Physio-Control LIFEPAK 9P User manual

r/pacemaker
0
brillator/mon
Service
Manual October
Corporate
11811 Post
Redmond,
Telephone: Toll Telex: Telefax:
No.
1993
Headquarters:
Willows
Office
Box
WA
Free:
206/867-4000
800/442-1142
990211 D PHYSIO
206/883-2988
Manual
805454-01
Road
Northeast
97006
98073-9706
USA
or
800/426-8047
RDMD
LIFEPAK
9P
defibrillator/monitor/pacemaker
Part
No.
Serial
No.
About by test,
monitor/pacemaker.
A
for
personnel.
features
well
This
Manual:
technical
troubleshoot,
separate
use
as
service
publication,
by
physicians,
It
of
the
operator-level
provides
This
personnel.
and
clinicians,
LIFEPAK
testing
Service
repair
the
Operating
step-by-step
9P
defibrillator/monitor/pacemaker
Manual
It
describes
the
LIFEPAK
and
emergency
instructions
and
maintenance.
is
intended
how
to
maintain,
9P
defibrillator/
Instructions,
care
for
all
for
use
is
intended
operating
as
Trademarks
Wa
rranty
PHYSIO-CONTROL®:
QUIK-LOOK®
CODE
Physio—Control
Redmond,
Refer with the contact
SUMMARY",
to
the
the
product.
Physio—Control
your
DERMA
Corporation,
WA
98073-9706.
product
Duplicate
PARTSLINE
local
Physio—Control
warranty
LIFEPAK® , FAST-PATCH®,
JEL®,
SAS
and
PARTSLINE"
11811
Willows
statement
copies
may
at
1-800-442-1142.
sales
(SHOCK
included be
or
ADVISORY
are
trademarks
Road
in
obtained
service
N.E.,
the
office.
LIFE*PATCH®,
SYSTEM)",
of
the
PO Box
accessory
in
the
USA
by
Outside
the
97006,
kit
shipped
calling
USA,
ii
All
Rights
Reserved
October
©
Physio—Control
1993
Corporation
Safety
information
General
information
Introduction Terms Warnings Symbols
Service Effective Configuration
Content
......................,...........
...........................................
.............................
Information
Publication
Overview
........................................
Gene
...........................
Dates
Information
........
............................
............................
K
eee
εν
ων ων
xii xii xiii xiv
XV Xvi xvii
xviii
©
October
Description
1993
Physio—Control
Corporation
Introduction
Physical
Functional
Theory Power
Supply
Power-On................
Regulation
Overvoltage
Output Battery
BATT
..........
Description
Description
of
Operation
PCB
Assembly
..
Protection
Isolation
.......................,........,....,,...
CHRG
....................................
Indicator
All
ュー
....................
...............................
ee
1-7
.................................
(803726)
ωωωρεον
...................
eee
εν
νενε
ων εν ο νεν
ον
..............................
..............................
Rights
Reserved
1-1
1-9 1-9
1-9 1-9
1-9
1-9 1-9
1-10
LIFEPAK
9P
defibrillator/monitor/pacemaker
Power
İnterconnect
Main
Display
Conversion Power Logic
Switching
External
Energy Low Capacitor Energy Energy Energy Serial
Analog
ChageLEDDrive..................................
Interface Pace
Pace
Pace Echo Pacing Arming Pace Isolated Switching Power
LED
Control
Power
Energy
Data
Pulse
Pulse
Pulse
Pulse
Leads
Pulse
Supply
Drive
PCB
Control
Storage
Transfer Dump Monitor
Voltage
Circuit
Power
Assembly
PreamD
System
Controller
Display/Recorder/Pacing
PCB/CRT
Video
Signal Frame Raster Raster
Raster CRT
Sweep Scan Voltage Generator
Bias
PCB
Assembly
.....................................
Monitor
Power
Charge
Charge
.................................,,...
Interface
................................
Supply
Input
...........................,...
Capacitor
Rate
Control
....................................
....................................
..........................,.,...
Evaluation
Pacing
.........................,........,.......
Regulator
PCB
Assembiy
Output
Wave
Amplitude
Width
Length
Drive
Shaper
Decoder
Converter
Off
Detector
...............................,,.,..
Generator
Supply
Controller
...............................
................................
(803724)
.............................
Charger
.............................
.....................,..,..,
...............,
.....................
...........................
(805472)
.............................
..............
.............................
........................
..........................
..........,...........,....,
..........................
...........................,.
.............................
(805474)
..........................,.......
Assembly
Amplifier
Generator
Oscillator
Generator
...................................
Generator
.................................
..........................
ーー
バトー
Interface
(803706)
..............................
.............................
...............................
.........................,...
.....................
....................
severe
ーー
こてこて
ーー
1-10 1-10 1-11
1-12
1-12
1-12 1-13 1-14 1-14 1-15 1-16 1-16
1-17
1-17 1-17 1-17 1-19
1-19
1-19 1-20 1-21 1-21 1-22 1-22 1-22 1-22
1-22
1-23
1-25
1-30
1-32
1-39 1-41
1-41 1-41 1-41 1-42 1-42
|
iv
|
All
Rights
Reserved
© October
1993
Physio—Control
Corporation
Operation
2
Introduction
Controls Option
Test
Menu
Setup
Menu
Info
Me
Introduction Performance PIP-Purpose
PIP-Scope
........................................
and
Indicators
Settings
......................................
...........................................
.........................................
...................,....................
Inspection
.........................................
and
Applicability
...............................
Procedure
(PIP)
................
............................
PIP-Definitions . ..........,...........................
PIP-Requirements
PIP-Required PIP-Test
Equipment PIP-Workstation PIP-Personnel
PIP-General
Instructions
PIP-Physical
PIP-AC PIP-Battery PIP-Battery PIP-Setup PIiP-Keypadlesi
PIP-Lead
PIP-Recorder PIP-ECG РР-ПпУ PIP-CRT PIP-Heart PIP-Rate PIP-Recorder PIP-Test PIP-Sync/Sync PIP-Recorder/CRT PIP-Defib PIP-Paddle
PIP-Charge
Ро\ег
Charge Power
Menu
Select
Display САГ
Gain/Aspect
Rate
Alarms/Alarm-Activated
Load
Disarm
Polarity
Time
....................................
Equipment
.............................
Verification
.......................
...................................,
................................,...,
............................,..
Inspection
.........
................................
iie
...................................
.............................,......
Settings
.....................
........
Speed/Baseline
...............................
aa
Centering
...................
................,....................
..........
..
Ratio
erre
............................
Record
Annotation/Real-Time
.................
Clock
................
a
rare
........................................
Annotation
Noise
and
.............................
...............................
CHARGE
LEDs
Check
..............
....................,..............
.....................................
erev
o
2-1
2-1 2-7 2-7 2-8 2-9
3-1 3-1 3-1 3-2 3-2 3-2 3-2 3-3 3-3
3-3 3-4
3-4
3-5 3-5 3-5 3-6 3-7
3-8
3-8
3-9 3-9 3-10 3-10 3-10 3-11 3-11 3-11 3-12 3-12 3-12 3-13
O
October
1993
Physio—Control
Corporation
All
Rights
Reserved
LIFEPAK
9P
defibrillator/monitor/pacemaker
PIP-Discharge PIP-Energy
PIP-Charge PIP-Delivered
PIP-CODE PIP-Pacemaker PIP-PacingRate PIP-Demand
PIP-Non-Demand PIP-Pacing PIP-Leakage
PIP-Pacemaker
PIP
Checklist
Test
and TCP-Purpose TCP-Scope TCP-Definitions
TCP-Reguirements..................
