1. I NTENDED USE
PUE 5.15, PUE 5.19 terminals are designed for scales that work with strain
gauge load cells. The terminals’ housing is made of stainless steel. PUE
5.15, PUE 5.19 are intended to be used in industry.
A big colourful screen of the terminal with a touch panel makes the software
operation much more comfortable since there is no need to use a keyboard.
PUE 5 terminal is a genuine device that consists of two units: the computer
and the weighing module, placed in one housing. Both of these units are
connected via an internal interface. The possibility of using common
operating systems allows external companies to create its software or to use
the existing one.
Such common devices as PC computers can be used with the terminal,
which is a great advantage while creating a network. Individual workstation
with the PC device is as well possible.
The terminal can work with 4 weighing platforms (option).
2. PRECAUTIONS MEASURES
A. Before the first use read the Service Manual carefully and follow the
instructions. Use the device as intended.
B. The weighed products should be placed in the central part of a scales
pan.
C. Do not use any means causing corrosion for cleaning.
D. The gross mass of weighed products must be lower than the scales
maximum load.
E. Great loads shouldn’t be left on the scales pan too long.
F. In case of a failure the device power supply should be instantly switched
off.
G. If the device is to be put out of operation it must be utilized in
accordance with the actual law regulations.
Page 6
- 6 -
3. WARRANTY CONDITIONS
A. RADWAG company will repair or replace the elements that turn out to
be faulty in terms of production or construction.
B. Agreeing on the ways of any defects elimination in case of their unclear
origin can be done only in presence of the producer and the user
representatives.
C. RADW AG is not responsible for any loss or damages in case of
unauthorized or incorrect execution of production or service processes.
D. Warranty does not cover:
• Mechanical defects caused by product exploitation other than
intended, defects of thermal and chemical origin, defects caused by
lightning, overvoltage in the power network or other random event,
• maintenance (the cleaning process).
E. Loss of warranty takes place if:
• A repair is carried out outside RADWAG sales office or authorized
service point,
• Service claims intrusion into mechanical or electronic construction
by unauthorized people,
• The device does not bear company’s protective stickers.
F. Detailed warranty conditions are listed on a service card.
G. Authorized Service phone number:
(0-48) 384 88 00 extension numbers: 106 and 107.
4. UNPACKING AND INSTALLATION
A. Take terminal out of its factory packaging.
B. Connect the terminal to a weighing platform and place it on a smooth,
even, thick surface away from any heat sources,
C. It is necessary to level scales. Turn right and left the levelling feet. The
correct levelling is obtained when an air bubble is placed in a centre of
the small ring.
Page 7
- 7 -
5. TERMINAL CONSTRUCTION
5.1. Overall dimensions
Fig.1 PUE 5.15 terminal dimensions
Page 8
- 8 -
Fig.2 PUE 5.19 terminal dimensions
Page 9
- 9 -
1
Cable gland of the strain gauge load cell (x4items)
2
USB M12 4 pin port
3
RS232 port
4
4WY port (option)
6
Panel USB port
7
Panel USB port
8
RJ45 ethernet port
9
USB M12 4 pin port
10
RS232, RS485 or 2xPROFIBUS (IN, OUT) port
11
4WE port (option)
12
Power supply cable gland
13
ON/OFF power switch-key
5.2. Description of ports
Fig.3 Back board view of PUE 5.15, PUE 5.19
5 Cable gland for the cable RS485 (option)
Page 10
- 10 -
Pin8 – NC
• Standard version ports:
- Power supply cable: through PG7 cable gland,
- Weighing platform cable: through PG11 cable gland,
- RS232, RS485: M12 8pin ports,
- Ethernet: RJ45 port,
- USBx2: M12 4pin port,
- USBx2: type A panel port
• Depending on the version of the terminal PUE 5.15, PUE 5.19 it is
possible to optionally equip it with the following ports:
Depending on how many extra modules have been installed the number of
ports and cable glands and their position may change.
6. WEIGHING MODULE MW-04
PUE 5 weighing terminal is equipped with weighing module MW-04. It can
serve as a fully functional scales without a screen. Its parameters are kept in
a permanent storage of the module. It communicates with an internal PC
module via RS232 (COM6) interface.
„MwManager” program serves to maintain weighing module MW-04. Its
description is to be found further down this User Manual.
7. I NTERFACES MODULE
Interfaces module allows connection of terminal ports and sockets that are
available on its back board. Weighing module MW-04 – RS232 (COM6) is
also connected to the Interfaces module.
Page 12
- 12 -
Fig.4 Interfaces module TOP view
7.1. Configuration of COM serial ports
Chart below presents assignation of COM ports detectable by Windows
operating system to proper ports of a terminal back board.
