2004 Koninklijke Philips
Electronics N.V. All rights
are reserved.
All other product names
are the property of their
respective owners.
Permission is granted to
copy and distribute this
document for educational
purposes.
Warranty
Philips Medical Systems
makes no warranty of any
kind with regard to this
material, including, but not
limited to, the implied
warranties or merchantability and fitness for a
particular purpose.
Philips Medical Systems
shall not be liable for
errors contained herein or
for incidental or consequential damages in
connection with the
furnishing, performance,
or use of this material.
WARNING
As with electronic equipment, Radio Frequency
(RF) interference between
the cardiograph and any
existing RF transmitting or
receiving equipment at the
installation site, including
electrosurgical equipment, should be evaluated
carefully and any limitations noted before the
equipment is placed in
service.
Radio frequency generation from electrosurgical
equipment and close proximity transmitters may
seriously degrade performance.
WARNING
Like all electronic devices,
this cardiograph is susceptible to electrostatic
discharge (ESD). Electrostatic discharge typically
occurs when electrostatic
energy is transferred to the
patient, the electrodes, or
the cardiograph.
ESD may result in ECG
artifact that may appear as
narrow spikes on the
cardiograph display or on
the printed report. When
ESD occurs, the cardiograph ECG interpretation
may be inconsistent with
the physician interpretation.
Philips Medical Systems
assumes no liability for
failures resulting from RF
interference between
Philips Medical Systems
medical electronics and
any radio frequency generating equipment at levels
exceeding those established by applicable standards.
CAUTION
The use of parts or accessories other than those
approved by Philips
Medical Systems may
compromise product
performance.
United States federal law
restricts this device to use
by or on the order of a
physician.
THIS PRODUCT IS NOT
INTENDED FOR HOME
USE.
Medical Dev ice
Directive
The PageWriter Trim
Cardiograph complies
with the requirements of
the Medical Device Directive 93/42/EEC and carries
the
PageWriter Trim Cardiograph Instructions for UseContents-7
Safety Symbols Marked on the Cardiograph
1Safety Summary
Safety Symbols Marked on the Cardiograph
SymbolNameDescription
AttentionSee PageWriter Trim Instructions for Use for
Type CFECG physio isolation is type CF, defibrillator proof.
Alternating currentIndicates that the cardiograph is receiving alternating
On/StandbyPressing the button with this symbol on it turns on the
information.
Electrical leakage current is suitable for all patient
applications including direct cardiac application.
currents.
cardiograph or puts the cardiograph into Standby
(power saving mode).
Do not touch exposed pins. Touching exposed pins can
Electrostatic Discharge
cause electrostatic discharge that can damage the
cardiograph.
Equipotential grounding postEquipotential grounding post used for establishing
common ground between instruments.
FuseCardiograph contains a 1.5 amp (250V) time-delay
fuse.
InputThe connector near this symbol receives an incoming
signal.
Serial NumberThe number next to this symbol is the serial number of
the cardiograph.
Product model numberThe number next to this symbol is the product model
number of the cardiograph
Entry of liquidsThe cardiograph is not protected against splashing
water.
Entry of liquidsThe PIM (Patient Interface Module) is protected
against splashing water. Water splashed against the
PIM from any direction shall have no harmful effect.
PageWriter Trim Cardiograph Instructions for Usei
Safety Symbols Marked on the Cardiograph Packaging
AC power indicator lightWhen lit, indicates that AC power is on. The battery is
charging when inserted into the cardiograph.
Safety Symbols Marked on the Cardiograph Packaging
SymbolDescription
Keep dry.
o
Ambient temperature range of 0
o
to 50
C (122o F) (non-condensing) for
transport and storage.
Atmospheric pressure range of 466 hPa to
1014 hPa for transport and storage.
C (32o.F)
Relative humidity range of 25% to 80%
(non-condensing) for transport and storage.
Move and store packaging this end up.
Fragile.
Sealed lead acid battery. Do not dispose of
in trash. Follow local regulations for
disposing of as small chemical waste.
Recycle the packaging materials after use.
iiPageWriter Trim Cardiograph Instructions for Use
Conventions Used in the Instructions for Use
Conventions Used in the Instructions for Use
WARNINGWarning statements describe conditions or actions that may result in personal injury or
loss of life.
CAUTIONCaution statements describe conditions or actions that may result in damage to equipment or
software.
NOTE Notes contain additional important information about a topic.
