LIFEPAK, LIFEPAK CR, LIFEPAK EXPRESS, and QUIK-COMBO are registered trademarks of Physio-Control, Inc. ADAPTIV, CODE-STAT,
CHARGE-PAK, QUIK-PAK, and Shock Advisory System are trademarks of Physio-Control, Inc. IrDA is a registered trademark of Infrared Data
Association. Specifications are subject to change without notice.
It is the responsibility of our customers to ensure that the appropriate person(s) who may use this defibrillator have
access to the information in this manual, including general safety information provided in Section 1.
Rx Only
Device Tracking
The U.S. Food and Drug Administration requires defibrillator manufacturers and distributors to track the
location of their defibrillators. If the device is located somewhere other than the shipping address or the device has
been sold, donated, lost, stolen, exported, destroyed, permanently retired from use, or if the device was not obtained
directly from Physio-Control, please do one of the following: register the device at http://www.physio-control.com, call
the device tracking coordinator at 1.800.426.4448, or use one of the postage-paid address change cards located in
the back of this manual to update this vital tracking information.
Revision History
These operating instructions describe the LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators with software
Version 3.0 or later.
TABLE OF CONTENTS1
1 Introduction
About Automatic External Defibrillators ............................................................................................... 1-2
Indications for Use........................................................................................................................ 1-2
Text Conventions.......................................................................................................................... 1-4
Safety Information ................................................................................................................................ 1-4
Supplies, Accessories, and Training Tools ........................................................................................... 5-8
Warranty Information ........................................................................................................................... 5-8
6 Defibrillator Operating Settings
Operating Settings and Setup Configuration ....................................................................................... 6-2
A Specifications
B Shock Advisory System
C User’s Checklist
D Electromagnetic Compatibility Guidance
Index
ivLIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
INTRODUCTION1
1 Introduction
This section provides background information about defibrillation and includes an overview of
LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillator features.
About Automatic External Defibrillatorspage 1-2
Safety Information1-4
Symbols1-6
About the LIFEPAK CR Plus and
1-10
LIFEPAK EXPRESS Defibrillators
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions1-1
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are automated external defibrillators
(AEDs). For many years, defibrillators have been used only by medical professionals to treat victims in
sudden cardiac arrest (SCA). Today, the ability of defibrillators to save lives is so widely recognized that
people once trained to do only cardiopulmonary resuscitation (CPR) can now use defibrillators.
After electrode pads are applied to the victim’s chest, the defibrillator analyzes the victim’s heart rhythm.
If a shockable rhythm is detected, the defibrillator will either deliver an intense pulse of electricity
(shock) to the heart muscle (fully automatic model) or direct the responder to deliver the shock
(semiautomatic model). The defibrillator delivers shocks through the electrode pads on the victim’s
chest.
When this pulse of electricity is delivered, it is called defibrillation. Defibrillation is recognized for treating
life-threatening heart beat irregularities, such as ventricular fibrillation, that cause SCA.
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are designed specifically for infrequent
use and for use by people whose only training is in CPR and in using AEDs.
Indications for Use
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are indicated for use on patients in cardiac
arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs
of circulation (for example, no pulse, no coughing, or no movement). With Infant/Child Reduced Energy
Defibrillation Electrodes, the LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators may be used on
children who are up to 8 years old or who weigh up to 25 kg (55 lb).
Contraindications
None known.
Why the Need for Defibrillators
The American Heart Association estimates that, in the USA alone, at least 250,000 people die each
year of cardiac arrest. Of these, about 10,000 people might have been saved had they received
immediate treatment from a defibrillator.
Sudden cardiac arrest is usually caused by a malfunction in the heart’s electrical system. Called
ventricular fibrillation, this critical condition prevents the heart from pumping blood throughout the body.
Ventricular fibrillation can cause death within seconds.
Defibrillation is a relatively simple procedure that involves placing electrode pads on a victim’s exposed
chest and delivering an electrical shock to the heart. The externally-delivered shock often restores the
heart’s electrical system to normal rhythm. Combined with CPR, defibrillation provides the most
effective care for victims in cardiac arrest.
1-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Introduction
Terminology
The following terms appear in this manual.
AEDAutomated External Defibrillator. A device that evaluates the victim’s heart
rhythm and delivers an electrical shock to the heart if a shockable rhythm is
detected.
Cardiac arrestThe termination of the heart’s pumping action resulting in the lack of a
heartbeat or pulse and breathing.
CPRCardiopulmonary resuscitation. This involves delivering rescue breathing and
chest compressions to a victim in cardiac arrest.
DefibrillationDelivery of an electrical shock to the heart for the purpose of reversing
ventricular fibrillation.
ECGElectrocardiogram. A composite picture of what is occurring electrically in the
heart.
FibrillationChaotic activity of the heart’s electrical system. This condition can occur in
the atria or the ventricles. When it occurs in the ventricles, they quiver in a
rapid, chaotic manner, preventing them from pumping blood to the body.
1 Introduction
Heart attackA nonspecific term referring to the death of heart muscle resulting from
interruption of blood supply, often confused with cardiac arrest.
ImpedanceResistance to the flow of electrical current through the body.
JouleThe basic unit of energy delivered by a defibrillator.
LEDLight emitting diodes.
Myocardial infarctionThe specific term for what is usually meant by heart attack; death of heart
muscle resulting from an interruption of the blood supply to that area of
myocardium.
Nonshockable rhythmA heart rhythm that is detected by the defibrillator that does not need a
shock, but may need CPR.
VictimIn this manual, the person suffering from cardiac arrest.
ResponderIn this manual, the person giving aid to a victim in cardiac arrest. Used
interchangeably with user.
SAS
Physio-Control patented Shock Advisory System
™
.
Shockable rhythmA heart rhythm that is detected by the defibrillator as requiring a shock, for
example, ventricular fibrillation.
UserIn this manual, the person giving aid to a victim in cardiac arrest. Used
Throughout this manual, special text characters are used to indicate labels and voice instructions:
Operating control labels:CAPITAL LETTERS such as ON/OFF and SHOCK.
instructions: ITALICIZED CAPITAL LETTERSsuch as STAND CLEAR.
Voice
SAFETY INFORMATION
This section provides important information to help you safely operate your defibrillator. Familiarize
yourself with all of the terms, warnings, and symbols presented in this section.
Safety Terms
You may encounter the following terms in this manual and while using your defibrillator:
DangerImmediate hazards that will result in serious personal injury or death to the user and/or
the victim.
WarningHazards or unsafe practices that could result in serious personal injury or death to the
user and/or the victim.
CautionHazards or unsafe practices that could result in minor personal injury to the user and/or
the victim, product damage, or property damage.
1-4LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Introduction
General Warnings and Cautions
WAR NIN GS!
Shock hazard.
The defibrillator delivers up to 360 joules of electrical energy. Unless properly used as described in this
manual, this electrical energy may cause serious injury or death. Do not attempt to operate this
defibrillator unless thoroughly familiar with this manual and the function of all controls, indicators,
connectors, and accessories.
Do not insert a finger or any object other than the CHARGE-PAK battery charger into the well of the
defibrillator.
Shock or fire hazard.
Do not immerse any portion of this defibrillator in water or other fluids. Avoid spilling any fluids on the
defibrillator or its accessories. Do not clean with ketones or other flammable agents. Do not autoclave
or sterilize this defibrillator or the accessories unless otherwise specified.
Possible fire or explosion.
Do not store this defibrillator in the presence of flammable gases or in direct contact with flammable
material.
Do not use this defibrillator in the presence of flammable gases or anesthetics. Use care when
operating this defibrillator close to oxygen sources (such as bag-valve-mask device or ventilator tubing).
Turn off gas source or move source away from patient during defibrillation.
Possible device shutdown.
When the attention symbol first appears on the readiness display, only a small number of shocks
and monitoring time are available. Always keep a CHARGE-PAK™ battery charger in the defibrillator.
Routinely check that the defibrillator is ready for use. Replace the CHARGE-PAK battery charger after
each use of the defibrillator.
Possible electrical interference with device performance.
Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference
(RFI) which could affect the performance of this device. RFI may result in improper device operation,
distorted ECG, failure to detect a shockable rhythm, or cessation of pacing. Avoid operating the device
near cauterizers, diathermy equipment, cellular phones, or other portable and mobile RF
communications equipment. Maintain equipment separation of at least 1.2 m (4 ft) and do not rapidly
key EMS radios on and off. Contact a technical support representative if assistance is required.
Possible electrical interference.
Using cables, electrodes, or accessories not specified for use with this device may result in increased
emissions or decreased resistance to electromagnetic interference which could affect the performance
of this defibrillator or of equipment in close proximity. Use only parts and accessories specified in this
manual.
Possible electrical interference.
