
- operator's parameters expo rt/import,
- Alibi and weighing repo rts export,
• connecting scale to PCL printer,
• connecting EPSON TM-T20 printer (featuring USB port).
The USB flash drive must support FAT files system.
23
Page 24
10. PERIPHERAL DEVICES
10.1. IM01.EX Communication Module
IM01.EX communication module is a device intended for operation outside
hazardous area, it is equipped with intrinsically safe circuits which can be
placed in zones endangered with explosion. The module can be connected, via
the intrinsically safe interface, with HX5.EX scale placed in the hazardous area.
This allows integration of HX5. EX scale with various accessories: barcode
scanners, printers, control buttons, light signalling towers, buzzers, PLC
controllers and other controlling/signalling devices, PCs operating outside
hazardous areas.
10.1.1. Connecting IM01.EX to the Mains
IM01.EX communication module is equipped with cable terminated with a plug
featuring a ground pin, plug type is conditioned by region/country. Connect the
cable to the wall outlet with a ground pin.
It is not allowed to connect the IM01.EX communication
module's plug to the wall outlet located in the hazardous
area.
10.1.2. Connecting IM01.EX Module and PUE HX5.EX Indicator
IM01.EX communication module is equipped with a connector for intrinsically
safe circuit, marked as HX5.EX, used to connect PUE HX5.EX indicator.
Connect the PUE HX5.EX indicator to the IM01.EX communication module via
a dedicated cable, PT0327.
Connect the IM01.EX communication module to devices
operating in hazardous area when there is no risk of explosive
atmosphere occurrence.
10.1.3. Activation of Connection Between the IM01.EX Module and
HX5.EX Scale
• Connect the IM01.EX communication module to the HX5.EX scale and to
the mains.
• Enter <Peripherals / IM01.EX communication module / Activation>
submenu and activate the IM01.EX communication module (
active, - module inactive).
- module
24
Page 25
• Activation of the IM01.EX communication module results with automatic
display of the following data:
Status
Software version Software version of the IM01.EX module.
Additional modules
10.2. Computer
Active connection status, the status takes the following values:
Connected, Not connected.
Submenu for configuration of optional additional modules of the
IM01.EX communication module (anybus module, analog outputs
module).
The weighing device can integrate with a computer. Active weighing device – computer connection is signalled by
pictogram (top bar of the home
screen). To configure scale-computer connection settings go to <
/
Peripherals / Computer> submenu.
10.2.1. Computer Port
Communication between the weighing device and the computer can be
established via the following ports:
RS232 (1)
RS232 (2)
RS232 (3)
RS232 (4)
RS485
Ethernet
Procedure:
RS232 ports of the scale (M12 5P connectors).
RS232 ports of scale-connected IM01.EX communication module (M12 8P
connectors).
RS485 port of scale-connected IM01.EX communication module (M12 4P
connector).
Ethernet port of scale-connected IM01.EX communication module (RJ45
connector).
• Enter <Peripherals / Computer / Port> submenu and set respective
option. <Port settings> submenu differs depending on the selected port:
Port: number of port for transmission protocol.
Timeout [ms]: time delay - time interval for disconnection, counted from
the moment of the most recent command sent from the peripheral
device to which the weighing device is connected.
Parameter allowing you to set address of scale connected with the comp uter.
Procedure:
• Enter <Peripherals / Computer / Address> submenu, <Address> edit
box is displayed.
• Enter the address and press
10.2.3. Weighing Printout Template
key to confirm changes.
Template of an individual printout designed using scale and sent to a computer.
Procedure:
• Enter <Peripherals / Computer / Weighing printout template>
submenu, <Weighing printout template> edit box is displayed.
• Modify the template and press
10.2.4. Continuous Transmission
key to confirm changes.
Parameter allowing you to activate 'scale' - 'computer' continuous transmission.
In order to send the content of <Weighing printout template> to a computer
continuously you must activate <Continuous transmission> parameter.
Procedure:
• Enter <Peripherals / Computer / Continuous transmission> submenu
and set respective value.
Parameter allowing you to set frequency of print of <Weighing printout
template> when continuous transmission is activated. Interval for printouts
is set in seconds with 0.1 [s] accuracy within 0.1 - 1 000 [s] range.
Procedure:
• Enter <Peripherals / Computer / Interval> submenu, <Interval> edit box
is displayed.
• Enter respective value and press
10.2.6. PUE HX5.EX and E2R System
key to confirm changes.
Establishing connection between the scale and the E2R System PC software.
E2R System is a modular solution designed to comprehensively manage
production processes, some stages of which consist in weighing.
<E2R> can be activated exclusively by an <Administrator>
user. In case of integration of the scale with <E2R System>
software, editing of databases on scales is disabled.
Procedure:
• Enter <Peripherals / Computer / E2R> submenu and set respective
option.
Options:
Connection with E2R System inactive.
Connection with E2R System active.
10.3. Printer
<Printer> submenu allows you to:
• set port for establishing communication with printer,
• select printer code page,
• define print start - <Prefix> parameter,
• define print end - <Suffix> parameter.
To set scale-printer communication parameters go to < / Peripherals /
Printer> submenu.
27
Page 28
10.3.1. Printer Port
Communication between the weighing device and the printer can be
established via the following ports:
RS232 (1)
RS232 (2)
RS232 (3)
RS232 (4)
USB
Ethernet
Pendrive
Procedure:
RS232 ports of the scale (M12 5P connectors).
RS232 ports of scale-connected IM01.EX communication module. (M12 8P
connectors).
USB port of type A of the scale-connected IM01.EX communication module.
(M12 4P connector).
Ethernet port of scale-connected IM01.EX communication module (RJ45
connector) used for connecting a network printer or a computer with
RADWAG-designed program, e.g. RAD KEY.
USB port of type A of the scale-connected IM01.EX communication module
(M12 4P connector). used for connecting a USB flash drive in order to print
weighings to a text file.
• Enter <Peripherals / Printer / Port> submenu and set respective option.
<Port settings> submenu differs depending on the selected port:
IP Address: IP address of the scale.
Port: number of port for transmission protocol.
Timeout [ms]: time delay - time interval for disconnection, counted from
the moment of the most recent command sent from the peripheral
device to which the weighing device is connected.
In order to provide correct integration of the weighing device with the printer
(correct printout of diacritical signs of a given language) it is necessary to make
sure that the code page of a sent printout is accordant with a code page
of a printer. There are two methods for obtaining code page accordance:
• by setting the right code page in the settings of a printer (read the user
manual of the printer) – it must be accordant with the printout code page
of a weighing device:
• by sending the control code from the weighing device, which automatically
sets the right code page of the printer (i.e. code page accordant with the
one of a weighing device).
Default code page value of the printer is 1250 – Central
European code page.
Exemplary weighing device settings for correct printout of Polish signs with use
of EPSON printer connected to RS232 port:
Prefix and suffix are controlling codes (given in a hexadecimal format) sent
to a printer at print start - <Prefix> parameter, and at print end - <Suffix>
parameter. Sending these codes allows to control globally information and
actions carried out at the beginning and/or at the end of each printout sent from
a weighing device to a printer. Most frequent usage:
• Prefix - sending info on print code page.
• Suffix - sending paper crop command in EPSON printers (in case
of printer equipped with an auto-cutter blade). Paper crop in EPSON
printer – code 1D564108.
Settings of <Prefix> and <Suffix> parameters are valid for all printouts sent via
the weighing device, e.g.: calibration reports, statistics and header, footer and
GLP printouts.
29
Page 30
10.3.4. Record of Measurement Data Onto a USB Flash Drive
Applicable in case of an active IM01.EX communication module
Scale program enables record of measurement data onto an external flash
drive.
Procedure:
• Plug the flash drive to the USB port of the IM01.EX communication
module, to do it use PT0084 cable.
• Enter <Peripherals / Printer / Port> submenu and set <Pendrive> value.
Return to weighing.
• Now the measurement data is saved to a text file on a USB flash drive
whenever you press
a scale software under printout.txt name.
key. The file is generated automatically by
Remove the USB flash drive from a weighing device's USB
port after about 10s counting from the last measurement
record, only after this time the data is saved. Next plug the
USB flash drive to a computer and read the saved text file.
The data can be printed using any computer-connected printer. It is possible
to overwrite the file with new data. New data is recorded to an existing file
therefore it is possible to continue recording measurement data to the file once
created.
USB flash drive must comprise <FAT files system>.
10.4. Barcode Scanner
The scale can integrate with a barcode scanner. The barcode scanner is used
to facilitate quick search for database records.
Enter <Communication> submenu and set baud rate for
a barcode scanner (by default 9600b/s). For detailed
description of 'scale' - 'barcode scanner' communication read
ANNEX 04 of this user manual.
10.4.1. Barcode Scanner Port
Communication between the weighing device and the barcode scanner can be
established via the following ports:
30
Page 31
RS232 (1)
RS232 (2)
RS232 (3)
RS232 (4)
USB
Procedure:
RS232 ports of the scale (M12 5P connectors).
RS232 ports of scale-connected IM01.EX communication module. (M12 8P
connectors).
USB port of type A of the scale-connected IM01.EX communication module.
(M12 4P connector).
• Enter <Peripherals / Barcode Scanner / Port> submenu and set
respective port.
10.4.2. Offset
Parameter allowing you to set the first significant code's character, characters
preceding the first significant character are skipped during comparis on search.
Procedure:
• Enter <Peripherals / Barcode Scanner / Offset> submenu and, using the
on-screen keyboard, enter a required value.
10.4.3. Length of Code
Parameter allowing you to set the number of code's characters to be taken into
account during search procedure.
Procedure:
• Enter <Peripherals / Barcode Scanner / Length of code> submenu and,
using the on-screen keyboard, enter a required value.
10.4.4. Prefix, Suffix
Parameter allowing you to edit <Prefix> and <Suffix> in order to provide
synchronization of the scale program with a barcode scanner.
In RADWAG-adopted standard, the prefix is 01 sign (byte)
hexadecimal format, the suffix is 0D sign (byte) hexadecimal
format.
Procedure:
• Enter <Peripherals / Barcode Scanner / Prefix> submenu and, using the
on-screen keyboard, enter a required value (hexadecimally).
31
Page 32
• Go to <Suffix> submenu and, using the on-screen keyboard, enter
a required value (hexadecimal format).
10.4.5. Field Selection
Parameter allowing you to specify, for which field the search is to be carried out
in particular databases.
Procedure:
• Enter <Peripherals / Barcode Scanner / Field selection> submenu, list
of fields for search is displayed.
Available values: None, Product, Operator, Customer, Packaging, Lot
number, Batch number.