Caution:
• Weighing module MW-04 communicates with a PC module by COM6
• Chart presents factory assignation of COM ports, detectable by
Windows operating system
• Any changes of COM ports assignation in Windows operating system
may result in faulty functioning of applications installed on a terminal.
Page 13
- 13 -
MW-04 module
board)
Socket Pins view COM
RS232
RS485
RS232
(J6 on
334Rxxxx
RS232 – COM6
RS232 – COM3
RS485 – COM2
RS232 – COM5
7.2. RS485 – COM2
COM2 can be configured as RS485 or RS232. By defult it is configured as
RS485 (standard positioning of the ports).
In this case COM2 available on the J7 port of 344Rxxxx interfaces module is
not active.
8. STARTUP
• Press ON/OFF button to turn on the device. The button is placed at the
back of the terminal housing.
• Wait for the operating system loading procedure to start.
• Operating system Windows activates automatically after the loading
procedure has finished.
9. STRUCTURE OF „MwManager” PROGMAM
„MwManager” is a basic application installed on a terminal. It operates a
weighing module. This application serves as a display and a control panel. It
enables the scales maintenance and configuration.
The program enables reading of the mass, taring, zeroing, filtering,
input/output simulation and dosing function simulation for a given weighing
platform. Besides it enables setting input/output functions, which may be
assigned to a particular platform.
Page 14
- 14 -
Caution:
1. The manual complies with 1.0.3.1version of„MwManager” program
(and the following versions) and with weighing module MW-04 – version
1.1 and following versions.
2. Pressing the Save button saves all the changes in the weighing module.
All temporary parameters that haven’t been saved permanently into the
module are marked with red colour.
3. The view of some windows of the „MwManeger” program depends on
the number of operated A/D converter connected with the weighing
platforms and their configuration in the weighing module MW-04.
9.1. MwManager start-up
To start the program press
icon, or use
START/PROGRAMS/MwManager menu of Windows system.
Wait to see a main window of the program.
Fig.5 Main window of the program
Page 15
- 15 -
9.2. Parameters edition
Parameters edition depends on a parameter type, to do it:
• Press
button and choose a demanded menu out of a list.
• Click a window of a given parameter and use a
program keyboard to write a value.
• Press a button responsible for decrease of a parameter value or
its increase
, .
• Press button to select a given option, press button to
unselect it.
Page 16
- 16 -
Functions of the key buttons:
-
Confirmation of a change of a parameter
-
Canceling change of a parameter
-
Scrolling the list of available parameter values up
-
Scrolling the list of available parameter values down
Information button, displays information about settings of a
-
given parameter
Saving the settings:
Changes will be stored in a weighing module after Savebutton
has been pressed. All temporary parameters that haven’t been saved
permanently will be marked with a red colour.
Procedure:
• Press Save button ,
• When you see a following message press <YES>
Page 17
- 17 -
• You will see the confirmation message:
• Press <OK>,
• Changes have been saved in a permanent storage of a weighing
module.
If any setting changes will be done but not saved the user may press
Refresh button
to view the present settings.
9.3. Weighment display
Symbols:
Fig.6 Weighment display view
- exact zero
- stable measurement result
Page 18
- 18 -
- weighing unit
- weighing platform number
Functions of the key buttons:
-
Zeroing
-
Taring
Selection of a weighing platform (when a module works
-
with more than one platform). Green colour indicates the
presently operated platform.
Caution:
Zeroing and taring function is available for a presently operated platform.
9.4. Application settings
Application settings bookmark contains: settings of connection with a
weighing module, list of available program interface languages and other
options
.
9.4.1. Connection settings
Connection settings
bookmark
, starts settings of connection with a weighing
module.
button, located in application settings
Page 19
- 19 -
Fig.7 Connection settings window
Press MW-04 button
to connect to module MW-04. This button
is located in a “device selection” bookmark.
Description:
Device selection
The device to which the user wants to connect
weighing module MW-01
weighing module MW-04
To be selected in case of cooperation with weighing module MW-04
Page 20
- 20 -
Transmission speed of RS485 communication port
By default 57600 bps
Connection types
RS 232 Connection via RS232 port
TCP/IP Connection via Ethernet network
RS 485 Connection for RS 485 network
Offline
RS232:
Port Choice of COM port number to which the module is physically connected
Speed
Parity Parity state. By default „none” (not editable value)
Data bits Number of data bits. By default 8 data bits (not editable value)
Stop bits Number of stop bits. By default 1 stop bit (not editable value)
TCP/IP:
IP address
Port
RS485:
Port Choice of COM port number to which a module is physically connected
Selection of interface connection with a weighing module
Offline mode is used for saving and edition of all indispensable
parameters. Saving and edition is carried out in a configuration file.
