Who this Book is For9
Confirm Fetal Life Before Using the Monitor10
Introducing the Avalon Family of Fetal Monitors11
2 What's New29
What's New in Release J.329
What's New in Release G.032
3 Basic Operation35
Supported Measurements35
Avalon FM20 and FM3036
Avalon FM40 and FM5039
Avalon CL Transducer System40
Getting to Know Your Avalon FM20/FM3040
Getting to Know Your Avalon FM40/FM5043
Transducers45
Getting to Know Your Avalon CL47
Cableless Transducers51
CL Pods55
Operating and Navigating56
Operating Modes63
Automatic Screen Layouts64
Settings64
Preparing to Monitor67
Switching the Monitor to Standby98
After Monitoring98
Disconnecting from Power99
Power On/Power Off Behavior99
Monitoring After a Power Failure99
Troubleshooting100
4 FM20/30 Battery Option101
External Power Supply M8023A102
Using Batteries102
Optimizing Battery Performance105
5 Alarms109
Alarm Mode110
Nurse Call Systems110
3
Visual Alarm Indicators110
Audible Alarm Indicators111
Acknowledging Alarms114
Acknowledging Disconnect INOPs114
Alarm Reminder114
Pausing or Switching Off Alarms115
Alarm Limits117
Reviewing Alarms118
Latching Alarms118
Testing Alarms120
Alarm Behavior at Power On121
Important Considerations183
Monitoring Triplets184
Separating FHR Traces184
"Standard" Separation Order184
"Classic" Separation Order185
Switching Trace Separation On and Off186
When Trace Separation is On187
When Trace Separation is Off188
Troubleshooting188
14 Fetal Heart Rate Alarms191
Changing Alarm Settings191
Changing Signal Loss Delay192
15 Monitoring Uterine Activity Externally193
What You Need193
External Toco Monitoring194
Toco Sensitivity195
Troubleshooting195
Testing Toco Transducers195
16 Monitoring Uterine Activity Internally197
What You Need197
Internal (IUP) Monitoring199
Troubleshooting199
17 Monitoring FHR Using DECG201
Misidentification of MHR as FHR201
What You Need203
5
Making Connections205
Monitoring DECG206
Suppressing Artifacts208
Troubleshooting209
Testing DECG Mode209
18 Monitoring Noninvasive Blood Pressure211
Introducing the Oscillometric Noninvasive Blood Pressure Measurement211
Preparing to Measure Noninvasive Blood Pressure212
Starting and Stopping Measurements214
Enabling Automatic Mode and Setting Repetition Time215
Enabling Sequence Mode and Setting Up The Sequence215
Choosing the Alarm Source216
Assisting Venous Puncture216
Calibrating NBP216
Troubleshooting217
19 Monitoring Maternal Temperature219
Measuring Tympanic Temperature219
20 Monitoring SpO2227
Selecting an SpO2 Sensor227
Applying the Sensor227
Connecting SpO2 Cables228
Measuring SpO2228
SpO2 Signal Quality Indicator (Fast SpO2 only)229
Assessing a Suspicious SpO2 Reading230
Understanding SpO2 Alarms230
Setting Up Tone Modulation231
Setting the QRS Volume231
21 Monitoring Maternal Heart / Pulse Rate233
Priority for Maternal Heart / Pulse Rate233
Misidentification of MHR for FHR234
MHR from MECG Electrodes235
Monitoring MECG Wave238
Pulse Rate from Toco MP240
Pulse Rate from SpO2241
Adjusting the Heart Rate / Pulse Alarm Limits241
Average Pulse Rate from Noninvasive Blood Pressure242
Testing MECG Mode242
6
22 Printing the ECG Waveform243
23 Paper Save Mode for Maternal Measurements249
24 Recovering Data251
Recovering Traces on Paper251
Recovering Traces on an OB TraceVue/IntelliSpace Perinatal System252
Recording Stored Data252
25 Care and Cleaning255
General Points255
Cleaning and Disinfecting256
Cleaning and Disinfecting Monitoring Accessories257
Cleaning and Disinfecting the Tympanic Temperature Accessories258
Sterilizing258
26 Maintenance259
Inspecting the Equipment and Accessories259
Inspecting the Cables and Cords259
Maintenance Task and Test Schedule260
Recorder Maintenance260
Cleaning the Print Head265
Disposing of the Monitor266
27 Accessories and Supplies267
Information on Latex267
Avalon CL Base Station268
Transducers268
Fetal Accessories268
MECG Accessories270
Noninvasive Blood Pressure Accessories270
SpO2 Accessories273
Tympanic Temperature Accessories277
Recorder Paper277
Batteries277
Monitoring After a Loss of Power307
ESU, MRI and Defibrillation307
Cardiac Pacemakers and Electrical Stimulators307
Fast Transients/Bursts307
Symbols on the System308
29 Default Settings Appendix311
Alarm and Measurement Default Settings311
Recorder Default Settings314
Index317
8
1Introduction
Who this Book is For
This book is for trained healthcare professionals using the Avalon FM20, FM30, FM40 and FM50
Fetal/Maternal monitors, and the cableless monitoring system Avalon CL. It describes how to set up
and use the monitor and transducers. Familiarize yourself with all instructions including warnings and
cautions before starting to monitor patients. Read and keep the Instructions for Use that come with
any accessories, as these contain important information about application and care and cleaning that is
not repeated in this book.
