6-1. Replacing the battery
6-2. Fuse replacement
6-3.Cleaning and Decontamination
1. Introduction
This Meter is a handheld and battery operated Digital Multi Meter (DMM)
with multi function which designed to meet IEC-61010-1 CAT III 1000V over
voltage category and double insulation. The entire outer surface of the case has
been formed with thermo plastic elastomeric, giving the main body, though
downsized, high resistance against the shock of a drop. This user’s manual
covers information for safety and caution. Please read relevant information
carefully and observe all the warnings and note strictly. The DMM is a general
measurement tool and widely used in the school, laboratory, factory etc.
2. Safety note
War ning
To avoid possible electric shock or personal injury and to avoid possible damage
to the meter or to the equipment under test, adhere to the following rule:
z Do not apply more than the rated voltage of marked on the meter
z Do not apply voltage between COM and OHM terminal, in the
resistance measuring state.
z Do not measure current with test lead inserted into voltage or OHM
terminal.
z Do not expose the instrument to the direct sun light, extreme
temperature and humidity or dew full.
z Inspect the test lead for damaged insulation or exposed metal.
z Before measuring current, check the Meter’s fuses and turn off power
to the circuit before connecting the meter to the circuit.
z Disconnect circuit power and discharge all high voltage capacitors
before testing continuity, diode, resistance, capacitance or current.
“FUNC” key is the function select key that acts with
trigger. Use the key as switch of DC/AC voltage,
DC/AC current, Resistance/Continuity and / .℃℉
Press “HOLD” to enter and exit the hold mode in any
mode. That act with trigger.
This key is act with trigger. Press this key once, the
maximum value is holding (Will displays ‘MAX’
symbol in the LCD).
After pressing the key, A/D will keep working and the
display value will up dated and kept the maximum
value. (The actual gained value is not the peak value.).
This key is used to control Backlight and acts as trigger.
When press the key, will enable Backlight for 15 sec.
Press the key again within 15 sec, Backlight will
disable.
● Dimension (L×W×H) & Weight: 205 mm×102 mm×58mm, Approx. 390g.
Accessory: 1.K-type thermocouple probe x 1pcs, 2. Socket for temperature
measurement x 1pcs, Test Lead x 1set, Carrying bag x 1pcs
4-2. Electrical Specification (at 23±5℃; <75% RH)
4.2.1 DC Voltage
Range Resolution Accuracy
200mV 0.1mV
2V 0.001V
20V 0.01V
200V 0.1V
1000V 1V
* Input impedance: 10MΩ
* Over load protection: 250V DC/AC(rms) at 200mV Range.
* Max input voltage: 1000VDC
±(0.5% rdg + 2dgt)
4.2.2 AC Voltage
Range Resolution Accuracy
200mV 0.1mV
2V 0.001V
20V 0.01V
200V 0.1V
700V 1V
* Input impedance:10MΩ
* Over load protection: 250V DC/AC(rms) at 200mV Range.
* Frequency response: 40~400Hz
* Average rectifier type.
* Max input voltage:750V AC rms
±(0.8% rdg + 3dgt)
±(1.0% rdg + 5dgt)
4.2.3 Resistance
Range Resolution Accuracy
200Ω 0.1Ω
2kΩ 0.001kΩ
20kΩ 0.01kΩ
200kΩ 0.1kΩ
2MΩ 0.001MΩ
20MΩ 0.01MΩ
* Over load protection:250V DC/ AC rms
±(0.8% rdg + 3dgt)
±(1.5% rdg + 5dgt)
4.2.4 Diode check
Range Resolution Function
* Over load protection:250V DC/ AC rms
0.001V Will display the forward drop
voltage.
5 4
4.2.5 Continuity
Range Function
* Over load protection:250V DC/ AC rms
If measured resistance less than 30Ω, will buzzer is
sounded.
4.2.6 DC Current
Range Resolution Accuracy
2mA 1μA
20.00mA 0.01mA
200.0mA 0.1mA
10.00A 0.01A
* Over protection:use the fuse(F250mA/1KV) at mA range and use the
fuse(F10A/1KV) at 10A range.
* Max input current::200mA at ‘mA’ jack and 10A at ‘10A’ jack. .
(1.0% rdg + 3dgt)
(2.0% rdg + 3dgt)
4.2.7 AC Current
Range Resolution Accuracy
2mA 1μA
20mA 0.01mA
200.0mA 0.1mA
10.00A 0.01A
* Over protection use the fuse (F250mA/1KV) at mA range, and use the
fuse(F10A/1KV) at 10A range.
* Max input current::200mA at ‘mA’ jack and 10A at ‘10A’ jack. .