TIP A Tip contains suggested information on using a particular feature.
Menu item
Button name
Menu items and button names appear in a bold no-serif font.
Example: Highlight the
Config button.
Important Patient and Safety Information
The PageWriter Trim cardiograph isolates all connections to the patient from electrical ground
and all other conductive circuits in the cardiograph. This reduces the possibility of hazardous
currents passing from the cardiograph through the patient’s heart to ground, and from other
equipment connected to the patient passing through the leads into the cardiograph to ground.
WARNINGFailure to follow these warnings could affect both patient and operator safety.
When operating the cardiograph on AC power, ensure that the cardiograph and all other
electrical equipment connected to or near the patient are effectively grounded.
Use only grounded power cords (three-wire power cords with grounded plugs) and
grounded electrical outlets. Never adapt a grounded plug to fit an ungrounded outlet by
removing the ground prong. Use the equipotential post when redundant earth ground is
necessary according to IEC 60601-1-1.
If a safe ground connection is not ensured, operate the cardiograph on battery power only.
The use of equipment that applies high frequency voltages to the patient (including
electrosurgical equipment and some respiration transducers) is not supported and may
produce undesired results. Disconnect the patient data cable from the cardiograph, or
detach the leads from the patient prior to performing any procedure that uses high
frequency surgical equipment.
Do not perform ST analysis on the R/T ECG screen display or on Rhythm reports when
the 0.5 Hz Baseline Wander filter is applied.
PageWriter Trim Cardiograph Instructions for Useiii
Important Patient and Safety Information
If abnormal ECG data appears on the printed report, and the abnormal data does not have
a physiological origin, perform the printer diagnostic test to assess printer performance
(see page 7-13).
When printing a Rhythm report, there may be a slight delay before the Rhythm report
begins to print on the cardiograph. Rhythm printing is not completed in real-time.
Pace pulse tick marks will not print on an Auto report that uses simultaneous acquisition.
WARNINGDo not touch accessible connector pins and the patient simultaneously.
Electrical shock hazard. Keep cardiograph, Patient Interface Module (PIM) and all
cardiograph accessories away from liquids. Do not immerse cardiograph, PIM, or other
accessories in any liquids.
Periodically inspect the patient data cable, lead wires, and AC power cord for any worn or
cracked insulation to ensure that no inner conductive material is exposed. Discard worn
accessories and replace them only with Philips Medical Systems accessories (see page 7-
20).
Keep the patient data cable away from power cords and any other electrical equipment.
Failure to do so can result in AC power line frequency interference on the ECG trace.
The Philips Medical Systems patient data cable (supplied with cardiograph) is an integral
part of the cardiograph safety features. Use of any other patient data cable may
compromise defibrillation protection, degrade cardiograph performance, and may result in
distorted ECG data.
Only qualified personnel may service the cardiograph or may open the cardiograph
housing to access internal cardiograph components. Do not open any covers on the
cardiograph. There are no internal cardiograph components that are serviced by the
operator.
Do not use this cardiograph near flammable anesthetics. It is not intended for use in
explosive environments or in operating rooms.
Do not touch the patient, the patient data cable, any unused patient leads, or the
cardiograph during defibrillation. Death or injury may occur from the electrical shock
delivered by the defibrillator.
The PageWriter Trim I is not recommended for diagnostic cardiograph use during
defibrillation. It does not provide less than 10 seconds of real-time data.
Always use electrode gel with reusable electrodes during defibrillation as ECG recovery
will be greater than 10 seconds. Philips Medical Systems recommends the use of
disposable electrodes at all times.
Ensure that the electrodes or lead wires do not come in contact with any other conductive
materials (including earth-grounded materials) especially when connecting or
disconnecting electrodes to or from a patient.
ivPageWriter Trim Cardiograph Instructions for Use
Important Patient and Safety Information
Connecting multiple medical electrical equipment to the same patient may pose a safety
hazard due to the summation of leakage currents. Any combination of instruments should
be evaluated by local safety personnel before being put into service.
Portable medical equipment such as X-rays and MRI may produce electromagnetic
interference that produces noise in the ECG signal. Move the cardiograph away from these
potential sources of electromagnetic interference.
Do not pull on the paper while an ECG report is being printed. This can cause distortion of
the waveform and can lead to potential misdiagnosis.
Only use the Philips Medical Systems AC power cord supplied with the cardiograph.
