RADWAG PUE C User Manual

User manual
of PUE C/31 H/EX indicator applicable in:
WPT……../EX, WTC……./EX.
BALANCES AND SCALES
RADWAG 26 – 600 Radom, Bracka 28, POLAND Phone +48 48 38 48 800, phone/fax: +48 48 385 00 10 Sales department +48 48 366 80 06
www.radwag.com
- 2 -
JUNE 2014
EXI-01-04-06-14-GB
- 3 -
TABLE OF CONTENTS
1. DESIGN AND INTENDED USE .......................................................................................... 4
2. SECURITY REQ UIREMENTS............................................................................................. 6
2.1. ATEX marking – symbols meaning .............................................................................. 7
2.2. Rating plates for indicator and scale ............................................................................ 8
2.3. Placement of information stickers ................................................................................ 9
3. START-UP ........................................................................................................................ 10
3.1. Scale placing ............................................................................................................. 10
3.2. Connecting power supplier ........................................................................................ 11
3.3. Switching on .............................................................................................................. 14
4. CLEANING ....................................................................................................................... 14
5. SERVICE AND REPAIRING ............................................................................................. 14
6. FUNCTIONS ..................................................................................................................... 15
7. KEYBOARD ..................................................................................................................... 16
8. FUNCTINS OF KEYS ....................................................................................................... 16
9. INSCRIPTIO NS AND PICTOGRAMS ............................................................................... 17
10. USER MENU ................................................................................................................... 17
11. BROWSING USER’S MENU ........................................................................................... 18
11.1.Keypad…. ................................................................................................................ 18
11.2.Return to weighing .................................................................................................... 19
12. WEIGHING...................................................................................................................... 19
12.1.Tarring….. ................................................................................................................ 19
12.2.Tarring by inscribing tare value ................................................................................. 20
12.3.Zeroing…. ................................................................................................................ 21
12.4.Weighing for two range scales .................................................................................. 21
12.5.Setting of basic unit for weighing .............................................................................. 22
12.6.Temporary weighing unit .......................................................................................... 23
13. MAIN PARAMETERS ..................................................................................................... 24
13.1.Adjustment of filtering level ....................................................................................... 24
13.2.Autozero.. ................................................................................................................. 25
13.3.Tare function operation ............................................................................................. 26
13.4.Median filter .............................................................................................................. 27
13.5.Determining the minimal mass .................................................................................. 28
14. OPERATION MODES ..................................................................................................... 28
14.1.Enabling operating modes (functions) ....................................................................... 28
14.2.Determining number of accessible modes ................................................................ 29
14.3.Counting parts of identical weight ............................................................................. 30
14.4.+/- control referring to the inscribed standard mass .................................................. 33
14.5.Per cent deviation with reference to standard mass .................................................. 35
14.5.1.Mass standard weight determined by the measuring process ................................ 35
14.5.2.Standard mass inscribed to scale memory ............................................................. 36
14.6.Automatic tare .......................................................................................................... 37
14.7.Measurement of the max force – peak hold .............................................................. 38
14.8.Totalizing. ................................................................................................................. 38
14.8.1.Operation mode activation procedure .................................................................... 38
14.8.2.Totalizing procedure .............................................................................................. 39
14.8.3.Storing the last sum of weighings .......................................................................... 40
15. ADJUSTMENT ................................................................................................................ 41
15.1.Adjustment procedure .............................................................................................. 41
15.2.Start mass estimation ............................................................................................... 43
16. ERRORS ......................................................................................................................... 45
17. RESPECTIVE STANDARDS .......................................................................................... 45
EXI-01-04-06-14-GB
- 4 -

