The non-stress test (NST) timer shows the elapsed time for the non-stress test. The timer counts up to
the time you set for the NST.
Setting NST Autostart/Autostop
You can set the recorder so that it starts automatically (NST Autostart) when the NST timer is started,
and stops automatically (
As default,
NST Autostart is On, and NST Autostop is Off.
NST Autostop) when the NST is complete (when the set run time has elapsed).
Viewing the NST Timer
9
You can configure the timer notification symbol, (the NST label), a progress bar and the elapsed time
to be displayed in the top left-hand corner of the screen. By default, the NST timer is not displayed on
the screen.
Alternatively, you can view the timer in the
To open the
Either
•Press the
Or
•Access the NST pop-up keys (see “Accessing the NST Setup Pop-up Keys” on page 152), and
press the
Timers window:
Timer SmartKey.
Timers key.
Timer Expiry Notification
Timers window.
When the timer expires, the color changes from blue to green, you hear a single tone, and a message
appears in the status line on the main screen.
The volume of the tone can be set in Configuration Mode.
151
9 Non-Stress Test Timer
Accessing the NST Setup Pop-up Keys
You control and set up the NST timer (for example, start, stop, or clear the timer, and set the run time)
using a selection of pop-up keys that you access via any one of three possible routes:
•Via the
•Via the
Timer SmartKey (Route 1)
Main Setup SmartKey (Route 2)
•Via the NST display area at the top left-hand corner of the screen (Route 3). When you touch the
NST display area, the NST pop-up keys become available at the bottom of the screen.
Via the Timer SmartKey (Route 1)
Press the
Timer SmartKey. The Timers window opens, and the pop-up keys for controlling/setting up
the NST timer appear (see “Pop-up Keys for NST Timer Setup” on page 152).
Via the Main Setup SmartKey (Route 2)
1Enter the Main Setup menu using the SmartKey.
2Select NST to enter the Setup NST menu. At the same time, the pop-up keys for setting up the NST
timer appear (see “Pop-up Keys for NST Timer Setup” on page 152).
Via the NST Display Area (Route 3)
Select the NST display area at the top left-hand corner of the screen (when so configured). The pop-up
keys for controlling/setting up the NST timer become available at the bottom of the screen (see “Pop-
up Keys for NST Timer Setup” on page 152).
Pop-up Keys for NST Timer Setup
Pop-Up KeysSelecting this pop-up key lets you:Comments
StartStart the timer.
StopStop the timer, allowing either restarting after a
pause (
Start) or clearing (Clear key).
Setup NSTEnter the Setup NST menu. From here you can set
the run time.
This pop-up key is not available with
Route 2, as the
Setup NST menu is
already open.
TimerReturn to the Timers window.This pop-up key is not available with
Route 1, as the
Timers window is already
open.
Runtime
The run time can be set from 10 to 60 minutes. See the Configuration Guide for details how to set the
run time.
Run Time
The run time can be set from 10 to 60 minutes. See the Configuration Guide for details how to set the
run time.
152
10Non-Stress Test Report
It is generally accepted that a non stress test (NST) allows you to assess fetal well-being. The monitor's
NST report process uses fetal ultrasound (but not DECG) heart rate traces and the Maternal Toco
trace to generate a printed report when criteria are met and it is an indication of the fetal well-being.
The American term Non Stress Test (NST) is used for antepartum testing. The interpretation
algorithm and rule set are equivalent to those implemented in OB TraceVue Revision G.xx or
IntelliSpace Perinatal Revision H.xx and higher, and are based on the 2008 NICHD guidelines.
An NST report is a diagnostic aid, but it does not replace the clinician’s judgment. The interpretation
and the appropriate clinical response remain with the clinician.
A fetus normally produces characteristic heart rate patterns. Average baseline variability and
acceleration of the FHR in response to fetal movement are considered reassuring signs. This test does
not take into account any form of external fetal stimulation.
For every active ultrasound fetal heart rate measurement, one NST report can reside in the monitor’s
memory. The reports are cleared when you discharge a patient and when you start a new NST report.
10
When the NST Report option is available and the
available ultrasound fetal heart rate measurements are displayed on the screen.
The minimum displayed information is:
•NST identification (by FHR number: 1, 2, 3)
•Current NST status (by color: inverse for "not started yet", white for "running", yellow for
"stopped", green for "finished")
Setting Up an NST Report
To set up NST Report functionality:
1Enter the Main Setup menu and select the NST Report.
2Or select the NST Report SmartKey.
3Press the Setup pop-up key.
NST Report feature is "on", the NST status for all
153
10 Non-Stress Test Report
Set your configuration options.
4
Select from:
NST Analysis choose from On or Off.
•
This switches the report feature on or off. This is linked to the NST timer. Both must be set to
for the NST report to function.
Report Recording choose from:
•
–
Manual - press the Record Report pop-up key to trigger a manual request.
After Recorder Stop - report is recorded as soon as recorder becomes idle.
–
Immediately - if a realtime recording is running, the monitor pauses it. The recording is
–
continued after the report has been recorded.
Average short term variability (STV) value is documented in [bpm] and [ms] if STV is configured as
part of the NST Report. This parameter is not considered as reassuring criteria.
NST Report Status Window
The NST Report window displays a detailed overview of the current NST status for any available
ultrasound fetal heart rate measurement. You can see:
•NST Status - whether it is ready, ongoing, or the time and date at which it was stopped, or at which
it was finished.
