RADWAG PUE 5.15, PUE 5.19 Service Manual

Page 1
Service manual
Weighing terminal PUE 5.15 PUE 5.19
For cooperation with strain gauge load cells
Manual number: ITKU-84-02-10-13-PL
MANUFACTURER OF ELECTRONIC WEIGHING EQUIPMENT
RADWAG Wagi Elektroniczne, 26 – 600 Radom, Bracka 28, POLAND Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10 Sales department: +48 (0-48) 366 80 06 export@radwag.com www.radwag.com
Page 2
- 2 -
OCTOBER 2013
Page 3
- 3 -
Table OF CONTENTS
1. INTENDED USE ................................................................................................................. 5
2. PRECAUTIONS MEASURES ............................................................................................. 5
3. WARRANTY CONDITIONS ................................................................................................ 6
4. UNPACKING AND INSTALLATION ................................................................................... 6
5. TERMINAL CONSTRUCTION ............................................................................................ 7
5.1. Overall dimensions ...................................................................................................... 7
5.2. Description of ports ..................................................................................................... 9
6. WEIGHING MODULE MW-04 ...........................................................................................11
7. INTERFACES MODULE ....................................................................................................11
7.1. Configuration of COM serial ports ...............................................................................12
7.2. RS485 – COM2 ..........................................................................................................13
8. STARTUP ..........................................................................................................................13
9. STRUCTURE OF „MwManager” PROGMAM ..................................................................13
9.1. MwManager start-up ...................................................................................................14
9.2. Parameters edition .....................................................................................................15
9.3. Weighment display .....................................................................................................17
9.4. Application settings ....................................................................................................18
9.4.1. Connection settings ...........................................................................................18
9.4.2. Language ...........................................................................................................21
9.4.3. Other .................................................................................................................22
9.5. Parameters .................................................................................................................23
9.5.1. User parameters ................................................................................................23
9.5.2. Communication settings ....................................................................................24
9.5.3. INPUT/OUTPUT functions. ................................................................................26
9.5.4. Available platforms preview ...............................................................................28
9.5.5. Available A/D converters preview .......................................................................29
9.6. Functions ....................................................................................................................30
9.6.1. Dosing ...............................................................................................................30
9.6.2. Checkweighing ..................................................................................................34
9.6.3. Input/Output state ..............................................................................................35
10. WEIGHMENT ..................................................................................................................36
10.1. Terms of use ............................................................................................................36
10.2. Scales zeroing .........................................................................................................37
10.3. Taring ......................................................................................................................37
10.4. Weighment for dual range scales .............................................................................38
10.5. Measuring unit change .............................................................................................38
11. SCALES PARAMETERS .................................................................................................39
11.1. Autozero function .....................................................................................................40
11.2. Median filter .............................................................................................................40
11.3. Filter .........................................................................................................................41
12. CHECKWEIGHING ..........................................................................................................42
12.1. LO threshold ............................................................................................................42
12.2. MIN/MAX threshold ..................................................................................................43
13. DOSING ...........................................................................................................................44
14. PARAMETERS SAVED TO FILE .....................................................................................46
14.1. Save to file ...............................................................................................................47
14.2. Read from file ..........................................................................................................48
15. OFFLINE M ODE ..............................................................................................................49
Page 4
- 4 -
16. ERROR MESSAGES .......................................................................................................52
17. CABLES DIAGRAMS ......................................................................................................52
17.1. USB cable for a printer .............................................................................................52
17.2. Cable RS232 for EPSON, CITIZEN printer ..............................................................53
17.3. Cable RS232 terminal – computer ...........................................................................53
17.4. Cable RS232, RS485 ...............................................................................................53
17.5. Ethernet cable ..........................................................................................................54
18. EXTRA MODULES SPECIFICATION ..............................................................................54
18.1. A/D converter ...........................................................................................................54
18.1.1. Technical specification of A/D converter ..........................................................56
18.1.2. Strain gauge load cells connection ..................................................................56
18.2. 4 inputs / 4 outputs module ......................................................................................59
18.2.1. Technical parameters of 4IN/4OUT module ....................................................59
18.2.2. Input/output schematic diagram ......................................................................60
18.2.3. 4IN/4OUT in sockets .......................................................................................60
18.2.4. 4IN/4OUT through cable gland ........................................................................60
18.3. Profibus ...................................................................................................................61
19. TECHNI CAL PARAMETERS ...........................................................................................63
20. EXTRA EQUIPMENT .......................................................................................................64
Page 5
- 5 -

1. I NTENDED USE

PUE 5.15, PUE 5.19 terminals are designed for scales that work with strain gauge load cells. The terminals’ housing is made of stainless steel. PUE
5.15, PUE 5.19 are intended to be used in industry. A big colourful screen of the terminal with a touch panel makes the software operation much more comfortable since there is no need to use a keyboard.
PUE 5 terminal is a genuine device that consists of two units: the computer and the weighing module, placed in one housing. Both of these units are connected via an internal interface. The possibility of using common operating systems allows external companies to create its software or to use the existing one. Such common devices as PC computers can be used with the terminal, which is a great advantage while creating a network. Individual workstation with the PC device is as well possible. The terminal can work with 4 weighing platforms (option).