You should be:
•Trained in the use of fetal heart rate (FHR) monitors.
•Trained in the interpretation of FHR traces.
1
•Familiar with using medical devices and with standard fetal monitoring procedures.
For information on how to configure and service the monitor, see the Configuration Guide and the
Service Guides, or contact your authorized service provider.
Your monitor may not have all of the features and options described in this guide. The exact
appearance of the monitor may differ slightly from that shown in the illustrations.
In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
•A caution alerts you to where special care is necessary for the safe and effective use of the
product. Failure to observe a caution may result in:
– minor or moderate personal injury,
–damage to the product or other property,
–possibly in a remote risk of more serious injury.
•Monitor refers to the entire fetal/maternal monitor. Display refers to the physical display unit.
Screen refers to everything you see on the monitor's display, such as measurements, alarms,
patient data, and so forth.
FM30 •Whenever a monitor’s identifier appears to the left of a heading or paragraph, it means that the
information applies to that monitor only. Where the information applies to all models, no
distinction is made.
9
1Introduction
Avalon CTS
Avalon CL
Whenever one of these identifiers appear to the left of a heading or paragraph, it means that the
information applies to that cableless monitoring system. Where the information applies to both
systems, no distinction is made.
Confirm Fetal Life Before Using the Monitor
Fetal monitoring technology available today is not always able to differentiate a fetal heart rate (FHR)
signal source from a maternal heart rate (MHR) source in all situations. Therefore, you should confirm
fetal life by independent means before starting to use the fetal monitor, for example, by palpation of fetal
movement or auscultation of fetal heart sounds using a fetoscope, stethoscope, or Pinard stethoscope.
If you cannot hear the fetal heart sounds, and you cannot confirm fetal movement by palpation,
confirm fetal life using obstetric ultrasonography. Continue to confirm that the fetus is the signal
source for the FHR during monitoring.
Be aware that:
•an MHR trace can exhibit features that are very similar to those of an FHR trace, even including
accelerations and decelerations. Do not rely solely on trace pattern features to identify a fetal
source.
•Fetal Movement Profile (FMP) annotations on a fetal trace alone may not always indicate that the
fetus is alive. The body of a deceased fetus can move and cause the monitor to annotate fetal body
movements.
Here are some examples where the MHR can be misidentified as the FHR.
•When using an ultrasound transducer:
–It is possible to pick up maternal signal sources, such as the aorta or other large vessels.
–Misidentification may occur when the MHR is higher than normal (especially when it is over
100 bpm).
•When using a fetal scalp electrode:
–Electrical impulses from the maternal heart can sometimes be transmitted to the fetal monitor
through a recently deceased fetus via the spiral scalp electrode cable, appearing to be a fetal
signal source.
–The recorded MHR (and any artifact) can be misinterpreted as an FHR (especially when it is
over 100 bpm).
•When Fetal Movement Profile (FMP) is enabled:
FMP annotations in the absence of fetal life may be a result of:
–Movement of the deceased fetus during or following maternal movement.
–Movement of the deceased fetus during or following manual palpation of fetal movement
(especially if the pressure applied is too forceful).
–Movement of the ultrasound transducer.
–The ultrasound transducer detecting a maternal movement source, such as the mother
coughing.
See also “Monitoring FHR and FMP Using Ultrasound” on page 155 and “Monitoring FHR Using
DECG” on page 195.
To reduce the possibility of mistaking the MHR for an FHR, it is recommended that you monitor both
maternal and fetal heart rates. The monitor's cross-channel verification (CCV) facility can help by
automatically detecting when an MHR coincides with an FHR. For further details, see “Cross-Channel
Verification (CCV)” on page 149.
10
1 Introduction
Introducing the Avalon Family of Fetal Monitors
The Avalon family of fetal monitors consists of the Avalon FM20, FM30, FM40 and FM50. While the
FM20/FM30 and the FM40/FM50 have different form factors, the method of operation is very
similar for all monitors. The Avalon fetal monitors also share transducers, accessories, software, and
are compatible with the Avalon CL, and Avalon CTS Cableless Fetal Transducer Systems.
Intended Use
The Philips Avalon FM20 (M2702A), FM30 (M2703A), FM40 (M2704A) and FM50 (M2705A) Fetal/
Maternal Monitors are intended for:
•non-invasive monitoring of fetal heart rates and movements.
•invasive monitoring of fetal Direct ECG and intrauterine pressure and for displaying and
recording of fetal and maternal ECG waves. (FM30 and FM50 only).
•displaying, storing, and recording patient data and parameter values, and for generating alarms
from fetal and maternal parameters.
•transmitting patient data and parameter values to a patient information and surveillance system
used by trained health care professionals.
•use in antepartum testing areas, in labor and delivery rooms, and during postpartum recovery in
the hospital environment. They are not intended for use in intensive care units or operating rooms.