On
•Elapsed time - the time that has elapsed since the NST began.
•Accelerations - the number of FHR accelerations detected so far.
•Baseline - the average baseline value.
•Variability - the average variability value.
•Short Term Variability - the current short term variability (STV) value.
•Decelerations - the number of FHR decelerations detected so far.
•FHR Availability - current statistical FHR availability value.
•Sinusoidal - the current status of sinusoidal rhythm detection.
For criteria not yet met, a white arrow symbol marks the overall status on the top line, and also appears
against every criterion not yet met. A yellow symbol indicates detection of severe or prolonged
decelerations.
The pop-up keys let you perform the following actions:
FHR1, FHR2, FHR3 - switch to the window showing the current NST status for the fetal heart rate.
•
Record Report - print the NST Report on paper.
•
Record Trace - record the trace episode that belongs to the current report. Depending on device
•
usage, the trace recording might be incomplete.
Setup - open the Setup NST Report window.
•
154
Example NST Report
FieldField Content
10 Non-Stress Test Report
Report Title, with FHR label
NST Report for FHR1 on 12 Oct. 2009
and date
Product InformationProduct DE53102345 G.01.70, OB A.04.24,
Toco DE52401090, FHR1 DE00002345 A.05.26
Patient InformationRogers, Alice
Age: 27
Week 34, Day 5
min
Start time, end time, Elapsed
time, configured runtime
Gestational Age:
Time: 11:34 – 12:06
Elapsed time: 32
Run time: 20 min
Overall one-line NST result
NST Criteria*: not met
summary
TitleTrace Interpretation Summary
Result AccelerationsAccelerations: 2
at: 11:59 12:02
Result: ContractionsContractions: 3
at: 11:57 12:00 12:04
Result: Baseline and
Variability
Baseline: 125 bpm (Range: 118-129 bpm)
Variability: 23 bpm (Range: 20-24 bpm)
Statistics: FHR availabilityFHR available: 95%
Result: DecelerationsDecelerations: 1
Result:
Sinusoidal Rhythm detected
at: 11:58
severe
prolonged
Sinusoidal: No
155
10 Non-Stress Test Report
FieldField Content
Result: Decelerations before
Reporting Period
Guideline/Criteria
Information
Events before Reporting Period:
Decelerations: 1
at: 11:38
severe
prolonged
This field is enabled if there were decelerations between the start of
NST and the start of the reporting period.
(*) Interpretation criteria based on guideline "NICHD 2008, v01"
User-defined criteria for CTG tracing:
•valid FHR for 90% of reporting period
•baseline heart rate between 120 bpm and 160 bpm
•at least 2 accelerations in 10 min
•not more than 1 decelerations
•moderate baseline variability (6-25 bpm)
Additional criteria:
•no severe or prolonged decelerations
•no sinusoidal pattern in reporting period
156
NST Criteria
The patient is monitored for a user-definable period of time (10-60 minutes in steps of 5 minutes). The
test is considered reassuring when the following criteria are met:
•The fetal heart rate is valid at least 90% (this is configurable) of the specified time span.
•The FHR features a user-defined minimum number of accelerations.
•The FHR features a user-defined maximum number of tolerated decelerations, and does not
include severe or prolonged decelerations, which are never tolerated.
•The average baseline fetal heart rate lies within the user-defined limits for low heart rate and high
heart rate over the whole time span.
•The FHR exhibits a moderate variability (user-defined) for the specified time span.
An NST Report is generated when the reassuring criteria are met the first time in the current
monitoring phase. When performing NST with twins or triplets, a separate NST Report is generated
for each fetus.
After the reassurance criteria have been met, the clinician can print the NST Report and then turn the
fetal monitor off, or may continue fetal monitoring and print the report at any time.
10 Non-Stress Test Report
Non-Reassuring Report
If the reassurance criteria are not met when the test has run for 90 minutes, or if you stop anytime
during the 90 minute period, then the test is stopped, and a report is generated stating the reassurance
criteria have not been met.
Nonreactive NST Test
If a nonreactive test occurs, and you then use acoustic stimulation, you must exercise caution in
interpreting the resulting traces, as artificial stimulation is not taken into account when calculating test
results.
157
10 Non-Stress Test Report
158
11Cross-Channel Verification
(CCV)
The cross-channel verification helps to reduce the possibility of misidentification of the maternal heart
rate for the fetal heart rate. It does this by comparing the measured fetal heart rate to the maternal
heart rate. If there are multiple fetal rates, they are also compared with each other and the maternal
heart rate.
Misidentification of Heart Rates
FHR detection by the monitor may not always indicate that the fetus is alive. Confirm fetal life before
monitoring, and continue to confirm that the fetus is the signal source for the recorded fetal heart rate
(see “Confirm Fetal Life Before Using the Monitor” on page 10).
11
To reduce the possibility of mistaking the maternal HR or pulse for FHR, or FHR1 for FHR2 or
FHR3, it is recommended that you monitor both maternal HR/pulse and the heart rates of all fetuses
(see “Monitoring FHR and FMP Using Ultrasound” on page 165, “Monitoring Twin FHRs” on
page 183, “Monitoring Triple FHRs” on page 191, and “Monitoring Maternal Heart / Pulse Rate” on
page 221).
Here are some examples where the maternal HR can be misidentified as the FHR, or one FHR for
another FHR (twins/triplets).
When using an ultrasound transducer:
–It is possible to pick up maternal signal sources, such as the maternal heart, aorta, or other
large vessels. Especially if the recorded maternal HR, and any other artifact is over 100 bpm.