Periodically inspect the AC power cord and AC power connector (rear of cardiograph, see
figure 1-2 on page 1-8) to ensure that both are in a safe and operable condition. If the AC
power cord or AC power connector is not in a safe or operable condition, operate the
cardiograph on battery power and contact Philips Medical Systems for service.
The cardiograph has been safety tested with the recommended accessories, peripherals,
and leads, and no hazard was found when the cardiograph is operated with cardiac
pacemakers or other stimulators.
Do not connect any equipment or accessories to the cardiograph that are not manufactured
or approved by Philips Medical Systems or that are not IEC 60601-1 approved. The
operation or use of non-approved equipment or accessories with the cardiograph is not
tested or supported, and cardiograph operation and safety are not guaranteed.
The list of cables and other accessories with which Philips claims compliance with the
emissions and immunity requirements of IEC standard 60601-1-2 are listed in “Supplies
and Ordering Information” on page 7-20.
WARNINGWhen using additional peripheral equipment powered from an electrical source other
than the cardiograph, the combination is considered to be a medical system. It is the
responsibility of the operator to comply with IEC 60601-1-1 and test the medical system
according to the requirements. For additional information contact Philips Medical
Systems.
WARNINGDo not use non-medical peripherals within 1.83 meters or 6 feet of a patient unless the
non-medical peripherals receive power from the cardiograph or from an isolation
transformer that meets medical safety standards.
Only install Philips Medical Systems software on the cardiograph. The installation or use
of software not approved by Philips Medical Systems is strictly prohibited and
cardiograph safety and performance are not guaranteed.
Only use Philips Medical Systems replacement parts and supplies with the cardiograph.
The use of non-approved replacement parts and supplies with the cardiograph is strictly
prohibited. Cardiograph safety and performance are not guaranteed when non-approved
replacement parts and supplies are used with the cardiograph.
PageWriter Trim Cardiograph Instructions for Usev
The PageWriter Trim CardiographIntended Use
Manual measurements of ECG intervals and magnitudes should be performed on printed
ECG reports only. Do not make manual measurements of ECG intervals and magnitudes
on the R/T ECG display since these ECG representations are scaled.
Only use patient electrodes that are approved by Philips Medical Systems. The use of non-
approved patient electrodes may degrade cardiograph performance.
The Philips Medical Systems warranty is applicable only if you use Philips Medical
Systems approved accessories and replacement parts. See “Supplies and Ordering
Information” on page 7-20 for more information.
Before using the Patient Cable Arm with the cardiograph cart, properly install the counter
balance on the cardiograph base.
Only use the shielded LAN cable provided with the PageWriter Trim cardiograph, Philips
Part Number 989803138021. Do not use any other LAN cables with the PageWriter Trim
cardiograph. Use of unapproved LAN cables may result in radiated emissions that exceed
the limit specified by CISPR11 Class B.
The combined maximum weight that can be placed on the cardiograph cart shelf and the
top surface of the cart cannot exceed 20 kg (44 lbs). Do not place more than the specified
weight on the cardiograph top surface and shelf.
Do not connect any device to the RS-232 port on the rear of the cardiograph when the
patient data cable is connected to a patient.
There are no cardiograph parts that can be sterilized.
The cardiograph is not intended for direct, or invasive cardiac monitoring purposes.
Excessive, repetitive use of the cardiograph keyboard and the cardiograph Trim Knob may
result in a risk of developing carpal tunnel syndrome.
Ensure that the patient data cable is tucked away from the cardiograph cart wheels when
transporting the cardiograph. Ensure that the patient data cable does not present a hazard
when pushing the cardiograph cart.
The PageWriter Trim Cardiograph
Intended Use
The intended use of the cardiograph is to acquire multi-channel ECG signals from adult and
pediatric patients from body surface ECG electrodes and to record, display, analyze, and store
these ECG signals for review by the user. The cardiograph is to be used in healthcare facilities
by trained healthcare professionals. Analysis of the ECG signals is accomplished with
algorithms that provide measurements, data presentations, graphical presentations, and
interpretations for review by the user.
The interpreted ECG with measurements and interpretive statements is offered to the clinician
on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the
patient, the results of the physical examination, the ECG tracings, and other clinical findings.
viPageWriter Trim Cardiograph Instructions for Use
Indications for UseThe Philips 12-Lead Algorithm
A qualified physician is asked to overread and validate (or change) the computer-generated
ECG interpretation.
