Physio-Control LUCAS User Manual

Chest Compression System
Instructions for Use
US version
100057-00 F, VALID FROM CO J2256

Important user information

All users must read and fully understand the entire Instructions for Use before operating LUCASChest Compression System.
The Instructions for Use should always be easily accessible to the users of LUCAS.
Always follow local and/or international guidelines for CPR when using LUCAS.
The use of other medical equipment or drugs in conjunction with external chest compressions may reduce the effect of compressions. Always consult the instructions for use for other equipment or drugs to check that they are appropriate to use in conjunction with the treatment and compressions being performed with LUCAS.
When LUCAS is used on a non intubated patient the device can safely be stopped to allow intermittent manual ventilation. Switch off the device by turning the ON/OFF knob to the Lock position, ventilate and turn the ON/OFF knob back to the Active position again. See Section 1.5 on page 7 for information about the ON/OFF knob positions.
REQUIRED SKILLS
JOLIFE AB strongly recommends that LUCAS Chest Compression System is only used by persons with basic medical skills such as:
First responders, ambulance personnel, nurses, physicians or medical staffs, who have:
completed a CPR course according to the American Heart Association or equivalent, AND
received training in how to use LUCAS
LUCAS can only be bought by or on the order of a licensed medical practitioner.
TRAD EMA RKS
LUCASTM is a trademark of JOLIFE AB.
DECLARATION OF CONFORMITY
LUCAS Chest Compression System complies with the requirements of the European Medical Device 93/42/EEC. It is marked with the CE-symbol:
MANUFACTURER, MAIN OFFICE
JOLIFE AB Scheelev. 17 SE-223 70 LUND Sweden Tel: +46 46 286 50 00
www.jolife.com www.lucas-cpr.com
DISTRIBUTOR, USA
Medtronic Emergency Response Systems 11811 Willows Road NE P.O. Box 97006 Redmond, WA 98073-9706 USA Tel: 425.867.4000 Fax: 425.867.4121 www.medtronic-ers.com
LUCAS Chest Compression System, Instructions for Use
ENGLISH (US)
Art. No. 100057-00 F, Valid from CO J2256 © Copyright J
OLIFE AB 2007. All rights reserved.
2

Table of Contents

IMPORTANT USER INFORMATION.......................................................................................... 2
1I
1.1 INTENDED USE ...............................................................................................................................5
1.2 CONTRAINDICATIONS..................................................................................................................5
1.3 LUCAS CHEST COMPRESSION SYSTEM...................................................................................5
1.4 LUCAS COMPONENTS .............................................................................................................. 6
1.5 ON/OFF-KNOB, POSITIONS ......................................................................................................7
1.6 DESCRIPTION................................................................................................................................8
1.7 SYMBOLS ON THE DEVICE........................................................................................................... 9
1.8 DELIVERED ITEMS ........................................................................................................................10
1.9 THE LUCAS TEAM.....................................................................................................................10
1.10 BACKGROUND ..............................................................................................................................11
1.11 CHEST COMPRESSIONS USING LUCAS....................................................................................11
1.12 SIDE EFFECTS ................................................................................................................................11
NTRODUCTION..............................................................................................................5
2WARNINGS AND PRECAUTIONS................................................................................. 12
2.1 ASSEMBLY....................................................................................................................................12
2.2 USING LUCAS CHEST COMPRESSION SYSTEM ....................................................................12
2.3 CONNECTION TO AIR SUPPLY................................................................................................... 13
2.4 ADJUSTING LUCAS TO THE PATIENT ..................................................................................... 13
2.5 HANDLING LUCAS CHEST COMPRESSION SYSTEM ............................................................ 13
3USING LUCAS ............................................................................................................14
3.1 ARRIVAL AT THE PATIENT .........................................................................................................14
3.2 UNPACKING AND CONNECTING THE AIR .................................................................................14
3.3 ASSEMBLY....................................................................................................................................16
3.4 ADJUSTMENT ...............................................................................................................................17
3.5 OPERATING LUCAS CHEST COMPRESSION SYSTEM ..........................................................18
3.6 DEFIBRILLATION ..........................................................................................................................19
3.7 TRANSPORTING THE PATIENT...................................................................................................19
3.8 CHANGING AIR SOURCES...........................................................................................................21
3.9 REMOVING LUCAS FROM THE PATIENT................................................................................22
4CARE AFTER USE......................................................................................................... 22
4.1 CLEANING ROUTINES.................................................................................................................22
4.2 ROUTINE CHECKS....................................................................................................................... 23
4.3 STORAGE .................................................................................................................................... 23
4.4 SERVICE ....................................................................................................................................... 23
5PACKING AWAY THE DEVICE .....................................................................................24
6T
A
WEEKLY CHECK AND CHECK AFTER USE, LUCAS............................................................................ 28
100057-00 F, ©JOLIFE AB 2007 3 LUCASTM CHEST COMPRESSION SYSTEM INSTRUCTIONS FOR USE
ECHNICAL SPECIFICATION....................................................................................... 25
PPENDIX A ......................................................................................................................... 28
APPENDIX B - LUCAS STABILIZATION STRAP................................................................ 29
INSTRUCTIONS FOR USE ..................................................................................................................... 29
1INTRODUCTION...........................................................................................................30
1.1 OVERVIEW ..................................................................................................................................30
1.2 CONTRAINDICATIONS............................................................................................................... 30
1.3 ASSOCIATED DOCUMENTATION ............................................................................................. 30
1.4 LUCAS STABILIZATION STRAP.............................................................................................. 30
1.5 DESCRIPTION .............................................................................................................................. 30
1.6 THE LUCAS TEAM..................................................................................................................... 31
2PREPARING LUCAS STABILIZATION STRAP FOR USE............................................ 32
3W
ARNINGS AND PRECAUTIONS ................................................................................32
3.1 USING LUCAS STABILIZATION STRAP ................................................................................. 32
4USING LUCAS STABILIZATION STRAP ................................................................... 33
4.1 INTRODUCTION...........................................................................................................................33
4.2 ATTACHING LUCAS STABILIZATION STRAP........................................................................33
5CARE AFTER USE......................................................................................................... 35
5.1 CLEANING ROUTINES ................................................................................................................ 35
5.2 ROUTINE CHECKS....................................................................................................................... 35
5.3 STORAGE .................................................................................................................................... 35
6PACKING AWAY THE STABILIZATION STRAP ..........................................................36
7T
A
WEEKLY CHECK AND CHECK AFTER USE, LUCAS STABILIZATION STRAP .................................. 37
ECHNICAL SPECIFICATION .......................................................................................36
PPENDIX C.......................................................................................................................... 37
4 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

1 Introduction

1.1 INTENDED USE

LUCAS Chest Compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
LUCAS must only be used in cases where manual chest compression would be used. LUCAS is only intended for temporary use.

