Legend for earthing and cabling diagramZ--20.1..................................
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--0.2
OPTIMUS_RF_1_d021_inh
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA
1.Product information
The Optimus family of generators for radiography is based on computer-controlled converter technology.
The converter operates in the non-audible frequency range.
Applicable options are essentially achieved by releasing software modules using customized PAL ICs.
Control between the internal Function Units (FUs) and the external online equipment takes place by a CAN bus.
Safety-relevant signals are transferred directly on the so-called “Signal bus”.
Units without any CAN interface are operated by the ”Adapter for 4 auxiliary units WA” option.
1.1.Applications
-- Radiography
-- Tomography
-- Fluoroscopy
1.2.Options
Component overview according to the commercial catalogue.
Only the versions in the current commercial catalogue can be ordered.
If an existing generator is to be upgraded the commercial department must order:
MGR0011 (upgrade of an existing configuration)
+ MGRxxxx
+S/N
The latest information on further tubes which are connectable is available at the service center Hamburg.
Note
When the generator is retrofitted it is important to use the screened cable 3 x 1.31mm
2
(0722 215 02054) as the
stator cable.
If necessary exchange the old stator cable.
OPTIMUS R/F(d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--3
3.Mechanical data
C
For installation dimensions and weights see drawings Z--1.1.
Transport data:
Weights [kg]Dimensions [cm]
aseNo.Contents
-- Generator cabinet
1
-- Operating panel
-- Cables
netgrosslengthwidthheight
1782262108284
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
1--tube version:
73
100
776780
2
2--tube version:
88
115
Contents: H.V. generator
OPTIMUS R/F(d/02.1)1--4
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS_RF_1_d021
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA
4.Environmental data
The environmental data comply with to PMS standard UXW 13600.
4.1.Electrical environment
Class S0 -- Dedicated mains supply, 3 phases and neutral. Thus single phase voltage is also available.
A low impedance, permanently installed connection, fed in by the step down transformer of the hospital to supply
large systems like in MR, CT and X--ray departments is required.
Note
Use always a mains cable with 4 wires and concentric PE--shield, type NYCY.
4.2.Climatic conditions
Ambient temperature10_C--40_C...................
Relative humidity15% -- 90%; no condensation.......................
-- External mains transformer PDU
9890 000 0260x is required
(requiresmodificationatthe
EMC-- filter of the kV power unit).
X-ray
installation
L2
L3
Caution!
Ensure the sequence of phases in the wall junction box corresponds to designations L1, L2, L3.
OPTIMUS R/F(d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
-- 400V / 440V / 460V / 480V
1--9
5.4.2. Calculating the mains resistances
Note
The cross section of lead l
must not exceed 25mm2. (See figure below).
3
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
If possible the sum of R
smaller than the R
requires.
XG
0,R1,R2
and R3should be
R
+R1+R2+R3<R
0
With higher internal mains resistances the generator
output is reduced correspondingly.
R
designates the mains resistance of the distributor transformer
0
R
depends on the length of lead l1between distributor transformer and mains distributor
1
L1
L2
L3
l
1
and on the selected cross section as well:
==>R
R
consists of upstream elements such as:
2
1=l1
× R
Cu
RCufrom table below
-- Emergency-OFF switch4.0mΩ.................
-- Earth-leakage circuit breaker5.5mΩ............
l
3
XG
MEX 100
L1
L2
L3
357J94
-- Fuse5.5mΩ.................................
-- Surge arrester WN23.0mΩ...................
R
depends on the length of lead l3between mains distributor and wall junction box
3
and on the selected cross section as well:
==>R
3=l3
× R
Cu
RCufrom table below
The resistances consider the go and return lines so that the calculation can be based on simple cable lengths.
Copper cross section
[mm
2
]
Resistance R
[mΩ/m]
Cu
162.19
251.4
351.0
500.7
700.5
950.38
1200.30
1500.24
Note
500m
Ω
is the absolute max. mains resistance.
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--10
OPTIMUS_RF_1_d021
OPTIMUS R/F
INTRODUCTION AND TECHNICAL DATA
5.4.3. Earth-leakage circuit breaker
To be provided between mains fuse and X-ray installation depending on local regulations.
Siemens earth-leakage circuit breaker:
-- Order No.: 5SZ3 466 OKG00
-- Rated fault current 30mA
-- Rated current 63A
-- Connection terminals for wire cross sections of up to 25mm
2
5.4.4. Emergency--OFF device
To be provided depending on local regulations.
There are 2 possibilities:
1.All the Emergency--OFF buttons are connected in series and looped into the switch--ON circuit (12VDC) of
the generator.
2.The Emergency--OFF circuit acts on an external mains contactor which switches OFF the power before it is
fed into the generator.
OPTIMUS R/F(d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--11
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
6.Tools
-- Service engineer standard tool kit
-- Service-PC:
IBM-compatible, 640kB RAM, 3.5” floppy disk drive, ≥ 1 serial port
-- Installation and service software OMA: 4512 116 022xx / OMB: 4512 1 16 023xx.
Supplied on two floppy disks within this generator service manual.
-- PC-hardkey (DIAGGEN):
Necessary to carry out the installation and to run the service software (special programmings, fault finding).
-- 0-modem cable:
Minimum length is distance between generator cabinet and operating desk.
Male 25-pole D-Sub connector at the generator side.
A 5m data cable of bucky controller can be used: 4512 130 5693x
Caution!
AMPLIMAT cables 9803 507 0xx02 (for hybrid measuring chambers 9803 509 xxxxx) with 3--Plus plugs at
both ends must be connected in the generator by the following adapter for each cable:
Adapter for AMPLIMAT cable: 4512 108 09042. The generator includes 1 adapter.
The hybrid measuring chambers 9803 509 xxxxx require connection (chassis) between contacts:
D--Sub end GND (13) <---- --> RF 0V (8)(generator input)
or
3--Plus end GND (N) <------> RF 0V (J)(generator input)
This connection is established by the adapter for the AMPLIMAT cable.
See drawing Z1--6 ”Basic interface”.
In case a hybrid measuring chamber 9803 509 xxxxx is not operated with the required
make sure to establish this connection (13 <-- -- --> 8) in the D--Sub connector!
For ALC measuring chambers 9890 000 016xx connection GND <------> RF 0V is not permitted.
Therefore, ALC measuring chambers AMPLIMAT cables 9890 000 017xx should always be used.
Operating desk
Note
Use the shortest cables. Noise immunity increases.
cable set10m9890 000 02411.......................
20m9890 000 02421.......................
30m9890 000 02431.......................
8.3.Manpower
At least two persons are necessary to insert the H.V. tank in the generator cabinet.
The weight of the 2--tube version is about 88kg.
OPTIMUS R/F(d/02.1)
OPTIMUS_RF_1_d021
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
1--15
OPTIMUS R/FINTRODUCTION AND TECHNICAL DATA
9.Planned maintenance
The technical documentation for carrying out maintenance work in compliance with the applicable regulations are
available at the responsible authority of Philips Medical Systems.
The importance of having maintenance implemented is pointed out to the operator in the operating instructions.
It must be guaranteed that the person carrying out maintenance work knows about the respective national
regulations and that this person observes these regulations throughout all steps of maintenance work.
E 1998 Philips Medical Systems
ALL RIGHTS RESERVED
OPTIMUS R/F(d/02.1)1--16
OPTIMUS_RF_1_d021
Loading...
+ 306 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.