–It is possible to pick up the same fetal heart rate simultaneously with multiple transducers.
NOTE
When an ultrasound transducer is connected to the monitor, but not applied to the patient, the
measurement may generate unexpected intermittent FHR readings.
159
11 Cross-Channel Verification (CCV)
When Fetal Movement Profile (FMP) is enabled:
The FMP annotations on a fetal trace alone may not always indicate that the fetus is alive. For
example, FMP annotations in the absence of fetal life may be a result of:
•Movement of the deceased fetus during or following maternal movement.
•Movement of the deceased fetus during or following manual palpation of fetal position (especially
if the pressure applied is too forceful).
•Movement of the ultrasound transducer.
When using a scalp electrode (DECG):
•Electrical impulses from the maternal heart can be transmitted to the fetal monitor through a
recently deceased fetus via the spiral scalp electrode, appearing to be a fetal signal source.
Cross-Channel Verification Functionality
The cross-channel verification functionality (CCV) of the fetal monitors compares all monitored heart
rates (maternal and fetal), and indicates automatically whether any two channels are picking up the
same signal, or monitoring similar values.
If the fetal monitor detects that any channels have the same or similar values, the
issued with an INOP tone that can have a configurable delay. In addition, yellow question marks
appear next to the numerics on the touchscreen that have the same or similar values. On the recording
trace there is also a question mark from the point where recorded traces continuously overlap.
Visual Aids for CCV Detection
Coincidence INOP appears on the screen of the fetal monitor.
Question mark appears on the screen of the fetal monitor next to the
numerics that show the same or similar values.
Question mark recorded on the trace from the point where two
measured values coincide.
Coincidence INOP is
160
11 Cross-Channel Verification (CCV)
Overview of Cross-Channel Comparisons
Measurements from Transducers
MeasurementTransducer
FHR (US)From Ultrasound or CL Ultrasound transducer
dFHR (DECG) From a fetal scalp electrode
aFHR (abdom. ECG)From the CL Fetal & Maternal Pod
+
MP transducer
FHR1 (US)
Pulse (Toco) From Toco MP, or CL Toco
Pulse (SpO
HR (MECG)From MECG electrodes
aHR (abdom. ECG)From the CL Fetal & Maternal Pod
Measurement Comparison Done by the Fetal Monitor for Cross-Channel Verification
* dFHR and aFHR always replace one of the fetal channels (1,2, or 3) and cannot be compared to the
channel it replaces. If you monitor for example twins with two ultrasound transducers, you see the
numerics FHR1 and FHR2 at the monitor. If you decide to replace the ultrasound transducer for
FHR2 with a fetal scalp electrode, the dFHR numeric is then shown as dFHR2.
161
11 Cross-Channel Verification (CCV)
Coincidence Examples
Coincidence of Maternal Pulse and FHR
When the maternal pulse and FHR are being monitored, and the measured values are very similar or
the same, the coincidence question mark is displayed on the monitor’s screen above both of the
corresponding numerics (in this case maternal pulse and FHR). Often the signal loss or coincidence
happens because the fetal or maternal movement displaced the ultrasound transducer, and a
repositioning of the transducer is necessary.
1Coincidence INOP
2Coincidence question mark
FHR1
above
3Coincidence question mark
above pulse from Toco MP
Pulse Delay
pulse rate traces have an averaging calculation of approximately 10 seconds and an overall delay
SpO
2
of approximately 12 seconds (depending on recorder speed). This differs from a non-averaged beat-tobeat MECG heart rate trace or an ultrasound heart rate trace calculation (having switched to the
maternal HR) with no significant delay. Note that Maternal Pulse from Toco has an averaging of
4 seconds and an overall delay of between 6 and 8 seconds.
162
11 Cross-Channel Verification (CCV)
The coincidence question mark is also printed on the trace paper next to the corresponding FHR and
maternal pulse.
1Printed coincidence
question mark on trace
2Fetal heart rate trace from
Ultrasound
3Maternal pulse trace from
SpO
2
Coincidence of Twins/Triplets FHRs
When both FHR1 and FHR2 are being monitored, and the measured values are very similar or the same,
the coincidence question mark is displayed on the monitor’s screen above both of the corresponding
numerics (in this case
FHR1 and FHR2).
1Coincidence INOP
2Coincidence question
mark above
3Coincidence question
mark above
FHR1
FHR2
163
11 Cross-Channel Verification (CCV)
The coincidence question mark is also printed on the trace paper next to FHR1 and FHR2.
1Printed coincidence question mark on
trace
2FHR1 and FHR2 traces
Recommended Actions for Coincidence INOP
1Confirm fetal life by palpation of fetal movement or auscultation of fetal heart sounds using a
fetoscope, stethoscope, or Pinard stethoscope.
2Manual determination of the maternal pulse and comparison with the fetal heart rate sound signals
from the loudspeaker.
3Reposition the transducer, or ensure that the fetal scalp electrode is placed correctly, until you
receive a clear signal and the monitor is no longer issuing the
4In case of difficulties deriving a stable maternal pulse reading using the Toco MP or CL Toco
transducer, use SpO
measurement from SpO
or the CL SpO2 Pod instead. In case of similar problems with the pulse
2
, use MECG instead. Reasons to switch the method for deriving a
2
maternal pulse or heart rate include: motion artifacts, arrhythmia, and individual differences in
pulse signal quality on the abdominal skin (via Toco MP).
