Philips Optimus RF User manual

X--RAY GENERATION
LIST OF DOCUMENTATION IN THIS BINDER:
SUBSYSTEM MANUAL OPTIMUS R/F
UNIT MANUAL Extension set for an additional tube assembly WG/GWB
UNIT MANUAL 26 V DC / 230 V AC Adapter
UNIT MANUAL Handswitch for OPTIMUS
UNIT MANUAL Extension Photomultiplier (SEV)
UNIT MANUAL Mains group EWD
UNIT MANUAL Power Distribution Unit (PDU)
Note: indicated document present
LIST OF ALL BINDERS FOR X--RAY GENERA
-- SUBSYSTEM MANUAL OPTIMUS R/F (this binder)
TION:
4512 984 25672 REV AA
OPTIMUS_RF_01_25672_AA
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
0.1
EW
Philips Medical Syst ems DMC GmbH
OPTIMUS R/F
SERVICE MANUAL
742
SUBSYSTEM
INTRODUCTION AND TECHNICAL DATA
INSTALLATION
1
2
9890 000 02182
FAULT FINDING
REPLACEMENT
PROGRAMMINGS
ADJUSTMENTS
3
4
5
6
358J92
CAN-controlled X-ray generator of the converter type
DMC Hamburg
Printed in Hamburg, Germany
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
ACCEPTANCE
SERVICE INFORMATION
PARTS LIST
SCHEMATIC DRAWINGS
7
8
P
Z1-
4512 984 25672 REV AA 1
OPTIMUS_RF_reg_25672_AA
WIRING DIAGRAMS
Z2-
OPTIMUS R/F
SERVICE MANUAL -- SUBSYSTEM
OPTIMUS R/F
Author: U. Wente
Type No: 9890 000 02182
Techn. No: Basis 4512 104 70625
Release: 3.5
In case there are any questions concerning this manual, please send this LOPAD via fax to 49/(0)40/5078 2481
File: OPTIMUS_RF_25672_AA
List of pages and drawings (LOPAD) Manual Order No: 4512 984 25672
released: 09/2002
0.1
1
3.1
3.2
_________________________________________
1--0.1 ... 0.2 (d/02.1)
1--1 ... 16 (d/02.1) Z--1.1 (01.0) A4 4512 982 0092.
Z--1.2 (01.0) A4 4512 982 0092. Z--1.3 (01.0) A4 4512 982 0099. Z--1.4 (01.0) A4 4512 982 0099. Z--6.1 (01.0) A4 4512 982 0099. Z--6.2 (01.0) A4 4512 982 0099. Z--7.1 (01.0) A3 4512 982 0088. Z--7.2 (01.0) A3 4512 982 0088. Z--7.3 (01.0) A3 4512 982 0088. Z--7.4 (01.1) A3 4512 982 0088. Z--7.5 (01.0) A3 4512 982 0088. Z--7.10 (01.0) A3 4512 982 0010. _________________________________________
2--0.1 ... 0.3 (c/02.1)
2--1 ... 86 (c/02.1) 2Z--2.0 3x (01.0) A4 4512 983 05591
2Z--2.1 (a/01.0) A4 4512 983 05601 2Z--2.2 (a/01.0) A4 4512 983 0561 1 2Z--2.3 (a/01.0) A4 4512 983 05621 2Z--2.4 (a/01.0) A4 4512 983 05631 2Z--2.5 (a/01.0) A4 4512 983 05641
_________________________________________
3--0.1 (b/02.0)
3--1 ... 93 (b/02.0) 3Z--1 (a/01.0) A4
3Z--21 (97.1) A4 _________________________________________
4--0.1 (a/02.1)
4--1 ... 25 (a/02.1) _________________________________________
5Z--1 (a/01.0) A3 5Z--2 (a/01.0) A3 _________________________________________
6--0.1 (b/02.0)
6--1 ... 24 (b/02.0) _________________________________________
7--0.1 (a/01.0)
7--1 ... 4 (a/01.0) _________________________________________
8--1 (00.0) FCO--Checklist 8--2 (00.0) FCO--Checklist _________________________________________
P--List OPTIMUS R/F (840061) _________________________________________
2Z--2.6 (01.0) A4 4512 983 05651 2Z--2.7 (01.0) A4 4512 983 05661 2Z--2.8 (01.0) A4 4512 983 05671 2Z--2.9 (01.0) A4 4512 983 05681 2Z--2.10 (01.0) A4 4512 983 05691 2Z--3 (97.0) A3 4512 983 05721 2Z--4 (97.0) A4 4512 983 05771 2Z--5 (97.0) A4 4512 983 05731 2Z--10 (02.0) A3 4512 983 05701 _________________________________________
4512 984 25672 REV AA
OPTIMUS_RF_lopad_25672_AA
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
3.1
Z0--1 (02.0)
Z1--1.1 (98.0) A3 Z1--1.2 (97.0) A3 4512 983 05561 Z1--2.1 (b/97.0) A3 4512 983 05581 Z1--2.2 (c/97.0) A3 4512 983 05931 Z1--2.3 (99.0) A3 Z1--3.3 (a/01.0) A3 Z1--4.1 (a/01.0) A3 Z1--4.2 (a/99.0) A3 Z1--5.1 (97.1) A3 4512 983 06551 Z1--6 (02.0) A3 4512 983 05531 Z1--11.1 (02.0) A3 4512 983 05521 Z1--11.2 (96.0) A4 Z1--13.2 (c/00.0) A3 Z1--14.1 (b/98.0) A3 4512 983 05541 Z1--14.2 (c/97.1) A3 4512 983 05551 Z1--15.1 (97.0) A3 4512 983 05571 Z1--15.2 (97.1) A3 4512 983 05711 _________________________________________
Z0--2 (02.0)
Z2--1.0 (96.0) A4 Z2--1.1 (d/99.0) A3 4512 983 06461 Z2--1.2 (e/99.0) A3 4512 983 06471 Z2--1.3 (c/98.0) A3 4512 983 06481 Z2--2.1.1 (97.0) A3 4512 983 06491 Z2--2.1.2 (97.0) A3 4512 983 06501 Z2--2.2.1 (96.0) A3 4512 983 06511 Z2--2.2.2 (a/97.0) A3 4512 983 06521 Z2--5.1 (a/02.0) A3 Z2--5.2 (b/02.0) A3 4512 983 0551 1 Z2--5.3 (a/02.0) A3 Z2--5.4 (02.0) A3 Z2--13 (d/00.0) A3 Z2--14.1 (b/99.0) A3 Z2--14.2 (a/96.0) A3 Z2--14.3 (a/96.0) A3 Z2--14.4 (97.0) A3 Z2--15.1 (97.0) A3 4512 983 05741 Z2--16 (97.0) A3 Z2--17 (97.0) A3 _________________________________________
(nur für Fabrik-- Auslieferung)
Service software No:
-- OMA : 4512 116 0220x
-- OMB : 4512 116 0230x _________________________________________
OPTIMUS R/F
E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
4512 984 25672 REV AA3.2
OPTIMUS_RF_lopad_25672_AA
OPTIMUS R/F
1. Product information 1--1..................................................
1.1. Applications 1--1..........................................................
1.2. Options 1--1..............................................................
1.2.1. Hardware options 1--1......................................................
1.2.2. Software options 1--2......................................................
2. Compatibility 1--3........................................................
2.1. Generator components 1--3.................................................
2.2. Tubes 1--3................................................................
3. Mechanical data 1--4......................................................
4. Environmental data 1--5...................................................
4.1. Electrical environment 1--5..................................................
4.2. Climatic conditions 1--5.....................................................
4.3. Emission 1--5.............................................................