10.4.6. Filtering
Parameter allowing you to declare search criteria.
Procedure:
• Enter <Peripherals / Barcode Scanner / Filtering> submenu, list
of search criteria is displayed.
Filtering criteria conditioned by field type:
Field selection Filtering
None *
Product Name, Code, Name 2, Code 2.
Operator Name, Code.
Customer Name, Code.
Packaging Name, Code.
Lot number **
Batch number **
*) - <Filtering> submenu hidden. Function inactive.
**) - <Filtering> submenu hidden. Function active.
10.4.7. Test
Parameter allowing you to verify if operation of a barcode scanner connected
to a scale is correct.
32
Page 33
Procedure:
• Enter <Barcode Scanner / Test> submenu, <Test> edit box is displayed,
it features ASCII field and HEX field.
• The code is scanned and entered to the ASCII field and HEX filed, next
the test result is displayed in the bottom part.
When:
•<Prefix> and <Suffix> declared in scale settings comply with <Prefix>
and <Suffix> of the scanned code, the test result is <Positive>.
• <Prefix> and <Suffix> declared in scale settings do NOT comply with
<Prefix> and <Suffix> of the scanned code, the test result is <Negative>.
10.5. Additional Display
Settings of connection between a scale and an external additional display.
10.5.1. Additional Display Port
Communication between the weighing device and the additional display can be
established via the following ports:
RS232 (1)
RS232 (2)
RS232 (3)
RS232 (4)
Procedure:
RS232 ports of the scale (M12 5P connectors).
RS232 ports of scale-connected IM01.EX communication module. (M12 8P
connectors).
• Enter <Peripherals / Additional display / Port> submenu and select
respective port.
Enter <Communication> submenu and set baud rate, mind
that the set value must be accordant with the additional
display (115200b/s).
10.5.2. Bottom Text Area Template
Additional display of WD-6 series features text area for extra information such
as date, tare value etc. Displayed data content is set using the weighing
device.
33
Page 34
Procedure:
• Enter <Peripherals / Additional display / Bottom text area template>
submenu, <Bottom text area template> edit box is displayed.
• Modify the template and press
key to confirm changes. The template
can comprise 30 characters maximum.
11. PRINTOUTS
<Printouts> submenu allows you to:
• define template for header, GLP and footer printout,
• define template for adjustment and dosing reports,
• create 10 non-standard printouts.
11.1. Print Mode: Header – GLP Printout - Footer
Printout comprises 3 basic sections:
Header section
GLP printout section
Footer section
Group of parameters allowing you to declare variables that are to be
printed on a header printout.
Group of parameters allowing you to declare variables that are to be
printed on a measurement printout.
Group of parameters allowing you to declare variables that are to be
printed on a footer printout.
For each section a customised list of printout variables has been specified. You
must set respective option either enabling or disabling particular variable
printing.
Variables list
HEADER GLP printout FOOTER
Dashes *
Working mode *
Date *
Time *
Scale type
Scale S/N
Operator *
Product *
Customer
Universal variable 1
Universal variable 2
Universal variable 3
Empty line *
Adjustment Report
Non-standard printout
Date
Time
Operator
Product
Customer
Packaging
Universal variable 1
Universal variable 2
Universal variable 3
Net
Tare
Gross
Current result *
Adjustment Report
Non-standard printout
Working mode
Date
Time
Scale type
Scale S/N
Operator
Product
Customer
Universal variable 1
Universal variable 2
Universal variable 3
Dashes *
Empty line *
Adjustment Report
Signature *
Non-standard printout
34
Page 35
*) – Variables accessible by default (marked with pictogram).
How to Operate Printouts
1. To print variables comprised within GLP printout section which are
enabled for printing and marked with
pictogram, press key
located on the weighing device panel.
2. To print variables comprised within Header and/or Footer section which
are enabled for printing and marked with
programmable button: <
Header printout> and/or <Footer
pictogram, press respective
printout>.
For detailed procedure informing you how to program
buttons, read section 13.2.
Variables description
Variable
Working mode Working mode name.
Scale type Declared scale type.
Scale S/N Serial number of the scale.
Operator Logged-in operator name.
Product Currently selected product name.
Customer Currently selected customer name.
Packaging Currently selected packaging names.
Date Current date.
Time Current time.
Net Net weight value in a basic unit (calibration/adjustment unit).
Tare Tare weight value in the current unit.
Gross Gross weight value in the current unit.
Universal variable 1 Value of universal variable 1.
Universal variable 2 Value of universal variable 2.
Universal variable 3 Value of universal variable 3.
Current result Measurement result in a current unit.
The program allows you to design 10 non-standard printouts. Each of them can
consist of approximately 1900 characters (letters, digits, special signs, spaces),
including:
• fixed texts,
• variables conditioned by a particular working mode and operator needs
(mass, date, checkweighing thresholds).
For list of printout variables read ANNEX 01 of this user
manual.
Procedure for adding non-standard printout:
• Enter <Printouts / Non-standard printouts> submenu.
• Press
is created, it is defined by the following data:
Name
Code
Project
button (add record) assigned to panel key, a new record
Non-standard printout name (43 characters maximum).
Non-standard printout code (15 characters maximum).
Text editor of printout template (1900 characters maximum).
For detailed information regarding text box read section 5.3.
36
Page 37
11.3. Adjustment Report
Group of parameters allowing you to declare variables that are to be printed
on an adjustment report printout. Adjustment report is automatically generated
at the end of each adjustment process.
Variables description
Project Project name (31 characters maximum).
Adjustment type Performed adjustment type.
Operator Logged-in operator name.
Project Project name printout.
Date Adjustment performance date.
Time Adjustment performance time.
Scale S/N Serial number of the scale.
Adjustment result
difference
Dashes Dashed line separating printout data and signature fields.
Signature An area for the signature of an operator carrying out the adjustment.
11.4. Dosing Report
Difference between mass of the adjustment weight that was measured
during the last adjustment and mass of a currently measured
adjustment weight.
Group of parameters allowing you to declare variables that are to be printed
on a dosing report printout. Dosing report is generated automatically after
either completion or abortion of each process.
Variables description
Variable Description
Start date Dosing process start date.
End date Dosing process end date.
Operator Operator carrying out the dosing process.
Customer Customer for whom the dosing process is carried out.
Product
Fast dosing threshold [DT1]
Dosing threshold [DT2]
Name of dispensed product selected from <Products>
database.
Mass value for rough dosing in case of automatic 2-stage
dispensing.
Enter to set target mass value to be dosed in case of an
automatic dispensing.
37
Page 38
Dosing correction Dosing correction value.
Dosing mass
Difference
Min Value of low threshold of manual dosing.
Max Value of max threshold of manual dosing.
Status
Dashes Dashed line separating printout data and signature fields.
Signature
Non-standard printout One of 10 non-standard printouts.
11.5. Formulation Report
Net value of mass that is to be dosed, given in an adjustment
unit.
Difference between the net value of mass that is to be dosed
and a value of automatic dosing threshold.
Status of the dosing process, values: Ongoing, Completed, Aborted.
An area for the signature of an operator carrying out the
dosing process.
Group of parameters allowing you to declare variables that are to be printed on
a formulation report printout. Formulation report is generated automatically
after either completion or abortion of each process.
Variables description
Status
Start date Formulation start date.
End date Formulation end date.
Formulation Performed formulation name.
Operator Operator performing formulation process.
Customer Customer for whom the formulation is carried out.
Ingredients quantity Number of formulation ingredients.
Measurements quantity Number of weighings performed within the formulation cycle.
Lot number Lot number assigned to a formulation.
Measurements List of weighings performed within the formulation cycle.
Target value Sum of declared nominal masses of the ingredients.
Sum Total weight value of the performed formulation.
Difference
Formulation performance status. Status values: Ongoing,
Aborted, Completed.
Difference between the sum and the value of expected target
weight.
38
Page 39
12. INPUTS / OUTPUTS
Some models of HX5.EX scale can be equipped with 4 inputs / 4 outputs.
It is possible to expand inputs / outputs range outside hazardous area, this
is done using IM01.EX communication module, connected to the indicator via
intrinsically safe interface. Standard IM01.EX communication module
is equipped with 4 inputs / 4 outputs. Optional IM01.EX communication
module is equipped with 12 inputs / 12 outputs.
In order to set inputs and outputs enter: < / Inputs/Outputs> submenu.
12.1. Input Setup
• Enter <Inputs/Outputs> submenu.
• Select <Inputs> parameter, list of available inputs is displayed.
Where:
Inputs 1 ÷ 4
Inputs 5 ÷ 8
Inputs 9 ÷ 16
4 inputs, available in HX5.EX scale.
4 inputs of standard scale-connected IM01.EX communication module.
8 inputs of non-standard scale-connected IM01.EX communication module.
• Edit selected input, list of functions to be assigned is displayed. Input
functions list is identical like key functions list, read section 13.2.
• Select the function you want to assign to the input, next go back to the
home screen.
By default all functions are assigned with <None> value.
12.2. Output Setup
Output gets activated at the moment of assigning it with a given function.
Unassigned outputs remain inactive.
Procedure:
• Enter <Inputs/Outputs> submenu.
• Select <Outputs> parameter, list of available outputs is displayed.
39
Page 40
Where:
Outputs 1÷4
Outputs 5÷8
Outputs 9÷16
4 outputs, available in HX5.EX scale.
4 outputs of standard scale-connected IM01.EX communication module.
8 outputs of non-standard scale-connected IM01.EX communication module.
• Edit selected output, list of functions to be assigned is displayed.
None Output inactive.
Stable Stable weighing result over LO threshold value.
MIN stable Stable weighing result below the MIN threshold.
MIN unstable Unstable weighing result below the MIN threshold.
OK stable Stable weighing result between MIN and MAX thresholds.
OK unstable Unstable weighing result between MIN and MAX thresholds.
MAX stable Stable weighing result over the MAX threshold.
MAX unstable Unstable weighing result over the MAX threshold.
Zero Zero weighing result („zero” marker).
! OK unstable Unstable weighing result out of OK threshold.
! OK stable Stable weighing result out of OK threshold.
MIN MIN threshold signalling.
OK OK threshold signalling.
MAX MAX threshold signalling.
Active process Ongoing process signalling.
• Select the function you want to assign to the output, next go back to the
home screen.
By default all functions are assigned with <None> value.
For detailed description on how to set MIN, MAX thresholds
refer to section 8.5.
40
Page 41
13. DISPLAY
You can customize the home screen and the displayed data. Customization
concerns the workspace (fields 1, 2, 3, 4) and the bottom area (5) featuring
programmable buttons with functions assigned to numeric keys of the operation
panel:
1
3
2
4
5
In order to customize the screen go to < / Display> submenu.