5If you cannot hear the fetal heart sounds, and you cannot confirm fetal movement by palpation,
confirm fetal life using obstetric ultrasonography.
Coincidence INOP.
+
MP
164
12Monitoring FHR and FMP
Using Ultrasound
To monitor a single FHR externally, you use an ultrasound transducer attached to a belt around the
mother's abdomen. The ultrasound transducer directs a low-energy ultrasound beam towards the fetal
heart and detects the reflected signal. Your monitor can also detect fetal movements and print the fetal
movement profile (
25th week of gestation for non-stress testing or routine fetal monitoring.
WARNING
Performing ultrasound imaging or Doppler flow measurements together with ultrasound fetal
monitoring may cause false FHR readings, and the trace recording may deteriorate.
FMP) on the trace. Monitoring using ultrasound is recommended from the
12
Technical Description
Fetal monitors use the ultrasound Doppler method for externally monitoring the fetal heart rate. Using
the Doppler method, the transducer (in transmitter mode) sends sound waves into the body which are
then reflected by different tissues. These reflections (Doppler echoes) are picked up by the transducer
(in listening mode). These Doppler echoes are amplified and sent to the monitor’s speaker through
which the fetal heart signal can be heard. In parallel the Doppler echoes are processed through an
autocorrelation algorithm to determine the fetal heart rate (FHR). The FHR is displayed on the
monitor’s numeric display and on the recorded trace.
Properly representing the fetal heart rate using a device that derives heartbeats from motion is a
formidable task and the limitations of the technology will be discussed shortly. Basic fetal cardiac
physiology may contribute to difficulties in obtaining a reliable ultrasound signal.
A heart rate pattern of a fetus is capable of extraordinary variation, ranging from a stable pattern with
minimal variation while the fetus is “asleep” to robust accelerations of 40-60 bpm above baseline rate
over a few seconds, or exaggerated variability when the fetus is active. Decelerations of the rate
60-80 bpm below baseline may develop even more abruptly than the accelerations. Beat-to-beat
arrhythmias may further exaggerate the amount of “variability” and can be seen at the bottom of
variable decelerations, or in the presence of fetal breathing movements which also tend to lower the
fetal heart rate. The recognition of these normal variations in fetal heart rate patterns will greatly assist
in the separation of genuine fetal information from the artifact.
165
12 Monitoring FHR and FMP Using Ultrasound
Limitations of the Technology
All tissues moving towards or away from the transducer generate Doppler echoes. Therefore, the
resulting signal that is provided to the monitor’s speaker, and for further fetal heart signal processing,
can contain components of the beating fetal heart wall or valves, fetal movements, fetal breathing or
hiccup, maternal movements such as breathing or position changes, and pulsating maternal arteries.
The fetal heart signal processing uses an autocorrelation algorithm to obtain periodic events such as
heart beats. If the signal is erratic such as from a fetal arrhythmia, the ultrasound device may have
trouble tracking the abrupt changes, and may misrepresent the true FHR pattern. Signals such as those
from moving fetal limbs are usually very strong, thereby masking the fetal heart signal. During
prolonged movements where the fetal heart signal is masked, the FHR appears blank on the numeric
display and as a gap on the recorded trace. Fetal position changes, maternal position changes, or
uterine contractions can move the fetal heart partly or fully out of the ultrasound beam resulting in
signal loss, or even picking up Doppler echoes from pulsating maternal arteries. In these cases a
maternal heart rate or sometimes even a rate resulting from the mixture of fetal and maternal signals
may be displayed on the monitor’s numeric display and on the recorded trace.
In contrast to the timely well-defined R-peak of an ECG signal obtained with a fetal scalp electrode,
the ultrasound Doppler signal from a fetal heart consists of multiple components from atria (diastole),
ventricles (systole), valves, and pulsating arteries. These components vary depending on fetal and
transducer position and angle, and are further modulated by factors such as fetal or maternal breathing.
These effects may produce what is called “artifact”. Optimal transducer positioning therefore is key to
minimizing these effects and thereby minimizing artifact.
Misidentification of Maternal HR as FHR
FHR detection by the monitor may not always indicate that the fetus is alive. Confirm fetal life before
monitoring, and continue to confirm that the fetus is the signal source for the recorded heart rate (see
“Confirm Fetal Life Before Using the Monitor” on page 10 and “Cross-Channel Verification (CCV)”
on page 159).
What You Need
•Ultrasound transducer
•Toco MP or CL Toco
•Ultrasound gel
•Transducer belt (and optional butterfly belt clip, if applicable)
+
MP transducer
166
12 Monitoring FHR and FMP Using Ultrasound
Cableless Monitoring - Important Considerations
When using an Avalon CL or Avalon CTS Fetal Transducer system with your monitor, note the
following:
Refer to “Cableless Status Indication” on page 95 for general rules regarding the use of cableless
transducers from an Avalon CL or Avalon CTS Cableless Fetal Transducer system.
CAUTION
Never use ultrasound transducers connected to more than one fetal monitor on the same patient.
•When using an Avalon CL or Avalon CTS you should be aware that FMP is not recommended
when the mother is likely to move, and you should disable Fetal Movement Profile (FMP) on the
fetal monitor (
page 170.
•With the Avalon CL Transducer System, you can monitor twins and triplets with cableless
transducers. The Avalon CTS System does not have this option.
