Prokit's Industries MT-1232 User Manual

MT-1232
3-3/4 Autorange Digital Multimeter
User’s Manual
1st Edition, 2013
©2013 Copyright by Prokit’s Industries Co., Ltd.
Contents
General------------------------------------------------------------------1 Open-Package Inspection------------------------------------------2 Safety Note -------------------------------------------------------------2 Instrument Panel& Button Function Description---------------5 Other Functions--------------------------------------------------------6 Properties----------------------------------------------------------------6 Instrument Maintenance--------------------------------------------16 Fault Elimination------------------------------------------------------17
General
This product equipped with the LCD display of text height 15mm, is a 3 3/4 digital multimeter which has the merits of clear reading, stable performance and high reliability. It could be used to measure DC voltage, AC voltage, DC current, AC current, resistance, capacity, frequency/duty cycle, diode and make on-and-off test. Meanwhile, it is available for unit symbol display, automatic/manual range switching, automatic power off and alarm function. The multimeter, due to its complete functions, performs with high measurement accuracy and convenient operation, troubleshooting
and is ideal for general electrical testing and
1
Open-package Inspection
Open the package box and take out the meter, check carefully if the following accessories are missing or damaged. If there is anything missing or damaged, please contact the distributor immediately. Digital multimeter 1 PC User’s manual 1 copy Test leads 1 pair Temperature Probe (K-Thermocouple) 1 PC Test Socket 1PC
Safety Note
The design of this meter is in accordance with IEC1010 clause (the safety standard issued by International Electrotechnical Commission). Prior to the operation of the instrument, please read the safety considerations before use.
1. When DC voltage is above 30V, AC voltage above 25V, current above 10mA, AC power line with inductive load or power line during electric fluctuation is measured, please beware of electric shock.
2. Prior to measurement, check if the measurement function switch is at the correct position. Check if the test lead is contacted reliably, connected correctly, and grounded well etc. in order to avoid electric shock.
3. Only if the meter is used with the matched test lead, does it meet the requirement of the safety standard. When the line of the test lead is damaged, it is necessary to replace with another one of the same model or the same electrical specification.
4. Don’t use other unconfirmed or disapproved fuses to replace the fuse inside the meter. Only the same model or same
2
specification fuse can be replaced. Before the replacement, the test leads must be removed from the measuring point and ensure there is no any signal at the input terminal.
5. Don’t use other unconfirmed or disapproved battery to replace the battery inside the meter. Only the same model or same electrical specification battery can be replaced. Before the replacement, the test lead must be removed from the measuring point and ensure there is no any signal at the input terminal.
6. When the electrical measurement is made, never let your body gets in touch with the ground directly, and don’t touch uncovered metal terminal, output port, lead clamp and etc. where earth potential may exist. Dry clothes, rubber shoes, rubber cushion and other insulating material are usually used to keep your body insulated against the ground.
7. Don’t store and use it in the high-temperature, high-humidity, inflammable and strong magnetic field environment.
8. It may damage to the meter and endanger the operator’s safety if the voltage value beyond the permitted ultimate voltage value is measured. The ultimate voltage value permitted for measurement is marked on the instrument panel, and never measure the value exceeding the standard. Don’t input the ultimate value out of regulation in order to avoid electric shock and the damage to the meter.
9. When the test lead is inserted into the current socket, don’t measure any voltage for fear that the meter should be damaged and the operator’s safety be endangered.
10. Don’t try calibrating or repairing the meter. When necessary, only the qualified professional personnel who have had special training or gained approval can make it.
11. During measurement, the requirement of measurement
3
function must be in accordance with LCD display. Please be sure to disconnect the line of the test lead with the measured object first and ensure there is no any input signal. It is forbidden to switch the function/range selection switch during measurement
12. When “ battery immediately to ensure the measurement precision.
13. It is not allowed to insert the test lead into the current terminal to measure voltage!
14. Please don’t change the circuits of the meter freely for fear that the meter will be damaged and the safety endangered.
15. Description of Safety Symbols
” is shown on LCD display, please replace
Warning!
Insulation
High Voltage! Danger!
Ground
Double
Low
Battery
In accordance with the
instruction of European
Trade Union
DCV
ACV
DCA
ACA
Fuse
4
Instrument Panel & Function Button Description
1. Instrument model number
2. LCD Display: Displays the measured data and unit.
3. Function Button
3.1 Hz/% (Frequency/Duty Cycle) Press this button to select the frequency or duty cycle mode. The measurement mode of voltage/frequency/duty cycle or current/frequency/duty cycle could be selected by pressing this button in AC/DC voltage or AC/DC current .