Transmission speed of RS232 communication port
By default 57600 bps
Device IP address, by default 192.168.0.2
Port set in the weighing module, by default 4001
Speed
Parity Parity state. By default „none” (not editable value)
Data bits Number of data bits. By default 8 data bits (not editable value)
Stop bits Number of stop bits. By default 1 stop bit (not editable value)
Address Weighing module IP address
Caution:
1. Weighing module MW-04 used in PUE 5 terminal communicates with an
internal PC module of the terminal via RS232 interface of COM6 port,
with the default transmission speed of 57600 bit per second.
2. Parameters <Connection settings> are not active in case of connection
with the weighing module.
Page 21
- 21 -
Description of buttons:
Connecting with a module. After the connection has been
established the button function switches from “Connect” to “Disconnect”
and its colour changes from red to green.
Disconecting from a module. In case of accidental
disconnection the button’s function switches to „Connect” and the colour
changes to red again.
Closing „MwManager” application.
9.4.2. Language
Language button
bookmark
, opens the window of language settings.
Fig.8 Language settings window
, which is located in application settings
After selection of a language version press Apply button in
order to save changes. Present version of the program offers following
languages:
• English
• Polish
Page 22
- 22 -
9.4.3. Other
Other button , which is located in application settings
bookmark
, lets to operate other available options.
Fig.9 Other options window
„Carry out a test connection during the application starting procedure”
– this option enables the program to automatically connect with the weighing
module. It connects either by its default type of connection or the recently
chosen one.
„Turn on a touch interface” – this option adapts a visual image of
”MwManager” program to make it able to work with PUE5 weighing
terminal, it turns on a touch panel and it turns off a mouse.
Page 23
- 23 -
After setting the options press Save button
in order to save
introduced changes.
9.5. Parameters
Parameters bookmark contains user parameters, weighing
module communication parameters, INPUT/OUTPUT functions and a
preview of available scales platforms and A/D converters.
9.5.1. User parameters
User parameters button
bookmark
opens a user parameters window. These
, which is located in Parameters
parameters are an active platform parameters, they can be edited by each of
the users.
Fig.10 User parameters window
List of user parameters:
Autozeroing - Turning on/off autozeroing function
Page 24
- 24 -
Beep - Sound signal (unavailable for MW-04 module)
Median filter value settings
None – median filter turned off
The moving average filter speed settings
None – filter turned off
Median filter -
Filter -
Current unit - Change of a current unit
Caution:
When few platforms are being operated by MW-04 module at one time, only
parameters for a selected platform are displayed and can be edited .
9.5.2. Communication settings
Communication button
bookmark
opens a weighing module parameters window.
, which is located in Parameters
These parameters can be viewed and edited by each of the users who
connect to the weighing module.
•Ethernet
Page 25
- 25 -
Fig.11 Ethernet communication parameters window
Number of seconds of the device inactivity that results in
from 0 to 300 [s].
Description of fields:
IP address
Subnet mask
Default gateway
Port
- Device IP address, by default 192.168.0.2
- Ethernet subnet mask, by default 255.255.255.0
- Ethernet default gateway, by default 192.168.0.1
- TCP communication port, by default 4001.
Timeout -
• RS 232/485
Fig.12 RS communication parameters window
Description of fields:
Weighing module address on RS485 network (a separate IP
-
Module address
Speed RS232 -
address is given to each of the network devices), default value
is 1. Values range from 1 to 254.
RS232 communication interface speed transmission. By default
57600 bps
disconnection of the device, number of seconds ranges
Page 26
- 26 -
Speed RS485 -
RS485 communication interface speed transmission. By default
57600 bps
To save changes of communication parameters it is necessary to execute
restaring procedure of a weighing module.
One must remember that new parameters have to be written in the
Connection settings window
. See point 9.4.1 of this manual.
Caution:
Internal weighing module MW- 04 used in the PUE5 terminal is physically
connected to the PC module of the terminal via RS232 interface of COM6
port and it works with a default transmission speed of 57600 bps.
9.5.3. INPUT/OUTPUT functions.
Weighing module MW-04 can have up to 4 inputs and 4 outputs.
INPUT/OUTPUT button
bookmark
opens a window allowing the user for configuration
, which is located in Parameters
of weighing module Inputs and Outputs. Select number of scales/platform for
which the function is to be realized.
