Quantum Q-Rad Operator’s Manual

Page 1
Q-Rad System
Operator’s Manual
Manual Part No. DC30-091 Revision F
Page 2
The original release of this manual dated 2011-02-24 (February 24, 2011) was drafted in the English language by Quantum Medical Imaging, LLC.
This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging, LLC. (QMI)
Copyright © 2012 QMI
Quantum Medical Imaging, LLC Division of Carestream
2002-B Orville Drive North Ronkonkoma, NY 11779-7661 USA Phone: (631) 567-5800 Fax: (631) 567-5074 E-mail: [email protected] www.quantummedical.net
Made in U.S.A.
Page 3

Revision History

Revision Date Type Of Modification
A 2011-02-24 Initial Release
B 2011-11-02 Added DX-D DR configuration
C 2011-08-30 Added QG-DIG-DRX integration option
D 2012-01-30 Incorporated ECO 2568, 2609
E 2012-02-10 Renamed Operator Manual, changed EU Representative
F 2012-05-09 Updated per IEC 60601-1: 2005, IEC 60601-1-3,
IEC 60601-2-54
Page
Number Rev
i thru iv F
1-1 thru 1-14 F
2-1 thru 2-18 F
Page
Number Rev
Page
Number Rev
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging
i
Page 4
Revision History
REVISIO N
A 2011-02-24--- Initial Release
B 2011-10-27All Various
DATE PAGES NOTES
THIS PAGE INTENTIONALLY LEFT BLANK
Revision F Q-Rad System - Operator’s Manual
ii Quantum Medical Imaging
Page 5

Table of Contents

Chapter 1: Safety Notices
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Warnings, Cautions, Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Intended Operator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Training. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Applicable Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Accompanying Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Environmental Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Electromagnetic Compatibility
(EN 60601-1-2:2007/IEC 60601-1-2:2007) . . . . . . . . . . . . . . . . . . . . . .1-8
Abbreviation Definition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
Chapter 2: General Information
Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Component Weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Generators (Cabinet) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Generators (Control Panel) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Tube Supports. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Upright Cassette Holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Heat Dissipation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Generators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
X-Ray Tube. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Tube Supports. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Upright Cassette Holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging iii
Page 6
Table of Contents
System Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Transport and Non-Operating Environment. . . . . . . . . . . . . . . . . . . . . .2-7
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Q-Rad Floor-Mounted System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Q-Rad “DRX Series” Integrated Radiography System . . . . . . . . . . . . . . .2-9
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
General Application Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Cleaning, Preventive Inspection and Maintenance . . . . . . . . . . . . . . . . . .2-11
Instructions for Disposal of Waste Equipment . . . . . . . . . . . . . . . . . . .2-11
Labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Power Disconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Detachable Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Revision F Q-Rad System - Operator’s Manual
iv Quantum Medical Imaging
Page 7
Chapter
C
HAPTER
1: S
AFETY
N
1
OTICES
1-1
Page 8
1-2
Page 9
Chapter 1 Safety Notices

General Safety Information

This product has been designed to meet stringent safety standards. All medical electrical equipment requires proper installation, operation, and maintenance (particularly with regard to safety).
It is vital that the user read, understand, note, and where applicable, strictly observe all Warnings, Cautions, Notes and Safety markings within this document and on the equipment, and that the user strictly follow all safety directions in this manual to help ensure the safety of users and patients.
Every reasonable precaution has been taken during manufacture to safeguard the health and safety of persons who will operate this equipment. The following precautions must be observed at all times.

