Philips M1754 User manual

Page 1
Cardiograph
Signal-Averaged ECG
M1754A
INSTRUCTIONS FOR USE
Page 2
Page 3
M1754A Signal-Averaged ECG (SAECG) Instructions for Use
3KLOLSV3DUW1R0 3ULQWHGLQ86$-DQXDU\
(GLWLRQ

Page 4

Notice

About This Edition
Publication number M1754-91908 Edition 3 Printed in USA
Edition History
October 1991 Edition 1 June 1992 Edition 2 January 2002 Edition 3
Copyright
Copyright © 2001 Philips Electronics North America Corporation
All rights are reserved. Reproduction in whole or in part is prohibited without the prior written consent of the copyright holder.
Philips Medical Systems Andover, MA 01810 USA
This informatio n is s ub ject to change without notice.
Warranty
Philips makes no warranty of any kind with regard to this material, included, but not limited to, the implied warranties or merchantability and fitness for a particular purpose .
Philips shall not be liable for errors co nta ined herein or for incidental or conseque ntial damages in connection with the furni s hi ng, performance, or use of this material.
This product may contain rem anufactured parts equivalent to new in performance or have had incidental use.
Philips assumes no li abilit y for f ailures resultin g from RF interference between Philips medical electronics and any radio frequency generating equipment at levels exceeding those established by applicable standards.
Classified by Underwriters Laborato ries , Inc. with respect to electrical shock, fire, and mechanical hazards only in accordance with UL 2601-1.
CAUTION
The manufacture r, importe r and seller are responsible for the effects on safety, reliability and performance only if: assembly operations, extensions, re-adjustments, modi fications or repair are carried out by persons authorized by them, and the electrical installation of the relevant room complies with all local regulations and the equipment is used in accordance with the instructions for use . Us e of accessories other than those recommended by Philips may compromise product performance.
THIS PRODUCT IS NOT INTENDED FOR HOME USE.
Responsibility of Manufacturer
Philips only considers it s elf responsib le for any effects on safety, reliability and performance of the PageWriter cardiograph and the battery support system if:
assembly operations, extensions, readjustments, modifications or repairs are don e by persons authorized by He wl ett Packard, and
the electrical installation of the relevant room or vehicle complies with the IEC or national requiremen ts , and
the instrument is used according to the instructions for use presented in this manual.
WARNING
As with electroni c e quipment, Radio Frequency (RF) interference between the
defibrillator and any existing RF transmitting or receiving equipment at the installat io n site, including electrosurgica l equipment, should evaluated carefully and a ny limitations noted before the equipm ent is placed in servic e.
Monitoring during quiescent periods of electrosurgery is possible but electromagnetic interference generated by electrosurgical tools during operation is sufficient to mask cardiac signals.
Radio frequency generat io n from electrosurgical equi pment and close prox imity transmitters may seriously degrade performance of the PageWriter cardiograph.
ii
Page 5
1

1Printing History

October 1991 Edition 1
June 1992 Edition 2
January 2002 Edition 3
iii
Page 6
1
127(

1Safety Summary

Safety Symbols Marked on the Cardiograph

The following symbols are used on the cardiograph or the cart.

Caution - See operating instructions Type CF, defibrillation protected Alternating current Equipotential (this is on the ground lug) The maximum weight that the cart can hold.

Conventions Used in This Manual

Notes contain additional information on M1754A usage.
VRIWNH\V

represents the temporary key labels that appear on the display.

represents keys on the front panel.

.H\
iv
Page 7

Contents

1. Getting Started

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Installing M1754A SAECG Software . . . . . . . . 1-1
Selecting the SAECG Application . . . . . . . . . . . 1-3
Preparing the Patient and Applying Electrodes . 1-4

2. Acquiring a Signal-Averaged ECG

Selecting the Template Beat . . . . . . . . . . . . . . . . 2-1
Editing the Final Report . . . . . . . . . . . . . . . . . ... .2-5
Adjusting the QRS Onset and Offset . . . . . . . . . 2-5
Editing and Printing a Stored SAECG . . . . . . . . 2-7