Page 27
- 27 -
Fig. 13 Inputs/Outputs configuration window
Page 28
- 28 -
None
Zeroing
Dosing Start
Stable weighing result over the LO mass, on a selected
platform
Stable weighing result over LO mass and below MIN
Unstable weighing result over LO mass and below MIN
threshold, for a selected platform
Stable OK
stabilny
Stable weighing result between MIN and MAX
thresholds, for a selected platform
thresholds, for a the selected platform
Stable weighing result over MAX threshold, for a
selected platform
Unstable weighing result over MAX threshold, for a
selected platform
•Inputs configuration
Inputs functions:
Inactive Input
Taring
Dosing Stop
Selected platform taring
Selected platform zeroing
Start of a dosing process for a selected platform
Stop of a dosing process for a selected platform
• Oututs configuration
Outputs functions:
None
Stable
Inactive Output
Stable MIN
MIN
OK
Stable MAX
MAX
threshold, for a selected platform
Unstable weighing result between MIN and MAX
Caution:
If the user sets a function for a given Output and if at the same time on this
Output a function of fast dosing or fine dosing is set then upon start of
a dosing and during its duration the outputs will be activated according to
dosing parameters settings. After the process has finis hed the set functions
will swap to Output.
9.5.4. Available platforms preview
Page 29
- 29 -
The button, which is located in Parameters bookmark
opens scales windows preview of all the scales being operated
by MW-04 module. For each platform the following information is displayed:
A/D converter(s) divisions, calibration coefficient and an input mass.
Caution:
Window preview depends on the number of A/D converters, the number of
platforms connected to the device and their configuration.
Fig.14 Example of 4 weighing platforms preview window
9.5.5. Available A/D converters preview
ADC button
, which is located in Parameters bookmark
opens a preview of reading units, calibration coefficient, mass,
correction coefficient and input mass of the available A/D converters.
Page 30
- 30 -
Caution:
Window preview depends on the number of A/D converters, the number of
platforms connected to the device and their configuration.
Fig.15 A/D converter divisions preview window
9.6. Functions
Dosing, checkweighing, state and Input/Output simulations are set in
functions bookmark
.
9.6.1. Dosing
, which is located in Functions bookmark Dosing button
opens a windows with settings for a dosing
process of a weighing platform (presently selected one).
Page 31
- 31 -
Fig.16 Dosing parameters window
•Bargraph
Dosing window contains a graphic bar that visualizes mass readout
within a weighing range of a weighing module.
Bargraph can be graduated upto 120% of an extreme dosing threshold if
a fine dosing threshold option has been selected. If a fine dosing
threshold option is turned off then the bargraph is graduated in
accordance with a fast dosing threshold.
Fig.17 Graduation of bargraph for a fast dosing threshold
Page 32
- 32 -
Fig.18 Graduation of bargraph for a quick and fine dosing threshold
Fig.19 Bargraph for a small amount of mass with a graduation option turned off
Fig.20 Bargraph for the same amount of mass with a graduation option turned on
• Dosing parameters
Fig.21 Dosing parameters settings window
Depending on the need dosing process may be carried out in two stages or
one stage.
Page 33
- 33 -
Mass value which finishes the 1st stage
case of one-stage dosing process.)
Choice of an output or few outputs
weighing platform).
Choice of an output or few outputs
weighing platform).
Przycisk funkcji, przypisanej do wejścia 1
Przycisk funkcji, przypisanej do wejścia 2
Przycisk funkcji, przypisanej do wejścia 3
Przycisk funkcji, przypisanej do wejścia 4
Fields description:
Fast dosing threshold Output number
of a dosing process.
(switching on to the second stage, not in
active during the 1st dosing process
stage (for a presently operated
Fine dosing threshold Output number
Mass value which finishes the 2nd stage
of a dosing process.
(the end of the dosing process.)
•Dosing parameters
active during the 1st dosing process
stage (for a presently operated
Dosing status window informs about a
present state of a dosing process on a
given weighing platform.
Description:
Dosing process state:
DOZING – dosing in progress
Dosing status
SUSPENDED – suspending the dosing process.
STOPPED – stopping the dosing process,
FINISHED – dosing is finished.
•Inputs simulation
Inputs simulation allows for simulation of an operation of a function
assigned to a particular Input. See point 9.5.3 of this manual.
Page 34
- 34 -
LO
threshold
Gross value of mass, checkweighing function is active if this
value is exceeded
Mass value for estimation of tolerance threshold.
value
•Dosing simulation
There are Stop dosing and Start dosing buttons in the bottom part of a
dosing simulation window. No matter what functions have been assigned
on Inputs, these buttons enable dosing process start and stop.