1. DESIGN AND INTENDED USE

Indicator of PUE C/31H/EX type, in cooperation w ith a weighi ng platf orm , is intended for fast and precise meas urements performed i n explosive areas . Tarring throughout the who le measuring range allo ws to determine the net mass of measured loads.
The PUE C/31H/EX series indicator consists of:
Main PCB (scale)
Stainless steel housing
Power supply connector
Input cable gland (from weighing platform)
Functional grounding clamp
Overlay
Indicator holder
View - dimensions
Dimensions of PUE C/31H/EX indicator.
EXI-01-04-06-14-GB
- 5 -
PUE C/31H/EX
o
o
125
1200
I/O marking +5V, E, AGND, REF+, REF-, +IN, -IN
Parameters identical for each clamp
KDB 06ATEX250
PUE C/31H/EX technical data:
Housing type stainless steel Ingress Protection Rating IP66/67 Display type LCD Keyboard Microswitch (500 000 cycles) OIML class III Maximum quantity of A/D converter divisions 8388608 Number of verification intervals 6 000 Maximum input voltage 19,5 mV Maximum voltage per verification interval Minimum voltage per verification interval Ambient temperature -10 Minimum load cell impedance Maximum load cell impedance Connection of load cell 4 or 6 cables + shield Number of ranges 1 or 2 ranges
Power supply - DC IN connector pin 1(4) pin 2(3)
Ui=9V Ii=0,24A Pi=0,6W
Ci≈0 Li≈0
3,25 µV
1 µV
C ≤ ta ≤ 40
C
DESIGNATION
EXI-01-04-06-14-GB
Uo=7,14V
Po=0,52W
Io=0,24A
Li≈0
Ci=6uF Lo=40uH Co=10nF
II 2G Ex ib IIC T 4
- 6 -

2. SECURITY REQUIREMENTS

Please read this manual carefully prior the first usage. Following the instruction, use the device as intended;
PUE C/31H/EX indicator, and scales with design based on the indicator, can be used in explosive z one 1 and 2 where ther e is a mixtures of gases, vapours and mists of II (second) explosion group and T1, T2, T3 or T4 temperature class. WPT………/EX and W T C………/EX s cales, b ecaus e of the constr uction of non-electric part of the platf orm and used materials, can be operated in 1 and 2 zone where t here are mixtures of gases , vapours and mists of IIA, IIB, or IIC group – according to „Technical Documentation” of the scale.
Anti-explosive safety of the PUE C/31H/EX indicator, and of scales with design based on the indicator is assured by:
Power supplied from the ZRi02 power supplier II (2) G [Ex ib] IIC KDB 06ATEX251 manufactured b y RADWAG, placed behind the hazardous area or from another adapter with appropriate parameters of intrinsically safe electrical circuit.
Intrinsically safe design of PUE C/31H/EX indicator complying with: EN 60079-0 and EN 60079-11 standards, confirmed by KDB 06ATEX250 certificate
Using load cells certified for conformity with ATEX directive (on anti-explosion safety)
Making non-electric parts pf the scale in compliance with EN 13463-1 standard
Observing recommendations in this manual by users
- marking of paragraphs crucial for anti-explosion safety
EXI-01-04-06-14-GB
- 7 -
Electrical equipment -
Explosive atmosphere:
Symbol of the
Gas explosiveness group -
Temperature class –
Equipment group:
Categories for group II
zone 2
Security level

2.1. ATEX m ar ki ng – symbols meaning

I – underground parts of mines II – all other places liable to be endangered by explosive atmospheres
equipment:
1 - the device assures a
very high protection level,
- to be operated in zones 0,1,2
2 - the device assures a
high protection level,
- to be operated in zones 1,2
3 - the device assures a
standard protection level,
- to be operated in
G – caused by gas,
vapor and mist.
D – caused by dust.
compatible with one or few kinds of anti­explosion designs
used anti-explosion design type
o – oil protection p – gas protection with an
excess pressure q – sand protection d – flame-proof protection e – strengthened design
intrinsically safe design
ia -
for operation in zone 0,1,2,
ib -
intrinsically safe design
for operation in zone 1,2,
examples:
-IIA: propane (T1)
-IIB: ethylene (T2)
-IIC: acetylene (T2)
:
benzene (T3) butane (T2) ethanol (T2)
hydrogen (T1)
surface max temperature of the equipment parts that are in touch with explosive
:
mixture
T1: 450 T2: 300 T3: 200 T4: 135 T5: 100 T6: 85
Temperature class of the gas or vapor auto ignition
T1: > 450 T2: > 300 T3: > 200 T4: > 135 T5: > 100 T6: > 85
Gas atmosphere:
- Ga
- Gb
- Gc Dust atmosphere:
- Da
- Db
- Dc
0
C
0
C
0
C
0
C
0
C
0
C
:
0
C
0
C
0
C
0
C
0
C
0
C
EXI-01-04-06-14-GB
- 8 -
PUE C/31H/EX indicator is a device of group II with „ ib ” pr otec tio n theref or e “Gb” protection symbol may be omitted.