13.1. Workspace
Workspace of your weighing device can comprise the following widgets: label,
text box, bar graph. Each working mode features default home screen widgets
set. Available fields (widgets) dimensions (width x height):
• Label – 1x1; 2x1.
• Text box - 1x1; 2x1.
• Bar graph – 1x1; 2x1.
2x1 field size equals size of two 1x1 fields (1 and 2 or 3 and 4). The program
automatically detects suitable widget size matching particular area. The
widgets are displayed in declared fields after completing data setup and leaving
'Settings' menu.
13.1.1. Label
Field displaying information updated in a real time in the course of weighing
device operation (selected product name, gross weight value, etc.).
<Workspace components> submenu for a field comprising <Label> widget
features the following options:
Information Enter this parameter to read info on selected widget type and dimensions.
Settings
Delete
Enter this parameter to specify which widget-assigned data is to be
displayed.
Enter this parameter to delete the widget. Upon entering, a respective
warning is displayed: <Delete?>. Press
key to confirm.
41
Page 42
Add
Option available only for those widgets that have not been added yet. Upon
entering <Add> submenu you can select <Label> widget of particular
dimensions.
Label data list (data accessibility is conditioned by a working mode).
Net
Tare
Gross
Operator
Product
Packaging
Customer
Lot number
Batch number
Universal variable 1
Universal variable 2
Universal variable 3
Date
Time
Date and Time
Thresholds
MIN threshold
MAX threshold
Number
Sum
Gross sum
Average
Min
Max
SDV
Net
Gross
Part mass
Reference mass
42
Page 43
Status
Formulation
Ingredient
- Default label settings.
13.1.2. Text Box
Field displaying weighing-related information. As an operator you can freely
program the field content (texts and variables of line 1 and line 2).
For detailed information regarding text box read section 5.3.2.
For list of printout variables read ANNEX 1 of this user
manual.
<Workspace components> submenu for a field comprising <Text box>
widget features the following parameters:
Information
Settings
Delete
Add
13.1.3. Bar Graph
Enter this parameter to read info on selected widget type and dimensions.
Enter this parameter to specify which widget-assigned data is to be
displayed. Upon entering this parameter line 1 and line 2 settings are
displayed. Lines 1, 2 can comprise 45 characters maximum.
Enter this parameter to delete the widget. Upon entering, a respective
warning is displayed: <Delete?>. Press key to confirm.
Option available only for those widgets that have not been added yet. Upon
entering <Add> submenu you can select <Text field> widget of particular
dimensions.
Bar graph is available for all working modes. It is a graphic visualisation
informing you how much of the weighing device capacity is used. It allows you
to observe whether the load stays within the specified Max capacity or is out of
it. Additionally it shows:
• position of Min and Max thresholds in <Weighing>, <Parts Counting>,
<Percent weighing> working modes.
• position of dosing thresholds in dosing process (read section 29.3 of this
manual).
• position of percent weighing range in formulation process (read section
30.4 of this manual).
<Workspace components> submenu for a field comprising <Bar graph>
widget features the following parameters:
43
Page 44
Information Enter this parameter to read info on selected widget type and dimensions.
This submenu provides two functions: 1. Bar graph type: Linear presentation
Settings
Delete
Add
of the weighing range. 2. Zoom: Enabling/disabling bar graph zoom in order
to provide more clear visualisation of the 'Min threshold' - 'Max threshold'
range.
Enter this parameter to delete the widget. Upon entering, a respective warning
is displayed: <Delete?>. Press key to confirm.
Option available only for those widgets that have not been added yet. Upon
entering <Add> submenu you can select <Bar graph> widget of particular
dimensions.
Bar graph operation:
Bar graph provides you with a linear presentation of the weighing range.
Min=0g
Max=0g
Additionally it shows where Min and Max thresholds are (providing that they
were declared).
• Visualisation of mass value lower than MIN value:
Min=1600g
Max=1800g
• Visualisation of mass value higher than MIN value and lower than MAX
value:
Min=1600g
Max=1800g
• Visualisation of mass value higher than MIN value and lower than MAX
value, <Zoom> option on:
Min=1600g
Max=1800g
• Visualisation of mass value higher than MAX value:
Min=1600g
Max=1800g
44
Page 45
MAX threshold value has to be greater than MIN threshold
value.
For detailed description on how to set MIN, MAX thresholds
refer to section 8.5.
13.2. Keys
Programmable numeric keys (F1 - F5). These are so called quick access keys
for triggering the most often performed operations. The keys are additionally
provided with graphic symbol which you can see in the bottom bar of the
display.
Procedure:
• Enter <Display / Buttons functions> submenu, edit a respective key.
• Select the function you want to assign to the key, next go back to the
home screen.
Key functions list (function accessibility is conditioned by a working
mode).
Function
Working mode parameters
Select product
Select formulation
Select packaging
Select customer
Set tare
Set MIN and MAX
Print header
Print footer
Statistics: Zero
Statistics: Print
Statistics: Print and zero
Working modes
45
Page 46
Edit labels quantity
Edit C labels quantity
Edit lot number
Edit batch number
Edit universal variable 1
Edit universal variable 2
Edit universal variable 3
Databases
Reports
Select operator
Print
Zero
Tare
Parameters
Change working mode
Change unit
Last digit
Set date
Set time
Set part mass
Determine part mass
Assign reference sample
Reference sample quantity - 5pcs.
Reference sample quantity – 10pcs.
Reference sample quantity – 20pcs.
Reference sample quantity – 50pcs.
Reference sample quantity – 100pcs.
Set reference sample mass
Determine reference sample mass
46
Page 47
Process start
Process stop
Process pause
Breakdown
None
- Default settings.
13.3. Default Screen Settings
Function allowing you to set default workspace values and default function
keys for a particular working mode.
Procedure:
• Enter <Display / Default screen settings> submenu, a respective
warning is displayed: <Continue?>.
• Press
key to confirm.
• Default workspace values and default function keys for a particular
working mode are set. Next, <Display> submenu is shown automatically.
14. PERMISSION LEVELS
<Permissions> submenu is available for operators logged as Administrator.
This group of parameters allows you to determine access rights for particular
operators.
In order to set permission levels enter < / Permissions> submenu.
14.1. Anonymous Operator
Parameter allowing you to assign unlogged weighing device operator (so called
anonymous operator) with permission level.
Procedure:
• Enter <Permissions / Anonymous operator> submenu.
• Set one of four options: None, Operator, Advanced operator,
Administrator.
47
Page 48
14.2. Date and Time
Default scale settings enable operator who is logged as Administrator
to modify date and time parameter. but the software allows to change
permission level, and as a result authorize other operators to access <Date
and time> parameter.
Procedure:
• Enter <Permissions / Date and time> submenu.
• Set one of four options: None, Operator, Advanced operator,
Administrator.
When <None> option is set, all unlogged operators can set
date and time.
14.3. Printouts
Default scale settings enable operator who is logged as Administrator to edit
printouts. but the software allows to change permission level, and as a result
authorize other operators to access <Printouts> parameter.
Procedure:
• Enter <Permissions / Printouts> submenu.
• Set one of four options: None, Operator, Advanced operator,
Administrator.
When <None> option is set, all unlogged operators can edit
printouts.
14.4. Databases Edition
Parameter allowing you to set permission levels enabling particular operators
to edit the following databases: products, formulations, packaging, customers,
universal variables.
Procedure:
• Enter <Permissions / Databases edition> submenu.
• Select respective database and set one of four options: None, Operator,
Advanced operator, Administrator.
When <None> option is set, all unlogged operators can edit
particular database.
48
Page 49
14.5. Select Database Record
Parameter allowing you to set permission levels enabling particular operators
to select given database records. 'Select database record' parameter is
available for the following databases: products, formulations, packaging,
customers, universal variables.
Procedure:
• Enter <Permissions / Select database record> submenu.
• Select respective database and set one of four options: None, Operator,
Advanced operator, Administrator.
When <None> option is set, all unlogged operators can select
particular database records.
15. UNITS
<Units> submenu allows you to:
• set particular units accessibility,
• set start unit,
• define two custom units,
• change gravitational acceleration value.
Option available for the Weighing mode exclusively
In order to set units enter < / Units> submenu.
Accessibility of particular units is conditioned by weighing
device status; i.e. it depends on the fact whether the given
scale is verified or not.
15.1. Units Availability
Parameter allowing to declare which units are to be accessible upon pressing
key.
Procedure:
• Enter <Units / Availability> submenu.
• Units list is displayed, mark units of your choice as available (
enabled
, - Unit disabled).
49
- Unit
Page 50
15.2. Start Unit
Upon setting start unit, the scale activates with the set start unit for these
modes where change of the unit is possible.
Procedure:
• Enter <Units> submenu, next enter <Start unit activation> parameter.
• Go to <Start unit> submenu and select start unit from the list.
• Go back to the home screen and restart the weighing device.
• After restart, the weighing device runs with the declared start unit.
15.3. Gravitational Acceleration
<Gravitational acceleration> parameter compensates changes of gravity
force being a result of different latitude and altitude when the selected unit
is „Newton” [N].
Procedure:
• Enter <Units / Gravitational acceleration> submenu, <Gravitational
acceleration> edit box is displayed.
• Enter gravitational acceleration value respective for the particular place
of use and press
key to confirm changes.
15.4. Custom Units
option available for non-verified scales exclusively
Parameter allowing you to declare two custom units. Displayed custom unit
value is a result of calculation, where obtained in the course of measurement
weight value is multiplied by a multiplier determined for this particular custom
unit. By default the custom units are marked as: [u1] – custom unit 1 and [u2]
– custom unit 2.
Procedure:
• Enter <Units / Custom unit 1> submenu, set the following parameters:
Name Unit name (3 characters maximum).
Multiplier Adjustment unit multiplier.
• Go to the home screen.
• Now you can use the custom unit, to do it press
50
key.
Page 51
Procedure for declaring <Custom unit 2> is analogous.
16. MISCELLANEOUS PARAMETERS
Parameters facilitating scale operation. To set these parameters go to:
<
/ Misc> submenu.
16.1. Interface Language
Parameter allowing you to set program menu language.
Procedure:
• Enter <Misc / Language> submenu and set the interface language.
Available languages: Polish, English, German, French, Spanish, Czech.
16.2. Display Brightness
Parameter allowing you to change display brightness, the brightness can be
changed within 0% - 100% range.
Procedure:
• Enter <Misc / Display brightness> submenu and set respective value.