Fetal Movement Off) if the mother is walking. See also “Fetal Movement Profile” on
The wireless symbol appears next to the measurement label, indicating that the
measurement is being made by a cableless transducer.
1FHR1
2Toco p a ramet e r
167
12 Monitoring FHR and FMP Using Ultrasound
WARNING
•During ambulant FHR monitoring, the chance of losing the signal or detecting the maternal heart
rate is higher than during stationary monitoring. The frequency of the patient's walk may be
detected, and mistaken for an FHR signal.
•Check the mother’s pulse periodically during monitoring and compare this with the FHR signal.
Beware of mistaking a “doubled” maternal heart rate for FHR. If a fetus is dead, there is a risk that
the maternal heart rate is monitored and misinterpreted as the fetal heart rate. Therefore, the
simultaneous monitoring of maternal heart rate (preferably, the maternal ECG) and the fetal heart
rate is encouraged.
•Do not interpret maternal movements as fetal movements.
•Artifacts: FMP artifacts are generated during fetal heart rate searching by changing the transducer
position, therefore the fetal monitors enable the FMP only after detecting a valid heart rate signal
for several seconds. FMP is not recommended when the mother is likely to move, and you should
disable Fetal Movement Profile (FMP) at the fetal monitor (
walking.
•Gaps in maternal heart rate detection can occur:
–if the transducer is not correctly positioned.
–due to the pulsation of uterine blood vessels.
–if the fetus moves.
Fetal Movement Off) if the mother is
Preparing to Monitor
Prepare for ultrasound monitoring using the list below. The standard procedures in use in your facility
determine the sequence of actions.
1Determine fetal position.
2Fasten the belt around the patient.
3Switch on the monitor and the recorder.
4Connect the transducer to a free socket. The signal quality indicator for the heart rate initially
displays an invalid signal.
5Apply a thin layer of ultrasound gel to the underside of the transducer.
CAUTION
Using ultrasound gel not approved by Philips may reduce signal quality and may damage the
transducer. This type of damage is not covered by warranty.
1Place the transducer on the abdomen, if possible over the fetal back or below the level of the
umbilicus in a full-term pregnancy of cephalic presentation, or above the level of the umbilicus in
a full-term pregnancy of breech presentation. Work the transducer in a circular motion to ensure
the gel layer makes good contact.
2When the transducer is connected correctly and you receive a good signal, the signal quality
indicator should be filled out. If an inadequate signal is produced, the signal quality indicator will
indicate a poor signal, and no numeric will appear on the screen.
168
12 Monitoring FHR and FMP Using Ultrasound
Adjust the audio volume of the monitor's loudspeaker to a clearly audible level, while moving the
3
transducer over the abdomen. When you have a good signal, secure the transducer in position
below the belt.
WARNING
Periodically compare the mother's pulse with the signal coming from the monitor's loudspeaker to
ensure that you are monitoring fetal heart rate. Do not mistake a doubled or elevated maternal HR for
FHR.
The ultrasound transducer may warm slightly (less than 1°C/1.8°F above ambient temperature) when
applied to the patient. When not applied, the transducer can reach a maximum temperature of 44°C/
112.2°F at an air temperature of 40°C/104°F.
Selecting Fetal Heart Sound
You can listen to the fetal heart sound from one ultrasound transducer at a time. When the fetal heart
sound is selected for an FHR channel, you see the audio source symbol next to the FHR numeric label
for that channel.
1FHR1
2Audio source symbol
To select the audio source for an FHR channel:
1Enter the Setup FHR1 menu for the channel you want to hear (FHR1 used as an example).
2Press Select Audio. It may take a few seconds for the audio source symbol to appear.
169
12 Monitoring FHR and FMP Using Ultrasound
Changing the Fetal Heart Sound Volume
The FHR volume symbol at the top right of the Fetal Heart Sound Volume window gives you an
indication of the current volume. To change the volume:
1Select the volume symbol. The volume scale pops up.
2Select the required volume from the volume scale.
1Fetal Heart Sound Volume
Fetal Movement Profile
The Fetal Movement Profile (FMP) parameter detects fetal movements with an ultrasound transducer
connected to the monitor. Only the fetus monitored on the FHR1 channel is monitored for FMP.
Once you have enabled FMP (see “Switching FMP On and Off” on page 171), it is triggered
automatically whenever:
•You connect an ultrasound transducer.
•A patient is discharged.
When FMP is enabled, the ultrasound transducer detects most fetal body movements. Eye movements
are not detected, and movement of the feet and hands may not be detected. Positioning or
repositioning of the transducer is recorded as fetal movement. Maternal movement, excessive fetal
breathing, or fetal hiccups may also be recorded as fetal movement (also in case of fetal demise or
during the second stage of labor). You can mark these artifacts on the trace paper using either the
remote event marker, or the event marker key as described in “Marking an Event” on page 60. FMP
should be interpreted with care, or disabled when the patient is ambulating or during the second stage
of labor. Ignore these movements when you interpret the FMP. When monitoring twins or triplets,
only the fetus monitored on the FHR1 channel is monitored for movement, but be aware that
movements recorded for FHR1 may also be caused by movement of the second or third fetus.
The fetal movement profile (FMP) appears as "activity blocks" along the top of the Toco scale, the
length of each block showing the duration of the activity.
170
FMP Statistics
FMP statistics are printed every ten minutes.
12 Monitoring FHR and FMP Using Ultrasound
1FMP enabled
2FMP started here
AIndication of current fetal movement
BThe FMP statistics are presented as two percentage figures:
The first figure shows the percentage of detected fetal movements in the previous ten
minutes.