TCP-Equipment TCP-Test
TCP-Workstation
TCP-Personnel TCP-Voltage TCP-Low
Monitoring
Defibrillator
TCP-Test TCP-Defibrillator TCP-Test
TCP-Charge
TCP-Output TCP-Display TCP-Power
TCP-Recorder
TCP-Recorder
TCP-CRT
TCP-ECG
TCP-Pacemaker TCP-Pacemaker
TCP
Checklist...........
Troubleshooting
Error
Codes
Control
Dump
Reduction
Energy
SUMMARY
Power-Up
......................,....,......
...........................,..,,....
.....................,...........
.................................
Report
.........................
..........................,...
......................
Pacing
Leads
Current
Calibration
and
Equipment
CheckS
Battery
Menu
Load
Time
Waveform
Calibration
Supply
Frequency
Output
...........................,...,...
Pacing
Off..................................
Leads
....................,..............,...
.................................,......
Applicability
......................................
..........................,....,.....
Threshold
Check
Charge
Access
Calibration
Calibration
...........................,...
..............................,...
Detect
Procedure
Requirements
Verification
Requirements
..................................
............................,.....
Check
................................
....................................
Current
(TCP)
...........................
........................
......................
.......................
Checks
.............................
...........:
...........................
...........................
........................,.,...,.
...............................
Calibration
Calibration
Frequency
Response
..........................
..............................
Response
......................
............,..,..........
.....................,.......,,,,...,
Calibration Current
............................
and
Waveform
TN
Aids
.................................
and
SERVICE
Indicator
......................
0s
...................
..................
ο ο ωρών
Output
κο
νε
ων
ες
e
.............
3-13 3-13
3-14 3-14
3-14 3-16 3-16 3-16
3-17 3-17 3-17
3-18
3-21 3-23
3-23 3-23 3-23 3-23
3-23 3-24
3-24 3-25 3-26 3-27 3-27 3-28 3-29 3-30 3-31 3-32
3-32 3-33 3-37
3-38
3-40
3-42
3-42
3-42 3-43
3-45 3-46
3-46
vi
All
Rights
Reserved
©
October
1993
Physio-Control
Corporation
Service
and
Maintenance
4
E Battery Recharging Battery
Guidelines
Battery
Recycling
General
Operation
Strip
Fuse Special
Devices Look Use Wear Transport Keep
Test
Disassembly
Battery
Front Strip Case Power
Main Display Power
Interconnect
Inspection
Exterior
Interior Tools Cleaning
External
Interior
PCB
Preparation
......................
Maintenance
Batteries
Maintenance
for
Replacement
Capacity
Batteries
Maintenance . ................................
from
AC
Chart
Recorder Loading Cleaning
Protecting
Static-Dissipative
Power
Power
Paper
..
Printouts
Replacement
Handling
(SSDs)
for
SSD
Symbol
Wrist
Strap
and
Store
Work
Area
Work
Area
Procedures
Pack
Removal
Panel
Bezel
Chart
Recorder
Separation
Conversion
Conversion Conversion
PCB
Removal
PCB/CRT
Supply
Removal
Pacing
Routinely
Techniques
Inspection . ...................................
Inspection
and
Materials
Procedures
Cleaning
Cleaning
Repair
Precautigns
for
.................................
.................................,
Guidelines
Check
..
......:..,....,...,.,.,.,...
Batteries
...................................
or
DC
Power
......................,...,,,.,,.,
.......................,..,.....,....
.................................
....................................
Procedures
......................................
.................................
Mat
.....................................
PCBs
Properly
Static-Free
Source
for
Static
.............................,
............................
..................
..............................
.....................,...........
and
Keypad
Removal
......................................
PCB
PCB PCB
..............,....................
Assembly
PCB
....................................
for
.................................
.............................,.......
.....................................
Storage
Assembly
...........................
Removal
Removal Reassembiy
................................
Removal
and
Removal
................................
Cleaning
..............................
or
Shipping
a
.....................
..................
Sensitive
.......................
Removal
Reassembly
......................
....................
....................
.......................
and
Repair
...................,
sessnese
eseseses
MARE
..........
...........
.............
4-1 4-1 4-1 4-2
4-2
4-3
4-4
4-4
4-4
4-4 4-5 4-5
4-5 4-5
4-6 4-6 4-6 4-6 4-6 4-6
4-6
4-7
4-7
4-7 4-8 4-10 4-10 4-10 4-11 4-12 4-12 4-12 4-12
4-13
4-13 4-13 4-14 4-14 4-14
4-15
4-16
4-17
©
October
1993
Physio-Control
Corporation
All
Rights
Reserved
vii
LIFEPAK
9P
defibrillator/monitor/pacemaker
Parts
Lists/
Schematics
Component Reference
Index
5
Introduction Parts
Component Schematic How
Supplies
6
........................................
List
...........................................
to
Order
Layouts
Diagrams
and
or
Exploded
....:..............................
Parts
..................................
Accessories
Views
...................
..............................
еее
еее
вееекекиныя
5-1 5-1
5-2 5-2 5-2
5-5.
Index-1
viii
All
Rights
Reserved
O
October
1993
Physio-Control
Corporation
List
of
Figures
Description
Operation
1
2
Figure
Figure
Figure
Figure Figure Figure
Figure Figure Figure Figure
Figure
Figure Figure
Figure
Figure Figure
Figure
Figure
Figure
1-1 1-2
1-3
1-4 1-5 1-6
1-7 1-8 1-9 1-10
1-11
1-12 1-13
2-1
2-2 2-3
2-4
2-5
2-6
LIFEPAK
LIFEPAK
Functional
Power
Interconnect
Main
Preamp Echo
Control Character Video
Printhead
Display Display
Front
Rear Test
Setup
Info
9P
9P
Conversion
PCB
Block
Timing
Period
Line
Cell
Intensity
Assembly
PCB/CRT Scanning
Panel
Controls
Panel
Controls,
Menu
...............................,
Menu
Menu
..............................,,
defibrillator/monitor/pacemaker
defibrillator/monitor/pacemaker Block
Diagram
PCB
Pacing
PCB
Diagram
............................
..............................
Timing
Locations
.......................
............................
.....................,..
Assembly
................
Block
Diagram
Block
Diagram
...................
....................
Block
Diagram
.
........
.......
....
..........................
and
Indicators
Indicators,
...........
and
Connectors
.
..............................
Testing
3
Figure Figure Figure Figure
Figure Figure
Figure
Figure Figure Figure
Figure Figure
Figure
Figure
Figure Figure Figure
3-1 3-2 3-3 3-4
3-5 3-6
3-7
3-8 3-9 3-10
3-11 3-12
3-13
3-14
3-15 3-16 3-17
Power-On
Test
Menu
Setup
Menu
Display
1mV
Example
Test
Test
Test
Test
Output
Display Power
Recorder Test
with
CAL
of
Connections
Measurement
Setup
Connector
Menu
Waveform
PCB
Supply
Calibration
Signal
Pacemaker
Pacing
Pulse
Display
................................
Pulse
CODE
for
................................
.........................
..............................
Flatline
Pacemaker
ECG
Trace
..............
...........................
SUMMARY
for
Ground
...........................
J3
..........................
Resistance
Leads
Printout
Detect
Current
.......
..........................
Assembly
Calibration
Setup
Calibration
Waveform
Potentiometer
...................
Adjustments
for
Function
Test
.....................
Location
Generator
Load
Setup
Locations
.....
.......