INTRODUCTION AND TECHNICAL DATA

INTRODUCTION AND TECHNICAL DATA
TEXT
Contents 1--0.1............................................................
5. Electrical data 1--6........................................................
5.1. Power data and mains conditions 1--6........................................
5.2. Power supply for applications 1--8...........................................
5.3. Operating data 1--8........................................................
5.4. Power supply 1--9.........................................................
5.4.1. Type of power supply 1--9..................................................
5.4.2. Calculating the mains resistances 1--10........................................
5.4.3. Earth-leakage circuit breaker 1--11............................................
5.4.4. Emergency--OFF device 1--11................................................
6. Tools 1--12................................................................
7. Traceable items 1--13......................................................
8. Preparation 1--14..........................................................
8.1. Installation material 1--14....................................................
8.2. Cables 1--14...............................................................
8.3. Manpower 1--15............................................................
9. Planned maintenance 1--16.................................................
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021_inh
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--0.1
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
DRAWINGS
Generator cabinet Z--1.1......................................................
Room layout Z--1.2...........................................................
Operating panel Z--1.3........................................................
Power Distribution Unit (PDU) Z--1.4............................................
Connection of generator Z--6.1................................................
Connection of generator with PDU Z--6.2.......................................
Connection diagram Z--7.1....................................................
Connection diagram Z--7.2...................................................
Connection diagram Z--7.3....................................................
Connection diagram Z--7.4....................................................
Earthing diagram Z--7.5.......................................................
Legend for earthing and cabling diagram Z--20.1..................................
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--0.2
OPTIMUS_RF_1_d021_inh
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA

1. Product information

The Optimus family of generators for radiography is based on computer-controlled converter technology. The converter operates in the non-audible frequency range.
Applicable options are essentially achieved by releasing software modules using customized PAL ICs. Control between the internal Function Units (FUs) and the external online equipment takes place by a CAN bus.
Safety-relevant signals are transferred directly on the so-called “Signal bus”. Units without any CAN interface are operated by the ”Adapter for 4 auxiliary units WA” option.

1.1. Applications

-- Radiography
-- Tomography
-- Fluoroscopy

1.2. Options

Component overview according to the commercial catalogue.
Only the versions in the current commercial catalogue can be ordered.
If an existing generator is to be upgraded the commercial department must order:
MGR0011 (upgrade of an existing configuration) + MGRxxxx +S/N
1.2.1. Hardware options + MGR0011 + S/N
-- Dual-speed rotor control 9890 000 0268x MGR2082.............................. ............
-- Mains transformer PDU: 400 -- 480V; 50 / 60Hz,
also for 400V mains supply without neutral lead N
with taps for 400 / 440 / 460 / 480V 9890 000 0260x.....................
-- Mains transformer: 190--390V; 50 / 60Hz with taps for 190 / 200 / 207 / 220 / 230 / 240 / 250 /
343 / 380 / 390V max. 50kW! 9803 720 8100x.......................
-- Adapter for 4 aux. units WA 9890 000 0231x MGR2131........................... ............
-- Adapter decade cable for 4 aux. units R/F WB 9890 000 0261x...........
-- Option rack 9890 000 0232x..........................................
-- Extension set for one additional tube 9890 000 0234x....................
-- Tube extension WG 9890 000 0238x..................................
-- Operating panel 9890 000 0240x......................................
-- Operating module Optimus 9890 000 0278x............................
-- Operating desk data cable 10m, 20m, 30m 9890 000 0241x / 2x / 3x..............
-- Stand for operating panel 9890 000 0244x MGR1482............................. ............
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--1
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
-- Wall mounting of operating panel 9890 000 0245x.......................
-- 26VDC / 230VAC adapter 9890 000 0246x MGR2281............................. ............
-- Handswitch for Optimus 9890 000 0249x...............................
-- Patient Data Organizer (PDO) 9890 000 0255x MGR2091......................... ............
-- Decade cable set 14 x 4m top decade AMP decade 9803 704 2010x....
-- Extension photo pick-up Optimus 9890 000 0258x.......................
1.2.2. Software options + MGR0011 + S/N
Software options are provided by the function key (see also 5Z--1, EZ 139 Central Unit D38). Additional hardware components are not required.
-- Automatic Exposure Control (AEC) 9890 000 0281x
-- Anatomic Programmed Radiography / Fluorography (APR/F) 9890 000 0282x
1)
1)
-- Automatic Tomo Time Input (TTI) 9890 000 0222x MGR2121............................. .......
-- Tomo Density Control (TDC) 9890 000 0223x MGR2122................................. .......
-- VARIOFOCUS 9890 000 0227x MGR2101............................................. .......
-- Area dose calculator 9890 000 0256x MGR2141........................................ .......
1) = Options only for base 9890 000 0218x
Options are always included in base 9890 000 0216x
MGR2171........................... .....
MGR2181.... .....
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--2
OPTIMUS_RF_1_d021
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA

2. Compatibility

2.1. Generator components

-- Base OPTIMUS 9890 000 0218x..............................
-- H.V. transformer R/F 1 tube, 50kW 9890 000 0270x.. .........
-- H.V. transformer R/F 2 tubes, 50kW 9890 000 0271x MGR2061 (Upgrade 1 ----> 2) tubes.. ........ ....
-- H.V. transformer R/F 1 tube, 65 / 80kW 9890 000 0272x.. ......
-- H.V. transformer R/F 2 tubes, 65 / 80kW 9890 000 0273x MGR2062 (Upgrade 1 -- --> 2) tubes.. ..... ....
-- 50kW extension -- R/F 9890 000 0274x........................
-- 65kW extension -- R/F 9890 000 0275x........................
-- 80kW extension -- R/F 9890 000 0276x........................
-- Firmware Rel. 3.5 9890 000 0251x............................