1.2 CONTRAINDICATIONS

Do NOT use LUCAS Chest Compression System in the following cases:
Too small adult patient: The suction cup is not
being completely compressed when it is lowered as far as possible.
Too large patient: The support legs of LUCAS
cannot be locked to the back plate without compressing the patient.
Patient is a child.
Patient with traumatic injury (wounds resulting
from sudden physical injury or violence).
Pregnant patients. The woman has to lie 10–15°
to one side to prevent vena cava syndrome (impaired venous return to the heart as the uterus compresses the inferior vena cava).
If there is no indication that chest compression
is likely to help the patient.

1.3 LUCAS CHEST COMPRESSION SYSTEM

LUCAS is a portable device for external cardiac compressions. It is stored and easily carried in a backpack, giving the user free hands until the device is needed.
The components of LUCAS are shown in the illustration in Section 1.4.
100057-00 F, ©JOLIFE AB 2007 5 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

1.4 LUCAS COMPONENTS

q
e
w
i
h
r
o
t
y
u
1. ON/OFF Knob
2. Hood
3. Patient strap
4. Release ring
5. Support leg
6. Claw lock
7. Back plate
8. Hose attachment
9. Bellows
10. Suction cup
11. Height adjustment handle
12. Air hose
13. LUCAS unique connector
14. Pressure pad
15. Upper part
s
a
g
f
d
6 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
1.5 ON/OFF-KNOB,
POSITIONS
1. Adjust: The suction cup
can be adjusted to fit the patient, using the height adjustment handles.
2.
Lock: The compression
mechanism is locked. This is used during defibrillation and when moving the patient.
3.
Active: This is the
operating position. When LUCAS Chest Compression System is connected to an air supply of specified capacity, it performs compressions at a rate of about 100 per minute.
100057-00 F, ©JOLIFE AB 2007 7 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
t
h

1.6 DESCRIPTION

1.6.1 MAIN PART S

LUCAS consists of an upper part (h) and a back plate (
u). The back plate is placed underneath the
patient to form a support for the external chest compressions.
The upper part contains a pneumatically driven piston rod, which acts on the patient’s chest via a pressure pad. The pressure pad is surrounded by a suction cup.
u
j
The support legs ( to the back plate prior to starting compressions.
t) of the upper part are fastened

1.6.2 CONNECTION TO AIR SUPPLY

LUCAS is powered by compressed air from a wall outlet or a cylinder. See Section 6 for specification of air sources.
The air hose is permanently mounted on LUCAS, and has a unique male connector at the open end. A suitable pressure regulator is delivered with LUCAS.
LUCAS requires no electrical supply and has no conducting parts on the outside, except the hose attachment, the claw lock bar, and the upper attachment of the bellows.

1.6.3 LUCAS STABILIZATION STRAP

LUCAS Stabilization Strap (j) is an accessory which is attached to LUCAS to prevent downward movement of LUCAS during operation. Please see Appendix B for instructions for use of LUCAS Stabilization Strap
.
8 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

1.7 SYMBOLS ON THE DEVICE

Symbol/ Location
Meaning
Caution – see instructions for use. All
users must read the entire Instructions for Use before operating LUCAS Chest Compression System.
Year of manufacture.
WARNING – Crushing hazard. The
pressure pad and suction cup are drawn into the bellows with high force when LUCAS is connected to an air supply. Keep all body parts away from the suction cup, the bellows and the height adjustment handles when connecting the air hose, and during operation.
WARNING – Crushing hazard. Keep
your fingers away from the claw locks when attaching the upper part to the back plate.
Caution – Do not lift by the straps.
The straps are only intended for securing the patient’s arms to LUCAS.
Place the suction cup immediately above the end of the sternum, as indicated in the figure. The suction cup should be centred over the sternum.
Pull up the release rings to release the support legs from the back plate.
100057-00 F, ©JOLIFE AB 2007 9 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
q

1.8 DELIVERED ITEMS

Before using LUCAS Chest Compression System, make sure that the following items are delivered:
1. LUCAS in carrying bag together with Instructions for Use.
2. Pressure regulator.
3. LUCAS Stabilization Strap (packed in carrying bag).

1.9 THE LUCAS TEAM

It is strongly recommended that qualified personnel work in teams of two. This enables one person to perform cardiopulmonary resuscitation (CPR) while the other person unpacks LUCAS Chest Compression System.
The Instructions for Use will refer to “The LUCAS team” (Figure have the required skills stated on page 2 .
q), consisting of two people who
10 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

1.10 BACKGROUND

1.11 CHEST COMPRESSIONS USING LUCAS

The single most common cause of death in western society is cardiovascular disease either on its own or as a contributory factor.
There is only a short space of time after the heart stops during which it is possible to reverse the situation and restore spontaneous circulation.
If the heart is not started quickly, the two most important organs, the heart and brain, will be damaged by the lack of oxygen.
Effective chest compressions can restore circulation to a level, which may be sufficient to prevent heart and brain damage during a short period of time.
For optimal results the compressions should be performed in such a way that the rib cage is held down for approximately half of the time, and subsequently released for the remaining half of the
1
.
time
It is very difficult to perform manual chest compressions effectively, even for experienced CPR practitioners, and especially in a vehicle, moving ambulance, during transport or when there is no firm surface under the patient
2 3
.
In situations where it is possible to use LUCAS Chest Compression System, many of the problems with manual chest compressions can be eliminated.
LUCAS works by pressing down very quickly on the rib cage and holding it down for 50% of the period/ cycle.
When following normal routines for CPR, the use of LUCAS has the following advantages over manual chest compression:
The device maintains consistent chest
compression over a long period of time.
One person becomes free to provide other care.
It is possible and safer for the user to give
effective chest compression even in a moving vehicle.

