The original release of this manual dated 2011-02-24 (February 24, 2011) was drafted in the
English language by Quantum Medical Imaging, LLC.
This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in
writing from Quantum Medical Imaging, LLC. (QMI)
This product has been designed to meet stringent safety standards. All medical
electrical equipment requires proper installation, operation, and maintenance
(particularly with regard to safety).
It is vital that the user read, understand, note, and where applicable, strictly observe
all Warnings, Cautions, Notes and Safety markings within this document and on the
equipment, and that the user strictly follow all safety directions in this manual to help
ensure the safety of users and patients.
Every reasonable precaution has been taken during manufacture to safeguard the
health and safety of persons who will operate this equipment. The following
precautions must be observed at all times.
Warnings, Cautions, Notes
The following samples show how warnings, cautions, and notes appear in this
document. The text explains their intended use.
Indicates injury or death is possible if the instructions
Warning
Caution
Note
The purpose of safety icons, such as those shown below, is to indicate at a glance the
type of caution, warning or danger.
WarningIonizing radiation: indicates the possibility of
WarningDangerous voltage: indicates the presence of high
WarningWarning, hot surface.
are not obeyed. Instructs users to refer to
documentation if displayed without warning text.
Indicates that damage to equipment is possible if the
instructions are not obeyed.
Notes provide advice and highlight unusual points. A
note is not intended as an instruction.
increased levels of radiation.
voltage.
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging1-3
Page 10
Chapter 1 Safety Notices
WARNING
Quantum Medical Imaging, LLC disclaims all responsibility from any injury
resulting from improper application of this equipment.
This equipment is sold to be used exclusively under the prescribed direction of a person who is licensed by law to operate equipment of this nature.
This equipment must be used in accordance with all safety procedures
described in this manual and must not be used for purposes other than
those described herein. In the United States, Federal law restricts this
device to sale, distribution, and use by or on order of a licensed physician.
Quantum Medical Imaging, LLC cannot assume responsibility for any malfunctioning of this equipment resulting from improper operation, maintenance, or repair, or from damage or modification of its components.
Failure to observe these warnings may cause serious injuries.
WARNING
X-rays are hazardous to both patient and operator unless established safe
exposure factors and operating instructions are observed.
Only qualified and authorized personnel shall operate this system. In this
context, qualified means those legally permitted to operate this equipment
in the jurisdiction in which the equipment is being used, and authorized
means those authorized by the authority controlling the use of the equipment. Full use must be made of all radiation protection features, devices,
systems, procedures and accessories.
It is important that everyone having anything to do with x-radiation be
properly trained and fully acquainted with the recommendations of the
National Council on Radiation Protection and Measurements as published
in NCRP Reports available from NCRP Publications, 7910 Woodmont Avenue, Suite 800, Bethesda, Maryland 20814-3095 (www.ncrp.com), and of
the International Commission on Radiological Protection (www.icrp.org),
and take adequate steps to protect against injury.
WARNING
X-ray equipment may cause injury if used improperly. The instructions contained in this manual must be read and followed when operating this unit.
Personal radiation monitoring and protective devices are available. You are
urged to use them to protect against unnecessary x-ray exposure.
Revision FQ-Rad System - Operator’s Manual
1-4Quantum Medical Imaging
Page 11
Chapter 1 Safety Notices
WARNING
Exterior surfaces of the X-ray Tube can rise in temperature to levels where
the operator or patient could suffer burns. Avoid contact with the tube
housing.
WARNING
Changes, additions or maintenance to the equipment carried out by persons without appropriate qualifications and training and/or using unapproved spare parts may lead to serious risk of injury and damage to the
equipment as well as making the warranty void.
Regulatory Compliance
This certified Quantum Medical Imaging, LLC medical device has been designed,
manufactured, and calibrated to comply with governing Federal Regulations 21 CFR
Subchapter J and the performance standards attendant thereto. Upon installation, all
certified products require the filing of Form FD-2579 “Report of Assembly of a
Diagnostic X-ray System” by the assembler (i.e., the installer) with the appropriate
agencies; the “Installation Quality Assurance Checklist” must also be completed and
properly distributed upon installation. A copy of each form (pink copy) is provided to
the user. The installation report is completed by the installer and returned to
Quantum Medical Imaging, LLC.