Warnings, Cautions, Notes

The following samples show how warnings, cautions, and notes appear in this document. The text explains their intended use.
Indicates injury or death is possible if the instructions
Warning
Caution
Note
The purpose of safety icons, such as those shown below, is to indicate at a glance the type of caution, warning or danger.
Warning Ionizing radiation: indicates the possibility of
Warning Dangerous voltage: indicates the presence of high
Warning Warning, hot surface.
are not obeyed. Instructs users to refer to documentation if displayed without warning text.
Indicates that damage to equipment is possible if the instructions are not obeyed.
Notes provide advice and highlight unusual points. A note is not intended as an instruction.
increased levels of radiation.
voltage.
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 1-3
Page 10
Chapter 1 Safety Notices
WARNING
Quantum Medical Imaging, LLC disclaims all responsibility from any injury resulting from improper application of this equipment.
This equipment is sold to be used exclusively under the prescribed direc­tion of a person who is licensed by law to operate equipment of this nature. This equipment must be used in accordance with all safety procedures described in this manual and must not be used for purposes other than those described herein. In the United States, Federal law restricts this device to sale, distribution, and use by or on order of a licensed physician.
Quantum Medical Imaging, LLC cannot assume responsibility for any mal­functioning of this equipment resulting from improper operation, mainte­nance, or repair, or from damage or modification of its components.
Failure to observe these warnings may cause serious injuries.
WARNING
X-rays are hazardous to both patient and operator unless established safe exposure factors and operating instructions are observed.
Only qualified and authorized personnel shall operate this system. In this context, qualified means those legally permitted to operate this equipment in the jurisdiction in which the equipment is being used, and authorized means those authorized by the authority controlling the use of the equip­ment. Full use must be made of all radiation protection features, devices, systems, procedures and accessories.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont Ave­nue, Suite 800, Bethesda, Maryland 20814-3095 (www.ncrp.com), and of the International Commission on Radiological Protection (www.icrp.org), and take adequate steps to protect against injury.
WARNING
X-ray equipment may cause injury if used improperly. The instructions con­tained in this manual must be read and followed when operating this unit. Personal radiation monitoring and protective devices are available. You are urged to use them to protect against unnecessary x-ray exposure.
Revision F Q-Rad System - Operator’s Manual
1-4 Quantum Medical Imaging
Page 11
Chapter 1 Safety Notices
WARNING
Exterior surfaces of the X-ray Tube can rise in temperature to levels where the operator or patient could suffer burns. Avoid contact with the tube housing.
WARNING
Changes, additions or maintenance to the equipment carried out by per­sons without appropriate qualifications and training and/or using unap­proved spare parts may lead to serious risk of injury and damage to the equipment as well as making the warranty void.

Regulatory Compliance

This certified Quantum Medical Imaging, LLC medical device has been designed, manufactured, and calibrated to comply with governing Federal Regulations 21 CFR Subchapter J and the performance standards attendant thereto. Upon installation, all certified products require the filing of Form FD-2579 “Report of Assembly of a Diagnostic X-ray System” by the assembler (i.e., the installer) with the appropriate agencies; the “Installation Quality Assurance Checklist” must also be completed and properly distributed upon installation. A copy of each form (pink copy) is provided to the user. The installation report is completed by the installer and returned to Quantum Medical Imaging, LLC.
Those responsible for the planning of x-ray equipment installations must be thoroughly familiar and comply completely with NCRP Report No. 49, “Structural Shielding Design and Evaluation for Medical Use of X-Rays and Gamma Rays of Energies up to 10 MeV”, as revised or replaced in the future. Those authorized to operate, test, participate in or supervise the operation of the equipment must be thoroughly familiar and comply completely with the currently established safe exposure factors and procedures described in publications such as Subchapter J of Title 21 of the Code of Federal Regulations, “Diagnostic X-Ray Systems and Their Major Components,” and NCRP Report No. 102, “Medical X-Ray, Electron Beam and Gamma Ray Protection for Energies Up to 50 MeV—Equipment Design and Use” as revised or replaced in the future.
This equipment must only be used in rooms that comply with all applicable laws or regulations that have the force of law, concerning electrical safety for this type of equipment.
Scheduled maintenance is essential to the assurance of continued integrity of this equipment with respect to regulatory compliance. The continuance of certified performance to the regulatory standard is incumbent upon the user's diligent conformance to recommended maintenance instructions.
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 1-5
Page 12
Chapter 1 Safety Notices
N
Complies with EC Directive 93/42/EEC for Medical Devices, FDA Center for Devices and Radiological Health (CDRH) - Title 21 CFR Subchapter J
EU Authorized Representative:
Carestream Health France 1, rue Galilée 93192 NOISY-LE-GRAND CEDEX France

Classification

This product has been classified as Class I, Type B. This equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or with nitrous oxide. Protection against Harmful Ingress of Water (Ordinary), enclosed equipment without protection against ingress of liquids.
MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE, MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1-1 AND CAN/CSA C22.2 NO. 601.1
The following symbols may be used for marking on this equipment or equipment documentation:
Non-ionizing Radiation Consult operating instructions
Earth (ground) Dangerous Voltage
Protective Earth (ground) Type B Equipment
Connection point for the neutral conductor on PERMANENTLY INSTALLED EQUIPMENT
Alternating current
Revision F Q-Rad System - Operator’s Manual
1-6 Quantum Medical Imaging
Page 13
Chapter 1 Safety Notices

Compatibility

The equipment described in this manual must only be used in combination with other equipment or components if these are expressly recognized by Quantum Medical Imaging, LLC as compatible.

Intended Operator

This equipment is intended to be installed, used and operated only in accordance with the safety procedures given within this manual for the purpose for which it was designed. Before attempting to work with this equipment, read, understand, note and strictly observe all warnings, cautions and safety markings on the equipment.
Users include those persons who actually handle the equipment and those who have authority over the equipment.

Training

Users of this equipment shall have received adequate training on its safe and effective use before attempting to work with the equipment. Training requirements may vary from country to country. The User shall make sure that training is received in accordance with local laws or regulations that have the force of law.