2. PRECAUTIONS MEASURES

A. Before the first use read the Service Manual carefully and follow the
instructions. Use the device as intended.
B. The weighed products should be placed in the central part of a scales
pan.
C. Do not use any means causing corrosion for cleaning.
D. The gross mass of weighed products must be lower than the scales
maximum load.
E. Great loads shouldn’t be left on the scales pan too long.
F. In case of a failure the device power supply should be instantly switched
off.
G. If the device is to be put out of operation it must be utilized in
accordance with the actual law regulations.
Page 6
- 6 -

3. WARRANTY CONDITIONS

A. RADWAG company will repair or replace the elements that turn out to
be faulty in terms of production or construction.
B. Agreeing on the ways of any defects elimination in case of their unclear
origin can be done only in presence of the producer and the user representatives.
C. RADW AG is not responsible for any loss or damages in case of
unauthorized or incorrect execution of production or service processes.
D. Warranty does not cover:
Mechanical defects caused by product exploitation other than intended, defects of thermal and chemical origin, defects caused by lightning, overvoltage in the power network or other random event,
maintenance (the cleaning process).
E. Loss of warranty takes place if:
A repair is carried out outside RADWAG sales office or authorized service point,
Service claims intrusion into mechanical or electronic construction by unauthorized people,
The device does not bear company’s protective stickers.
F. Detailed warranty conditions are listed on a service card.
G. Authorized Service phone number:
(0-48) 384 88 00 extension numbers: 106 and 107.

4. UNPACKING AND INSTALLATION

A. Take terminal out of its factory packaging. B. Connect the terminal to a weighing platform and place it on a smooth,
even, thick surface away from any heat sources,
C. It is necessary to level scales. Turn right and left the levelling feet. The
correct levelling is obtained when an air bubble is placed in a centre of the small ring.
Page 7
- 7 -

5. TERMINAL CONSTRUCTION

5.1. Overall dimensions

Fig.1 PUE 5.15 terminal dimensions
Page 8
- 8 -
Fig.2 PUE 5.19 terminal dimensions
Page 9
- 9 -
1
Cable gland of the strain gauge load cell (x4items)
2
USB M12 4 pin port
3
RS232 port
4
4WY port (option)
6
Panel USB port
7
Panel USB port
8
RJ45 ethernet port
9
USB M12 4 pin port
10
RS232, RS485 or 2xPROFIBUS (IN, OUT) port
11
4WE port (option)
12
Power supply cable gland
13
ON/OFF power switch-key

5.2. Description of ports

Fig.3 Back board view of PUE 5.15, PUE 5.19
5 Cable gland for the cable RS485 (option)
Page 10
- 10 -
Pin8 – NC
Standard version ports:
- Power supply cable: through PG7 cable gland,
- Weighing platform cable: through PG11 cable gland,
- RS232, RS485: M12 8pin ports,
- Ethernet: RJ45 port,
- USBx2: M12 4pin port,
- USBx2: type A panel port
Depending on the version of the terminal PUE 5.15, PUE 5.19 it is possible to optionally equip it with the following ports:
- Extra weighing platforms
- Ethernet cable through the PG9 cable gland
- Cable RS485 through the PG9 cable gland
- 4WE/4WY cable through the PG9 cable gland
- 4 IN/4OUT in M12 5 pin connector
- Profibus communication module
Ports typology
Pin1 – B (RS485) Pin2 – RxD Pin3 – TxD
RS232, RS485
Pin4 – A (RS485) Pin5 – GND Pin6 - +5VDC Pin7 – NC
PROFIBUS IN
(male)
PROFIBUS OUT
(female)
RS232
Pin1 – NC Pin2 – A Pin3 – NC Pin4 – B Pin5 – NC
Pin1 - +5V Pin2 – A Pin3 – GND Pin4 – B Pin5 – NC
Pin1 – NC Pin2 – RxD Pin3 – TxD Pin4 – NC Pin5 – GND Pin6 - +5VDC
Page 11
- 11 -
Pin1 – Vcc
Pin4 – GND
Pin1-WE1
Pin7-GND
Pin1-WY1
Pin7-GND
USB
Ethernet RJ45 Standard RJ45
USB panel USB A Standard USB A
4WE
4WY
Pin2 – D­Pin3 – D+
Pin2-WE2 Pin3-WE3 Pin4-WE4 Pin5-COMM Pin6-+24VDC
Pin2-WY2 Pin3-WY3 Pin4-WY4 Pin5-COMM Pin6-+24VDC
Caution:
Depending on how many extra modules have been installed the number of ports and cable glands and their position may change.

6. WEIGHING MODULE MW-04

PUE 5 weighing terminal is equipped with weighing module MW-04. It can serve as a fully functional scales without a screen. Its parameters are kept in a permanent storage of the module. It communicates with an internal PC module via RS232 (COM6) interface.
„MwManager” program serves to maintain weighing module MW-04. Its
description is to be found further down this User Manual.

7. I NTERFACES MODULE

Interfaces module allows connection of terminal ports and sockets that are available on its back board. Weighing module MW-04 – RS232 (COM6) is also connected to the Interfaces module.
Page 12
- 12 -
Fig.4 Interfaces module TOP view

7.1. Configuration of COM serial ports

Chart below presents assignation of COM ports detectable by Windows operating system to proper ports of a terminal back board.
Caution:
Weighing module MW-04 communicates with a PC module by COM6
Chart presents factory assignation of COM ports, detectable by
Windows operating system
Any changes of COM ports assignation in Windows operating system may result in faulty functioning of applications installed on a terminal.
Page 13
- 13 -
MW-04 module
board)
Socket Pins view COM
RS232 RS485
RS232
(J6 on
334Rxxxx
RS232 – COM6
RS232 – COM3 RS485 – COM2
RS232 – COM5

7.2. RS485 – COM2

COM2 can be configured as RS485 or RS232. By defult it is configured as RS485 (standard positioning of the ports). In this case COM2 available on the J7 port of 344Rxxxx interfaces module is not active.