........
.
.
©
October
1993
Physio-Control
Corporation
All
Rights
Reserved
ix
LIFEPAK
9P
defibrillator/monitor/pacemaker
Service
and
Maintenance
Paris
Lists/
Schematics
4
5
Figure Figure
Figure
Figure
Figure
Figure Figure
Figure Figure
Figure
Figure
Figure
Figure
Figure
Figure Figure
Figure
Figure
Figure Figure Figure
Figure Figure
Figure Figure Figure
Figure
Figure
Figure
Figure
Figure
Figure Figure Figure
Figure Figure
4-1 4-2
4-3
4-4
5-1
5-2 5-3
5-4 5-5
5-6
5-7
5-8
5-9
5-10
5-11 5-12
5-13
5-14
5-15 5-16 5-17
5-18 5-19
5-20 5-21 5-22
5-23
5-24
5-25
5-26
5-27
5-28 5-29 5-30
5-31 5-32
Front
Panel
Removal Connector Connections
Power
LIFEPAK
Shipping
LIFEPAK
Final
Interconnect
Main
Power
Power Interconnect Keypad Rear Strip
Transfer
Paddle
Display Power
Wire
Power
Locations
for
Conversion
9P
defibrillator/monitor/pacemaker
Assembly
9P
defibrillator/monitor/pacemaker
Assembly
Diagram
PCB
Assembly
Supply
Conversion
Pacing
Assembly
Panel
Membrane
Chart
Recorder
Relay
Assembly
PCB/CRT
Supply
Harness
Conversion
PCB/Battery
Battery High
Main
Main Interconnect
Main AC
AC
Test
Power Power
Power
Power
Main
Main
Storage
Storage
ECG
Sensor/Power
Voltage
Connector
PCB/Display
PCB/Recorder
Pacing
Assembly
PCB/Patient
Receptacle/Ground
..............................
Receptacle
Load
Contact/Resistor Conversion Conversion
Harness
Wire
...............................
Conversion
Harness
Conversion
Wire
Harness
PCB/Sternum
PCB/Apex
Capacitor Capacitor
Out
Wire
.......................
..........,............
Discharging
PCB
Capacitors
Assembly
on
..........
......................
.........................
...................,.,..
.......................
PCB
Assembly
PCB
PCB
.........................
Switch
Assembly
Assembly
..........................
Assembly
PCB/AC
Wire
Receptacle/Ground
..................,.........
PCB/Power
Harness
Supply
Cable
Cable
PCB/Keypad
Connector
Wire
Wire
Harness
PCB/Reed PCB/Dump
PCB/Charge
...........................
PCB/Charge
...........................
Relay
Relay
Wire
Positive/Relay Negative/Relay
Harness
................
Assembly
Assembly
............
...........
................
...............
....................
.................
Supply
................
PCB
Wire
Harness . 5-75
Cable
Assembly
Assembly
Assembly
...........
..........
Cable
Wire
Harness
Harness
..........
.................
Wire
Harness
Assy
Wire
Harness
Relay
Wire
Relay
Negative
Relay
Positive
Wire
Harness
Harness
Wire
........
..........
Harness
Wire
Harness . 5-86
.....................
the
.......
.....
.......
..
4-8
4-9
4-11
4-17
5-11 5-12
5-22
5-31
5-40
5-50 5-55 5-56 5-58
5-60
5-61
5-64
5-73
5-74
5-76
5-77 5-78
5-79 5-80 5-81
5-81
5-82 5-82
5-83
5-84
5-84
5-85
5-85
5-86
5-87
All
Rights
Reserved
©
October
1993
Physio-Control
Corporation
Figure Figure
Figure
Figure Figure
Figure
Figure Figure Figure
5-33 5-34
5-35
5-36 5-37
5-38
5-39 5-40 = Pushbutton 5-41
Main
PCB/Speaker
Power
Connector
Supply
Receptacle/Pacing
Recorder/Bracket Defibrillation
Test
Load
High
Test
Load
High
Switch
High
Voltage
Wire
Harness
Bracket/Ground
Ground
Adapter
Voltage
Voltage
Connector
Wire
......................
Wire
Harness
Wire
Wire
Harness Cable
.............
Wire
Harness
Cable
Assembly
Harness
Harness
........
(W1)
(W2)
..............
Assembly
.......
....
...
.....
.....
5-88 5-89
5-90
5-91 5-93
5-95
5-95 5-96 5-97
Component
Reference Diagrams
6
Figure
Figure Figure
Figure Figure Figure Figure
Figure Figure Figure
Figure
Figure
Figure Figure
6-1
6-2 6-3
6-4 6-5 6-6 6-7
6-8 6-9 6-10
6-11
6-12
6-13
6-14 Figure6-15 Figure
Figure Figure
Figure Figure Figure Figure Figure
6-16
6-17 6-18
6-19 6-20 6-21 6-22 6-23
Buffer
A/D D/A
Dual Counter Counter Font
DEFip-FOD D D
D
J-KFlip-Flop
Microprocessor Multiplexer Mulüvibrator
Multivibrator
Oscillator
Pulse-Width
Pulse-Width
Display Video Shift Analog
...................................
Converter Converter
D/A
...........................,......
.......................,..,,..,...
EPROM
Flip—Flop Flip-Flop
Flip-Flop
RAM
RAM
Register
Switch
............................
............................
Converter
.............................
..............................
..............................
.........................,....
....................,.......
...........................
...............................
.............................
.............................
................................
Modulator
Modulator
.............................
..............................
....................,...,....
............................
........................
eee
cee
ees
.....................
.....................
6-3
6-3 6-4
64 6-5 6-5 6-6
6-6 6-7 6-7
6-8
6-8
6-9 6-10
6-10 6-11
6-11 6-12
6-12 6-13 6-13 6-14 6-14
©
October
1993
Physio—Control
Corporation
All
Rights
Reserved
LIFEPAK
9P
defibrillator/monitor/pacemaker
List
of
Tables
Description
Operation
Testing
1
2
3
Table
Table Table
Table
Table
Table
Table
Table Table
Table
Table
Table
Table Table
Table
Table
1-1
1-2 1-3 1-4
2-1
2-2
3-1
3-2 3-3
3-4
3-5
3-6
3-7 3-8
3-9
3-9
LIFEPAK
9P
defibrillator/monitor/pacemaker
Specifications . ..........................
Main
PCB
Functional
Preamp
Display/Recorder/Pacing
Front
Rear
PIP
Delivered Maximum
Maximum
Test
Power
Low Defibrillator
Current
Error
Pulse
YO
Lines
Panel
Panel
Test
Equipment
and
Calibration
Conversion
Battery
and
Codes
WidthS
..............................
Controls
Controls,
Energy
Leakage
Leakage
Thresholds Calibration
Waveform
............................,,
Circuits
.......................
and
Indicators
................
Microprocessor
Indicators
...........
and
Connectors
........................
Tolerances
Current
Current
Test
PCB
.................
for
120Vac
for
230Vac
Equipment
Voltage
.....................
Measurement
Levels
Checks
................
and
60Hz
and
50Hz
...........
........
Ranges
..
...
1-3
1-23 1-27
1-32
2-3
2-6
3-2
3-14 3-20
3-20
3-24
3-26
3-27 3-31
3-44
3-46
xii
Service and
Maintenance
Parts
Lists/
Schematics
Component
Reference
4
5
6
Table
4-1
Table
4-2
Table
5-1
Table
5-2
Table
5-3
Table
5-4
Table
6-1
Inspection
Tools
LIFEPAK
Major Defibrillation Reference
Supplies
Component
Ali
Rights
Techniques
and
Materials
9P
defibrillator/monitor/pacemaker
Assemblies
Adapter
Designator
and
Accessories
Reference
Reserved
......................
for
Cleaning
.......................