9.6.2. Checkweighing
Turning on
option and pressing button opens
checkweighing settings window for a presently selected weighing platform.
Fig.22 Checkweighing parameters window
Description of the field s:
Min
threshold
Max
threshold
- MIN threshold value is signaled below the Min threshold
value
- OK threshold is signaled between Min threshold - Max threshold values.
- MAX threshold value is signaled over the Max threshold
Functions signalling within thresholds:
Page 35
- 35 -
MINOK
MAX
Caution:
Checkweighing signalling is available after the Output functions have been
set. See point 9.5.3 of this manual.
9.6.3. Input/Output state
Window of Inputs signaling which also allows configuration of Outputs state
opens upon activation of functions
Input/Output state
button.
option and pressing
Fig.23 Inputs/Outputs states window
Input/output numbers in program accord with the numeration in a module.
Active input/output
Inactive input/output
Simulation of output operation is possible after this output number has been
pressed. This very output activates immediately if only no other function has
been assigned to it.
Output activity simulation is available in a dosing window.
Page 36
- 36 -
10. WEIGHMENT
Put a product to be weighted on a scales pan. Read a measurement when a
marker
Caution:
In case of cooperation of weighing module MW-04 with more than one
weighing platform the user must select a correct platform in a weighment
display in order to get the correct mass reading.
10.1. Terms of use
In order to assure long device usage and accurate mass measurements the
user should:
•avoid applying mechanical shocks while loading a scales pan:
has been displayed.
• put loads precisely in the centre of a scales pan (erros of eccentric
weighment are defined by PN-EN 45501 standard, point 3.5 and 3.6.2):
• not load a scales pan with concentrated force:
Page 37
- 37 -
• avoid side loads, especially side shocks:
10.2. Scales zeroing
In order to zero mass readout of a presently selected platform the user has
to press
button which is to be found in a weighment display of a
„MwManager” program (the top right hand corner) or he has to initiate
zeroing function defined for a given input (See point 9.5.3 of this manual).
On a display mass readout with zero value will be shown next to symbols:
and .
Zeroing means estimating a new zero point which will be recognized by scale
as an exact zero. Zeroing is possible only when display states are stable.
Caution:
Zeroing a display state is possible only within the range upto ±2% of the
maximum scales load. If zeroing value is greater than ±2% of the maximum
load then the message Err2 is displayed.
10.3. Taring
To estimate net mass for a presently selected platform the user needs to put
packaging on a device and wait for a readout to stabilize, next (s)he may
press button
. The other possibility is to initiate tare function defined
for a given input (See point 9.5.3 of this manual).
On a display mass readout with zero value is be shown next to symbols : Net
. The scales has been tared.
i
Page 38
- 38 -
While using tare function the user has to pay attention to maximal weighing
range. It cannot be exceeded. The readout which is a sum of tarred mass will
be displayed after the load and the package have been taken off. Minus sign
will be seen in front of a displayed value.
Caution:
Taring process cannot be executed when a display shows minus mass value
or a zero mass value. In such a case the message Err3 is displayed.
10.4. Weighment for dual range scales
1st range weighing proceeds to 2nd range weighing automatically without
any interference of the user (immediately after the 1
maximum has been exceeded). The 2
display of a
sign in the top left hand corner of the display. After a load
nd
range weighing is signaled with a
has been taken off the indication is zero. The weighment accuracy is the 2
st
range weighing
nd
range accuracy up to the moment of getting back to zero.
Fig.24 2nd range scales window
Return from the 2nd range weighing to the 1st range weighing proceeds
automatically after the load has been taken off and after the scales has
entered AUTOZERO zone – instead of the 2
displayed. The scale returns to the 1
st
nd
range weighing symbol is
range weighing accuracy mode.
10.5. Measuring unit change
To change a measuring unit for a presently selected platform the user has to
press a Change
button, which is to be found in in users
parameters window. To open user parameters window the user needs to
open „MwManager” program window first.
Page 39
- 39 -
Fig. 25 User parameters window
Possible choices:
• When a main unit of scales is [kg], the user can select following units:
[kg, lb, oz, ct, N, g], for verified scales [lb, oz, N] are unavailable;
• When a main unit of scales is [g], the user can select following units: [g,
kg, lb, oz, ct, N], for verified scales [lb, oz, N] are unavailable;
11. SCALES PARAMETERS
The user can adapt scale to an outdoor environmental conditions (filters
order) or his own needs (active autozero)
These parameters are to be found in parameters bookmark
. They are available for each of the platforms, and for each of
the platforms they can be edited.
List of parameters:
• Autozeroing,
• Median filter,
• Filter.
