PHYWE Cobra4 Oxygen Operating Instructions Manual

Operating Instru
with the applicable
EC regulations
PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 37079 Göttingen
Germany
Tel. +49 (0) 551 604-0 Fax +49 (0) 551 604-107 E-mail info@phywe.de Internet www.phywe.de
Cobra4 Sensor-Unit Oxygen
12676-00
The unit complies
Fig. 1: 12676-00 Cobra4 Sensor-Unit Oxygen
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS
2 PURPOSE AND CHARACTERISTICS
3 FUNCTIONAL AND OPERATING ELEMENTS
4 OPERATING INSTRUCTIONS
5 HANDLING
6 TECHNICAL DATA
7 SCOPE OF SUPPLY
8 ACCESSORIES
8 WARRANTY
9 DISPOSAL
1 SAFETY INSTRUCTIONS
Read the operating instructions thoroughly and com-
Only use the unit for its intended purpose.
The unit is intended solely for use in dry rooms where
Protect the Cobra4 Sensor-Unit Oxygen against dust,
Ensure that no liquid can penetrate the housing. Other-
When using the unit outdoors, ensure that no rain or
Do not open the unit.
Connect the data output of the Cobra4 Sensor-Unit Oxy-
Based on today's state of software development, it is im-
pletely prior to starting the unit. Please read also the op­erating instructions of the measurement data recording device that is used (Cobra4 Wireless-Link, USB-Link, Mobile-Link, or Junior-Link). This is important for your own protection and for avoiding damage to the unit.
there is no risk of explosion.
moisture, and vapours. Clean the unit with a slightly moist, lint-free cloth. Do not use any aggressive cleaning agents or solvents.
wise, the Cobra4 Sensor-Unit Oxygen may be damaged.
splash water can penetrate the housing.
gen only to the intended measurement data recording devices.
possible to guarantee that a product is completely free from errors or defects.
This is why PHYWE Systeme GmbH & Co. KG does not accept any liability for damage that results from the in­stallation or use of this unit.
1
www.phywe.de, © All rights reserved 12676-00 / 4513
2 PURPOSE AND CHARACTERISTICS
The Cobra4 Sensor-Unit Oxygen is used to measure the oxygen content in the gaseous state and the oxygen concen­tration in dissolved form. Depending on the type of application, the Cobra4 Sensor­Unit Oxygen can be connected to one of the following meas­urement data recording devices for the transfer of the meas­urement data:
Cobra4 Wireless-Link (12601-00) for the wireless data transfer to a computer,
Cobra4 USB-Link (12610-00) for the data transfer to a computer via a USB cable,
Cobra4 Junior-Link (12615-00) for the data transfer via USB during student experiments,
hand-held Cobra4 Mobile-Link (12620-10).
The front of the unit is equipped with an interface via which the connection to the measurement data recording devices is realised. Mechanically secure locking is ensured by way of the lock knob and the corresponding hole.
3 FUNCTIONAL AND OPERATING ELEMENTS
3.1 Operating elements
The Cobra4 Sensor-Unit Oxygen does not have any manual operating elements. It is controlled by way of the hand-held Cobra4 Mobile-Link or via a computer if it is connected to a Cobra4 USB-Link, Cobra4 Wireless-Link, or Cobra4 Junior­Link, as soon as the connection between the Sensor-Unit and one of the three measurement data recording devices has been established.
3.2 Measurement inputs
The front of the unit is equipped with a connector for connect­ing the oxygen probe.
5 HANDLING
This section describes the start-up of the Cobra4 Sensor-Unit Oxygen and the recording of measurement data. Please read this section thoroughly in order to avoid failures or operating errors.
5.1 Starting the Cobra4 Sensor-Unit Oxygen
Software/device software update Ensure that the current versions of the "measure" software package and of the firmware of the Cobra4 module are in­stalled. Cost-free software update at:
www.phywe.com > Downloads > Software:
"measure" measurement software
PHYWE firmware A cost-free update of the software is also possible via the "measure" DVD.
Topping-up the electrolyte solution The lower end of the electrode can be unscrewed. This cap covers the actual electrode. The entire cavity between the cap and electrode must be filled with the electrolyte liquid (KCl). Ensure that no air bubbles are entrapped. The elec­trode comes supplied in an unfilled state. In addition, the liq­uid evaporates gradually. This is why the liquid must be topped up from time to time.