8. STARTUP

Press ON/OFF button to turn on the device. The button is placed at the
back of the terminal housing.
Wait for the operating system loading procedure to start.
Operating system Windows activates automatically after the loading
procedure has finished.

9. STRUCTURE OF „MwManager” PROGMAM

„MwManager” is a basic application installed on a terminal. It operates a
weighing module. This application serves as a display and a control panel. It enables the scales maintenance and configuration.
The program enables reading of the mass, taring, zeroing, filtering, input/output simulation and dosing function simulation for a given weighing platform. Besides it enables setting input/output functions, which may be assigned to a particular platform.
Page 14
- 14 -
Caution:
1. The manual complies with 1.0.3.1 version of „MwManager” program (and the following versions) and with weighing module MW-04 – version
1.1 and following versions.
2. Pressing the Save button saves all the changes in the weighing module. All temporary parameters that haven’t been saved permanently into the module are marked with red colour.
3. The view of some windows of the „MwManeger” program depends on the number of operated A/D converter connected with the weighing platforms and their configuration in the weighing module MW-04.

9.1. MwManager start-up

To start the program press
icon, or use
START/PROGRAMS/MwManager menu of Windows system.
Wait to see a main window of the program.
Fig.5 Main window of the program
Page 15
- 15 -

9.2. Parameters edition

Parameters edition depends on a parameter type, to do it:
Press
button and choose a demanded menu out of a list.
Click a window of a given parameter and use a
program keyboard to write a value.
Press a button responsible for decrease of a parameter value or
its increase
, .
Press button to select a given option, press button to
unselect it.
Page 16
- 16 -
Functions of the key buttons:
-
Confirmation of a change of a parameter
-
Canceling change of a parameter
-
Scrolling the list of available parameter values up
-
Scrolling the list of available parameter values down
Information button, displays information about settings of a
-
given parameter
Saving the settings:
Changes will be stored in a weighing module after Save button
has been pressed. All temporary parameters that haven’t been saved permanently will be marked with a red colour.
Procedure:
Press Save button ,
When you see a following message press <YES>
Page 17
- 17 -
You will see the confirmation message:
Press <OK>,
Changes have been saved in a permanent storage of a weighing
module.
If any setting changes will be done but not saved the user may press
Refresh button
to view the present settings.

9.3. Weighment display

Symbols:
Fig.6 Weighment display view
- exact zero
- stable measurement result
Page 18
- 18 -
- weighing unit
- weighing platform number
Functions of the key buttons:
-
Zeroing
-
Taring
Selection of a weighing platform (when a module works
-
with more than one platform). Green colour indicates the presently operated platform.
Caution:
Zeroing and taring function is available for a presently operated platform.

9.4. Application settings Application settings bookmark contains: settings of connection with a

weighing module, list of available program interface languages and other
options
.

9.4.1. Connection settings

Connection settings
bookmark
, starts settings of connection with a weighing
module.
button, located in application settings
Page 19
- 19 -
Fig.7 Connection settings window
Press MW-04 button
to connect to module MW-04. This button
is located in a “device selection” bookmark.
Description:
Device selection
The device to which the user wants to connect
weighing module MW-01
weighing module MW-04
To be selected in case of cooperation with weighing module MW-04
Page 20
- 20 -
Transmission speed of RS485 communication port By default 57600 bps
Connection types
RS 232 Connection via RS232 port
TCP/IP Connection via Ethernet network
RS 485 Connection for RS 485 network
Offline
RS232:
Port Choice of COM port number to which the module is physically connected
Speed
Parity Parity state. By default „none” (not editable value)
Data bits Number of data bits. By default 8 data bits (not editable value)
Stop bits Number of stop bits. By default 1 stop bit (not editable value)
TCP/IP:
IP address
Port
RS485:
Port Choice of COM port number to which a module is physically connected
Selection of interface connection with a weighing module
Offline mode is used for saving and edition of all indispensable parameters. Saving and edition is carried out in a configuration file.
Transmission speed of RS232 communication port By default 57600 bps
Device IP address, by default 192.168.0.2 Port set in the weighing module, by default 4001
Speed
Parity Parity state. By default „none” (not editable value)
Data bits Number of data bits. By default 8 data bits (not editable value)
Stop bits Number of stop bits. By default 1 stop bit (not editable value)
Address Weighing module IP address
Caution:
1. Weighing module MW-04 used in PUE 5 terminal communicates with an internal PC module of the terminal via RS232 interface of COM6 port, with the default transmission speed of 57600 bit per second.
2. Parameters <Connection settings> are not active in case of connection with the weighing module.
Page 21
- 21 -
Description of buttons:
Connecting with a module. After the connection has been
established the button function switches from “Connect” to Disconnect” and its colour changes from red to green.
Disconecting from a module. In case of accidental
disconnection the button’s function switches to „Connect” and the colour
changes to red again.
Closing „MwManager” application.