Available values: 0%, 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, 90%,
100% (set by default).
16.3. Date and Time
Parameter allowing you to set current date and time, and to specify date and
time format.
Procedure:
• Enter <Misc / Date and time> submenu and change the settings. Refer to
the below table:
Parameter Description
Date
Time
Enter this parameter to set current date.
Enter this parameter to set current time.
51
Page 52
Enter this parameter to set date format. Options: YYYY.MM.DD (set by
Enter this parameter to set time format. Options: HH:mm:ss 24H
(set by default), 1 HH:mm:ss 12H, HH-mm-ss 24H, HH-mm-ss 12H,
HH.mm.ss 24H, HH.mm.ss 12H.
*) - Date format symbols: Y – year, M – month, D – day.
**) - Time format symbols: HH – hour, mm – minute, ss – second, 24H – 24-hour mode,
12H – 12-hour mode.
16.4. Backlight Standby Time
Parameter allowing you to set time interval, in [min], after which the screen
goes black.
Procedure:
• Enter <Misc / Backlight standby time> submenu and set respective
value.
Available values: None (set by default), 0.5, 1, 2, 3, 5.
16.5. Automatic Shutdown
Parameter allowing you to set time interval, in [min], after which the weighing
device shuts down automatically. If the indication is stable during the declared
time interval, the device is shut down automatically. Shutdown function
is inactive and the device cannot be turned off when any of the processes
is ongoing or when the menu is operated.
Procedure:
• Enter <Misc / Automatic shutdown> submenu and set respective value.
Available values: None (set by default), 1, 2, 3, 5, 10.
16.6. Default Operator Settings
Parameter allowing you to restore default operator settings.
Procedure:
• Enter <Misc / Restore default operator settings> submenu, a respective
warning is displayed: <Continue?>.
52
Page 53
• Press key to confirm. Message <Please wait...> is displayed, the
operation is in progress.
• After completed operation <Misc> submenu is displayed.
17. ADJUSTMENT
In order to ensure the highest weighing accuracy, it is recommended
to periodically introduce a corrective factor of indications to weighing device
memory, the said factor must be referred to the reference mass. Adjustment
has to be carried out prior first weighing or if the ambient temperature has
changed dynamically. Prior adjustment unload the weighing pan.
17.1. External Adjustment
External adjustment is carried out using an external mass standard of the right
accuracy and weight value, which value depends on scale type and capacity.
Correction is carried out semi-automatically, successive process stages are
signalled with prompts.
Procedure:
• Enter <Adjustment / External adjustment> submenu. Message
<Remove weight> is displayed.
• Unload the weighing pan and press
Please wait...> is displayed.
• Upon completed start mass determination, message <Put weight>
is displayed along with particular value of mass standard assigned to the
scale.
• Load the weighing pan with the required adjustment weight and press
key.
• Upon completed procedure, message <Remove weight> is displayed.
• Take the weight off the weighing pan, the scale displays the
<Adjustment> submenu.
17.2. User Adjustment
User adjustment can be carried out using any adjustment weight, mass
of which ranges between: 0.3 Max - Max. To run user adjustment procedure
enter <Adjustment / User adjustment> submenu. User adjustment and
external adjustment procedures are likewise with one exception, before user
adjustment start, an edit box for entering mass of the used adjustment weight is
opened.
option available for non-verified scales exclusively
key, message <Adjustment;
53
Page 54
17.3. Start Mass Determination
If the weighing device does not require adjustment or you do not have suitable
amount of adjustment weights, you can only determine start mass.
Procedure:
• Enter <Adjustment / Start mass determination> submenu. Message
<Remove weight> is displayed.
• Unload the weighing pan and press
key, message <Start mass
determination; Please wait...> is displayed.
• Upon completion of start mass determination procedure, the scale
displays the <Adjustment> submenu.
17.4. Adjustment Report
Adjustment report is automatically generated at the end of each adjustment
process, next it is sent to port selected for <Peripherals / Printer>. To declare
report content go to <Printouts / Adjustment report> submenu.
For instruction on how to declare adjustment report settings read section
'Printouts'.
Report example:
----------Adjustment report---------Adjustment type External
Operator Nowak Jan
Date 2018.04.10
Time 13:22:28
Scale S/N 123456
Adjustment result difference 0.0g
------------------------------------Signature
.....................................
18. SCALE DATA
Scale data menu provides information on the weighing device and its program.
The parameters serve informative purposes: scale S/N, scale type, software
version, product code, settings printout.
In order to send weighing device settings (all parameters) to printer port, select
<Settings printout> parameter.
54
Page 55
19. WORKING MODES - GENERAL INFORMATION
The weighing device features the following working modes:
Weighing
Parts counting
Percent weighing
Dosing
Formulations
19.1. Running Working Mode
• Go to the home screen and press
key, <Working modes>
submenu providing list of available working modes is displayed.
• Select the working mode you need to operate, the home screen is
displayed automatically, wherein the top bar of the screen features
pictogram of the selected mode.
19.2. Working Mode Accessibility
You can declare which working modes are to be accessible for an operator
upon pressing
Procedure:
key.
• Enter <Working modes / Accessibility> submenu and make given
working modes accessible.
Where:
Working mode accessible.
Working mode inaccessible.
20. WORKING MODES - Local Settings
In order to set working modes' parameters enter < / Working Modes>
submenu. Particular working modes’ settings feature specific functions.
The functions enable adapting mode operation to your individual needs. Some
of the specific functions are global, i.e. they are applied for all accessible
working modes; refer to the table below:
55
Page 56
Readout
Proximity sensors
Save mode
Auto Threshold
Result control
Autotare
Labelling mode
Statistics
-
-
-
-
-
-
-
-
-
-
-
Remaining specific functions referring directly to a given working mode are
described further down this user manual.
On-screen button
(working mode parameters) that is to be
found in each working mode home screen, provides direct
access to settings of the given working mode.
20.1. Readout
Readout submenu comprises functions allowing you to adjust your weighing
device to ambient conditions of a given workstation.
Filter
Value Release
Autozero
Ambient
conditions
Last digit
Enter this parameter to adjust your weighing device to ambient conditions.
The higher filter level, the longer the indication takes to stabilise. Available
values: Very fast, Fast, Average, Slow, Very slow.
Enter this parameter to adjust rate of stabilisation of the measurement
result. Depending on the selected option, weighing time is either shorter or
longer. Available values: Fast, Fast plus reliable, Reliable.
Enter this parameter to enable automatic control and correction of zero
indication. There are, however, some cases when this function can be a
disturbing factor for the measuring process. e.g. very slow placing of a load
on the weighing pan (load adding, e.g. pouring, filling). In such case, it is
recommended to disable the function. Options:
- function enabled.
Parameter relating to ambient and environmental conditions of the
workstation. Enter this parameter and set 'unstable' value if the ambient
conditions are unfavourable (air drafts, vibrations). Available values:
Stable, Unstable.
Enter this parameter to enable/disable display of the last digit (placed on
the right of the decimal point) of the weighing result. Available values:
Always: all digits are displayed. Never: last digit disabled. When stable:
last digit displayed only when the result is stable.
- function disabled,
56
Page 57
20.2. Save Mode
Parameter allowing you to set mode of sending data from the weighing device
to a peripheral device.
Procedure:
• Enter <Working modes> menu and select respective working mode.
• Enter <Save mode> submenu and select respective save mode.
Save mode options list:
Manual each stable
Manual first stable
Automatic first stable
Automatic last stable
Semi-automatic each stable
Semi-automatic first stable
20.3. Auto Threshold
Manual printout of each stable weighing result above <Auto
threshold>.
Manual printout of the first stable weighing result above
<Auto threshold>.
Automatic printout of the first stable weighing result above
<Auto threshold>.
Automatic printout of the last stable weighing result detected
upon the weight value gets below <Auto threshold>.
Manual printout of each weighing result, where the weight
value is above -LO- threshold; this option requires awaiting for
the stable weighing result.
Manual printout of the first weighing result, where the weight
value is above -LO- threshold; this option requires awaiting for
the stable weighing result.
<Auto threshold> parameter allows you to configure the function of automatic
operation. The next measurement is saved when mass indication gets below
the set net value of <Auto threshold> parameter.
Procedure:
• Enter <Working modes> submenu and select respective working mode.
• Enter <Auto threshold> submenu and provide respective value, next
press
20.4. Result Control
key to confirm changes.
When your weighing device operates with 'result control' mode on, the printout
is carried out only when mass of load placed on the weighing pan is comprised
within MIN, MAX thresholds.
57
Page 58
If the 'result control' mode is on, respective pictograms are displayed on the
right of the weighing result window,
Procedure:
, , .
• Enter <Working modes> menu and select respective working mode.
• Enter <Result control> submenu and set respective option.
Where:
Each weighing is recorded.
Weighings between MIN, MAX thresholds are recorded.
For detailed description on how to set MIN, MAX thresholds
refer to section 8.5.
For MIN=0 and MAX=0
20.5. Autotare
, , pictograms are inactive.
Autotare function allows you to quickly determine net weight for loads with
different tare values, which loads are measured one after another. When
<Autotare> function is on, the operation cycle takes the following steps:
• Make sure that the weighing pan is empty and press zeroing key.
• Load the weighing pan with product packaging (packaging weight value
must be greater than the set value of <Auto threshold> parameter).
• Upon indication stabilization, packaging mass is automatically tared
pictogram is displayed in the upper part of the display).
(
• Put/pour/dispense product that is to be packed into the packaging.
• Unload the weighing pan (remove both product and packaging).
• When the weighing device detects weight value lower than <Auto
threshold> value, the entered tare value gets automatically deleted (
pictogram is no longer displayed).
• When you load the weighing pan with a new product packaging,
analogous set of operations proceeds.
20.6. Labelling Mode
Labelling system is designed to print labels for product marking. The program
generates standard labels, for single products, and cumulative labels, for bulk
boxes.
58
Page 59
Custom functions of <Labelling mode> submenu:
Labels quantity
C labels quantity
Automatic C label triggering For detailed description read section 20.6.1.
20.6.1. Automatic C Label Triggering
Parameter for declaring number of labels to be printed
on the scale-connected printer.
Parameter for declaring number of C labels to be printed
on the scale-connected printer.
Automatic C Label Triggering function allows to trigger printout of C label,
to do it you must define <Mode> and <Threshold> parameters.
Procedure:
• Enter <Working modes> submenu and select respective working mode.
• Go to <Labelling mode / Automatic C label triggering / Mode>
parameter and set respective option, where:
None
Mass
Number
Manual printout of C label *, performed upon pressing either or button.