+
•monitoring in a bath or shower. (Avalon CL cableless transducers Toco
ECG/IUP only)
•transport situations in healthcare facilities, for healthcare facilities outside hospitals, such as
doctors’ offices, and for use in private households. (FM20 and FM30 only)
WARNING
The fetal/maternal monitors are not intended for:
•use during defibrillation, electro-surgery, or magnetic resonance imaging (MRI).
•ECG measurements on patients connected to electrical stimulator or with cardiac pacemakers.
•use of the invasive measurements IUP and fetal DECG, use of the patient module (M2738A) and
use of the Avalon CL system in domestic establishments and those connected directly to the
public low-voltage supply network that supplies buildings used for domestic purposes.
WARNING
No modification of the fetal monitors, transducers, and the Avalon CL base station is allowed.
MP, Ultrasound, and
CAUTION
US federal law restricts this device to sale by, or on the order of, a physician.
11
1Introduction
Connecting the Monitor to AC Mains
WARNING
•Always use the supplied power cord with the earthed mains plug to connect to an earthed AC
mains socket. Never adapt the mains plug from the fetal monitor to fit an unearthed AC mains
socket.
•Check that the line frequency is correctly configured in the
•FM20/FM30 only: The protective earth conductor is required for EMC purposes. It has no
protective function against electric shock. Double and/or reinforced insulation protects this
device against electric shock.
•Do not use AC mains extension cords or multiple portable socket-outlets.
Indications for Use
Avalon Fetal/Maternal Monitor FM20
Indicated for use by trained health care professionals whenever there is a need for monitoring of the
physiological parameters uterine activity, heart rate, oxygen saturation, noninvasive blood pressure,
pulse rate, and temperature of pregnant women and the fetal heart rates of single fetuses, twins, and
triplets in labor and delivery rooms, in antepartum testing areas, in private households and during
transports in healthcare facilities.
Avalon Fetal/Maternal Monitor FM30
Indicated for use by trained health care professionals whenever there is a need for monitoring of the
physiological parameters uterine activity, heart rate, ECG, oxygen saturation, noninvasive blood
pressure, and pulse rate and temperature of pregnant women and the fetal heart rates of single fetuses,
twins, and triplets in labor and delivery rooms, in antepartum testing areas, in private households and
during transports in healthcare facilities.
Global Settings menu.
Avalon Fetal/Maternal Monitor FM40
Indicated for use by trained health care professionals whenever there is a need for monitoring of the
physiological parameters uterine activity, heart rate, oxygen saturation, noninvasive blood pressure,
and pulse rate and temperature of pregnant women and the fetal heart rates of single fetuses, twins,
and triplets in labor and delivery rooms and in antepartum testing areas.
Avalon Fetal/Maternal Monitor FM50
Indicated for use by trained health care professionals whenever there is a need for monitoring of the
physiological parameters uterine activity, heart rate, ECG, oxygen saturation, noninvasive blood
pressure, and pulse rate and temperature of pregnant women and the fetal heart rates of single fetuses,
twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Safety Information
In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to
observe a warning may result in death or serious injury to the user or patient.
•A caution alerts you to where special care is necessary for the safe and effective use of the product.
Failure to observe a caution may result in minor or moderate personal injury or damage to the
product or other property, and possibly in a remote risk of more serious injury.
12
1 Introduction
WARNING
The fetal/maternal monitors are not intended for:
•use during defibrillation, electro-surgery, or magnetic resonance imaging (MRI).
•ECG measurements on patients connected to electrical stimulator or with cardiac pacemakers.
•use of the invasive measurements IUP and fetal DECG, use of the patient module (M2738A) and
use of the Avalon CL system in domestic establishments and those connected directly to the
public low-voltage supply network that supplies buildings used for domestic purposes.
WARNING
No modification of the fetal monitors, transducers, and the Avalon CL base station is allowed.
CAUTION
US federal law restricts this device to sale by, or on the order of, a physician.
WARNING
•Always use the supplied power cord with the earthed mains plug to connect to an earthed AC
mains socket. Never adapt the mains plug from the fetal monitor to fit an unearthed AC mains
socket.
•Check that the line frequency is correctly configured in the Global Settings menu.
•FM20/FM30 only: The protective earth conductor is required for EMC purposes. It has no
protective function against electric shock. Double and/or reinforced insulation protects this
device against electric shock.
•Do not use AC mains extension cords or multiple portable socket-outlets.
WARNING
•To avoid magnetic interference affecting the mode of the pacemaker, ensure that the Avalon CL
base station does not come into close contact with implanted pacemakers.
•This equipment generates, uses, and radiates radio-frequency energy, and if it is not installed and
used in accordance with its accompanying documentation, may cause interference to radio
communications. Operation of this equipment in a residential area may cause interference, in
which case the users must take whatever measures may be required to correct the interference.
13
1Introduction
WARNING
Do not change the date and time setting, if the fetal monitor is connected to a Philips OB TraceVue/
IntelliSpace Perinatal system via a LAN -setup. The monitor uses the OB TraceVue/IntelliSpace
Perinatal system date and time, including daylight saving time changes. As long as the fetal monitor is
connected to the OB TraceVue/IntelliSpace Perinatal system via the LAN-setup (locomotive icon
displayed on the monitor's screen), the option to change the date and time settings at the fetal monitor
are disabled, this is not valid for RS232 connections, or the connection to other systems.