3. Understanding the Printed SAECG Report

4. Configuring the M1754A SAECG Application

Configuring Your Cardiograph to Start Up
in the SAECG Application . . . . . . . . . . . . . . 4-1
Configuring the SAECG Application . .. . . . . . . 4-3
Patient ID Fields . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Configuration Tables . . . . . . . . . . . . . . . . . . . . . 4-5
Acquisition Fields . . . . . . . .. . . . . . . . . . . . . . . . 4-5
Measurement Fields . . . . . . . . . . . . . . . . . . . . . . 4-6

5. Troubleshooting

Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Glossary Index
contents 1
Page 8

Figures

1-1. Placing Electrodes for an SAECG. . . . . . . . 1-5
2-1. A Template Screen. . . . . . . . . . . . . . . . . . . .2-2
2-2. An Averaging Screen . . . . . . . . . . . . . . . . . 2-4
2-3. A Measurements Screen . . . . . . . . . . . . . . . 2-5
3-1. The SAECG Report. . . . . . . . . . . . . . . . . . . 3-2

Tables

1-1. SAECG Electrode Positions . . . . . . . . . . . . 1-5
3-1. SAECG Report Annotations . . . . . . . . . . . . 3-3
4-1. Acquisition Configuration Fields . . . . . . . . 4-5
4-2. Measurements/Processing Configuration
Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
5-1. SAECG Testing Problems . . . . . . . . . . . . . 5-2
contents 2
Page 9

1Getting Started

1Introduction

This guide contains information about using the M1754A Signal-Averaged ECG (SAECG) cardiograph application to acquire a Signal-Averaged ECG for use in the detection of late potentials. There are also instructions on configuring the cardiograph for M1754A SAECG application operation. Refer to the Page Writer XLi Cardiograph Operating Guide for operating and installation instructions for your cardiograph.

1Installing M1754A SAECG Software

Be sure that the M1700A XLi software has been installed before installing any other applications in your cardi ogr aph. To ins t all th e M175 4A S AECG so ft ware, perform the following steps.
1. Turn the cardiograph to Standby from the front panel.
2. Insert the M1754A SAECG application disk in the disk drive in the front of the cardiograph.
3. Prepare the SAECG patient module and attach it to the cardiograph. a. Put shorting plugs in the LA and LL positions. b. Attach lead wires to the remaining positions on the SAECG patient module.
1-1
Page 10
*HWWLQJ6WDUWHG
127(
,QVWDOOLQJ0$6$(&*6RIWZDUH
The SAECG patient module can also be used for XLi applications. If you wish to routinely acquire XLi ECGs as well as SAECGs, you may attach lead wires in place of the shorting plugs for the LA and LL leads on the SAECG patient module. Then attach the LA and LL electrodes on the normal positions.
You will achieve shorter averaging times if you use shorting plugs in the LA and LL lead positions.
4. Turn the cardiograph on. The message Load SAECG application? appears.
5. Press . The copyright display appears briefly.
<H V
(c) 1991 Hewlett-Packard, Corazonix Corp
SAECG (M1754A - A.01.00)
Then the following display appears.
******
SAECG (M1754A - A.01.00)
The asterisks gradually extend across the display until the software install a tion is complete. The following display appears.
Auto 3x4
Report Format 01/26/92 01:23AM
You can now select the SAECG application as described in the next section.
T o conf i gure your cardiograph for your requirements, refer to Chapter 4.
1-2
Page 11
*HWWLQJ6WDUWHG
Reformat Timed Cont Prev Delayed Exit
6HOHFWLQJWKH6$(&*$SSOLFDWLRQ

1Selecting the SAECG Application

When the SAECG application is installed, you can either use your cardiograph to acquire a standard (XLi) ECG or you can acquire Signal-Averaged ECGs. To select SAECG opera t i on from the Prev iew Plus application, perform the following steps.
1. Press until you see the following display.
0HQX
Prev+
2. Press .
3. In the Applications menu, press .
([LW
6$(&*
• If you haven't already attached the SAECG patient module, you will be instructed to do so.
The following display appears.
SAECG
Start 01/26/92 01:23AM
4. Attach the electrodes to the patient as described in the next section.
5. Check the signal quality by viewing the preview screen and the SAECG patient
module display.
The next chapter, Chapter 2, describes how to acquire the ECG and create the final SAECG report.
1-3
Page 12
*HWWLQJ6WDUWHG
127(
127(
3UHSDULQJWKH3DWLHQWDQG$SSO\LQJ(OHFWURGHV