Those responsible for the planning of x-ray equipment installations must be
thoroughly familiar and comply completely with NCRP Report No. 49, “Structural
Shielding Design and Evaluation for Medical Use of X-Rays and Gamma Rays of
Energies up to 10 MeV”, as revised or replaced in the future. Those authorized to
operate, test, participate in or supervise the operation of the equipment must be
thoroughly familiar and comply completely with the currently established safe
exposure factors and procedures described in publications such as Subchapter J of
Title 21 of the Code of Federal Regulations, “Diagnostic X-Ray Systems and Their
Major Components,” and NCRP Report No. 102, “Medical X-Ray, Electron Beam and
Gamma Ray Protection for Energies Up to 50 MeV—Equipment Design and Use” as
revised or replaced in the future.
This equipment must only be used in rooms that comply with all applicable laws or
regulations that have the force of law, concerning electrical safety for this type of
equipment.
Scheduled maintenance is essential to the assurance of continued integrity of this
equipment with respect to regulatory compliance. The continuance of certified
performance to the regulatory standard is incumbent upon the user's diligent
conformance to recommended maintenance instructions.
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging1-5
Page 12
Chapter 1 Safety Notices
N
Complies with EC Directive 93/42/EEC for Medical Devices,
FDA Center for Devices and Radiological Health (CDRH) - Title 21 CFR Subchapter J
EU Authorized Representative:
Carestream Health France
1, rue Galilée
93192 NOISY-LE-GRAND CEDEX
France
Classification
This product has been classified as Class I, Type B. This equipment is not suitable
for use in the presence of a flammable anesthetic mixture with air or with oxygen
or with nitrous oxide. Protection against Harmful Ingress of Water (Ordinary),
enclosed equipment without protection against ingress of liquids.
MEDICAL ELECTRICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK, FIRE,
MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 60601-1-1 AND
CAN/CSA C22.2 NO. 601.1
The following symbols may be used for marking on this equipment or equipment
documentation:
Connection point for the
neutral conductor on
PERMANENTLY INSTALLED
EQUIPMENT
Alternating current
Revision FQ-Rad System - Operator’s Manual
1-6Quantum Medical Imaging
Page 13
Chapter 1 Safety Notices
Compatibility
The equipment described in this manual must only be used in combination with
other equipment or components if these are expressly recognized by Quantum
Medical Imaging, LLC as compatible.
Intended Operator
This equipment is intended to be installed, used and operated only in accordance
with the safety procedures given within this manual for the purpose for which it
was designed. Before attempting to work with this equipment, read, understand,
note and strictly observe all warnings, cautions and safety markings on the
equipment.
Users include those persons who actually handle the equipment and those who
have authority over the equipment.
Training
Users of this equipment shall have received adequate training on its safe and
effective use before attempting to work with the equipment. Training
requirements may vary from country to country. The User shall make sure that
training is received in accordance with local laws or regulations that have the
force of law.
Applicable Standards
The Q-Rad Radiographic System complies with:
•IEC 60601-1: 2005
•IEC 60601-1-2: 2007
•IEC 60601-1-3: 2008
•IEC 60601-1-6:2010
•IEC 62366: 2007
Accompanying Documentation
The documentation consists of a System Operator’s manual (this document),
System Service manual (DC30-109) and related documentation, depending on
system configuration, as shown in Chapter 2, GENERAL INFORMATION, Table
2-1. The documentation shall be kept with the system for easy reference.
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging1-7
Page 14
Chapter 1 Safety Notices
Environmental Protection
This equipment contains certain materials and chemical compounds incidental to the
manufacture of electrical and electronic equipment, and improper “end-of-life”
disposal of such equipment can result in environmental contamination. Therefore, this
equipment should not be disposed of as ordinary household waste, but should instead
be delivered to a designated electrical and electronic waste disposal or recycling
center. For further information on disposing of electrical and electronic waste, contact
the cognizant authority within the jurisdiction.
Electromagnetic Compatibility
(EN 60601-1-2:2007/IEC 60601-1-2:2007)
The Q-Rad Radiographic System is intended for use in the electromagnetic
environment specified below. As such, Q-Rad Radiographic System must be installed
and put into service according to the information provided in the accompanying
Service Manual. Unauthorized changes to this product Do not make any modifications
Portable and mobile RF communications equipment can affect medical electrical
equipment. It is therefore recommended that the operation of equipment of this type
such as mobile telephones, cordless microphones and other similar mobile radio
equipment be restricted from the vicinity of this device.