This device may cause electromagnetic interference (EMI) especially during charge and energy
transfers. EMI may affect the performance of equipment operating in close proximity. If possible, verify
the effects of defibrillator discharge on other equipment prior to using the defibrillator in an emergency
situation.
1 Introduction
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions1-5
Using other manufacturers’ cables or electrodes may cause the defibrillator to perform improperly and
invalidates the safety agency certification. Use only the parts and accessories specified in this manual.
Using damaged or expired equipment or accessories may cause the defibrillator to perform improperly
and may injure the victim or the user.
Safety risk and possible equipment damage.
Monitors, defibrillators, and their accessories (including electrodes and cables) contain ferromagnetic
materials. As with all ferromagnetic equipment, these products must not be used in the presence of the
high magnetic field created by a Magnetic Resonance Imaging (MRI) device. The high magnetic field
created by an MRI device will attract the equipment with a force sufficient to cause death or serious
personal injury to persons between the equipment and the MRI device. This magnetic attraction may
also damage the equipment. Skin burns will also occur due to heating of electrically conductive
materials, such as patient leads and pulse oximeter sensors. Consult the MRI manufacturer for more
information.
CAUTION!
Possible equipment damage.
This defibrillator may be damaged by mechanical or physical abuse, such as immersion in water or
dropping the defibrillator. If the defibrillator has been abused, remove it from use and contact a qualified
technician.
1-6LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Introduction
OK
SYMBOLS
The following symbols may appear in this manual and on the LIFEPAK CR Plus and
LIFEPAK EXPRESS defibrillators. For more information about the readiness display symbols, refer to
Section 2, "Getting Started".
SymbolExplanation
OK Indicator. The defibrillator is ready for use.
Attention. On the readiness display—the internal battery is low. Refer to page 2-4 for
more information about the readiness display.
Attention. On the CHARGE-PAK battery charger—consult the operating instructions.
Refer to page 5-4 for more information about the CHARGE-PAK battery charger.
Attention. On the safety warning—consult the operating instructions. Refer to
page 1-5 for more information about the warnings and cautions.
1 Introduction
Attention. On the electrode pads—consult the operating instructions. Refer to
page 2-6 for more information about electrode pads.
Warning, high voltage.
CHARGE-PAK battery charger indicator. On the readiness display—CHARGE-PAK
battery charger needs to be replaced.
Wrench indicator. There is a condition that prevents or could prevent normal
defibrillator operation. Refer to page 5-7 for more information.
CHARGE-PAK battery charger.
This end up.
Fragile/breakable.
Handle with care.
Protect from water.
Power On/Off button.
Type BF patient connection.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions1-7
Not intended for use on children who are less than eight years of age or who weigh
less than 25 kg (55 lb).
Infant/Child Reduced Energy Electrodes are not compatible with QUIK-COMBO
defibrillation and therapy cables. To use Infant/Child Electrodes, connect Infant/Child
Electrodes directly to the AED.
Not intended for use on adults.
Physio-Control electrodes are latex-free.
Arrow indicates ON/OFF button location.
Symbol denoting a defibrillator and identifies the shock button.
Lot code.
Do not reuse—single use only.
Use By date shown: yyyy-mm-dd.
Refer to instructions for recycling procedure, page 5-8.
Refer to instructions for disposal procedure, page 5-2.
Do not dispose of this product in the unsorted municipal waste stream. Dispose of this
product according to local regulations. See http://recycling.medtronic.com for
instructions on the proper disposal of this product.
Mark of conformity according to the European Medical Device Directive 93/42/EEC.
Canadian Standards Association certification for Canada and the United States.
1-8LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
SymbolExplanation
!USA
35°C
15°C
95°F
59°F
50°C
0°C
122°F
32°F
YYYY
For USA audiences only.
Atmospheric pressure limitation.
Relative humidity range 5% to 95%.
Store in a cool, dry location (0° to 50°C or 32° to 122°F).
Recommended storage temperature: 15° to 35°C (59° to 95°F). Storage at extreme
temperatures of -40° or 70°C (-40° or 158°F) is limited to seven days. If storage at
these temperatures exceeds one week, the electrode shelf-life will be reduced.
Introduction
1 Introduction
Rx Only or
Rx Only
MIN
CAT or CAT
REF or REF
Date of manufacture.
CAUTION - FIRE HAZARD
Do not disassemble, heat above 100°C (212°F), or incinerate battery
CAUTION - FIRE HAZARD
Do not crush, puncture, or disassemble battery
Federal (U.S.) law restricts this device to sale by, or on the order of, a physician.
Manufacturer’s Item Number.
Catalog number used for placing orders.
Reorder number.
SN or SN
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions1-9
ABOUT THE LIFEPAK CR PLUS AND LIFEPAK EXPRESS DEFIBRILLATORS
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are designed for indoor and outdoor use.
Each has two models available—fully automatic and semiautomatic. After the electrode pads are
applied, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected,
delivers a shock without any responder assistance. The semiautomatic model evaluates the heart
rhythm but requires the responder to press the shock button if a shockable rhythm is detected. Both
models have voice instructions that guide the responder through the defibrillation process.
Capabilities and Features
The following paragraphs introduce specific features found in the defibrillators.
Accessories
The defibrillator arrives with an installed CHARGE-PAK battery charger, one preconnected QUIK-PAK™
electrode packet, and the operating instructions. Refer to Section 5 for other accessories.
Automated Operation
Voice instructions guide the responder through the defibrillation process.
The
fully automatic defibrillator requires no operator interaction beyond placing the electrode pads on
the victim. If the defibrillator detects a shockable rhythm, it warns the responder prior to delivering any
shock, and then delivers a shock without operator interaction.
semiautomatic defibrillator has a highly visible shock button that the responder must press when
The
the defibrillator issues a shock voice instruction.
Automatic Self-Tests
The defibrillator tests itself each week and every time you turn it on. In addition, every month, the
defibrillator performs a more extensive self-test. This self-test checks the defibrillator’s circuitry to verify
that it is ready for use.
ClearVoice™ Technology
ClearVoice technology was created specifically for portable medical devices. This technology
incorporates how the human ear interprets audio prompts and instructions within real world cardiac
arrest response scenes such as shopping malls, on the freeway, or in an emergency room. ClearVoice
technology minimizes distortion and enhances speech intelligibility so the user can clearly understand
audio and instructional prompts in a chaotic and stressful environment.
Customized Setup
The defibrillator is shipped ready to use with the preprogrammed ADAPTIV™ biphasic escalating
energy protocol and other operating settings. The operating settings are configured in accordance with
customer order. Refer to Section 6, "Defibrillator Operating Settings".
Data Management
The defibrillator digitally stores data when it is turned on and the electrode pads are successfully
applied to the victim. The stored data includes date and time, ECG data, and the number of shocks. The
defibrillator also stores the results of the automatic self-tests.
®
Stored data can be transferred to a PC by means of a serial infrared link, the IrDA
port. A data transfer
and management program running on the PC transfers event and test data from the defibrillator.
1-10LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Introduction
Defibrillation Electrodes (Pads)
When applied to the victim, Physio-Control QUIK-PAK defibrillation electrodes (pads) work with the
defibrillator to monitor the heart rhythm and identify when a shock should be delivered. If victim care is
transferred to emergency medical personnel, these electrode pads can be disconnected from the
defibrillator and reconnected to other AEDs or defibrillators that are compatible with QUIK-COMBO
®
electrodes.
For infants or children who are less than eight years of age or weigh less than 25 kg (55 lb), use
Infant/Child Reduced Energy Defibrillation Electrodes. These electrodes reduce the energy delivered by
the AED to the victim by approximately 75%. Keep all electrode pads with the AED.
Defibrillation Waveform
The defibrillation shock, using ADAPTIV Biphasic technology, is delivered in the form of a biphasic
truncated exponential (BTE) defibrillation waveform.
Heart Rhythm Analysis
The Physio-Control patented Shock Advisory System™ evaluates the victim’s heart rhythm. Refer to
"Appendix B" for further information.
Motion Detection
1 Introduction
This patented motion system detects victim or operator motion that could affect the heart rhythm
evaluation. Heart rhythm evaluation is interrupted if the defibrillator detects motion.
Readiness Display
This easy-to-read visual display indicates if the defibrillator is ready for use or if it needs attention.
SafeGuard™ Power System
The SafeGuard power system offers a dual layer of security as the CHARGE-PAK battery charger helps
keep the rechargeable internal lithium battery at its optimum level. The internal battery supplies power
to operate the defibrillator. It is important to keep a CHARGE-PAK battery charger in the defibrillator,
even when the defibrillator is stored. Refer to "Replacing the CHARGE-PAK Battery Charger and the
QUIK-PAK Electrode Packet" on page 5-2 for more information.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions1-11
UNPACKING AND INSPECTING YOUR LIFEPAK CR PLUS OR LIFEPAK EXPRESS
DEFIBRILLATOR
To help ensure the integrity of your defibrillator and to verify that it is complete, perform the initial
inspection as follows:
1 Remove your defibrillator and examine the outside for signs of damage that may have occurred
during shipping.