Applicable Standards

The Q-Rad Radiographic System complies with:
• IEC 60601-1: 2005
• IEC 60601-1-2: 2007
• IEC 60601-1-3: 2008
• IEC 60601-1-6:2010
• IEC 62366: 2007

Accompanying Documentation

The documentation consists of a System Operator’s manual (this document), System Service manual (DC30-109) and related documentation, depending on system configuration, as shown in Chapter 2, GENERAL INFORMATION, Table 2-1. The documentation shall be kept with the system for easy reference.
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 1-7
Page 14
Chapter 1 Safety Notices

Environmental Protection

This equipment contains certain materials and chemical compounds incidental to the manufacture of electrical and electronic equipment, and improper “end-of-life” disposal of such equipment can result in environmental contamination. Therefore, this equipment should not be disposed of as ordinary household waste, but should instead be delivered to a designated electrical and electronic waste disposal or recycling center. For further information on disposing of electrical and electronic waste, contact the cognizant authority within the jurisdiction.

Electromagnetic Compatibility (EN 60601-1-2:2007/IEC 60601-1-2:2007)

The Q-Rad Radiographic System is intended for use in the electromagnetic environment specified below. As such, Q-Rad Radiographic System must be installed and put into service according to the information provided in the accompanying Service Manual. Unauthorized changes to this product Do not make any modifications
Portable and mobile RF communications equipment can affect medical electrical equipment. It is therefore recommended that the operation of equipment of this type such as mobile telephones, cordless microphones and other similar mobile radio equipment be restricted from the vicinity of this device.
Use of accessories, transducers and cables other than those specified in the accompanying documents, may result in increased emissions or decreased immunity of the equipment.
Guidance and manufacturer's declaration - electromagnetic emissions
The Q-Rad Radiographic System is intended for use in the electromagnetic environment spec­ified below. The customer or the user of the Q-Rad Radiographic System should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment -
guidance
RF emissions CISPR 11
Group 1 The Q-Rad Radiographic System
uses RF energy only for their internal functions. Therefore, the RF emis­sions are very low and are not likely to cause any interference in nearby electronic equipment.
Revision F Q-Rad System - Operator’s Manual
1-8 Quantum Medical Imaging
Page 15
Chapter 1 Safety Notices
Guidance and manufacturer's declaration - electromagnetic emissions
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Class A The Q-Rad Radiographic System is
suitable for use in all establishments
Class A
Complies
other than domestic and those directly connected to the public low­voltage power supply network that supplies buildings used for domestic purposes.
Guidance and manufacturer's declaration - electromagnetic immunity
The Q-Rad Radiographic System is intended for use in the electromagnetic environment specified below. The customer or the user of the Q-Rad Radiographic System should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
Compliance
level
±6 kV contact ±8 kV air
Electromagnetic environment
- guidance
Floors should be wood, con­crete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 1-9
Page 16
Chapter 1 Safety Notices
Guidance and manufacturer's declaration - electromagnetic immunity
Voltage dips, short interruption, and voltage variations on power supply input lines IEC 60601-4-11
Power frequency (50/60 Hz) IEC 61000-4-8
NOTE: U
is the A.C. mains voltage prior to application of the test level.
T
< 5 % U
T
(>95 % dip in UT) for 0.5 cycle
< 5 % U
T
(>95 % dip in UT) for 0.5 cycle
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Q-Rad Radiographic Sys-
40 % U
T
(60 % dip in UT) for 5 cycles
70 % U
T
(30 % dip in UT) < 5 % U
T
(> 95 % dip in UT) for 5 s
40 % U
T
(60 % dip in UT) for 5 cycles
70 % U
T
(30 % dip in UT) < 5 % U
T
(> 95 % dip in UT) for 5 s
tem requires continued opera­tion during power mains interruptions, it is recommended that the Q-Rad Radiographic System be powered from an uninterruptible power supply or battery.
3 A/m 3 A/m Power frequency magnetic
fields should be at levels char­acteristic of a typical location in a typical commercial or hospital environment
Guidance and manufacturer's declaration - electromagnetic immunity
The Q-Rad Radiographic System is intended for use in the electromagnetic environment specified below. The customer or the user of the Q-Rad Radiographic System Q-Rad Radiographic System should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Q-Rad Radiographic System, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Revision F Q-Rad System - Operator’s Manual
1-10 Quantum Medical Imaging
Page 17
Chapter 1 Safety Notices
Pd 2,1
Pd 2,1
Pd 3,2
Guidance and manufacturer's declaration - electromagnetic immunity
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2,5 GHz
3 Vrms
3 V/m
, 80 MHz to 800 MHz
, 800 MHz to 2,5 GHz
where P is the maximum output power rat­ing of the transmitter in watts (W) accord­ing to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
a
survey
ance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following sym­bol:
, should be less than the compli-
b
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele­phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Q-Rad Radiographic System is used exceeds the applicable RF compliance level above, the Q-Rad Radiographic System should be observed to verify normal operation. If abnor­mal performance is observed, additional measures may be necessary, such as re-orienting or relocat­ing the Q-Rad Radiographic System.
b
Over the frequency range 150 kHz to 80 kHz, field strengths should be less than 3 V/m.
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 1-11
Page 18
Chapter 1 Safety Notices
Pd 2,1
Pd 2,1
Pd 3,2
Recommended separation distances between
portable and mobile RF communications equipment and the
Q-Rad Radiographic System
The Q-Rad Radiographic System is intended for use in an electromagnetic environment in which radi­ated RF disturbances are controlled. The customer or the user of the Q-Rad Radiographic System Q­Rad Radiographic System can help prevent electromagnetic interference by maintaining a minimum dis­tance between portable and mobile RF communications equipment (transmitters) and the Q-Rad Radio­graphic System as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation dis­tance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Revision F Q-Rad System - Operator’s Manual
1-12 Quantum Medical Imaging
Page 19
Chapter 1 Safety Notices