Automatic printout of C label, triggered upon exceeding of the total mass value
of the single labels, set in <Threshold> parameter.
Automatic printout of C label, triggered upon exceeding single labels quantity,
set in <Threshold> parameter.
*) Manual printout of C labels is carried out using one of the two below
programmable buttons:
Printout with zeroing (weighings qty. and total mass values are zeroed).
Printout without zeroing (weighings qty. and total mass values are not zeroed).
In case of automatic printout of C labels, counter zeroing
function is always on (weighings qty. and total mass values
are always zeroed).
• To confirm introduced modifications, press key and go to
• Set the value that is to trigger C label printout, mind that:
− in case of <Mode> parameter set to <Mass> value, you must provide
total mass value, upon obtaining of which you want the C label
printout to be triggered,
59
Page 60
−in case of <Mode> parameter set to <Counter> value, you must
provide quantity value, upon obtaining of which you want the C label
printout to be triggered.
• Press
20.7. Statistics
key to confirm introduced modifications.
All statistical data is updated in an ongoing manner. The statistical data can be
updated globally (regardless of the weighted product) or individually for each
product that is either weighed or selected from the database.
Procedure:
• Enter <Working modes> submenu and select respective working mode.
• Go to <Statistics> parameter and set respective option.
Where:
Global Update of statistical data carried out globally.
Product
Update of statistical data carried out individually for each product, either
weighed or selected from the database.
21. WORKING MODE – WEIGHING
<Weighing> is a standard working mode enabling you to carry out the
weighing operation along with record of the result to the database.
21.1. Home Screen
60
Page 61
21.2. Local Settings
In order to access local settings, press operation panel key to which
<
Working mode parameters> pictogram is assigned:
Readout For detailed description read section 20.1.
Save mode For detailed description read section 20.2.
Auto Threshold For detailed description read section 20.3.
Result control For detailed description read section 20.4.
Autotare For detailed description read section 20.5.
Labelling mode For detailed description read section 20.6.
Statistics For detailed description read section 20.7.
22. WORKING MODE – PARTS COUNTING
<Parts counting> is a working mode enabling you to determine quantity
of small pieces of the same mass, which determination is done on the basis
of reference weight value of single piece, either determined using the weighing
device or taken form the database.
22.1. Home Screen
22.2. Local Settings
In order to access local settings, press operation panel key to which
<
Working mode parameters> pictogram is assigned:
61
Page 62
Readout For detailed description read section 20.1.
Save mode For detailed description read section 20.2.
Auto Threshold For detailed description read section 20.3.
ACAI For detailed description read section 22.2.1.
Minimum Reference Sample Mass For detailed description read section 22.2.2.
Result control For detailed description read section 20.4.
Autotare For detailed description read section 20.5.
Labelling mode For detailed description read section 20.6.
Statistics For detailed description read section 20.7.
22.2.1. Function of Automatic Correction of Reference Sample Mass
<ACAI> is a specific function enabling you to correct weight value of a single
piece by means of the weighing device program.
Procedure:
• Enter <Working modes / Parts counting / ACAI> submenu and set
respective option.
Where:
Function of automatic correction of reference sample mass disabled.
Function of automatic correction of reference sample mass enabled.
<ACAI> function gets activated for <Parts counting> mode at the moment
of determination of reference sample quantity, it is signalled by display
of
pictogram at the home screen top bar.
There are four conditions of ACAI function operation implemented into
the weighing device program:
1. the weighing result must be stable,
2. pieces quantity must increase,
3. pieces quantity upon adding cannot increase more than twice,
4. pieces quantity upon adding must be comprised within ± 0.3 tolerance
of total value.
If you decide that the reference sample features the right amount of pieces, you
can record weight value of a single piece to weighing device memory and
deactivate the function by pressing
key.
62
Page 63
22.2.2. Minimum Reference Sample Mass
Prior determining single piece weight value you can declare „Minimum
reference sample mass, i.e. minimum weight value for the total mass of all
pieces loaded onto the weighing pan, expressed in reading units.
Procedure:
• Enter <Working modes / Parts counting / Minimum reference sample
mass> submenu and set respective option.
Available values: 1 d, 2 d, 5 d, 10 d.
Total mass of all the pieces loaded onto the weighing pan
must not be lower than value declared in „Minimum reference
sample mass” parameter. Unless this condition is met, the
weighing device displays a message: <Sample mass too low>.
22.3. Setting Reference Sample Mass by Entering Mass of a Single Part
• Press button (Set part mass), <Reference sample mass> edit box
featuring an on-screen keyboard is displayed.
• Enter respective value and press
key to confirm, <Parts counting>
working mode is run with automatically set mass of a single part.
If the value of entered single part mass is greater than max
capacity value, then the following message is displayed:
<Value too high>.
If the value of entered single part mass is lower than 0.1 of the
reading unit, then the following message is displayed: <Value
too low>.
22.4. Setting Reference Sample Mass by Determining Mass of a Single
Part
If the parts are to be weighed in a container, first put the container on
a weighing pan and tare it.
• Press
button (Determine part mass), <Reference sample quantity>
edit box featuring an on-screen keyboard is displayed.
• Enter respective value and press
key to confirm, the following
message is displayed: <Number of parts to be loaded: xx> (where xx –
previously entered value).
63
Page 64
• Load the weighing pan with declared amount of parts. When the indication
is stable (
pictogram is displayed), press key to confirm the
mass.
• Single part mass is calculated automatically, <Parts counting> working
mode is run and the weighing device displays quantity of single parts
(pcs).
Total weight value of all parts loaded onto the weighing pan
cannot be greater than the max capacity value.
Total weight value of all parts loaded onto the weighing pan
cannot be lower than the value determined by „Minimum
reference sample mass” parameter. Unless this condition
is met, the weighing device displays a message: <Sample
mass too low>.
Single part mass must be equal or greater than 0.1 reading
unit. Unless this condition is met, the weighing device
displays a message: <Single part mass too low>.
22.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part
from Database
Upon selecting a product from database, single part mass assigned to the
product (<Part mass> entry) is entered automatically.
Procedure:
• Enter <Parts counting> mode and press operation panel key to which
pictogram (product database) is assigned, next select given product
from the list.
Weight value of a single part must be declared for the selected
product. To declare weight value of a single part, enter
products database and edit the selected product.
22.6. Entering Reference Sample Mass to Weighing Device Memory
Reference sample mass of single part can be entered to products database.
Procedure:
• Press
button, select the given product.
• Determine reference sample mass (read sections 22.3 and 22.4 of this
user manual).
64
Page 65
• Press button (Assign reference sample), reference sample mass for
the product is recorded in <Part mass> entry.
23. WORKING MODE - PERCENT WEIGHING
Percent weighing is a working mode enabling you to compare the measured
sample with the reference mass. The result is expressed in [%]. Reference
mass can be either determined by weighing or entered to weighing device
memory by an operator.
23.1. Home Screen
23.2. Local Settings
In order to access local settings, press operation panel key to which
<
Working mode parameters> pictogram is assigned:
Readout For detailed description read section 20.1.
Save mode For detailed description read section 20.2.
Auto Threshold For detailed description read section 20.3.
Result control For detailed description read section 20.4.
Autotare For detailed description read section 20.5.
Labelling mode For detailed description read section 20.6.
Statistics For detailed description read section 20.7.
65
Page 66
23.3. Reference Sample Mass Determined by Weighing
If the reference sample is to be weighed in a container, first put the container
on a weighing pan and tare it.
• Press
button (Determine reference sample mass), the following
message is displayed: <Put reference sample>.
• Load the weighing pan with reference mass. When the indication is stable
pictogram is displayed) press key to confirm the value.
(
• Now (instead of weight value of the measured load) difference between
values of loaded mass and reference mass is displayed in [%].
23.4. Reference Sample Mass Entered to the Weighing Device Memory
• Now (instead of weight value of the measured load) difference between
values of loaded mass and reference mass is displayed in [%].
23.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part
from Database
Upon selecting a product from database, reference sample mass of single part
assigned to the product (<Reference sample mass> entry) is entered
automatically.
Procedure:
• Enter <Percent weighing> mode and press operation panel key to which
pictogram (products database) is assigned, next select given product
from the list.
• Difference between values of loaded mass and reference mass assigned
to a selected product is displayed in [%].
24. WORKING MODE – DOSING
<Dosing> is a working mode enabling you to dispense product. Two options
are available, manual and automated dispensing.
66
Page 67
24.1. Home Screen
24.2. Local Settings
In order to access local settings, press operation panel key to which
<
Working mode parameters> pictogram is assigned:
Procedure:
• Enter <
• Select <Dosing mode> parameter and set one of the dosing modes:
Manual dosing Manual dosing with MIN, MAX thresholds.
Automatic dosing Automatic 1-, 2-stage dosing with automated activation of dosing outputs.
/ Working modes / Dosing> submenu.
• List of available dosing functions changes depending on the selected
dosing mode.
List of automatic dosing functions
Fast dosing threshold [DT1]
Dosing threshold [DT2]
Fast dosing output
Dosing outputs
Chute outputs
Permission to start Enter to send signal triggering process START.
Enter to set mass value for rough dosing in case of automatic
2-stage dispensing.
Enter to set target mass value to be dosed in case of an
automatic dispensing.
Enter to declare outputs for rough dosing in case of automatic
2-stage dispensing.
Enter to declare outputs for precise dosing in case
of automatic 2-stage dispensing.
Enter to activate chute; signalling informing that the silos
is being emptied switches on.
67
Page 68
Permission to empty
Chute delay [s]
Record delay [s]
Request cycles quantity
Request mass
Automatic taring
Dosing correction
Readout For detailed description read section 20.1.
Auto Threshold For detailed description read section 20.3.
Enter to send signal triggering the process of silos emptying
(e.g. opening of flap).
Enter to set process end delay, counted from the moment
when mass value gets below the set <Auto threshold>
value.
Enter to set delay of record of dosed mass after the dosing
threshold gets reached. The parameter value is declared in
[s], value range is 0 [s] (function inactive) - 300 [s].
Enter to force query regarding number of dosing process
cycles, i.e. to specify how many times the whole process is to
be repeated.
Enter to force query regarding value of mass that is to be
dosed after process start.
Enter to enable/disable automatic taring at the moment
of dosing process start.
Enter to correct the dosed mass, this function allows to
optimise the dosing process, change of pressure of dosed
material in the silos is accounted for. Parameters: Constant correction – specifying global (constant) value of correction
to be applied during each process; Maximum correction –
specifying maximum value of correction, which can be
determined automatically during the process; Measurements quantity for correction calculation – specifying how many
of the most recent measurements is to be analysed in order to
automatically calculate the correction during the process.