2 Check the remaining contents in the box against the sales order.
3 View the
This indicates your defibrillator is ready for use. If the
Physio-Control representative.
4 Notice the Use By date.
The Use By date is located below the readiness display. This date tells you when the electrode
packet and battery charger must be replaced.
5 Check the defibrillator speaker by performing the following:
Note: This is only a speaker check. Do not respond to the voice instructions.
OK symbol in the readiness display.
OK symbol is not visible, contact your local
•Press the
ON-OFF button to open and turn on your defibrillator.
Confirm that the voice instructions sound.
•Press and hold the
ON-OFF button for approximately 2 seconds to turn off your defibrillator.
Three tones will sound.
6 Close and latch the lid. Do not reopen the lid unless necessary.
Doing so will reduce battery power.
If you have any questions about your defibrillator, please call your local Physio-Control representative.
CAUTION!
After completing an initial inspection, do not open the lid unnecessarily. Each time you open the
lid, the defibrillator turns on and internal battery power is reduced. After 30 minutes of cumulative
on time, the CHARGE-PAK indicator appears on the readiness display indicating the
CHARGE-PAK battery charger and the QUIK-PAK electrode packet should be replaced.
Save the shipping container and inserts in case you need to reship the defibrillator in the future.
2-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Getting Started
WHERE TO LOCATE YOUR LIFEPAK CR PLUS OR LIFEPAK EXPRESS
DEFIBRILLATOR
The defibrillator should be easy to reach in a location free of obstacles. This could include a location
near existing emergency equipment, such as fire extinguishers and first-aid kits. When considering
location, avoid areas that expose the defibrillator to moisture, dust, or extreme temperatures.
Recommended storage temperature is 15° to 35°C (59° to 95°F). Storage at higher temperatures will
shorten the life of the battery and electrodes.
WARNING!
Possible fire or explosion.
Do not store this defibrillator in the presence of flammable gases or in direct contact with flammable
material.
Although the defibrillator and electrodes are designed to withstand environmental temperature
fluctuations between -40° to 70°C (-40° to 158°F), storage at extreme temperatures of -40° or 70°C
(-40° or 158°F) is limited to one week. If storage at these temperatures exceeds one week, the
electrode shelf-life will be reduced. Refer to Appendix A, page A-4 environmental specifications
information. You can place your defibrillator on a stable surface, or you can mount it using the wall
mount bracket accessory. Contact your local Physio-Control representative.
2 Getting Started
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions2-3
This section introduces you to the controls, indicators, and labels on the LIFEPAK CR Plus and
LIFEPAK EXPRESS defibrillators.
Outside Controls, Indicators, and Labels
Controls, indicators, and labels on the outside of the defibrillator are identified in Figure 2-1 and
described in Table 2-1.
Figure 2-1 Outside Controls, Indicators, and Labels
Table 2-1 Outside Controls, Indicators, and Labels
FeatureDescription
Readiness DisplayThere are four indicators that can appear when your defibrillator is turned off.
These indicators allow you to determine, just by looking at the defibrillator,
whether it’s ready for use or needs attention. They include the following:
The OK indicator appears when the defibrillator is turned off and
ready for use.
The CHARGE-PAK indicator appears when the CHARGE-PAK battery
charger needs to be replaced or is not installed in the defibrillator. If
needed, the defibrillator can be used in an emergency.
The attention indicator appears when the internal battery is not fully
charged. When this indicator first appears, the internal battery can
power the defibrillator for a minimum of 6 shocks or 42 minutes.
The wrench indicator appears when a condition prevents or could
prevent the defibrillator from operating normally.
LidThe top of the defibrillator.
ON-OFF buttonThe ON-OFF button opens the defibrillator lid and turns the defibrillator on.
Pressing and holding the button for approximately 2 seconds after the lid is
open turns off the defibrillator.
2-4LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Table 2-1 Outside Controls, Indicators, and Labels (Continued)
Quick reference card
Use By date
Electrode packet
Electrode packet
release handle
Electrode packet
anchor pin
FeatureDescription
Getting Started
CHARGE-PAK batter y
charger
The CHARGE-PAK battery charger delivers a trickle charge to the internal
battery. The battery charger can provide a charge for approximately twoyears,
as long as the defibrillator is not used.
IrDA portThe Infrared Data Association defines specifications for infrared wireless
communications. The IrDA port provides wireless communications for
transferring data from your defibrillator to a PC.
Carrying HandleThe carrying handle is used to transport the defibrillator.
Safety WarningsSafety warnings provide important information concerning the defibrillator’s
use and service.
Serial Number LabelThe serial number label includes the defibrillator identification number.
Inside Features
The inside features of the defibrillator are designed to make it easy to use during a cardiac arrest event.
When you press the
packet and its release handle as shown in Figure 2-2. Table 2-2 describes the inside features.
ON-OFF button, the lid opens, the defibrillator turns on, and you see the electrode
2 Getting Started
Figure 2-2 Inside Features
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions2-5
Quick reference cardThis card provides abbreviated graphic directions for using your
defibrillator to treat a victim in cardiac arrest.
Use By dateUse By date shown (yyyy-mm-dd) can be viewed through the
defibrillator lid when it is closed.
Electrode packetThe QUIK-PAK electrode packet is preconnected to the defibrillator.
This packet contains a set of electrode pads.
Electrode packet release handle When you pull this handle, the electrode packet tears open.
Electrode packet anchor pinThis pin securely positions the electrode packet to the defibrillator.
After you pull the electrode packet release handle and tear open the electrode packet, you will see the
features shown in Figure 2-3.
Figure 2-3 Inside Features After Releasing the QUIK-PAK Electrode Packet
2-6LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Getting Started
Table 2-3 Inside Features After Releasing the QUIK-PAK Electrode Packet
FeatureDescription
SpeakerThis projects the voice instructions that lead you through the
defibrillation process.
Electrode indicatorsThe electrode indicators flash red until the pads are applied to the
victim’s exposed chest. When the pads are successfully applied, the
indicators turn solid green and the defibrillator can perform an
analysis.
In addition, electrode indicators briefly flash when the defibrillator
performs an automatic self-test.
Blue plasticThe plastic liner protects the conductive adhesive gel until the
electrode pads are used.
Electrode padsThe electrode pads are applied to the victim’s exposed chest; they
transfer the defibrillation energy (shock) to the victim. The electrode
pads must be removed from the blue plastic before applying them to
the victim.
2 Getting Started
SHOCK buttonThe SHOCK button is only provided on the semiautomatic model.
When pressed, this button delivers a shock to the victim. You cannot
deliver a shock to a victim unless the defibrillator instructs you to do
so.
Electrode connectorThe electrode connector is used to connect the electrode pads to
the defibrillator. To aid in victim transport, the connector can be
unplugged from the defibrillator and plugged into another AED or
defibrillator equipped for QUIK-COMBO electrodes.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions2-7
To help ensure safe use of the defibrillator, completely familiarize yourself with the following warnings
and cautions.
WAR NIN GS!
Incorrect rhythm interpretation.
Performing CPR or otherwise handling or transporting the victim while the defibrillator is evaluating the
heart rhythm can cause an incorrect or delayed diagnosis. Keep the victim as still as possible while the
defibrillator is attached and do not transport the victim.
Shock hazard.
When instructed “Do not touch patient,” “Stand by,” or “Everyone clear,” remain still, do not touch the
defibrillator, patient, electrode pads or any material in contact with the patient. Make sure no one is
touching the patient when the defibrillator shocks the patient.
Shock hazard.
To remove an unwanted charge, disconnect the electrode cable from the defibrillator, wait for the
defibrillator to automatically remove the charge, or turn off the defibrillator.
Possible fire, burns, and ineffective energy delivery.
During defibrillation, material in contact with the electrode pads can cause electrical sparks, skin burns,
and divert important defibrillating energy away from the heart. Place electrode pads so that they adhere
to the skin completely. Do not allow the electrode pads to touch each other, medication patches,
dressings, metal objects, other electrodes, or any other material on the patient’s chest.
During defibrillation, air pockets between the skin and electrode pads can cause skin burns. To help
prevent air pockets, make sure electrode pads completely adhere to the skin. Do not use damaged,
expired, or dried-out electrode pads.
CAUTION!
Possible equipment damage.
Before using the defibrillator, disconnect all equipment from the patient that is not
defibrillator-protected.
RESPONDING TO A SUDDEN CARDIAC ARREST EMERGENCY
If not treated, sudden cardiac arrest (SCA) will cause death. In an SCA situation, it is important to
remember to immediately call for help and activate your emergency response system.