Use of accessories, transducers and cables other than those specified in the
accompanying documents, may result in increased emissions or decreased immunity
of the equipment.
Guidance and manufacturer's declaration - electromagnetic emissions
The Q-Rad Radiographic System is intended for use in the electromagnetic environment specified below. The customer or the user of the Q-Rad Radiographic System should assure that it
is used in such an environment.
uses RF energy only for their internal
functions. Therefore, the RF emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
Revision FQ-Rad System - Operator’s Manual
1-8Quantum Medical Imaging
Page 15
Chapter 1 Safety Notices
Guidance and manufacturer's declaration - electromagnetic emissions
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Class AThe Q-Rad Radiographic System is
suitable for use in all establishments
Class A
Complies
other than domestic and those
directly connected to the public lowvoltage power supply network that
supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration - electromagnetic immunity
The Q-Rad Radiographic System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Q-Rad Radiographic System should assure that it is used in such an environment.
Immunity testIEC 60601
test level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Compliance
level
±6 kV contact
±8 kV air
Electromagnetic environment
- guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging1-9
Page 16
Chapter 1 Safety Notices
Guidance and manufacturer's declaration - electromagnetic immunity
Voltage dips,
short interruption,
and voltage
variations on power
supply
input lines
IEC 60601-4-11
Power frequency
(50/60 Hz)
IEC 61000-4-8
NOTE: U
is the A.C. mains voltage prior to application of the test level.
T
< 5 % U
T
(>95 % dip in UT)
for 0.5 cycle
< 5 % U
T
(>95 % dip in UT)
for 0.5 cycle
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Q-Rad Radiographic Sys-
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
< 5 % U
T
(> 95 % dip in UT)
for 5 s
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
< 5 % U
T
(> 95 % dip in UT)
for 5 s
tem requires continued operation during power mains
interruptions, it is recommended
that the Q-Rad Radiographic
System be powered from an
uninterruptible power supply or
battery.
3 A/m3 A/mPower frequency magnetic
fields should be at levels characteristic of a typical location in
a typical commercial or hospital
environment
Guidance and manufacturer's declaration - electromagnetic immunity
The Q-Rad Radiographic System is intended for use in the electromagnetic environment specified
below. The customer or the user of the Q-Rad Radiographic System Q-Rad Radiographic System
should assure that it is used in such an environment.
Immunity testIEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Q-Rad Radiographic System,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
Revision FQ-Rad System - Operator’s Manual
1-10Quantum Medical Imaging
Page 17
Chapter 1 Safety Notices
Pd2,1
Pd2,1
Pd3,2
Guidance and manufacturer's declaration - electromagnetic immunity
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
3 Vrms
3 V/m
, 80 MHz to 800 MHz
, 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d
is the recommended separation distance in
metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey
ance level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
, should be less than the compli-
b
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Q-Rad Radiographic System is used exceeds the applicable RF compliance
level above, the Q-Rad Radiographic System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Q-Rad Radiographic System.
b
Over the frequency range 150 kHz to 80 kHz, field strengths should be less than 3 V/m.
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging1-11
Page 18
Chapter 1 Safety Notices
Pd2,1
Pd2,1
Pd3,2
Recommended separation distances between
portable and mobile RF communications equipment and the
Q-Rad Radiographic System
The Q-Rad Radiographic System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Q-Rad Radiographic System QRad Radiographic System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Q-Rad Radiographic System as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power
of transmitter
W
0,010,120,120,23
0,10,380,380,73
11,21,22,3
103,83,87,3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz80 MHz to 800 MHz800 MHz to 2,5 GHz
Revision FQ-Rad System - Operator’s Manual
1-12Quantum Medical Imaging
Page 19
Chapter 1 Safety Notices
Abbreviation Definition
The following abbreviations and acronyms may be found in this document. Their
definition is explained below.
AECAutomatic Exposure Control
BtuBritish Thermal Unit
CRComputed Radiography
dDistance
DRDigital Radiography
HFHigh Frequency
hPaHecto-Pascal
hrHour
kgKilogram
kHzKilohertz
KVAKilovolt-Ampere
lbPound
mMeter
MHzMegahertz
mmMillimeter
OCPOperator Control Panel
PPower (max.)