WARNING
Short range radio connections are subject to interruption due to interference from other radio sources
in the vicinity, including microwaves, bluetooth devices, WLAN devices (802.11b,g,n) and cordless
phones. Depending on the strength and duration of the interference, the interruption may occur for an
extended period. A loss of connection, due to moving out-of-range, interference, or for other reasons,
is indicated with a
SpO
channel configuration is important, refer to the Configuration Guide for details.
No Host Monitoring INOP (here the host is the fetal monitor) on the NBP or
Pods, or a cl NBP Disconnect or cl SpO₂ Disconnect INOP at the fetal monitor. Correct
2
CAUTION
Check the fetal monitors housing for damage before you start to monitor as part of your safety
precautions.
WARNING
When connecting devices for acquiring measurements, always position cables and NBP tubing
carefully to avoid entanglement or potential strangulation.
CAUTION
Using recorder paper that is not approved by Philips can result in accelerated paper fading and can
damage the thermal line printhead. This type of damage is not covered by warranty.
WARNING
Never immerse the base station in liquid. You must protect it against water sprays or splashes. Place
the base station where there is no chance of contact with, or falling into water or other liquids.
14
1 Introduction
CAUTION
Avoid the use of pulsating water jets in the bath or shower while monitoring, as these can be
misinterpreted as an incorrect (or totally artificial) heart rate.
Toco Baseline drift: The accuracy specified for baseline drift cannot be guaranteed for underwater
usage. When using transducers under warm water the temperature increase causes a significant baseline
change due to internal pressure increase. The depth under water at which the Toco transducer is used
also has an effect on the Toco baseline, as the water pressure increases with depth. After immersion,
allow one to two minutes for the pressure to stabilize, then adjust the Toco baseline (between
contractions), and check it frequently.
When using the transducers underwater, the radio transmission range is reduced, and signal loss may
occur.
WARNING
•Always use the supplied power cord with the earthed mains plug to connect the external power
supply M8023A (option #E25) to an earthed AC mains socket. Never adapt the mains plug from
the power supply to fit an unearthed AC mains socket.
•Do not use AC mains extension cords or multiple portable socket-outlets. If a multiple portable
socket-outlet without an approved isolation transformer is used, the interruption of its protective
earthing may result in enclosure leakage currents equal to the sum of the individual earth leakage
currents, so exceeding allowable limits.
•Do not connect any devices that are not supported as part of a system.
•Any non-medical device placed and operated in the patient’s vicinity must be powered via an
approved isolation transformer that ensures mechanical fixing of the power cords and covering of
any unused power outlets.
WARNING
Do not touch the charging contacts for the cableless transducers at the Avalon base station while you
are touching the patient.)
CAUTION
Condition the battery with a monitor not currently in use. The monitor switches off automatically
when there is no battery power left.
15
1Introduction
WARNING
Use only Philips batteries part number M4605A. Use of a different battery may present a risk of fire
or explosion.
Do not open batteries, or dispose of them in fire, or cause them to short circuit. They may ignite,
explode, leak or heat up, causing personal injury.
If battery leakage should occur, avoid contact with skin. Refer to qualified and authorized service
personnel.
Dispose of used batteries promptly and in an environmentally-responsible manner. Do not dispose
of the battery in normal waste containers. Consult your hospital administrator to find out about local
arrangements. Do not expose batteries to liquids.
Do not crush, drop or puncture batteries - mechanical abuse can lead to internal damage and internal
short circuits which may not be visible externally.
If a battery has been dropped or banged against a hard surface, whether damage is visible externally or
not:
•discontinue use
•dispose of the battery in accordance with the disposal instructions above.
Keep batteries out of the reach of children.
CAUTION
Do not disassemble, heat above 100°C (212°F) or incinerate the batteries, to avoid the risk of fire and
burns. Keep batteries out of the reach of children and in their original package until you are ready to
use them.
If battery leakage should occur, use caution in removing the battery. Avoid contact with skin. Refer to
qualified and authorized service personnel.
Do not install or use pre-damaged batteries.
WARNING
Alarm systems of the monitor and those of the connected OB system are independent and not
synchronized.
WARNING
In INOP only mode, no fetal/maternal patient alarms are enabled or indicated.
16
1 Introduction
WARNING
Do not rely exclusively on the audible alarm system for fetal monitoring. Adjustment of alarm volume
to a low level or off during monitoring may result in a dangerous situation. Remember that the most
reliable method of fetal monitoring combines close personal surveillance with correct operation of
monitoring equipment.
WARNING
Be aware that the monitors in your care area may each have different alarm settings, to suit different
scenarios. Always check that the alarm settings are appropriate before you start monitoring.
CAUTION
In order to ensure that the settings are reset to user defaults for a new patient, always discharge the
previous patient from the fetal monitor.
WARNING
Performing ultrasound imaging or Doppler flow measurements together with ultrasound fetal
monitoring may cause false FHR readings, and the trace recording may deteriorate.
CAUTION
Never use ultrasound transducers connected to more than one fetal monitor on the same patient.