3-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Using the Defibrillator
Basic Steps for Using the LIFEPAK CR Plus or LIFEPAK EXPRESS Defibrillator
Responding to an SCA emergency using the defibrillator involves these basic steps:
Determine if the victim is in SCA. A person in SCA will not
respond when you try to shake him or her.
Check for breathing by listening next to the victim’s mouth and
looking for chest movement.
Use your defibrillator only if the victim is not responding, not
moving, and not breathing normally or not breathing at all. If in
doubt, use your defibrillator.
Place your defibrillator near the victim and on the side next to
you. Press the
ON/OFF button to open the lid and turn on your
defibrillator. Remain calm. Your defibrillator will guide you
through the defibrillation process.
Expose the victim’s chest. If the chest is excessively hairy,
quickly shave the hair in the area where you will place the pads.
If the chest is dirty or wet, wipe the chest clean and dry. If there
are medicine patches on the victim’s chest, remove them.
Hold down the left side of the electrode packet with one hand
and pull the red packet handle down with the other. The
electrode packet tears open.
Tear open the packet completely to remove the pads. A small
piece of the packet will remain attached to your defibrillator.
3 Using the Defibrillator
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions3-3
Separate the electrode pads, one at a time, from the blue
plastic. Use these pads on adults or children 8 years of age or
more, who weigh 25 kg (55 lb) or more. For infants or children
who are less than 8 years of age or who weigh less than 25 kg
(55 lb), special electrodes are needed. Refer to page 5-8 for
more information.
WARNING!
If you cannot determine a child’s age or weight, or if special
infant/child electrodes are not available, proceed with the existing
electrode pads, and continue on to the next step.
Apply the electrode pads to the victim’s bare chest (exactly as
shown in the picture on the pads). Be sure to press firmly so that
the pads completely adhere to the victim’s chest.
Note: Be sure you do not place the electrode pads over an
implanted device such as an implanted pacemaker or ICD.
An indication of an implant is a protrusion in the chest skin
and a scar. If you are in doubt, apply the pads as shown on
the labels.
Listen to voice instructions and do not touch the victim unless
instructed to do so.
If the defibrillator heart rhythm analysis determines that a shock
is needed, the defibrillator will announce
and then instruct you to
PRESS FLASHING BUTTON to administer a
shock (semiautomatic model) or it will announce
SHOCK
, and then automatically administer a shock without
PREPARING TO SHOCK,
PREPARING TO
requiring further action (fully automatic model).
Do not touch the victim while a shock is delivered.
Regardless of which model you have, continue to follow the
voice instructions.
Do not remove the pads or disconnect them from the defibrillator
until emergency medical personnel arrive. If the victim starts
moving, coughing, or breathing regularly, place the victim in the
recovery position (as instructed in CPR training) and keep him
or her as still as possible.
3-4LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Using the Defibrillator
What to Do After Emergency Medical Personnel Arrive
When emergency medical personnel arrive, tell them what actions you have taken. Tell them how long
the victim has been unconscious, if you delivered shocks, the number of shocks delivered, and if you
performed CPR.
Do not worry if you cannot recall precisely what happened. Your defibrillator makes a digital recording of
heart rhythms and shocks that can be transferred to a computer at a later time. Refer to Section 4 for
information on transferring victim data.
Without removing the electrode pads from the victim, emergency medical personnel can disconnect the
electrode pads from the defibrillator and reconnect them to another defibrillator or AED that has a
compatible QUIK-COMBO cable.
To disconnect the electrode pads:
1 Pull the electrode cable straight out from the defibrillator.
2 Remove the electrode packet anchor pin from the slot in the defibrillator.
3 Press the
ON-OFF button and close the lid to turn off the defibrillator.
What to Do After Using Your Defibrillator
After you use your defibrillator to respond to an SCA emergency, complete the following tasks:
1 If the defibrillator is turned on, press and hold the
ON-OFF button for approximately 2 seconds to turn
it off.
2 Clean the defibrillator and its accessories according to the instructions provided in Table 5-1,
page 5-7. Use only the cleaning agents identified in Table 5-1.
3 Transfer data, if desired.
4 Replace the CHARGE-PAK battery charger. (Refer to page 5-4.)
5 Install a new QUIK-PAK electrode packet. (Refer to page 5-6.)
6 Close the lid and verify that the
OK symbol appears in the readiness display, indicating that the
defibrillator is ready for use. If the attention symbolappears after you replace the battery
charger, the internal battery needs additional time to reach an adequate charge capacity.
7 Dispose of the used electrode pads, any unused spare electrode pads, and the battery charger.
(Refer to "Recycling Information" on page 5-8.)
VOICE INSTRUCTIONS AND TONES
3 Using the Defibrillator
Defibrillator voice instructions provide clear, step-by-step instructions for responding to a victim in
cardiac arrest. In addition, your defibrillator may emit sounds that alert you to the actions that the
defibrillator is performing.
Note: A few seconds may pass between voice instructions and tones. Always wait for further
instructions before taking action.
Note: Some voice instructions will repeat during the defibrillation process.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions3-5
This section explains problems you may encounter while using the LIFEPAK CR Plus or
LIFEPAK EXPRESS defibrillator. For information about keeping your defibrillator in a state of readiness,
refer to Section 5.
Table 3-1 Troubleshooting During Victim Use
ProblemPossible CauseWhat To Do
CHECK PADS FOR GOOD
CONTACT
CONNECTOR
or CHECK
voice
instructions are heard
Defibrillator cannot
deliver the required
shock
Voice instructions sound
faint or distorted
MOTION DETECTED and
STOP MOTION voice
instructions are heard
Inadequate connection to the defibrillator •Be sure that the connector is
completely inserted.
Electrode pads are not properly adhered
to the victim
•Firmly press the pads on the
victim’s skin.
•Clean, shave, and dry the
victim’s skin before placing
pads on skin.
Electrode pads are dry, damaged, or
•Replace the pads.
have passed the expiration date
Electrode pads are not removed from the
blue plastic
•Remove the pads from the
blue plastic and apply them
to the victim’s chest.
Defibrillator internal battery power is low•Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.
Defibrillator internal battery power is low•Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.
Victim movement because of location•Move the victim to a stable
location, if possible.
Victim movement because of breathing•Stop CPR during analysis.
•Check victim for normal
breathing.
Vehicle motion•Stop vehicle during analysis,
if possible.
Electrical/radio frequency interference•Move communication or
other suspected devices
away from the defibrillator
when possible.
3-6LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Using the Defibrillator
Table 3-1 Troubleshooting During Victim Use (Continued)
ProblemPossible CauseWhat To Do
Defibrillator does not
deliver voice
instructions or beeping
tones after you open it
(turn it on)
The readiness display
is blank
Depleted internal battery•Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.
•Replace the CHARGE-PAK
battery charger as soon as
possible. After the
OK symbol
appears on the readiness
display, return the defibrillator
to service.
•Contact authorized service
personnel.
Speaker system failure•Administer CPR if the victim
is not responding, not
breathing normally, or not
moving.
•Contact authorized service
personnel.
The defibrillator has been turned on•Normal condition when the
defibrillator is in use.
Operating temperature is too low or too
high
•Operate the defibrillator
within 0° to 50°C (32° to
122°F).
LCD not operating properly•Contact authorized service
personnel.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions3-7
This section describes the data that the LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators store
when used in an SCA event.
This manual does not provide instructions for how to hand off defibrillation data to the emergency
medical system or hospital personnel. Because this process varies from area to area, check with the
emergency medical system administration for information and directions.
Overview of Data Storagepage 4-2
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions4-1
Each time you use the LIFEPAK CR Plus or LIFEPAK EXPRESS defibrillator, it digitally saves data
about the victim that can be transferred to a PC. This data can be provided to emergency medical
personnel or hospital personnel to aid in case review for quality control, training, and research
purposes. We recommend that you become familiar with their local requirements for reporting a use of
the defibrillator and for providing use data. For assistance in retrieving data from the defibrillator, contact
your local Physio-Control sales or service representative.
Data Stored by Your Defibrillator
Whenever you turn on your defibrillator and connect it to a victim, it automatically stores data about the
victim. When this data is transferred to a data management system for review, three reports are
available: an Event Log, Continuous ECG, and a Summary. Table 4-1 describes these reports.
Table 4-1 Patient Reports
Report TypeDescription
Event LogA chronological log of all events. An event is a condition noted by the defibrillator.
Events are listed on page 4-3.
Continuous ECGTwenty minutes of the victim’s ECG rhythm beginning when the victim is
connected to the defibrillator and ending when the defibrillator is turned off.
SummaryCombines the Event Log and a sampling of continuous ECG rhythms associated
with certain events.