2.2. Rating plates for indicator and scales

Rating plate of the scale (exemplary)
Rating plate of the PUE C/31H/EX indicator
1 – Manufacturer’s Logo 2 – Name and phone number of the manufacturer 3 – Scale type, indicator type 4 – Factory number 5 – Metrological parameters of the scale
6 – Technical Documentation number 7 – Type approval certificate number (for verified scales) 8 – Ambient temperature 9 – geographical zone, ie. latitude and long itu de data and the hei ght abo v e
the level – only verified scales
EXI-01-04-06-14-GB
- 9 -
Protective stickers
Rating plate of the PUE C/31H/EX indicator
Protective sticker Power supply marking
Overlay with ATEX marking
10 – WEEE symbol 11 – ATEX marking relating to the scale (see point 2.1) 12 – EX marking stating that the device is intended for operation in
explosive areas 13 – accuracy class marking 14 – certificate number of used load cell 15 – CE marking with two last digits of the year of production and with a
number of notified body for NAWI directives 16 – ATEX certificate number relating to an indicat or
17 – CE marking with the number of a notified body supervising EX device
production
18 – ATEX marking relating to the indicator (see point 2.1) 19 – manufacturer website address

2.3. Placement of information stickers

EXI-01-04-06-14-GB
- 10 -
Functional grounding mark
Protective sticker Verification sticker
Rating plate of the scale

3. START-UP

Please read this manual carefully prior the start-up. Following the instruction, use the device as intended.

3.1. Scale placing

The scale should be unpacked outside the hazardous area.
Before moving the scale to the place of operation prepare a cable
for earthing the device against electrostatic charges.
NOTICE!! In order to protect the indicator against gathering electrostatic charge it is necessary to earth/zero its housing. Connect the earthing lead/wire to cable connection points
marked with „
Situate scales and an indicat or in hazardous area and connect earthing/zeroing.
EXI-01-04-06-14-GB
” . See the manual for details about ea rthing.
- 11 -
NOTICE!
Intrinsicaly safe circuit parameters for power supplier used for PUE C31H/EX indicator
Uo
8,61
V
Io
0,24
A
Po
0,52
W
Um
255
V
Lo
200
uH
Co
5,7
uF
Li
0
uH
Ci
0
uF
Mounting the indicator, placing the weighing platform and earthing/zeroing should be performed when the explosive atmosphere cannot appear.
Indicators with cable can be mounted on the rack, on the wall or on the table using a special mounting kit. Weighing platform should be placed on even and stable ground far from sources of hit and draughts. It should be also levelled out using the feet.

3.2. Connecting power supplier

1. Place the power supplier outside the hazardous area.
2. Connect the plug that terminates the supply cable (up to 20m) to the socket marked as DC IN.
3. Connect the supplier to mains (230V AC).
4. In case of not using the scale unplug it from the mains.
NOTICE!
Scales with PUE C/31H/EX indicators can be supplie d o n ly from ZRi02 power suppliers
II (2) G [Ex ib] IIC KDB 06ATEX251 of RADWAG production, placed outside the hazardous area, or from another supplier characterized by below presented parameters of intrinsically safe electrical circuit.
EXI-01-04-06-14-GB
- 12 -
Safe area
Hazardous area EX
AC 230v 50Hz
Safe area
Hazardous area EX
AC 230v 50Hz
ZRi02 power
ZRi02 power
Scale with the indicator on a cable
Scale with the indicator on a pillar
supplier
cable up to 20m
supplier
cable up to 20m
Supplier connection
EXI-01-04-06-14-GB
- 13 -
1 – earthing terminal
2 – power supply socket description
3 – power supply socket 4 – power supply socket
5 – screened load cell cable
Power connector parameters
EXI-01-04-06-14-GB
- 14 -

3.3. Switching on

Press the ON/OFF key – keep pressing for about 1 sec.
After switching on, wait for the test completion.
The indication shall equal zero and the following pictograms shall be
seen:
- zero indication
- equilibrium
kg - weighing mode
If the indication does not equal zero press the zeroi ng key
.
Ready to use.

4. CLEANING

Cleaning can be performed only after unplugging from mains (230V AC). Use agents and materials that do not cause gathering electrostatic charges on elements of the scale!
Do not use abrasive agents and solvents because it can cause damage of the casing.
Cleaning should be performed when the appearance of explosive gases is impossible

5. SERVICE AND REPAIRING

EXI-01-04-06-14-GB
- 15 -
In case of any problems with appropriate operation of the indicator, contact the manufacturer’s service centre or any service centre authorised by RADWAG. In case of any fault or dysfunction the device should be passed to the manufacturer’s s ervice centre or, if this is not possible, the fault should be reported in order to settle the conditions of repair.
Any intrusion (alteration, repair etc.) in the indicator or scale based on PUE C/31H/EX indicator, by unauthorised persons will cause a forfeiture of validity of certificates, declarations and manufacturer’s warranty.

