Top Cover
Front Cover
Center Beam Cover
Back Cover Removal
Left and Right Side Cover Removal
Left and Right Side Bottom Cover
Left and Right Bottom Cover
Left and Right Side Cover Assembly
Back Cover
Center Beam Cover
Front Cover
Upper Arm Cover
Rear Flat Belt Terminal Cover
Upper Arm Cam Cover
Top Cover
Warning: This service manual is for use by Precor trained service
providers only.
If you are not a Precor Trained Servicer, you must not attempt to
service any Precor Product; Call your dealer for service.
This document contains information required to perform the majority of
troubleshooting, and replacement procedures required to repair and
maintain this product.
This document contains general product information, software diagnostic
procedures (when available), preventative maintenance procedures,
inspection and adjustment procedures, troubleshooting procedures,
replacement procedures and electrical block and wiring diagrams.
Acrobat Reader Users:
To move directly to a procedure, click the appropriate procedure in the
bookmark section to the left of this page. You may drag the separator
bar between this page and the bookmark section to change the size of
the page being viewed.
The generator performs three functions in the AMT. First, by
controlling the amount of electrical current applied from the generator
to a load circuit, the user’s pedaling resistance is controlled.
Second, the generator is used to charge the AMT’s internal battery.
Lastly, one of the generators three phase output windings is monitored
to determine when the unit is in use and when it is idle. This system
also determines the step rate by determining the operating speed
(output frequency) of the monitored generator winding. There is also a
magnetic brake that applies resistance to horizontal motion and a
sensor that measures horizontal motion (stride length).
Warning and Caution Statements and General Safety
Guidelines
Warning statements indicate a particularly dangerous activity. Warning
statements you will find in this manual include:
♦
Because this is a self powered unit, it will either be necessary to
either equip the unit with the optional external power supply or
have an assistant pedal on the unit while voltage measurements are
being taken. Because of the danger of working on the unit while it
is in motion using the optional external power supply is strongly
recommended.
♦
When the unit is used, stairarms are in motion; the generator will
operate and produce potentially hazardous voltages even when the
battery is disconnected.
♦
Caution statements are intended to prevent damage to the AMT as a
result of the current activity. Caution statements included in this
manual are listed below:
•
When it is necessary to lift or move the AMT, ensure that the AMT
has adequate support and that you use proper lifting techniques.
When the rear platform is removed, the AMT may be lifted from the
rear and moved like a wheelbarrow.
•
To remove power from the AMT, the optional external power supply
(when equipped) must be disconnected from the wall outlet and the
red (positive) wire must be disconnected from the battery. Always
ensure that the AMT external power supply is unplugged from the
wall outlet and the red (positive) wire is removed from the
battery when you inspect or adjust the AMT, or when you isolate,
remove, or replace an AMT component.
•
Removing the covers exposes high voltage components and
potentially dangerous machinery. Exercise extreme caution when
you perform maintenance procedures with the cover(s) removed.
During service operations you will be very close to moving
machinery and voltage bearing components. When you perform
maintenance procedures with the covers removed, remove jewelry
(especially from ears and neck), tie up long hair, remove neck
ties, and do not wear loose clothing.
•
Exercise caution when touching any wire or electrical component
during AMT operation.
•
A pinching hazard exists when the unit is operated. It is
possible to seriously pinch a finger. The AMT can be mechanically
locked by inserting a screwdriver through the primary sheave and
frame. See the illustration below. Locking the AMT will prevent
accidental pinching.
A pinching hazard exists when the unit is operated. It is possible to
seriously pinch a finger. The AMT can be made immobile by lift the
stairarm pedals and removing the tension from stairarm flat belts.
Removing the tension from the flat belts will prevent accidental
pinching.
Safety
Safety guidelines you should know and follow include:
Read the owner’s manual and follow all operating instructions.
♦
Operate the AMT on a solid, level surface with the unit properly
leveled. The AMT is properly leveled when all five feet are in
contact with the floor. Visually check the AMT before beginning
service or maintenance operations. If it is not completely assembled
or is damaged in any way, exercise extreme caution while operating
and checking the AMT.
♦
When operating the AMT, do not wear loose clothing. Do not wear
shoes with heels or leather soles. Check the soles of your shoes and
remove any embedded stones. Tie long hair back.
♦
Do not rock the unit. Do not stand or climb on the handlebars,
display enclosure, or cover.
♦
Do not set anything on the handlebars, display enclosure, or cover.
Never place liquids on any part of the AMT, while performing
service.
♦
To prevent electrical shock, keep all electrical components away
from water and other liquids.
♦
Do not use accessory attachments that are not recommended by the
manufacturer-such attachments might cause injuries.
♦
AMT’s equipped with Cardio Theater PVS units will have external
power supply and coaxial cable routed through the bottom of the unit
to the top of the display console. Cord management must be
maintained.
General Information
For the latest exploded view, part number and part pricing information,
visit the Precor dealer website at www.precor.com/connection.
The P30 Console diagnostic software consists of the following modes:
♦
Display Test
♦
Keyboard Test
♦
Heart Rate Test
♦
Brake Test
♦
RPM Test
♦
Battery Test
♦
Stride Position Test
Procedure:
1 The P30 Console uses the standard access codes to provide access to
the various software features. Use the PAUSE key and the numeric
keypad to enter the access code. The standard access codes use all
sequential key presses. The allowable delay between key presses is
short. If too much time is taken between key presses or the wrong
key is pressed the access procedure will be aborted. If the access
is aborted, it will be necessary to start over from the beginning.
Figure 2: P30 Software Access
2 See P30 Diagnostic Software Menu Table (on page 14).
Press and hold the
CLEAR key for 5
seconds to return t
Hardware
Hardware Validation will scroll across the
display followed by DISPLAY TEST.
The upper most group of LED’s will illuminate on the
display. Check the display to ensure that all LED
To display the remaining LED groups. Check each
display group to ensure that all LED segments are
illuminated.
A representation of all of the keys on the console will
be displayed. Pressing a key on the console will
cause the illuminated representation of that key to
turn off. Press all of the keys on the console to
ensure that all of the keys are functioning.
Grasp both of the heart rate grips on the handlebar,
after a couple of seconds the heart rate will be
displayed in the heart rate and smart rate displays.
Heart Rate
Test
Press the CLEAR
key to return to
Hardware
Validation
Use chest strap transmitter or a test transmitter to
test the wireless heart rate function, after a couple of
seconds the heart rate will be displayed in the heart
The following tests will be will be available. Use the
and keys to toggle between tests. Press the OK
key to select.