Abbreviation Definition

The following abbreviations and acronyms may be found in this document. Their definition is explained below.
AEC Automatic Exposure Control
Btu British Thermal Unit
CR Computed Radiography
dDistance
DR Digital Radiography
HF High Frequency
hPa Hecto-Pascal
hr Hour
kg Kilogram
kHz Kilohertz
KVA Kilovolt-Ampere
lb Pound
m Meter
MHz Megahertz
mm Millimeter
OCP Operator Control Panel
P Power (max.)
RF Radio Frequency
VVolt
WWatt
WEEE Waste Electrical and Electronic Equipment
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 1-13
Page 20
Chapter 1 Safety Notices
THIS PAGE INTENTIONALLY LEFT BLANK
Revision F Q-Rad System - Operator’s Manual
1-14 Quantum Medical Imaging
Page 21
Chapter
C
HAPTER
2: G
ENERAL
I
NFORMATION
2
2-1
Page 22
2-2
Page 23
Chapter 2 General Information

Instructions for Use

The instructions for using Quantum’s Q-Rad Radiographic System, including an explanation of the function of controls, displays and signals, and the sequence of operation, are contained in each of the system component user’s manual indicated in Tabl e 2- 1.
WARNING: This equipment must be operated with reasonable care. Manufacturer’s equipment recommendations described in this manual mus t be observed.
WARNING: Use a focal spot-to-film distance as large as possible in order to keep the absorbed dose to the patient as low as reasonably achievable.
WARNING: During each exposure, the individual operating the X-ray machine must exit the radiographic operatory and stand behind a leaded wall.
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-3
Page 24
Chapter 2 General Information

Overview

This chapter describes the configurations of the Q-Rad Radiographic System, which typically includes Quantum Medical Imaging HF Series X-ray Generators, radiographic tables, free-standing tube support, upright cassette holder, collimator, X-ray tube, ion chamber (when equipped with AEC option), and image receptor.
Detailed instructions for successful integration of the compatible Quantum components are published in the Service Manuals for each of the various system components.
Table 2-1. Q-Rad System Components User/Service Documents
Component Type Model(s) User Manual P/N Service Manual
P/N
X-ray Generator, Single-phase, non­Stored Energy
X-ray Generator, Single-phase, Stored Energy
X-ray Generator, Three-phase
TechVision Supple­mentary Generator Control Panel
Radiographic Table, Elevating Tabletop
Radiographic Table, Non-Elevating Table­top
QG-25, QG-25-5, QG-32, QG-32-5, QG-40, and QG-40-5
QG-20-SE, QG-32-SE, QG-40-SE, QG-50-SE
QG-32-2, QG-32-3, QG-40-2, QG-40-3, QG-50, QG-50-2, QG-65, and QG-80
QG-TV-Q, QG-TV-IP-Q, QG-TV-F, QG-TV-IP-F, QG­TV-C, QG-TV-IP-C
QT-750 DC30-006 DC30-007
QT-740 DC30-006 DC30-007
Odyssey - DC30-010 Quest - DC30-015 Q-Vision - DC30-101
Odyssey - DC30-010 Quest - DC30-015 Q-Vision - DC30-101
Odyssey - DC30-010 Quest - DC30-015 Q-Vision - DC30-101
DC30-048 (Quantum non-integrated systems, DC30-120 (Q-Vision non-integrated sys­tems) DC30-126 (Carestream integrated systems)
Q-Vision/Odyssey ­DC30-011 Quest - DC30-063
Q-Vision/Odyssey ­DC30-011 Quest - DC30-063
Q-Vision/Odyssey ­DC30-011 Quest - DC30-063
DC30-049
Radiographic Table, Non-Elevating Table­top
Tube Support, Floor/ Wall Mounted
Tube Support, Free­Standing (Floor) Mounted
QT-730 DC30-003 DC30-003
QS-500 DC30-034 DC30-005
QS-550 DC30-034 DC30-009
Revision F Q-Rad System - Operator’s Manual
2-4 Quantum Medical Imaging
Page 25
Chapter 2 General Information
Table 2-1. Q-Rad System Components User/Service Documents
Component Type Model(s) User Manual P/N Service Manual
P/N
Radiographic Upright Cassette Holder,
Collimator, Manual Progeny, MC150
Collimator, Automatic Progeny, Linear II
Collimator, Program­mable
X-ray tube Toshiba E7239X, E7242X,
DX-D 500
Digital DRX Series Integration Software (QG-DIG-DRX) (Optional)
n
Q-Rad
QW-420, QW-420-S, QW­420-D, QW-400
Huestis, 150MC Ralco R108
Progeny, Linear IV
Huestis, CM-32211-QMI CM-21-002 CM-21-002
E7252X, E7254X, E7869X
Varian RAD-14, RAD-60, RAD-92
Dunlee DA 1092, DA 1036
QG-DIG-DRX DC30-123 DC30-123
DC30-039 DC30-040
00-02-1078 CM-21-001 MTR108
00-02-1559 00-02-1517
Refer to data sheet pro­vided with tube
00-02-1078 CM-21-001 MTR108
00-02-1559 00-02-1517
Refer to data sheet provided with tube
Dose Area Product (DAP) meter (Optional)
Vacutec VacuDAP OEM DAP-std-fl-
tw_BA011.3.doc
DAP-std-fl­tw_BA011.3.doc
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-5
Page 26
Chapter 2 General Information