1Preparing the Patient and Applying Electrodes

To prepare the patient and apply the electrodes, perform the following steps.
1. Relax the patient.
• Check that the patient is comfortable.
• Reassure the patient that the procedure is safe. It is essential to relax the patient as much as possible during the procedure. This is
because the quality of the ECG signal depends on reducing tension in the patient.
2. Prepare the electrode positions.
• Shave the electrode positions if necessary.
• Cleanse the skin with your institution's cleansing soluti on .
• Rub the skin gently with a gauze pad until the skin is slightly red. Electrode site preparation is especially important for the detection of the low level
signals involved in late potentials.
1-4
Page 13
3. Place the electrodes as shown in Figure 1-1.
*HWWLQJ6WDUWHG
3UHSDULQJWKH3DWLHQWDQG$SSO\LQJ(OHFWURGHV
AHA IEC Position Y+
Y­Z+
Z­X+
X-
RL LA LL
Y+
In the lower, left abdominal region on the left iliac crest
Y-
On the superior aspect of the manubrium.
Z+
At the fourth in tercostal space, at the le ft s ternal margin (V 2 po s ition)
Z-
On the back, directly opposite the Z+ electrode
X+
At the same level as the Z axis along the mid-axillary line on the pati en t’s left si d e
X-
At the same level as the Z axis along the mid-axillary line but on the patient’s right side
N
Opposite the Y+ electrode on the right iliac crest
L
On the left fo re arm (if used)
F
On the left leg (i f u se d )
1-5
Page 14
*HWWLQJ6WDUWHG
3UHSDULQJWKH3DWLHQWDQG$SSO\LQJ(OHFWURGHV
1-6
Page 15

2Acquiring a Signal - Averaged ECG

0HQX
6WDUW
This chapter contains information on how to acquire a Signal-Averaged ECG. There are three steps to acquiring a Signal-Averaged ECG.
1. Selecting a template beat.
2. Acquiring and averaging ECG data.
3. Verifying the onset and offset of the QRS complex.

2Selecting the Template Beat

You must confirm the template beat to start the signal averaging process. Perform the following steps to select a template beat.
1. Press until the following menu is displayed.
SAECG Start 01/26/92 01:23AM
2. Press . The message Determining template beat... appears. Then the following display appears.
Select the template. Select Reject Match Noise Exit
2-1
Page 16
$FTXLULQJD6LJQDO$YHUDJHG(&*
6HOHFWLQJWKH7HPSODWH%HDW
The template beat with the matching and noise windows appears on the preview screen as shown in Figure 2-1.
2-2
3. To confirm the template beat, press .
6HOHFW
• If you wish to place the template on another lead, use the and keys.
• If you wish to acquire a new template beat, press .
• If you wish to adjust the matching window, press .
5HMHFW
0DWFK
You can obtain best results when the matching window contains the fast rising portion of the R wave.
If you wish to adjust the noise window, press .
1RLVH
Be sure that the noise window is on a flat part of the signal and that the T wave is not inside the window.
Page 17
$FTXLULQJD6LJQDO$YHUDJHG(&*
'RQH
6HOHFWLQJWKH7HPSODWH%HDW
When you press or , perform the following steps to adjust the
0DWFK 1RLVH
windows. a. Use and keys to adjust the position of the windows.
b. Press . The following display returns.
Select the template. Select Reject Match Noise Exit
After you have selected the template beat, the message Averaging... appears and the preview screen displays a continuous ECG sig nal for leads X, Y, and Z as shown in Figure 2-2. The message remains on the screen until the end condition is met. The end condition is configured on your cardiograph. The default end
condition is a noise level of less than 0.3
µV, but the end condi tion could be
configured to a specified number of beats or a selected noise level. The number of beats and the noise level are displayed on the preview screen during averaging.
2-3
Page 18
$FTXLULQJD6LJQDO$YHUDJHG(&*
127(
'RQH
6HOHFWLQJWKH7HPSODWH%HDW
Y o u can press at any time to end the averaging process be fore it meets the configured end condition.
2-4
Once the averaging process is complete, the message Filtering averaged beats... appears.
Page 19
$FTXLULQJD6LJQDO$YHUDJHG(&*
(GLWLQJWKH)LQDO5HSRUW

2Editing the Final Report

After the ECG signals are averaged and filtered, the final SAECG information is displayed on the preview screen. You can adjust the QRS onset and offset before printing and storing the SAECG report.