CThe second figure shows the percentage of detected fetal movements since the start of
recording.
To mark the start of the FMP statistic, FMP is printed on the paper with an arrow.
The FMP detection activates after about half a minute of steady heart rate signals (signal indicator halffull, or full) to minimize transducer positioning artifact. You will notice this deliberate delay:
•When a new patient is admitted. A patient discharge restarts the FMP statistics from zero.
•When you connect an ultrasound transducer.
Switching FMP On and Off
You can switch FMP on and off from any FHR channel. For example, to set it from the FHR1
channel:
1Enter the Setup FHR1 menu.
2Select Fetal Movement to switch between On and Off.
3Return to the main screen.
171
12 Monitoring FHR and FMP Using Ultrasound
Troubleshooting
ProblemPossible CausesSolutions
Erratic trace
Erratic display
Signal quality indicator is
continuously poor
Questionable FHRRecording maternal HR by mistakeReposition transducer
FHR not recordedFHR is less than 50 bpm or over 240 bpmIf membranes are ruptured, using a fetal scalp
FHR1 Equip Malf or FHR2 Equip Malf or FHR3 Equip Malf INOP displayed. See “Patient Alarms and INOPs” on
FHR1 Signal Loss or FHR2 Signal Loss or FHR3 Signal Loss INOP
displayed.
FHR1 Unplugged or FHR2 Unplugged or FHR3 Unplugged INOP
displayed.
If you suspect the transducer is malfunctioningTest the transducer.
Fetal arrhythmiaConsider monitoring FHR using DECG after
Obese patient
Transducer position not optimalReposition transducer until signal quality
Belt looseTighten belt
Too much gelRemove excess
Very active fetus-
Insufficient gelUse enough gel to ensure the transducer
Transducer position not optimalReposition transducer until signal quality
FHR less than 50 bpm (and the FHR is
audible)
Recording periodic signals when the
transducer is not applied to the patient
Recorded FHR appears to be suspiciously
higher, or suspiciously lower, than real FHR.
In very rare cases, half- or double-counting
of the FHR can occur.
the rupture of membranes.
indicator shows a good signal (at least
half-full).
makes good contact with the mother's skin.
indicator shows a good signal (at least
half-full).
If membranes are ruptured, using a fetal scalp
electrode (FM30 and FM50 only) allows
measurement of FHR down to 30 bpm.
Confirm fetal life
Disconnect all NON-USED ultrasound
transducers, as continuous, regular
mechanical, or electromagnetic influences can
result in an artificial trace.
If you have reason to question the validity of
the recorded FHR, always verify FHR by
independent means (by auscultation, for
example). Measure maternal pulse by
independent means.
electrode (FM30 and FM50 only) allows
measurement of FHR down to 30 bpm.
If FHR is outside of the specified range, verify
FHR by independent means.
page 129.
172
Testing Ultrasound Transducers
If any of the following tests fail, repeat the test using another transducer. If the second transducer
passes the tests, confirming that the first transducer is defective, contact your service personnel.
If the second transducer also fails the tests, contact your service personnel.
12 Monitoring FHR and FMP Using Ultrasound
To test an ultrasound transducer:
1Switch on the monitor and the
recorder.
2Connect the transducer to the fetal
monitor.
3Select the fetal heart sound for this
channel.
4Increase the loudspeaker volume
to an audible level.
5Holding the transducer in one
hand, move your other hand
repeatedly towards and then away
from the surface.
You can test all ultrasound transducers, including the cableless ones, as described above.
Additional Information
Artifact in Fetal Heart Rate Measurement
How to detect it and reduce its occurrence using the Avalon Fetal Monitor
The ultrasound derived FHR measurement technique in Avalon fetal monitors, like all other
ultrasound fetal monitors’ FHR measurement techniques, has limitations that can lead to
misrepresentation of the fetal heart rate pattern and potential misinterpretation of the fetal condition.
An incorrect interpretation of the trace may lead to either unnecessary interventions, or to failure to
detect fetal distress, and the need for intervention. Thus, the on-going evaluation of the recorded trace
requires regular confirmation that the trace represents the true FHR. Specific situations requiring such
confirmation include the following:
6Check that a noise is heard from
the loudspeaker.
•After starting a measurement or changing a transducer
•After maternal position changes, for example during pushing with contractions
•When the tracing shows abrupt changes in baseline rate, variability, or pattern (decelerations to
accelerations) especially in the second stage of labor
•When the baseline maternal heart rate is within about 15 bpm of the FHR
173
12 Monitoring FHR and FMP Using Ultrasound
•When you are unable to determine a baseline rate, and variability occurs between consecutive
contractions
There are several ways to verify the source and/or accuracy of the recorded fetal heart rate pattern.
These include:
Verification of the FHR with:
•An obstetric stethoscope
•Ultrasound imaging
•A fetal scalp electrode
Verification of the maternal heart rate:
•Using pulse oximetry - for a maternal heart rate pattern displayed simultaneously with the FHR
(Cross-Channel Verification (CCV) feature)
•Using Maternal ECG - for a maternal heart rate pattern displayed simultaneously with the FHR
(CCV feature)
•Manual determination of the maternal pulse
Whenever possible measure the maternal pulse rate to make use of the monitor’s Cross-Channel
Verification (CCV) feature, especially during the second stage of labor, or when the maternal pulse is
elevated over 100 bpm. The Avalon fetal monitor provides a Toco MP or CL Toco
for maternal pulse detection and the creation of a maternal heart rate pattern plotted on the same
recorder as the FHR pattern. In case of difficulties deriving a stable maternal pulse reading using the
Toco MP or CL Toco+ MP transducer, use SpO2 or MECG instead.