6. FUNCTIONS

Filtering
Autozero
Minimal mass
Counting pieces
+/- control
Deviation in per cents
Peak hold
Automatic tare
Tare memory
Manual tare inscription
User’s calibration
Totalizing
It is possible to disable selected functions thus making the scales adjusted to one’s own needs. Access may be permitted to some particular functions only if required. The user menu allows enabling/disabling a given parameter, see the further part of the manual.
EXI-01-04-06-14-GB
- 16 -
Information window
Display window
Function keys – „keypad”

7. KEYBOARD

8. FUNCTINS OF KEYS

NOTICE:
Pressing wherein the change is valid until setting parameters procedure is completed. See the further part of the manual for details.
EXI-01-04-06-14-GB
Power on/off – keep it pressed for about 1 sec
Function key (selection of operation mode)
Confirmation of changes
- zeroing
- tarring
and k eys results in change of particular keys functions,
- 17 -
FIL
PCS
HiLo
PCS
kg (g)
Net
P1 rEAd
P 1.1 Fil
| 2 P 1.2 Auto
|
YES
P 1.3 tArA
|
no
P 1.4 Fnnd
|
no
P2 Prnt
P2.2 S_Lo
|

9. INSCRIPTIONS AND PICTOGRAMS

O.n. Inscription Description
1.
2.
3.
4. Auto
5. toP
6. Add
7.
8.
9.
10.
11.
12. Min
13. OK.
14. Max
Filtering level Parts counting +/- control in relation to reference mass Control and correction of zero indication of scal e Measurement of max. force influencing weighing pan Totalizing
Scale in autozero range (indication = precise zero) Result of measurement is stable (ready for readout) Scale in operating mode counting pieces Scale in operating mode weighing Scale is tarred.
+/- control in relation to reference mass – setting lower threshold or mass below first range
+/- control in relation to reference mass: mass of load between set threshold
+/- control in relation to reference mass – setting upper threshold or mass above upper threshold.

10. USER MENU

The menu is divided into 5 basic groups. Each group has its individual name starting with the capital letter P.
EXI-01-04-06-14-GB
- 18 -
P3 Unit
P3.1 StUn
|
kg
P4 Func
P4.1 FFun
|
ALL
P4.2 Funi
|
no
P4.3 PcS
|
no
P4.4 HiLo
|
no
P4.5 PrcA
|
no
P4.6 Prcb
|
no
P4.7 AtAr
|
no
P4.8 toP
|
no
P4.9 Add
|
no
P6 CAL
P6.1 St_u
|
* FUNCTION *
P6.2 uCAL
|
* FUNCTION *

11. BROWSING USER’S MENU

User moves in the menu using the scale keyboard.

11.1. Keypad

EXI-01-04-06-14-GB
+
entering the main menu
inscribing tare value
+
selecting parameter of a given menu level or changing the active parameter value
entering selected submenu or activation of a parameter for modification
change confirmation
leaving a function without changes or moving one level higher in the m enu
- 19 -

11.2. Re tu rn to weighing

NOTICE:
Changes introduced into the scale memory will be permanently saved after return to weighing.
Press F key
several times, keep pressing until message
<SAvE?> is seen.
Return to weighing
Upon noticing the message press: key – for changes confirmation or
key – for resigning. After completion of this procedure the scale returns
to weighing.

12. WEIGHING

Place a load on a weighing pan. After noticing stability pictogram,

12.1. Tarring

In order to determ ine a net mass , deposit a package of load on a pan and
, read the weighing result.
wait for the indication to b e stable, next press TARE key -
(indication
goes back to zero, Net symbol is displayed on the left side of the display).
After depositing a load on a pan, the display indicates net mass. Tarring can be done repeatedly in full weighing range of the scale. When using tare
EXI-01-04-06-14-GB
- 20 -
Select the digit value
Select the digit to be set
function pay attention to maximum weighing range, it cannot be exceeded. After removing the load and the package, display shows indication of sum of tarred weights with minus sign.
NOTICE:
Tarring may not be performed when minus value or zero is displayed. In such a case the display shows <Err3> message, short sound signal is heard.