9.4.2. Language

Language button
bookmark
, opens the window of language settings.
Fig.8 Language settings window
, which is located in application settings
After selection of a language version press Apply button in
order to save changes. Present version of the program offers following languages:
English
Polish
Page 22
- 22 -

9.4.3. Other

Other button , which is located in application settings
bookmark
, lets to operate other available options.
Fig.9 Other options window
„Carry out a test connection during the application starting procedure”
– this option enables the program to automatically connect with the weighing module. It connects either by its default type of connection or the recently chosen one.
„Turn on a touch interface” – this option adapts a visual image of ”MwManager” program to make it able to work with PUE5 weighing
terminal, it turns on a touch panel and it turns off a mouse.
Page 23
- 23 -
After setting the options press Save button
in order to save
introduced changes.

9.5. Parameters

Parameters bookmark contains user parameters, weighing
module communication parameters, INPUT/OUTPUT functions and a preview of available scales platforms and A/D converters.

9.5.1. User parameters

User parameters button
bookmark
opens a user parameters window. These
, which is located in Parameters
parameters are an active platform parameters, they can be edited by each of the users.
Fig.10 User parameters window
List of user parameters:
Autozeroing - Turning on/off autozeroing function
Page 24
- 24 -
Beep - Sound signal (unavailable for MW-04 module)
Median filter value settings None – median filter turned off
The moving average filter speed settings None – filter turned off
Median filter -
Filter -
Current unit - Change of a current unit
Caution:
When few platforms are being operated by MW-04 module at one time, only parameters for a selected platform are displayed and can be edited .

9.5.2. Communication settings

Communication button
bookmark
opens a weighing module parameters window.
, which is located in Parameters
These parameters can be viewed and edited by each of the users who connect to the weighing module.
Ethernet
Page 25
- 25 -
Fig.11 Ethernet communication parameters window
Number of seconds of the device inactivity that results in
from 0 to 300 [s].
Description of fields:
IP address
Subnet mask
Default gateway
Port
- Device IP address, by default 192.168.0.2
- Ethernet subnet mask, by default 255.255.255.0
- Ethernet default gateway, by default 192.168.0.1
- TCP communication port, by default 4001.
Timeout -
RS 232/485
Fig.12 RS communication parameters window
Description of fields:
Weighing module address on RS485 network (a separate IP
-
Module address
Speed RS232 -
address is given to each of the network devices), default value is 1. Values range from 1 to 254.
RS232 communication interface speed transmission. By default 57600 bps
disconnection of the device, number of seconds ranges
Page 26
- 26 -
Speed RS485 -
RS485 communication interface speed transmission. By default 57600 bps
To save changes of communication parameters it is necessary to execute restaring procedure of a weighing module. One must remember that new parameters have to be written in the
Connection settings window
. See point 9.4.1 of this manual.
Caution:
Internal weighing module MW- 04 used in the PUE5 terminal is physically connected to the PC module of the terminal via RS232 interface of COM6 port and it works with a default transmission speed of 57600 bps.

9.5.3. INPUT/OUTPUT functions.

Weighing module MW-04 can have up to 4 inputs and 4 outputs.
INPUT/OUTPUT button
bookmark
opens a window allowing the user for configuration
, which is located in Parameters
of weighing module Inputs and Outputs. Select number of scales/platform for which the function is to be realized.
Page 27
- 27 -
Fig. 13 Inputs/Outputs configuration window
Page 28
- 28 -
None
Zeroing
Dosing Start
Stable weighing result over the LO mass, on a selected platform
Stable weighing result over LO mass and below MIN
Unstable weighing result over LO mass and below MIN threshold, for a selected platform
Stable OK stabilny
Stable weighing result between MIN and MAX thresholds, for a selected platform
thresholds, for a the selected platform
Stable weighing result over MAX threshold, for a selected platform
Unstable weighing result over MAX threshold, for a selected platform
Inputs configuration
Inputs functions:
Inactive Input
Taring
Dosing Stop
Selected platform taring
Selected platform zeroing
Start of a dosing process for a selected platform
Stop of a dosing process for a selected platform
Oututs configuration
Outputs functions:
None Stable
Inactive Output
Stable MIN
MIN
OK
Stable MAX
MAX
threshold, for a selected platform
Unstable weighing result between MIN and MAX
Caution:
If the user sets a function for a given Output and if at the same time on this Output a function of fast dosing or fine dosing is set then upon start of a dosing and during its duration the outputs will be activated according to dosing parameters settings. After the process has finis hed the set functions will swap to Output.

9.5.4. Available platforms preview

Page 29
- 29 -
The button , which is located in Parameters bookmark
opens scales windows preview of all the scales being operated by MW-04 module. For each platform the following information is displayed: A/D converter(s) divisions, calibration coefficient and an input mass.
Caution:
Window preview depends on the number of A/D converters, the number of platforms connected to the device and their configuration.
Fig.14 Example of 4 weighing platforms preview window

9.5.5. Available A/D converters preview

ADC button
, which is located in Parameters bookmark
opens a preview of reading units, calibration coefficient, mass, correction coefficient and input mass of the available A/D converters.
Page 30
- 30 -
Caution:
Window preview depends on the number of A/D converters, the number of platforms connected to the device and their configuration.
Fig.15 A/D converter divisions preview window

9.6. Functions

Dosing, checkweighing, state and Input/Output simulations are set in
functions bookmark
.