1. Unscrew the lower end.
2. Fill the cap with the liquid.
4 OPERATING INSTRUCTIONS
Combined with the Cobra4 Wireless-Link, Cobra4 USB-Link, Cobra4 Junior-Link, and Cobra4 Mobile-Link, the Cobra4 Sensor-Unit Oxygen fulfils the technical requirements that are summarised in the current directives the of the European Community. The unit must be used under the supervision of an expert and in an electromagnetically controlled environ­ment at research, teaching, and training facilities (schools, universities, institutes, and laboratories). This means that, in such an environment, radio transmission devices, e.g. mobile phones, should not be used in the direct vicinity of the unit. The connected cables must not be longer than 2 m. Electrostatic charges or similar electromagnetic phenomena (HF, bursts, indirect lightning discharge, etc.) may affect the unit so that it will not work within the specified data range. The following measures reduce or eliminate the potential in­terferences: avoid carpets; provide equipotential bonding; perform the ex­periments on a conductive, earthed surface, use shields and shielded cables. Do not use radiofrequency transmitters (ra­dio sets, mobile phones) in the direct vicinity of the unit.
3. Screw the filled cap back onto the electrode. Use a cloth to wipe off any liquid that leaks out.
Preparation Connect the oxygen probe (see the list of accessories) to the connector at the front of the Sensor-Unit. Connect the Co­bra4 Sensor-Unit Oxygen with one of the four possible measurement data recording devices via the 15-pin snap-on connector. The Sensor-Unit is immediately ready for opera­tion since it is supplied with power via the measurement data recording device. Please refer to the corresponding operating instructions for more information concerning the operation and control of the measurement data recording devices.
Calibration It is not necessary to recalibrate the sensor for every meas­urement. Especially if only the change in oxygen content is to be demonstrated, an already existing calibration (the last calibration remains automatically stored in the sensor) or the factory settings are usually sufficient. However, if absolute values are required for your measurement, e.g. if you are de­termining the oxygen content of bodies of water, or if you want to optimise the accuracy of your measurement, we rec­ommend calibrating the sensor. See section 5.3.
2
www.phywe.de, © All rights reserved 12676-00 / 4513
5.2 Recording of measurement data
Measurement: Two types of measurements can be performed with the Sen­sor-Unit Oxygen: Measurement of the oxygen content, e.g. in the air (% by volume), and a measurement of the oxygen concentration in an aqueous solution (mg/l or %).
a) Measurement of oxygen in the air
Remove the protective cap and select the measuring
channel "Gas". It takes approximately 1 minute for the value to settle.
b) Measurement of dissolved oxygen
Remove the protective cap and select the measuring channel "Fluid".
Plunge the oxygen probe into the solution so that the temperature sensor is also covered (approxi­mately. 5-6 cm). Do not immerse it completely in the liquid. The handle is not waterproof.
Temperature sensor
Record the oxygen concentration by starting a measurement via the software/Mobile-Link or read the current values off of the display/user interface
a) With the Cobra4 Mobile-Link:
Select "Menu", "Calibration", "One point" and confirm by ticking.
b) With the Wireless-Link, USB-Link, or Junior-Link and via
the "measure" software In the navigator, double-click the channel "Concentra­tion", open the tab "Calibration", and then click "Apply".
2. Dissolved oxygen
In this case, a one-point-calibration will be performed in which the electrode is held in air that is saturated with water (see "Info" at the end of the paragraph). This corresponds to a saturated solution of the oxygen in water.
Procedure Select the measuring channel "Fluid". Fill the (supplied) calibration bottle with approximately 5 ml of distilled water. Push the oxygen electrode through the seal of the screw cap and screw the cap on tightly. Attention, the sensor must not plunge into the fluid. Then, shake the bottle with the electrode gently for approximately 10 seconds in or­der to saturate the air in the calibration bottle with water va­pour. Under these conditions, the oxygen content is assumed to be 100% saturation. After one minute, there should be an equilibrium (constant measurement value) and the calibration can be performed (either with the Mobile-Link, see c), or with another device and the software, see d)).
of the software.
Ensure that the water sample that is to be ana- lysed is slightly stirred in a continuous manner. This is important since the sensor consumes oxy­gen during the measurement. Without stirring, it would seem as if the oxygen concentration de­creases.
Measuring principle: This sensor type is an amperometric Clark sensor with a platinum cathode and a silver anode. A fixed voltage is applied at the platinum electrode. When oxygen diffuses through the membrane and to the cathode, the latter will be reduced:
1/2 O2 + H2O + 2
e–
2 OH–
The reference electrode (anode) is subject to oxidation:
Ag + Cl– → AgCl + e–
Correspondingly, a current flows and this current is propor­tional to the oxygen diffusion and, thereby, to the oxygen concentration. The current is converted into a voltage signal that is then received by the sensor.