3.2 REL (Relative Value
Measurement): The relative value measurement of Capacitance function could be conducted by pressing this button
3.3 D/¤ :( Data hold/ Backlight): Press the left side of this
button, the reading is locked; press it again (the left side), the lock will release and back to normal measurement status. Press the right side of this button over 3 seconds, the backlight will come on, press this button(the right side) again the backlight turns off.
3.4 S: (Function Switch): Press this button, the function could
switch between DC/AC and
4. Knob Switch: Used to change the measurement function and range.
5. Input Terminals
5.1. Current, Voltage, Diode, Resistance, Capacitance,
Frequency, Buzzer, Temperature “-” Input terminal.
5
5.2.10A“+”input terminal.
5.3. Voltage, Diode, Resistance, Capacitance, Frequency,
Buzzer, Temperature and“+” Input terminal with current less than 400mA.
Other Functions Auto Power off
During measurement, the meter will automatically shut down (enter sleeping mode) to save power if function buttons and knob switch are not operated for 15 minutes. In auto power off mode, press any function buttons or rotate the knob switch, the instrument will get into the auto power on mode (working mode)
Properties General Feature
1-1 Display: LCD 1-2 Max Display: 3999(3 3/4) counts automatic polarity display
and unit display 1-3 Measuring Method: Dual integral A/D converter 1-4 Sampling Range: Approx. 3 times / sec. 1-5 Over Range Indication: Display “OL” 1-6 Low Battery Indication: “ 1-7 Operation Environment: (0~40)˚C Relative Humidity:<80% 1-8 Storage Environment: (0~50)˚C Relative Humidity:<80% 1-9 Power: 2 pcs 1.5V batteries (“AA” battery) 1-10 Dimension (size): 147×78×41mm 1-11 Weight: Approx. 183g 1-12 Accessories: User’s Manual (1 pc), color box (1 pc), 10A
test leads (1 pair), K-Thermocouple, Test Socket(1 pc)
” symbol appears;
6
Technical Features
2-1. Accuracy: ± (a%× reading +digits),at (23±5)˚C, relative
humidity<75%. One year calibration guarantee from the
time dispatched from the factory. 2-2. Technical Specification
2-2-1. DCV
A) Turn the knob switch to “ B) The default state of the meter is automatic range status,
which shows “AUTO” symbol
C) Put the test lead in contact with the testing point. The voltage
and polarity of the point where the red pen is contacted will be displayed on the screen.
Caution:
1. Don’t measure voltage over 600V. Otherwise, there is a danger
of the meter being damaged.
2. When measuring high voltage, special attention should be
given to personal safety and avoid your body getting in touch with high voltage circuit.
Range Accuracy Resolution 400mV 100uV 4V 1mV 40V 10mV 400V 600V ±(1.0%+4d) 1V
Input Impedance: 400m >40M; 10M at other Ranges.
Overload Protection: 600V DC or 600V AC Peak Value.
±(0.5%+4d)
“Range
100mV
7
2-2-2. ACV
A) Insert black test lead into the hole of “COM” and red test
lead into “ B) Rotate function switch to “
select the AC measurement mode. C) The default state of the meter is in automatic range status,
which shows “AUTO” symbol D) Put the test lead in contact with the testing point. The voltage
of the point where the red test lead is contacted will be
displayed on the screen.
Caution:
1. Don’t measure voltage over 600V. ; Otherwise, the meter will
be damaged.
2. When measuring high voltage, special attention should be
given to personal safety and avoid your body getting in touch
with high voltage circuit.
Range Accuracy Resolution 4V 1mV 40V 10mV 400V 600V ±(1.0%+6d) 1V
Input Impedance: >10M
Overload Protection: 600V DC or 600V AC Peak Value
Frequency Response: (50~200) Hz
Display: Average value response (RMS of sine wave).
“, press “SELECT” button to
±(0.8%+6d)
100mV
8
2-2-3. DCA
A) Insert the black test lead into the “COM” input terminal and
red test lead into the “
or 10A input terminal (Max 10A). B) Rotate function switch to Current. The default state of the
meter is in automatic range status, which shows “DC”
symbol. Then connect the test lead to the tested circuit in
series, the tested current value and the current polarity of
the point where the red pen is contacted will be displayed on
the screen simultaneously.
Caution:
1. If “OL” is displayed on LCD, it indicates the tested current
value has exceeded the present range limit, please select
higher range to complete the measurement.
2. The Max input value is 400mA or 10A. (Depending on the
terminal where the red test lead is contacted)
Range Accuracy Resolution 400uA 0.1uA 4000uA 1uA 40mA 10uA 400mA 10A ±(1.2%+10d) 10mA
Max measurement voltage drop: Full Range mA is 0.4V, A is
100mV;
Max input current: 10A (less than 15 seconds);
Overload Protection: 0.4A/250V restorable fuse, 10A/250V
fuse.
±(1.0%+10d)
“input terminal. (Max 400mA),
100uA
9