9.6.1. Dosing

, which is located in Functions bookmark Dosing button
opens a windows with settings for a dosing
process of a weighing platform (presently selected one).
Page 31
- 31 -
Fig.16 Dosing parameters window
Bargraph
Dosing window contains a graphic bar that visualizes mass readout within a weighing range of a weighing module.
Bargraph can be graduated upto 120% of an extreme dosing threshold if a fine dosing threshold option has been selected. If a fine dosing threshold option is turned off then the bargraph is graduated in accordance with a fast dosing threshold.
Fig.17 Graduation of bargraph for a fast dosing threshold
Page 32
- 32 -
Fig.18 Graduation of bargraph for a quick and fine dosing threshold
Fig.19 Bargraph for a small amount of mass with a graduation option turned off
Fig.20 Bargraph for the same amount of mass with a graduation option turned on
• Dosing parameters
Fig.21 Dosing parameters settings window
Depending on the need dosing process may be carried out in two stages or one stage.
Page 33
- 33 -
Mass value which finishes the 1st stage
case of one-stage dosing process.)
Choice of an output or few outputs
weighing platform).
Choice of an output or few outputs
weighing platform).
Przycisk funkcji, przypisanej do wejścia 1
Przycisk funkcji, przypisanej do wejścia 2
Przycisk funkcji, przypisanej do wejścia 3
Przycisk funkcji, przypisanej do wejścia 4
Fields description:
Fast dosing threshold Output number
of a dosing process. (switching on to the second stage, not in
active during the 1st dosing process stage (for a presently operated
Fine dosing threshold Output number
Mass value which finishes the 2nd stage of a dosing process.
(the end of the dosing process.)
Dosing parameters
active during the 1st dosing process stage (for a presently operated
Dosing status window informs about a present state of a dosing process on a given weighing platform.
Description:
Dosing process state: DOZING – dosing in progress
Dosing status
SUSPENDED – suspending the dosing process. STOPPED – stopping the dosing process, FINISHED – dosing is finished.
Inputs simulation
Inputs simulation allows for simulation of an operation of a function
assigned to a particular Input. See point 9.5.3 of this manual.
Page 34
- 34 -
LO threshold
Gross value of mass, checkweighing function is active if this value is exceeded
Mass value for estimation of tolerance threshold.
value
Dosing simulation
There are Stop dosing and Start dosing buttons in the bottom part of a dosing simulation window. No matter what functions have been assigned on Inputs, these buttons enable dosing process start and stop.

9.6.2. Checkweighing

Turning on
option and pressing button opens
checkweighing settings window for a presently selected weighing platform.
Fig.22 Checkweighing parameters window
Description of the field s:
Min threshold
Max threshold
- MIN threshold value is signaled below the Min threshold
value
- OK threshold is signaled between Min threshold - Max threshold values.
- MAX threshold value is signaled over the Max threshold
Functions signalling within thresholds:
Page 35
- 35 -
MIN OK
MAX
Caution:
Checkweighing signalling is available after the Output functions have been set. See point 9.5.3 of this manual.

9.6.3. Input/Output state

Window of Inputs signaling which also allows configuration of Outputs state
opens upon activation of functions
Input/Output state
button.
option and pressing
Fig.23 Inputs/Outputs states window
Input/output numbers in program accord with the numeration in a module.
Active input/output
Inactive input/output
Simulation of output operation is possible after this output number has been pressed. This very output activates immediately if only no other function has been assigned to it.
Output activity simulation is available in a dosing window.
Page 36
- 36 -

10. WEIGHMENT

Put a product to be weighted on a scales pan. Read a measurement when a marker
Caution:
In case of cooperation of weighing module MW-04 with more than one weighing platform the user must select a correct platform in a weighment display in order to get the correct mass reading.

10.1. Terms of use

In order to assure long device usage and accurate mass measurements the user should:
avoid applying mechanical shocks while loading a scales pan:
has been displayed.
put loads precisely in the centre of a scales pan (erros of eccentric weighment are defined by PN-EN 45501 standard, point 3.5 and 3.6.2):
not load a scales pan with concentrated force:
Page 37
- 37 -
avoid side loads, especially side shocks:

10.2. Scales zeroing

In order to zero mass readout of a presently selected platform the user has
to press
button which is to be found in a weighment display of a
„MwManager” program (the top right hand corner) or he has to initiate zeroing function defined for a given input (See point 9.5.3 of this manual).
On a display mass readout with zero value will be shown next to symbols:
and . Zeroing means estimating a new zero point which will be recognized by scale as an exact zero. Zeroing is possible only when display states are stable.
Caution:
Zeroing a display state is possible only within the range upto ±2% of the maximum scales load. If zeroing value is greater than ±2% of the maximum load then the message Err2 is displayed.

10.3. Taring

To estimate net mass for a presently selected platform the user needs to put packaging on a device and wait for a readout to stabilize, next (s)he may
press button
. The other possibility is to initiate tare function defined
for a given input (See point 9.5.3 of this manual). On a display mass readout with zero value is be shown next to symbols : Net
. The scales has been tared.
i
Page 38
- 38 -
While using tare function the user has to pay attention to maximal weighing range. It cannot be exceeded. The readout which is a sum of tarred mass will be displayed after the load and the package have been taken off. Minus sign will be seen in front of a displayed value.
Caution:
Taring process cannot be executed when a display shows minus mass value or a zero mass value. In such a case the message Err3 is displayed.