>
Page 40
- 40 -
11.1. Autozero func tion
For ensuring accurate balance readaouts „AUTOZERO” function has been
introduced. This function automatically controls and corrects zero readouts.
When this function is active several measuring results are compared at
regular intervals. If value differentiating these results is lower than declared
AUTOZERO range, e.g. 1 reading unit, than zeroing is automatic. Markers of
stable result -
and zero readout - are displayed.
When AUTOZERO function is active each measurements start with an exact
zero which is useful. Although there are cases when this function can be a
drawback. One of such examples might be putting load on a pan very slowly
(e.g. pouring the load). Zero readout correction system in such a case might
distort readout of a real mass of the load.
Procedure:
• Enter user parameters window
,
• Select or unselect <Autozeroing> parameter
- autozeroing function on
- autozeroing function off
11.2. Median filter
Median filter eliminates short-term impulse noise (e.g. mechanical shock).
Page 41
- 41 -
Procedure:
• Enter user parameters window
• Select <Median filter> parameter, press
• Select a desired setting out of the available list
Available parameters:
None - medain filter turned on
0.5, 1, 1.5, 2, 2.5 - medain filter turned off
11.3. Filter
Moving average filter adjusts scales to outdoor environmental conditions.
Procedure:
• Enter user parameters window
• Select <Filter> parameter, press
• Select a desired setting out of the available list
Available parameters:
none, very fast, fast, medium, slow
Caution:
The higher the order of a filter is the longer time of measurement stabilization
it needs.
Page 42
- 42 -
Lo
> 0 <
MIN
MAX
OK
HI
Próg Min
Próg Max
12. CHECKWEIGHING
Checkweighing is a function that enables precise weighment of a sample for
which minimum and maximum weighing threshold, so called checkweighing
thresholds, have been determined (LO – to little mass of the sample, HI – to
big mass of the sample, OK – correct mass of the sample).
These states are presented through light signaling or through steering of
peripheral devices systems.
Fig.26 States range for checkweighing function
Caution:
Starting procedure of checkweighing function and its signaling have been
presented in point 9.6.2 of this manual.
12.1. LO threshold
<LO threshold> parameter detemines net mass upon exceeding which the
checkweighing function activates.
Procedure:
• Click parameter window:
• Enter LO threshold value.
Page 43
- 43 -
• Press button to close dialog window
• Press Save button to save introduced changes into a
permanent storage of a module.
12.2. MIN/MAX threshold
<MIN threshold> parameter determines net mass threshold of a
checkweighing function. For the determined net mass the state swaps
between Min and OK.
<MAX threshol d > parameter determines net mass threshold of
checkweighing function. For the determined net mass the state swaps
between OK and MAX.
Outputs signaling starts over the set net LO threshold value.
Dosing function enables precise weighment of a given
product up to desired value.
<Fast dosing threshold> parameter stands for net mass value of fast
(rough) dosing below which one or more outputs are active. (Outputs are
assigned to fast dosing).
<Fine dosing threshold> parameter stands for net mass value of slow
dosing below which (but at the same time not below mass value of fast
dosing threshold) one or more outputs are active. (Outputs are assigned to
fine dosing).
• Changes will be confirmed with the following message:
• If any threshold values will be changed but not saved than present
settings can be read out, to do it the user has to use a Readout
button.
• Readout will be confirmed with a following message:
Page 46
- 46 -
Caution:
Description of a dosing function and its parameters are to be found in the
point 9.6.1 of this manual.
14. PARAMETERS SAVED TO FILE
„MwManager” program allows saving set parameters into *.sav file format.
This function can be useful when saving module settings as a safety copy in
case of a weighing module breakdown. Parameters saved on a safety copy
can be used for further configuration purposes with a greater number of
weighing modules.
Fig.27 An example of a window with available options Save to file and Read from
file.
File format:
MW04_(factory number)_RRRR-MM-DD_HH-MM.sav
Page 47
- 47 -
14.1. Save to file
Procedure:
• Press Save to file button after the weighing module
parameters have been set
• In an operating system window select file destination folder and press
Save
button.
Fig.28 „Save as” system window
• The following message will inform about correctly saved parameters.
Page 48
- 48 -
Caution:
Preview of an operating system window depends on the installed operating
system therefore it may differ from the one presented in Fig.28.
14.2. Read from file
Procedure:
• In order to read the parameters press Read from file button
• Open operating system window, see Fig 29, and select a previously
saved file, next press Open
button.
Fig. 29 „Open” window
• In Parameters groups selection window select Read all parameters
option or Read chosen parameters option and press <OK>.
