Precision Medical Vac Trap User Manual

U S E R M A N U A L
Vac Trap
Vacuum Safety Trap Bottle
Model: PM9000 (shown)
SAVE THESE INSTRUCTIONS
CAUTION
Federal (USA) law restricts this device to sale by or on the order of a physician.
www.precisionmedical.com
RECEIVING / INSPECTION
Remove the Precision Medical, Inc. Vac Trap from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENDED USE
To protect suction regulators and central vacuum systems from liquid contamination.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Vac Trap. This is provided for your safety and to prevent damage to the Vac Trap. If you do not understand this manual, DO NOT USE the Vac Trap and contact your Provider.
SAFETY INFORMATION - wARNINGS AND CAUTIONS
wARNING
Indicates a potentially hazardous situation which, if not avoided, may result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
CAUTION
Used without the safety alert symbol, indicates a potentially hazardous situation which, if not avoided, could result in property damage.
wARNING
Use this Vac Trap only for its “Intended Use” as
described in this manual.
CAUTION
Only personnel instructed and trained in its use should operate this Vac Trap.
Be sure all connections are tight and leak free.
TYPICAL INSTALLATION DIAGRAM
CAUTION
The Vac Trap must be in the vertical position to function
.
Vacuum Regulator
Vac Trap
Tubing
Suction Collection Canister
PARTS LIST
Description Part #
Float Assembly 503234 Jar 1731 Tubing Nipple 9073
MAINTENANCE
Before each use, inspect the Vac Trap for any damage; DO NOT USE if damaged.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical ,Inc. Return Policy is available on the Internet at www.precisionmedical.com.
Manuals available on our Website;
www.precisionmedical.com
DISPOSAL INSTRUCTIONS
Dispose of the Vac Trap in accordance with the local regulations.
Please Recycle
wARNING
Dispose of the Vac Trap using Biohazard precautions.
TROUBLESHOOTING
Problem Probable Cause Remedy
Float Stuck 1. Float neck dirty 1. Remove foreign
substance
No Suction 1. Leaking
connections
2. Suction Source
not turned “ON”
3. Jar full
1a. Check
connections & reconnect
1b. Tighten Jar
2. Turn Suction
Source “ON”
3. Empty / clean jar
CLEANING
1. Unscrew the Jar.
2. Remove the Float Assembly.
Gas Sterilization:
Ethylene Oxide mixtures can be used at
temperatures up to 135° F. Follow the procedures
from the manufacturer of the sterilizer.
CAUTION
Exceeding the recommended sterilization temperatures may shorten the life of the Vac Trap.
• DO NOT autoclave.
LIMITED wARRANTY
AND
LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that the Medical Vacuum Safety Trap will be free of defects in workmanship and/or material for the following period:
One (1) year from date of shipment. Should any failure to conform to this warranty appear within the
applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with
Precision Medical, Inc.’s instructions and standard industry practice,
and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES. The representative of Precision Medical, Inc. or any retailers are not
authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
Precision Medical, Inc.
300 Held Drive
Northampton, PA 18067 USA
Toll Free Tel: 800-272-7285
Toll Free Fax: 800-353-1240
Tel:(+001) 610-262-6090
Fax:(+001) 610-262-6080
www.precisionmedical.com
ISO 13485 Certified
502056 Rev11 01/12 (6M) Printed in USA