PIE 235 Operating Instructions Manual

Page 1
Copyright © 2016 All rights reserved • 235-9002 - Rev C 30 March 2016
PIE Model 235
Process Voltage Calibrator
Operating Instructions
PIE Model 235
Process Voltage Calibrator
Operating Instructions
Intuitive EZ-DIAL Double Click Menu makes it easier to setup than other calibrators. Turn on the backlight to easily see the display in dark areas of the plant. Connections diagrams are indicated on the display for each function along with a labeled connector panel on the top.
• Calibrate with Confidence
Accurate to ±0.02% of Reading + 0.01% Full Scale. Out of range or over current is indicated by an audible alarm, flashing display and the OVERLOAD LED will light.
• Source three ranges of mV & V dc plus percent of 1 to 5 volts
With the 235 you can check, calibrate and measure all your voltage, millivolt and pH signal instruments in your plant. Source -500.00 to
999.00 mV, 1.000 to 5.000 V, 0.0 to 100.0% of 1 to 5 V; 0.000 to 20.000 V.
• Power your loops with over 1,000 Ohm loads
The 20.000 V range has an extended span of up to 24.000 Volts giving you a portable source for powering your loops.
• Read DC volts
The 235 can measur
e from -500.00 to +999.00 mV,
0.000 to 5.000 V, 0.0 to 100.0 % of 1 to 5 V; 0 to
20.000 V and 0.00 to 60.00 V DC. Use it to check loop power supplies, I/V converters, 1 to 5 Volt signals, and other voltages.
• Save time calibrating recorders and monitors in 4 to 20 mA loops
Other brands of calibrators require that you break the loop to calibrate voltage inputs. With the PIE 235 you simply clip the test leads across the voltage inputs in a live 4 to 20 mA loop without disconnecting any
wires.
• Simulate pH probes into transmitters & analyzers
Use the pH simulator to verify proper operation of pH devices before you place a probe into a calibrated buffer.
• Quickly set any three outputs plus automatic stepping & ramping
Easily set any value with the adjustable “DIAL” plus store any three output settings for instant recall with the EZ-CHECK™ switch. 2, 3, 5 & 11 steps automatically increment output in 100%, 50%, 25% or 10% of span plus continuous ramp. Set step/ramp time to 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 & 60 seconds.
• Designed for you by experienced calibrator manufacturers
PIE Calibrators are designed and built by members of the same team that designed and built the calibrators manufactured by Fluke* under the Altek* label. The 235 improves upon other brands by including a rubber boot, backlit display with larger digits, higher accuracy and more ranges for flexibility.
www. .com
information@itm.com1.800.561.8187
Page 2
LO
V
2 1
1.000
+
-
q EZ-CHECK™ SWITCH
SOURCE: Instantly output two preset voltages by
moving the EZ-CHECK™ switch to the “LO” position or “HI” position.
For fast three point checks select the “SET” position. The PIE Model 235 will remember the last “SET” value, even with the power off. These values can easily be changed to suit the calibration requirements. The values stored in the HI and LO positions are also used for Auto Stepping.
READ: Slide the switch to the SET position. The PIE Model 235 will display the current voltage from the process output. Slide the switch to HI and the highest voltage measured since turn-on or reset will be displayed; slide the switch to LO and the lowest voltage measured since turn-on or reset will be displayed.
w SOURCE/OFF/READ Switch
Select “SOURCE” to output in volts, millivolts, % of 1 to 5 Volts or pH. Select “READ” to read volts or millivolts. Select “OFF” to turn the unit off.
e EZ-DIAL™ KNOB
SOURCE: Turn the knob to adjust the output
level. Turn clockwise to increase the output, counter clockwise to decrease the output in
0.001 V, 0.01 mV, 0.1% or 0.001 pH steps at a time. Push down and turn the EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store desired EZ-Check™ HI/LO points in SOURCE mode. Continue to press and hold the knob for two more seconds to start the automatic stepping or ramping.
READ: Press and hold to transfer the current temperature into the EZ-Check™ HI/LO points. This clears the HI/LO voltage readings which will update as the input signal changes.
Double click the knob to get into the PIE Model 235 Configuration Mode. Use configuration to select voltage or pH, Range, Backlight On/Off, Step Size, Step Time and Auto Off On/Off.
CHANGING BATTERIES
Low battery is indicated by a battery symbol on the display. Approximately one to four hours of typical operation remain before the PIE Model 235 will automatically turn off. To change the batteries; remove the rubber boot, remove the battery door from the back of the unit by sliding the door downward. This allows access to the battery compartment. Replace with four (4) “AA”
1.5V batteries being careful to check the polarity. Replace the battery door and replace the boot. All stored configuration options (Range, EZ-CHECK Memories, etc.,) are reset to factory settings when the batteries are removed.