12.2. Tarring by inscribing tare value

It is possible to inscribe a tare value. Procedure:
Press simultaneously
and keys
On the display you will see:
– select the digit to be set
– select the digit value
Using
Press
The scale returns to the weighing mode and the value of tare is
displayed with the “–“ sign,
Tare value can be inscribed any time in weighing mode.
and keys set tare value,
,
EXI-01-04-06-14-GB
- 21 -
NOTICE:
Tarring by inscribing tare value cannot be performed when the tare value has already been implemented into the scale memory. In such a case the display shows <Err3> message, short sound signal is heard.

12.3. Zeroing

In order to zero the indic ation press zeroing k ey - . Zero value is shown on a display next to symbols:
and .
The new zero point is treated b y the scale as the precise zero. Zeroing is possible while stable status is shown on the display.
NOTICE:
Display status zeroing is possible only in range up to ±2% of maximum scale capacity. If zeroed value is bigger than ±2% of maximum capacity, display shows Err 2 message, short sound signal is heard.

12.4. Weighing for two range scales

Switching from the I range to the II range is carried out automatically after exceeding the switching point value (Max of the I range).
Weighing in the II ra nge is s igna lled by a pictogram, located in the upper left-hand corner of a display.
After unloading, the scale returns to zero. Weighing in the II range takes place until unloading the pan.
II range weighing automatically turns to I range weighing after taking the load off the weighing pan when the scale returns to AUTOZERO zone – the
pictogram is seen.
EXI-01-04-06-14-GB
- 22 -

12.5. Setting a basic weighing unit

The user may set a weighing unit to be active after turning on the scales.
Procedure:
Following point 11 of this manual enter <P3.Unit> submenu, next:
Press key several times to view available units:
Selection:
A. When the basic unit is set to [kg], users can toggle between: [kg, lb,
N]. For verified scales [lb] is not accessible.
B. When the basic unit is set to [g], users can toggle between: [g, ct].
For verified scales [lb] is not accessible.
Upon unit selection press following win do w :
key, wait for the scales to return to the
EXI-01-04-06-14-GB
- 23 -
Save changes and return to weighing:
See point - 11.2. – return to weighing.
NOTICE:
Upon activation the scale operates with a set basic unit.

12.6. Temporary weighing unit

This function allows users to choose unit for indication. The unit is valid until the scale is switched off or until another unit is chosen. In order to perform the following procedure, operation mode <P4.2.Funi> must be accessible, see point 14.1.1 – Enabling operating modes.
Procedure:
Press F key
, next:
Confirm selection of a given unit and wait for the scale to return to weighing.
Selection:
A. When the basic unit is set to [kg], users can toggle between: [kg, lb
– not accessible for verified scales, N].
B. When the basic unit is set to [g], users can toggle between: [g, ct,
lb – not accessible for verified scales].
EXI-01-04-06-14-GB
- 24 -

13. M AIN PARAMETERS

User can adjust the scale to external ambient conditions (setting filters) or individual requirements (autozero, tare value). Respective parameters can be found in <P1.rEAd> parameters group.

13.1. Adjustment of filtering level Procedure:

Following point 11 of this manual enter <P1.rEAd> submenu, next:
1 - 4 - filtering levels to be selected (depending on the ambient
conditions)
Return to weighing:
See point - 11.2. – return to weighing.
NOTICE:
The higher filtering level the longer stabilisation time.
EXI-01-04-06-14-GB
- 25 -

13.2. Autozero

„AUTOZERO” function has been introduced in order to ensure precise indication. The intended use of this function is automatic control and correction of zero indication.
When function is active comparison of results takes place at specified time intervals. If results differ by value smaller than declared AUTOZERO range, e.g. one division, scale is zeroed aut omatically, stable state is marked –
and zero indication – is displayed.
When AUTOZERO function is enabled, each result starts from precise zero. However there are some cases when this function can disturb measurement, e.g. very slow process of load depositing on the pan (load pouring). In such case, correction system for zero indication can also correct indication of real mass of load.
Procedure:
Following point 11 of this manual enter <P1.rEAd> submenu, next:
AUTO no - autozero disabled AUTO YES - autozero enabled
Return to weighing:
See point - 11.2. – return to weighing.
EXI-01-04-06-14-GB
- 26 -
tArA
AtAr
-
automatic tare function on - stored in scale memory
tArA
no
-
automatic tare function off (scale tarring by means of
tArA
tArF
-
tare memory function – stores last value of tare in

13.3. Tare fu n ction operation

This function enables the user to set appropriate parameters (depending on needs) for tare function.
Procedure:
Following point 11 of this manual enter <P1.rEAd> submenu, next:
after unplugging it from mains (see point 15.5 for description);
T key )
scale memory. It is automatically displayed after starting the scale.
Return to weighing:
EXI-01-04-06-14-GB
See point - 11.2. – return to weighing.
- 27 -

