36
9.9 Electromagnetic environmental declaration
Guidance and manufacturer’s declaration - electromagnetic emissions
The LUCAS device is intended for use in the electromagnetic environment specified below. The customer or the
operator of the device must make sure that it is used in the correct environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 The LUCAS device uses limited radio frequency
energy (Bluetooth) only during data transmission
after use. This makes its radio frequency emissions
low and not likely to cause interference with other
electronic equipment near the LUCAS device.
RF emissions CISPR 11 Class B The LUCAS device is suitable for use in all buildings
including domestic homes and places directly
connected to the public low-voltage Power Supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
The LUCAS device is intended for use in the electromagnetic environment specified below. The customer or the
operator of the device must make sure that it is used in the correct environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors must be wood, concrete
or ceramic tile. If there is
synthetic material on the floor,
the relative humidity must be
30% or more.
Electrical fast
transient / Burst
IEC 61000-4-4
+/- 2 kV for Power Supply lines
+/- 1 kV for input/output lines
+/- 2 kV for Power Supply
lines
n/a. for input/output lines
The mains power quality must
be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV differential mode
+/- 2 kV common mode
+/- 1 kV differential mode
n/a. for common mode
The mains power quality must
be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
Power Supply input
lines IEC 61000-4-11
<5 % UT (>95 % dip in UT) for
0.5 cycle
40 % UT (60 % dip in UT) for
5 cycles
70 % UT (30 % dip in UT) for
25 cycles
<5 % UT (>95 % dip in UT)) for
5 sec
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
The mains power quality must
be that of a typical commercial
or hospital environment. If the
user of the device requires
continued operation during
power mains interruptions,
Jolife AB recommends that
the device is energized from a
Power Supply or Battery that
cannot be interrupted.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m 30 A/m The power frequency magnetic
fields must be at levels that
are characteristic of a typical
location in a typical commercial
or hospital environment
NOTE: UT is the a.c. mains voltage prior to application of the test level.
The following Essential performance was applied for EMC testing (IEC/EN 60601-1-2: 2007):
The EUT shall continuously perform compression at the intended rate.