Note: Alkaline batteries are supplied and recommended. Purchase the optional Ni-MH rechargeable batteries for maximum battery life.
Basic Operation
The PIE Model 235 has two banana jacks mounted in the top end of the housing.
Connections
2
-
+
1
www. .com
information@itm.com1.800.561.8187
Page 3
Configuration
Configure the Calibrator
Move w POWER SWITCH to “SOURCE” or
“READ”. Setup
Double click the
e DIAL KNOB at any time the
unit is on and the following displays will appear for 15 seconds: Turn the
e DIAL KNOB to move through the two
pages of menus. Press the e DIAL KNOB to toggle between OFF and ON or to scroll through the settings.
EXIT - exits this menu immediately and saves any changes. Menu will automatically exit after 15 seconds of inactivity).
FUNCTION - pressing the knob will toggle between V and pH.
RANGE - pressing the knob will cycle through the Voltage ranges of 20V, 5V, 999mV and 60V.
UNITS - pressing the knob will toggle between V and % only in the 5 V RANGE. The 5 V RANGE is preset with EZ-CHECK LO and HI at 1.000 (0.0%) and 5.000 V (100.0%).
DOUBLE CLICK
EZ-DIAL KNOB
FOR CONFIGURATION
V#.##
AUTO OFF - If AUTO OFF is ON, the unit will
turn off after 30 minutes of inactivity to save battery life. If AUTO OFF is OFF the unit will stay on until the POWER SWITCH is moved to the off position.
BACKLIGHT - If BACKLIGHT is ON the backlight will light all the time the unit is powered up. For maximum battery life turn the backlight off when using the calibrator in areas with enough ambient light to read the display.
STEPS/RAMP - pressing the knob will cycle through 2, 3, 5, 11 and RAMP. The endpoints of the steps or ramp are based on the values stored in the HI and LO EZ-CHECK outputs.
2 steps will automatically switch between the values stored in the HI & LO EZ-CHECK (0 & 100%).
3 steps between the HI, Midpoint and LO EZ-CHECK (0, 50 & 1
00%).
5 steps between the HI and LO EZ-CHECK in 25% increments (0, 25, 50, 75 & 100%).
11 steps between the HI and LO EZ-CHECK in 10% increments (0, 10, 20...80, 90 &100%).
RAMP continuously ramps up and down between the HI and LO EZ-CHECK outputs.
STEP/RAMP TIME - pressing the knob will cycle through 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds.
Note: All settings are remembered even with the power off. Removing the batteries resets the values to factory defaults.
MAIN
> EXIT (1/2)
FUNCTION V (pH only in SOURCE) RANGE
20V 5V, 999mV, (60V READ Only)
UNITS V % (Only in 5V RANGE)
FEATURES
> EXIT (2/2) AUTO OFF ON OFF BACKLIGHT ON OFF STEPS/RAMP 2 3 5 11 RAMP STEP/RAMP TIME 5 6 7 8 9 10 15 20 25 30 60 (STEP/RAMP settings only in SOURCE)
www. .com
information@itm.com1.800.561.8187
Page 4
SOURCE
Choose this function to provide an output from -500.00 to 999.00 mV, from 0.000 to 20.000 V, from
1.000 to 5.000 V or from 0.0% to 100.0% of 1 to 5 Volts. The source current is a nominal 20 mA to provide the driving power to your voltage receivers.
Move the power switch
w to SOURCE then Double Click the EZ-DIAL knob to get into the Menu.
Turn the knob to scroll through the settings and press the knob to make your selection. Select V for the FUNCTION and 999 mV, 20 V or 5 V for the RANGE.
Connect the output leads of the PIE Model 235 to the inputs of the device being calibrated, making sure to check polarity. Red lead to the plus (+) input and black lead to the minus (-) input.
Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI and LO. You may also select any third output setting (such as mid-range) using the SET position on the EZ-CHECK switch. The output is adjusted in 0.001 V, 0.01 mV or 0.1% increments by turning the knob
e. Press and turn the knob for faster dialing with 0.100 V, 1.00 mV or 10.0% increments.
The red OVERLOAD indicator will light, a beeper will sound, and the display will flash if excessive voltage or current is detected by the calibrator.
Sourcing DC Volts & Millivolts
Voltage Receiver Input
Controller
Transmitter
Computer
Recorder or Logger
DCS
2 1
+
-
HI
5.000V
www. .com
information@itm.com1.800.561.8187
Loading...
+ 8 hidden pages