Your defibrillator can store up to two records: one for the current victim and one for the previous victim.
When you use your defibrillator
to free up storage.
The Complete Record for the current victim includes the Continuous ECG and Event Log. If you treat a
second victim, the first victim’s Complete Record will be reformatted into a Summary report. If you treat
a third victim, all of the first victim’s data will be deleted and the second victim’s Complete Record will be
reformatted into a Summary Report. Refer to Table 4-2.
Table 4-2 Defibrillator Patient Records
Complete RecordSummary
Current Victim
Previous Victim
If you turn your defibrillator on and off without attaching electrode pads to a victim, the defibrillator will
not create a new record and the records in the defibrillator will not be altered. The defibrillator deletes
previous data only after it is connected to a new victim.
After you transfer data records to a PC, the defibrillator will disallow repeat transmissions. However,
service personnel may access device records, if necessary.
, it is important that you transfer this data as soon as possible after use
✔✔
✔
Test and Service Data
Your defibrillator stores a test log consisting of the most recent auto-tests, power cycles, and
CHARGE-PAK battery charger replacements. The test log lists the test results and any errors detected.
The test log data is available only to service personnel or to users through the data management
system.
4-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Data Storage
Event and Test Log
Table 4-3 lists the types of events that may be annotated on event and test log reports.
Table 4-3 Event and Test Log Reports
Event LogTes t Log
Power OnSelf-Test Power On
Connect ElectrodesSelf-Test Pass/Fail
Patient ConnectedUser Power On
Initial Rhythm*CHARGE-PAK Changed
Analysis X*CHARGE-PAK
Shock AdvisedFault Log
Charge Complete
SHOCK X-XXXJ*
Shock X Abnormal
4 Data Storage
No Shock Advised
CPR Prompt
Stop CPR Prompt
Check Patient*
Charge Removed
Low Battery
Motion
Analysis Stopped*
Out of Event Memory
Out of Waveform Memory
Power Off
*These events include ECG samples in the Summary Report.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions4-3
This section explains how to help keep your LIFEPAK CR Plus or LIFEPAK EXPRESS defibrillator in
good working condition. Cared for properly, your defibrillator is built to give you many years of service.
Maintaining a State of Readinesspage 5-2
5 Caring for Your Defibrillator
Replacing the CHARGE-PAK Battery Charger
5-2
and the QUIK-PAK Electrode Packet
Cleaning Your Defibrillator5-7
Storing Your Defibrillator5-7
Recycling Information5-8
Supplies, Accessories, and Training Tools5-8
Warranty Information5-8
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions5-1
Your LIFEPAK CR Plus or LIFEPAK EXPRESS defibrillator does not require routine maintenance. It
performs an automatic self-test once a week and every time you turn it on. The electrode indicators
briefly flash during the test. If the automatic self-test detects a condition that requires attention, the
symbol in the readiness display will fade and either the
CHARGE-PAK symbol, the ATTENTION symbol, or
the WRENCH symbol will appear, depending on the type of condition detected.
On a regular basis, you should do the following:
•Check to make sure that the
OK symbol is visible in the readiness display.
•Check the Use By date on the electrode packet (visible through the defibrillator lid in the upper right
corner) and all other electrode packets. If the date has passed, replace the electrode packet and the
battery charger. Refer to "Replacing the CHARGE-PAK Battery Charger and the QUIK-PAK
Electrode Packet" below for more information.
•Check other emergency supplies that may be stored with your defibrillator.
When establishing your local inspection schedule, consider how often your defibrillator will be used and
how familiar the operators are with using a defibrillator. For example, if the defibrillator is used only
rarely, monthly inspections may be appropriate. An inspection checklist is provided in Appendix C.
OK
REPLACING THE CHARGE-PAK BATTERY CHARGER AND THE QUIK-PAK
ELECTRODE PACKET
The CHARGE-PAK battery charger is a replaceable, nonrechargeable battery cell that charges your
defibrillator’s internal battery. The internal battery provides the power for your defibrillator. To prevent
damage to the internal battery, always keep the battery charger in place, including during storage or
shipping.
The QUIK-PAK electrode packet contains the pads that transfer the defibrillation energy to the victim.
The packet should remain connected to the defibrillator and unopened until required for an SCA
emergency. QUIK-PAK electrode pads are not reusable.
When installed, these two accessories allow your defibrillator to stand by for use for approximately two
years. The electrode packet Use By date is programmed into the battery charger. When the date is
reached, the CHARGE-PAK symbol appears in the readiness display, indicating both the battery
charger and electrode packet need to be replaced.
WARNING!
Possible device shutdown.
Always replace the CHARGE-PAK battery charger and QUIK-PAK electrodes at the same time.
The CHARGE-PAK battery charger and QUIK-PAK electrodes are a set and have the same expiration
date. Always replace the CHARGE-PAK battery charger and QUIK-PAK electrodes at the same time to
keep the replacement cycles in sync.
Use the Physio-Control replacement kit to replace the CHARGE-PAK battery charger and the
QUIK-PAK electrode packet as follows:
•After using the defibrillator
•If the CHARGE-PAK symbol appears in the readiness display
•When the Use By date is reached or passed
5-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Caring for Your Defibrillator
The replacement kit includes a CHARGE-PAK battery charger, one or two QUIK-PAK electrode packets,
and a CHARGE-PAK battery discharger. The discharger depletes a used battery charger so that it is
ready for recycling or disposal.
Note: Always keep the CHARGE-PAK battery charger with its original defibrillator. When a
CHARGE-PAK battery charger is inserted into a defibrillator, it becomes internally linked to that
device. If the CHARGE-PAK battery charger is inserted into a second device, the device will not
function correctly. If this occurs, obtain a new replacement kit and replace both the CHARGE-PAK
battery charger and the QUIK-PAK electrodes at the same time.
Follow instructions provided in the replacement kit for battery charger and electrode packet
recycling/disposal.
WARNING!
Possible explosion or fire.
The CHARGE-PAK battery charger is not rechargeable. Do not attempt to recharge, open, crush, or
burn the battery, or it may explode or catch fire.
5 Caring for Your Defibrillator
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions5-3
1 Press the release latch (in the direction of the arrow) to
remove the used battery charger. The battery charger
springs outward from the defibrillator.
2 Insert the new battery charger into your defibrillator and
push until you hear it click into position.
3 Confirm that the symbol disappears and that the OK
symbol appears in the readiness display.
Note: If the symbol appears after you replace the
battery charger, the internal battery is very low and needs
time to charge. It may take up to three days if you had the
defibrillator on for a long time or if you delivered many
shocks. The OK symbol appears when the internal battery
is charged.
Remember: If the defibrillator is needed for an emergency,
attempt to use it even if the symbol is visible.
CAUTION!
Keep the defibrillator at temperatures between 0° – 50°C (32° – 122°F) while the new battery charger
charges the internal battery. The internal battery may not charge effectively at lower temperatures.
Temperatures exceeding 50°C (122°F) for longer than seven days can permanently damage the
internal battery.
5-4LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
To discharge and dispose of a used battery charger:
1 Insert the discharger into the used battery
2 Let the discharger fully deplete the battery
Caring for Your Defibrillator
charger.
Note: Do not attempt to remove the discharger
once it is in place.
5 Caring for Your Defibrillator
charger. Wait at least 9 days.
3 Place the used battery charger in the trash or
recycle it.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions5-5
defibrillator lid (voice instructions will
sound).
2 Press and hold down the
ON/OFF
button for 2 seconds to turn off the
defibrillator and save battery power.
3 Remove the outdated or used
electrode packet:
a Unplug the electrode connector
from the connector receptacle.
b Slide the anchor pin from the slot.
c Discard the outdated or used
electrode packet according to
local regulations.
4 Install the new electrode packet:
a Slide the anchor pin into the slot.
b Plug the electrode connector into
the receptacle.
c Ensure that the new electrode
packet is centered on the
defibrillator and is tucked behind
the lip before closing the lid.
d Close the lid. Confirm that the
packet Use By date is visible
5-6LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
through the upper right-hand
corner of the lid.
Caring for Your Defibrillator
STORING YOUR DEFIBRILLATOR
Always store your defibrillator according to the temperature ranges recommended in Appendix A,
"Specifications".
WAR NIN G!
Possible fire or explosion.
Do not store this defibrillator in the presence of flammable gases or in direct contact with flammable
material.
CLEANING YOUR DEFIBRILLATOR
CAUTION!
Possible equipment damage.
Do not clean any part of the defibrillator or accessories with bleach, bleach dilution, or phenolic
compounds. Do not use abrasive or flammable cleaning agents. Do not steam, autoclave, or
gas-sterilize the defibrillator or its accessories.