Assemblies
Key
....................
Diagrams
and
Repair
............
...................
....
..............
© October
1993
Physio-Control
4-13
4-14
5-3 5-4 5-4
5-5
6-1
Corporation
SET
Introduction
Terms
This
safety
information
on
the
equipment recommended medical
Refer Facilities, guidelines environments.
Certain yourself
equipment.
to
NFPA
and
on
terms
with
Danger:
Warning:
to
alert
precautions
(National
NFPA
70-1990, National
the
standards
are
used
in
their
definitions
Immediate
or
death.
Hazards
personal
or
injury
includes
both
Fire
this
hazards
unsafe
terms
operating
in
care,
Protection
and
practices
manual,
and
which
practices
or
death.
and
symbols
and
service
use
and
handling
Association)
Electrical
for
health-care
or
on
the
equipment.
significance.
will
result
which
could
used
in
personnel
of
this
specialized
99-1990,
Code,
for
instruments
in
severe
result
this
manual
of
Health
specific
Familiarize
personal
in
severe
Care
and
injury
or
©
October
1993
Physio—Control
Caution:
Note:
Corporation
Hazards personal
Points
instrument
concerning
or
unsafe
injury
of
particular
operation.
the
All
Rights
practices
or
product
interest
Additional
subject
Reserved
under
which
could
damage.
for
more
information
discussion.
result
efficient
or
in
minor
convenient
or
explanation
xiii
LIFEPAK
9P
defibrillator/monitor/pacemaker
Warnings
Shock
Possible Paddle
Possible Explosion
Hazard
Arcing
Damage
Fire
or
and
Following result
cautions
When
pacemaker
paddles. electrical perform of
the
When standard (STERNUM arcing paddle
Use bag-valve-mask
anesthetics.
are
in
death,
not
appearing
charged
discharges
Unless
energy
this
procedure
LIFEPAK
discharging
paddles
paddle
and
formation
electrode
care
when
descriptions
severe
and
discharged,
discharged
may
9P
defibrillator/monitor/pacemaker
the
are
on
surfaces
operating
devices
of
general
injury
or
product
in
this
section
the
up
to
360J
of
properly
cause
personal
unless
defibrillator
securely
the
of
you
and
left,
APEX
pits
on
paddle
can
cause
this
device
or
ventilator
hazards
damage.
are
found
LIFEPAK
electrical
as
described
injury
are
thoroughly
into
the
internal
properly
patient
stored
paddle
electrode
skin
close
to
tubing)
and
unsafe Specific
throughout
9P
defibrillator/monitor/
energy
in
or
death.
familiar
in
on
the
surfaces.
burns
oxygen
and
flammable
practices
through
this
manual,
Do
not
and
all
test
load,
the
paddle
right).
during
sources
warnings
the
manual.
the
attempt
with
accessories.
make
This
Pitted
defibrillation.
(such
gases
that
could
and
defibrillator
this
to
the
operation
sure
the
storage
helps
prevent
or
damaged
as
and
area
Safety
Electrical
Possible
Hazard
Hazard
Equipment
Failure
Shock
Possible
or
Fire
Equipment
Damage
and
Hazard
Do
not
mount location other
Do frayed and
inspections
which eventually
Do spilling
Do damage. instructions
where
mounting
not
operate
wires,
possible
on
the
wires
causes
not
immerse
any
not
sterilize
Do
fluids
clearly
the
product
it
cannot
location.
equipment
or
loose
snap
equipment
cables
and
enter
the
the
wire
any
portion
on
the
this
product.
not
autoclave
approve
directly
harm
the
using
damaged
fittings
failure.
wires.
terminals.
instrument
Perform
Pay
strands
of
the
Sterilization
or
gas
it.
above
the
patient
to
if
it
cables
may
cause
frequent particular Repeated
break.
instrument
or
accessories.
environments
sterilize
accessories
patient.
should
interference
in
Place
fall
and
wires.
electrical
attention
flexing
water
at
or
unless
the
product
from
its
Broken
or
loss
and
visual
to
the
point
these
points
other
fluids.
can
cause
manufacturer
shelf
or
or
of
signal
at
serious
in
a
Avoid
xiv
All
Rights
Reserved
©
October
1993
Physio-Control
Corporation
Safety
Shock
Risk
or
Fire
Hazard
Do
not Table work
Make NFPA
Code.
substitute
5-4
shown
improperly.
sure
(National
accessories. on
that
all
Fire
page
5-5.
equipment
Protection
Use
only
recommended
Substitution
. .
is
interconnected
Association)
may
cause
safely
70-1990,
accessories
the
instrument
in
accordance
National
Electrical
listed
in
t
:
with
Symbols
Note:
accessory
applications grounding
accessory
Any
or
all
equipment:
>
ATTENTION
+ Db
L
>|[@
TT
„L
Within
certain
equipment
laboratory.
requirements
equipment.
of
the
following
Static
Sensitive
Additional
DANGER
ATTENTION
Defibrillation
Defibrillation
governmental
must
It
is
important
in
your
location.
symbols
Device
information
high
voltage
refer
protected,
protected,
jurisdictions,
be
labeled
that
you
Check
after
interconnecting
may
appear
(SSD)
in
Section
present
to
manual
type
CF
type
BF
by
an
verify
and
leakage
in
this
6,
Component
for
more
patient
patient
all
interconnected
approved
observe
current
this
manual
connection
connection
testing
and
instrument
or
References
information
the
required
|
on
with
the
©
October
1993
Physio—Control
Corporation
Protective
Fusible
Equipotentiality
Off
On
(Power:
ECG
(Power:
OO Qe
ground
Link
disconnection
connection
Output
All
Rights
(earth)
connector
Reserved
terminal
to
from
the
AC
the
mains)
mains)
LIFEPAK
9P
defibrillator/monitor/pacemaker
General
Service Information
Information
Before technician Service
If
technician
in operating required, a qualified Procedure adjustments
appropriate
attempting
and
the
instrument
Section
to
should
should
3,
within
be
Maintenance.
has
complete
beginning
specifications.
(TCP)
in
must
be
training
clean
or
repair
familiar
been
technician
Section
performed
and
with
dropped,
the
on
page
3.
experience.
the
Performance
3-1,
If
can
Component
only
any
assembly
information
damaged,
to
calibration
perform
by
or
Inspection
confirm
or
the
replacement
service
in
this
instrument,
provided
abused, a qualified
whether
more
Test
and
personnel
in
Section
Procedure
the
instrument
extensive
Calibration and
internal
qualified
testing
the
4,
(PIP)
is is
by
Use
of
non-Physio-Control
adapter devices
For
assistance
1-800-442-1142.
representative.
may
in
void
servicing
In
other
All
Rights
Reserved
defibrillation
Safety
Agency
the
instrument
countries
contact
electrodes,
Certifications
in
the
the
batteries,
US,
call
local
Physio-Control
©
October
and
1993
accessories,
warranty.
Physio-Control
Physio-Control
or
at
Corporation
=
Effective
Publication
Dates
The
effective
Title
Trademark
Table
of
Contents
List
of
List
of
publication
and
Warranty
Figures
Tables
date
for
each
page
Page
ii
ii
ix
xii
of
this
thru
thru
manual
|
viii
xi
is
listed
October
October
October
October
below.