1.12 SIDE EFFECTS

Bruising and soreness of the chest is common during use of LUCAS External Cardiac Compressor.
The International Liaison Committee on Resuscitation (ILCOR) states the following side effects of CPR
4
:
1
American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation 2005;112: 24, Supplement; December 13, 2005.
2
Hightower D, Thomas SH, Stone CK et al. Decay in Quality of Closed­Chest Compressions Over Time. Ann Emerg Med 1995;26:300-303.
3
Sunde K, Wik L, Steen P A: Quality of Mechanical, manual Standard and active compression-decompression CPR on the arrest site and during transport in a manikin model. Resuscitation 1997;34:235-242.
Rib fractures and other injuries are common but
acceptable consequences of CPR given the alternative of death from cardiac arrest.
After resuscitation, all patients should be
reassessed and re-evaluated for resuscitation­related injuries.
4
2005 International Concensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations. Resuscitation 2005;67:195.
100057-00 F, ©JOLIFE AB 2007 11 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
2 Warnings and
precautions
Important
This section contains safety information. All users
must read this section before using LUCAS Chest
Compression System and observe the safety
information at all times during use.

2.2 USING LUCAS CHEST COMPRESSION SYSTEM

WARNING – IMPROPER USE
Improper use of the device can cause serious injury
to the patient and ineffective chest compressions.
J
OLIFE AB strongly recommends that the equipment
only be used by personnel who have undertaken
training in use and handling of LUCAS Chest
Compression System.
Sections 2.1 to 2.5 summarize warnings and precautions during assembly and use of LUCAS Chest Compression System.
Additional hazards are described in relevant sections of the Instructions for Use. All users must observe the safety information in
Cautions during use of LUCAS Chest Compression
System.
Warnings and

2.1 ASSEMBLY

WARNING – LETHAL PINCH HAZARD
If the air is connected during the assembly
procedure, and the position, the compressions may seriously injure the patient or create conditions leading to death of the
Always follow the described sequence of air
connection and assembly.
ON/OFF-knob is in the Active
patient.
Other personnel may assist those with the
necessary training in using LUCAS.
WARNING – DEVICE STATUS
If your LUCAS unit does not appear as in the
illustration in Section 1.4, or if the device does not
function in the same way as in training, or according
to specification:
Discontinue use and continue with manual chest
compressions.
WARNING – MALFUNCTION
If there is air leakage, interruption in compressions
or insufficient compressions, if the frequency is
wrong, or there is jarring, or there is something else
that cannot be classified as normal running:
Disconnect the air hose immediately, turn the
OFF
knob to Adjust and remove the device. Start
manual chest compressions.
ON/
Caution - Keep hose and connectors clean
The mechanical parts of LUCAS may be damaged if sand or soil enters the air hose, resulting in reduced
performance.
Always keep the hose and connectors clean, do not
leave these parts on the ground.
WARNING – SUCTION CUP SLIDING ON GEL
If there is gel on the patient’s chest (e.g. from
ultrasound examination), the position of the suction
cup may change during use.
Remove any gel left before applying the suction cup.
WARNING – PATIENT INJURY OR DEATH
Do not leave the patient or device unattended while
LUCAS Chest Compression System is active.
12 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

2.3 CONNECTION TO AIR SUPPLY

WARNING – INTERRUPTING AIR SUPPLY TO
LIFE SUPPORT DEVICES
If other life support equipment is disconnected from
the air supply, serious injury or death may result.
Never remove attachments to air outlets without
first checking that this will not put a patient at risk.
WARNING – EMPTY AIR SUPPLY
When the air in the cylinder runs out, the pressure
falls and compressions delivered by LUCAS
force.
Immediately start manual chest compressions and
ensure that the air cylinder is exchanged. Always
change air cylinder if the pressure gauge is on red.

2.4 ADJUSTING LUCAS TO THE PATIENT

lose
2.5 HANDLING LUCAS CHEST COMPRESSION
YSTEM
S
WARNING – CRUSHING HAZARD
Avoid placing your hands or other body parts on, or
below the suction cup, when LUCAS is active.
WARNING – CRUSHING HAZARD
Keep your fingers away from the claw locks,
particularly when lifting the patient from the floor to
the stretcher with LUCAS in position.
WARNING – INCORRECT HEIGHT
ADJUSTMENT
If the suction cup presses down too hard or too
lightly on the rib cage, the patient’s circulation will
be compromised.
Turn th e
ON/OFF knob to Adjust and adjust the
height of the suction cup immediately.
WARNING – INCORRECT POSITION
If the suction cup is not positioned correctly in
relation to the sternum, there is an increased risk of
damage to the rib cage and internal organs. In
addition, the patient’s circulation will be
compromised.
Make sure that the suction cup is positioned
according to instructions in Section 3.4 before
starting compressions.
If the position changes during operation,
immediately turn the
adjust the position of the suction cup.
ON/OFF knob to Adjust and
100057-00 F, ©JOLIFE AB 2007 13 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

3 Using LUCAS

w
e
LUCAS Chest Compression System should be used to treat patients following the same routines as for manual CPR.
This section describes the use of LUCAS. It is divided into nine steps:
1. Arrival at the patient
2. Unpacking and connecting the air
3. Assembly
4. Adjustment
5. Operating
6. Defibrillation
7. Transporting the patient
8. Changing air sources
9. Removing LUCAS from the patient

3.1 ARRIVAL AT THE PATIENT

Upon arrival at the patient, confirm cardiac arrest by determining level of consciousness, breathing and pulse (Figure
If the patient has suffered a cardiac arrest, commence cardiopulmonary resuscitation (CPR) with the LUCAS team immediately.
If CPR is already being carried out when you arrive, assist as required. Do not interrupt CPR.
w).

3.2 UNPACKING AND CONNECTING THE AIR

r
14 100057-00 F, ©JOLIFE AB 2007
On arrival, one of the LUCAS team members should immediately start to unpack and connect LUCAS Chest Compression System.

