PHYWE Cobra4 Sensor-Unit Tesla Operating Instructions Manual

Operating instruct
i
Cobra4 Sensor-Unit Tesla, ± 1 Tesla
guidel
ines.
12652-00
PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 D-37079 Göttingen
Phone +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de
The unit complies with the corresponding EC
CONTENTS
Fig. 1: 12652-00 Cobra4 Sensor-Unit Tesla.
1 SAFETY PRECAUTIONS
2 PURPOSE AND CHARACTERISTICS
3 FUNCTIONAL AND OPERATING ELEMENTS
4 NOTES ON OPERATION
5 HANDLING
6 TECHNICAL SPECIFICATIONS
7 PARTS SUPPLIED
8 ACCESSORIES
9 NOTES ON THE GUARANTEE
10 WASTE DISPOSAL
Read these operating instructions carefully and com- pletely before putting your Cobra4 Sensor-Unit Tesla into operation. In addition, also similarly read the operating instructions for the transmission device (Cobra4 Wire­less-Link, Cobra4 USB-Link or Cobra4 Mobile-Link) that is to be used for the acquisition of measured values be­fore operating the Sensor-Unit. With these measures you protect yourself and your equipment.
Only use the instrument for the purpose for which it was designed.
Only use the instrument in dry rooms in which there is no risk of explosion.
Protect your Cobra4 Sensor-Unit Tesla from dust, mois- ture and vapours. Clean it with a slightly moist, lint-free cloth. Aggressive cleaning agents and solvents are un­suitable.
Take care that no liquid penetrates in through the open- ings in the housing, as the Cobra4 Sensor-Unit Tesla would then suffer damage.
When using the Sensor-Unit outdoors, ensure that nei- ther rainwater nor spray penetrate into the housing.
Only use the experimental set-up for the use for which it is intended.
Do not open the Sensor-Unit..
Only connect the data output of the Cobra4 Sensor-Unit
The present-day state of software development makes it
Tesla to one of the three measured data recording in­struments that are intended for it.
1
www.phywe.com, © All rights reserved 12652-00 / 3611
impossible to guarantee that a product is free of faults. PHYWE Systeme GmbH & Co. KG therefore does not take on any liability for damages that occur during the installation or the use of the instrument.
2 PURPOSE AND CHARACTERISTICS
Cobra4-Unit Tesla serves to measure the magnetic field strengths of constant and alternating fields in combination with a tangential or axial Hall probe.
According to the application, Cobra4 Sensor-Unit Tesla can be connected to one of the following measured data re­cording instruments and pass measured data to it:
to Cobra4 Wireless-Link (12601-00) for wireless trans- mission to a computer,
to Cobra4 USB-Link (12610-00) for data transmission to a computer via a USB-cable,
to the hand-held measuring instrument Cobra4 Mobile- Link (12620-00).
The interface to measured data recording instruments is on the front side of the Sensor-unit, whereby mechanically se­cure engagement is ensured by a locking mushroom and a hole.
For measurements of magnetic field strengths, a tangential (13610-02) or an axial (13610-01) Hall probe can be con­nected to the Sensor-Unit, according to the application. It has a 5-pin DIN socket for connection of the Hall probe.
In combinations of the Sensor-Unit with USB-Link (12610.00), EM-disturbances can end the measurement pro­gramme. A restart (reset) of the device must then made by unplugging and re-plugging the USB-cable. Measurement must then be started anew in the “Measure” programme.
In extreme cases, USB-communication can be re-established by restarting the computer.
5 HANDLING
This section describes how to put the Sensor-Unit into opera­tion and record measured data. Please read carefully through it to avoid failure or faulty operation.
5.1 Putting the Cobra4 Sensor-Unit Tesla into operation
Check that the current version of the measure software package and the Cobra4 module are installed.
Updating of the software is free via: www.phywe.com… software … measure measure main programme measure Cobra4 module Free updating of the software can also be made with the measure DVD.