Date
1993
1993
1993
1993
Safety
Information
Terms
Warnings
Symbols
General
1 2 3
4 5
6
Information
Service Effective Configuration
Content
Overview
Description Operation Testing
Service Part
and
Lists/Schematics
Component
Publication
Information
Maintenance
Reference
Dates
xii
xiii
xiv
XV xvi
xvi xvii xviii
xix
1-1 2-1 3-1
4-1 5-1
6-1
thru
thru
thru thru thru
thru thru
thru
xv
xix
1-42 2-10 347
4-17 5-97
6-14
October October
October October October
October October October
October
October October October
October October
October
1993
1993
1993 1993 1993
1993 1993 1993
1993
1993 1993 1993
1993 1993
1993
©
October
1993
Physio-Control
Corporation
Index
Index-1
|
All
Rights
Reserved
thru
Index-7
October
1993
LIFEPAK
9P
defibrillator/monitor/pacemaker
Configuration
Information
Assembly
LIFEPAK Main Power Power Interconnect Keypad Rear
Strip
Transfer
Paddle
Display
Power Power Battery High
Main
Main Interconnect
Main
AC AC Test
Power Power
Power
Power Main Main
Storage
Storage
ECG
Main
Power
Connector
Recorder/Bracket
9P
PCB
Assembly Supply Conversion
Assembly
Panel
Switch
Chart
Recorder
Relay
Assembly
PCB/CRT
Supply
Conversion
Sensor/Power
Voltage
PCB/Monitor
PCB/Recorder
PCB/Patient
Receptacle/Ground Receptacle
Load
Contact/Resistor
Conversion Conversion
Conversion
Conversion PCB/Sternum PCB/Apex
Capacitor Capacitor
Out
Wire
PCB/Speaker
Supply
Receptacle/Pacing
This
manual
numbers.
is
current
The
assemblies
Name
defibrillator/monitor/pacemaker
PCB
Assembly
PCB
Assembly
Pacing
PCB
Assembly
Membrane
Assembly
Assembly
Assembly
PCB/AC
Connector
Pacing
Wire
Harness
Bracket/Ground
Receptacle/Ground
PCB/Power
Supply
PCB
Cable
SupplyPCB/Battery
PCB
Cable
Assembly
Cable
Assembly
Assembly
PCB/Keypad
Connector
Harness
Wire
Wire
Harness
(W
Wire
PCB/Reed PCB/Dump
PCB/Charge PCB/Charge
Relay
Relay
Wire
Wire
Assembly
Relay
Relay Relay Harness
Harness Positive/Relay Negative/Relay
(W20)
Wire
Hamess
Wire
Cable
Gound
Wire
Harness
Wire
Harness
(W6)
Cable
Assembly
Harness
(W9)
10) Harness
Wire
Wire
Negative Positive
(W17)
Wire
Harness
Wire
Harness
(W21)
Harness
Assembly
(W24)
Wire
(W4)
(W5)
(W8)
(W11)
Harness
Harness
(W16)
(W22)
Harness
(W3)
Wire
Wire
(W18)
(W19)
(W23)
with
appear
Wire
Harness
(W7)
(W12)
(W13)
Harness
Harness
the
listed
(W1)
(W15)
in
(W2)
(W14)
revision
the
level
same
of
the
order
as
in
PartNumber
805460 805474-01 803726-01 803724-03 805472-02 805469 805464-00
804189-00 800240-14 802901-06 803706 803783-30 803783-23 803783-24 803756-13
803733-00
803773-00 805465-00
803783-00
803783-04 803783-05
803783-06
803783-07 803783-08
803783-09 803783-10 803783-11 803783-12 803783-13 803783-14 803783-15 803783-16 803783-29
802925-03
803783-19
following
Table
5-1.
Rev
F4 B8 Orig7 D6 B4 B3 Orig3 B2 U10 F
AS BS BS B5 D4
Orig]
Orig1 Orig3
B5
B5 B5 B5
B5 B5
B5 B5
B5
B5
B5 B5 BS B5 B5 C6
B5
part
xviii
All
Rights
Reserved
©
October
1993
Physio-Control
Corporation
Assembiy
Defibrillation Test
Load
Test
Load Pushbutton High
Voltage
Content
Name
Adapter
High
Voltage
High
Volage
Switch
Wire
Connector
Overview
Wire
Harness
Wire
Harness
Harness
Cable
This
Assembly
manual
contains
the
following
information:
PartNumber 803747
803774-03 803774-04 803774-05 803756-09
-
Rev B2
Al Al Al D4
Section
Section
Section
Section
1
2
3
4
Description: power
supplies,
PCB
circuit
major
circuits.
appropriate
Operation:
It
identifies
“This
section instrument; Instructions
This
section
PCB
functions,
description
Circuit
schematics
This
section Controls, is
not
such
Indicators,
intended
detailed
manual.
Testing/Troubleshooting: Procedure closed-case be
duplicated
(TCP)
testing;
additional
Service Procedures referenced This repairing
(PIP)—seguential
check
and
in
this
section
it
also
includes a checklist.
information
and
Maintenance:
for
removing
to
numbered
section
includes
the
instrument.
of
the
used
describes
instructions
details
and
is
accompanied
names
in
Section
in
these
familiarizes
to
instruct
instructions
This
section
steps
equipment.
during
testing.
calibration
to
support
This
all
major
parts
in
the
how
the
instrument
instrument
by a block
block
outputs
diagram
diagrams
5.
the
user
with
and
Connectors,
the
operator
are
provided
contains
to
follow A PIP
The
Test
and
The
Troubleshooting
and
in
in
the
when
performing
checklist
and
more
troubleshooting.
section
for
provides
subassemblies.
Final
Assembly
inspecting,
cleaning,
Disassembly/Assembly
The
works.
also
basic
screen
the
the
Input
are
described.
to
illustrate
appear
equipment
menu
clinical
separate
Performance
an
is
provided Calibration extensive
instrument
Aids
procedures
drawing
in
maintaining,
signals,
Each
the
in
the
function.
options.
use
of
the
Operating
Inspection
operational
which
can
Procedure
includes
are
Section
5. and
O
October
Section
Section
1993
Physio-Control
5
6
Parts Accessories, Diagrams
Component consists (ICs).
a
Corporation
Lists/Schematics:
Illustrated
for
all
Reference
of
Component
The
ICs
in
large
asterisk
K.
This
Parts
repairable
assemblies.
Diagrams:
Reference
Section 6 are
All
Rights
section
Lists,
PCB
An
Diagrams
identified
Reserved
contains a list
Component
aid
to
for
in
the
of Layouts,
troubleshooting,
selected
schematics
Supplies
and
and
this
Integrated
in
Section 5 with
Schematic
section
Circuits
xix
Introduction
This theory The
e e
section
section
Physical
describes
of
operation
is
Description
divided
Functional assemblies
the
general
of
the
LIFEPAK
into
three
describes
Description
features,
parts:
briefly
describes
specifications,
9P
defibrillator/monitor/pacemaker.
general
features
the
function
and
lists
functions,
specifications
of
the
major
and
Physical
Description
e © Theory
component
The
LIFEPAK provides (CRT)
monitor monitoring, pacing, detected paddles,
The sealed
and
through or
instrument
lead-acid temporary instrument battery
The
continually
front
discharge
rear-panel
and
the
real-time
of
Operation
level.