Belt Speed (Treadmills Only)
Brake (Non-Treadmills Only)
Stride Height (AMT12 Open Stride Only)
SPM (Strides Per Minute - Elliptical Only)
RPM (Bikes Only)
SPM (Steps Per Minute - Climbers Only)
Incline/ Crossramp
Battery (Non-treadmills and Self Powered
Only)
Stride Position (AMT12, Only)
Stride Count Reset (AMT12)
The speed of the running belt is controlled by the
SPEED keys during this test.
The following values are displayed; time (that the
running belt has operated during this test), current (i
amps through the drive motor), belt speed (in MPH)
and power bits (PWM value).
This test is valuable in diagnosing treadmill load
Belt Speed
Test (Treadmill
only)
Press the CLEAR
key to return to
Hardware
Validation
Brake Test
(Self Powered
Units Only)
Press the CLEAR
key to return to
Hardware
conditions such as running bed and/or running belt
condition. As load is increased, the speed is kept at
constant value (as set via the SPEED keys) and the
current and power bits will vary proportionately to the
load.
It is suggested that you perform this test on a
treadmill known to be in excellent operating
condition. Note the values displayed at a fixed speed
(such as 3 MPH) and your body weight. Those
values may then be used to benchmark the operati
The console will display the power bits (PWRB).
Changing the resistance , keys will change the
power bit settings.
SPM/ RPM
(Strides/
Revolutions
Per Minute
NonTreadmills
only)
Incline/
Crossramp
Press the CLEAR
key to return to
Hardware
Validation
Press the CLEAR
key to return to
Machine Test
The console will display the Pulse. The pulse is
showing a count of zero crossings from the
generator. RPM shows a software averaged version
of pulse.
Use , keys to view the following RPM readings.
PULSE
MIN
Two numbers will be displayed, the percent of lift and
A/D number related to the current lift position. The
A/D (analog to digital) number represents the lift
potentiometer reading. The A/ D number will be in
the range of 0 to 255 or 0 to 32769.
The lift may be operated using the
INCLINE/CROSSRAMP keys. The incline
percentage and A/D numbers will change
accordingly. Care must be taken while operating the
lift in this mode because the normal upper and lower
range limits are not in effect. It is possible to operate
the lift to its physical limits and may cause the lift
The console will display the battery test. Battery test
Battery Test
Press the OK key
Press the CLEAR
key to return to
(NonTreadmills)
Stride Position
Test
AMT12 (Open
Stride Only)
Hardware
Press the CLEAR
key to return to
Hardware
Press the CLEAR
STRIDE POSITION TEST will scroll across the
display.
The low, high and actual stride positions will be
displayed. The low display shows the lowest position
reading achieved during the test. The high display
shows the highest position reading achieved during
the test. The actual display changes with the current
Note: If power is lost or the unit
is no longer being pedaled
during the rest process, the
update process will be
interrupted, displaying a error 70
(Belt replacement, LPCA not
read). If a error is displayed, the
Stride Count Reset process
must be repeated.
Incline/Crossr
amp
Press the key
Press the CLEAR
key to return to
Hardware
Validation
Press the key
Two numbers will be displayed, the percent of lift and
A/D number related to the current lift position. The
A/D (analog to digital) number represents the lift
potentiometer reading. The A/ D number will be in
the range of 0 to 255 or 0 to 32769.
The lift may be operated using the
INCLINE/CROSSRAMP keys. The incline
percentage and A/D numbers will change
accordingly. Care must be taken while operating the
lift in this mode because the normal upper and lower
range limits are not in effect. It is possible to operate
the lift to its physical limits and may cause the lift
User Detected or User Not Detected will be
Auto Stop
(Treadmill
Only)
Press the CLEAR
key to return to
Hardware
A magnet is attached to the running deck and a
sensor is attached to the frame. If the sensor detects
motion from the running deck, User Detected will be
displayed. If the sensor does not detect motion from
the running deck, User Not Detected will be
The information display will access the following data:
♦
Odometer
♦
Hour Meter
♦
U-Boot Software
♦
U-Base Software
♦
Lower Software
♦
Metrics Board Software
♦
Stride Dial Software
♦
Serial Number
♦
Usage log
♦
Error Log
Procedure
1 The AMT 885, 835, 833, and 832 uses the standard access codes to
provide access to the various software features. Use the PAUSE key
and the numeric keypad to enter the access code. The standard access
codes use all sequential key presses. The allowable delay between
key presses are short. . If too much time is taken between
key
presses or the wrong key is pressed the access procedure will be
aborted. If the access is aborted, it will be necessary to start
over from the beginning.
2 See, P30 Information Table (see "P30 Information Table" on page 20).
The odometer will display HORIZONTAL distance
in miles or KM depending on Club parameter
settings and VERTICAL distance in steps.
Note: The odometer data is
stored in non-volatile memory
on the upper PCA. If the upper
PCA is replaced the odometer
data will be lost.
The operating time of the unit will be displayed as
12345 HOURS. The operating time is defined as
total amount of time that the unit has operated in
program modes. The hour meter is also used to
The quantity of errors in the log will be displayed.
Press the OK key.
The most recent error will be displayed first.
e
hour meter when the error occurred.
display.
exit the DIAGS-
Usage log
Use the , keys to move
Press the OK key.
Use the , keys to move
Press the CLEAR key to
exit the USAGELOG
Use the , keys to move
A message will scroll describing the program, the
number of times and the number of minutes the
Error Log
Use the , keys to move
through the list of errors.
If you wish to clear the error
log, press and hold the
QUICK START key.
The error messages will list the error name, the
odometer reading when the error occurred and th
The message HOLD TO CLEAR ERRORS will be
displayed. The error log will be cleared when the
message NO ERRORS is displayed.
Please note that the ERROR
LOG may also be accessed at
any time by pressing and
holding the CLEAR key for
four seconds. If the error log
does not contain any errors,
the message STUCK KEY will
This procedure allows you to change the following club settings:
♦
Select Language
♦
Select Units
♦
Set Max Workout Time
♦
Set Max Pause Time
♦
Set Cool Down Time
Procedure1 The AMT 885, 835, 833, and 832 uses the standard access codes to
provide access to the various software features. Use the PAUSE key
and the numeric keypad to enter the access code. The standard access
codes use all sequential key presses. The allowable delay between
key presses are short. . If too much time is taken between key
presses or the wrong key is pressed the access procedure will be
aborted. If the access is aborted, it will be necessary to start
over from the beginning.
2 See P30 Setting Club Parameters Table (on page 24).
P30 Setting Club Parameters Table
Club
Settings
With the banner scrolling,
press keys
PAUSE,5,6,5,1,5,6,5,
DIAGS-SET CLUB PARAMETERS will scroll
across the display followed by select
LANGUAGE.