Component Weights

Generators (Cabinet)
Single-Phase, non-Stored Energy (SE) units: 135.2 kg (298.0 lb)
32, 40, and 50 kW, SE units: 195.5 kg (431.0 lb)
Three-Phase units: 172.4 kg (380.0 lb)
Generators (Control Panel)
Odyssey (without pedestal): 2.3 kg (5.0 lb) Odyssey (with pedestal): 11.3 kg (25.0 lb) Quest (without pedestal): 1.8 kg (4.0 lb) Quest (with pedestal): 10.9 kg (24.0 lb) Q-Vision: 5.9 kg (13.0 lb) TechVision Option: 4.3 kg (9.5 lb)
without internal High Speed Starter (HSS)
176.0 kg (388.0 lb) with internal High Speed Starter (HSS)
Tables
Model QT-730: 79.4 kg (175.0 lb) Model QT-740: 158.8 kg (350.0 lb) Model QT-750: 226.8 kg (500.0 lb)
Tube Supports
Model QS-550: 272.2 kg (600.0 lb) Model QS-500: 230.4 kg (508.0 lb)
Upright Cassette Holders
Model QW-420: 102.1 kg (225.0 lb) Model QW-420-S: 102.1 kg (225.0 lb) Model QW-420-D: 90.7 kg (200.0 lb) Model QW-400: 45.4 kg (100.0 lb)
Revision F Q-Rad System - Operator’s Manual
2-6 Quantum Medical Imaging
Page 27

Heat Dissipation

Generators
Cabinet Typical Heat Output: 3600 BTU per hour Control Panel Typical Heat Output: 130 BTU per hour
X-Ray Tube
Max. Heat Output (600 khu tube): 2000 BTU per hour
Tables
QT-730 Typical Heat Output: 40 BTU per hour (from receptor) QT-750 Typical Heat Output: 400 BTU per hour QT-740 Typical Heat Output: 100 BTU per hour
Tube Supports
QS-500/QS-550: 200 BTU per hour
Upright Cassette Holders
Typical Heat Output: 50 BTU per hour
Chapter 2 General Information
NOTE: The above calculations are typical, based upon average usage. Air
circulation around the electronic cabinets must be provided. This includes the sides, back, front, and top of cabinet.

System Operating Environment

Ambient Temperature: 10–40 °C (50–104 °F) Relative Humidity: 15–75 % Atmospheric Pressure: 700–1060 hPa

Transport and Non-Operating Environment

Ambient Temperature: -25 to +55 °C (-13 to 131 °F) Relative Humidity: 5–95 % Atmospheric Pressure: 500–1060 hPa
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-7
Page 28
Chapter 2 General Information

System Configuration

The Quantum Q-Rad System consists of a combination of integrated radiographic components, typically including an X-ray generator, X-ray tube and tube support, collimator, radiographic table, and an upright cassette holder. Different types of image receptors may be utilized with the Q-Rad System, including computed radiography (CR) receptors, as well as conventional analog film cassettes.