Page 49
- 49 -
Fig.30 Parameters groups selection window.
• If parameters have been read correctly the following message will be
displayed:
15. OFFLINE MODE
Offline mode allows for operation of a selected options of a program without
the need of connecting it to a weighing module. Through this indispensable
parameters can be saved without a physical connection with the device.
Page 50
- 50 -
Fig.31 Offline mode window preview.
Procedure:
• Press Connection settings
in Applicatio n settings
button, which is to be found
bookmark
• In Connection types settings window select Offline option
• Press Connect button
• „Offline” message will be displayed on the scales screen.
Page 51
- 51 -
• Set chosen parameters and save the configuration to file, see
description in point 14.1.
Page 52
- 52 -
16. ERROR MESSAGES
Err2
- Value beyond zeroing range
Err3
Err8
NULL
FULL
HI
LH
- Value beyond taring range
- Tarring/zeroing time limit exceeded
- Zero value from the load cell
- Measuring range exceeded
- Display state exceeded
Input mass error, readout beyond range
(from -5% to +15% of an input mass)
17. CABLES DIAGRAMS
Scales of STANDARD made design can cooperate with:
• a computer,
• KAFKA, EPSON receipt printers,
• CITIZEN, ZEBRA label printers,
17.1. USB cable for a printer
Fig.32 USB cable for a printer - PT0087
Page 53
- 53 -
17.2. Cable RS232 for EPSON, CITIZEN printer
Fig.33 Cable RS232 for Citizen, Epson printers - PT0019
17.3. Cable RS232 terminal – computer
Fig.34 Cable RS232 terminal – computer - PT0020
17.4. Cable RS232, RS485
Fig.36 Cable RS232, RS485 – wires colors
Page 54
- 54 -
Caution:
These are colours for cables of “M12” standard. The figure shows just an
example cable type.
17.5. Ethernet cable
Fig.35 P0212 Ethernet cable
18. EXTRA MODULES SPECIFICATION
Scales with PUE5 terminal can optionally be expanded with an extra modules
positively influencing the device functionality:
• Extra A/D converter module
• 4 Inputs / 4 outputs module
• Profibus DP interface.
18.1. A/D converter
A/D converters influence PUE 5 terminal functionality. Additional weighing
modules can work with PUE 5 thanks to them. A/D converter board is fixed
onto weighing module MW-04, inside PUE 5 gauge. It is possible to fix up to
4 A/D converter boards. On a housing lid an additional cable gland is
installed through which load cell cable is led out. Parameters of all load cells
are identical.
Page 55
- 55 -
Fig.38 A/D converter board
Fig.39 Weighing module MW-04
Page 56
- 56 -
Maximal number of divisions from A/D
converter
18.1.1. Technical specification of A/D converter
8 388 608
OIML class III
Number of verification intervals 6000e
Max increase of signal 19.5mV
Max voltage per 1 verification interval 3.25µV
Min voltage per 1 verification interval 0.4µV
Min impedance of strain gauge load
cells
Max impedance of strain gauge load
cells
Strain gauge load cells feeding voltage 5V
Strain gauge load cells type 4 or 6 cables + screen
80 Ω
1200 Ω
18.1.2. Strain gauge load cells connec tion
•6-wire load cells
For 6-wire strain gauge load cells do the connection as follows:
Fig.40 6-wire strain gauge load cell connection
Page 57
- 57 -
See strain gauge load cell
connection rules
REF+
SENSE +
REF-
SENSE -
IN+
OUTPUT+
IN-
OUTPUT-
+5V
INPUT+
AGND
INPUT-
See the strain gauge load cell
connection rules
REF+ - Connect to +5V
REF-
-
Connect to AGND
IN+
OUTPUT+
IN-
OUTPUT-
+5V
INPUT+
AGND
INPUT-
Interface on load
cell board
Strain gauge load
cell signal
Notes
E SCREEN
•4-wire load cells
For 4-wire load cells do the connection as follows:
Connect REF+ with +5V and REF- with AGND using the LiY 0,34mm2
cable.
Interface on load
cell board
E SCREEN
Fig.41 4-wire load cell connection
Strain gauge load
cell signal
Notes
Page 58
- 58 -
Permanent galvanic
producer
•Strain gauge load cell screen connection rules.
While connecting signal cable screen of a strain gauge load cell the
following connection rules must be obeyed in order to ensure correct
weighment.
In both cases (scales platforms with 6-wire and 4-wire signal cable) the
same rule of connection is obligatory:
No galvanic connection of
the screen and the
construction of the scales
arm
Scales with metal head
connected with platform via
signal cable
Compact mechanical
construction of scales (e.g.
