Posey Roll Belt User Manual

4 (1)
Posey Roll Belt User Manual

 

 

ROLL BELTS

APPLICATION

 

 

Applicable Products: REF 1201, 1231, 1231Q, 1231QXL

INSTRUCTIONS

 

DESCRIPTION OF PRODUCT: TheRollBeltisdesignedtoallowthepatientfreedomtorollfromsidetosideorsitupinbed.Bedapplicationonly. Rx ONLY

1231

INDICATIONS FOR USE:

 

• Patients assessed to be at risk of injury from a fall.

 

• Patients needing freedom to roll from side to side or sit up in bed.

CONTRAINDICATIONS:

DO NOT use on a patient who is or becomes highly aggressive, combative, agitated, or suicidal.

DO NOT use on patients with: ostomy, colostomy, or G-tubes; hernias, severe Cardio Obstructive Pulmonary Disease (COPD); or with post-surgery tubes, incisions or monitoring lines. These could be disrupted by a restraint.

ADVERSE REACTIONS

Severe emotional, psychological, or physical problems may occur: if the applied device is uncomfortable; or if it severely limits movement. If symptoms of these problems ever appear for any reason, get help from a qualified medical authority and find a less restrictive, product or intervention.

POSEY ROLL BELTS

REF 1201 Flame Retardant

REF 1231 Cotton, connecting straps for quick-release ties

REF 1231Q Cotton, connecting straps w/quick-release buckles

REF 1231QXL Cotton, connecting straps w/quick-release buckles

APPLICATION INSTRUCTIONS: (Increased Movement)

Before use, check device for damage. Discard if you have any questions about patient safety.

1.Open the belt and center it across the mattress with the flannel side toward the small of the patient’s back.

2.Position the patient in bed with the belt at waist level.

3.Secure the short strap to a movable part of the bed frame, with a quick-release tie or buckle, out of the patient’s reach.

4.Bring the long strap over, around the patient’s waist and back behind the patient through the slot in the back pad.

5.Secure the long strap to a movable part of the bed frame, with a quick-release tie or buckle, out of the patient’s reach. The belt

must be snug, but not interfere with breathing. To check for proper fit, slide an open hand (flat) between the belt and the patient.

Heed these warnings to reduce the risk of serious injury or death:

BED SAFETY

• ALWAYS use Hospital Bed Safety Workgroup (HBSW) (http://www.fda.gov/ cdrh/beds/modguide.html) compliant side rails in the UP position and fill ALL gaps to reduce the risk of entrapment.

Use side rail covers and gap protectors to help prevent the patient’s body from going under, around, through or between the side rails. A failure to do so may result in serious injury or death if a patient becomes suspended or entrapped. Posey offers a full range of side rail pads and/or gap protectors to cover gaps.

6.

Check that the straps are secure and will not change position,

• There is a risk of chest compression or suffocation, if the patient’s

 

loosen, or tighten if the patient pulls on them, or if the bed is

body weight is suspended off the mattress (figs. 2 and 3).

 

adjusted.

• Monitor per facility policy to ensure that the patient cannot slide

APPLICATION INSTRUCTIONS (Decreased Movement):

down, or fall off the mattress and become suspended or

1.

Open the belt and center it across the mattress with the flannel

entrapped (figs. 2 and 3).

 

side toward the small of the patient’s back.

 

2.Position the patient in bed with the belt at waist level.

3.Secure the short strap to a movable part of the bed frame, with a quick-release tie or buckle, out of the patient’s reach.

4.

Bring the long strap over,

 

FIG. 2

 

 

 

FIG. 3

 

around the patient’s waist and

 

 

 

 

 

 

 

in front of the patient through

 

• STOP USE AT ONCE: if the patient has a tendency to slide

5.

the slot in the front (fig. 1).

 

forward or down in the device; or is able to self-release.

Secure the long strap to

 

NOTE: A restraint with a pelvic piece will reduce the risk of

 

movable part of the bed frame,

FIG. 1

 

sliding, or of the patient pulling the device over his or her head.

 

with a quick-release tie or

 

 

 

See Posey Catalog.

 

 

 

 

 

buckle, out of the patient’s reach. The belt must be snug, but not

 

 

 

 

 

 

 

 

 

 

 

interfere with breathing. To check for proper fit, slide an open hand

 

 

 

 

 

 

(flat) between the belt and the patient.

 

 

 

 

 

 

6. Check that the straps are secure and will not change position,

 

ADDITIONAL SAFETY AND LAUNDERING

 

loosen, or tighten if the patient pulls on them, or if the bed is

 

 

adjusted.

 

 

 

INSTRUCTIONS ON OTHER SIDE

 

 

PoseyCompany•5635PeckRoad,Arcadia,CA91006-0020USA•www.posey.com

 

 

 

MDSS GmbH

 

 

 

EC

REP

 

 

Phone:1.800.447.6739or1.626.443.3143•Fax:1.800.767.3933or1.626.443.5014

 

Schiffgraben 41

 

 

 

 

 

D-30175 Hannover, Germany

© 2009 Posey Company. All rights reserved.

 

 

 

 

 

I9240 040609