RFRadio Frequency
VVolt
WWatt
WEEEWaste Electrical and Electronic Equipment
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging1-13
Page 20
Chapter 1 Safety Notices
THIS PAGE INTENTIONALLY LEFT BLANK
Revision FQ-Rad System - Operator’s Manual
1-14Quantum Medical Imaging
Page 21
Chapter
C
HAPTER
2: G
ENERAL
I
NFORMATION
2
2-1
Page 22
2-2
Page 23
Chapter 2 General Information
Instructions for Use
The instructions for using Quantum’s Q-Rad Radiographic System, including an explanation of the function
of controls, displays and signals, and the sequence of operation, are contained in each of the system
component user’s manual indicated in Tabl e 2- 1.
WARNING: This equipment must be operated with reasonable care.
Manufacturer’s equipment recommendations described in this manual mus t
be observed.
WARNING: Use a focal spot-to-film distance as large as possible in order to
keep the absorbed dose to the patient as low as reasonably achievable.
WARNING: During each exposure, the individual operating the X-ray
machine must exit the radiographic operatory and stand behind a leaded
wall.
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging2-3
Page 24
Chapter 2 General Information
Overview
This chapter describes the configurations of the Q-Rad Radiographic System, which typically includes
Quantum Medical Imaging HF Series X-ray Generators, radiographic tables, free-standing tube support,
upright cassette holder, collimator, X-ray tube, ion chamber (when equipped with AEC option), and image
receptor.
Detailed instructions for successful integration of the compatible Quantum components are published in
the Service Manuals for each of the various system components.
Table 2-1. Q-Rad System Components User/Service Documents
Ambient Temperature:-25 to +55 °C (-13 to 131 °F)
Relative Humidity:5–95 %
Atmospheric Pressure:500–1060 hPa
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging2-7
Page 28
Chapter 2 General Information
System Configuration
The Quantum Q-Rad System consists of a combination of integrated radiographic components, typically
including an X-ray generator, X-ray tube and tube support, collimator, radiographic table, and an upright
cassette holder. Different types of image receptors may be utilized with the Q-Rad System, including
computed radiography (CR) receptors, as well as conventional analog film cassettes.
Q-Rad Floor-Mounted System
The Q-Rad Floor-Mounted system is configured around either a free-standing,
floor-mounted tube support (Model QS-550) or a wall/ceiling mounted tube support (Model QS-500).
Because the free-standing version does not require wall attachment for support, it provides enhanced
freedom of movement in the patient vicinity for the operator.
Recumbent examinations are accomplished utilizing an elevating (Model QT-750) or the
non-elevating (Model QT-740 and QT-730) radiologic tables. Standing (weight-bearing) examinations
are performed using an upright, non-tilting (Model QW-420 Series) chest stand (cassette holder).
The X-ray generator is available in various power output and operator control panel types, with the
typical configuration including a 80 kW output generator with an Odyssey operator control.
STORED ENERGY
QG-40-SE, QG-50-SE
THREE-PHASE
QG-40-3, QG-50-3, QG-65-3, QG-80-3
GENERATOR
CONTROL PANEL
SUPPLEMENTAL
GENERATOR
CONTROL PANEL
(TECHVISION)
TABLEQT1QT-730, QT-740, QT-750
TUBE SUPPORTQS1QS-500, QS-550
WALL STANDQW1QW-420, QW-420-D, QW-420-S, QW-400
OCP1QUEST SERIES
QG-400, QG-500
ODYSSEY SERIES
QG-5000, QG-6500, QG-8000
Q-VISION SERIES
QGV-65, QGV-80
OCP2QG-TV-IP-F
: QG-20-SE, QG-32-SE,
: QG-32-2, QG-32-3, QG-40-2,
: QG-200, QG-250, QG-320,
: QG-3200, QG-4000,
: QGV-32, QGV-40, QGV-50,
X-RAY TUBETU1TOSHIBA
E7869X
VARI A N
DUNLEE
: E7239X, E7242X, E7252X, E7254X,
: RAD-14, RAD-60, RAD-92
: DA 1092, DA 1036
Revision FQ-Rad System - Operator’s Manual
2-8Quantum Medical Imaging
Page 29
Chapter 2 General Information
Table 2-2. Q-Rad Floor Mounted System
COMPONENTCOMPONENT
COMPONENT MODEL#
DESIGNATION
COLLIMATORQC1PROGENY
LINEAR II, LINEAR IV
RALCO
HUESTIS
DAPQD1VacuDAP-OEM DISPLAY + ION CHAMBER,
VacuDAP ION CHAMBER
ION CHAMBERIN1AID
IMAGE
RECEPTOR
RC1GRID CABINET
: ICX 1153
BUCKY CABINET
KLB10-08013
: MC150, MC150 PINNACLE,
: R 108/464 DHHS
: 150MC (MANUAL), CM 32211-QM1
: R50-17GC
: PROGENY KLB10-08011,
Q-Rad “DRX Series” Integrated Radiography System
The Q-Rad DRX Series Integrated Digital Radiography (DR) System combines the Q-Rad System’s HF
Series X-ray Generator and Carestream DRX-1 Digital Imaging Systems by way of a common user
interface (QG-DIG-DRX Software) running on the Carestream DRX-1 Digital Imaging System’s
Workstation computer.