2Adjusting the QRS Onset and Offset

The QRS complex on the final report has its onset and offset marked for signal averaging measurements. It is essential that the onset and offset in any individual lead or in the vector are properly placed to produce accurate measurements. If you must adjust the QRS onset and offset on the final report, perform the following steps.
1. When averaging is finished, the following message appears on the display.
Adjust the onset and offset. Print Onset Offset Lead Exit
At the same time the Measurements screen, as shown in Figure 2-3, appears on the preview screen.
2-5
Page 20
$FTXLULQJD6LJQDO$YHUDJHG(&*
$GMXVWLQJWKH4562QVHWDQG2IIVHW
2. Ensure that a flexible disk for storage is in the cardiograph’s disk drive.
3. If the QRS complex’s onset and offset are correctly placed by default and you want a printed report, press . The message Printing report ...
3ULQW
appears before the report is stored.
• If you wish to change the waveform's scale, use the and keys.
• If you wish to move the onset, press .
• If you wish to move the offset, press .
2QVHW
2IIVHW
The following display appears.
Adjust the onset Done Exit
a. Use the and keys to move the onset or offset windows left or right. b. Press when the QRS onset and offset are correctly placed.
• If you wish to view another lead, press . The following display
'RQH
/HDG
appears.
Choose another lead Vector X Y Z Exit
2-6
m If you wish to view the X, Y, or Z lead, press , , or respec-
; < =
tively. You can then change the onset and offset on the leads if desired.
m If you wish to display the combined vector lead on the preview screen for
editing or viewing, press .
9HF WR U
After selecting a lead you can adjust the onset and offset independently for all leads plus the combined vector lead.
Page 21
$FTXLULQJD6LJQDO$YHUDJHG(&*
127(
'RQH
0HQX
([LW
(GLWLQJDQG3ULQWLQJD6WRUHG6$(&*
After printing, the report will be st o red. If your appl icatio n is configured for manual file naming, you will be prompted to provide a file name. a. To store a report with a specified file name, type a name with up to eight characters and with a three-digit extension; for exampl e, W enw orth.001. You can also use the patient ID for a file name if the ID is under eight characters.
Manually enter the file name if you intend to use this file on a Corazonix Predictor for additional study.
b. Press . The message Storing report... appears, then the display returns to the SAECG display.

2Editing and Printing a Stored SAECG

You can edit and print a stored SAECG report. To retrieve a stored SAECG report so you can edit and print the report, perform the following steps from the SAECG application.
1. Insert the storage disk that has the SAECG report you wish to edit or print into the cardiograph disk drive.
2. Press until the following display appears.
Print Edit Config
3. Press . The following display appears.
12345678 12/01/91 01:07 PM Select Next Previous Exit
2-7
Page 22
$FTXLULQJD6LJQDO$YHUDJHG(&*
3UHYLRXV
(GLWLQJDQG3ULQWLQJD6WRUHG6$(&*
4. Press or until the correct patient ID number appears.
1H[W
3UHYLRXV
5. Press . The following display appears.
Adjust the onset and offset. Print Onset Offset Lead Exit
You can now adjust the onset and offset and print the SAECG report as described earlier in Editing the Final Report. If you wish, you can also print the SAECG report without editing it.
You can delete stored SAECGs from the Preview + or XLi menus the same way ECGs are deleted.
2-8
Page 23

3Understanding the Printed SAECG Report

This chapter contains information about the printed SAECG report. The report is printed on the PageWriter XLi printer. The SAECG report gives a concise summary of the SAECG test.
3-1
Page 24
8QGHUVWDQGLQJWKH3ULQWHG6$(&*5HSRUW
3-2
Page 25
8QGHUVWDQGLQJWKH3ULQWHG6$(&*5HSRUW
Table 3-1. SAECG Report Annotations
Description A B C D
E F G H I J K
Patient ID number
Date and Time
Patient Name
Department,
Room number,
Operator
Age and Sex
Comments
Configured Settings
Measurements
Unfiltered Lead Waveforms
Absolute Filtered Lead Waveforms
Vector Size
3-3
Page 26
8QGHUVWDQGLQJWKH3ULQWHG6$(&*5HSRUW
3-4
Page 27