3.2.1 UNPACKING LUCAS

1. Place the bag on the floor/ground with the straps downwards and the top opposite you (Figure
2. Take hold of the straps on the side of the cover and open up the cover so that the entire bag unfolds (Figure
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
r).
r).

3.2.2 CONNECTING THE AIR

WARNING – INTERRUPTING AIR SUPPLY TO
LIFE SUPPORT DEVICES
If other life support equipment is disconnected from
the air supply, serious injury or death may result.
Never remove attachments to air outlets without
first checking that this will not put a patient at risk.
WARNING – CRUSHING HAZARD
The suction cup is drawn into the bellows with high
force when the air hose is connected.
Keep all body parts away from the suction cup, the
bellows and the height adjustment handles when
connecting the air hose.
1. Confirm that the
Adjust position.
2. If not already connected, attach the air hose to the connector.
3. Attach the connector to a wall outlet or to a portable air cylinder.
4. If using a pressure regulator, open the air valve.
ON/OFF knob is in the
100057-00 F, ©JOLIFE AB 2007 15 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
t
u
y

3.3 ASSEMBLY

WARNING – LETHAL PINCH HAZARD
If the air is connected during the assembly
procedure, and the position, the compressions may seriously injure the patient or create conditions leading to death of the
Always follow the described sequence of air
connection and assembly.
1. Take the back plate out of the bag and approach the patient (Figure
2. Instruct those carrying out CPR to interrupt chest compression.
3. Work in a pair, one person on each side of the patient.
4. Take hold of the patient’s arms. One of the LUCAS team should support the patient’s head.
5. Lift up the patient’s upper body and lay the back plate below the armpits. Ensure that the patient’s arms are outside the back plate (Figure
6. Continue manual chest compressions.
7. Take the upper part of LUCAS out of the bag. Lift it out by holding on to the handles of the support legs. Ensure that the support legs have reached their outer position (Figure
8. Pull up once on the release rings to check that the claw locks are open.
9. Interrupt manual chest compression.
10. Place the upper part of LUCAS over the patient’s chest so that the claw locks of the support legs will engage with the back plate (Figure
11. Start by pressing on the support leg nearest to you and then the one on the other side so the support legs lock against the back plate.
12. Check by pulling upwards, that both support legs have locked against the back plate.
y).
i).
ON/OFF-knob is in the Active
patient.
t).
u).
i
16 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
o

3.4 ADJUSTMENT

3.4.1 INTRODUCTION

To achieve effective compressions, it is very important that the suction cup is positioned correctly on the patient (Figure
The lower edge of the suction cup should be positioned immediately above the end of the sternum. The suction cup should be centred over the sternum (Figure
a).

3.4.2 ADJUSTMENT PROCEDURE

o).
a
s
d
>0 mm
WARNING – MINIMUM PATIENT SIZE
If the suction cup cannot be completely compressed and lowered enough to make the pressure pad touch
the patient's chest, the patient is too small for
LUCAS to be used.
Continue manual compressions.
To adjust the position of LUCAS:
1. Set the
2. Lower the suction cup with the height­adjustment handles until the pressure pad inside the suction cup touches the patient’s chest without compressing the chest (Figure
If there is a distance between the pressure pad inside the suction cup and the sternum, LUCAS cannot be used on the patient(Figure
Note: By simultaneously holding up the outer rim of the suction cup you will more easily see or feel when the pressure pad touches the chest.
3. If the suction cup is not correctly positioned in relation to the patient, adjust the position of LUCAS by pulling on the support legs (Figure determines whether the position is correct.
ON/OFF knob to Adjust .
s).
d).
f). The person assembling the device
f
100057-00 F, ©JOLIFE AB 2007 17 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
g
ACTIVE
LOCK

3.5 OPERATING LUCAS CHEST COMPRESSION SYSTEM

3.5.1 STARTING COMPRESSIONS

When the position of the suction cup is correctly adjusted according to Section 3.4, LUCAS can be switched on:
1. Turn the
will now give chest compressions.
WARNING – PATIENT INJURY OR DEATH
Do not leave the patient or device unattended while
2. Check that the device is working as it should
regarding frequency and compression.
3. When you want to stop chest compressions,
turn the without changing the respective positions of the patient and device.

3.5.2 APPLYING LUCAS STABILIZATION STRAP

When LUCAS is applied to the patient and performing compressions under the supervision of one person of the LUCAS team, attach the Stabilization Strap to LUCAS (Figure detailed instructions, please refer to Appendix B.
ON/OFF knob to Active . LUCAS
LUCAS is active.
ON/OFF knob to the Lock position,
g). For
18 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

3.6 DEFIBRILLATION

WARNING – INCORRECT ECG ANALYSIS
Compressions interfere with ECG analysis.
Switch off LUCAS before ECG analysis. Make the
interruption as short as possible
WARNING – INADEQUATE COMPRESSION
Patient’s movements may change adjustment of
Verify position of suction cup after defibrillation,
re-adjust if necessary.
suction cup.
h
j
Note: Self-adhesive electrodes should be used during defibrillation as these make it easier to work with LUCAS. Position the defibrillator electrodes and wires so they are not under the suction cup; if electrodes are already on the patient, you may need to apply new electrodes.
Defibrillation can be performed when LUCAS is applied to the patient (Figure
1. Apply electrodes and perform the defibrillation according to manufacturer’s instruction for the defibrillator.
2. After defibrillation or analysis of the outcome verify the position of the suction cup and re­adjust if necessary.
h).