When QG-DIG-DRX Software is implemented, all user interaction (except PREP and EXPOSE) with the
HF Series X-ray Generator is accomplished via the Digital Imaging System’s touchscreen computer,
thereby eliminating the need for a separate generator Operator Control Panel.
The Q-Rad DRX Series Integrated System consists of either a fixed DR Detector, and/or a Portable DR
Detector, and a Workstation Computer. The DR System is connected to the Quantum X-Ray Generator
by way of a software interface module.
Table 2-3. Q-Rad System, DRX Series Integration
COMPONENTCOMPONENT
DESIGNATION
GENERATORQG1SINGLE-PHASE
STORED ENERGY
QG-40-SE, QG-50-SE
THREE-PHASE
QG-40-3, QG-50-3, QG-65-3, QG-80-3
(PORTABLE, WIRELESS)
CARESTREAM DRX-1 IMAGING SYSTEM
(FIXED, 4343R)
Revision FQ-Rad System - Operator’s Manual
2-10Quantum Medical Imaging
Page 31
Chapter 2 General Information
Intended Use
The Quantum Q-Rad System provides diagnostic quality images to aid the physician with diagnosis. The
system can be used to perform radiographic exposures of the skeleton (including skull, spinal column, and
extremities), chest, abdomen, and other body parts. The Q-Rad Radiographic System is not indicated for
use in mammography.
General Application Specification
Patient Population:
Age:Newborn to geriatric
Weight:Up to 294.8 kg (650.0 lb)
Region of Body:Produces images of any bone in the human body,
including the hand, wrist, arm, foot, ankle, knee, leg, or
spine.
Operator Requirements:
Education/Training:This equipment is sold to be used exclusively under the
prescribed direction of a person who is licensed by law
to operate equipment of this nature.
Application:
Location:Professional use only, indoor use only.
Physical Environment:Refer to Q-Rad System operating and non-operating
specifications in this chapter.
Frequency of Use:Intermittent operation, limited primarily by loading
characteristics specific to tube type used in the system.
Mobility:Stationary
Cleaning, Preventive Inspection and Maintenance
Instructions for cleaning, preventive inspection and routine maintenance of Q-Rad Radiographic System,
are included in each of the system component user’s and service manuals indicated in Table 2- 1 .
Instructions for Disposal of Waste Equipment
In the European Union, this symbol indicates that when the last user wishes to
discard this product, it must be sent to appropriate facilities for recovery and
recycling. Contact your local representative for additional information on the
collection and recovery programs available for this product.
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging2-11
Page 32
Chapter 2 General Information
REGULATORY
LABEL
CONFIGURATION
LABEL
GENERATOR CABINET
GENERATOR
COMPLIANCE
LABEL
Labeling
Each component of the Q-Rad System has its own rating plate/serial label. The following table provides the
general location of the rating plate/serial label on each of the various types of components that could be
incorporated in a Q-Rad System.