4Configuring the M1754A SAECG Application

This chapter describes how to conf i gur e you r cardi o grap h fo r use w ith the M17 54A SAECG application. Tables at the end of this chapter show which operating parameters are configurable.
4Configuring Your Cardiograph to Start Up in the SAECG
Application
If you wish to configure your cardiograph to start up in the SAECG application, perform the following steps from the main XLi application.
1. Press until the following display appears
0HQX
Transmit Store Config CheckDisk Files
4-1
Page 28
&RQILJXULQJWKH0$6$(&*$SSOLFDWLRQ
127(
&RQILJXULQJ<RXU&DUGLRJUDSKWR6WDUW8SLQWKH6$(&*$SSOLFDWLRQ
2. Press . The main configuration menu will appear:
&RQILJ
Configuration Global ID Transmit Files Exit
If this is not the initial configuration, you might need to type the configured password before you can make any changes to the configuration.
3. Press and the following display appears.
*OREDO
Initial Interpretation? Reasons Enter Choose Previous Exit
4. Press to step through the global configuration fields until the following
(QWHU
field appears.
Power on application? XLi Enter Choose Previous Exit
5. Press until SAECG appears on the display.
6. Press and the SAECG setting will be accepted. The next configuration
&KRRVH
(QWHU
parameter will appear.
7. Press when you have finished changing the main configuration settings.
([LW
Your cardiograph will now start up in the SAECG application when you turn it on.
4-2
Page 29
&RQILJXULQJWKH0$6$(&*$SSOLFDWLRQ
&RQILJXULQJWKH6$(&*$SSOLFDWLRQ

4Configuring the SAECG Application

The M1754A SAECG application has default operating parameters when it is installed on your cardiograph. To customize the SAECG operating parameters, perform the following steps from the SAECG application.
1. From the main SAECG menu, press until the following menu is displayed.
0HQX
Print Edit Config
2. Press . The SAECG Configuration menu is displayed.
&RQILJ
SAECG Configuration ID Acquis Meas Exit
3. Press one of the softkeys on the SAECG configuration menu to configure the SAECG application for your institution.
• If you wish to determine which ID parameters are requested each time you enter patient informatio n , press .
• If you wish to choose Acquisitio n par a met e rs, press .
• If you wish to choose Measurements parameters, press .
,'V
$FTXLV
0HDV
The following sections describe how to configure ID, Acquisition, and Measurements operating parameters.
4-3
Page 30
&RQILJXULQJWKH0$6$(&*$SSOLFDWLRQ
3DWLHQW,')LHOGV

4Patient ID Fields

You don’t have to step through all the possible patient ID information each time you enter patient ID. You can turn off the ID fields that you don’t use. The only information you must include for storage is the patient ID number.
The following list shows each patient ID configuration option. You can choose either Yes to include or No to exclude that item from the patient ID information printed on the ECG. Yes is the default choice for all.
• Allow NAME entry?
• Allow AGE entry?
• Allow SEX entry?
• Allow OPERATOR entry?
• Allow DEPARTMENT entry?
• Allow ROOM entry?
• Allow COMMENTS entry?
4-4
Page 31
&RQILJXULQJWKH0$6$(&*$SSOLFDWLRQ
A
T
T