+
MP transducer
When either of these parameters is utilized, the monitor will automatically and continuously perform a
CCV of the maternal heart rate pattern against the FHR pattern displayed on the monitor. If the
patterns and rates are similar, the CCV provides an alarm that both rates are probably from the same
source (i.e., they both represent the maternal heart rate pattern and the fetus is not being monitored).
Repositioning the ultrasound transducer will usually correct this, but it may be necessary to apply a
fetal scalp electrode. Advising the mother to temporarily cease pushing during contractions may help
to more rapidly resolve any uncertainty in this situation.
Doubling: The autocorrelation algorithm can display a doubled fetal or maternal heart rate if the
duration of diastole and systole are similar to each other, and if the heart rate is below 120 bpm.
Doubling, usually brief, is accompanied by an abrupt switch of the trace to double the baseline value.
Halving: With fetal tachycardia (above 180 bpm) and some interference from breathing or maternal
arteries the autocorrelation algorithm may only recognize every second beat resulting in a halved rate
for a limited time. If the actual FHR is above the maximum limit of the monitor (240 bpm), the
algorithm will also half-count. Halving is accompanied by an abrupt switch of the trace to exactly half
the prior baseline value. This switch may simulate an FHR deceleration and be referred to by clinicians
as a “false deceleration.”
Switching to maternal heart rate (also referred to as "Maternal Insertion"): The fetal heart can
move partly or fully out of the ultrasound beam and the autocorrelation algorithm may then pick up
and display the maternal heart rate. Depending on the signal mix in the ultrasound signal, switching to
the maternal heart rate may mimic several conditions with the potential for erroneous interpretation
and response as follows:
•The switch to the maternal heart rate may simulate an FHR deceleration (i.e., a decrease of
the fetal heart rate, and be referred to by clinicians as a “false deceleration”).
174
12 Monitoring FHR and FMP Using Ultrasound
•The maternal heart rate may simulate a normal fetal heart rate pattern (i.e., it may mask an
FHR deceleration or fetal demise).
Especially during pushing with contractions in the second stage of labor, the maternal heart rate
may increase to the point where it may equal or exceed the fetal rate. Here the maternal trace may
mimic a normal fetal trace while the fetus may be having decelerations or fetal demise has
occurred. This change from fetal to maternal heart rate pattern may not be at all obvious unless
CCV is used and represents the most dangerous pitfall of all the artifacts because fetal distress may
go unrecognized.
•The maternal heart rate may simulate an FHR acceleration, which is an increase of the fetal
heart rate.
During expulsive efforts, the maternal heart rate normally accelerates and may be at or above the
normal FHR range.
•The FHR may display gradual appearing decelerations. Generally, the “false decelerations”
described above are abrupt. Rarely, combinations of “noisy/erratic signal” associated with changes
in maternal and/or fetal rate or movement will produce more gradual appearing “false
decelerations” but these are usually short-lived with an abrupt return to an obviously stable FHR
baseline.
“Noisy/Erratic” signals: With mixed or weak signals the tracing may reveal very brief episodes of
erratic recorded traces. These represent the autocorrelation algorithm finding brief sequences of
apparent and persistent heartbeats amidst a mixed or weak signal. These erratic recorded traces are
commonplace, especially in association with fetal or maternal movement. During prolonged periods
of such noisy/erratic signals, the fetus is not being adequately monitored.
Drop out: With mixed or weak signals there may be no heart rate tracing at all. These episodes reflect
that if the algorithm does not find an apparent and persistent heartbeat amidst a mixed or weak signal,
it will not print a heart rate on the tracing. Brief episodes of drop out are commonplace, especially in
association with fetal or maternal movement. During prolonged periods of drop out, the fetus is
not being adequately monitored.
Multiple Fetuses
With multiple fetuses, the potential to experience these artifacts is increased. Positioning of the
transducer is even more critical. Ultrasound scanning should be used to help with positioning of
individual transducers. See also “Monitoring Twin FHRs” on page 183 and “Monitoring Triple FHRs”
on page 191.
175
12 Monitoring FHR and FMP Using Ultrasound
Obtaining a Good Heart Signal
To successfully position the ultrasound transducer, first determine the fetal position using palpation.
Position the transducer over the strongest audible fetal heart sound from the monitor’s speaker and
wait at least six seconds after each transducer adjustment to verify a good signal quality displayed
on the Signal Quality Indicator and a consistent FHR numeric display. Having determined the position
that provides a strong fetal signal, fix the transducer on the abdomen with the belt.
If the quality of the signal or the appearance of the heart rate trace from the ultrasound transducer is
questionable, the transducer should be repositioned as described above. Alternatively, the use of an
ultrasound scanner will greatly facilitate the determination of the optimal site for the ultrasound heart
rate transducer. Factors during the second stage of labor that may influence the quality of the FHR
tracing obtained with ultrasound include:
•Uterine contractions
•Changing contour of the maternal abdomen
•Maternal body movement - positioning
•Maternal expulsive efforts - pushing
•Maternal tachycardia/accelerations with contractions
•Fetal decelerations, Fetal tachycardia
•Delayed return of the fetal heart rate from a deceleration
•Descent of the fetus in the birth canal
•Rotation of the fetus in the birth canal
In some cases during the second stage of labor, a good and reliable ultrasound FHR signal may not be
obtainable, and the use of a fetal scalp electrode must be considered (fetal ECG).
