Prokit's Industries MT-1132 User Manual

MT-1132
3-1/2 Digital Multimeter
User’s Manual
1st Edition, 2012
©2012 Copyright by Prokit’s Industries Co., Ltd.
SAFETY INFORMATION
This multimeter has been designed according to IEC 1010 concerning electronic measuring instruments with an overvoltage category (CAT III) and pollution 2. Follow all safety and operating instructions to ensure that the meter is used safely and kept in good operating condition. Full compliance with safety standards can be guaranteed only with test leads supplied. If necessary, they must be replaced with the type specified in this manual.
SAFETY SYMBOLS
Important safety information, refer to the operating manual.
Dangerous voltage may be present. Earth ground
Double insulation (Protection class II).
Fuse must be replaced with rating specified in the manual.
MAINTENANCE
Before opening the case, always disconnect test leads from all
energized circuits.
For continue protection against fire; replace fuse only with the
specified voltage and current ratings: F10A/250V (Quick Acting)
Never use the meter unless the back cover is in place and
fastened completely
Do not use abrasives or solvents on the meter. To clean it
using a damp cloth and mild detergent only.
1
DURING USE
Never exceed the protection limit values indicated in
specifications for each range of measurement.
When the meter is linked to measurement circuit, do not touch
unused terminals.
Never use the meter to measure voltages that might exceed
500V above earth ground in category III installations.
When the value scale to be measured is unknown, set the
range selector at the highest position.
Before rotating the function / range selector to change
functions, disconnect test leads from the circuit under test.
When carrying out measurements on TV or switching power
circuits always remember that there may be high amplitude voltages pulses at test points, which can damage the meter.
Always careful when working with voltages above 60V DC or
30V AC rms. Keep fingers behind the probe barriers while measuring.
Before attempting to insert transistors for testing, always be
sure that test leads have been disconnected from any measurement circuits.
Components should not be connected to the hFE socket when
making voltage measurements with test leads.
Never perform resistance measurements on live circuits.
GENERAL DESCRIPTION
The meter is a handheld 3 1/2 digital multimeter for measuring DC and AC voltage, DC current, Resistance, Diode, Transistor and Continuity Test with battery operated.
2
FRONT PANEL DESCRIPTION
Display
1. 3 1/2 digit, 7 segment, 15mm high LCD.
2. Backlight button
3. Function/range selector Rotate this selector to select function and desired range
4. Data hold button Press this button, the display will keep the last reading and "
" symbol will appear on the LCD until pushing it again.
5. "10A" jack
6. “COM" jack
7. Voltage, resistance, current (except 10A) measurements, Continuity check and Diode test.
8. Power on/off button
9. LED test socket
3
SPECIFICATIONS
Accuracy is specified for a period of one year after calibration and at 18°C to 28°C(64°F to 82°F) with relative humidity to 80%.
GENERAL
Maximum voltage between : CAT III 600V Terminals and earth ground Fuse protection : F 10A/250V Power : 1.5V, AAx2 Display : LCD, 1999 counts, updates 2-3/sec. Measuring method : Dual-slope integration A/D converter Over range indication : Only figure "1" on the display Polarity indication : "-" displayed for negative polarity Operating environment : 0 to 40ºC Storage temperature : -10ºC to 50ºC Low battery indication : " Size : 147mm x 78mm x 41mm Weight : Approx. 183g
DC VOLTAGE
Range Resolution Accuracy 200mV 100uV ±(0.5% of rdg + 2 digits) 2V 1mV ±(0.5% of rdg + 2 digits) 20V 10mV ±(0.5% of rdg + 2 digits) 200V 100mV ±(0.5% of rdg + 2 digits) 500V 1V ±(0.8% of rdg + 5 digits) Overload Protection: 250V rms. For 200mV range and 500V DC or rms. AC for other ranges.
" appears on the display
4
DC CURRENT
Range Resolution Accuracy 200uA 0.1uA ±(1.0% of rdg + 5 digits) 2mA 1uA ±(1.0% of rdg + 5 digits) 20mA 10uA ±(1.0% of rdg + 5 digits) 200mA 100uA ±(1.0% of rdg + 5 digits) 10A 10mA ±(2.5% of rdg + 5 digits)
Overload Protection: 0.2A/250V restorable fuse.
10A/250V fuse
AC VOLTAGE
Range Resolution Accuracy 200V 100mV ±(1.2% of rdg+10digits)
500V 1V ±(1.2% of rdg+10 digits) Overload Protection: 500V DC or rms. AC for all ranges. Frequency range: 40Hz to 400Hz. Response: Average responding, calibrated in rms. of a sine wave.
DIODE & CONTINUITY
Range Description
If continuity exists (about less than 50Ω), built-in buzzer will sound Show the approx. forward voltage drop of the diode
Overload Protection: 250V DC or rms. AC.
5