17
1Introduction
WARNING
•Ensure that the conductive parts of the fetal scalp electrode and the maternal leg plate electrode do
•Indication of the heart-rate may be adversely affected by the operation of cardiac pacemaker pulses
•During ambulant FHR monitoring, the chance of losing the signal or detecting the maternal heart
•Check the mother’s pulse periodically during monitoring and compare this with the FHR signal.
•Do not interpret maternal movements as fetal movements.
•Artifacts: FMP artifacts are generated during fetal heart rate searching by changing the transducer
not contact other conductive parts, including earth.
or by cardiac arrhythmias.
rate is higher than during stationary monitoring. The frequency of the patient's walk may be
detected, and mistaken for a FHR signal.
Beware of mistaking a “doubled” maternal heart rate for FHR. In the case of a dead fetus, there is
a risk that the maternal heart rate is monitored and misinterpreted as the fetal heart rate.
Therefore, the simultaneous monitoring of maternal heart rate (preferably, the maternal ECG) is
encouraged.
position, therefore Philips fetal monitors enable the FMP only after detecting a valid heart rate
signal for several seconds. FMP is not recommended when the mother is likely to move, and you
should disable Fetal Movement Profile (FMP) at the fetal monitor (FMP Off) if the mother is
walking.
•Gaps and maternal heart rate detection can occur:
–if the transducer is not correctly positioned.
–due to the pulsation of uterine blood vessels.
–if the fetus moves.
CAUTION
Using ultrasound gel not approved by Philips may reduce signal quality and may damage the
transducer. This type of damage is not covered by warranty.
WARNING
Periodically compare the mother's pulse with the signal coming from the monitor's loudspeaker to
ensure that you are monitoring fetal heart rate. Do not mistake a doubled or elevated MHR for FHR.
WARNING
Do not catheterize if placenta previa is diagnosed, or if uterine bleeding from an undetermined source
is present.
18
1 Introduction
WARNING
Never attempt to connect the fetal scalp electrode to anything other than the correct DECG adapter
cable.
WARNING
The fetal/maternal monitor is not a diagnostic ECG device. In particular, the display of fetal/maternal
ECG is intended only for evaluating signal quality for fetal/maternal heart rate as derived from the
ECG waveform.
When in doubt, it can be used to identify sources of compromised signal quality, such as noise or
muscle artifacts. It can subsequently be used to verify the result of measures taken to resolve them (e.g.
checking ECG cable connections or adapting the fetal
ArtifactSuppress configuration).
The safety and effectiveness of the displayed fetal/maternal ECG waveform (i.e. P, QRS and T
segments) for evaluation of fetal/maternal cardiac status during labor have not been evaluated.
WARNING
Intravenous infusion: Do not use the NBP cuff on a limb with an intravenous infusion or arterial
catheter in place. This could cause tissue damage around the catheter when the infusion is slowed or
blocked during cuff inflation.
Skin Damage: Do not measure NBP in cases of sickle-cell disease or any condition where skin
damage has occurred or is expected.
Unattended measurement: Use clinical judgment to decide whether to perform frequent unattended
blood pressure measurements in cases of severe blood clotting disorders because of the risk of
hematoma in the limb fitted with the cuff.
WARNING
Do not use the thermometer in the presence of flammable anesthetics, such as a flammable anesthetic
mixture with air, oxygen or nitrous oxide.
19
1Introduction
WARNING
•Never apply the probe to the patient when the probe is not connected to the unit.
•Always use a single-use probe cover to limit patient cross-contamination.
•Measurement errors or inaccurate readings may result when:
•Insert the probe slowly and carefully to avoid damage to the ear canal and the tympanic
•Inspect the probe for damage, holes, tears, or sharp edges to avoid injuring the skin.
•Always ensure that the used probe cover is removed before attaching a new probe cover.
CAUTION
•Do not immerse the probe in fluids, or drop fluids on the probe.
•Do not use a probe cover that has been dropped or is damaged.
•Do not autoclave. To prevent damage to the unit, probe and accessories, refer to the cleaning
–probe covers other than the specified probe covers are used (see “Tympanic Temperature
Accessories” on page 267)
membrane.
procedures in the “Cleaning and Disinfecting” on page 246 chapter.
CAUTION
Do not use OxiCliq disposable sensors in a high humidity environment, or in the presence of fluids,
which may contaminate sensor and electrical connections causing unreliable or intermittent
measurements. Do not use disposable sensors when there is a known allergic reaction to the adhesive.
WARNING
Proper Sensor Fit: If a sensor is too loose, it might compromise the optical alignment or fall off. If it
is too tight, for example because the application site is too large or becomes too large due to edema,
excessive pressure may be applied. This can result in venous congestion distal from the application site,
leading to interstitial edema, hypoxemia and tissue malnutrition. Skin irritations or lacerations may
occur as a result of the sensor being attached to one location for too long. To avoid skin irritations and
lacerations, periodically inspect the sensor application site and change the application site regularly.
Venous Pulsation: Do not apply sensor too tightly as this results in venous pulsation which may
severely obstruct circulation and lead to inaccurate measurements.