A list of available languages will be displayed,
Set
Language
Use the , keys to toggle
between the available
languages.
Note: When the default
language is changed to
anything other than English,
the submenus will remain in
English.
Set Units
Set Max
Workout
Time
Press the OK key to select
desired language and exit to
Use the , keys to toggle
Press the OK key to select
desired unit exit to Club
Use the , keys to select
the maximum time a user
can remain in a program.
Press the OK key to save
selection and exit to Club
UNITS will be displayed.
U.S (miles per hour) and METRIC (kilometers
SET MAX WORKOUT TIME will be displayed.
As the , keys are pressed, the displayed
value increments or decrements between 1 and
240 minutes, and NO LIMIT. If an arrow key is
held down then the value changes slowly at first
SET MAX PAUSE TIME will be displayed.
Set Max
Pause Time
Use the , keys to select
the maximum time the unit
will remain in paused before
resetting.
As the , keys are pressed, the displayed
value increments or decrements between 1 and
120 seconds. If an arrow key is held down then
the value changes slowly at first and then at an
As the , keys are pressed, the displayed
value increments or decrements between 0 and 5
Set Cool
Use the , keys to select
the cool down time.
Down Time
Press the OK key to save
selection and exit to Club
SET METRIC DEFAULTS will be displayed.
A list of available default settings will be
displayed, including:
OFF
% COMPLETE
TIME IN ZONE
Use the , keys to select
Set Metric
Defaults
the default settings that will
be displayed during a
workout.
Set
Crossramp
Auto-Level
Press the OK key to save
selection and exit to Club
Use the , keys to select
height the crossramp will
return to when after a
workout.
Press the OK key to save
selection and exit to Club
SEGMENT TIME REMAINING
AVERAGE SPEED
CALS/MIN
CALS/HR
WATTS
METS
TARGET HR
AVE HEART RATE
CROSSRAMP AUTO-LEVEL will be displayed.
As the , keys are pressed, the displayed
value increments or decrements between
XRAMP 1 and 120 seconds. If an arrow key is
held down then the value changes slowly at first
Use the , keys to select
the either Show Programs or
Hide Programs.
Press the OK key to save
selection and exit to Club
Pressing the , keys
changes the active column.
Pressing the corresponding
, resistance keys
changes the resistance
Press the OK key to save
program and exit to Club
Press the PAUSE key exit
will be displayed, including:
SHOW PROGRAMS
SET CUSTOM PROGRAMS will be displayed.
When this function is initiated, the currently
stored custom program (#1 or #2) is displayed in
the matrix. The first column is active; all the
LEDs in that column blink. The resistance value
for that column is displayed in the message bar.
Documenting P30 Software Problems
When a problem is found with the software in the upper or lower PCA,
record the information listed below.
When a problem occurs, record the following information:
The action taken by the user just before the problem occurred.
♦
Problem-related information (such as how far into the program the
problem occurred, the work level being used when the problem
occurred, error code displayed, etc.).
♦
The frequency of occurrence.
Replacing the Upper PCA
1 Set the treadmill circuit breaker in the off position and unplug the
treadmill’s line cord from the AC outlet. (TRM only)
2 The PCA’s in the console are static sensitive. They can be damaged
if proper static prevention equipment is not used. Attach an antistatic wrist strap to your arm, and then connect the ground lead of
the wrist strap to the treadmill’s frame ground.
3 Remove the two screws that fasten the dash transition cover and
remove the cover.
Figure 3: Dash Transition Cover – (TRM only)
4 Remove the four screws that fasten the console to the dash mounting
7 Remove the Heart Rate PCA. See Replacing the P30 HR PCA (on page 37).
8 Disconnect the Heart Rate Cable from the Heart Rate PCA.
9 Disconnect the Data cable, Auto Stop cable (Treadmill only), Stop
Key cable (Treadmill only), Ground wire and the CSAFE cable (if in
use) from the Upper PCA. Remove the console from the maintenance
access hook and place it on a flat work surface.
14 Install the replacement upper PCA into the console using the five
screws removed in step 14. Torque to 10 inch pounds.
Note: Do not use an electric screw driver or over tighten the screws. Over
tightening may damage the console. This type of damage is not covered
under warranty.
15 The white wire in the Metrics cable and the yellow wire in the
Machine Controls and HR cables indicate pin 1. Align the white wire
in the Metrics cable and yellow wires in the Machine Controls and HR
cables with the pin 1 markings on the upper PCA.
Note: If pin 1 is not marked on the Upper PCA refer to the pin 1 reference
below.
Note: Do not use an electric screw driver or over tighten the screws. Over
tightening may damage the console. This type of damage is not covered
19 Place the P30 console on the maintenance access hook.
under warranty.
20 Reconnect the Data cable, Auto Stop cable (Treadmill only), Stop Key
cable (Treadmill only), Ground wire and the CSAFE cable (if in use)
to the Upper PCA.
21 Reconnect the Heart Rate Cable to the Heart Rate PCA.
22 Replace the Heart Rate PCA. See Replacing the P30 HR PCA (on page 37).
23 Replace the access cover with the hardware removed in step 6.
24 Tilt the console back against the mounting plated. While tilting the
console back feed the excess cable into weldment tube making sure
that the cables will not become pinched.
25 Fasten the console to the dash mounting plate with the four screws
removed in step 4.
26 Fasten the dash transition cover using the two screws removed in
step 3.
27 Check treadmill operation as per procedure, Checking the AMT Operation
(on page 125).
Replacing the P30 Metrics PCA
1 Access the consoles upper PCA and Metric PCS as per procedure,
2 Disconnect the Metric cable (J1) from the Metric PCA board.
3 Remove the four screws that fasten the Metric PCA board to the
Replacing the Upper PCA (on page 28).
console and remove the Metric PCA and the four spacers.
4 Replace the four spacers removed in step 3 onto the console mounting
studs. The spacers must be used to insure the Metrics LED’s are
flush with the display window. If spacers are not used the Metrics
LED’s will apply excessive pressure on the Metrics board and may
cause damage to the Metrics board.
5 Install the replacement Metric PCA into the console using the four
screws removed in step 3.
6 The white wire in the Metrics cables indicate pin 1. Align the white
wire in the Metrics cables with the pin 1 markings on the Metrics
PCA.
7 Connect the Metrics cable to the Metric PCA.
8 Replace the P30 Rear Cover removed step 11 and secure the cover with
the screws 6 screws. Torque to 10 inch pounds.
Note: Do not use an electric screw driver or over tighten the screws. Over
tightening may damage the console. This type of damage is not covered
under warranty.