Q-Rad Floor-Mounted System

The Q-Rad Floor-Mounted system is configured around either a free-standing, floor-mounted tube support (Model QS-550) or a wall/ceiling mounted tube support (Model QS-500). Because the free-standing version does not require wall attachment for support, it provides enhanced freedom of movement in the patient vicinity for the operator.
Recumbent examinations are accomplished utilizing an elevating (Model QT-750) or the non-elevating (Model QT-740 and QT-730) radiologic tables. Standing (weight-bearing) examinations are performed using an upright, non-tilting (Model QW-420 Series) chest stand (cassette holder).
The X-ray generator is available in various power output and operator control panel types, with the typical configuration including a 80 kW output generator with an Odyssey operator control.
Table 2-2. Q-Rad Floor Mounted System
COMPONENT COMPONENT
COMPONENT MODEL#
DESIGNATION
GENERATOR QG1 SINGLE-PHASE: QG-25-1, QG-32-1, QG-40-1
STORED ENERGY QG-40-SE, QG-50-SE THREE-PHASE QG-40-3, QG-50-3, QG-65-3, QG-80-3
GENERATOR CONTROL PANEL
SUPPLEMENTAL GENERATOR CONTROL PANEL (TECHVISION)
TABLE QT1 QT-730, QT-740, QT-750
TUBE SUPPORT QS1 QS-500, QS-550
WALL STAND QW1 QW-420, QW-420-D, QW-420-S, QW-400
OCP1 QUEST SERIES
QG-400, QG-500 ODYSSEY SERIES QG-5000, QG-6500, QG-8000 Q-VISION SERIES QGV-65, QGV-80
OCP2 QG-TV-IP-F
: QG-20-SE, QG-32-SE,
: QG-32-2, QG-32-3, QG-40-2,
: QG-200, QG-250, QG-320,
: QG-3200, QG-4000,
: QGV-32, QGV-40, QGV-50,
X-RAY TUBE TU1 TOSHIBA
E7869X VARI A N DUNLEE
: E7239X, E7242X, E7252X, E7254X,
: RAD-14, RAD-60, RAD-92
: DA 1092, DA 1036
Revision F Q-Rad System - Operator’s Manual
2-8 Quantum Medical Imaging
Page 29
Chapter 2 General Information
Table 2-2. Q-Rad Floor Mounted System
COMPONENT COMPONENT
COMPONENT MODEL#
DESIGNATION
COLLIMATOR QC1 PROGENY
LINEAR II, LINEAR IV RALCO HUESTIS
DAP QD1 VacuDAP-OEM DISPLAY + ION CHAMBER,
VacuDAP ION CHAMBER
ION CHAMBER IN1 AID
IMAGE RECEPTOR
RC1 GRID CABINET
: ICX 1153
BUCKY CABINET KLB10-08013
: MC150, MC150 PINNACLE,
: R 108/464 DHHS
: 150MC (MANUAL), CM 32211-QM1
: R50-17GC
: PROGENY KLB10-08011,

Q-Rad “DRX Series” Integrated Radiography System

The Q-Rad DRX Series Integrated Digital Radiography (DR) System combines the Q-Rad System’s HF Series X-ray Generator and Carestream DRX-1 Digital Imaging Systems by way of a common user interface (QG-DIG-DRX Software) running on the Carestream DRX-1 Digital Imaging System’s Workstation computer.
When QG-DIG-DRX Software is implemented, all user interaction (except PREP and EXPOSE) with the HF Series X-ray Generator is accomplished via the Digital Imaging System’s touchscreen computer, thereby eliminating the need for a separate generator Operator Control Panel.
The Q-Rad DRX Series Integrated System consists of either a fixed DR Detector, and/or a Portable DR Detector, and a Workstation Computer. The DR System is connected to the Quantum X-Ray Generator by way of a software interface module.
Table 2-3. Q-Rad System, DRX Series Integration
COMPONENT COMPONENT
DESIGNATION
GENERATOR QG1 SINGLE-PHASE
STORED ENERGY QG-40-SE, QG-50-SE THREE-PHASE QG-40-3, QG-50-3, QG-65-3, QG-80-3
GENERATOR CON­TROL PANEL
SUPPLEMENTAL GENERATOR CON­TROL PANEL (TECH­VISION)
TABLE QT1 QT-730, QT-740, QT-750
OCP1 Q-CONNECT (WITH DRX-1 WORKSTATION)
QG-DIG-DRX
OCP2 QG-TV-IP-C, QG-TV-IP-F
COMPONENT MODEL#
: QG-32-1, QG-40-1
: QG-32-SE,
: QG-32-2, QG-32-3, QG-40-2,
:
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-9
Page 30
Chapter 2 General Information
Table 2-3. Q-Rad System, DRX Series Integration
COMPONENT COMPONENT
DESIGNATION
TUBE SUPPORT QS1 FLOOR-MOUNTED: QS-500, QS-550
WALL STAND QW1 QW-420, QW-420-D, QW-420-S
X-RAY TUBE TU1 TOSHIBA
E7869X VARIAN DUNLEE
COLLIMATOR QC1 PROGENY
LINEAR II, LINEAR IV RALCO HUESTIS (PROGRAMMABLE)
DAP QD1 VacuDAP-OEM DISPLAY + ION CHAMBER,
VacuDAP ION CHAMBER
ION CHAMBER IN1 AID
IMAGE DETECTOR CABINET
RC1 FIXED DETECTOR
: ICX 1153
Q-FDR-T (TABLE) PORTABLE DETECTOR STAND), Q-DRT-T (TABLE)
COMPONENT MODEL#
: E7239X, E7242X, E7252X, E7254X,
: RAD-14, RAD-60, RAD-92
: DA 1092, DA 1036
: MC150, MC150 PINNACLE,
: R 108/464 DHHS
: 150MC (MANUAL), CM-32211-QMI
: Q-FDR (WALL STAND),
: Q-DRT (WALL
DR IMAGING DETEC­TOR/SYSTEM
QR1, QR2 CARESTREAM DRX-1 IMAGING SYSTEM
(PORTABLE, WIRELESS) CARESTREAM DRX-1 IMAGING SYSTEM (FIXED, 4343R)
Revision F Q-Rad System - Operator’s Manual
2-10 Quantum Medical Imaging
Page 31
Chapter 2 General Information