Table 5-1 Cleaning Methods
ItemCleaning MethodCleaning Agent
Exterior case, readiness
display, and crevices
CHARGE-PAK battery
charger
Electrode PadsNone, do not remove electrode
Clean with damp sponge or
cloth
Nonabrasive soap and water
Quaternary ammonium compounds
Rubbing (isopropyl) alcohol
Peroxide (peracetic acid) solutions
NoneNone, dispose of/recycle after use
None, dispose of/recycle after use
pads from the packet
5 Caring for Your Defibrillator
Carrying caseWipe with damp cloth or spongeWater
Quick Reference CardWipe with damp cloth or spongeWater
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions5-7
Do not disassemble the defibrillator. It contains no operator-serviceable components and dangerous
high voltages may be present. Contact authorized service personnel for repair.
The WRENCH indicator appears on the readiness display if the defibrillatorrequires service. Contact
authorized service personnel only. In the USA, contact Physio-Control at 1.800.442.1142. Outside the
USA, contact your local Physio-Control representative. Be prepared to provide the following information:
•Model number and MIN number (part number)
•Serial number
•Description of the problem based on your observations
RECYCLING INFORMATION
Recycle the defibrillator and its accessories at the end of their useful lives.
Recycling Assistance
Items should be recycled according to national and local regulations. Contact your local Physio-Control
representative for assistance or refer to http://recycling.medtronic.com for instructions on disposing of
this product.
Preparation
Items should be clean and contaminant-free prior to being recycled.
Recycling of Disposable Electrodes
After using disposable electrodes, follow your local clinical procedures for recycling.
Packaging
Packaging should be recycled according to national and local regulations.
SUPPLIES, ACCESSORIES, AND TRAINING TOOLS
If possible, we recommend that you have an extra QUIK-PAK Electrode Packet and CHARGE-PAK
Battery Charger on hand. In addition, there are other useful accessories available. For example, you
can buy Infant/Child Reduced Energy Defibrillation Electrodes for use with your defibrillator on children
who are less than 8 years of age or weigh less than 25 kg (55 lb). Contact your local Physio-Control
representative.
WARRANTY INFORMATION
Refer to the warranty statement included in the accessory kit shipped with this product. For duplicate
copies, contact your local Physio-Control representative. In the USA, call 1.800.442.1142. Outside the
USA, contact your local Physio-Control representative.
5-8LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
DEFIBRILLATOR OPERATING SETTINGS6
This section introduces the operating settings that are adjustable on the LIFEPAK CR Plus and
LIFEPAK EXPRESS defibrillators.
Operating Settings and Setup Configurationpage 6-2
6 Defibrillator Operating Settings
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions6-1
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators have several operating settings. The
settings range from defining the time and date used by the defibrillator, to the energy sequence and
protocol of the shocks delivered. This group of operating settings are the defibrillator’s setup
configuration.
Table 6-1 identifies operating settings in the setup configuration, describes each setting, including
possible options, and identifies the preset defaults. These settings can be specially configured in
accordance with customer order.
Tab le 6-1 Operating Settings
Operating SettingsDescription Default Setting
Device IDDevice ID is a unique identifier (ID) assigned to
each defibrillator used for tracking the location of
defibrillators. When you transfer event data from
the defibrillator to a PC, the defibrillator ID will be
included in the transferred data.
Energy SequenceEnergy sequence defines the energy levels used
by the defibrillator.
The energy level choices are: 150, 175, 200, 225,
250, 275, 300, 325, 360.
Energy ProtocolEnergy protocol determines how the defibrillator
delivers successive shocks. There are two options
for this setting: flexible or fixed.
Flexible sequence means the energy delivered for
a shock increases only if an analysis immediately
following a shock results in another
SHOCK ADVISED
decision. For example, if the defibrillator energy
sequence is set up as 200, 300, 360, flexible
sequence means that the energy delivered for the
first shock is 200 joules. If the arrhythmia is
terminated by shock 1 and the next analysis results
NO SHOCK ADVISED decision, the energy will not
in a
increase for the next shock. However, if the
arrhythmia is not terminated by shock 1 and the
next analysis results in a
SHOCK ADVISED decision,
the energy will increase to 300 joules, and so on.
Fixed sequence means that the energy delivered
after the first shock of 200 joules increases from
200 to 300, and then to 360 joules, regardless of
the post-shock ECG rhythm and subsequent
analysis.
Serial Number
Level 1 – 200 joules
Level 2 – 300 joules
Level 3 – 360 joules
Flexible
CPR Time 1
CPR Time 2
The CPR Time 1 and CPR Time 2 settings define
the time interval for performing CPR after a shock
or after a no shock advised decision.
CPR Time 1–120 sec
CPR Time 2–120 sec
The choices for CPR Time 1 and CPR Time 2 are
15, 30, 45, 60, 90, 120, and 180 seconds.
6-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Defibrillator Operating Settings
Tab le 6-1 Operating Settings (Continued)
Operating SettingsDescription Default Setting
Device Date
Device Time
The device date and device time are used to time
stamp patient reports and to control when the
automatic self tests are performed. The date and
time can be set when you retrieve data from the
defibrillator.
Turn-On PromptThis setting determines whether you hear only
tones, or tones and the voice prompt,
HELP NOW
, when you open the lid and the
CALL FOR
defibrillator turns on. Choices are Voice and Tones.
Voice Prompt VolumeThis setting sets the voice prompt volume
to MEDIUM or HIGH.
Pulse PromptThe pulse prompt setting determines whether the
defibrillator prompts you to check the victim’s pulse
(appropriate for medically trained users) or to
check the victim for signs of circulation such as
breathing and movement (appropriate for lay
users). Choices are Check Pulse, Check
Breathing, or Check Circulation.
Stack ShocksWhen set to OFF, the Stack Shocks option
eliminates the analysis after each shock and
inserts prompting for CPR after each (a single)
shock. This eliminates the three-shock stack. CPR
is prompted regardless of the ECG rhythm after the
shock. The CPR time following the shock is
determined by the CPR Time 1 setting selected.
Choices for Stack Shocks option are ON or OFF.
When set to ON, an analysis will occur after
shocks, and up to three shocks in a row may be
delivered (three-shock stack).
Pacific Standard Time
and Date
Voic e
High
Per customer order
Off
6 Defibrillator Operating Settings
Pulse CheckWhen set to Never, the Pulse Check option
Never
removes all prompting for pulse checks. The other
Pulse Check settings available allow pulse checks
only after every No Shock Advised (NSA) decision,
after the second NSA decision and thereafter, or
Always (after shocks, NSA and CPR).
Motion DetectionThe motion detection setting is used to determine if
On
motion detection is active or not active during
analysis.
When motion detection is On, the defibrillator stops
analysis for up to 10 seconds if it detects any victim
motion. The defibrillator notifies the responder of
the problem. The defibrillator will resume analysis
after 10 seconds, even if motion is still present.
When motion detection is Off, analysis is not
inhibited, regardless of any victim motion.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions6-3
Note: Table values are nominal for a 360 joule energy setting.
Figure A-2 Biphasic Waveform
Shock Advisory System: An ECG analysis system that advises whether a shock is appropriate, meets
rhythm recognition criteria specified in the American Association of Medical
Instrumentation standard DF39.
The device allows a defibrillation shock only if the Shock Advisory System
advises defibrillation.
Device Capacity:LIFEPAK CR Plus defibrillator — Thirty (30) full discharges or 210 minutes
of “ON time” with a fully charged device.
LIFEPAK EXPRESS defibrillator — Twenty (20) full discharges or
140 minutes of “ON time” with a fully charged device.
A-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Charge Ready Time:
Specifications
Appendixes
System Recharge
Times:
Timing
Parameter
Analysis timeLess than 10
200 Joules360 JoulesComments
seconds
Less than 10
seconds
Analysis time applies if no
motion is detected. If the
device detects motion,
analysis may be delayed by
up to 10 seconds.
Defibrillator
charge time
Less than 9
seconds
Less than 15
seconds
Defibrillator charge time
applies when the device is
fully charged or has been
discharged up to 15 times.
Power on to
charge ready
time
Less than 26
seconds
Less than 32
seconds
Power on to charge ready
time applies when a
shockable rhythm is detected
and no motion is detected,
and when the device is fully
charged or has been
discharged up to 15 times.
Recharge times with a fully discharged device:
Able to deliver 6 shocks or provide 42 minutes of operating time after
24 hours of recharge and 20 shocks or 140 minutes of operating time after
72 hours of recharge time with a new CHARGE-PAK battery charger at
temperatures above 15°C (59°F).
Minimum 20 minutes of ECG stored for the current victim.
Summarized data stored for the previous victim.
Report Types:Continuous ECG—A continuous ECG report for the victim.
Summary—A summary of critical resuscitation events and ECG waveform
segments associated with these events.
Event Log report—A report of time stamped markers, which reflect operator
and device activity.