9 Replace the P30 Option Cap back cover removed step 12 and secure the
cover with the screws 2 screws. Torque to 10 inch pounds.
Note: Do not use an electric screw driver or over tighten the screws. Over
tightening may damage the console. This type of damage is not covered
10 Place the P30 console on the maintenance access hook.
11 Reconnect the Data cable, Auto Stop cable (Treadmill only), Stop Key
cable (Treadmill only), Heart Rate ground wire and the CSAFE cable
(if in use) from the Upper PCA.
12 Reconnect the Heart Rate Cable to the Heart Rate PCA.
13 Replace the Heart Rate PCA. See Replacing the P30 HR PCA (on page 37).
14 Replace the access cover..
15 Tilt the console back against the mounting plated. While tilting the
console back feed the excess cable into weldment tube making sure
that the cables will not become pinched.
16 Fasten the console to the dash mounting plate with the four screws.
17 Fasten the dash transition cover using the two screws (TRM only)
18 Check treadmill operation per Checking the AMT Operation (on page 125).
Replacing the P30 HR PCA
1 Set the treadmill circuit breaker in the off position and unplug the
treadmill’s line cord from the AC outlet.
2 The PCA’s in the console are static sensitive. They can be damaged
if proper static prevention equipment is not used. Attach an antistatic wrist strap to your arm, and then connect the ground lead of
the wrist strap to the treadmill’s frame ground.
3 Remove the HR PCA access panel on the back of the console.
8 Replace the HR PCA access panel removed in step 3.
9 Check treadmill operation per Checking the AMT Operation (on page 125).
Flash Programming the Upper PCA
This unit utilizes an upper PCA software system that is capable of
onsite upper PCA software uploading (re-programming). The software
upload may be accomplished with the use of a PDA (palm device) or a
laptop computer.
If a palm device is to be used it must use OS software version 3.2 or
higher. Currently, we have only tested devices manufactured by Palm,
but other manufacturer’s palm devices may also function correctly.
If a palm device is to be used, the appropriate software must be
available in the palm device. A PDA containing the appropriate software
may be ordered from Precor or it may be downloaded to the PDA from your
desktop or laptop computer via the hot sync function. With the PDA
inserted in the hot sync cradle, download to Precor IFP (in field
programmer) directory.
If a laptop computer is to be used, the computer’s DB9 serial port will
be used for the upload.
1 The P30 must be powered down before the upload procedure can be
initiated. Ensure that the P30 has not been used for a sufficient
time to allow the lower PCA to completely discharge. The light
emitting diode on the lower PCA will go out when the power supply is
discharged.
2 If the CSAFE port is in use, temporarily disconnect the RJ-45 cable
from the CSAFE port. If the CSAFE port is not in use, temporarily
remove the plastic plug from the CSAFE port.
3 Connect the PDA or computer interface cable to the CSAFE port.
4 Select the software file to be uploaded on the palm device or
computer.
5 Start pedaling the P30 or power up using the external power supply.
When the P30 powers up the upload will commence. You must continue
to pedal until the upload is complete, approximately 2 minutes.
6 Stop pedaling or disconnect the external power supply when the
upload is complete, and allow the lower PCA to completely discharge.
The light emitting diode on the lower PCA will go out when the power
supply is discharged.
7 Start pedaling the P30, after it has been allowed to power down, the
P30 will now be operating on the newly uploaded software.
8 Thoroughly, check the P30’s function per Checking the AMT Operation (on
Troubleshooting the Keypad and Upper PCA........... 41
Troubleshooting the Handheld Heart Rate............ 44
P30 AMT '11 - Upper Display does not Illuminate.... 48
Troubleshooting the Keypad and Upper PCA
Procedure
Note: The white or yellow wire on the cables shown below denotes pin 1.
When these cables are inserted into their connectors, the white or yellow wire
must align with the pin 1 designation on the PCA.
1 Set the treadmill’s on/off switch in the off position (Treadmill
only). Access the upper electronics and machine controls per
Replacing the Upper PCA (on page 28), steps 1 through 12. Place the console
on the maintenance mounting hook and reconnect all cables.
2 If the message STUCK KEY is displayed when the unit is turned on for
treadmill’s or pedaling on the RBK, UBK, AMT, and EFX, skip to step
23.
3 If a key does not function, skip to step 18.
4 If the display does not illuminate, continue with step 5.
5 If the upper PCA is not illuminating, skip to step 12.
6 If the metrics PCA is not illuminating, continue with step 7.
7 Remove the four screws that fasten the metrics PCA to the display
face and rotate it so that the front of the PCA is visible. For
treadmill’s set the on/off switch in the on position for AMT’s,
EFX’s, UBK’s, RBK’s and CLM’s start pedaling.
Note: There are four spacers between the Metric board and the mounting
studs. These must be saved and reinstall with the Metric board. If they are not
used the Metric board or display face may become damaged.
8 With a DC voltmeter, measure between TP4 (+8V) and TP1 (DGND) for 8
Vdc and between TP2 (+5V) and TP1 (DGND) for 5Vdc.
9 If 5 Vdc is not present on TP2 and 8 Vdc is present on TP4, replace
the metrics PCA.
10 If 8 Vdc is not present on TP4 and the upper PCA is illuminating
normally, replace the cable between the upper PCA and the metrics
PCA.
11 If you have performed steps 7 - 10 and the metrics PCA still does
not illuminate, contact Precor customer support for assistance.
12 For treadmill’s set the on/off switch in the on position, for AMT’s,
EFX’s, UBK’s, RBK’s and CLM’s start pedaling.
13 With a DC voltmeter, measure between pins 2 and 4 of J9 (JTAG
Connector) for 5 Vdc and between pins 6 and 7 of J6 (Option Cap) for
8 Vdc.
14 If 5 Vdc is not present on J9 and 8 Vdc is present on J6, replace
the upper PCA.
15 If 8 Vdc is not present on J6, temporarily replace the upper PCA to
power control module (treadmill’s) or Lower PCA (AMT’s, EFX’s,
UBK’s, RBK’s and CLM’s), cable with a known good cable.
16 If the upper PCA illuminates normally, replace the upper PCA to
power control module cable permanently. If the upper PCA still does
not illuminate, replace the power control module (treadmills) or
Lower PCA (AMT’s, EFX’s, UBK’s, RBK’s and CLM’s).
17 If you have performed steps 12 - 16 and the upper PCA still does not
illuminate, contact Precor customer support for assistance.
18 For treadmill’s set the on/off switch in the on position for AMT’s,
EFX’s, UBK’s, RBK’s and CLM’s start pedaling.
