PIE PIECAL 322 Operating Instructions Manual

Page 1
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
Model 322
Automated Thermocouple Calibrator
Operating Instructions
Model 322
Automated Thermocouple Calibrator
Operating Instructions
Easy to use
With the PIECAL 322 you can check & calibrate all your thermocouple instruments and measure thermocouple sensors.
Take it with you into the shop, plant or field
Carry it without worry - it comes protected with a rubber boot and rugged, low profile switches. Easy to operate even in the dark areas of the plant with the backlit display.
• Calibrate directly in temperature (°C & °F)
Stop carrying around a millivolt source and thermocouple tables. The PIECAL 322 works with the thermocouples you use including types J, T, E and K. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer “DIAL” plus store any three temperatures for instant recall with the EZ-CHECK™ switch.
• Calibrate quickly with automatic output stepping
Choose between 2, 3, 5, 11 steps and ramp to automatically increment the output in 100%, 50%, 25%, 10% of span or continuously ramp between span and zero. Select the step time to match your system from 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
• Compatible with all process instruments
Connect directly to the thermocouple inputs of smart transmitters, PLCs, DCS and multichannel recorders and verify their outputs or displays.
• Measure Thermocouple Sensors
Trouble shoot sensor connections and find broken wires. The PIECAL 322 measures the thermocouple in degrees C or F.
• Calibration Lab Accurate & Stable
The internal cold junction thermistor is accurate to ±0.1°C and is traceable to NIST. The sensor is thermally bonded to an isothermal mass which includes brass blocks with screw terminals for connection of bare thermocouple wires along with a miniature thermocouple connector for fast connections. The circuitry uses an extremely stable voltage reference and low drift components which make the PIECAL 322 more accurate than most other handheld and benchtop thermocouple calibrators.
Copyright © 2017 All rights reserved • 322-9002 - Rev I 3 May 2017
Page 2
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
q EZ-CHECK™ SWITCH
SOURCE: Instantly output two preset thermocouple
temperatures by moving the EZ-CHECK™ switch to the “LO” position or “HI” position.
For fast three point checks select the “SET” position. The PIECAL 322 will remember the last “SET” value, even with the power off. These values can easily be changed to suit the calibration requirements. The temperatures stored in the HI and LO positions are also used for Auto Stepping.
READ: Slide the switch to the SET position. The PIECAL 322 will display the current temperature from the thermocouple sensor. Slide the switch to HI and the highest temperature measured since turn-on or reset will be displayed; slide the switch to LO and the lowest temperature measured since turn-on or reset will be displayed.
w SOURCE/OFF/READ Switch
Select “SOURCE” to output in °C, °F, or millivolts. Select “READ” to read a thermocouple sensor or millivolts. Select “OFF” to turn the unit off.
e EZ-DIAL™ KNOB
SOURCE: Turn the knob to adjust the output level. Turn clockwise to increase the output, counter clockwise to decrease the output in 0.1° steps at a time. Push down and turn the EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store desired EZ-Check™ HI/LO points in SOURCE mode. Continue to press and hold the knob for two more seconds to start the automatic stepping or ramping.
READ: Press and hold to transfer the current temperature into the EZ-Check™ HI/LO points. This clears the HI/LO temperature readings which will update as the temperature changes.
Double click the knob to get into the PIECAL 322 Configuration Mode. Use configuration to select °C or °F, T/C Type, Backlight On/Off, Step Size, Step Time and Auto Off On/Off, CJC & mV.
CHANGING BATTERIES
Low battery is indicated by “BAT” on the display. Approximately one to four hours of typical operation remain before the PIECAL 322 will automatically turn off. To change the batteries; remove the rubber boot, remove the battery door from the back of the unit by sliding the door downward. This allows access to the battery compartment. Replace with four (4) “AA” 1.5V batteries being careful to check the polarity. Replace the battery door and replace the boot. All stored configuration options (T/C Type, EZ-CHECK Memories, etc.,) are reset to factory settings when the batteries are removed.
Note: Alkaline batteries are supplied and recommended. Purchase the optional Ni-MH rechargeable batteries for maximum battery life.
Basic Operation
LO
TYPE J
°F
THERMOCOUPLE
CALIBRATOR
J, T, E, K, mV
READ
SOURCE
HI
SET
LO
MAX
READ
MIN
OVERLOAD
Push & Hold
to
Store/Ramp
Push & Turn
for
Fast Dialing
Double Click
for Options
OFF
322.1
T/C
+
-
Simulating or reading thermocouples requires the use of thermocouple or extension grade thermocouple wire. Plug thermocouple wires into the miniature thermocouple jack or place bare thermocouple wires onto the brass block under the screws.
Connections
The PIECAL 322 has two banana jacks mounted in the top end of the housing. These are not temperature compensated and are to be used only for millivolt signals.
Page 3
Practical Instrument Electronics
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
Configuration
Configure the Calibrator
Move w POWER SWITCH to “SOURCE” or “READ”.
Setup
Double click the
e DIAL KNOB at any time the
unit is on and the following displays will appear for 15 seconds: Turn the e DIAL KNOB to move through the three pages of menus. Press the e DIAL KNOB to toggle between OFF and ON or to scroll through the settings.
EXIT - exits this menu immediately and saves any changes. Menu will automatically exit after 15 seconds of inactivity.
UNITS - pressing the knob will toggle between °C and °F.
FUNCTION - pressing the knob will toggle between T/C and V. Select T/C for thermocouple or V for the -13.000 to 80.000 mV range.
T/C TYPE - pressing the knob will cycle through T/C types J, K, E & T.
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
V#.##
COLD JUNC - Automatic COLD JUNCTION
(Cold Junction Compensation) may be turned on or off. It is recommended that CJC be left on (default). CJC should only be turned off if an external cold junction compensator or ice bath is used with the PIECAL 322.
AUTO OFF - If AUTO OFF is ON, the unit will turn off after 30 minutes of inactivity to save battery life. If AUTO OFF is OFF the unit will stay on until the POWER SWITCH is moved to the off position.
BACKLIGHT - If BACKLIGHT is ON the backlight will light all the time the unit is powered up. For maximum battery life turn the backlight off when using the calibrator in areas with enough ambient light to read the display.
STEPS/RAMP - pressing the knob will cycle through 2, 3, 5, 11 and RAMP. The endpoints of the steps or ramp are based on the values stored in the HI and LO EZ-CHECK outputs.
2 steps will automatically switch between the values stored in the HI & LO EZ-CHECK (0 & 100%).
3 steps between the HI, Midpoint and LO EZ-CHECK (0, 50 & 100%).
5 steps between the HI and LO EZ-CHECK in 25% increments (0, 25, 50, 75 & 100%).
11 steps between the HI and LO EZ-CHECK in 10% increments (0, 10, 20...80, 90 &100%).
RAMP continuously ramps up and down between the HI and LO EZ-CHECK outputs.
STEP/RAMP TIME - pressing the knob will cycle through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
Note: All settings are remembered even with the power off. Removing the batteries resets the values to factory defaults.
MAIN
> EXIT (1/2) FUNCTION T/C V UNITS °C °F T/C TYPE
J K E T
COLD JUNC ON OFF
FEATURES > EXIT (2/2) AUTO OFF ON OFF BACKLIGHT ON OFF STEPS/RAMP 2 3 5 11 RAMP STEP/RAMP TIME 5 6 7 8 9 10 15 20 25 30 60