13.4. Median filter

The role of the median filter is to eliminate short lasting interference (e.g. mechanical shocks).
Procedure:
Following point 11 of this manual enter <P1.rEAd> submenu, next:
Fnnd no - median filter disabled Fnnd YES - median filter enabled
Return to weighing:
See point - 11.2. – return to weighing.
EXI-01-04-06-14-GB
- 28 -
Select the digit to be set
Select the digit value

13.5. De termining the minimal load S_Lo parameter works for automatic tare.

The rule is that no automatic tare is performed until the indication drops below S_Lo value (gross).
Procedure:
Following point 11 of this manual enter <P2.Prnt> submenu, next:
Return to weighing:
See point - 11.2. – return to weighing.

14. OPERATION MODES

14.1. En abling operating modes (functions)

In this group of parameters users declare functions they want to use after pressing F key -
Procedure:
Following point 11 of this manual enter <P4.Func> submenu, next:
EXI-01-04-06-14-GB
while in weighing mode.
- 29 -
no - mode disabled YES - mode enabled
NOTICE:
Enabling procedure for other operation modes is analogous to the one described above.
Return to weighing:
See point - 11.2. – return to weighing.

14.2. Determining number of accessible modes

The user may decide whether upon pressing F key modes are to be accessible (option <ALL>), or only one, selected out of the list.
Procedure:
Following point 11 of this manual enter <P4.Func> submenu, next:
all the operation
EXI-01-04-06-14-GB
- 30 -
Return to weighing:
See point - 11.2. – return to weighing.

14.3. Counting parts of identical weight

The standard solution is equipped with option of counting small pieces of the same mass. If parts counting is to be performed in a container, its weight must be recorded into the scale memory (the container must be tarred).
NOTICE:
1. Counting pieces mode does not operate with other functions,
2. Parts counting function is not active after scale’s restart.
Procedure:
Enter <PcS> submenu (first make the operation mode accessible - see point 14.1):
EXI-01-04-06-14-GB
- 31 -
The blinking value on the display informs of quantity of a sample. By means of
key, set quantity of a sample and press key to
confirm:
If option <LASt> has been chosen than the most recent weight value of a single piece is displayed for 3 sec., the program proceeds to Parts Counting mode automatically setting the previously displayed value.
If option <FrEE> has been chosen, the program displays the following window:
EXI-01-04-06-14-GB
- 32 -
Pilsujący symbol
Blinking symbol
Using and keys, enter the demanded quantity of sample, where:
- selection of the digit to be set, - selection of the digit
value,
Press
key for confirmation,
• <LoAd> message is displayed, the following window is seen:
If pieces are to be weighed in a container, it is necessary to tare it,
next declared number of pieces must be put onto the weighing pan, their mass confirmed after stabilization of the indication (symbol on a display):
The program automatically calculates weight of a single element and proceeds to the Parts Counting mode displaying number of weighed elements (pcs):
NOTICE:
1. If the user presses the pan, the message -Lo- is indicated for a few seconds and the scale automatically returns to weighing.
2. In order to obtain reliable results, it is recommended to load the weighing pan with elements, mass of which is greater than 5 reading units.
EXI-01-04-06-14-GB
key when there are no elements on the
- 33 -
Select the digit to be set
Select the digit value
3. If a single element weight is not greater than a reading unit, then message <Err5> is shown on the scale display (see point 19. Error Messages), short sound signal is heard, next the scale returns to weighing.
Function deactivation:
Press F key
twice.

14.4. +/- control referring to the inscribed standard mass The program allows entering checkweighing threshold values (Min, Max.

Procedure:
Enter <HiLo> function (first make the operation mode accessible - see point 14.1):
The program proceeds to window for setting the lower threshold (Min):
EXI-01-04-06-14-GB
- 34 -
load mass below the I threshold
load mass between the
load mass above the II threshold
Select the digit to be set
Select the digit value
Press key to confirm, the program proceeds to window for setting the upper threshold (Max):
Press
key to confirm, the program automatically returns to
weighing with threshold values recorded into the memory.
During setting the threshold values the following cases take place:
thresholds
NOTICE:
If a user erroneously enters a value of the lower threshold higher than the upper one, the scale indicates an error message and returns to weighing.
Function deactivation:
Press F key
twice.
EXI-01-04-06-14-GB
- 35 -
Blinking symbol

14.5. Per cent deviation with reference to standard mass

This program allows to meas ure deviation in per cent with reference to a standard mass. The standard mass can be estimated by weighing (PrcA function) or inscribed by a user (PrcB function).