Scales with head on a mast),
head in a metal housing.
connection of the screen
and the construction of the
scales arm, done by the
POINT B POINT B
POINT B E
Point B – electrical connection with housing of a terminal (e.g. threaded
pin, threaded hole to be used for screwing of a soldering eye),
E – soldering point of A/D converter board
• Test of galvanic connection of a strain gauge load cell screen with
a platform construction
To carry out a test ohmmetter is needed.
Page 59
- 59 -
Inputs number
4
Inputs type
OptoMOS
Max switching current
0,2A DC
Max forward voltage
50V DC
Outputs number
4
Outputs type
Optoinsulated
Steering voltage range
5 -24V DC
18.2. 4 inputs / 4 outputs module
Fig.42 4Inputs/4Outputs module – description of cable outlets
4IN/4OUT module is installed inside a terminal on weighing module MW-04
board (Fig. 39).
Configuration for an optional made design of a terminal.
Signals are fed on hermetic M12 8P sockets, one socket for inputs, the other
one for outputs.
Chart below presents signals distribution for each pin of a socket.
Configuration for an optional made design of a terminal.
Signals are fed on two cables, one cable for inputs, the other one for outputs.
Chart below presents signals distribution for each wire of a cable.
Page 61
- 61 -
1
IN1
1
OUT1
2
IN2
2
OUT2
3
IN3
3
OUT3
4
IN4
4
OUT4
5
COM
5
COM
6
+24V
6
+24V
7
GND
7
GND
Cable for INPUTS Cable for OUTPUTS
WIRE NUMBER SIGNAL WIRE NUMBER SIGNAL
18.3. Profibus
Configuration for an optional made design of a terminal.
PROFIBUS module is installed on an interface board, inside PUE 5 gauge.
Signals are fed on hermetic M12 5P PROFIBUS IN, PROFIBUS OUT
sockets (type B ports coding - PROFIBUS)
Fig.44 Interfaces board with an installed PROFIBUS module.
Page 62
- 62 -
Chart below presents signals distribution for each pin of a socket.
Pin1 – NC
PROFIBUS IN
(male)
PROFIBUS OUT
(female)
Pin2 – A
Pin3 – NC
Pin4 – B
Pin5 – NC
Pin1 - +5V
Pin2 – A
Pin3 – GND
Pin4 – B
Pin5 – NC
Caution:
For a correct work reconfiguration of jumpers on a PC module is required –
JP1 jumpers
Fig.45 Jumpers configuration on a PC module for PROFIBUS
Page 63
- 63 -
Min impedance of a strain gauge load
cell
Max impedance of a strain gauge load
cell
modules marked with the CE sign.
Optional IN/WY
4 INputs, 4 OUTputs
19. TECHNICAL PARAMETERS
Housing Stainless steel
IP rating (degree of protection) IP65
PUE 5.15 PUE 5.19
Display
Power 100÷240VAC 50-60Hz
Temperature range
Max number of load cell reading units 8 388 608
OIML class III
Number of verification intervals 6000e
Max increase of signal 19.5mV
Max voltage per 1 verification interval 3.25µV
Min voltage per 1 verification interval 0.4µV
Strain gauge load cells feeding voltage 5V
Strain gauge load cells type 4 or 6 cables + screen
Processor Intel® Atom™ D525 Dual-Core 1.8 GHz
RAM 2 GB DDR3 800 MHz
Data storage HDD SATA lub SSD SATA
Graphics storage max. 224 MB
Ethernet 10/100/1000 Mbit
Standard interfaces 2xRS232, RS485, 4xUSB 2.0, Ethernet
Optional interfaces
TFT 15’’÷15,6” IR
with touch panel
Work temperature: 0°C to +40°C
Storage temperature: -20°C to +60°C
80 Ω
1200 Ω
Pofibus, or other communication
TFT 19’’IR
with touch panel
Page 64
- 64 -
Extra A/D converter
-
Metrological parameters like in
Inputs/Outputs module
-
Additional 4 inputs / 4 outputs
Profibus DP interface
-
Slave mode operation
20. EXTRA EQUIPMENT
the main platform
Page 65
-65-
MANUFACTURER
OF ELECTRONIC WEIGHING EGUIPMENT
RADWAG WAGI ELEKTRONICZNE
26 – 600 Radom, ul. Bracka 28
Phone: +48 48 38 48 800, fax. + 48 48 385 00 10
Sales department: + 48 48 366 80 06
ee--mmaaiill:: eexxppoorrtt@@rraaddwwaagg..ccoomm
wwwwww..rraaddwwaagg..ccoomm