19 If none of the keys on the display are functioning, check the stop
switch cable connection to the upper PCA. (Treadmill Only).If the
stop switch is not connected or the stop switch is not functioning,
none of the display keys will operate. This feature insures that the
treadmill has a functioning stop switch when it is in use (Treadmill
Only).
20 If a particular key is not functioning, perform the keyboard test in
Accessing the P30 Diagnostic Software (on page 13). If the test verifies
that the key is not functioning, replace the console face or the
machine controls.
21 If the console face or the machine controls has been replaced and
the same key or control is still not functioning, replace the upper
PCA.
22 If you have performed steps 18 - 22 and the same key is still not
functioning, contact Precor customer support for assistance.
23 Remove the keypad cable from the upper PCA.
24 For treadmill’s set the on/off switch in the on position for AMT’s,
EFX’s, UBK’s, RBK’s and CLM’s start pedaling.
25 If the STUCK KEY message is no longer displayed, replace the console
face.
26 Set the treadmill’s on/off switch in the off position (Treadmills
Only).
27 Remove the Machine Control cable from the upper PCA.
28 For treadmill’s set the on/off switch in the on position for AMT’s,
EFX’s, UBK’s, RBK’s and CLM’s start pedaling.
29 If the STUCK KEY message is no longer displayed, replace the machine
Controls assembly.
30 If the STUCK KEY is still being displayed with the keyboard cable
disconnected, replace the upper PCA.
31 If you have performed steps 23 - 30 and the STUCK KEY message is
still being displayed, contact Precor customer support for
assistance.
Troubleshooting the Handheld Heart Rate
Circuit Description
The hand held heart rate system is actually a dual system, that is, it
can accept a heart rate signal from either the hand held heart rate
contacts on the unit’s handlebar or from a Polar heart rate chest strap
transmitter. The PCA is configured for hand held priority. That is, if
both a chest strap and hand heart rate signal are being received, the
system will accept the hand held signal and ignore the chest strap
signal. If a hand held signal is not being received, the system will
accept the chest strap signal.
Note: There are four typical failure modes for the hand held/chest strap heart
rate system. They are: hand held is normal - no chest strap reading; no hand
held reading - chest strap normal; no hand held or chest strap reading or
constant or intermittent readings when neither hand held or chest strap are in
use.
Figure 17: Hand Held - Chest Strap Heart Rate PCA
Normal hand held reading - No chest strap reading
1 Access the diagnostic program, per procedures Accessing the P30
Diagnostic Software (on page 13), P80 Settings (on page 51), or Accessing
the Diagnostic Software. Advance to the heart rate display portion
of the diagnostic program. Verify that a chest strap signal is not
being accepted with either a Polar heart rate test transmitter or a
known good chest strap transmitter. If this reading is good, skip to
step 3.
2 Using a known good Polar heart rate chest strap, verify that the
heart rate operates with the known good chest strap. If the known
good Polar chest strap does correct the problem, replace the
original chest strap transmitter.
3 If the above procedures did not correct the problem, replace the
4 Access the diagnostic program, per procedures Accessing the P30
Diagnostic Software (on page 13), P80 Settings (on page 51), or Accessing
the Diagnostic Software. Advance to the heart rate display portion
of the diagnostic program. Verify that a hand held signal is not
being accepted by firmly grasping both the right and left hand held
contacts on the handlebars. Cover as much of the top and bottom
contact surface area with your hands as possible (without moving
your hands), you should receive a heart rate reading within ten
seconds.
5 If the hand held signal is now being accepted, something in the near
vicinity is radiating RF (radio frequency) energy that is being
received by the chest strap portion of the heart rate PCA.
6 If a hand held signal still not being accepted, skip to step 8.
7 Replace the heart rate PCA with a 300812-101 (or higher) heart rate
PCA. 300812-101 and higher versions of heart rate PCA are less
susceptible to radiated RF energy.
8 Access the diagnostic program, per procedures Accessing the P30
Diagnostic Software (on page 13), P80 Settings (on page 51), or Accessing
the Diagnostic Software. Advance to the heart rate display portion
of the diagnostic program. Verify that a hand held signal is not
being accepted by firmly grasping both the right and left hand held
contacts with the opposite hands, right hand on the left handlebar
contacts and left hand on the right handlebar contacts. Cover as
much of the top and bottom contact surface area with your hands as
possible, you should receive a heart rate reading within ten
seconds. If a hand held signal is still not being accepted, skip to
step 10.
9 If a hand held signal was accepted in step 8, the hand held contact
wiring is reversed. The harness that connects to the hand held
contacts in the handlebar is segregated into two groups. One group
has blue shrink wrap around it and the other group has black shrink
wrap around it. The blue group must go to the right hand contacts
and the black group must go to the left hand contacts. If necessary,
rewire the hand held contacts as described above and test as
described in step 4.
10 If the wiring is correct, for the following measurements. With an
ohmmeter measure between the lower right contact pin on the J1
connector and the lower right hand held heart rate contact on the
handlebar. The reading should be 1 Ω (ohms) or less. Measure between
the upper right contact pin on the J1 connector and the upper right
hand held heart rate contact on the handlebar. The reading should be
1 Ω (ohms) or less. Measure between the upper left contact pin on the
J1 connector and the upper left hand held heart rate contact on the
handlebar. The reading should be 1 Ω (ohms) or less. Measure between
the lower left contact pin on the J1 connector and the lower left
hand held heart rate contact on the handlebar. The reading should be
1 Ω (ohms) or less. If any of the above readings are greater than 1
Ω (ohms), replace the heart rate PCA to handlebar wire harness.
No hand held reading - No chest strap reading
11 Access the diagnostic program, per procedures Accessing the P30
Diagnostic Software (on page 13), P80 Settings (on page 51), or Accessing
the Diagnostic Software. Advance to the heart rate display portion
of the diagnostic program. Verify that neither a chest strap signal
or a hand held signal is being accepted with either a heart rate
test transmitter or a chest strap transmitter.
12 Check the plug/connector connections on both the heart rate PCA
(J4), and upper PCA (J1).
13 If neither a chest strap signal or a hand held signal is being
accepted, measure between the ground and 5 Vdc pins on J4 for 5 Vdc.
If 5 Vdc is present, replace the heart rate PCA.
14 If 5 Vdc is not present, remove the connector from J4 of the heart
rate PCA. Measure between the ground and 5 Vdc pins of the connector
(just removed from the heart rate PCA) for 5 Vdc. If 5 Vdc is
present, replace the heart rate PCA. If the 5 Vdc is not present,
measure between the corresponding pins of J1 on the upper PCA (red
and black wires). If 5 Vdc is not present replace the upper PCA. If
5 Vdc is present, replace the upper PCA to heart rate PCA cable.
15 Constant or intermittent readings when neither the hand held or
chest strap is in use.
