Precision Medical Continuous Intermittent Vacuum Regulator User Manual

USER MANUAL
CONTINUOUS INTERMITTENT
MODEL SERIES: PM3300 (shown) PM3400
SAVE THESE INSTRUCTIONS
CAUTION
300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 ISO 13485 Certied www.precisionmedical.com
Federal (USA) law restricts this device to sale by or on the order of a physician.
CONTENTS
RECEIVING / INSPECTION .......................................................... 2
INTENDED USE ............................................................................ 2
READ ALL INSTRUCTIONS BEFORE USING ............................. 2
EXPLANATION OF ABBREVIATIONS ......................................... 2
SAFETY INFORMATION - WARNINGS AND CAUTIONS ............ 2
SPECIFICATIONS ......................................................................... 3
OPERATING INSTRUCTIONS ...................................................... 4
PARTS DESCRIPTION ................................................................. 5
PARTS LIST .................................................................................. 5
REPAIR KITS ................................................................................ 6
DISASSEMBLY INSTRUCTIONS ................................................. 6
ASSEMBLY INSTRUCTIONS ....................................................... 6
VACUUM REGULATOR CLEANING ILLUSTRATION .................. 7
CLEANING / DECONTAMINATION .............................................. 8
MAINTENANCE ............................................................................ 8
RETURNS ..................................................................................... 8
DISPOSAL INSTRUCTIONS ......................................................... 9
TROUBLESHOOTING .................................................................. 9
LIMITED WARRANTY ................................................................ 10
DECLARATION OF CONFORMITY ............................................ 11
CONTINUOUS INTERMITTENT
1
RECEIVING / INSPECTION
Remove the Precision Medical, Inc. Vacuum Regulator from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENDED USE
The devices are intended to control and show the amount of vacuum from a central vacuum system used in various medical suctioning procedures.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Vacuum Regulator. This is provided for your safety and to prevent damage to the
Vacuum Regulator. If you do not understand this manual, DO NOT USE the Vacuum Regulator and contact your Provider.
EXPLANATION OF ABBREVIATIONS
l/min Liters Per Minute mmHg Millimeters of Mercury inHg Inches of Mercury kPa Kilopascal
SAFETY INFORMATION - WARNINGS AND CAUTIONS
WARNING
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
CAUTION
Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
CONSULT ACCOMPANYING DOCUMENTS
Symbol indicates the device complies with the requirements of Directive 93/42/EEC concerning medical devices and all applicable International Standards.
(On CE marked Devices ONLY)
There are no components in this product made with natural rubber latex.
WARNING
DO NOT use this Vacuum Regulator for anything other than its Intended
Use. Personal injury and/or damage to Regulator may result from misuse.
Only personnel instructed and trained in its use should operate this
Vacuum Regulator.
2
CONTINUOUS INTERMITTENT
SPECIFICATIONS
GAUGE RANGE:
GAUGE ACCURACY:
Analog: ± 5% of MAX Digital/Analog, Dual Gauge:
Digital Display: ± 1% of Full Scale Analog Gauge: ± 5% of MAX within ref. Indicator
VACUUM PORTS: 1/8 NPT Female
MODES:
PM3300: 0 - 200 mmHg - Full Vacuum
*PM3300E: 0 - 200 mmHg (0 - 26 kPa) *PM3300EHV: 0 - 300 mmHg (0 - 40 kPa)
PM3300HV: 0 - 300 mmHg - Max Vacuum
PM3400: 0 - 150 mmHg
*PM3400E: 0 - 150 mmHg (0 - 20 kPa)
*Counterclockwise direction
REG. - (Regulated) provides an adjustable,
continuous vacuum level OFF - No Vacuum INT. - (Intermittent) provides an adjustable vacuum
level that cycles between ON and OFF
FLOW: Models
PM3300:
PM3400: 50 l/min
MAXIMUM FLOW
IS OBTAINED WITH A VACUUM SOURCE OF 21” Hg (71.1 kPa)
Mode Max Flow
REG
51 l/min
MAXIMUM VACUUM:
PM3300: REG. Mode @ Full Vac-396 mmHg (53 kPa)
PM3300HV: REG. Mode @ Max Vac-396 mmHg (53 kPa)
PM3400: Restricted to 170 mmHg ± 10 mmHg (1.3 kPa)
INTERMITTENT CYCLE TIME: Factory set at sixteen (16) seconds ON
and eight (8) seconds OFF ( Reference only)
Operating Environmental Limits: 0°F to 122°F (-18°C to 50°C)
Recommended Environmental Operating Limits:
Storage Environmental Limits: Temperature Range: -4°F to 140°F (-20°C to 60°C)
Humidity: Max 95% Noncondensing
Battery: 3 Volt Lithium, AA (Digital Vacuum Gauge Models ONLY)
Specications are subject to change without prior notice.
CONTINUOUS INTERMITTENT
55°F to 85°F (13°C to 29°C)
3