Indications for Use
The cardiograph is to be used where the clinician decides to evaluate the electrocardiogram of
adult and pediatric patients as part of decisions regarding possible diagnosis, potential
treatment, effectiveness of treatment, or to rule out causes for symptoms.
The Philips 12-Lead Algorithm
The PageWriter Trim Cardiograph software uses the Philips 12-Lead Algorithm. The
algorithm in the software analyzes the morphology and rhythm on each of the 12 leads and
summarizes the results. The set of summarized measurements is then analyzed by the
clinically-proven ECG Analysis Program.
12-lead Reports may include or exclude ECG measurements, reasons, or analysis statements.
Intended Use
The intended use of the Philips 12-Lead Algorithm is to analyze multi-channel ECG signals
from adult and pediatric patients with algorithms that provide measurements, data
presentations, graphical presentations, and interpretations for review by the user.
The interpreted ECG with measurements and interpretive statements is offered to the clinician
on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the
patient, the results of the physical examination, the ECG tracings, and other clinical findings.
A qualified physician is asked to overread and validate (or change) the computer-generated
ECG interpretation.
Indications for Use
The Philips 12-Lead Algorithm is to be used where the clinician decides to evaluate the
electrocardiogram of adult and pediatric patients as part of decisions regarding possible
diagnosis, potential treatment, effectiveness of treatment, or to rule out causes for symptoms.
PageWriter Trim Cardiograph Instructions for Usevii
1
1Getting Started
Welcome to the PageWriter Trim Cardiograph! With its intuitive operation, clearly labeled
PIM and lead wires, and convenient cart, it is the ideal cardiograph for processing large
volumes of ECGs quickly and easily.
The PageWriter Trim supports multiple ECG acquisition modes, display settings, and report
formats. Signal quality indicators provide instant feedback to the user to help ensure a quality
ECG each and every time.
This PageWriter Trim Cardiograph Instructions for Use provides comprehensive information
on the cardiograph and all of its components. It is intended to be used with the other materials
included in the PageWriter Trim Cardiograph Learning Kit.
NOTE Read and complete the materials included in the PageWriter Trim Cardiograph Learning Kit before
using the cardiograph. Read all Patient Safety information and pay close attention to all warnings and
cautions. For more information, see “Important Patient and Safety Information” on page -iii.
1-1
Getting StartedPageWriter Trim Cardiograph Learning Kit
PageWriter Trim Cardiograph Learning Kit
Philips Medical Systems provides detailed instructional and reference materials in the
PageWriter Trim Learning Kit
The PageWriter Trim Learning Kit contains the Getting Started Guide, a Quick Help Card,
and the User Documentation and Interactive Training CDs.
Figure 1-1The PageWriter Trim Cardiograph Learning Kit
A
B
C
About the PageWriter Trim Learning Kit
Getting Started Guide (A)
This guide includes important information that must be read before operating the
cardiograph. It includes an overview of cardiograph features and functions, assembly and
setup instructions, how to configure the cardiograph, and how to order supplies.
Quick Help Card (B) This card provides quick reference information about basic
cardiograph features, lead placement, and signal quality indicators. It is also provided as a
PDF file on the PageWriter Trim User Documentation CD.
D
PageWriter Trim Cardiograph Interactive Training Program CD (C)
The interactive training CD contains a computer-based training program about the proper
recording of ECGs and how to operate the cardiograph.
Use this program to train the entire ECG staff in the proper operation of the cardiograph.
To run the interactive training program:
XInsert the CD into a PC CD-ROM drive. The CD only works on PCs with a Windows
operating system.
The interactive training program starts automatically.
1-2PageWriter Trim Cardiograph Instructions for Use
Getting StartedAttaching the Cardiograph to the Cart
PageWriter Trim User Documentation CD (D)
The User Documentation CD contains the following files:
–Philips 12-Lead Algorithm Physician’s Guide
This PDF file describes the Philips 12-Lead Algorithm, and lists all of the interpretive
statements included in the 0A criteria.
–PageWriter Trim Instructions for Use
This PDF file includes complete information on all aspects of using and maintaining
the cardiograph.
–PageWriter Trim Quick Help Card
A PDF version of the Quick Help Card provided in the Learning Kit.
–PageWriter Trim Cardiograph XML Schema (English only)
The schema for the XML ECG data exported from the PageWriter Trim cardiograph.
To view user documentation:
XInsert the CD into a PC CD-ROM drive. The CD only works on PCs with a Windows
operating system.