16 Verify that a ferrite core is clamped around the heart rate PCA to
upper PCA cable. Constant or intermittent heart rate readings when
neither heart rate system is in use is caused by something in the
near vicinity radiating RF energy that is being received by the
chest strap portion of the heart rate PCA. Replace the heart rate
PCA with a 300812-101 (or higher) heart rate PCA. 300812-101 and
higher versions of heart rate PCA are less susceptible to radiated
RF energy.
1 Because this is self powered unit, the display will not illuminate
until it is used or the optional external power supply is equipped.
If the optional external power supply is equipped, the display
should be constantly illuminated.
2 If the optional external power supply is not equipped, skip to step
5.
3 Disconnect the optional external power supply from the AMT and
measure between the inner and outer sleeves of the power supply’s
output jack with a DC voltmeter. You should measure approximately 18
VDC.
4 If the voltage measured in step 3 was significantly low, replace the
optional external power supply. If the voltage measured in step 3
was 0 Vdc, disconnect external power supply from its AC outlet and
measure the voltage at the AC outlet. If the AC outlet voltage is
normal replace the optional external power supply. If the AC outlet
voltage is significantly low or 0 Vdc, the AC system must be
inspected by an electrician.
5 Troubleshoot the generator per Procedure, Troubleshooting the Generator
(on page 147).
6 If the generator was found to be good, the problem will be in either
the lower PCA, upper PCA or the upper to lower PCA interconnect
cable.
Warning: Because this is a self powered unit, it will either be
necessary to either equip the unit with the optional external power
supply or have an assistant pedal on the unit while voltage
measurements are being taken. Because of the danger of working on the
unit while it is in motion using the optional external power supply is
strongly recommended.
7 Remove the right side cover and disconnect the interconnect cable
8 The following voltage reading must be taken while the unit is in
motion. Extreme care must be taken to keep meter wires, hands, etc.
clear of all moving parts. Using a DC voltmeter, measure the voltage
between TP24 (7.5V) and TP14 (DGND). The voltage measured should be
approximately 7.5 Vdc. If the voltage is significantly low, replace
the lower PCA.
9 Reconnect the interconnect cable to the lower PCA and repeat the
voltage measurement in step 8. The voltage measured should be
approximately 7.5 Vdc. If the voltage is significantly low, the
problem is in the upper PCA or the upper to lower PCA interconnect
cables.
Figure 18: Lower PCA Test Points
10 Troubleshoot the upper to lower PCA interconnect cables as per
procedure, Troubleshooting the Interconnect Cable (on page 150).
11 If the upper to lower interconnect cable is found to be good,
replace the upper PCA.
12 If you have performed all of the above tests and are unable to
resolve the problem, contact Precor customer support.
1 The Welcome screen will be the first screen you see when you
approach the P80 console. Press the Pause key and continue holding
it down while you double press (prior to 1.1 version software
release) or single press (1.1 version or after software release) in
sequential order an X configuration on the Volume and Channel key
pad (Channel Up, Volume Down, Volume Up, Channel Down).
Release the Pause key when done pressing the arrows.
Note: The keypad X configuration sequence should be not faster than 1
second between each key stroke. Any speed faster than 1 second may result
in an unsuccessful access to the sign-in screen.
Figure 20: P80 Sign In Access
2 The sign-in screen will be displayed. Type in the technician access
,5,1,7,6,5,7,6,1, and then touch OK. The Settings menu will be
displayed. See Settings Table, P80 Settings Table (on page 53).
3 Use the System mode to configure settings in ways that benefit the
users and the facility. Changes made to these settings save to the
fitness equipment. To select a Setting touch on an item in the
Settings menu.
System information settings
provide basic information
including equipment models
and serial numbers. Most of
these settings are for
informational purposes only
The Equipment Usage
settings provide information
on individual usage as well
as cumulative equipment
usage. This information and
these settings are accessible
only to administrators and
registered service
Use the System Settings to
test your equipment and set
parameters that benefit your
users and your facility. The
System Settings menu is
visible only to administrators
and registered service
Touch About
Touch
Equipment
Usage
Touch
System
Settings
See Procedure, P80
About (on page 54)
See Procedure, P80
Equipment Usage (on
page 57)
See Procedure, P80
System Settings (on
page 58)
Software
Downloads
Allows software to be
automatically downloaded.
It is recommended that this
remain ON at all times.
Contact Precor Customer
Service before this setting is
disabled.
Touch
Software
Downloads
Touch radio button next
to On to Enable
Automatic Downloads.
The default is No Available
Updates. When Software
Downloads is enabled and a
USB flash drive is installed
that contains a valid software
package, Available Updates
will be displayed.
The log indicates current
Touch
Available
Updates
Touch
Touch next to any of the
available updates.
Touch View to see the
details of the update.
Touch BACK to
updates list.
Touch Install to load the
update into the console.
See Procedure,
Updating P80 Softwa
Touch BACK to
P80 About
1 Refer to Procedure, P80 About (on page 54) to access the About
menus.
2 About settings provide basic information including equipment models
and serial numbers. Most of these settings are for informational
purposes only and cannot be modified. See About Table. To select a
About item, touch on an item in the About menu.
Provides a list of the m
recent hardware and
software events, and
equipment status. If no
events have occurred
since the last time the
event log was cleared,
the event log screen will
be empty.
Number of events
recorded by equipment
Set by Precor and stored
Touch
Event Log
None
None
Details of the Event will
be listed.
Touch REFRESH to
update the list in the log.
Touch SAVE to save the
log to a Flash drive
device.
Touch BACK to
None
None
Lower Serial
Number
Stored in the console's
memory. Note: This
information is not
Use this setting to
save your
equipment settings
to a USB flash drive,
or to import settings
from a USB flash
Touch
Manage
Settings
Updating P80 Software
The P80 console is driven by software with innovative features which
enhances the user’s experience during a workout. Periodically the
software will need to be updated with improvements to the functionality
and features. Procedure will explain how to update the software at the
customer’s location using a USB Flash drive device for P80 consoles.
A USB Flash drive device with pre-loaded software or a web site
location where you can upload the software onto a USB Flash drive
device will be provided by PRECOR.
Procedure to update (Flash) software for P80 consoles:
1 The Welcome screen will be the first screen you see when you
approach the P80 console. Press the Pause key and continue holding
it down while you double press (prior to 1.1 version software
release) or single press (1.1 version or after software release) in
sequential order an X configuration on the Volume and Channel key
pad (Channel Up, Volume Down, Volume Up, Channel Down).