Intended Use

The Quantum Q-Rad System provides diagnostic quality images to aid the physician with diagnosis. The system can be used to perform radiographic exposures of the skeleton (including skull, spinal column, and extremities), chest, abdomen, and other body parts. The Q-Rad Radiographic System is not indicated for use in mammography.

General Application Specification

Patient Population:
Age: Newborn to geriatric Weight: Up to 294.8 kg (650.0 lb) Region of Body: Produces images of any bone in the human body,
including the hand, wrist, arm, foot, ankle, knee, leg, or spine.
Operator Requirements:
Education/Training: This equipment is sold to be used exclusively under the
prescribed direction of a person who is licensed by law to operate equipment of this nature.
Application:
Location: Professional use only, indoor use only. Physical Environment: Refer to Q-Rad System operating and non-operating
specifications in this chapter.
Frequency of Use: Intermittent operation, limited primarily by loading
characteristics specific to tube type used in the system.
Mobility: Stationary

Cleaning, Preventive Inspection and Maintenance

Instructions for cleaning, preventive inspection and routine maintenance of Q-Rad Radiographic System, are included in each of the system component user’s and service manuals indicated in Table 2- 1 .

Instructions for Disposal of Waste Equipment

In the European Union, this symbol indicates that when the last user wishes to discard this product, it must be sent to appropriate facilities for recovery and recycling. Contact your local representative for additional information on the collection and recovery programs available for this product.
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-11
Page 32
Chapter 2 General Information
REGULATORY
LABEL
CONFIGURATION
LABEL
GENERATOR CABINET
GENERATOR COMPLIANCE LABEL

Labeling

Each component of the Q-Rad System has its own rating plate/serial label. The following table provides the general location of the rating plate/serial label on each of the various types of components that could be incorporated in a Q-Rad System.
Table 2-4. Q-Rad System Components Labeling Guide
Component Type Model(s) Label Location
Q-Rad System Q-Rad
X-ray Generator Single-phase,
non-Stored Energy QG-25, QG-25-5, QG-32, QG-32-5, QG-40, and QG-40-5
Single-phase, Stored Energy QG-20-SE, QG-32-SE, QG-40-SE, QG-50-SE
Three-phase QG-32-2, QG-32-3, QG-40-2, QG-40-3, QG-50, QG-50-2, QG-65, and QG-80
Revision F Q-Rad System - Operator’s Manual
2-12 Quantum Medical Imaging
Page 33
Chapter 2 General Information
compliance label
Table 2-4. Q-Rad System Components Labeling Guide
Component Type Model(s) Label Location
TechVision
Supplementary Gener-
ator Control Panel
Radiographic Table Fixed Non-Float Tabletop
Radiographic Table Non-Elevating Tabletop
QG-TV-F, QG-TV-IP-F, QG-TV-C, QG-TV-IP-C
QT-730
QT-740
Elevating Tabletop QT-750
Tube Support,
Floor Mounted
QS-500 QS-550
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-13
Page 34
Chapter 2 General Information
Table 2-4. Q-Rad System Components Labeling Guide
Component Type Model(s) Label Location
Radiographic Upright Cassette Holder
Collimator, Manual PROGENY MC-150
QW-420, QW-420-S, QW-420-D QW-400
HUESTIS 150MC RALCO R108/484 DHHS
Collimator, Automatic PROGENY Linear II,
Linear IV HUESTIS CM-32211-QM1
Revision F Q-Rad System - Operator’s Manual
2-14 Quantum Medical Imaging
Page 35
Chapter 2 General Information
X-RAY TUBE
COMPLIANCE LABEL
LOCATIONS
REAR VIEW
COMPLIANCE
LABEL/DATAPLATE
CABINET (FRONT)
Table 2-4. Q-Rad System Components Labeling Guide
Component Type Model(s) Label Location
X-ray tube
VARI A N:
RAD-14, RAD-60, RAD-92
TOSHIBA:
DUNLEE:
Digital Image Detector
Cabinet
E7239X, E7242X, E7252X, E7254X, E7869X
DA 1092, DA 1036
Q-DRT Q-DRT-T Q-FDR Q-FDR-T
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-15
Page 36
Chapter 2 General Information
REAR VIEW
COMPLIANCE
LABEL
Table 2-4. Q-Rad System Components Labeling Guide
Component Type Model(s) Label Location
Integration Option Q-Rad DRX Series
(QG-DIG-DRX)
Revision F Q-Rad System - Operator’s Manual
2-16 Quantum Medical Imaging
Page 37
Chapter 2 General Information