3.7 TRA NS PORTING THE PATIENT

3.7.1 SECURING THE PATIE NTS ARMS

The patient’s arms can be fixed to the device during transport, using the straps on the support legs (Figure patient, especially if not in a hospital environment.
j). This makes it easier to move the
WARNING – BLOCKED INTRAVENOUS
CATHETER
Do not tighten the patient strap/straps if drugs are
administered to the patient via an intravenous
catheter.
Caution – Do not lift by the straps
The straps are only intended for securing the
patients arms to LUCAS.
Apply the straps tightly enough to secure the arms, but do not overtighten as this may reduce blood circulation to the patient’s hands.
100057-00 F, ©JOLIFE AB 2007 19 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
k
l
LOCK

3.7.2 LIFTING THE PATIEN T

At least three people should lift the patient and LUCAS – one should support the patient’s head and there should be one person on each side.
Plan carefully before lifting the patient. Consider which equipment needs to be moved and where to place the stretcher.
Those at the side should lift with one hand beneath the claw locks of the back plate. The other hand should be used to lift the patient’s belt, trouser or under the thigh.
Use appropriate lifting technique to avoid injuries when lifting the patient (Figure
k).
WARNING – CRUSHING HAZARD
Do not insert your fingers in the claw locks.
To lift a patient:
1. Secure the patient’s arms with the straps on the support legs.
2. Turn the compressions while lifting the patient, for example when: – lifting to a stretcher – moving from one stretcher to another
3. When the patient has been placed on the stretcher (or other surface), verify correct position of the suction cup on the patient.
4. Turn the
Once the patient is placed on the stretcher, LUCAS may be in active state during all horizontal lifts and movements.
ON/OFF knob to Lock , to pause
ON/OFF knob to Active .

3.7.3 MOV ING HORIZONTALLY

When the situation absolutely demands that the patient be moved while undergoing chest compression, the patient’s chest should be horizontal throughout the entire movement (Figure
l).
20 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

3.7.4 MOVING PATIE NT ON TILTED STRETCHER

In situations where conditions do not allow the chest to remain horizontal:
1(
LOCK
1. Turn the compressions (Figure
2. As soon as the patient is in horizontal position, check the position of the suction cup and adjust if necessary (see Section 3.4).
3. Turn the
ON/OFF knob to Lock to pause
1().
ON/OFF knob to Active .

3.7.5 AMBULANCE TRANSPORT

Make sure that the patient is fixed when using LUCAS Chest Compression System during ambulance transport. The ambulance cot should have four-point restraints for the patient’s upper body in addition to the two lower-body restraints.

3.8 CHANGING AIR SOURCES

WARNING – INTERRUPTED COMPRESSIONS
There may be excessive changeover time if pressure
regulators have to be mounted.
Always use reserve cylinders with pre-fitted
pressure regulators.
To change air source:
1. Set the
2. Close the valve on the used air cylinder.
3. Disconnect the air hose from the air source.
4. Open the valve on the new air source.
5. Attach the air hose to the new air source.
6. Adjust the height of the suction cup on the patient’s chest, as described previously on page 17.
7. Turn the
8. Turn the continue compressions.
ON/OFF knob to Adjust .
ON/OFF knob to Lock . ON/OFF knob to Active to
100057-00 F, ©JOLIFE AB 2007 21 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

3.9 REMOVING LUCAS FROM THE PATIENT

To remove LUCAS Chest Compression System from the patient after treatment:
1. Switch off the device by turning the knob to
2. Raise the suction cup and place the height adjustment handle in its uppermost position.
3. If available, close the air valve, and disconnect the air hose from the air source.
4. To detach LUCAS Stabilization Strap, disconnect the support cushion straps from the device straps.
5. Disconnect the upper part of LUCAS by placing one finger in each release ring of the handles on the support legs and pulling these up so the upper part releases from the back plate. Lay the upper part carefully on the floor.
6. If the patient’s condition allows it, remove the back plate.
Before packing LUCAS or LUCAS Stabilization Strap into the carrying bag, clean and, if required, disinfect the product(s). See instructions in Section 4.
Adjust .
ON/OFF

4 Care after use

This section describes the routines that should be performed by the user after each application of LUCAS Chest Compression System, before packing it into the carrying bag. In addition, LUCAS should be serviced once a year by service personnel authorised by J

4.1 CLEANING ROUTINES

Caution – Do not immerse in water or
The compression mechanism may be damaged by
liquid entering the hood. Do not immerse the device
in water or disinfectant. Do not splash liquid over the

4.1.1 NORMAL CLEANING PROCEDURE

1. Attach the upper part to the back plate.
2. If a LUCAS Stabilization Strap has been used, clean it according to its Instructions for Use.
3. Clean all outer surfaces of LUCAS with a soft cloth, wetted in warm water containing a mild cleaning agent.
4. If necessary, remove the patient straps and clean separately (see Section 4.1.2).
5. Remove the upper part. Clean the claw locks on the support legs and the back plate.
6. Wipe all surfaces with a clean moist cloth.
OLIFE AB.
disinfectant
hood.

4.1.2 REMOVING THE PATIEN T STRAPS

To remove the patient straps from the support legs:
Open the velcro lock and pull out the strap from
the retaining rings on the support leg.
To re-attach the patient straps:
1. Thread the patient strap through the rings on the support leg.
2. Fold the patient strap so that the symbol is visible as shown in the illustration in Section 1.4.
3. Press the strap parts firmly together.
22 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

4.1.3 DISINFECTION

4.3 STORAGE

WARNING – HAZARDS DURING DISINFECTION
Always wear appropriate protective clothing during
disinfection of LUCAS. Follow the handling instructions from the manufacturer of the
disinfectant.
The normal cleaning procedure is sufficient after “normal” soiling. If LUCAS is blood-stained or if an infectious patient has been treated, the device should also be disinfected using 45% isopropyl alcohol with added detergent.

4.1.4 DISINFECTION PROCEDURE

After cleaning LUCAS, disinfect it using the following procedure:
1. Wipe all outer surfaces of the device using a cloth wetted with the disinfectant solution.
2. Allow LUCAS to dry before packing into the bag.

4.2 ROUTINE CHECKS

Store LUCAS in the carrying bag, in a location where it is easily accessible to all those who know how to use the device (see packing instructions in Section 5). The device should not be exposed to dirt or moisture during storage.