Release the Pause key when done pressing the arrows. Note: The
keypad X configuration sequence should be not faster than 1 second
between each key stroke. Any speed faster than 1 second may result
in an unsuccessful access to the sign-in screen.
5 The P80 software will automatically access the USB Flash Drive files
and look for a valid software update. While the USB Flash Drive is
being validated, No Updates Available, will be displayed on the P-80
console screen. Once an update file has been identified and
validated, Updates Available, will be displayed on the P-80 console
screen. Touch the spot on the P80 console screen where Updates
Available is displayed to select.
6 A software package or list of software packages will be displayed on
the P80 console screen. Touch the spot on the P80 console screen
displaying the software package required. Note: PRECOR customer
service will specify which software package is required when the
service is requested to update P-80 Console software. If you are
unsure of which software package should be installed, contact PRECOR
customer service.
7 Touch the spot on the screen displaying Install and the P80 console
will start the software installation process. The software
installation process can take up to 35 minutes and the console will
automatically reboot when it has finished.
8 Once the download as started it is safe to remove the USB Flash
Drive device from the P80 USB port.
9 When the P80 has completed rebooting the Welcome screen will be
displayed. The installation of the software update is now complete.
10 Verify the correct version of software has been installed. This can
be found by accessing the main menu using the instructions provided
in steps 1 and 2, then selecting About. The software version will
be displayed under Installed Bundle Version and should match the
package selected in step 5.
P80 System Tests
1 Refer to Procedure P80 Settings (on page 51), to access the Equipment
Usage menus.
2 The systems tests allow you to check specific functions of the
equipment.
3 To select a Systems Test touch on an item in the Systems Test menu.
4 Touch BACK to Settings.
at 1 mph; count will be reset to 0, and
then count up with each step.
3 Touch STOP to end the test. Touch
1 Touch
, then Touch
.
2 The belt will start a 0.5 mph, using the
machine control paddle to increase by
0.1 mph increments. The console will
display motor current as the belt speed
increases or decreases.
3 Touch STOP to end the test. Touch
1 Touch
START.
, and then Touch
2 Use the machine control paddle to change
the lift. The console will display the
analog to digital (A/D) incline position
number and the incline position in
percent of incline.
3 Touch STOP to end the test. Touch
1 Touch
AMT Stride
Touch START.
Position Test
2 The High, Low and Dynamic positions will
be displayed. The low position display
shows the lowest position reading
AMT Only
achieved during the test. The high
position display shows the highest
position reading achieved during the
test. The dynamic position display
changes with current stride position.
Stride on the AMT to the maximum
horizontal movement in both directions.
At maximum stride length the low stride
display should read 0 and the high
the resistance up and down. The
resistance changes will be displayed in
power bits.
3 Touch STOP to end the test. Touch
Battery Test
SPM
RPM Test
AMT, RBK,
UBK, EFX
AMT, EFX
RBK, EFX
1 Touch
START.
, and then Touch
2 The console will display the voltage of
the battery.
3 Touch STOP to end the test. Touch
1 Touch
, and then
.
2 The console will display Pulse. The
pulse is showing a count of zero
crossings from the generator. SPM shows
a software averaged version of pulse.
An averaged pulse per minute will
display as Minimum and a Maximum pulse.
3 Touch STOP to end the test. Touch
1 Touch
, and then
.
2 The console will display Pulse. The
pulse is showing a count of zero
crossings from the generator. RPM shows
a software averaged version of pulse.
3 Touch STOP to end the test. Touch BACK
1 Touch
CrossRamp
START.
2 Use the machine control paddle to change
the CrossRamp. The console will display
the analog to digital (A/D) and Glitches
EFX Only
incline position number. Available
battery voltage to the lift system will
be displayed.
ports currently active, there should be
six. Insert a USB Flash drive into the
USB port and then Touch CLOSE, and then
Touch USB Test again. The number of
active USB ports should be increased by
one.
3
Touchscreen
Test
Touchscreen
Calibration
All
All
1
.
2 Place your finger on one corner of the
screen and run it around the edges
multiple times.
3 The line drawn by your finger should be
very close to previous passes. Be sure
that you do not touch anything else with
your other hand.
4
1 Touch
.
2 A + will be displayed on a white screen.
Touch the +, the spot you just touched
will disappear and then be replaced with
another + in a different location on the
white screen.
3 Continue to touch the + as they appear
on the screen until the test
automatically exits to the Systems Test
screen. You cannot exit this test until
all the + have been touched and the
Backlight Test
1 Touch
2 The backlight will display three levels
of brightness, 100%, 60%, and 0% in
All
succession in 1 second intervals,
returning to the System Tests menu when
RGB Test
All
1 Touch
2 The full screen should move through a
succession of five colors, requiring a
touch anywhere on the screen to advance
to the next. These are Red, Green, Blue, Black, White, returning to the
the speaker mounted in the headphone
jack and while providing a graphic of a
scrolling bar on the screen. When the
test is complete it will automatically
Numeric
Backlight Test
Numeric Display
Test
Heart Rate
Step Count Reset
All
All
All
1 Touch
.
2 This tests the back lights of the lower
display just above the paddle controls.
The back lights of the display will
illuminate in sequence and then
automatically exit to the Systems Test
1 Touch
.
2 This will test the numeric LCD display
just above the paddle controls. All the
numeric characters will be displayed in
sequence and then automatically exit to
1 Touch
.
2 Grasp both of the heart rate grips on
the handlebar, after a couple of seconds
the heart rate will be displayed.
3
1 Touch
.
The screen will display:
Current total step count
Step Count Reset
Odometer Value at the last reset.
1 During this step it is necessary to
AMT12 Only
pedal the AMT while the Step Count is
being reset, touch START.
The Current Total Step Count will go to "0".
The Step Count Reset will increment up.
The Odometer will display the total miles
Configuring your audio, video, and network infrastructure requires
expertise. Precor strongly recommends that you work with a qualified
contractor to set up this infrastructure.
This setting provides the following information:
♦
Network Type
•
Wired (Default)
•
Wireless (This is a selectable option but is not currently
supported)
♦
Configuration
♦
Status
♦
IP Address
♦
Preva Server
To view connectivity information:
1 Refer to Procedure,P80 Settings (on page 51) to access the
Connectivity menus.
2 To select Connectivity Setting touch on an item in the Connectivity
P80 Connectivity Set Up and Registration Procedure
1 Before you attempt to connect to the Precor Preva Server, check the
status field and ensure that the Connected is displayed. If the
Status field shows Not Connected, you will need to diagnose your
networking issue. Continue with Step 2. If the Status Field shows
Connected go to step 3.
2 Either the Ethernet cable is not connected or there is a problem
with the network connection. Check the connection and then check
the Status field again. If the Status field shows Connected continue
with step 3. If the Status field still shows Not Connected contact
your IT expert.