Configuration Tables

Table 4-1 and Table 4-2 show configurable operations on the M1754A SAECG application. The setting choices that are shown in bold type are the default settings. These settings are adequate for most applications.
cquisition Fields
able 4-1 shows Acquisition configuration choices.
able 4-1. Acquisition Configuration Fields
Parameter Set ting Choices Comments
Fiducial point?
Correlation coefficient? (0.01- 0.99)
Average to?
Noise level? (0 - 100
Number of beats? (1 - 9999)*
µV)*
Left Center Right
0.99
Type a number between
0.01 and 0.99
Noise level
Beats
0.3 µV
Type a number between
0.0 and 100.0
200
Type a number between 1 and 9999
The peak of the QRS can be moved from the center to either the right or left third of the display and printed report. The noise window is moved by the same amount as the fiducial point. This sets how closely the acquired beats must match the template. A value of 0.99 is as close as possible. The end of the averaging process is controlled by either a total number of accepted beats, or an average noise level of the three leads. The noise level required before the process completes. * This field appears only if Noise level is selected in the Average to? field. The total beats that must be collected before the process completes. * This field appears only if Beats is selected in Average to? field.
3DWLHQW,')LHOGV
4-5
Page 32
&RQILJXULQJWKH0$6$(&*$SSOLFDWLRQ
3DWLHQW,')LHOGV
4-6
Page 33
4Troubleshooting
127(
This chapter contains information about troubleshooting the M1754A operation. Cardiograph troubleshooting is covered in more detail in the
PageWriter XLi Cardiograph Operating Guide and the PageWriter XLi Cardiograph Instructions for Use.
Be sure that you have the SAECG patient module attached to the cardiograph for use with the M1754A application.
ECG technique is particularly important for ensuring a good recording. An extremely noisy ECG signal can take too long to average. Three problems that could prevent you from acquiring an SAECG include:
• Muscle artifact
• Bad lead contact
• Electrical interference
Chapter 1 includes more information on ECG technique.
5-1
Page 34
7URXEOHVKRRWLQJ
(UURU0HVVDJHV

4Error Messages

Error messages appearing on the display instruct you as to what action to take. If the error is something th at you can correct, the message in structs you what to do.
5-2
Page 35