9P
defibrillator/monitor/pacemaker,
dc
defibrillation,
screen, a strip
heart
rate
display
CODE
SUMMARY
patient
FAST-PATCH
operates
battery,
operation
plugged
during
into
recharges,
panel
pushbuttons
pushbuttons
pushbuttons
clock.
provides
synchronous
chart
and
cable
electrodes,
disposable
on
either
accessible
ac
power
ac
power
even
provide
are
located
control
the
circuit
descriptions
cardioversion, a Cathode
recorder,
alarms,
critical
3-lead
multiple
event
record.
QUIK-LOOK
defibrillation
ac
power
or
internal
from
the
bottom
failure
and
with
control
on
the
QRS
or
disconnection.
the
rear
panel
the
front
panel
for
most
defibrillator
audio
volume, a calibration
to
the
major
shown
in
Figure
1-1,
Ray
electrocardiogram
energy
selection,
The
ECG
external
signal
defibrillation
electrodes.
battery
of
the
power
functions. paddles.
power.
instrument,
With
switch
power
off.
Four
the
on,
Charge
pulse,
Tube
(ECG)
is
The
allows
the
and
©
October
1993
Physio-Control
Corporation
All
Rights
Reserved
1-1
LIFEPAK
9P
defibrillator/monitor/pacemaker
El
bisi
ğ
Е
Е
Figure
1-1
LIFEPAK
The
CRT
monitor
of
25
+1mm/second.
selected and menus configuration
The the the screen. clip-on
defibrillation/ECG Shock
paddle defibrillation
lead,
pacing
allow
defibrillator
patient
selected
Defibrillation
posterior
Advisory
wells,
9P
screen
ECG
information,
access
information.
delivers a controlled
with
selectable
energy
or
electrodes
Adapter
an
internal
energy
defibrillator/monitor/pacemaker
provides a no-fade
The
monitor
gain,
heart-rate
sync
markers,
to
power-on
energy
and
the
available
pulses
pediatric
delivery.
and
test
load
may paddles,
(with
cable).
screen
violations,
default
levels
energy
be
delivered
the
is
accessible
ECG
displays
status
and
operation
parameters,
monophasic
ranging
are
through
or
optional
optional
When
the
paddles
for
display
heart
messages,
status.
diagnostic
dc
from 0 to
displayed
FAST-PATCH
Defibrillation
testing
with a sweep
rate,
QRS
indicator,
defibrillator
Additional
tests,
defibrillating
360 on
standard
are
stored
and
pulse
joules.
the
monitor paddles, disposable
Adapter
in
confirming
rate
and
to
Both
or
the
All
Rights
Reserved
©
October
1993
Physio-Control
Corporation
The
recorder
heart
rate,
automatically
occurs.
prints
the
event
summary
can
pacing
When
stored
print
parameters,
prints
for
the
CODE
CODE
and
recorded
the
ECG
trace,
and
SYNC
approximately
SUMMARY
SUMMARY
ECG
trace
time
and
date,
(if
activated).
16
seconds
pushbutton
critical
event
data.
ECG
lead
and
The
recorder
when a heart
is
pressed,
record
the
data,
including
size,
rate
violation
recorder
Description
1
ECG
ECG INPUT
PATIENT COMMON
MONITOR SWEEP
FREQUENCY
ECG
HEART
HEART
1
mV
ECG
Table
MONITOR
LEAD
CABLE
SPEED
ECG
Leads:
Paddles:
SIZE
RATE
RATE
Cal
OUTPUT
1-1
LIFEPAK
SELECTIONS
LENGTH
MODE
DISPLAY
REJECTION
SIZE
RESPONSE
METER
ALARM
Pacing
Demand at a selected
by
and
The connector
monitor/pacemaker
is
any
cardiac
rate.
ac
line
selectable
mode
rate
power
are
in
depends
regardless
activity
fuses, a ground
accessible
specifications
9P
defibrillator/monitor/pacemaker
Std,
Paddles,
Isolated
disposable Total
length
_
With
notch
ground
measured
peak
with
102mm
25mm/s
Non-diagnostic.
1.0
to
2.2
to
Adjusts
ECG
out.
Three-digit outside Heart
rate
User-selectable
160/90.
rate
alarms
Momentary
the
ECG
1V/mV
either
demand
on
are
at
cardiac
of
shown
the
activity
cardiac
on
the
connector,
rear
panel.
are
I,
II,
ITI.
ECG
via
OUIK-LOOK
defibrillation/ECG
4.0m
(13ft):
filter
engaged,
and
65dB
minimum
at
60Hz.
Common
respect
(4in)
40Hz 20Hz
amplitude
this
meter
Other
to
isolated
wide x 76mm
(-3dB) (-3dB)
of
ECG
readout
range
are
pushbutton
displays
do
not
is
disabled
alarm
limit
options
disabled
limits.
on
input.
at
x1.0
gain
or
non-demand
sensing;
activity.
display
and
The
listed
in
defibrillator
electrodes,
cable
3.1m
100dB
with
mode
ground.
(3in)
trace
on
rates
yield
valid
during
Three
available
during
rear
pacing.
panel
(asynchronous)
non-demand
Stimulus
along
an
LIFEPAK
Table
markers
with
ECG
1-1.
Specifications*
or
(10ft),
minimum
respect
tall,
to
isolated
range
for
non-fade.
.
monitor,
from
20
to
systole
pacing.
high/low
tones
through
simulates a 1mV
mode
followed
pacing
signal
9P
paddles,
3-lead
leads
with
patient
strip
300bpm.
settings:
the
current
output
defibrillator/
FAST-PATCH
patient
0.9m
respect
ground
cable
chart
recorder,
Heart
or
heart
Setup
signal
mode.
paces
cable.
(3ft).
to
chassis when
input
210V
and
rates
rate
display.
150/40, 120/60,
Menu.
Heart
pulse
to
©
October
1993
Physio-Control
Corporation
All
Rights
Reserved
1-3
LIFEPAK
9P
defibrillator/monitor/pacemaker
Table
DEFIBRILLATOR
ENERGY
External
Internal
CHARGE
PADDLE CHARGE
CHARGE
SYNC SYNC
DEFIBRILLATOR ACCURACY
DEFIBRILLATOR
I(amps)
1-1
CORD
INDICATOR
60
LIFEPAK
SELECT
Paddles:
Paddles:
CONTROLS
LENGTH
INDICATORS
TIME
ENERGY
WAVEFORM
9P
defibrillator/monitor/pacemaker
1,
2,
3,
4,
1,
Independent
paddle.
3m
Flashing
increasing progress. completed
Charge
Synchronizes
Inverted
synchronizer
+15% +30%
Sms
50
Ohm
Load
5, 6, 7,
2,
3,
4,
5, 6, 7,
(10ft)
lamps
Upon
to
360J
triangle
at
500
over
monophasic
360
Joule
8,
9,
8,
9,
momentary
on
paddle
stored
energy
full
charge,
tone
sounds.
in
less
defibrillator
marker
trigger
the
point
range
pulse
Ifamps)
10,
20,
10,
20,
pushbutton
and
display
energy
than
10s
pulse
on
displayed
with
of
25
to
(Edmark)
80
Specifications*
30,
50,
100,
200, 300,
30,
50J
controls
front
panel
on
monitor
available
with a fully-charged
to
patient-generated
ECG
respect
1000
on
front
pushbutton
indicate
is
displayed
waveform
to
patient’s
25
QRS
Ohm
A
(cont.)
3601
panel
and
along
with
charge
battery.
QRS
in
and
complex.
identifies
complex.