4 If the Status shows Connected and the IP Address is not present, it
will be necessary to either have the P80 software assist in
acquiring the IP Address (Recommended) or enter the IP address
manually. Touch Configuration from the Connectivity menu.
5 Touch on the radio button next to Automatic, and then touch
6 The P80 will attempt to connect to the network, and when successful
will acquire the networking information necessary for registration.
If the console was already connected, it will skip straight to the
‘Network Settings’ screen and display the current network
information. Once a good IP Address has been acquired, the P80 can
now be registered. Continue with step 12.
7 Note: Manually setting up the network configuration is not
recommended. Precede with the following procedure only with
assistance from the facilities IT staff. Touch CONFIGURATION form
the Connectivity menu.
8 Touch on radio button next to Manual, and then touch SETTINGS.
Site Code – This is given at the time of the dispatch.
•
Base Serial – The base serial number is located on the exercise
equipment.
19 Friendly Name - It is recommended that you have a list of the
equipment with friendly names already assigned, such as Elliptical
1, Treadmill 22, and so forth. Touch NEXT.
20 Enter the Site Code using the on-screen keypad, then touch NEXT.
21 Enter the Base Serial Number using the on-screen keypad, then touch
NEXT.
22 Enter the Friendly Name using the on-screen keypad, then touch NEXT.
23 The next screen displayed will be the Registration Summary screen.
Review the screen to ensure the information entered is correct
before completing the registration. If the information is not
correct touch BACK until the screen with incorrect information is
displayed. Make the corrections and then touch NEXT unit you are
back at the Registration Summary Screen.
24 Touch Register to submit the data.
25 If the registration was successful a screen with current software
version will be displayed. Touch FINISH to the Systems Settings
menu, if a newer software version is available, the console will
automatically start the download.
26 Touch BACK to Settings.
P80 Display
The display settings allow you to configure Measurement units, Standby
Mode Delay, Time Zone, and Closed Captioning.
1 Refer to Procedure, P80 Settings (on page 51) to access the Display
menus.
2 To select a Display Settings touch on an item in the Display menu,
see .
3 Touch BACK to Settings, P80 Display Settings Table (on page 80).
The following procedure discusses how to set up the channel guide and
how to configure audio and video playback.
Note: The correct region must be set before you scan for available channels.
To confirm the settings, touch Region, a list of regions will be displayed.
Confirm that the radio button next to the appropriate region for your area is
selected. If the highlighted region is incorrect, touch on the radio button next to
appropriate region. Touch OK, and then touch t BACK to the TV Settings
menu.
To scan for available channels:
1 Touch Channel Guide from the TV Settings menu.
2 Touch Scan Channel.
3
To name a channel:
4 Touch a channel from the channel guide list.
5 Touch Channel Name. A key board will be displayed.
6 Type in the desired channel name.
7 Touch OK.
The console scans for available channels and when the scan is finished
all available channels will be displayed on the screen.
To delete a channel from the Channel Guide:
1 On the Channel Guide screen, touch the listing for the channel you
3 When asked, Are you sure? Verify that you want to proceed, then
touch YES.
4 Touch BACK to Settings.
P80 Workout Limits
The Workout Limit settings give you control over how the equipment is
used. Use these settings to create reasonable limitations for equipment
usage including speed, incline, resistance, and duration.
1 Refer to Procedure, P80 Settings (on page 51) to access the Workout
2 To select a Workout Limits setting, touch on an item in the Workout
Maximum Workout Duration
Maximum Pause Time
Summary Time Out
Resistance Range (RBK, UBK
Speed Limit (TRM only)
Incline Limit (TRM only)
Auto Stop (TRM only)
Channel Guide
Region
Default Channel
Primary DNS – Manual only
Secondary DNS – Manual only
Gateway – Manual only
Connectivity
P80 Cloning Procedure
The following procedure will explain how to save existing P80 console
settings onto a USB Flash Drive and then import the saved settings into
another P80 console.
Note: Settings from one P80 console will need to be configured manually and
imported to a flash drive device before cloning and exporting to additional P80
consoles.
1 The Welcome screen will be the first screen you see when you
approach the P80 console. Press the Pause key and continue holding
it down while you double press (prior to 1.1 version software
release) or single press (1.1 version or after software release) in
sequential order an X configuration on the Volume and Channel key
pad (Channel Up, Volume Down, Volume Up, Channel Down).
Release the Pause key when done pressing the arrows. Note: The
keypad X configuration sequence should be not faster than 1 second
between each key stroke. Any speed faster than 1 second may result
in an unsuccessful access to the sign-in screen.
Figure 29: P80 Sign In Access
2 The sign-in screen will be displayed. Type in the technician access
code and then touch OK. The Settings menu will be displayed. See
Settings Table.
3 Use the System mode to configure settings in ways that benefit the
users and the facility. Changes made to these settings save to the
fitness equipment. To select a Setting touch on an item in the
Settings menu.
4 The Settings screen will be displayed. Select Systems Settings from
the Settings menu.
5 From the Systems Settings menu select Manage Settings.
8 When the gray inactive Save to USB Flash Drive Icon becomes active
and lights up, select Save to USB.
Note: Depending on the type of Flash Drive device used, it may take up to a
minute for the USB Flash Drive Icons to become active (light up) after the USB
Flash Drive device has been inserted into the P80 console.
Figure 32: Cloning Procedure- Save To USB
9 You will see a scrolling in process bar and then Settings exported
17 Touch OK to return the Settings menus.
18 Exit the service mode as described in steps 14.
19 Repeat the described process steps 11 though 16 for any remaining
P80 consoles.
Replacing the P80 Console
Required Tools
♦
2- 7/16 in open end wrenches
♦
¹/₄-inch hex wrench
♦
SAE ⁵/₃₂-inch hex wrench
♦
SAE¹/₂-inch box-end wrench
♦
#2 Phillips screwdriver
♦
Wire cutter
Console Mounting - Removal
1 Remove the rear vent cover. Do not use a sharp tool, such as a flat
bladed screwdriver, to pry up the cover, as you can damage the
covers and possibly components inside the console. When the cover
starts to come loose, gently unsnap it from the P80 and set it aside
where it will not be scratched.
3 Remove the console by lifting it off the mounting hook.
Console Hook and Notch
1 Install the replacement console. In the back of the P80 console
there is a large black steel support called the Armor. At the
bottom of Armor there is a notch about ¼ inches in depth and about
1½ inches across. Set the console notch over the over the mounting
hook.
Note: Before releasing the console insure the notch is securely seated
over the mounting hook.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.