Connect your Cobra4 Sensor-Unit Tesla to one of the three measured data recording instruments via the engaging 15-pin plug connection; the Sensor-Unit is ready to use after a brief starting phase (self-calibration), as voltage is supplied by the measured data recording instrument. Further information on the control of the measured data re­cording instruments is given in their operating instructions..
3 FUNCTIONAL AND OPERATING ELEMENTS
3.1 Operating elements
Cobra4 Sensor-Unit Tesla has no manual operating ele­ments. Operation of it is made via the Cobra4 Mobile-Link hand-held measuring instrument, or via a computer to which either a Cobra4 USB-Link or a Cobra4 Wireless-Link has been connected, as soon as it has been connected to the se­lected one of the three allowable measured data recording instruments.
4 NOTES ON OPERATION
Together with Cobra4–Wireless-Link, Cobra4–USB-Link and Cobra4–Mobile-Link, Cobra4–Sensor-Unit Tesla fulfils the technical requirements that are compiled in current European Community Guidelines. The characteristics of this product qualify it for the CE mark.
This instrument is only to be operated under specialist super­vision in a controlled electromagnetic environment in re­search, educational and training facilities (schools, universi­ties, institutes and laboratories). This means that in such an environment no mobile phones etc. are to be used in the immediate vicinity. The individual connecting leads are each not to be longer than 2 m.
The instrument can be so influenced by electrostatic charges and other electromagnetic phenomena that it no longer func­tions within the given technical specifications. The following measures reduce or do away with disturbances: Avoid fitted carpets; ensure potential equalization; carry out experiments on a conductive, earthed surface, use screened cables, do not operate high-frequency emitters (radios, mo­bile phones) in the immediate vicinity.
6 TECHNICAL SPECIFICATIONS
(Typical for 25°C) Operating temperature range 5...40°C
Relative humidity < 80%
Measuring range, constant field
- ± 1 T ; Resolution ± 1 mT
- ± 100 mT ; Resolution ± 100 µT
- ± 10 mT ; Resolution ± 10 µT
- Each measurement range can be
compensated up to the final value of the particular measurement range.
- Accuracy approx. ± 2 % of measure- ment range final value
Measuring range, alternating field from 15 Hz to 1 kHz
- 1 T ; Resolution 1 mT
- 100 mT ; Resolution 100 µT
- 10 mT ; Resolution 10 µT
- Alternating fields are measured with a
root-mean-square rectifier (RMS)
- No measurement range compensation
- Accuracy approx. ± 5 % of measure-
ment range final value
Scan rate, maximum (Hz) 5 Dimensions (L x W x H in mm) 60 x 70 x 35 Weight (g) 100
2
www.phywe.com, © All rights reserved 12652-00 / 3611
7 PARTS SUPPLIED
The extent of delivery is:
Cobra4 Sensor-Unit Tesla 12652-00
Operating instructions
8 ACCESSORIES
Hall probe, axial 13610-00 Hall probe, tangent., prot. cap 13610-02
9 NOTES ON THE GUARANTEE
We guarantee the instrument supplied by us for a period of 24 months within the EU, or for 12 months outside of the EU. Excepted from the guarantee are damages that result from non-observance of the operating instructions, from improper handling of the instrument or from natural wear. The manufacturer can only be held responsible for the func­tion and technical safety characteristics of the instrument, when maintenance, repairs and alterations to the instrument are only carried out by the manufacturer or by personnel who have been explicitly authorized by him to do so.
10 WASTE DISPOSAL
The packaging consists predominately of environmentally compatible materials that can be passed on for disposal by the local recycling service.
PHYWE Systeme GmbH & Co. KG Abteilung Kundendienst (Customer Service) Robert-Bosch-Breite 10 D-37079 Göttingen
Phone +49 (0) 551 604-274 Fax +49 (0) 551 604-246
Should you no longer require this product, do not dispose of it with the household refuse.
Please return it to the address below for proper waste disposal..
3
www.phywe.com, © All rights reserved 12652-00 / 3611