Load
360
Joule
APEX
charge
aT
“TAN
20
10
K
)
‘amps
PIA
0 2
50
Ohm
Load DC
0 2
100
Ohm
Load
N
DS
~
4
6 8
Defibrillator
4
6
DC
Defibrillator
|
10
Output
8
10
Output
12
Waveform
100
12
Waveform
Ohm
14
Load
360
14
16
T
(msec)
Joule
18
T
(msec)
o
"IT
Ш
30
20
10
0
-10
0 2
25
Ohm
Load
\
\
4
6
DC
Defibrillator
8
Output
10 12
Waveform
14
T
(msec)
1-4
All
Rights
Reserved
© October
1993
Physio—Control
Corporation
Description
1
Table
THERMAL
PAPER
Size: Speed:
DELAY
FREQUENCY
ECG
Leads
ECG
Leads
Paddles:
ANNOTATION
CODE Record
NON-INVASIVE
OUTPUT RATE OUTPUT
OUTPUT
REFRACTORY
MODE
SUMMARY
ACCURACY
Pacing
40
110 130
150
1-1
ARRAY
LIFEPAK
RECORDER
RESPONSE (non-diagnostic):
(diagnostic):
Critical
PACEMAKER
RATE
WAVEFORM CURRENT
PERIOD
Rates:
to
100ppm
to
120ppm
to
140ppm
to
170ppm
9P
defibrillator/monitor/pacemaker
Event
50mm
25mm/s
ECG Non-diagnostic
Time, test
Digitally
40
Monophasic, 0
Refractory
Demand
wide x 30m
prints 8 seconds
1.0
to
40Hz
0.05
to
100Hz
2.2
to
20Hz
date,
load
discharges,
stored
to
170ppm.
+5%
over
to
200mA + 5%
3mA
(whichever
300ms
250ms 220ms 200ms
or
(100ft)
(diagnostic
(-3dB).
(-3dB).
(-3dB).
ECG
lead,
pacemaker
record
entire
range.
truncated
or
2mA
is
greater)
Periods:
non-demand.
after
first
available
ECG
of
ECG
exponential
(whichever
appearing
gain,
heart
parameters.
and
device
current
over
the
Specifications*
on
the
monitor.
via
Setup
Menu).
rate,
defibrillation
parameters.
pulse
(15
to
25%
is
greater)
range
of
100
at
to
700%
1500
(cont.)
parameters,
droop).
and
+7%
or
O
October
GENERAL
AC
INPUT
BATTERY BATTERY
1993
Physio—Control
OPTIONS
TYPE CAPACITY
Corporation
120
or
240Vac
with
adjustment
Sealed
A
lead-acid,
new,
fully-charged
shutdown:
Number
Minutes
Minutes
All
Rights
nominal
in
Setup
16Vdc
of
360J
discharges:
of
monitoring:
of
pacing:
Reserved
line
voltage
Menu.
nominal, 3 Ahr.
battery
will
without
provide
Typical
adjustment.
one
75
90 70
75
of the
Minimum
40
50
50
or
following
60Hz
prior
to
LIFEPAK
9P
defibrillator/monitor/pacemaker
Table
GENERAL
LOW
BATTERY
BATTERY
SERVICE
POWER
SIZE
Height:
Width:
Depth:
Weight:
POWER
STANDARD
ELECTRODE
1-1
(cont.)
BATTERY
CHARGE
CHARGE
INDICATOR
CONSUMPTION
CORD
PADDLE
LIFEPAK
INDICATOR
LENGTH
AREA
9P
INDICATOR
TIME
defibrillator/monitor/pacemaker
Advises after fully
Indicator
24hrs
Message
improper
160W
charging.
35.2cm
29.7cm
31.0cm
13.2kg
3m
operator
warning
charged
Number
Number
BATT
to
full
SERVICE
operation
maximum
(13.9in)
(11.Tin)
(12.2in)
(29lbs)
(10ft)
to
connect
(LOW
battery
of
of
BATTERY-CONNECT
360J
discharges
360J
discharges
CHRG
capacity.
indicates requiring
while
ac
are:
before
after
illuminates
self-diagnostic
service
monitoring
Specifications*
power.
warning:
warning:
when
attention.
with
82cm?
Battery
AC
battery
routines
recorder
(cont.)
capacities
POWER)
Typical
is
on
before
for a new,
50 20
25
charging.
have
and
defibrillator
and
Minimum
5
detected
ENVIRONMENTAL
ATMOSPHERIC
RELATIVE
TEMPERATURE
HUMIDITY
PRESSURE
RANGE
797
0
to
0
to
0
to
-30
to
500mm
95%
80%
45°C
to
65°C
Hg
(non-condensing) (non-condensing)
(2
to
(-22
(-570
113°F) to
149°F)
to
11,000ft).
at 0 to at
operating.
storage.
35
34°C
to
45°C
(32
(95
to
94°F).
to
113°F).
1-6
All
Rights
Reserved
©
October
1993
Physio-Control
Corporation
Description
1
Functional Descri
ption
The
overall
shown
mechanical Recorder, Board Pacing,
The
Keypad input. Transfer Battery
The
Power 90Vac
Power
supplied
The
Power power defibrillation present
function
in
Figure
assemblies
Transfer
(PCB)
Main,
and
The
Strip
Relay
Pack
Supply
to
270Vac
Supply
to
the
Conversion
on/off
on
the
of
the
1-2.
The
(Keypad,
Relay,
assemblies
and
Display).
Rear
Panel
Chart
Recorder
and
Paddle
provides
PCB
Power
switching,
charging,
Power
temporary
PCB
at
SOHz
converts
Conversion
PCB
and
Conversion
LIFEPAK
defibrillator/monitor/pacemaker
Paddles,
(Power
Membrane
Assemblies
functions
or
60Hz
the
performs a variety
generation
stored
9P
Rear
Panel
and
Battery
Supply,
Assembly
backup
ac
PCB.
Power
Switch
help deliver
over
the
without
line
power
PCB.
of
additional
energy
defibrillator/monitor/pacemaker
includes
Membrane
Pack)
Conversion,
Assemblies
provides
power.
full
range
input
line
to
of
power
transfer. A test
Switch,
and
five
receive
printed
defibrillator
of
voltage
the
nominal
power-handling
output.
input
supply
load
is
six
Strip
Chart
Printed
Interconnect
voltages
selection.
Circuit
operator
The
pulses.
from
The
The
+18.8Vdc
functions:
voltages,
circuit
is
also
The
Interconnect electrical It
also
panel
The Controller, input The QRS are
the
System allowing the
The
5-inch deflection produce a raster
connection
contains
LEDs.
Main
PCB
and
from
both
System
detection,
all
System
appropriate
Display
Controller
implemented
microprocessor
microprocessor
the
Display/Recorder/Pacing
PCB/CRT
diagonal
and
incorporates
circuitry
has the
rate
CRT
Pacing
the
Cathode
scan.
PCB
provides
between
three
Display/Recorder/Pacing
3-lead
calculation in
the
messages
the
Main
to
produce
major
patient
directs
System
and
to
the
Assembly
Ray
vertical
the
sections
cable
the
overall
and
alarms,
Controller
the
ECG
Display/Recorder/Pacing
microprocessor
and
recorder
contains a Monitor
Tube
(CRT).
and
all
the
pacing
PCB
and
the
audible
of
and
horizontal
tones
circuitry:
Interface.
QUIK-LOOK
operation
and
synchronous
software.
data
are
transmitted
annotations.