4.4 SERVICE

LUCAS should be serviced once a year to check that it is functioning properly. Use the original package when sending LUCAS. For this purpose, store the package and padding that came with LUCAS when you first received it.
WARNING – DO NOT OPEN
Do not, under any circumstances, open the casing of
LUCAS, change or modify any external or internal
parts of LUCAS.
Unless otherwise stated, all service and repairs must
only be done by service personnel authorised by
J
OLIFE AB.
After each use and once a week the following checks should be performed. Sign each item on the sheet attached in Appendix A:
1. Check that the device is clean.
2. Check that the air hose does not have any cracks and fits tightly to LUCAS.
3. Check that the connector(s) is/are present and undamaged.
4. Check that the
Adjust position.
5. Pull up the release rings so that the claw locks are open for assembly. Attach the upper part to the back plate.
6. Check that the compression mechanism can be adjusted from the uppermost to the lowermost position without any jarring or sluggishness.
7. Attach the air hose to a air supply and switch the device on for 1-2 minutes. Check that it is operating normally. Calculate the number of compressions per minute. 100 compressions/ min ±10 is acceptable.
ON/OFF knob is in the
If the above conditions are not adhered to, this may
lead to patient or user injury or death.
Contact your distributor or the manufacturer for current information on where to send LUCAS for maintenance (see page 2 for addresses).
100057-00 F, ©JOLIFE AB 2007 23 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
5Packing away the
device
2)
After cleaning and, if necessary, disinfection, pack LUCAS Chest Compression System into the carrying bag:
1. Place the back plate in the larger pocket at the bottom of the bag.
2. Lay the upper part in the bag with the hood towards the open end.
3. Place the support cushion of LUCAS Stabilization Strap in front of the suction cup.
4. Roll up the air hose and place it on top of the device (Figure
5. Close the bag.
2)).
24 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
2!
6 Technical
specification

6.1 OPERATION

Power source Air
2@
2#
Air consumption
Air hose Permanently attached, 10.8 feet (3.3 m)
Approved air sources for use with LUCAS
All pressure regulators supplied by your distributor are approved for use with LUCAS.
13.7 gallons/minute (52 litres/minute)
in length, with unique male connector.
Pressure regulators or wall outlets for air (medical grade/breathing quality) should meet the following specifications:
1. Nominal supply pressure: 46-87 PSI (3.17-6 bar).
2. Minimum required free flow rate in relation to nominal pressure according to Figure shown: Nominal supply pressure 66.7 PSI (4.6 bar).
3. Pressure drop at minimum free flow rate obtained from Figure not exceed the value obtained from Figure
2@. The shaded area in
Figure
2# illustrates an unacceptably
high pressure drop for nominal supply pressure 66.7 PSI (4.6 bar) and above.
The performance of the air supply should be verified according to the procedures in Section 6.1.1 and 6.1.2. If you do not know how to do this, please contact a clinical engineer or your distributor, for help.
2!. Example
2!, must
100057-00 F, ©JOLIFE AB 2007 25 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE
2$

6.1.1 CHECKING A PRESSURE REGU LATOR

If a flow rate performance diagram exists:
1. Check that the nominal output pressure for the pressure regulator is within 46-87 PSI (3.17-6 bar).
2. Find the regulator output pressure ( Figure
2$ and obtain the minimum free flow
rate (
L/min) from the diagram. The example in
Figure
2$ shows a regulator output pressure of
66.7 PSI (4.6 bar).
3. From the flow rate performance diagram for the pressure regulator, obtain the output pressure for the regulator at the minimum free flow rate. Calculate the pressure drop.
4. Check that the calculated pressure drop does not exceed the maximum allowable pressure drop (
ΔP) at the nominal pressure (Bar),
obtained from Figure
If no flow rate performance diagram exists, follow the instructions in Section 6.1.2 to check your pressure regulator.
2%.
Bar) in
2%

6.1.2 CHECKING A WALL OUTLET

1. Measure and verify that the nominal supply pressure is within 46-87 PSI (3.17-6 bar).
2. Find the nominal supply pressure ( Figure
2% and obtain the minimum free flow
rate (
L/min) from the diagram. The example in
Figure
2% shows a nominal supply pressure of
66.7 PSI (4.6 bar).
3. Measure the supply free flow rate and restrict the flow to the minimum required free flow rate obtained from Figure
4. Measure the supply pressure at the minimum required free flow rate. Calculate the pressure drop.
5. Check that the calculated pressure drop does not exceed the maximum allowable pressure drop (
ΔP) at the nominal pressure (Bar),
obtained from Figure 2%.
2%.
Bar) in
26 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

6.2 COMPRESSIONS

Patients suitable for treatment
Depth of compression
Frequency 90-110 comp/min.
Adult patients with a sternum height of
7.5 inches to 11.9 inches (190 to 303 mm) and a maximal width of 17.7 inches (449 mm) who fit into the device.
The device is not suitable for use with children.
Compresses the sternum by 2 inches (5 cm)
with air at +59 °F to +95 °F (+15 ºC to +35ºC)

6.3 ENVIRONMENT

Temperature range – operation
Temperature range – storage
+41 °F to +104 °F (+5 °C to +40 °C)
-22 °F to +140 °F (-30 °C to +60 °C)

6.4 DIMENSIONS AND WEIGHT

Dimensions in carrying bag 25.6 × 13 × 9.8 inches
(H × W × D) (65 × 33 × 25 cm)
We ig h t 13.9 lbs (6.3 kg)
100057-00 F, ©JOLIFE AB 2007 27 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

Appendix A

WEEKLY CHECK AND CHECK AFTER USE, LUCAS

Use this protocol for the weekly check on the device and the check after use. Store the completed checklist. Copying of this sheet is permitted.
Date:
Action Approved
signature
Check that the device is clean.
Check that the air hose is not cracked and fits tightly to LUCAS.
Check that the connector(s) is/are present and undamaged.
Check that the position.
Pull the release rings so that the claw locks are open for assembly. Attach the upper part to the back plate.
Check that the compression mechanism can be adjusted from the uppermost to the lowermost position without any jarring or sluggishness.
Attach the air hose to a air supply. Switch the device on for 1–2 minutes. Check that it is operating normally. Calculate the number of compressions per 60 seconds. 100 compressions/ min ± 10 is acceptable.
ON/OFF knob is in the Adjust
Adjustments carried out
Serious problem
If any error cannot be corrected take the device out of operation and indicate the problem in the “Serious problem” column. Then immediately contact your distributor for advice or to order a new product (www.jolife.com).
When carrying out these checks always use air from a central air supply or separate air cylinder. Do not use up air from a air cylinder intended for use in emergency situations.
28 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE

Appendix B - LUCAS Stabilization Strap

LUCAS Stabilization Strap
Instructions for Use
US version
100057-00 F, ©JOLIFE AB 2007 29 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABIL IZATION STRAP

1 Introduction

1.1 OVERVIEW

LUCAS Stabilization Strap is intended to be used in conjunction with LUCAS Chest Compression System to provide means of stabilizing LUCAS relative to the patient when LUCAS is active.
LUCAS Stabilization Strap will prevent downward movement of LUCAS during operation.
LUCAS Stabilization Strap can be used in most situations where LUCAS Chest Compression System is used.