2.2. Tubes

Recommended standard tubes:
-- RO 17 50
-- SRO 25 50
-- SRO 33 100
Further compatible tubes:
-- RO 30 -- SRO 09 51 -- SRO 20 55
-- RO 12 30 -- SRO 13 30 -- SRO 22 50
-- RO 16 48 -- SRO 20 50 -- SRO 32 100
-- RO 30 50
Compatible tube housings:
-- ROT 350
-- ROT 351
The latest information on further tubes which are connectable is available at the service center Hamburg.
Note
When the generator is retrofitted it is important to use the screened cable 3 x 1.31mm
2
(0722 215 02054) as the stator cable. If necessary exchange the old stator cable.
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--3

3. Mechanical data

C
For installation dimensions and weights see drawings Z--1.1.
Transport data:
Weights [kg] Dimensions [cm]
aseNo.Contents
-- Generator cabinet
1
-- Operating panel
-- Cables
net gross length width height
178 226 210 82 84
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
1--tube version:
73
100
77 67 80
2
2--tube version:
88
115
Contents: H.V. generator
OPTIMUS R/F(d/02.1)1--4
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS_RF_1_d021
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA

4. Environmental data

The environmental data comply with to PMS standard UXW 13600.

4.1. Electrical environment

Class S0 -- Dedicated mains supply, 3 phases and neutral. Thus single phase voltage is also available.
A low impedance, permanently installed connection, fed in by the step down transformer of the hospital to supply large systems like in MR, CT and X--ray departments is required.
Note
Use always a mains cable with 4 wires and concentric PE--shield, type NYCY.

4.2. Climatic conditions

Ambient temperature 10_C--40_C...................
Relative humidity 15% -- 90%; no condensation.......................
Relative atmospheric pressure 70kPa -- 110kPa...........

4.3. Emission

Heat dissipation max. 1200W; average per hour........................
Noise level, normal operation 46dBA............
Noise level, maximum power operation 55dBA....
EMC IEC 950..................................
To avoid any possible annoying noise of the implemented fans it is advisable to install the generator cabinet outside the examination room.
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--5

5. Electrical data

withexternaltransformerPDU(option
)
I
3+1x1
0
2
(L1,L
2,L
3,PE)

5.1. Power data and mains conditions

Voltage
50kW 65kW 80kW
Mains voltage 3 x 400V ±10% (415V
3 x 400 / 440 / 460V ±10% *
6%
+
3 x 480V
=*=
The following connection cables are recommended:
|
--10%
+6%
/ 380V
*
-- 5 %
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
)
nput:
Output: 4 x 4mm
4+1x4mm 3 x 1,5mm
mm
2
Generator supply 3 x 400V
2
Device supply 3 x 220V (option)
2
Switch control and temperature supervision
3 x 190 ... 343V ±10% **
** = with external transformer; max. 50kW (option)
Frequency
50kW 65kW 80kW
Mains frequency 49 ... 61Hz
OPTIMUS R/F(d/02.1)1--6
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS_RF_1_d021
OPTIMUS R/F
(
)
INTRODUCTION AND TECHNICAL DATA
Max. mains current
Voltage
50kW 65kW 80kW
Exposure: 400V 145A 190A 230A
440V 135A 180A 215A
460V 125A 170A 210A
480V 120A 160A 205A
190V 300A -- --
Short--time power consumption
100kVA 132kVA 160kVA
[I x U x 3]
Fluoro: 400V -- 8A
480V -- 7A
Fuse protection
slow blow
Connected load [I
xUx3]
Fuse
35A 50A
100A at 240V --
25kVA 35kVA
Emergency power supply:
static
(Inverter)
Short--time power consumption
[I x U x 3]
dynamic
(Diesel generator
with flywheel mass)
Connected load
[I
xUx3]
Fuse
Mains resistance
Voltage
50kW 65kW 80kW
400V 300m 200m
440V 350m 240m
460V 350m 240m
480V 400m 300m
480V valid for DOD only 300m 240m 180m
Note
500m
is the absolute max. mains resistance.
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--7
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA

5.2. Power supply for applications

Generator power
Supply 50kW 65kW 80kW
Output 1 230V / 400V; generator switched and protected ----> I
Output 2 230V / 400V; generator switched and protected ----> I
Output 3 230V / 400V; generator not switched and protected ----> I
Sum value I
max(1+2+3)
10A
Only with external transformer PDU
Output 4 127V / 220V generator switched and protected ----> I
Output 5 double socket 127V / 15A to phase L1;
generator not switched and protected ----> I
Sum value I
max (4 +5)
16A

5.3. Operating data

Generator power
1
2
3
4
5
Data
50kW 65kW 80kW
Exposure: Tube current 1 ... 650mA 1 ... 900mA 1 ... 1100mA
Tube voltage 40 ... 150kV in kV-- or %--steps
mAs product 0,5 ... 850mAs
Exposure time 1ms ... 6s / 16s
Exposure frequency 12exp./s
Interfacing option for door contact, external radiation warning indicator
Fluoro: Tube current 0,25 ... 6mA
Tube voltage 40 ... 110kV
kV/mA curves 3
Setting time 1s
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--8
OPTIMUS_RF_1_d021
OPTIMUS R/F

5.4. Power supply

5.4.1. Type of power supply
3 phase WYE
INTRODUCTION AND TECHNICAL DATA
L1
L2
L3 (N)
3 phase DELTA, balanced earth or floating
L1
L2
L3
X-ray installation
X-ray installation
-- 400V
-- 440V / 460V / 480V with external mains transformer PDU 9890 000 0260x.
-- Neutral not required if the external mains transformer PDU 9890 000 0260x is ordered.
-- 190V ... 343V with external mains transformer 9803 720 8100x (max. 50kW).
-- External mains transformer PDU 9890 000 0260x is required.
-- 400V / 440V / 460V / 480V
-- 190V ... 343V with external mains transformer 9803 720 8100x (max. 50kW).
3 phase DELTA, grounded
L1
-- External mains transformer PDU 9890 000 0260x is required (requires modification at the EMC-- filter of the kV power unit).
X-ray installation
L2
L3
Caution! Ensure the sequence of phases in the wall junction box corresponds to designations L1, L2, L3.
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
-- 400V / 440V / 460V / 480V
1--9
5.4.2. Calculating the mains resistances
Note
The cross section of lead l
must not exceed 25mm2. (See figure below).
3
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
If possible the sum of R smaller than the R
requires.
XG
0,R1,R2
and R3should be
R
+R1+R2+R3<R
0
With higher internal mains resistances the generator output is reduced correspondingly.
R
designates the mains resistance of the distributor transformer
0
R
depends on the length of lead l1between distributor transformer and mains distributor
1
L1 L2 L3
l
1
and on the selected cross section as well:
==> R
R
consists of upstream elements such as:
2
1=l1
× R
Cu
RCufrom table below
-- Emergency-OFF switch 4.0m.................
-- Earth-leakage circuit breaker 5.5m............
l
3
XG
MEX 100
L1 L2 L3
357J94
-- Fuse 5.5m.................................
-- Surge arrester WN 23.0m...................
R
depends on the length of lead l3between mains distributor and wall junction box
3
and on the selected cross section as well:
==> R
3=l3
× R
Cu
RCufrom table below
The resistances consider the go and return lines so that the calculation can be based on simple cable lengths.
Copper cross section
[mm
2
]
Resistance R
[m/m]
Cu
16 2.19 25 1.4 35 1.0 50 0.7 70 0.5
95 0.38 120 0.30 150 0.24
Note
500m
is the absolute max. mains resistance.
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--10
OPTIMUS_RF_1_d021
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA
5.4.3. Earth-leakage circuit breaker
To be provided between mains fuse and X-ray installation depending on local regulations.
Siemens earth-leakage circuit breaker:
-- Order No.: 5SZ3 466 OKG00
-- Rated fault current 30mA
-- Rated current 63A
-- Connection terminals for wire cross sections of up to 25mm
2
5.4.4. Emergency--OFF device
To be provided depending on local regulations.
There are 2 possibilities:
1. All the Emergency--OFF buttons are connected in series and looped into the switch--ON circuit (12VDC) of the generator.
2. The Emergency--OFF circuit acts on an external mains contactor which switches OFF the power before it is fed into the generator.
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--11
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA

6. Tools

-- Service engineer standard tool kit
-- Service-PC:
IBM-compatible, 640kB RAM, 3.5” floppy disk drive, 1 serial port
-- Installation and service software OMA: 4512 116 022xx / OMB: 4512 1 16 023xx.
Supplied on two floppy disks within this generator service manual.
-- PC-hardkey (DIAGGEN):
Necessary to carry out the installation and to run the service software (special programmings, fault finding).
-- 0-modem cable:
Minimum length is distance between generator cabinet and operating desk. Male 25-pole D-Sub connector at the generator side. A 5m data cable of bucky controller can be used: 4512 130 5693x
-- Mains resistance measuring instrument
-- Dose measuring instrument
-- mAs--meter
-- Multimeter
-- Digital oscilloscope with 2-beam memory
-- Recommended PLCC extraction tool (AMP 822154--1): 2422 487 89772
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--12
OPTIMUS_RF_1_d021
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA

7. Traceable items

The following items have serial numbers of the following format when delivered ex factory:
1. Generator cabinet 6 digit serial number........
2. H.V. tank 7 digit serial number................
3. Operating desk 8 digit serial number...........
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--13
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA

8. Preparation

Connection of the generator: see drawing Z--6.1................
Connection of the generator with PDU: see drawing Z--6.2.......
Operating panel: see drawing Z--1.3..........................
Connection diagram: see drawing Z--7.1/.2/.3/.4.......................
Earthing diagram: see drawing Z--7.5.........................
Legend for earthing and cabeling: see drawing Z--7.10...........

8.1. Installation material

To be ordered from the service department of PMS Hamburg:
-- Wall junction box 4512 103 7538x.............................
2
inclusive connection block (25mm
-- Relay for radiation warning indicator 4512 100 4523x............
1 interface relay with a floating contact (230V/1A) is included in the scope of delivery of the generator.
) for mains supply and connection block (10mm2) for unit supply.

8.2. Cables

H.V . cables
with O3 / O3 plugs: 9806 402 6xx02................................
length: 6m -- 30m in steps of 2m...........................................
capacity: 155pF/m.........................................
diameter: 16.5mm........................................
The cable length is indicated by the 9th and 10th digit of the numeric code.
Thermal contact cable
-- 2-wire screened for 1 excess temperature switch 4512 100 66151.............
-- 10-wire screened for additional supervision like
temperature alarm switch, buzzer, selection indicator 0722 215 19005.........
Stator cable
3×1.31mm
Note
The above described cables are part of the pre--installed systems.
2
, screened 0722 215 02054.....................................
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--14
OPTIMUS_RF_1_d021
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA
AMPLIMAT cable
with D-Sub and 3--Plus plug:
12m 9890 000 01721......................
16m 9890 000 01731......................
20m 9890 000 01741......................
24m 9890 000 01751......................
Caution! AMPLIMAT cables 9803 507 0xx02 (for hybrid measuring chambers 9803 509 xxxxx) with 3--Plus plugs at both ends must be connected in the generator by the following adapter for each cable:
Adapter for AMPLIMAT cable: 4512 108 09042. The generator includes 1 adapter.
The hybrid measuring chambers 9803 509 xxxxx require connection (chassis) between contacts:
D--Sub end GND (13) <---- --> RF 0V (8) (generator input)
or
3--Plus end GND (N) <------> RF 0V (J) (generator input)
This connection is established by the adapter for the AMPLIMAT cable. See drawing Z1--6 ”Basic interface”.
In case a hybrid measuring chamber 9803 509 xxxxx is not operated with the required
AMPLIMAT cable 3--Plus / 3--Plus 9803 507 0xx02... ....
but with
AMPLIMAT cable D--Sub / 3--Plus 9890 000 017xx... ....
make sure to establish this connection (13 <-- -- --> 8) in the D--Sub connector!
For ALC measuring chambers 9890 000 016xx connection GND <------> RF 0V is not permitted. Therefore, ALC measuring chambers AMPLIMAT cables 9890 000 017xx should always be used.
Operating desk
Note
Use the shortest cables. Noise immunity increases.
cable set 10m 9890 000 02411.......................
20m 9890 000 02421.......................
30m 9890 000 02431.......................

8.3. Manpower

At least two persons are necessary to insert the H.V. tank in the generator cabinet. The weight of the 2--tube version is about 88kg.
OPTIMUS R/F (d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--15
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA

9. Planned maintenance

The technical documentation for carrying out maintenance work in compliance with the applicable regulations are available at the responsible authority of Philips Medical Systems.
The importance of having maintenance implemented is pointed out to the operator in the operating instructions.
It must be guaranteed that the person carrying out maintenance work knows about the respective national regulations and that this person observes these regulations throughout all steps of maintenance work.
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--16
OPTIMUS_RF_1_d021
Loading...
+ 306 hidden pages