Test Log report—A device self test activity report.
Capacity:Minimum 200 time stamped Event Log markers.
Communications:Wireless transfer to a personal computer.
Data Review:Physio-Control provides an array of tools to meet customer needs for data
viewing and analysis.
A-6LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Specifications
CLINICAL SUMMARY: DEFIBRILLATION OF VENTRICULAR FIBRILLATION AND
VENTRICULAR TACHYCARDIA
Background
Physio-Control conducted a multi-centered, prospective, randomized and blinded clinical trial of
biphasic truncated exponential (BTE) shocks and conventional monophasic damped sine wave (MDS)
shocks. Specifically, the equivalence of 200 J and 130 J BTE shocks to 200 J MDS shocks
1
was tested.
Methods
Ventricular fibrillation (VF) was induced in 115 patients during evaluation of implantable cardioverter
defibrillator function and 39 patients during electrophysiologic evaluation of ventricular arrhythmias.
After 19±10 seconds of VF, a customized defibrillator delivered an automatically randomized shock.
Efficacy was based on success of this shock. To demonstrate equivalence of test shocks to control
shocks, the 95% upper confidence limit of the difference in efficacy (95UCLD), control minus test, was
required to be less than 10%.
Appendixes
Results
Ventricular Fibrillation
The efficacy of the 200 J BTE shocks was demonstrated to be at least equivalent to the efficacy of 200 J
MDS shocks (95UCLD=2%). The difference is success rates of 200 J MDS minus 200 J BTE shocks
was -10% (exact 95% confidence interval from -27% to 4%). The 130 J BTE shocks were not
demonstrated equivalent to 200 J MDS shocks (95UCLD=22%). However, neither was their efficacy
significantly lower than that of the 200 J MDS shocks (statistical power limited by small sample sizes).
For all shock types, hemodynamic parameters (oxygen saturation and systolic and diastolic blood
pressure) were at or near their pre-induction levels by 30 seconds after successful shocks.
Shock
200 J MDS61/68 (90%) 80 to 96%
200 J BTE39/39 (100%)91 to 100%
130 J BTE39/47 (83%)69 to 92%
Ventricular Tachycardia
Seventy-two episodes of ventricular tachycardia (VT), induced in 62 patients, were treated with
randomized shocks. High rates of conversion were observed with biphasic and monophasic shocks.
Sample sizes were too small to statistically determine the relationship between success rates of the
waveforms tested.
Ventricular Fibrillation
st
1
Shock Success
Exact 95% Confidence Interval
1
S.L. Higgins et al., “A comparison of biphasic and monophasic shocks for external defibrillation,” Prehospital Emergency Care, 2000, 4(4):305-13.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating InstructionsA-7
In this double-blinded study, the efficacy of the 200 J BTE shocks was demonstrated to be at least
equivalent to the efficacy of 200 J MDS shocks for defibrillation of short duration, electrically-induced
VF. However, the comparison of efficacy of 130 J biphasic and 200 J monophasic shocks for VF was
inconclusive. All waveforms tested provided a high rate of termination of VT. The VT sample sizes were
too small to statistically determine the relationship between VT success rates of the waveforms tested.
Compared to conventional shocks for VF, we found no positive or negative effect of biphasic shocks for
VF on hemodynamic parameters following the defibrillating shock. It is possible that, compared to 200 J
monophasic shocks, 200 J biphasic shocks will in some cases enable earlier termination of VF.
Therefore, we conclude that biphasic shocks for VF delivered at conventional energy levels have the
potential to improve outcome in resuscitation of victims with sudden cardiac arrest.
A-8LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
APPENDIX B
SHOCK ADVISORY SYSTEMB
Appendixes
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
The Shock Advisory System (SAS) is an ECG analysis system built into the LIFEPAK CR Plus and
LIFEPAK EXPRESS defibrillators that advises the operator if it detects a shockable or nonshockable
rhythm. This system makes it possible for individuals not trained to interpret ECG rhythms to provide
potentially-lifesaving therapy to victims of ventricular fibrillation or pulseless ventricular tachycardia. The
SAS contains the following features:
•Electrode contact determination
•Automated interpretation of the ECG
•Operator control of shock therapy
•Motion detection
Electrode Contact Determination
The victim's transthoracic impedance is measured through the defibrillation electrodes. If the baseline
impedance is higher than a maximum limit, it is determined that the electrodes are not in sufficient
contact with the victim or not properly connected to the defibrillator. ECG analysis and shock delivery
are inhibited. The operator is advised to connect electrodes any time electrode contact is inadequate.
Appendixes
Automated Interpretation of the ECG
The Shock Advisory System is designed to recommend a shock if it detects the following:
•
Ventricular fibrillation — with a peak-to-peak amplitude of at least 0.08 mV
Ventricular tachycardia — defined as having a heart rate of at least 120 beats per minute, QRS
•
width of at least 0.16 seconds, and no apparent P waves.
Pacemaker pulses may prevent advisement of an appropriate shock, regardless of the victim’s
underlying rhythm. The SAS is designed to recommend no shock for all other ECG rhythms including
pulseless electrical activity, idioventricular rhythms, bradycardia, supraventricular tachycardias, and
normal sinus rhythms.
ECG analysis is performed on consecutive 2.7 second segments of ECG. The analysis of two out of
three segments must agree before a decision (
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillator SAS performance for adult, pacemaker and
pediatric ECGs is summarized in the following table.
Table B-1 LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance Table for Adult ECGs
Rhythm Class
Shockable:
ECG Test1
Sample Size
168>90% sensitivity100.0% [98.6%]
coarse VF
Shockable:
65>75% sensitivity84.6% [77.3%]
shockable VT
SHOCK ADVISED or NO SHOCK ADVISED) is made.
Observed Performance
Performance Goal2,
3
Sensitivity or Specificity [LCL]
4
Nonshockable:
NSR
Nonshockable:
144>99% specificity for
100.0% [98.4%]
NSR (AHA)
43>95% specificity100.0% [94.8%]
asystole
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating InstructionsB-1
Table B-1 LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance Table for Adult ECGs (Continued)
Rhythm Class
Nonshockable:
ECG Test1
Sample Size
Performance Goal2,
531>95% specificity95.9% [94.5%]
Observed Performance
3
Sensitivity or Specificity [LCL]
4
all other rhythms
Intermediate:
29Report only96.6% [87.2%] sensitivity
fine VF
1
From Physio-Control ECG database. Each sample is run 10 times asynchronously.
2
Association for the Advancement of Medical Instrumentation. DF39-1993 Standard for Automatic External Defibrillators
and Remote-Control Defibrillators. Arlington, VA: AAMI;1993.
3
Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting
Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety. AHA Task Force on
Automatic External Defibrillation, Subcommittee on AED Safety and Efficacy. Circulation, 1997, Vol. 95, 1677-1682.
4
LCL = 90% exact one-sided lower confidence limit
VF = ventricular fibrillation
VT= ventricular tachycardia
NSR = normal sinus rhythm
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators were also tested using paced rhythms
recorded at high-fidelity from victims with implanted pacemakers. The high-fidelity pacemaker spikes
were also added to samples of ventricular fibrillation to test the defibrillator’s ability to reach a shock
decision in the case of ventricular fibrillation with an implanted, active pacemaker. The results are
summarized in the following table.
Table B-2 LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance with Active Pacemakers
Rhythm Class
Shockable:
ECG Test
Sample Size
Performance GoalObserved Performance
35>90% sensitivity91.4% [81.9%]
Coarse VF
Nonshockable:
35>95% specificity100.0% [93.6%]
Paced rhythms
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators were also tested using ECGs acquired
from hospitalized pediatric victims ranging in age from < 1 day old to 17 years old. The results are
summarized in the following table.
Table B-3 LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance Table for Pediatric ECGs
Rhythm Class
Shockable:
ECG Test
Sample Size
1
Performance Goal
90>90% sensitivity100.0% [97.5%]
Observed Performance
2
Sensitivity or Specificity [LCL]
3
coarse VF
Shockable:
11>75% sensitivity54.5% [31.8%]
shockable VT
Nonshockable:
424>99% specificity 100.0% [99.5%]
NSR
Nonshockable:
95>95% specificity100.0% [97.6%]
asystole
B-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
Shock Advisory System
Table B-3 LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator SAS Performance Table for Pediatric ECGs (Continued)
Rhythm Class
Nonshockable:
ECG Test
Sample Size
1
Performance Goal
433>95% specificity99.3% [98.5%]
Observed Performance
2
Sensitivity or Specificity [LCL]
3
all other rhythms
Intermediate:
4Report only100.0% [56.2%] sensitivity
fine VF
Intermediate:
7Report only42.9% [17.0%] specificity
other VT
1
From Physio-Control ECG database.