Ambient Temperature: At elevated ambient temperatures be careful with measurement sites that are
not well perfused, because this can cause severe burns after prolonged application. All listed sensors
operate without risk of exceeding 41°C on the skin if the initial skin temperature does not exceed
35°C.
Extremities to Avoid: Avoid placing the sensor on extremities with an arterial catheter, an NBP cuff
or an intravascular venous infusion line.
20
1 Introduction
WARNING
•For fully conscious pediatric or adult patients, who have a normal function of perfusion and
sensory perception at the measurement site:
To ensure skin quality and correct optical alignment of the sensor, inspect the application site
when the measurement results are suspicious or when the patient complains about pressure at
the application site, but at least every 24 hours. Correct the sensor alignment if necessary.
Move the sensor to another site, if the skin quality changes.
•For all other patients:
Inspect the application site every two to three hours to ensure skin quality and correct optical
alignment. Correct the sensor alignment if necessary. If the skin quality changes, move the
sensor to another site. Change the application site at least every four hours.
•Injected dyes such as methylene blue, or intravascular dyshemoglobins such as methemoglobin
and carboxyhemoglobin may lead to inaccurate measurements.
•Inaccurate measurements may result when the application site for the sensor is deeply pigmented
or deeply colored, for example, with nail polish, artificial nails, dye or pigmented cream.
•Interference can be caused by:
–High levels of ambient light (including IR warmers) or strobe lights or flashing lights (such as
fire alarm lamps). (Hint: cover application site with opaque material.)
–Another SpO
sensor in close proximity (e.g. when more than one SpO2 measurement is
2
performed on the same patient). Always cover both sensors with opaque material to reduce
cross-interference.
–Electromagnetic interference, especially at perfusion indicator values below 1.0 or signal
quality indicator below medium.
–Excessive patient movement and vibration.
WARNING
With pulse oximetry, sensor movement, ambient light (especially strobe lights or flashing lights) or
electromagnetic interference can give unexpected intermittent readings when the sensor is not
attached. Especially bandage-type sensor designs are sensitive to minimal sensor movement that might
occur when the sensor is dangling.
CAUTION
If you measure SpO2 on a limb that has an inflated noninvasive blood pressure cuff, a non-pulsatile
SpO
INOP can occur. If the fetal monitor is configured to suppress this alarm there may be a delay of
2
up to 60 seconds in indicating a critical status, such as sudden pulse loss or hypoxia.
21
1Introduction
WARNING
The fetal/maternal monitor is not a diagnostic ECG device. In particular, the display of fetal/maternal
ECG is intended only for evaluating signal quality for fetal/maternal heart rate as derived from the
ECG waveform.
When in doubt, it can be used to identify sources of compromised signal quality, such as noise or
muscle artifacts. It can subsequently be used to verify the result of measures taken to resolve them
(e.g., checking ECG cable connections).
The safety and effectiveness of the displayed fetal/maternal ECG waveform (i.e., P, QRS and T
segments) for evaluation of fetal/maternal cardiac status during labor have not been evaluated.
WARNING
The fetal/maternal monitor is NOT intended for use during defibrillation, electro-surgery, or MRI.
Remove all transducers, sensors, and accessories before performing electro-surgery, defibrillation, or
MRI, otherwise harm to the patient or the user can result.
WARNING
•No alarm is possible when Toco MP or CL Toco+MP transducer is the source of the pulse rate.
•No QRS tone is audible when Toco MP or CL Toco
+
MP transducer is the source of the pulse
rate.
•The Toco MP or CL Toco+MP transducer signal is significantly less reliable if the patient is up
and moving about, or is pushing during the second stage of labor.
WARNING
No alarm is possible when noninvasive blood pressure is the source of the pulse rate.
CAUTION
Only use Philips paper. Using paper other than Philips paper may result in the failure to recover traces.
CAUTION
Ensure that you admit each patient by name, including other patient identification information, and
discharge the patient when you have finished monitoring, so that you can identify which trace period
(entry in the patient list) refers to which patient.
22
1 Introduction
WARNING
•Do not operate the monitor if it is wet. If you spill liquid on the monitor, contact your service
personnel or Philips service engineer.
•Do not perform underwater monitoring (for example, in a bath or shower) using wired
transducers.
•Place the monitor where there is no chance of contact with, or falling into water or other liquid.
•Do not dry equipment using heating devices such as heaters, ovens (including microwave ovens),
hair dryers and heating lamps.
•Do not put equipment or accessories in autoclave (for sterilization).
CAUTION
Solutions: Do not mix disinfecting solutions (such as bleach and ammonia) as hazardous gasses may
result.
Skin contact: To reduce the risk of skin irritations, do not allow a cleaning or disinfecting agent to
leave residues on any of the equipment surfaces - wipe it off with a cloth dampened with water, after
allowing the appropriate time for the agent to work, or before applying to a patient.
Hospital policy: Disinfect the product as determined by your hospital's policy, to avoid long term
damage to the product.
Local requirements: Observe local laws governing the use of disinfecting agents.