Range Resolution Accuracy
2nF 0.001nF
20nF 0.01nF
200nF 0.1nF
2µF 0.001µF
20uF 0.01µF
200µF 0.1µF
* Over load protection:use the fuse[F250mA/1KV]
(4.0%rdg +5dgt)
(5.0%rdg + 10dgt)
4.2.10 hFE
Range Function
hFE
* Over load protection:use the fuse [F250mA/1KV ]
hFE: 1~1000
4.2.11 Frequency
Range Resolution Accuracy
20kHz 1 Hz
* Sensitivity: 200mV -5V AC rms
* Over load protection:250V DC/AC rms
±(1.5%rdg+5dgt)
4.2.12 Inductance
Range Resolution Accuracy
2mH 0.001mH
20mH 0.01mH
200mH 0.1mH
2H 1mH
20H 10mH
* Over load protection:use the fuse [ F250mA/1KV]
(4.0% rdg + 8dgt)
(4.0%rdg + 3dgt)
(4.0%rdg + 3dgt)
(4.0%rdg + 3dgt)
(4.0%rdg + 5dgt)
5. Measurement operation
5-1 DC & AC voltage measurement
To avoid harms to you or damage to the meter from electric
shock. Please do not attempt to measure voltage higher than
DC 1000V/AC 700Vrms although readings may be obtained.
The DC voltage ranges are 200.0mV, 2.000V, 20.00V, 200.0V and 1000V.
The AC voltage ranges are 200.0mV, 2.000V, 20.00V, 200.0V and 750.0V.
To measure DC/AC voltage:
1. Insert the red test lead into the “VΩ” input terminal and the black test lead
into the COM terminal.
2. First, default mode is DC measurement mode, you can select the AC voltage
mode by press ‘FUNC“ key.
3. Set the rotary switch to proper DC/AC mV or V range.
4. Connecting the test lead across with the object be measured. The measured
value will be showed on the LCD display.
6. If the “OL” symbol is displayed on the LCD, please set the rotary switch to
the higher range.
Note:
In the range of 200mV in DC/AC and 2V AC, even in the absence of input or
connectivity test lead, there also have something display on LCD. In this case,
short-circuit "V" and "COM", the LCD display will back to zero.
5-2. Resistance measurement
To avoid damage to the Meter or to the equipment under
test, disconnect power and discharge all high-voltage
capacitors before resistance measurement.
The resistance range: 200Ω, 2.000kΩ, 20.00kΩΩ, 200.0k, 2.000MΩ, 20.00MΩ
To measure resistance, connect the meter as follows:
1. Insert the red test lead into the ”VΩ” terminal and the black test lead into the
COM terminal.
2. Set the rotary switch to proper resistance range.
3. Connect the test lead across with the object be measured. The measured value
will be display on the LCD
Note:
The test lead can add 0.1Ω to 0.2Ω of error to resistance measurement. To obtain
precision reading in low-resistance measurement, that is the range of 200.0Ω,
short the input terminal before measuring. Then the contact resistance will be
displayed on the LCD. You can subtract the contact resistance value from the
measured value.
z For high-resistance measurement (>10MΩ), it is normal taking several
second to obtain stable reading.
z The LCD display “OL” indicating open-circuit for the tested resistor or the
resistor value is higher than the maximum range of the meter.
5-3. Diode/Continuity check
To avoid damage to the Meter or to the equipment under
test, disconnect power and discharge all high-voltage
capacitors before Diode/Continuity check.
1. Insert the red test lead into the “VΩ” terminal and the black test lead into the
COM terminal.
2. Set the rotary switch to
3. For forward voltage drop reading on any semiconductor component, place the
red test lead on the component anode and place the black test lead on the
component cathode. The measured value show on the LCD.
4. Select the continuity check mode by the “FUNC” key.
The buzzer may be sound if the resistance of a circuit under test is less than
30Ω.
Ω position.
8 9
5-4. Capacitance measurement
To avoid damage to the Meter or to the equipment under
test, disconnect power and discharge all high-voltage
capacitors before measuring capacitance. You can use the
DC voltage function to confirm that the capacitor is
discharged.
Capacitance ranges are 2.000nF, 20.00nF, 200.0nF, 2.000uF, 20.00uF,
200.0uF.
To measure capacitance, connect the Meter as follows:
1. Set the rotary switch to “ F” proper position, according to about
capacitance value of capacitor under measured.
2. For user conveniently, there has a Multi Function Socket as accessory. First,
please insert the Socket to the “
and then insert the capacitor to the two holes on the Socket. The pin marked
“Cx/Lx/Temp” is pointed to the “” input terminal, and the pin marked
“COM” is pointed “COM” input terminal respectively. The measured value
will be displayed on the LCD.