14.5.1. Mass standard weight determined by the measuring process Procedure:

Enter <PrcA> function (first make the operation mode accessible - see point 14.1):
Message <LoAd> is displayed, the following window is seen:
On a weighing pan place the load, mass of which is to be taken as a
The scale display shows indication 100,000%,
From this moment display will indicate weight deviation of load placed
EXI-01-04-06-14-GB
standard, upon stabilization of the indication (symbol press
key for confirmation,
displayed)
on a pan, wherein the said deviation is referred to the mass of standard (in %):
- 36 -
Blinking symbol
Function deactivation:
Press F key

14.5.2. Standard mass inscribed to scale memory Procedure:

Enter <Prcb> function (first make the operation mode accessible - see point 14.1):
The program proceeds to the following window:
twice.
Using and keys, set the mass standard weight,
Press
The scale display will show indication 0,000%,
From this moment display will indicate weight deviation of load placed
EXI-01-04-06-14-GB
where:
- selection of the digit to be set, - selection of the digit
value
key for confirmation,
on a pan, wherein the said deviation is referred to the mass of standard (in %).
- 37 -
Function deactivation:
Press F key
twice.

14.6. Automatic tare

This function is useful for fast net mass determination of weighed load when the tare value for each next load to be weighed is different. In case when the function is active the cycle of scales operation looks as follows:
press zeroing key when the pan is empty,
place the container for pieces,
when indication is stable, automatic tarring of the container mass
gets started (Net marker appears in the upper part of the display),
place a sample in the container,
display indicates net mass of a sample,
remove the sample together with the container,
scale displays zero indication (zeroing),
place a container for the next sample. When indication is stable
automatic tarring gets started (Net marker appears in the upper part of the display),
place next sample in the container.
Activation procedure:
Function deactivation:
Press F key
EXI-01-04-06-14-GB
twice.
- 38 -

14.7. Measurement of the max force – peak hold Procedure:

Enter <toP> function (first make the operation mode accessible - see point 14.1):
Max pictogram visible in upper part of the display is a confirmation of toP function activation:
Apply a force to the weighing pan. The display of scale will latch the maximum value of the force.
Remove loads from the pan.
Before next measurement press
.
Function deactivation:
Press F key
twice.

14.8. Totalizing

The program enables totalizing the mass of weighed loads and the totalized mass printout (for EX scales the printout is disabled).

14.8.1. Operation mode activation procedure

EXI-01-04-06-14-GB
- 39 -
Enter <Add> function (first make the operation mode accessible - see point 14.1)
P” letter visible on the left of the display is a confirmation of <Add> function activation:

14.8.2. Totalizing procedure

Following point 14.8.1 enter <Add> function,
Place the first load on a weighing pan. If the load is to be weighed in a
container, it is necessary to tare it before putting the first load into it, its mass must be confirmed by pressing
the indication (symbol
on a display),
Sum of weighings is displayed („▲” pictogram in the upper right-hand corner of the display).
key after stabilization of
Take off the load , the scale indication returns to ZERO, letter P appears on the left,
Place next load on the pan,
EXI-01-04-06-14-GB
- 40 -
Upon indication stabilization press key, sum of the first and second weighing is displayed („▲” pictogram in the upper right-hand corner of the display):
In order to finish the procedure of totalizing, press
(either with a
load on the pan or after taking it off):
If the
key is pressed again while load is being placed on a weighing pan, the sum of previously performed measurements is deleted, <unLoAd> message is displayed - unload the pan to return to ZERO . Letter „P” will appear. The scale is ready for the next totalizing procedure.
If the
key is pressed again while load is not being placed on a weighing pan, the sum of previously performed measurements is deleted, the scale is ready for the next totalizing procedure.

14.8.3. Storing the last sum of weighings

After interrupting the procedure of totalizing (e.g. turning off the scale) it is possible to restart with the last sum of weighings. In order to do it:
enter <Add> function (see point 14.8.1),
last sum of weighings is seen on the display, the displayed value
comes from the scale memory,
press
key in order to continue totalizing, the indication returns to ZERO and the letter „P” appears on the display; next load may be placed on the scale.
press
, or key in order to finish totalizing, the letter „P” appears on the display; the scale is ready for the next totalizing procedure.
EXI-01-04-06-14-GB
- 41 -
NOTICE:
In case of a display range overflow in totalizing, the program displays error <5-FULL>. The user must either take the load off the pan and press
key in order to complete totalizing or replace the load with ano ther one (of lower mass) not resulting in display range overflow.
Function deactivation:
Press F key
twice.