10.4. Weighment for dual range scales 1st range weighing proceeds to 2nd range weighing automatically without

any interference of the user (immediately after the 1 maximum has been exceeded). The 2
display of a
sign in the top left hand corner of the display. After a load
nd
range weighing is signaled with a
has been taken off the indication is zero. The weighment accuracy is the 2
st
range weighing
nd
range accuracy up to the moment of getting back to zero.
Fig.24 2nd range scales window
Return from the 2nd range weighing to the 1st range weighing proceeds
automatically after the load has been taken off and after the scales has entered AUTOZERO zone – instead of the 2
displayed. The scale returns to the 1
st
nd
range weighing symbol is
range weighing accuracy mode.

10.5. Measuring unit change

To change a measuring unit for a presently selected platform the user has to
press a Change
button, which is to be found in in users
parameters window. To open user parameters window the user needs to
open „MwManager” program window first.
Page 39
- 39 -
Fig. 25 User parameters window
Possible choices:
When a main unit of scales is [kg], the user can select following units:
[kg, lb, oz, ct, N, g], for verified scales [lb, oz, N] are unavailable;
When a main unit of scales is [g], the user can select following units: [g,
kg, lb, oz, ct, N], for verified scales [lb, oz, N] are unavailable;

11. SCALES PARAMETERS

The user can adapt scale to an outdoor environmental conditions (filters order) or his own needs (active autozero)
These parameters are to be found in parameters bookmark
. They are available for each of the platforms, and for each of
the platforms they can be edited.
List of parameters:
Autozeroing,
Median filter,
Filter.
>
Page 40
- 40 -

11.1. Autozero func tion

For ensuring accurate balance readaouts „AUTOZERO” function has been introduced. This function automatically controls and corrects zero readouts.
When this function is active several measuring results are compared at regular intervals. If value differentiating these results is lower than declared AUTOZERO range, e.g. 1 reading unit, than zeroing is automatic. Markers of stable result -
and zero readout - are displayed.
When AUTOZERO function is active each measurements start with an exact zero which is useful. Although there are cases when this function can be a drawback. One of such examples might be putting load on a pan very slowly (e.g. pouring the load). Zero readout correction system in such a case might distort readout of a real mass of the load.
Procedure:
Enter user parameters window
,
Select or unselect <Autozeroing> parameter
- autozeroing function on
- autozeroing function off

11.2. Median filter

Median filter eliminates short-term impulse noise (e.g. mechanical shock).
Page 41
- 41 -
Procedure:
Enter user parameters window
Select <Median filter> parameter, press
Select a desired setting out of the available list
Available parameters:
None - medain filter turned on
0.5, 1, 1.5, 2, 2.5 - medain filter turned off

11.3. Filter

Moving average filter adjusts scales to outdoor environmental conditions.
Procedure:
Enter user parameters window
Select <Filter> parameter, press
Select a desired setting out of the available list
Available parameters:
none, very fast, fast, medium, slow
Caution:
The higher the order of a filter is the longer time of measurement stabilization it needs.
Page 42
- 42 -
Lo
> 0 <
MIN
MAX
OK
HI
Próg Min
Próg Max

12. CHECKWEIGHING

Checkweighing is a function that enables precise weighment of a sample for which minimum and maximum weighing threshold, so called checkweighing thresholds, have been determined (LO – to little mass of the sample, HI – to big mass of the sample, OK – correct mass of the sample).
These states are presented through light signaling or through steering of peripheral devices systems.
Fig.26 States range for checkweighing function
Caution:
Starting procedure of checkweighing function and its signaling have been presented in point 9.6.2 of this manual.

12.1. LO threshold <LO threshold> parameter detemines net mass upon exceeding which the

checkweighing function activates.
Procedure:
Click parameter window:
Enter LO threshold value.
Page 43
- 43 -
Press button to close dialog window
Press Save button to save introduced changes into a
permanent storage of a module.

12.2. MIN/MAX threshold <MIN threshold> parameter determines net mass threshold of a

checkweighing function. For the determined net mass the state swaps between Min and OK.
<MAX threshol d > parameter determines net mass threshold of
checkweighing function. For the determined net mass the state swaps between OK and MAX.
Outputs signaling starts over the set net LO threshold value.
Procedure:
Click <Min threshold> parameter window or <Max threshold> parameter window,
Enter a threshold value
Page 44
- 44 -
Press button to close a dialog window,
Press Save
button to save introduced changes into a
permanent storage of a module.

13. DOSING

Dosing function enables precise weighment of a given product up to desired value.
<Fast dosing threshold> parameter stands for net mass value of fast
(rough) dosing below which one or more outputs are active. (Outputs are assigned to fast dosing).
<Fine dosing threshold> parameter stands for net mass value of slow
dosing below which (but at the same time not below mass value of fast dosing threshold) one or more outputs are active. (Outputs are assigned to fine dosing).
Procedure:
Click <Fast dosing threshold> parameter window or <Fine dosing threshold> parameter window,
Page 45
- 45 -
Enter a threshold value
Press button to close the dialog window,
Press Save
button to save introduced changes into a
permanent storage of a module.
Changes will be confirmed with the following message:
If any threshold values will be changed but not saved than present
settings can be read out, to do it the user has to use a Readout
button.
Readout will be confirmed with a following message:
Page 46
- 46 -
Caution:
Description of a dosing function and its parameters are to be found in the point 9.6.1 of this manual.