Heart Rate Sound
The heart rate sound emitted by the device is a representation of movement that, in most cases,
permits accurate auscultation of the FHR corresponding to the FHR displayed on the monitor and rate
pattern depicted on the trace recording. On occasion, the user may hear an FHR that differs from the
FHR display and the recorded trace. This may occur in situations where the fetal heart moves partly
out of the transducer ultrasound beam. In these cases, the user may hear the FHR emitted from the
monitor’s speaker, even though another periodic signal (usually the maternal heart rate) has become
stronger. The autocorrelation algorithm will display the stronger maternal heart rate, despite the
persistence of a weaker fetal signal. These occurrences are usually very brief and, if persistent, can be
addressed by repositioning the transducer.
176
Signal Quality Indicator
The signal quality on the Avalon fetal monitor is indicated by a triangle on the touchscreen that is
displayed in one of three ways:
12 Monitoring FHR and FMP Using Ultrasound
1Completely filled triangle, indicating good signal
quality (good/full).
2Half-filled triangle, indicating limited signal quality.
This condition may indicate a weak or ambiguous
signal. If this status persists, reposition the
transducer (acceptable/medium).
3Empty triangle, indicating insufficient signal
quality. No FHR is displayed on the monitor’s
numeric display or the recorded trace. If this status
persists, reposition the transducer (poor/no signal).
Examples of Artifacts
When monitoring the maternal ECG, a beat-to-beat maternal heart rate trace is printed alongside the
FHR recorded trace. When monitoring the maternal SpO
heart rate trace is printed.
Following are recorded trace examples of complaints received regarding inaccurate output from the
Avalon monitors. Scaling is 3 cm/min and 30 bpm/cm.
Double-Counting
Baseline Rate120
Baseline VariabilityModerate
AccelerationsPresent
DecelerationsNot apparent
derived pulse rate, a filtered and averaged
2
177
12 Monitoring FHR and FMP Using Ultrasound
Double-Counting
ContractionsExcessive, coupling, hypertonus
ArtifactDouble-Counting
CommentReassuring tracing. The excessive uterine activity should prompt discontinuation of any
oxytocic agent.
RemediationThe true fetal rate can be confirmed by auscultation or by fetal scalp electrode.
Half-Counting
Baseline Rate120
Baseline VariabilityModerate
AccelerationsPresent
DecelerationsNot apparent
ContractionsMinimal
ArtifactHalf-counting, noise, drop out
CommentReassuring tracing. The half-count at 4-5 minutes into the tracing may simulate a fetal
deceleration, but the abruptness and the lack of any compensatory changes when the
normal rate returns suggests that this is half-counting. Insertion of the maternal heart rate
(see below) may produce a similar pattern. Note also very brief episodes of half-counting,
maternal insertion, and signal dropout.
RemediationAuscultation or the application of a direct scalp electrode, if feasible, will reveal the true
fetal heart rate.
178
12 Monitoring FHR and FMP Using Ultrasound
Maternal-Switching (Maternal Insertion)
Baseline Rate170 - Tachycardia
Baseline VariabilityModerate
AccelerationsUnable to determine
DecelerationsAbsent
ContractionsAbsent
ArtifactMaternal insertion, noise
CommentThe fetus has an elevated baseline rate of about 170 bpm with minimal to moderate
variability. The ability to assess fetal status is limited because about half of the tracing
displays the maternal heart rate.
RemediationThe application of a maternal transducer (ECG or pulse oximeter) will likely resolve any
possible confusion with the tracing. Repositioning the transducer may produce a more
reliable tracing. Consideration must also be given to applying a fetal scalp electrode.
Noisy/Erratic Signal and Dropout
Baseline Rate140
Baseline VariabilityModerate
AccelerationsPresent
DecelerationsAbsent
ContractionsMinimal
ArtifactNoisy signal, drop-out
179
12 Monitoring FHR and FMP Using Ultrasound
Noisy/Erratic Signal and Dropout
CommentReassuring tracing. Note that there is episodic drop out of the signal with discontinuity of the
fetal tracing.
RemediationEither improving the position of the transducer or the application of a fetal scalp electrode
will reduce the amount of artifact in the tracing.
Antepartal and Intrapartal Fetal Monitoring, 3rd Edition (2007)
By Michelle L. Murray, PhD, RNC
Springer Publishing Company, ISBN 0-8261-3262-6
Page 2, Table 2: Limitations of Continuous EFM
Item 15: “The US may detect maternal aortic wall movement and the maternal HR will be printed. A
failure to recognize the lack of an FHR may delay appropriate management.”
Page 38, “Solving Equipment Problems”, Table 3: The Ultrasound Transducer
JOGC (Journal of Obstetrics and Gynecology Canada)
Volume 29, Number 9, September 2007
Chapter 2: Intrapartum Surveillance
Page S35: “Methods of Electronic Fetal Monitoring”
“… Among its disadvantages are the need for readjustment with maternal or fetal movements and the
following: the transducer may record the maternal pulse, it may be difficult to obtain a clear tracing in
obese women or those with polyhydramnios, artifact may be recorded, and there may be doubling or
halving of the fetal heart rate when it is outside of the normal range.”