1.2 CONTRAINDICATIONS

Do NOT use LUCAS Stabilization Strap under circumstances where it is judged that it might obstruct or delay any medical treatment of the patient.

1.5 DESCRIPTION

LUCAS Stabilization Strap consists of a support cushion with two straps (called support cushion straps), and two separate device straps (see Figure 1). The device straps are attached to the support legs of LUCAS Chest Compression System with buckles.
An extra buckle on each of the device straps is used to connect to the support cushion straps. To minimize set-up time, the device straps should always be attached to the support legs of LUCAS Chest Compression System.
q
w
w
r

1.3 ASSOCIATED DOCUMENTATION

Refer to LUCAS Chest Compression System, Instructions for Use for details of LUCAS and instructions for its use.

1.4 LUCAS STABILIZATION STRAP

LUCAS Stabilization Strap is a portable product for stabilizing LUCAS Chest Compression System to the patient during compressions. It can conveniently be stored and transported in the same carrying bag as LUCAS. LUCAS Stabilization Strap is shown in Figure 1.
e
e
r
Figure 1. LUCAS Stabilization Strap.
LUCAS Stabilization Strap consists of the following parts (see Figure 1):
1. Support cushion
2. Support cushion strap
3. Buckle
4. Device strap

1.5.1 DELIVERED ITEMS

The LUCAS Stabilization Strap package contains the following items:
One support cushion.
Two device straps.
Instructions for Use in relevant language
versions.
30 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABI LIZATION STRAP

1.5.2 USING LUCAS STABILIZATION STRAP

When compressions with LUCAS Chest Compression System have been started, the support cushion is placed under the patient’s neck, and the support cushion straps are fastened to the device straps. See instructions in Section 4.

1.6 THE LUCAS TEAM

Two persons are required when attaching LUCAS Stabilization Strap, and at the same time monitoring the operation of LUCAS.
The Instructions for Use will refer to “The LUCAS team”, consisting of two people who have the required skills stated on page 2 of the Instructions for Use.
Figure 2. LUCAS Stabilization Strap attached to LUCAS Chest Compression System.

1.5.3 STORAGE

LUCAS Stabilization Strap should be stored in the LUCAS carrying bag. See packing instructions in Section 6.

1.5.4 SYMBOLS ON THE DEVICE

Symbol Meaning
Caution – see instructions for use. All users must read
the entire Instructions for Use before using LUCAS Stabilization Strap.
Year of manufacture.
Figure 3. The LUCAS team.
100057-00 F, ©JOLIFE AB 2007 31 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABIL IZATION STRAP
2Preparing LUCAS
3 Warnings and Stabilization Strap for use
Before the first use of LUCAS Stabilization Strap, attach both device straps to the support legs, using the following procedure:
1. Fold the device strap around the support leg.
2. Fasten the buckle on the inside of the support
leg, and check that it is fixed.
3. Make sure that you can move the extra buckle
on the device strap from side to side on the outside of the support leg.
q-w e
precautions
Important
This section contains safety information. All users
must read this section before using LUCAS
Stabilization Strap and observe the safety
information at all times during use.
Section 3.1 summarizes warnings and precautions during assembly and use of LUCAS Stabilization Strap.
Additional hazards are described in relevant sections of the Instructions for Use. All users must observe the safety information in
Cautions during use of LUCAS Stabilization Strap.

3.1 USING LUCAS STABILIZATION STRAP

WA R N IN G – IMPROPER USE
Improper use of LUCAS Stabilization Strap can cause
serious injury to the patient and ineffective chest
compressions.
Wa rn ing s and
r
Figure 4. Attaching the device straps (1-3), LUCAS Chest Compression System with both device straps attached (4).
J
OLIFE AB strongly recommends that LUCAS
Stabilization Strap only be used by personnel who
have undertaken training in use and handling of
LUCAS Chest Compression System.
Other personnel may assist those with the
necessary training in using LUCAS.
WARNING – PATIENT INJURY OR DEATH
The use of LUCAS Stabilization Strap does not mean
that you can pay less attention to the patient.
Do not leave the patient or LUCAS Chest
Compression System unattended while LUCAS is
active.
Particularly, make sure that LUCAS does not move on
the patient. If the position of LUCAS is changed,
correct the position according to LUCAS Chest
Compression System, Instructions for Use.
WARNING – IMPAIRED VENTILATION OF
PATIENT
Do not use LUCAS Stabilization Strap if this impairs
the ventilation of the patient.
32 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABI LIZATION STRAP
4Using LUCAS
Stabilization Strap