14. PARAMETERS SAVED TO FILE

„MwManager” program allows saving set parameters into *.sav file format.
This function can be useful when saving module settings as a safety copy in case of a weighing module breakdown. Parameters saved on a safety copy can be used for further configuration purposes with a greater number of weighing modules.
Fig.27 An example of a window with available options Save to file and Read from
file.
File format:
MW04_(factory number)_RRRR-MM-DD_HH-MM.sav
Page 47
- 47 -

14.1. Save to file Procedure:

Press Save to file button after the weighing module
parameters have been set
In an operating system window select file destination folder and press
Save
button.
Fig.28 „Save as” system window
The following message will inform about correctly saved parameters.
Page 48
- 48 -
Caution:
Preview of an operating system window depends on the installed operating system therefore it may differ from the one presented in Fig.28.

14.2. Read from file Procedure:

In order to read the parameters press Read from file button
Open operating system window, see Fig 29, and select a previously
saved file, next press Open
button.
Fig. 29 „Open” window
In Parameters groups selection window select Read all parameters
option or Read chosen parameters option and press <OK>.
Page 49
- 49 -
Fig.30 Parameters groups selection window.
If parameters have been read correctly the following message will be
displayed:

15. OFFLINE MODE

Offline mode allows for operation of a selected options of a program without
the need of connecting it to a weighing module. Through this indispensable parameters can be saved without a physical connection with the device.
Page 50
- 50 -
Fig.31 Offline mode window preview.
Procedure:
Press Connection settings
in Applicatio n settings
button, which is to be found
bookmark
In Connection types settings window select Offline option
Press Connect button
„Offline” message will be displayed on the scales screen.
Page 51
- 51 -
Set chosen parameters and save the configuration to file, see
description in point 14.1.
Page 52
- 52 -

16. ERROR MESSAGES

Err2
- Value beyond zeroing range
Err3 Err8 NULL FULL HI
LH
- Value beyond taring range
- Tarring/zeroing time limit exceeded
- Zero value from the load cell
- Measuring range exceeded
- Display state exceeded
Input mass error, readout beyond range
­(from -5% to +15% of an input mass)

17. CABLES DIAGRAMS

Scales of STANDARD made design can cooperate with:
a computer,
KAFKA, EPSON receipt printers,
CITIZEN, ZEBRA label printers,

17.1. USB cable for a printer

Fig.32 USB cable for a printer - PT0087
Page 53
- 53 -

17.2. Cable RS232 for EPSON, CITIZEN printer

Fig.33 Cable RS232 for Citizen, Epson printers - PT0019

17.3. Cable RS232 terminal – computer

Fig.34 Cable RS232 terminal – computer - PT0020

17.4. Cable RS232, RS485

Fig.36 Cable RS232, RS485 – wires colors
Page 54
- 54 -
Caution:
These are colours for cables of “M12” standard. The figure shows just an example cable type.

17.5. Ethernet cable

Fig.35 P0212 Ethernet cable

18. EXTRA MODULES SPECIFICATION

Scales with PUE5 terminal can optionally be expanded with an extra modules positively influencing the device functionality:
Extra A/D converter module
4 Inputs / 4 outputs module
Profibus DP interface.

18.1. A/D converter

A/D converters influence PUE 5 terminal functionality. Additional weighing modules can work with PUE 5 thanks to them. A/D converter board is fixed onto weighing module MW-04, inside PUE 5 gauge. It is possible to fix up to 4 A/D converter boards. On a housing lid an additional cable gland is installed through which load cell cable is led out. Parameters of all load cells are identical.
Page 55
- 55 -
Fig.38 A/D converter board
Fig.39 Weighing module MW-04
Page 56
- 56 -
Maximal number of divisions from A/D converter

18.1.1. Technical specification of A/D converter

8 388 608
OIML class III
Number of verification intervals 6000e
Max increase of signal 19.5mV
Max voltage per 1 verification interval 3.25µV
Min voltage per 1 verification interval 0.4µV
Min impedance of strain gauge load cells
Max impedance of strain gauge load cells
Strain gauge load cells feeding voltage 5V
Strain gauge load cells type 4 or 6 cables + screen
80 Ω
1200 Ω

18.1.2. Strain gauge load cells connec tion

6-wire load cells
For 6-wire strain gauge load cells do the connection as follows:
Fig.40 6-wire strain gauge load cell connection
Page 57
- 57 -
See strain gauge load cell
connection rules
REF+
SENSE +
REF-
SENSE -
IN+
OUTPUT+
IN-
OUTPUT-
+5V
INPUT+
AGND
INPUT-
See the strain gauge load cell
connection rules
REF+ - Connect to +5V
REF-
-
Connect to AGND
IN+
OUTPUT+
IN-
OUTPUT-
+5V
INPUT+
AGND
INPUT-
Interface on load
cell board
Strain gauge load
cell signal
Notes
E SCREEN
4-wire load cells
For 4-wire load cells do the connection as follows:
Connect REF+ with +5V and REF- with AGND using the LiY 0,34mm2 cable.
Interface on load
cell board
E SCREEN
Fig.41 4-wire load cell connection
Strain gauge load
cell signal
Notes
Page 58
- 58 -
Permanent galvanic
producer
Strain gauge load cell screen connection rules.
While connecting signal cable screen of a strain gauge load cell the following connection rules must be obeyed in order to ensure correct weighment. In both cases (scales platforms with 6-wire and 4-wire signal cable) the same rule of connection is obligatory:
No galvanic connection of the screen and the construction of the scales arm
Scales with metal head connected with platform via signal cable
Compact mechanical construction of scales (e.g. Scales with head on a mast), head in a metal housing.
connection of the screen and the construction of the scales arm, done by the
POINT B POINT B
POINT B E
Point B – electrical connection with housing of a terminal (e.g. threaded
pin, threaded hole to be used for screwing of a soldering eye),
E – soldering point of A/D converter board
Test of galvanic connection of a strain gauge load cell screen with a platform construction
To carry out a test ohmmetter is needed.
Page 59
- 59 -
Inputs number
4
Inputs type
OptoMOS
Max switching current
0,2A DC
Max forward voltage
50V DC
Outputs number
4
Outputs type
Optoinsulated
Steering voltage range
5 -24V DC