Touch display: To clean and disinfect the touch-enabled display, disable the touch operation by
switching off the monitor during the cleaning procedure, or by selecting and holding the Main Screen
key until the padlock symbol appears on it, indicating that touch operation is disabled. Select and hold
again to re-enable touch operation.
WARNING
Do not put device and accessories in autoclave (for sterilization).
WARNING
Schedule: Failure on the part of the responsible individual hospital or institution employing the use of
this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure
and possible health hazards.
In case of problems: If you discover a problem with any of the equipment, contact your service
personnel, Philips, or your authorized supplier.
Electric shock hazard: Do not open the monitor housing. Refer all servicing to qualified service
personnel.
23
1Introduction
WARNING
To avoid contaminating or infecting personnel, the environment or other equipment, make sure you
disinfect and decontaminate the monitor appropriately before disposing of it in accordance with your
country's laws for equipment containing electrical and electronic parts. For disposal of parts and
accessories such as thermometers, where not otherwise specified, follow local regulations regarding
disposal of hospital waste.
WARNING
Reuse: Disposable accessories and supplies intended for single use, or single use only, and are
indicated as such on their packaging. Never reuse disposable accessories and supplies, such as
transducers, sensors, electrodes and so forth that are intended for single use, or single patient use only.
Approved accessories: Use only Philips-approved accessories.
Packaging: Do not use a sterilized accessory if its packaging is damaged.
Protection against electric shocks: The transducers and accessories listed in this chapter are
not defibrillator proof.
Electro-Surgery, Defibrillation and MRI: The fetal/maternal monitors are not intended for use
during defibrillation, electro-surgery, or MRI. Remove all transducers, sensors, and accessories before
performing electro-surgery, defibrillation, or MRI, otherwise harm can result.
WARNING
Explosion Hazard: Do not use in the presence of flammable anesthetics, such as a flammable
anesthetic mixture with air, oxygen, nitrous oxide, or in oxygen rich environment. Use of the devices in
such an environment may present an explosion hazard.
WARNING
•Do not use additional AC mains extension cords or multiple portable socket-outlets. If a multiple
portable socket-outlet is used, the resulting system must be compliant with IEC/EN 60601-1 A1:
2012.
•Do not connect any devices that are not supported as part of a system.
•Do not use a device in the patient vicinity if it does not comply with IEC/EN 60601-1 A1: 2012.
The whole installation, including devices outside of the patient vicinity, must comply with IEC/
EN 60601-1 A1: 2012. Any non-medical device, including a PC running an OB TraceVue system,
placed and operated in the patient's vicinity must be powered via a separating transformer
(compliant with IEC/EN 60601-1 A1: 2012) that ensures mechanical fixing of the power cords
and covering of any unused power outlets.
•Do not use USB devices with own power supplies unless an appropriate separation device is used
(either between USB interface and device or between device and power).
24
1 Introduction
CAUTION
•FM20/FM30 only: Although this is an electrical Class II device, it has a protective earth
conductor which is needed for EMC purposes.
•Always use the supplied power cord with the three-prong plug to connect the monitor to AC
mains. Never adapt the three-prong plug from the power supply to fit a two-slot outlet.
CAUTION
The use of accessories, transducers and cables other than those specified may result in increased
electromagnetic emissions or decreased electromagnetic immunity of the device.
WARNING
Do not use cordless/mobile phones or any other portable RF communication system within the
patient vicinity, or within a 1.0 m radius of any part of the fetal monitoring system.
WARNING
For paced patients: The radiated SRR power of the CL SpO2 and NBP Maternal Cableless
Measurement Devices, the CL Transmitter, and other sources of radio-frequency energy, when used in
very close proximity of a pacemaker, might be sufficient to interfere with pacemaker performance.
Due to shielding effects of the body, internal pacemakers are somewhat less vulnerable than external
pacemakers. However, caution should be exercised when monitoring paced patients.
In order to minimize the possibility of interference, avoid positioning and wearing the Cableless
Measurement Devices in very close proximity to a pacemaker. Consult the pacemaker manufacturer
for information on the RF susceptibility of their products
CAUTION
Fetal parameters, especially ultrasound and ECG, are sensitive measurements involving small signals,
and the monitoring equipment contains very sensitive high gain front-end amplifiers. Immunity levels
for radiated RF electromagnetic fields and conducted disturbances induced by RF fields are subject to
technological limitations. To ensure that external electromagnetic fields do not cause erroneous
measurements, it is recommended to avoid the use of electrically radiating equipment in close
proximity to these measurements.
CAUTION
The device should not be used adjacent to, or stacked with, other equipment unless otherwise
specified.
25
1Introduction
WARNING
•Leakage currents: If several items of equipment used to monitor a patient are interconnected,
•ECG electrodes: NEVER allow ECG electrodes to contact other electrical conductive parts,
WARNING
The fetal/maternal monitors are NOT intended for use during defibrillation, electro-surgery, or MRI.
Remove all transducers, sensors, and accessories before performing electro-surgery, defibrillation, or
MRI, otherwise harm to the patient or the user can result.
WARNING
The fetal/maternal monitors are not intended for use for ECG measurements on patients connected
to external electrical stimulator or with cardiac pacemakers.
the resulting leakage current may exceed allowable limits.
including earth.