4.1 INTRODUCTION

This section gives instructions for attaching LUCAS Stabilization Strap to LUCAS Chest Compression System during treatment of a patient with LUCAS.
Please refer to LUCAS Chest Compression System, Instructions for Use, for details on the use of LUCAS.
Always work in a pair, one person on each side of the patient.
Figure 5. The LUCAS team, working on each side of the patient.
4.2 ATTACHING LUCAS STABILIZATION
TRAP
S
WARNING – DO NOT DELAY LIFE SUPPORT
ACTIONS
Before unpacking and attaching LUCAS Stabilization Strap, make sure that all life-supporting actions have
been taken. Particularly, LUCAS Chest Compression
System must be applied to the patient and
performing compressions under the supervision of
one person of the LUCAS team.
If the application of LUCAS Stabilization Strap
threatens to delay the treatment of the patient, it
should be applied later or not at all.
Begin by taking out LUCAS Stabilization Strap from the LUCAS carrying bag. Follow the instructions below to attach the Stabilization Strap to LUCAS (see Figure 6).
100057-00 F, ©JOLIFE AB 2007 33 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABIL IZATION STRAP
1. Lift the patient’s head carefully and place the
support cushion under the patient’s neck, as close to the patient’s shoulders as possible.
2. Move the buckles on the device straps towards
the patient’s head.
3. Connect the buckles on the support cushion
straps to the device straps. Make sure that the straps are not twisted.
q
w-e
r-t y
WARNING – LIFTING THE PATIENT
LUCAS Stabilization Strap must never replace a
person supporting the patient’s head. Always lift a
patient and LUCAS according to LUCAS Chest
Compression System, Instructions for Use.
WAR NI NG – TRIPPING ON STRAPS
Mind your step to avoid tripping on the support
cushion straps, or unintentionally tightening the
straps, when lifting the patient.
4.2.1 IF THE POSITION OF THE SUCTION CUP IS
NOT CORRECT
Refer to LUCAS Chest Compression System, Instructions for Use, for details on adjusting the position of LUCAS relative to the patient.
1. Turn the position.
2. Release the support cushion straps from LUCAS.
3. Adjust to correct position of the suction cup on the sternum.
4. Start compressions.
5. Extend the support cushion straps by pulling the buckles to the end positions of the straps.
6. Re-attach LUCAS Stabilization Strap, see Section 4.2.
ON/OFF knob to the (Adjust)
Figure 6. Attaching LUCAS Stabilization Strap to LUCAS.
WARNING – IMPAIRED VENTILATION OF
PATIENT
Never tighten the support cushion straps so much
that this impairs the ventilation of the patient.
4. Tighten the support cushion straps firmly, while holding the support legs of LUCAS. Make
sure that the position of LUCAS External Cardiac Compressor and the position of the suction cup on the patient, do not change.
5. Verify that the support cushion straps are properly fastened to the support legs of LUCAS.
6. Visually check the position of the suction cup: The lower edge of the suction cup should be positioned immediately above the end of the sternum. The suction cup should be centered over the sternum. Refer to Section 4.2.1 if the suction cup needs to be adjusted
q-w
e-r
Figure 7. Adjusting the position of the suction cup.
34 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABI LIZATION STRAP

5 Care after use

5.1.3 DISINFECTION PROCEDURE

After cleaning LUCAS Stabilization Strap, disinfect it using the following procedure:
This section describes the routines that should be performed by the user after each use of LUCAS Stabilization Strap, before packing it into the carrying bag.

5.1 CLEANING ROUTINES

5.1.1 NORMAL CLEANING PROCEDURE

1. Clean all surfaces of LUCAS Stabilization Strap using a soft cloth, wetted in warm water containing a mild cleaning agent.
2. Wipe all surfaces with a clean moist cloth.
3. If necessary, remove the device straps from LUCAS and clean separately as described in step 1. After cleaning, re-attach the device straps to the support legs (see Section 2).

5.1.2 DISINFECTION

WARNING – HAZARDS DURING
DISINFECTION
Always wear appropriate protective clothing during
disinfection of LUCAS Stabilization Strap. Follow the
handling instructions from the manufacturer of the
disinfectant.
1. Wipe all outer surfaces of the device using a cloth wetted with the disinfectant solution.
2. Allow LUCAS Stabilization Strap to dry before packing into the bag.

5.2 ROUTINE CHECKS

After each use and once a week, the following checks should be performed. Sign each item on the sheet attached in Appendix C:
1. Check that LUCAS Stabilization Strap is clean.
2. Verify that the device straps are fitted correctly as shown in Figure 4.
3. Make sure that the support cushion straps are extended, to speed up the next application of LUCAS Stabilization Strap.

5.3 STORAGE

Always store LUCAS Stabilization Strap together with LUCAS in the carrying bag (See packing instructions in Section 6). LUCAS Stabilization Strap should not be exposed to dirt or moisture during storage.
The normal cleaning procedure is sufficient after “normal” soiling. If LUCAS Stabilization Strap is blood-stained or if an infectious patient has been treated, the device should also be disinfected using 45% isopropyl alcohol with added detergent.
100057-00 F, ©JOLIFE AB 2007 35 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABIL IZATION STRAP
6 Packing away the
7 Technical
Stabilization Strap
After cleaning and, if necessary, disinfection, pack LUCAS Stabilization Strap together with LUCAS in the bag:
1. Pack LUCAS according to instructions in LUCAS Chest Compression System, Instructions for Use.
2. To speed up the next application of LUCAS Stabilization Strap, extend the support cushion straps by pulling the buckles to the end positions of the straps.
3. Lift the air hose and place LUCAS Stabilization Strap in front of the suction cup. Put the rolled­up air hose back on top of the Stabilization Strap.
4. Close the bag.
specification

7.1 PATIENTS SUITABLE FOR TREATMENT

Patients suitable for treatment

7.2 ENVIRONMENT

Temperature range – use
Temperature range – storage
Refer to technical specifications for LUCAS Chest Compression System on page 27
41 °F to +122 °F (+5 °C to +50 °C)
-22 °F to +140 °F (-30 °C to +60 °C)
Figure 8. LUCAS Stabilization Strap packed with LUCAS.
36 100057-00 F, ©JOLIFE AB 2007
LUCAS
TM
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABI LIZATION STRAP

Appendix C

WEEKLY CHECK AND CHECK AFTER USE, LUCAS STABILIZATION STRAP

Use this protocol for the weekly check on LUCAS Stabilization Strap and the check after use. Store the completed checklist. Copying of this sheet is permitted.
Date:
Action Approved, signature Adjustments carried
out
Check that all parts of the Stabilization Strap are clean.
Verify that the device straps are fitted correctly as shown in Figure 2.
Make sure that the support cushion straps are extended.
If any error cannot be corrected, take LUCAS Stabilization Strap out of operation and indicate the problem in the “Serious problem” column. Then immediately contact your distributor for advice or to order a new product (www.jolife.com).
Serious problem
100057-00 F, ©JOLIFE AB 2007 37 LUCASTM CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABIL IZATION STRAP
38 100057-00 F, ©JOLIFE AB 2007
TM
LUCAS
CHEST COMPRESSION SYSTEM, INSTRUCTIONS FOR USE—APPENDIX B, LUCAS STABI LIZATION STRAP
Loading...