Note:
z To measuring large capacitance of capacitor, stabilize the reading will take
some time.
z To increase the accuracy of capacitance measurement when measuring
under 5nF capacitance, you can Connecting 0.15uF Capacitor in parallel
with the capacitor be measured , or use relative measurement mode to
automatically subtract the capacitance build-in equalized capacitance and
residual capacitance of PCB lead wire from the result.
5-5. Transistor check
Do not attempt to apply DC/AC voltage more than 250V
between the two ‘COM’ and ‘hFE’ terminal to avoid damage
To check and measure transistor please use the Transistor Socket.
1. Set the rotary switch to the hFE position.
2. Use the special multi-function socket, and then it’s ‘hFE’ pin must be
3. Transistor( TO-92 package type) 3 pins must be correctly inserted to the
to the meter from electric shock.
inserted to the “hFE” input terminal and “COM” pin should be inserted to
the “COM” terminal.
‘e’,’b’,’c’ insert hole of the multi-function socket respectively according
to the NPN or PNP type. The hFE value will displayed on the LCD.
5-6. Frequency measurement
To avoid damage the Mete or electric shock the voltage
must less than 250V DC or AC rms between ”COM” and”
hFE”
1. Set the rotary switch to“20KHz”position.
2. Insert the red test lead into the “Hz” input terminal and the black test lead
into the “COM” terminal.
” input terminal and “COM” terminal,
3. Connect the test leads across with the circuit under testing. The
measured value shown on the LCD display.
Note: Input signal level must be higher than 0.2V (it is sensitivity).
5-7. DC/AC mA/10A measurement
When the voltage more than 250V between open-circuit
and ground, do not try to measure the current. If the
measurements were blown fuse, this may damage the
instrument or injury to yourself. In order to avoid damage
instrumentation or test equipment, before measurement
current, please check the instrument fuse. The input
socket, the functional profile and range should be use
correctly when measure current. When the test leader was
inserted at the current input socket, do not attempt to the
other side of the test leader connect any circuit.
DC Current range: 2.000mA, 20.00mA, 200.0mA, and 10A AC Current
range are2.000mA, 20.00mA, 200.0mA and 10A.
1. Turn off power to the circuit and discharge all high-voltage capacitors
before measuring capacitance, Set the rotary switch to the proper
DC/AC mA position.
2. Break the current path to be tested. Connect the red test lead to the more
positive side of the break and the black test lead to the more negative
side of the break.
3. Turn on power to the circuit the measured value show on the LCD.
4. If the “OL” symbol displayed on the LCD, please set the rotary switch
to the higher range.
5. When the currency between 200mA and 10A , insert the red test lead
into the input terminal marked as “10A”.
6. The measuring procedure is same as that of 1-4 section.
Note:
z For safety sake, the measuring time for high current should be less than
10 second for each measurement and the interval time between two
measurements should be greater than 5 minutes.
z When current measurement has been completed, disconnect the
connection between the testing lead and the circuit under test.
5-8. Inductance measurement
The inductance range is 2.000mH, 20.00mH, 200.0mH, 2.000H and 20.00H.
To measure inductance, connect the meter as follows:
1. Insert the red test lead into the ”Lx” terminal and the black test lead into the
COM terminal.
2. Set the rotary switch to proper inductance range.
Connect the test lead across with the inductor being measured. The
inductance value will be displayed on the LCD.
10
5-9. Temperature measurement
To measure temperature, please use the Multi-Function Socket.
1. Set the rotary switch to the “TEMP” range, then the “OL” displayed on the
LCD.
2. Insert the Socket to the two “COM” and “TEMP” terminals.
3. Insert the K-type temperature probe to the two insert holes of Socket. User
should take note of it’s polarity.
4. The measured temperature value will be displayed on the LCD.
Note:
The two “Temp” & “COM” pins of Socket must be direct at the two “TEMP”
& “COM” terminal respectively.
6. Maintenance
6-1. Battery replacement
When meter display the battery must be replace to maintain normal
operation.
1. Disconnect and remove all test probes from any live source and meter.
2. Open the battery cover on the bottom case by screwdriver.
3. Remove old battery and snap new one into battery holder
6-2. Fuse replacement
Replacing the defective fuse should the done according to the following
procedure.
1. Remove the test lead and turn off power.
2. Open the battery cover on the bottom.
3. Remove the defective fuse and insert a new fuse of the same size and
rating.
4. Replace the bottom case and reinstall all the screw.
6-3.Cleaning and Decontamination
The meter can be cleaned with soft clean cloth to remove any oil, grease or
grim. Do not use liquid solvent or detergent.
11
PROKIT’S INDUSTRIES CO., LTD
Certificate Number: TW98/12323QA
http//:www.prokits.com.tw
e-mail: pk@mail.prokits.com.tw
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.