26
Device Classification and Names within this Manual
New cableless measurements for fetal and maternal monitoring supplement the Avalon fetal monitor
family. The Avalon CL solution consists of the Avalon CL base station, the Avalon CL transducers, and the IntelliVue CL Pods.
The IntelliVue CL Pods are only used for maternal measurements within the Avalon CL solution.
To differentiate between the CL transducers and the CL Pods, the CL Pod are also referred to as
maternal measurement Pods. The following table provides an overview of all the devices within the
solution.
1
Avalon FM20/FM30 and FM40/
FM50
Wired Transducers
FM20/
FM30
M2702A and M2703A
FM40/
FM50
M2704A and M2705A
US
transducer
wired
Avalon CL Base Station
Avalon CL Transducers
Avalon CL
Base Station
866074
Avalon CL
Toco + MP
Transducer
cableless
IntelliVue CL Pods
IntelliVue
CL NBP
Pod
cableless
M2736A
865216
866075
27
1Introduction
Avalon FM20/FM30 and FM40/
FM50
Wired Transducers
Toco/ To co
MP
transducer
wired
M2734A and M2734B
To co +
transducer
with ECG/
IUP
capability
wired
M2735A
Avalon CL Base Station
Avalon CL Transducers
Avalon CL
US
Transducer
cableless
866076
Avalon CL
ECG/IUP
Transducer
cableless
IntelliVue CL Pods
IntelliVue
CL SpO2
Pod
cableless
865215
Patient
Module for
ECG/IUP
866077
M2738A
28
2What's New
This section lists the most important new features and improvements to the fetal monitors and their
user interface introduced with Release J.3.
You may not have all of these features, depending on the fetal monitor configuration purchased.
What's New in Release J.3
Avalon CL Transducer System
The Avalon CL system provides cableless monitoring with the Avalon FM20/FM30 and FM40/FM50
with the same functionality and performance as the wired measurement devices (e.g. twin and triplets
monitoring). The Avalon CL has a straight-forward handling and operating concept. The cableless
transducers are assigned by simply docking them at the base station, no further configurations are
necessary.
The Avalon CL system includes the following features:
2What's New
•Cableless monitoring of twins and triplets
•Cableless maternal measurement Pods CL SpO
•Maternal pulse from a CL Toco+MP (Smart Pulse) transducer
•A cableless ECG/IUP transducer measuring IUP and fetal/maternal ECG
•Watertight cableless transducers that can be used to monitor in water
•Patient call that pages an ambulating mother with an audible signal emitted by the worn CL
transducers.
•Out of range audible signal emitted by the worn CL transducers to inform an ambulating mother
that she is approaching the limit of the active signal area of reach.
•A transducer finder LED on all cableless transducers to help identify the assigned transducer
and CL NBP
2
Support For Use of Maternal Cableless Measurement Devices
The IntelliVue measurement pods are patient-worn, battery powered measurement devices for SpO2
and NBP. The devices provide measurement values on the built-in display and communicate them to
the fetal monitor using the wireless short range radio (SRR) interface of the Avalon CL base station.
Maternal Temperature Measurement
To measure maternal temperature, the new optional Tympanic Temperature device (866149) is
available for the Avalon fetal monitors. The measurement data is:
•fully documented at the local recorder and the OB system,
•can be displayed as a numeric on the screen,
•and is printed out on the recorder trace strip.
29
2What's New
Manually Entered Maternal Temperature Measurements
Manually measured temperatures can be entered manually into the fetal monitor and stored in the
database. They are also displayed as a numeric on screen, and are printed out on the recorder trace if
required.
New Design for the User Interface
The user interface for the fetal monitors has been redesigned to bring the presented information into
the foreground, letting the structural elements such as keys and frames retreat into the background.
Additionally special regard was given to making the "look and feel" similar to that of standard software
products.
New SmartKeys
•The Start ECG SmartKey is renamed to Record ECG .
•With the
transducers.
•With the
the
connected Avalon CL system.
•With the
externally measured maternal temperature values.
Call Patient SmartKey you can now page patients who are ambulating wearing cableless
Tele Info SmartKey you can call up the Tele Info window on the fetal monitor display. In
Tele Info window you can control and view the status of the cableless transducers from the
Enter Temp SmartKey a pop-up window opens showing a numeric pad for entering
Coincidence INOP Tone
When the cross-channel verification detects that the signal of the maternal heart rate coincides with the
fetal heart rate, the
INOP tone has a configurable delay.
Coincidence INOP is issued with a tone at the fetal monitor. The Coincidence
Increased Internal Back-up Memory
The internal back-up memory is now able to store data at least 3 hours.
USB Interface
An optional USB interface allows the use of USB printers, bar code readers, and input devices such as
a keyboard or mouse.
Flexible Nurse Call Interface
An optional Flexible Nurse Call interface allows now the connection of a nurse call device to the fetal
monitors.
DHCP Support
The DHCP support offers an alternative to BOOTP. DHCP (dynamic host configuration protocol)
enables the fetal monitors to request an IP address (internet protocol address) from the connected
network (OB TraceVue/IntelliSpace Perinatal) automatically.
30
Loading...
+ 282 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.