Table 2-4. Q-Rad System Components Labeling Guide
Component TypeModel(s)Label Location
Q-Rad SystemQ-Rad
X-ray GeneratorSingle-phase,
non-Stored Energy
QG-25, QG-25-5,
QG-32, QG-32-5,
QG-40, and QG-40-5
Single-phase, Stored
Energy
QG-20-SE, QG-32-SE,
QG-40-SE, QG-50-SE
Three-phase
QG-32-2, QG-32-3,
QG-40-2, QG-40-3,
QG-50, QG-50-2,
QG-65, and QG-80
Revision FQ-Rad System - Operator’s Manual
2-12Quantum Medical Imaging
Page 33
Chapter 2 General Information
compliance label
Table 2-4. Q-Rad System Components Labeling Guide
Component TypeModel(s)Label Location
TechVision
Supplementary Gener-
ator Control Panel
Radiographic TableFixed Non-Float Tabletop
Radiographic TableNon-Elevating Tabletop
QG-TV-F,
QG-TV-IP-F,
QG-TV-C,
QG-TV-IP-C
QT-730
QT-740
Elevating Tabletop
QT-750
Tube Support,
Floor Mounted
QS-500
QS-550
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging2-13
Page 34
Chapter 2 General Information
Table 2-4. Q-Rad System Components Labeling Guide
Component TypeModel(s)Label Location
Radiographic Upright
Cassette Holder
Collimator, ManualPROGENY MC-150
QW-420,
QW-420-S, QW-420-D
QW-400
HUESTIS 150MC
RALCO R108/484 DHHS
Collimator, AutomaticPROGENY Linear II,
Linear IV
HUESTIS CM-32211-QM1
Revision FQ-Rad System - Operator’s Manual
2-14Quantum Medical Imaging
Page 35
Chapter 2 General Information
X-RAY TUBE
COMPLIANCE LABEL
LOCATIONS
REAR VIEW
COMPLIANCE
LABEL/DATAPLATE
CABINET (FRONT)
Table 2-4. Q-Rad System Components Labeling Guide
Component TypeModel(s)Label Location
X-ray tube
VARI A N:
RAD-14, RAD-60,
RAD-92
TOSHIBA:
DUNLEE:
Digital Image Detector
Cabinet
E7239X, E7242X,
E7252X, E7254X,
E7869X
DA 1092, DA 1036
Q-DRT
Q-DRT-T
Q-FDR
Q-FDR-T
Q-Rad System - Operator’s ManualRevision F
Quantum Medical Imaging2-15
Page 36
Chapter 2 General Information
REAR VIEW
COMPLIANCE
LABEL
Table 2-4. Q-Rad System Components Labeling Guide
Component TypeModel(s)Label Location
Integration OptionQ-Rad DRX Series
(QG-DIG-DRX)
Revision FQ-Rad System - Operator’s Manual
2-16Quantum Medical Imaging
Page 37
Chapter 2 General Information
Power Disconnection
Refer to the following table for Q-Rad System component power disconnection methods.
Table 2-5. DX-D 500
n
Q-Rad System Component Power Disconnection Method
Component TypeModelMeans of Power
Disconnect
X-ray Generator,
Single-phase, nonStored Energy
X-ray Generator,
Single-phase, Stored
Energy
X-ray Generator, Threephase
TechVision Supplementary Generator Control
Panel
Radiographic Table,
Non-Elevating Tabletop
Radiographic Table,
Non-Elevating Tabletop
QG-25, QG-25-5,
QG-32, QG-32-5,
QG-40, and QG-40-5
QG-20-SE, QG-32-SE,
QG-40-SE, QG-50-SE
QG-32-2, QG-32-3,
QG-40-2, QG-40-3,
QG-50, QG-50-2,
QG-65, and QG-80
Unplug Line Cord (ACS
Power Supply & Junction
Box)
ON/OFF circuit breaker
CB1 on generator cabinet
Unplug Line Cord
Carestream Digital
Imaging System
DRX-1Quick Reference Card for
the Carestream DRX-1
System, document 8G8318
Carestream DRX-1 System
Safety and Regulatory
Information with Hardware
User’s Guide, document
7H8166
Carestream DRX-1 System
Battery Charger User’s
Guide, document 8H6193
Detachable Parts and Accessories
Instructions for detachable parts and accessories of Q-Rad Radiographic System are included in each of
the system component user’s manual indicated in Tab le 2 -1 .
<End of Document>
Revision FQ-Rad System - Operator’s Manual
2-18Quantum Medical Imaging
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.