Power Disconnection

Refer to the following table for Q-Rad System component power disconnection methods.
Table 2-5. DX-D 500
n
Q-Rad System Component Power Disconnection Method
Component Type Model Means of Power
Disconnect
X-ray Generator, Single-phase, non­Stored Energy
X-ray Generator, Single-phase, Stored Energy
X-ray Generator, Three­phase
TechVision Supplemen­tary Generator Control Panel
Radiographic Table, Non-Elevating Tabletop
Radiographic Table, Non-Elevating Tabletop
QG-25, QG-25-5, QG-32, QG-32-5, QG-40, and QG-40-5
QG-20-SE, QG-32-SE, QG-40-SE, QG-50-SE
QG-32-2, QG-32-3, QG-40-2, QG-40-3, QG-50, QG-50-2, QG-65, and QG-80
QG-TV-Q, QG-TV-IP-Q, QG-TV-F, QG-TV-IP-F, QG-TV-C, QG-TV-IP-C
QT-730 ON/OFF circuit breaker
QT-740 Unplug Line Cord
ON/OFF circuit breaker CB1 on generator cabinet
ON/OFF circuit breaker CB1 on generator cabinet
ON/OFF circuit breaker CB1 on generator cabinet
ON/OFF circuit breaker CB1 on generator cabinet
CB1 on generator cabinet
Radiographic Table, Elevating Tabletop
Tube Support, Floor/Wall Mounted
Tube Support, Free­Standing (Floor) Mounted
QT-750 Unplug Line Cord
QS-500 ON/OFF circuit breaker
CB1 on generator cabinet
QS-550 ON/OFF circuit breaker
CB1 on generator cabinet
Q-Rad System - Operator’s Manual Revision F Quantum Medical Imaging 2-17
Page 38
Chapter 2 General Information
Table 2-5. DX-D 500
n
Q-Rad System Component Power Disconnection Method
Component Type Model Means of Power
Disconnect
Radiographic Upright Cassette Holder, Non­Tilting
Collimator, Manual Progeny, MC150
Collimator, Automatic Progeny, Linear II
X-ray tube Toshiba E7239X, E7242X,
Dose Area Product Meter (optional)
QW-420, QW-420-S, QW-420-D, QW-400
Huestis, 150MC Ralco, R108
Progeny, Linear IV Huestis Programmable, CM-32211-QMI
E7252X, E7254X, E7869X Varian RAD-14, RAD-60, RAD-92 Dunlee DA 1092, DA1036
Vacutec Display Unit: VacuDAP OEM (QD1-2) Vacutec Ion Chamber: Vac­uDAP OEM (QD1-1)
ON/OFF circuit breaker CB1 on generator cabinet
ON/OFF circuit breaker CB1 on generator cabinet
Unplug Line Cord (ACS Power Supply & Junction Box)
ON/OFF circuit breaker CB1 on generator cabinet
Unplug Line Cord
Carestream Digital Imaging System
DRX-1 Quick Reference Card for
the Carestream DRX-1 System, document 8G8318 Carestream DRX-1 System Safety and Regulatory Information with Hardware User’s Guide, document 7H8166 Carestream DRX-1 System Battery Charger User’s Guide, document 8H6193

Detachable Parts and Accessories

Instructions for detachable parts and accessories of Q-Rad Radiographic System are included in each of the system component user’s manual indicated in Tab le 2 -1 .
<End of Document>
Revision F Q-Rad System - Operator’s Manual
2-18 Quantum Medical Imaging
Loading...