List of manual dosing functions
Min Enter to set the value of low threshold of manual dosing.
Max Enter to set the value of max threshold of manual dosing.
Confirm ingredients
manually
Automatic taring
Readout For detailed description read section 20.1.
Auto Threshold For detailed description read section 20.3.
Mark to enforce manual confirmation of dosed mass by means of
pressing
Enter to enable/disable automatic taring at the moment of dosing
process start.
key (the function is enabled for manual dosing mode).
68
Page 69
24.3. Automatic Dosing Procedure
• Save general parameters of Dosing mode to scale memory, the
parameters are to be found upon pressing a key to which
pictogram is
assigned.
• In case of selecting a product (that is to be dosed) form the database the
following message is displayed: <Fast dosing threshold, dosing
threshold, dosing outputs, fast dosing outputs will be obtained from
the product data. Continue?> (
- press to select, - press to
cancel).
• Press key to which
is locked except for
When:
pictogram (process start) is assigned, scale keypad
- stop, - pause, - breakdown keys.
1. Dosing outputs are declared incorrectly (e.g. lack of active dosing output)
the following message is displayed: <Dosing outputs determined incorrectly>, next the process is cancelled.
2. Mass value on the weighing pan (in silos) is greater than declared <Auto
threshold> value or when the weighing result is unstable then the
following message is displayed: Unstable weighing result or mass above „Auto threshold”. The process status changes to PS=Pause and
remains so until stable weighing result is obtained and/or weighing pan
emptied.
• Upon process start, bar graph of dosed material mass is displayed with
the following data:
Top bar pictograms:
process in progress (blinking pictogram).
69
Page 70
Workspace data:
DT1 Fast dosing threshold value.
DT2 Dosing threshold value.
Process status, the status takes the following values: -3.000kg – amount left to be
dosed; OK – target dosing mass obtained; Pause – process inhibited: a) by
means of
PS
Product: Name of a product that is to be dosed.
Tare: Tare value (mass of container, silos, etc.).
Buttons that are assigned to operation panel keys:
b) the start signal is awaited for; c) container empting signal is awaited for; Taring
– taring process in progress; 00:00:05 [s] - automatic countdown of time delay for
record of dosed mass, or automatic countdown of time delay for process end;
Chute – pending for silos empting; Completed – dosing completed; Aborted –
dosing aborted.
Local parameters (button inactive during the ongoing process).
Product selection.
Process start.
Process stop.
Breakdown stop.
(pause) button, to restart the process press button,
Bar graph functioning, 1-stage automatic dosing:
• Visualisation of mass value lower than the value of [DT2] threshold:
DT1=0g
DT2=1800g
• Visualisation of mass value higher than the value of [DT2] threshold:
DT1=0g
DT2=1800g
Bar graph functioning, 2-stage automatic dosing:
• Visualisation of mass value lower than the value of [DT1] threshold:
DT1=1600g
DT2=1800g
70
Page 71
• Visualisation of mass value higher than the value of [DT1] threshold and
lower than the value of [DT2] threshold:
DT1=1600g
DT2=1800g
• Visualisation of mass value higher than the value of [DT2] threshold:
DT1=1600g
DT2=1800g
• Upon exceeding the value of fast dosing threshold, fast dosing outputs are
activated.
• Upon exceeding the value of dosing threshold, dosing outputs are
activated and PS=OK process status is displayed.
• In case of active time delay, countdown of declared time starts.
• After countdown completion, PS=Completed status is displayed (process
completed). Blinking pictogram
24.4. Manual Dosing Procedure
is turned off.
• Save general parameters of Dosing mode to scale memory, the
parameters are to be found upon pressing a key to which
pictogram
is assigned.
• Press key to which
is locked except for
When:
pictogram (process start) is assigned, scale keypad
- stop, - pause, - breakdown keys.
1. Dosing outputs are declared incorrectly (e.g. lack of active MIN or MAX
outputs) the following message is displayed: <Dosing outputs determined incorrectly>, next the process is cancelled.
2. Mass value on the weighing pan (in silos) is greater than declared <Auto
threshold> value or when the weighing result is unstable then the
following message is displayed: Unstable weighing result or mass above „Auto threshold”. The process status changes to PS=Pause and
remains so until stable weighing result is obtained and/or weighing pan
emptied.
• Upon process start, bar graph of dosed material mass is displayed with
the following data:
71
Page 72
Top bar pictograms:
process in progress (blinking pictogram).
Workspace data:
Min
Max
PS
Product:
Tare:
Value of low threshold of manual dosing.
Value of max threshold of manual dosing.
Process status, the status takes the following values: -3.000kg – amount left to be
dosed; OK – target dosing mass comprised between MIN and MAX thresholds;
Pause – process inhibited by
button; Taring – taring process in progress; Completed – dosing completed;
Aborted – dosing aborted.
Name of a product that is to be dosed.
Tare value (mass of container, silos, etc.).
(pause) button, to restart the process press
Buttons that are assigned to operation panel keys:
Local parameters (button inactive during the ongoing process).
Product selection (button inactive during the ongoing process).
Process start.
Process stop.
Breakdown stop.
• If dosed mass value is between MIN and MAX thresholds then PS=OK
status is displayed.
• In case of active <Confirm ingredients manually> function, the user
must confirm mass value manually by pressing
key.
72
Page 73
• Press key to confirm dosed mass, PS=Completed status is
displayed (process completed). Blinking pictogram
24.5. Dosing Process Report
is turned off.
Dosing report is automatically generated at the end of each dosing process,
next it is sent to port selected for <Peripherals / Printer>. To declare report
content go to <Printouts / Dosing report> submenu. For instruction on how to
declare adjustment report settings read section 'Printouts'.
Report example:
---------------Dosing Report-------------Start date 2018.07.10 13:21:40
End date 2018.07.10 13:23:28
Operator Nowak Jan
Fast dosing threshold [DT1] 2.800 kg
Dosing threshold [DT2] 3.000 kg
Dosing mass 3.018 kg
Difference 0.018 kg
Status Completed
Report on each completed process is saved to <Dosing reports> database,
the file name is a combination of date and hour of process completion and
of process status (list of dosing report data – read section 26.2.3).
25. WORKING MODE – FORMULATIONS
Formulations is a working mode enabling you to make multi-ingredient
mixture.
25.1. Home Screen
73
Page 74
25.2. Local Settings
In order to access local settings, press operation panel key to which
Working mode parameters> pictogram is assigned:
<
Request multiplier
Request cycles
quantity
Request lot number Enter to provide series number for formulation.
Confirm ingredients
manually
Automatic taring
Readout For detailed description read section 25.1.
Proximity sensors For detailed description read section 25.2.
25.3. New Formulation
Enter to force query regarding formulation multiplier, i.e. value by
which the weight, high deviation and low deviation values of <mass>
type are to be multiplied.
Enter to force query regarding number of formulation process cycles,
i.e. to specify how many times the whole process is to be repeated.
Enter to enforce manual confirmation for each weighing by pressing
of
Enter to enable automatic taring at the moment of formulation start,
and after weighing of each ingredient.
key.
• Enter < / Databases / Formulations> submenu.
• Press operation panel key to which
(add) pictogram is assigned.
New record is automatically edited. For list of new formulation data read
section 31.6.3 of this manual.
• Enter <Ingredients> submenu, add formulation ingredients one by one,
do it by pressing key to which
List of ingredient data:
Name Ingredient name (43 characters maximum).
Code Ingredient code (16 characters maximum).
Mass Target mass of the ingredient.
Deviation type Declared deviation type: unit of mass or value in [%].
High deviation High deviation of ingredient mass in 'Formulations' mode.
Low deviation Low deviation of ingredient mass in 'Formulations' mode.
Weight entered manually
Platform Assigning weighing platform number to an ingredient.
Manual mode for entering ingredient mass (ingredient is not
weighed).
(add) pictogram is assigned.
Formulation ingredients are acquired from <Products>
database.
74
Page 75
• Modification of ingredient settings is possible upon editing of the selected
ingredient, to do it press key to which
(edit) pictogram is assigned.
When
1. you declare low deviation of value greater than the declared ingredient
mass then the following message is displayed: <Value too high>. Enter
correct low deviation value.
2. you declare high deviation of value greater than the declared ingredient
mass then the following message is displayed: <Value too high>. Enter
correct high deviation value.
• Enter the required data and press button, the new ingredient
is added to the formulation.
• Enter the last ingredient and go to the home screen, do it by pressing
key.
25.4. Formulation Procedure
• Select the formulation, do it by pressing key to which (select
formulation) pictogram is assigned.
• Save general parameters of 'Formulations' mode to scale memory,
the parameters are to be found upon pressing a key to which
pictogram is assigned.
• Press key to which
When:
(process start) pictogram is assigned.
1. The declared value of formulation multiplier causes exceeding of the max
capacity value, then the following message is displayed: <Formulation multiplier value too high>. The process gets cancelled.
2. The declared value of ingredient mass exceeds the max capacity value,
then the following message is displayed: <Ingredient mass value higher than max capacity value>. The process gets cancelled.
3. Sum of ingredient mass and high deviation values exceeds the max
capacity value, then the following message is displayed: <High deviation value too high>. The process gets cancelled.
4. The weighing result is unstable then the following message is displayed:
<Unstable weighing result>. The process status changes to PS=Pause
and remains so until stable weighing result is obtained.
75
Page 76
• Upon process start, bar graph of weighed ingredient mass is displayed
with the following data:
Top bar pictograms:
process in progress (blinking pictogram).
Workspace data:
LD Low deviation value.
HD High deviation value.
Process status, the status takes the following values: -3.000kg – amount left
to be weighed; [1/3] – cycle 1 of 3; OK – target mass obtained; Pause –
PS
Formulation Formulation name.
Ingredient
Buttons that are assigned to operation panel keys:
process inhibited: a) by means of
press
process in progress; Completed – formulation process completed; Aborted –
formulation process aborted.
Local parameters (button inactive during the ongoing process).
Select formulation (button inactive during of the ongoing process).
Process start (button inactive during the ongoing process).
Process stop.
Process pause.
button, b) stable weighing result is awaited for, Taring – taring
(pause) button, to restart the process
76
Page 77
Bar graph operation:
• Visualisation of mass value lower than the value of [LD], low deviation:
LD=1.600kg
HD=1.800kg
• Visualisation of mass value higher than the value of [LD], low deviation,
and lower than the value of [HD], high deviation:
LD=1.600kg
HD=1.800kg
• Visualisation of mass value higher than the value of [HD], high deviation:
LD=1.600kg
• Weigh expected amount of each ingredient.