2
Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting
Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety. AHA Task Force on
Automatic External Defibrillation, Subcommittee on AED Safety and Efficacy. Circulation, 1997, Vol. 95, 1677-1682.
3
LCL = 90% exact one-sided lower confidence limit.
Control of Shock Therapy
The Shock Advisory System causes the defibrillator to charge automatically when it detects the
presence of a shockable rhythm. When a shockable rhythm is detected, the defibrillator automatically
delivers a shock or instructs the user to deliver the shock by pressing the shock button.
Appendixes
Note: If the shock button is not pressed within 15 seconds, the system disarms and repeats the
analysis. The system also disarms if the patient impedance decreases suddenly or the patient
impedance goes outside the acceptable range for analysis. If none of these occur, the shockable rhythm
decision is not revised during charging or prior to shock.
Motion Detection
The Shock Advisory System detects victim motion independent of ECG analysis. A motion detector is
designed into the defibrillator.
"Defibrillator Operating Settings".
A number of activities can create motion, including CPR, rescuer movement, patient movement, vehicle
movement, and some internal pacemakers. If variations in the transthoracic impedance signal exceed a
maximum limit, the Shock Advisory System determines that patient motion of some kind is present. If
motion is detected, the ECG analysis is inhibited. The operator is advised by a displayed message, a
voice prompt, and an audible alert. After 10 seconds, if motion is still present, the motion alert stops and
the analysis always proceeds to completion. This limits the delay in therapy in situations where it may
not be possible to stop the motion. However, the rescuer should remove the source of motion whenever
possible to minimize the chance of artifact in the ECG.
There are two reasons why ECG analysis is inhibited when the motion alert occurs, and why the rescuer
should remove the source of the motion whenever possible:
•Such motion may cause artifact in the ECG signal. This artifact may occasionally cause the Shock
Advisory System to reach an incorrect decision.
•The motion may be caused by a rescuer's interventions. To reduce the risk of inadvertently shocking
a rescuer, the motion alert prompts the rescuer to move away from the patient. This will stop the
motion and ECG analysis will proceed.
MOTION DETECTION can be configured to be ON or OFF. Refer to Section 6,
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating InstructionsB-3
Table D-1 Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are intended for use in the electromagnetic
environment specified below. The customer or the user of the defibrillator should ensure that the defibrillator is
used in such an environment.
Group 1The defibrillator uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class BThe defibrillator is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
Not Applicable
Not Applicable
buildings used for domestic purposes.
Essential Performance
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators maintain safe and effective performance of the
defibrillation therapy and patient monitoring functions when operated in the electromagnetic environment
specified in Tables 2 through 4.
Limitations Affecting Immunity to Electromagnetic Disturbances
The level of protection from electromagnetic disturbances is limited by several factors, including requirements
for protection from third-party defibrillators, patient safety isolation, and maintenance of adequate
signal-to-noise ratios for processing patient signals.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating InstructionsD-1
Table D-2 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are intended for use in the electromagnetic
environment specified below. The customer or the user of the defibrillator should ensure that the defibrillator is
used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
IEC 60601
Test Level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
(>95% dip in UT)
for 0.5 cycle
40% U
T
(60% dip in UT)
for 5 cycles
70% U
T
(30% dip in UT)
for 25 cycles
<5 % U
T
(>95% dip in UT)
for 5 s
Compliance Level
±8 kV contact
±15 kV air
Electromagnetic Environment –
The defibrillator is suitable for use in a
dry environment.
Not ApplicableNot Applicable
Not ApplicableNot Applicable
Not ApplicableNot Applicable
Guidance
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
D-2LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
is the a.c. mains voltage prior to application of the test level.
T
3 A/m3 A/mPower frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Electromagnetic Compatibility Guidance
d1.2 P=
d2.3 P=
d7.7 P=
Table D-3 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are intended for use in the electromagnetic
environment specified below. The customer or the user of the defibrillator should ensure that the defibrillator is
used in such an environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the
defibrillator, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
10 Vrms
Not ApplicableNot Applicable
*
Not ApplicableNot Applicable
150 kHz to 80 MHz
*
10 V/m
80 MHz to 2.5 GHz
80 MHz to 800 MHz
Radiated RF
IEC 61000-4-3
in ISM bands
10 V/m
80 MHz to 2.5 GHz
for specified frequencies in the range
800 MHz to 2.5 GHz
870 MHz to 900 MHz
Where P is the maximum output power rating of the
transmitter in watts (w) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
†
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
‡
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
*
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to
13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
†
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are
intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently
brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in these frequency ranges.
‡
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitter, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the defibrillator is used exceeds the applicable RF compliance level above, the defibrillator should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
defibrillator.
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating InstructionsD-3
Table D-4 Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the
LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators
Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators
The LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of the defibrillator can
help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the defibrillator as recommended below, according to the
maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum
output power of
transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 80 MHz to 870 MHz,
900 MHz to 2.5 GHz
870 MHz to 900 GHz
W
0.01Not Applicable0.120.230.77
0.1Not Applicable0.380.732.43
1Not Applicable1.22.37.7
10Not Applicable3.87.324.3
100Not Applicable122377
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to
2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if
it is inadvertently brought into patient areas.
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
D-4LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions
INDEX
Index
A
About Automatic External
Defibrillators 1-2
About the Defibrillators 1-10
Accessories 5-8
ADAPTIV 1-10, 1-11
AED 1-2, 1-3
Attention Indicator 2-4, 5-2
B
Basic Steps 3-3
Blue Plastic 2-6, 2-7
C
Cardiac Arrest 1-3
Caring for the Defibrillator 5-1
Carrying Handle 2-4, 2-5
CHARGE-PAK 1-10, 2-4, 2-5
CHARGE-PAK Indicator 2-4, 5-2
Cleaning Methods 5-7
Cleaning the Defibrillator 5-7
Clinical Summary A-7
Continuous ECG 4-2
Controls 2-4
CPR 1-3
CPR Time 6-2
D
Data Storage 4-2
Data Storage Characteristics A-5
Data Stored by the
Defibrillator 4-2
Data Transfer
Current Record 4-2
Deleting Data 4-2
Previous Record 4-2
Storing Data 4-2
Recycling/Disposal 5-8
Electrode Pads 2-6, 2-7
Electrodes, Infant/Child 1-11
Energy Protocol 6-2
Energy Sequence
Energy Levels 6-2
Environmental Specifications A-4
Event and Test Log 4-3
Event and Test Log Reports 4-3
Event Log 4-2
F
Fibrillation 1-3
G
Getting Started 2-1
H
Heart Attack 1-3
I
Impedance 1-3
Indications for Use 1-2
Indicators 2-4
Infant/Child Electrodes 1-11
Inside Features 2-5, 2-6
Inspecting the Defibrillator 2-4
IrDA Port 1-10, 2-4, 2-5
J
Joule 1-3
L
Labels 2-4
LED 1-3
Lid 2-4
LIFEPAK Defibrillator
Accessories 1-10
Automated Operation
Fully Automatic 1-10
Semiautomatic 1-10
Automatic Self-Test 1-10
LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating InstructionsIndex-1
Obtaining Authorized Service 5-8
OK Indicator 2-4, 5-2
ON-OFF Button 2-4
Opening the Electrode
Packet 3-3
Operating Settings 6-2
About 6-1
Operating Settings and Setup
Configuration 6-2
Operator Settings
CPR Time Settings 6-2
Device Date 6-3
Device ID 6-2
Device Time 6-3
Energy Protocol 6-2
Energy Sequence 6-2
Pulse Check 6-3
Pulse Prompt 6-3
Stack Shocks 6-3
Time Zone 6-4
Voice Prompt Volume 6-3
Troubleshooting 3-6
Recycling Information 5-8
Regular Maintenance 5-2
Reports Stored 4-2
Responder 1-3
Responding to a Sudden Cardiac
Arrest Victim 3-2
S
Safety Information 1-4
Safety Terms 1-4
Safety Warnings 2-4, 2-5
SAS 1-3
Serial Number Label 2-4, 2-5
Service 5-8
Shock Advisory System B-1
Shock Button 2-6, 2-7
Shockable Rhythm 1-3
Speaker 2-6, 2-7
Specifications A-1
Stack Shocks 6-3
Storing the Defibrillator 5-2, 5-7
Summary 4-2
Supplies 5-8
Symbols 1-7
T
Te r m in o l o gy 1 - 3
Test and Service Data 4-2
Text Conventions 1-4
Time Zone 6-4
Training Tools 5-8
Treating a Victim 3-3
Troubleshooting 3-6
Troubleshooting During Victim
Use 3-6
U
Unpacking and Inspecting 2-2
Use By Date 2-5, 2-6
User 1-3
User Interface Specifications A-3
Using the Defibrillator 3-1, 3-3