It
uses
circuitry
Power
and
the
The
of
the
to
PCB
electromagnetic deflection
and
the
Conversion
to
drive
the
Preamp,
other
The
microprocessor,
select
the
Preamp
paddles
that
inputs.
circuits.
cardioversion
current
from
the
and
drives
beam
circuitry
PCB.
front
System
receives
state
of
format
the
to
©
October
1993
Physio-Control
Corporation
All
Rights
Reserved
1-7
LIFEPAK
9P
defibrillator/monitor/pacemaker
POWER
ENERGY
SUPPLY
ASSEMBLY
BATTERY
PACK
STORAGE
CAPACITOR
PADDLE
ASSEMBLY
O:
PACING
ELECTRODES
PATIENT
PCB
RELAY
ASSEMBLY
y
>
CABLE
POWER
INTERCONNECT
MAIN
ASSEMBLY
CONVERSION
ASSEMBLY
CHARGER
CIRCUITRY
PCB
ASSEMBLY , | |
DRIVER
|
GENERATOR Y *
PACING
CIRCUITAY
PCB
PREAMP
PACING!
TONE
OPTICAL
PCB
POWER
MONITORING/
CONDITIONING
VOLTAGE
EVALUATION
CIRCUITRY
Za
eme
, '
νο
i
ro
1
a
st
a
INTERFACE
CHARGE
CONTROLLER
LED
DRIVERS
SYSTEM
o
CONTROLLER
|
©
788
O
KEYPAD
ASSEMBLY
REAR
MEMBRANE
SWITCH
ASSEMBLY
=>
PANEL
1-8
STRIP
CHART
RECORDER
ASSEMBLY
Figure
<
1-2
LIFEPAK
DISPLAY/
PING
PACING
R “INTERFACE
9P
defibrillator/monitor/pacemaker
INTERFACE
>
All
|
Rights
Reserved
RASTER
FRAME
VIDEO
1
VIDEO
2
Functional
October
©
Block
1993
é
© *
DISPLAY
PCB/CRT
ASSEMBLY
REAR
PANEL
Diagram
Physio—Control
Corporation
Description
1
Theory
of
Operation
Power Assembly
Power-On
Supply
(803726)
PCB
This
section
Power
PCB/CRT circuits
selected
in
The
the
When
CR1.,
into
Conversion
in
Section
Power
Power
ac
Resistor
and
R12
and
is
coupled
Q1,
Increasing Q2
and a voltage
overrides
base
drive
contains
detailed
PCB,
Assemblies.
Section 5 while
integrated
circuits
6.
Supply
PCB
Conversion
current
turn
causing
Q1
flows
divider
on
Q1.
with
saturation.
current
drop
and
terminates
current
pins 1 and
through
to
Q2.
circuit
Interconnect
Refer
to
schematic
reading
(indicated
converts
PCB
and
for
through
R1,
Current
across
J1,
R3,
and
builds
2.
the
T1
parallel
the
primary
descriptions
Pacing
the
circuit
by a ste
input
ac
voltage
charging
it
is
conditioned
R4
in
series
in
T1
primary
Capacitor
primary
produces
resistors
current.
of
the
PCB,
Main
diagrams
descriptions.
the
C6
of
on
the
schematic)
to a nominal
battery.
by
RT1,
with
parallel
windings
couples
voltage
R11
and
Transformer
Power
these
Supply
PCB,
and
electrical
Information
dc
L1,
resistors (pins 4 and
the
current
on
the
R12.
O2
T2
PCB,
Display
about
is
provided
voltage
C2, C3,
activation
and
R11
6)
build-up
base
of
sustains
for
Regulation
Overvoltage
Output
Protection
Isolation
Battery
When
battery
and
couples current regulation.
If
together
CR13
short
protection
Diode
protect
necessary
subsides
the
regulation
with
sets a voltage
the
Power
circuit.
CR11
it
from
because
PCB.
The
16Vdc, output. instrument on.
As
The
is
the
battery
voltage
the
voltage
to
proper
circuit
CR13,
R17,
threshold
Supply
isolates
the
battery
the
the
lead-acid
battery
receives a constant
connected
approaches
exceeds
from
T2
levels,
fails,
and
PCB
circuit
Power
voltage
battery
battery
to
ac
is
power
the
voltage
pins 4 and 5 to
Q4
turns
Silicon
R18,
which,
is
Controlled
protects
when
output.
Supply
PCB
when
connected
wired
directly
charging
and
full
charge,
ac
the
threshold
T2
pins 2 and
off,
providing
Rectifier
the
load
exceeded,
Removing
output
power
in
the
is
parallel
to
the current
rear
panel
supply
set
by
VR1,
current
(SCR)
from
damage.
triggers
ac
power
from
the
removed.
with
the
Power
Supply
whenever
mains
operates
Q4
7.
When
feedback
CR12,
SCR
resets
filter
network
This
Power
the
power
in
constant
turns
Diode
CR12
the
is
Supply
PCB
switch
on
for
to
to
is
©
October
1993
Physio—Control
Corporation
All
Rights
Reserved
1-9
LIFEPAK
9P
defibrillator/monitor/pacemaker
BATT
CHRG
Power
PCB
Conversion
Assembly
(803724)
Power
Control
Indicator
voltage the Power charging a severely
A indicator panel the
The additional charging,
The control
Control
Control Charger
mode.
battery
Supply
battery-charging
mains
indicator
Power
Power
instrument
circuit
Input,
Current
at
approximately
PCB
whenever
power
from
lighting
Conversion
voltage
and
Control
generation
defibrillation
are:
Low
circuits.
LED
switch
operation.
limiting
+4.5A. sources depleted
the
circuit
current
battery.
signal
ac
power
is
on.
if
the
PCB
contains
and
energy
regulates
Switching
Energy
Charge
sets
the
maximum
This
into a dead
illuminates
is
connected
An
interlock
battery
The
Power
is
circuitry
monitoring,
transfer.
battery
circuits
Supply,
Rate,
charge
limit
is
implemented
short, a requirement
the
front
panel
to
the
instrument
in
the
battery
not
installed.
for
power
energy
Refer
voltage
directly
Logic
and
Energy
storage
to
Figure
for
the
affected
Power
Storage
current
when
BATT
and
harness
on/off
switching,
capacitor
1-3,
circuits
by
the
Monitor,
Capacitor
available
the
when
CHRG
the
rear
prevents
page
1-11.
which
Power
External
to
The
Front Panel
the
Power
power
which
gate
PWR
SWBATT
circuits,
When
V
BATT
the
PWR low-to-high disabling Control connector: of
Q11).
Control
on), V BATT
in
turn
of
Q11
SW
MONITOR
voltage
the
power
is
SW
the
circuit
ON
momentary
circuit.
passes
causes
by
R52,
Q10
which
for
switch
divided
transition
R60
by
MONITOR
Power
may
(connects
Control
be
pushbutton
When
the
to
the
to
conduct.
keeps
line
is
low
and
the
rest
of
the
is
pressed a second
R108
and
R109,
line
high.
and
activates
circuit.
confirmed
to
the
gate
provides
power
switch
gate
of
Q11,
A portion
Q10
conducting.
the
Power
Power
the
by
Conversion
time
forward-biasing
The
System
Logic
The
functional
monitoring
of
Q10)
the
start-up
is
first
causing
of V BATT
Q11
During
Control
circuit
PCB
(turning
microprocessor
Power
two
and
R57
power
CR44
Monitor
integrity
pins
on
(connects
pressed
to
conduct,
is
held
this
time
provides
Assembly
off),
and
reads
circuit,
of
the
the
test
to
signal
for
(turning
on
the
the
the
switching
the
Power
the
gate
1-10
All
Rights
Reserved
©
October
1993
Physio-Control
Corporation
Loading...
+ 237 hidden pages