HD=1.800kg
In the case of trying to accept unstable weighing result, the
following message is displayed: <Unstable measurements>.
• After weighing of the last ingredient, PS=Completed status is displayed
(process completed).
• New formulation process may begin.
25.5. Formulation Report
Formulation report is automatically generated at the end of each formulation
process, next it is sent to port selected for <Peripherals / Printer>. To declare
report content go to <Printouts / Formulation report> submenu. For
instruction on how to declare adjustment report settings read section 'Printout s'.
Report example:
-------------Formulation report-------------Start date 2018.07.10 13:21:40
End date 2018.07.10 13:23:28
Operator Nowak Jan
Target value 3.000 kg
Sum 3.018 kg
Difference 0.018 kg
Status Completed
Report on each completed process is saved to <Formulation reports>
database, the file name is a combination of date and hour of process
completion and of process status (for the list of dosing report data read section
27.2.4).
26. DATABASES
Weighing device software comprises the following databases:
In order to configure databases, enter < / Databases> submenu.
Databases editing is available for operators logged as
Administrator.
26.1. Database Export
Applicable in case of an active IM01.EX communication module
Export of weighing device databases carried out using USB flash drive.
Procedure:
• Enter selected database.
• Activate function keys, to do it press
key.
• Plug the flash drive to the USB port of the IM01.EX communication
module, to do it use PT0084 cable.
• Press operation panel key to which
pictogram (export) is assigned.
• The program automatically saves exported data to a USB flash drive file,
respective messages are displayed to confirm successfully completed
operation.
78
Page 79
• The file name is correlated with the name of exported database:
Product database stores names of all products that can be weighed, counted,
controlled.
List of parameters defined for a product:
Name Product name (43 characters maximum).
Code Product code (15 characters maximum).
80
Page 81
Name 2 Additional name (43 characters maximum).
Code 2 Additional code (15 characters maximum).
3)
Min
Max
3)
Low limit of weighing within ranges (Result control).
High limit of weighing within ranges (Result control).
Tare Tare value (set automatically when the product is selected).
1)
Mass
Deviation type
Low deviation
High deviation
4)
4)
4)
Fast dosing threshold
Dosing threshold
2)
Dosing outputs 2)
Fast dosing outputs
Dosing correction
2)
Nominal value of product mass.
Declared deviation type: unit of mass or value in [%].
Low deviation of ingredient mass in 'Formulations' mode.
High deviation of ingredient mass in 'Formulations' mode.
Mass value for rough dosing in case of automatic 2-stage
2)
dispensing.
Target mass value to be dosed.
Outputs for precise dosing in case of automatic 2-stage
dispensing.
2)
Outputs for rough dosing in case of automatic 2-stage dispensing.
Constant dosing correction value.
Price Unit price of a product.
VAT VAT value given in [%].
Shelf-life time in days Shelf-life time of a product (number of days).
Label Template of a single label assigned to a product.
C Label Template of a cumulative label assigned to a product.
Manual mode for entering ingredient mass (ingredient is not
Weight entered manually
Platform
4)
Assigning weighing platform number to an ingredient.
4)
weighed).
Variable name is conditioned by the working mode. For Weighing and Dosing modes
1)
the variable name is „Mass”. For Parts Counting mode the variable name is „Part mass”.
For Percent Weighing mode the variable name is „Reference sample mass”.
2) Variables available for product in Dosing working mode.
3) Variables NOT available for product in Formulations working mode.
4) Variables available for product exclusively in Formulations working mode.
26.6.3. Formulations
Formulations database features list of formulations that can be carried out
automatically by particular ingredients weighing.
81
Page 82
List of parameters defined for a formulation:
Name Formulation name (43 characters maximum).
Code Formulation code (15 characters maximum).
Ingredients
Target value Preview of total mass of formulation.
26.6.4. Packaging
Defining formulation ingredients, and preview of added formulation
ingredients quantity.
Packaging database features list of product containers. When carrying out
weighing process, upon selection of particular packaging, a respective tare
value is triggered automatically. The tare value is displayed with minus sign.
List of parameters defined for a packaging:
Name Packaging name (43 characters maximum).
Code Packaging code (15 characters maximum).
Mass Packaging weight value (set automatically when the packaging is selected).
26.6.5. Customers
Customers database features a list of names of customers for whom the
measurements are carried out.
List of parameters defined for a customer:
Name Customer name (43 characters maximum).
Code Customer code (15 characters maximum).
TIN Tax identification number (15 characters maximum).
Address Customer address (43 characters maximum).
Postal code Customer postal code (7 characters maximum).
City Customer city (43 characters maximum).
Discount Discount for a customer given in [%].
Label Template of a label assigned to a customer.
26.6.6. Labels
Labels database features a list of label templates that can be assigned
to a product or a customer in order to activate operation in labelling mode.
82
Page 83
List of parameters defined for a label:
Name Label name.
Code Label code.
Project * Label project.
*) An example of how to make and send label template to a scale memory is to be found in ANNEX
02 of this user manual.
26.6.7. Universal Variables
Universal variables database comprises universal variables templates. You can
assign the templates to function keys. This allows you to easily enter any
text/number/letter, that you want to print, to weighing device memory.
Values of 3 universal variables , , , entered to weighing device
memory will be saved to a completed weighing record.
List of parameters defined for a universal variable:
Code Universal variable code (15 characters maximum).
Name Universal variable name (43 characters maximum).
Value
27. REPORTS
Weighing device software comprises the following reports:
Universal variable value, it is to be printed and/or saved to weighing record
(32 characters maximum).
In order to preview, export and delete reports, enter < / Reports>
submenu.
Database of Alibi reports is protected against deletion.
83
Page 84
27.1. Deleting Weighings Report
• Enter reports database.
N/a in case of Alibi reports database
• Activate function keys, to do it press
• Press key to which
pictogram (delete all records of report database)
key.
is assigned. Message <Delete all records?> is displayed.
• Press
key to confirm. All database records get deleted.
27.2. Reports Preview
27.2.1. Weighings
Each weighing result sent from the weighing device to a printer is saved
to weighings report. You can preview data of particular weighings.
List of parameters defined for a completed weighing
Date Performed weighi ng date.
Time Performed weighing time.
Stable Stable weighing result marker
Result Weighing result given in specific unit (%, pcs).
Mass Net weight.
Tare Tare value.
Operator Operator name.
Product Product name.
Customer Customer name.
Lot number Lot number (16 characters maximum).
Batch number Batch number (16 characters maximum).
Result control Weighing result range.
Min Minimum weighing result threshold (result control).
Max Maximum weighing result threshold (result control).
Universal variable 1 Value of universal variable 1.
Universal variable 2 Value of universal variable 2.
Universal variable 3 Value of universal variable 3.
84
Page 85
27.2.2. Alibi
Each weighing result sent from the weighing device to a printer is saved to Alibi
report. You can preview data of particular weighings.
List of parameters defined for a completed weighing
Date Performed weighi ng date.
Time Performed weighing time.
Result Weighing result given in specific unit (%, pcs).
Mass Net weight.
Tare Tare value.
27.2.3. Dosing reports
Dosing report is generated automatically after completion of each dosing
process. You can preview data for particular reports.
Status
Start date Date of dosing process start.
End date Date of dosing process end.
Operator Operator carrying out the dosing process.
Customer Customer for whom the dosing process is carried out.
Dosing mass Dosed net weight in an adjustment unit.
Dosing threshold Target mass value to be dosed.
Fast dosing threshold
Correction Current value of correction in an automatic dosing.
Difference
Min Value of low threshold of manual dosing.
Max Value of max threshold of manual dosing.
27.2.4. Formulations reports
Information regarding correctness of dosing process realisation.
Status values: Aborted, Completed.
Mass value for rough dosing in case of automatic 2-stage
dispensing.
Difference between the net value of mass that is to be dosed and
a value of automatic dosing threshold.
Formulation report is generated automatically after completion of each
formulation process. You can preview data for particular reports.
85
Page 86
Status
Start date Formulation start date.
End date Formulation end date.
Formulation Performed formulation name.
Operator Operator performing formulation process.
Customer Customer for whom the formulation is carried out.
Ingredients quantity Number of formulation ingredients.
Measurements quantity Number of weighings performed within the formulation cycle.
Lot number Lot number assigned to a formulation.
Measurements List of weighings performed within the formulation cycle.
Target value Sum of declared nominal masses of the ingredients.
Sum Total weight value of the performed formulation.
Difference Difference between the sum and the value of expected target weight.
Formulation performance status. Status values: Ongoing, Aborted,
Completed.
28. IMPORT / EXPORT
Import/export option facilitates the following:
• archiving of reports,
• copying databases between weighing devices of the same series,
• copying parameters between weighing devices of the same series.
Import/export operation can be carried out by means of USB flash drive
comprising <FAT files system>.
Procedure:
• Plug the flash drive to the USB port of the IM01.EX communication
module, to do it use PT0084 cable.
• The weighing device detects USB flash drive automatically, as a result
<Import / Export> box is displayed.
28.1. Data Export
Function designed to allow export of databases and/or user parameters.
<Export> submenu comprises the following functions:
Applicable in case of an active IM01.EX communication module
Applicable in case of an active IM01.EX communication module
86
Page 87
• All databases,
• Products,
• Operators,
• Formulations,
• Packaging,
• Customers,
• Non-standard printouts,
• Universal variables,
• Weighings,
• Alibi,
• Dosing processes,
• Parameters.
When you run <All databases> function, files storing particular databases data
are saved to USB flash drive, the files names are correlated with databases
names.
The files extensions are specific (read section 'Databases Export', 26.1),
file-stored data is encoded therefore the files content is not readable for
standard computer programs. Data stored in weighings and Alibi reports files
can be read using special RADWAG-designed PC software.
28.2. Data Import
Applicable in case of an active IM01.EX communication module
<Import> function allows you to copy databases and operator's parameters
between weighing devices of the same series. It is both quick and reliable
method of entering data error-free. <Import> submenu comprises the following
functions:
• All databases,
• Products,
• Operators,
• Formulations,
• Packaging,
• Customers,
• Non-standard printouts,
• Universal variables,
• Parameters.
Reports data cannot be imported.
87
Page 88
29. ERROR MESSAGES
30. ANNEX 01 – Printout Variables
30.1. Variables List
Each defined variable must be inserted in between curly
bracket: {x}, where x – variable number.
List of variables defining non-standard printout templates and data displayed
within the workspace.