18.2. 4 inputs / 4 outputs module

Fig.42 4Inputs/4Outputs module – description of cable outlets
4IN/4OUT module is installed inside a terminal on weighing module MW-04 board (Fig. 39).
Two optional versions are available:
4IN/4OUT fed on hermetic socket
4IN/4OUT fed through cable glands.

18.2.1. Technical parameters of 4IN/4OUT module

Input parameters
Outputs parameters
Page 60
- 60 -
Inputs diagram
Outputs diagram
Pin1-IN1
Pin7-GND
Pin1-OUT1
Pin7-GND

18.2.2. Input/output schematic diagram

Fig.43 Input/output schematic (demonstrating) diagram

18.2.3. 4IN/4OUT in sockets

Configuration for an optional made design of a terminal. Signals are fed on hermetic M12 8P sockets, one socket for inputs, the other one for outputs. Chart below presents signals distribution for each pin of a socket.
Pin2-IN2 Pin3-IN3
4IN
4OUT
Pin4-IN4 Pin5-COMM Pin6-+24VDC
Pin2-OUT2 Pin3-OUT3 Pin4-OUT4 Pin5-COMM Pin6-+24VDC

18.2.4. 4IN/4OUT through cable gland

Configuration for an optional made design of a terminal. Signals are fed on two cables, one cable for inputs, the other one for outputs. Chart below presents signals distribution for each wire of a cable.
Page 61
- 61 -
1
IN1
1
OUT1
2
IN2
2
OUT2
3
IN3
3
OUT3
4
IN4
4
OUT4
5
COM
5
COM
6
+24V
6
+24V
7
GND
7
GND
Cable for INPUTS Cable for OUTPUTS
WIRE NUMBER SIGNAL WIRE NUMBER SIGNAL

18.3. Profibus

Configuration for an optional made design of a terminal. PROFIBUS module is installed on an interface board, inside PUE 5 gauge. Signals are fed on hermetic M12 5P PROFIBUS IN, PROFIBUS OUT sockets (type B ports coding - PROFIBUS)
Fig.44 Interfaces board with an installed PROFIBUS module.
Page 62
- 62 -
Chart below presents signals distribution for each pin of a socket.
Pin1 – NC
PROFIBUS IN
(male)
PROFIBUS OUT
(female)
Pin2 – A Pin3 – NC Pin4 – B Pin5 – NC
Pin1 - +5V Pin2 – A Pin3 – GND Pin4 – B Pin5 – NC
Caution:
For a correct work reconfiguration of jumpers on a PC module is required – JP1 jumpers
Fig.45 Jumpers configuration on a PC module for PROFIBUS
Page 63
- 63 -
Min impedance of a strain gauge load cell
Max impedance of a strain gauge load cell
modules marked with the CE sign.
Optional IN/WY
4 INputs, 4 OUTputs

19. TECHNICAL PARAMETERS

Housing Stainless steel
IP rating (degree of protection) IP65
PUE 5.15 PUE 5.19
Display
Power 100÷240VAC 50-60Hz
Temperature range
Max number of load cell reading units 8 388 608
OIML class III
Number of verification intervals 6000e
Max increase of signal 19.5mV
Max voltage per 1 verification interval 3.25µV
Min voltage per 1 verification interval 0.4µV
Strain gauge load cells feeding voltage 5V
Strain gauge load cells type 4 or 6 cables + screen
Processor Intel® Atom™ D525 Dual-Core 1.8 GHz
RAM 2 GB DDR3 800 MHz
Data storage HDD SATA lub SSD SATA
Graphics storage max. 224 MB
Ethernet 10/100/1000 Mbit
Standard interfaces 2xRS232, RS485, 4xUSB 2.0, Ethernet
Optional interfaces
TFT 15’’÷15,6” IR
with touch panel
Work temperature: 0°C to +40°C
Storage temperature: -20°C to +60°C
80 Ω
1200 Ω
Pofibus, or other communication
TFT 19’’IR
with touch panel
Page 64
- 64 -
Extra A/D converter
-
Metrological parameters like in Inputs/Outputs module
-
Additional 4 inputs / 4 outputs
Profibus DP interface
-
Slave mode operation

20. EXTRA EQUIPMENT

the main platform
Page 65
- 65 -
MANUFACTURER
OF ELECTRONIC WEIGHING EGUIPMENT
RADWAG WAGI ELEKTRONICZNE
26 – 600 Radom, ul. Bracka 28
Phone: +48 48 38 48 800, fax. + 48 48 385 00 10
Sales department: + 48 48 366 80 06
ee--mmaaiill:: eexxppoorrtt@@rraaddwwaagg..ccoomm
wwwwww..rraaddwwaagg..ccoomm
Loading...