5.3 Calibration
Ensure to perform the calibration at a temperature that is similar to the temperature of the sample. Like in the case of a measurement, wait for approximately 1-2 minutes after the connection of the sensor until the measurement value has settled. Then, perform the calibration.
1. Oxygen in air
Select the measuring channel "Gas". Hold the sensor in the ambient (outdoor) air since, in this case, the oxygen content will be constantly at 20.9%. De­pending on the Link device that you use, follow the steps un-
Fig. 2: Calibration
c) With the Cobra4 Mobile-Link:
Select "Menu", "Calibration", "One point" and confirm by ticking.
d) With the Wireless-Link, USB-Link, or Junior-Link and via
the "measure" software In the navigator, double-click the channel "Concentra­tion", open the tab "Calibration", and then click "Apply".
Info: The calibration in water-saturated air corresponds to the calibration in a saturated solution since the oxygen in the electrode first passes through the electrolyte solution before it reaches the cathode. The oxygen concentration in water­saturated air and the oxygen concentration in the electrolyte solution will be the same. Depending on the partial pressure of the oxygen in the air, it will correspond to Henry's law. This relationship applies to the sensor in air and to the sensor in the solution.
der a) or b) respectively.
3
www.phywe.de, © All rights reserved 12676-00 / 4513
5.4 Storage
Short-term: Leave the electrolyte solution in the cap and place the elec­trode in distilled water in between the measurements. If pos­sible, leave the measuring instrument switched on.
Several days: Rinse the outside of the electrode with distilled water, dry it, and put on the protective cap. This prevents the electrolyte solution from evaporating.
Long-term: Unscrew the cap and remove the electrolyte solution com­pletely. Clean the electrode and the cap with distilled water and dry both. Shake the water off the membrane. Screw the dry cap loosely onto the electrode (do not close it com­pletely!). Store in a dry and light-protected place.
5.5 Care
Membrane Ensure not to damage the membrane and follow the storage instructions in order to increase the service life of the elec­trode. Creases or holes in the membrane are a sign that it does not work properly any more. If the membrane is soiled or if it shows signs of damage (this might be indicated by incorrect measurement values), use the supplied spare cap with a membrane.
Cathode and anode They must be polished with the supplied polishing cloths from time to time. To do so, remove the cap so that the anode and cathode can be accessed. Polish them with slight pressure and a clean cleaning cloth (included in the scope of supply) in order to remove any deposits from the surface. Rinse them with distilled water and then with alcohol, shake the liquid off, and screw the cap back on loosely.
6 TECHNICAL DATA
(typical data for 25°C) Operating temperature range: 5 to 40°C Rel. humidity: < 80%
O2, dissolved Measuring range 0...20 mg/l
0...200% Accuracy 1%
Resolution 0.01 mg/l Max. sampling rate 100 Hz O2, gaseous Measuring range: 0...30 % Accuracy 1%
Resolution 0.1% Max. sampling rate 100 Hz
Connector: Sub-D, 15-pin-type Dimensions (L x W x H) 110 x 70 x 30 mm Weight: 100 g
7 SCOPE OF SUPPLY
The following items are included in the scope of supply:
Cobra4 Sensor-Unit Oxygen 12676-00
8 ACCESSORIES
For information concerning the use of the individual ac­cessories, see the section 5.1 to 5.4.
Oxygen probe 12676-11 including 50 ml of the electrolyte liquid, a protective cap, a
calibration bottle, a spare membrane, and cleaning strips for the cathode and anode
Electrolyte liquid (KCL-solution)
8 WARRANTY
We give a warranty of 24 months for units supplied by us in­side the EU, and a warranty of 12 months outside the EU. The following is excluded from the warranty: Damage that is due to non-compliance with the operating instructions, im­proper use, or natural wear.
The manufacturer can only be held liable for the function and safety-relevant properties of the unit, if the maintenance, ser­vice, and modifications of the unit are performed by the manufacturer or by an institution that is expressly authorised by the manufacturer.
9 DISPOSAL
The packaging mainly consists of environmentally-friendly materials that should be returned to the local recycling sta­tions.
Do not dispose of this product with normal household waste. If this unit needs to be disposed of, please return it to the address that is stated below for proper disposal.
PHYWE Systeme GmbH & Co. KG Customer Service Robert-Bosch-Breite 10 37079 Göttingen Germany
Tel. +49 (0) 551 604-274 Fax +49 (0) 551 604-246
4
www.phywe.de, © All rights reserved 12676-00 / 4513
Loading...