No. Description
{0} Standard printout in an adjustment unit
{1} Standard printout in a current unit
{2} Date
{3} Time
{4} Date and Time
{6} Net weight in a current unit
{7} Net weight in an adjustment unit
{8} Gross weight
{9} Tare
{10} Current unit
{11} Adjustment unit
{12} Min threshold
{13} Max threshold
{14} Lot number
88
Page 89
{15} Statistics: Number
{16} Statistics: Sum
{17} Statistics: Average
{18} Statistics: Min
{19} Statistics: Max
{20} Statistics: Gross sum
{21} Gross weight value in the current unit
{22} Net weight in [lb] unit
{23} Result control (status)
{24} Net price
{25} Gross price
{26} Total net price
{32} Serial number
{33} Reading unit
{34} Range (Max capacity)
{35} Parts counting: Reference sample mass
{36} Percent weighing: Reference sample mass
{37} Statistics: Standard deviation
1)
{39}
Universal variable: Value
{41} Batch number: Value
{45} Parts counting: Reference sample quantity
1)
{49}
Universal variable: Name
{50} Product: Name
{51} Product: Code
{52} Product: Name 2
{53} Product: Code 2
{54} Product: Mass
{55} Product: Price
{56} Product: Tare
{57} Product: Min
{58} Product: Max
{59} Product: Shelf-life time in days
{60} Product: VAT
{62} Product: Expiry date (current date + shelf-life time in days)
{75} Operator: Name
{76} Operator: Code
{77} Operator: Permissions
{80} Packaging: Name
89
Page 90
{81} Packaging: Code
{82} Packaging: Mass
{85} Customer: Name
{86} Customer: Code
{87} Customer: TIN
{88} Customer: Address
{89} Customer: Postal code
{90} Customer: City
{91} Customer: Discount in%]
{100} Dosing Report: Start date
{101} Dosing Report: End date
{102} Dosing Report: Fast dosing threshold
{103} Dosing Report: Dosing threshold
{104} Dosing Report: Dosing correction
{105} Dosing Report: Dosing mass
{106} Dosing Report: Difference
{107} Dosing Report: Status
{120} Formulation report: Formulation name
{121} Formulation report: Formulation code
{122} Formulation report: Start date
{123} Formulation report: End date
{124} Formulation report: Ingredients quantity
{125} Formulation report: Measurements quantity
{126} Formulation report: Sum
{127} Formulation report: Difference
{128} Formulation report: Status
For {39}, {49} variables, each database entry (1,2-n) must be
formatted as follows: Entry 1 - {39:1}, {49:1}, Entry 2 - {39:2},
{49:2}, etc.
30.2. Variables Formatting
You can format numeric variables, text variables and dates which are to be
printed or displayed in the grey workspace.
90
Page 91
Formatting types:
• variables with left justification,
• variables with right justification,
• determining quantity of characters for printing / displaying,
• determining quantity of decimal places for numeric variables,
• converting the format of date and hour,
• converting numeric variables into EAN13 code,
• converting numeric variables and date into EAN128 code.
Special formatting characters:
Character Description Example
, Sign separating variable from
format item.
- Either minus sign or left
justification.
: Either sign proceeding format
item or time separator (inserted
between hour, minute and
second).
F Either sign formatting digits to a
string of “-ddd.ddd…” format
(where: d - single digit, minus for negative values only) or
determination of decimal places
quantity.
V Formatting mass and quantities
being derivatives of mass in
EAN13 code.
T Formatting mass and quantities
being derivatives of mass in
EAN128 code.
/ Date separator separating days,
months and years.
. Date separator separating day,
month and year, and time
separator separating hour,
minute and second.
- Dash as a date separator
between day, month and year,
or as a time separator between
hour, minute and second.
{7,10} - Net weight value of fixed length (10
characters), given in an adjustment unit, right
justification.
{7,-10} - Net weight value of fixed length (10
characters), given in an adjustment unit, left
justification.
{7:F3} - Net weight value given in an adjustment
unit, always with three decimal places.
{3:HH:mm:ss 24H} – Current time in format:
hour-minute-second.
{7:F2} - Net weight value given in an adjustment
unit, always with two decimal places.
{7,9:F2} - Net weight value of fixed length
(9 characters), given in an adjustment unit,
always with two decimal places, right
justification.
{7:V6.3} - Net mass in a form of EAN13 code
(6-character code) with three decimal places.
{7:T6.3} - Net mass in a form of EAN128 code
with three decimal places.
{2:yyyy/MM/dd} – Current date in format: year /
month / day.
{2:yyyy.MM.dd} – Current date in format:
year.month.day.
{3:HH.mm.ss 24H} – Current time in format:
hour.minute.second.
{2:yyyy-MM-dd} – Current date in format:
year-month-day.
{3:HH-mm-ss 24H} – Current time in format:
hour-minute-second.
91
Page 92
Formatted variables application:
CODE DESCRIPTION
{6:V6.3} Net weight value in a current unit in EAN 13 code (6-character code)
{6:V7.3} Net weight value in a current unit in EAN 13 code (7-character code)
{7:V6.3} Net weight value in an adjustment unit in EAN 13 code (6-character code)
{7:V7.3} Net weight value in an adjustment unit in EAN 13 code (7-character code)
{8:V6.3} Gross weight value in an adjustment unit in EAN 13 code (6-character code)
{8:V7.3} Gross weight value in an adjustment unit in EAN 13 code (7-character code)
{16:V6.3} Total net weight value in EAN 13 code (6-character code)
{16:V7.3} Total net weight value in EAN 13 code (7-character code)
{20:V6.3} Total gross weight value in EAN 13 code (6-character code)
{20:V7.3} Total gross weight value in EAN 13 code (7-character code)
{21:V6.3} Gross weight value in a current unit in EAN 13 code (6-character code)
{21:V7.3} Gross weight value in a current unit in EAN 13 code (7-character code)
{24:V6.3} Net amount due in EAN 13 code (6-character code)
{24:V7.3} Net amount due in EAN 13 code (7-character code)
{25:V6.3} Gross amount due in EAN 13 code (6-character code)
{25:V7.3} Gross amount due in EAN 13 code (7-character code)
{26:V6.3} Total net amount due in EAN 13 code (6-character code)
{26:V7.3} Total net amount due in EAN 13 code (7-character code)
{6:T6.3} Net weight value in a current unit in EAN 128
{7:T6.3} Net weight value in an adjustment unit in EAN 128
{8:T6.3} Gross weight value in an adjustment unit in EAN 128
{16:T6.3} Total net weight value in EAN 128 code
{20:T6.3} Total gross weight value in EAN 128 code
{22:T6.3} Net weight value (lb) in EAN 128 code
{55:T6.2} Product price in EAN 128 code
{2:YYMMDD} Date in EAN 128 code
{62:YYMMDD} Product expiry date in EAN 128 code
92
Page 93
31. ANNEX 02 – Label Template
Label template can be created:
• via scale, using available variables,
• via Label Editor R02 PC software.
For detailed description concerning creation of label
template via PC software, read user manual of „Label Editor
R02”.
In order to work in label scale mode, assign a label with template to a particular
product or customer.
31.1. Creation of Label Template via the Scale
• Enter <Databases / Labels> submenu and select respective entry.
• Enter <Project> parameter, <Project> edit box is displayed.
• Modify selected label project and press
31.2. Sending Label Template to Scale Memory
key to confirm.
• Copy label template with *.lb extension, made using „Label Editor R02”,
onto the USB flash drive.
• Connect the USB flash drive to USB port of an active IM01.EX
communication module.
• Enter <Databases / Labels> submenu and select respective entry.
• Enter <Project> parameter, <Project> edit box is displayed.
• Press
key that is a counterpart of pictogram (import), box with
USB flash drive content is displayed.
• Select previously copied label template with *.lb extension, the scale
program inserts selected template into the edit box.
• Press
31.3. Assigning a Label with a Template to a Product
key to confirm.
• Enter <Databases / Products> submenu and select respective entry.
• Enter <Label> parameter, label database with list of existing labels
is displayed.
• Select respective label, it is automatically assigned to a product.
93
Page 94
31.4. Assigning a C Label with a Template to a Product
• Enter <Databases / Products> submenu and select respective entry.
• Enter <C Label> parameter, label database with list of existing C labels
is displayed.
• Select respective label, it is automatically assigned to a product.
31.5. Assigning a Label with a Template to a Customer
• Enter <Databases / Customers> submenu and select respective entry.
• Enter <Label> parameter, label database with list of existing labels
is displayed.
• Select respective label, it is automatically assigned to a customer.
31.6. Printing a Label with a Template
• Go to the home screen, select a product or a customer with a label
assigned from a database.
• Load the weighing pan, wait for a
pictogram to display, press
key.
• The label is printed via scale-connected printer.
The label can be assigned to a product or a customer. Label
template assigned to the most recently selected database
entry (a product or a customer) is printed on a scale-
connected printer upon pressing
key.
32. ANNEX 03 -ZEBRA Printer Configuration
Transmission parameters Printout data regarding RS232
Baud rate – 9600 b/sec
Parity control – none
Bits qty – 8 bits
Stop bits – 1 bit
For detailed procedure specifying how to print info printout and how to modify
printer settings read user manual of ZEBRA printers.
Serial port : 96, N, 8, 1
94
Page 95
33. ANNEX 04 - Barcode Scanner Configuration
1. For communication with barcode scanners, RADWAG scales use RS232
interfaces and simplex transmission (one-way direction), without
handshaking. For this purpose, use of second line of the cable is sufficient.
Used scanners should be equipped with RS232 interface, both hardware
and software handshaking must be disabled.
2. Transmission parameters can be set for both weighing devices and
scanners. For both devices the following parameters must comply: baud
rate, data bits quantity, parity control, stop bits quantity; e.g. 9600,8,N,1 –
baud rate 9600 bit/s, data 8-bits, no parity control, 1 stop bit.
3. Barcode scanners can send additional information apart from the expected
barcode e.g. symbology (type of barcode). Due to the fact that RADWAG
devices and software do not use such information, it is advisable to
disable it.
4. Some RADWAG systems can omit unnecessary coded information by
using parameters that mark the beginning and the length of the code
subjected to analysis.
5. A special protocol is required in order the code be received by RADWAG
equipment. It is required to program an appropriate prefix and suffix. In
RADWAG-adopted standard, the prefix is 01 sign (byte) hexadecimal
format, the suffix is 0D sign (byte) hexadecimal format.
6. Most barcode scanners allow to enable/disable different symbologies
(barcode types).
7. Programming of scanners is performed by reading special barcodes.
8. Scanners marketed together with RADWAG systems are always
configured according to the rules above.
Barcode with required prefix and suffix in
hexadecimal format