15. ADJUSTMENT

Only for verified scales
In order to ensure perfect weighing accuracy it is required to introduce into the scale memory a factor correcting the indications – it is so called adjustment. Adjustment should be performed when the user starts weighing or when dynamic change of temperature occurs. Before starting the procedure, loads should be removed from the pan.

15.1. Adjustment procedure Procedure:

Following point 11 of this manual enter <P6.CAL> submenu, next:
EXI-01-04-06-14-GB
- 42 -
The following messages are displayed:
A new start mass is adjusted during this period of time. After that a
mass of calibration weight is shown (e.g. 3 000kg).
Put the calibration weight on the weighing pan and press key, adjustment process starts, the following message is seen:
The <unLoAd> message informs about adjustment process completion:
EXI-01-04-06-14-GB
- 43 -
Take off the load, message <donE> is displayed for 1 sec., the scale shows the adjustment submenu name:
Adjustment process may be interrupted at any time by pressing key, the following message informs about abortion of a process:
NOTICE:
1. It must be remembered to perform calibration when the pan is unloaded!
2. If the adjustment takes more than 15 sec. the program displays an error message <Err8>, short sound signal is heard. Press
key and perform the adjustment again simultaneously maintaining as stable ambient conditions as possible!
Save changes and return to weighing:
See point - 11.2. – return to weighing.

15.2. Start mass estimation

If the scale does not require the f ull calibration pr ocess, or if the user lacks demanded number of mass standards, it is possi ble to adjust only a new start mass.
Procedure:
Following point 11 of this manual enter <P6.CAL> submenu, next:
EXI-01-04-06-14-GB
- 44 -
the program displays the following set of messages:
upon completion of the start mass estimation process the scale shows
the parameter name:
start mass estimation process may be interrupted at any time by key, the following message informs about abortion of a
pressing process:
NOTICE:
If the start mass estimation takes more than 15 sec. the program displays an error message <Err8>, short sound signal is heard. Press
key and perform the adjustment again simultaneously maintaining as stable ambient conditions as possible!
Save changes and return to weighing:
See point - 11.2. – return to weighing.
EXI-01-04-06-14-GB
- 45 -
Err2
-
Value beyond the zero range
Err3
-
Value beyond the tare range
Err4
-
Calibration mass or start mass beyond the acceptable Err5
-
Mass of a single piece lower than the scale division.
Err7
-
Power down time was to short (should be over 3s)
Err8
-
Operation time exceeded (tarring, zeroing, start mass null
-
Zero value from the AD converter
FULL2
-
Measurement range overflow
LH
-
Start mass error, the mass on the weighing platform start mass)
5–FULL
-
Display range overflow in totalizing

16. ERRORS

range (±1% for weight, ±10 for start mass).
estimation, adjustment)
is beyond the acceptable range (from -5% to +15% of
NOTICE:
1. The user is informed about occurrence of errors Err2, Err3, Err4, Err5, Err8, null by means of a short sound signal (1 second).
2. Error FULL2 is signalled by a sound signal lasting until the over excessive load is taken off the pan.

17. RESPECTIVE STANDARDS

PUE C/31H/EX indicator is manufactured in accordance with the following standards:
- EN 61326-1:2006 Electrical equipment for measurement, control and
laboratory use - EMC requirements -- Part 1: General requirements,
- EN 61010-1:2011 Safety requirements for electrical equipment for measurement, control, and laboratory use -- Part 1: General requirements,
- EN 60079-0:2013 + A11:2014-03 Explosive atmosph er es -- Part 0: Equipment – General requirements,
- EN 60079-11:2012 Explosive atmospheres -- Part 11: Equipment protection by intrinsic safety "i".
EXI-01-04-06-14-GB
- 46 -
EXI-01-04-06-14-GB
- 47 -
MANUFACTURER
OF ELECTRONIC EQUIPMENT
RADWAG WAGI ELEKTRONICZNE
POLAND, 26 – 600 Radom, Bracka 28
Phone: +48 48 38 48 800, phone/fax: + 48 48 385 00 10
Sales dpt.: + 48 48 366 80 06
www.radwag.com
EXI-01-04-06-14-GB
Loading...