The user documentation CD main menu opens automatically. Click on a blue button or the
file name to open the file.
NOTE Adobe Acrobat Reader does not need to be installed on the PC to view the PDF files on the CD. If
PDF files from the CD are saved to a PC hard drive, Acrobat Reader will need to be installed on the
PC in order to view the files. For a free install go to: www.adobe.com.
1
If the menu does not automatically appear, open the CD in Windows Explorer.
2Double-click the file menu.pdf on the CD. The PDF menu appears. Any of the files on the
CD may be printed or saved to a PC hard drive.
Attaching the Cardiograph to the Cart
The PageWriter Trim cardiograph is available with an optional cart that includes a storage
shelf and a holder for the PIM (Patient Interface Module).
CAUTIONFollow the procedure below to ensure that the cardiograph is securely fastened to the cart before
use.
PageWriter Trim Cardiograph Instructions for Use1-3
Getting StartedAttaching the Cardiograph to the Cart
To attach the cardiograph to the cart:
Align the front feet of the cardiograph with the front locking holes on the cart. Align the
1
rear feet of the cardiograph with the rear screw holes on the cart. Lower the cardiograph
onto the cart.
2Slide the cardiograph back to lock the front and rear feet into place. Insert the screws
through the bottom of the cart and through the screw holes. Tighten the screws with a
Phillips head screwdriver.
1-4PageWriter Trim Cardiograph Instructions for Use
Getting StartedPageWriter Trim I Cardiograph Parts
PageWriter Trim I Cardiograph Parts
The following sections show front, side, and rear views of the PageWriter Trim I cardiograph.
Figure 1-1PageWriter Trim I Cardiograph and Cart (Front View)
A
G
H
B
C
D
E
F
A Patient Interface Module (PIM) EPIM Leads
BPrinter paper/report storage slot FWheel Brake
C Storage ShelfG Control Panel
D AC Power CordH Printer Paper Drawer
CAUTIONAlways lock the wheel brake (F) when the cart is not in use. Press down on the wheel brake to set or
to release the wheel brake.
PageWriter Trim Cardiograph Instructions for Use1-5
Getting StartedPageWriter Trim I Cardiograph Parts
Figure 1-2PageWriter Trim I Cardiograph (Rear View)
IK
NM
J
LO
IReset Button M Battery Door
JAC Power Cord ConnectorN PIM Connector
KEquipotential Post O Serial Connector (not supported)
LFuse Door
WARNINGDo not connect a LAN cable connector to the PIM connector.
Do not plug a telephone connector into the PIM connector.
1-6PageWriter Trim Cardiograph Instructions for Use
Getting StartedPageWriter Trim II and III Cardiograph Parts
PageWriter Trim II and III Cardiograph Parts
The following sections show front, side, and rear views of the PageWriter Trim II and III
cardiographs.
Figure 1-1PageWriter Trim II and III Cardiograph and Cart
(Front View)
G
A
H
I
J
B
C
D
E
F
A Patient Interface Module (PIM) FWheel Brake
BPrinter paper/report storage slot G Control Panel
C Storage Shelf H Keyboard
D AC Power Cord ITrim Knob
EPIM LeadsJPrinter Paper Drawer
CAUTIONAlways lock the wheel brake (F) when the cart is not in use. Press down on the wheel brake to set or
to release the wheel brake.
PageWriter Trim Cardiograph Instructions for Use1-7
Getting StartedPageWriter Trim II and III Cardiograph Parts
Figure 1-2PageWriter Trim II and III Cardiograph (Rear View)
KL
U
T
M
Q
RS
NO
P
KReset ButtonQ Battery Door
LPC Card Eject ButtonRPIM Connector
M PC Card SlotSSmartCard Reader (USB) Connector
N AC Power ConnectorTBarcode Reader or Magnetic Card
Reader Connector
O Equipotential PostU Serial Connector (not supported)
PFuse Door
WARNINGDo not connect the LAN cable connector into the PIM connector.
Do not plug a telephone connector into the PIM connector.
1-8PageWriter Trim Cardiograph Instructions for Use
Getting StartedInstalling the Battery
Installing the Battery
The cardiograph is shipped with one removable battery. Install the battery before plugging the
cardiograph into AC power.
Figure 1-3Installing the Battery
To install the battery:
1
Open the battery door on the rear of the cardiograph. Push the recessed tab in and pull it
down to open the battery door.
2Insert the battery into the compartment with the connector facing down and the pull tab
facing out.
3Push in the pull tab to lock it.
4Reattach and close the battery door.
5Attach the AC power cord to the rear of the cardiograph.
PageWriter Trim Cardiograph Instructions for Use1-9
Getting StartedInstalling the Battery
6Plug the AC power cord into a grounded electrical outlet. Check that the green AC power
indicator light is on (front of cardiograph).
7Charge the battery for at least 24 hours before mobile use.
Battery Charging Indicator, PageWriter Trim II and III
On the PageWriter Trim II and III, a lightening bolt icon () appears on the Battery
Level Indicator on the Status Bar when the cardiograph is plugged into AC power and the
battery is charging. Check the Battery Power Indicator on the Status Bar to ensure that the
battery is fully charged, see “PageWriter Trim II and III Status Bar” on page 1-27. The
cardiograph can operate on AC power while the battery is charging, but the battery will
charge at a slower rate.
Battery Charging Indicator, PageWriter Trim I
On the PageWriter Trim I, the following battery icon () flashes on the LCD display
when the cardiograph is plugged into AC power and the battery is charging, see “LCD
Display” on page 1-24. The battery level indicator displays the following battery icon
() when the battery is fully charged. The cardiograph can operate on AC power while
the battery is charging, but the battery will charge at a slower rate.
1-10PageWriter Trim Cardiograph Instructions for Use
Getting StartedLoading the Printer Paper
Loading the Printer Paper
Replace the printer paper when a red stripe appears on the printed ECG report. Only use
Philips Medical Systems replacement printer paper. For part number and ordering
information, see page 7-20.
Tips for loading printer paper
Always load less than 100 sheets of printer paper into the paper tray
Ensure that the entire first page of the new paper roll is fully draped over the roller before
closing the printer door
PageWriter Trim II and III:
Ensure that the paper size configured for the cardiograph is the same size paper being
loaded into the paper drawer (see page 5-18)
Figure 1-4Loading Printer Paper
A
A Paper sensor hole
To change the printer paper:
Open the paper drawer on the front of the cardiograph.
1
2Insert a new pack of printer paper with the printed side facing up. Ensure that no more
than 100 sheets are being inserted into the paper tray.
3Ensure that the paper sensor hole (A) is positioned as shown in Figure 1-4.
4Drape the entire first sheet over the roller. Ensure that the perforated edge of the paper
aligns with the edge of the paper drawer.
5Close the paper drawer.
6Tear off the first sheet as shown in Figure 1-4.
PageWriter Trim Cardiograph Instructions for Use1-11
Getting StartedPatient Interface Module (PIM)
Patient Interface Module (PIM)
The Patient Interface Module (PIM) is a hand-held device that connects to the cardiograph.
The lead wires on the PIM attach to the electrodes placed on the patient. The exterior of the
PIM is labeled and color-coded for quick and easy lead identification.
Figure 1-5Patient Interface Module
A
B
CD
A Lead Wire LabelingC Limb Lead Wires
BPatient Data CableD Precordial (Chest) Lead Wires
1-12PageWriter Trim Cardiograph Instructions for Use
Getting StartedPatient Interface Module (PIM)
Inserting the Lead Wires into the PIM
The exterior of the PIM is color-coded for quick and easy lead identification. The lead wires
that are included with the cardiograph must be inserted into the correct lead wire connector on
the PIM.
Figure 1-6PIM connector and lead wire labeling
C
B
A
A Lead Wire ConnectorC Lead Wire Labeling
BLead Wire
To insert the lead wires into the PIM:
Locate the lead wires in the cardiograph packaging.
1
2Match the lead wire to the correct lead wire connector on the PIM.
3Insert each lead wire into the correct lead wire connector on the PIM.
4Push down on the lead wire to ensure that there is no gap between the lead wire and the
connector.
PageWriter Trim Cardiograph Instructions for Use1-13
Getting StartedPatient Interface Module (PIM)
Connecting the PIM to the Cardiograph
Connect the patient data cable on the PIM to the PIM connector port () on the rear panel
of the cardiograph.
To connect the PIM to the cardiograph:
XConnect the patient data cable to the PIM connector port on the rear of the cardiograph.
WARNINGTo ensure safety and prevent damage to the system, only connect the patient data cable
to the PIM connector port on the rear of the cardiograph.
1-14PageWriter Trim Cardiograph Instructions for Use
Loading...
+ 188 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.