4Glossary

applications
Software used for a specific task. The PageWriter XLi has the SAECG application and the Preview Plus application.
combined vector lead
A combination of three orthogonal leads. It is derived from the sum of the squares of the three leads, and then the square root of that sum.
configuration
The preset operating parameters for the cardiograph. When the software is installed, the cardiograph defaults to a preset configuration which may be customized any time.
correlation coefficient
A number that shows how closely that the acquired beats must match the template beat. O indicates the least close match; 1 indicates the closest match.
end condition
The point when the designated number of beats are acquired or the level of noise is reached in order to have enough clear ECG data to average.
Glossary-1
Page 36
*ORVVDU\
fiducial point
A reference point based on the peak of the QRS complex. This adjustment enables you to average on different parts of the waveform.
filtered lead scaling
The size at which the filtered leads are displayed.
filter frequency
Frequency on the high pass filter above which an ECG signal is recorded. For example, a filter frequency of 80 1Iz means that frequency components below 80 Hz are removed.
flexible disk
A disk which can be inserted in the disk drive to store data such as ECG and SAECG reports or the configuration. Software is installed from a flexible disk.
function keys
The line of keys lettered F1 to F5 at the top of the keyboard that correspond to the labels displayed at the bottom of the screen.
matching window
The area of the template beat that will be correlated to each new beat.
0HQX
Cardiograph key that changes the menu selections displayed on the cardiograph’s front panel display.
Glossary-2
key
Page 37
*ORVVDU\
noise window
The area of the template beat used to determine the amount of noise in the signal. This should be the smoothest part of the beat.
operating parameters
Configured settings that permit the SAECG application to function in a desired manner.
password
Private code word that limits access to the cardiograph’s configuration software to those persons knowing the code word. Passwords prevent accidental or unauthorized changes to cardiograph configuration.
Preview Plus
Philips name for optional software application which enhances th e standard preview screen.
preview screen
Philips term for the screen which, when installed on the cardiograph, shows the ECG traces as they will appear on the printed ECG report. The preview screen is required for the SAECG application.
QRS onset and offset
The beginning and end of the QRS complex.
SAECG patient module
The remote patient unit that contains all of the cardiograph’s SAECG data acquisition electronics. The patient module connects to the patient data cable and to the leads attached to the patient.
Glossary-3
Page 38
*ORVVDU\
signal averaging
Averaging performed on an ECG waveform to reduce random noise.
Signal-Averaged ECG (SAECG)
A report produced by the M1754A to aid in the detection of late potentials.
softkeys
The function keys labelled in the lower portion of the cardiograph’s front panel display and physically positioned underneath the display. These keys change functions when they are pressed. The matching display label also changes.
template beat
A typical beat used to match other beats dur ing aver aging . When selecting a template beat, choose a clean beat without a lot of noise.
vector scaling
The size at which the combined leads are displayed.
Glossary-4
Page 39
Index
6
6503, 5-2
A
acquiring a Signal-Averaged ECG, 2-1 Acquis, 4-3 acquisition configuration, 4-5 adjusting the onset and offset , 2-6
Adjust the onset., 2-6 Adjust the onset and offset., 2-5
AGE entry, 4-4 application, 1-1, Glossary-1 application disk, 1-2 Applications menu, 1-3 artifact, muscle, 5-1 Attach SAECG patient module., 5-2 automatic file naming , 2-7 , 4- 6 automatic scaling
filtered leads, 4-6 unfiltered leads, 4-6
averaging, 2-3, 2-5
shorter times, 1-2
Averaging..., 2-3 Averaging screen, 2-3
B
beat
template, 2-1, Glossary-4
beats, number of, 4-5, Glossary-1
C
changing. See adjusting, editing
matching window, 2-2 noise window, 2-2 the onset and offset, 2-5, 2-6 the scale, 2-6
Choose another lead., 2-6 combined vector lead, 2-6, Glossary-1 COMMENTS entry, 4-4 Config, 4-2 configuration, Glossary-1
acquisition, 4-5 measurements, 4- 6 patient ID, 4- 4
start up, 4-2 Configuration menu, 4-2 configured settings, 3-3 configuring your cardio gra ph, 1-2, 4-1-6 copyright display, 1-2 Corazonix Predictor correlation coefficient, 4-5, Glossary-1 correlation window. See matching window customizing operating parameters, 4-1
®, 2-7
D
DEPARTMENT entry, 4-4 Determining template be at. . .,2-1 disk
application, 1-1, 1-2
storage, 2-6, 2-7 disk drive, cardiograph, 1-1, 2-6, 2-7 Done, 2-3, 2-4, 2-6, 2-7
E
ECG
noisy, 5-1
technique, 5-1 Edit, 2-7 editing. See ad justing, changin g
a lead, 2-6
combined vector lead, 2-6
final report, 2-5
stored SAECG, 2-7 electrical interference, 5-1 electrodes, 1-3
placing, 1-5
positions, 1-5
site preparation, 1-4 end condition, 2-3, Glossary-1 error messages, 5-2 Exit, 2-3, 2-8, 4-3
F
fiducial po int, Glossary - 2 fiducial point configuration, 4-5 file name, 2-7 file naming
automatic, 4-6
manual, 4- 6 filtered lead
scale, 4-6
scaling, 4-6, Glossary-2
waveforms, 3-3 filter frequency, 4-6, Glossary-2 Filtering averaged beats..., 2-4 final report
editing, 2-5 flexible disk, 2-6, Glossary-2 function keys, Glossary-2
H
high pass filter frequency, 4-6 Hi-Res ECG. See SAECG
I
identification, patient. See patient ID ID fields
configuring, 4-4
turning off, 4-4 iliac crest, 1- 5
Index-1
Page 40
installing M1754A SAECG software, 1-1 interferenc e, electrical, 5-1
L
LA
electrode, 1-5
position, 1-1 late potentials, 1-1, 1-4, Glossary-4 lead
contact, 5-1
orthogonal, Glossary-1
viewing another, 2-6 Lead, 2-6, 2-8 lead off , 5-2 LL
electrode, 1-5
position, 1- 1
Load SAECG application?, 1-2 M
M1754A SAECG
application, 1-1
configuring, 4-1-6
installing software, 1-1 manual file naming, 2-7, 4-6 manubrium, 1-5 Match, 2-2, 2-3 matching window, 2-2, Glossary-2
adjusting, 2-2 Meas, 4-3 measurements, 3- 3
configuration, 4-6 Measurements screen, 2-5
Menu
, 1-3, 2-1, 2-7, Glossary-2 mid-axillary line, 1-5 mid-clavicular l ine , 1 -5 muscle artifact, 5-1
N
NAME entry, 4-4 naming a file
automatically, 2-7, 4-6 manually, 2-7, 4-6
No heartbeat. Check connections., 5-2 Noise, 2-2, 2-3 noise level, 2-3, 5-2
configuring, 4-5
noise window, 2-2, Glossary-2
adjusting, 2-2
noisy signal, 5-2 No room to load application., 5-2 number of beats, 2-3
O
Offset, 2-8
Onset, 2-8 onset and offset
adjusting, 2-6
operating parameters, Glossary-3
customizing, 4-1, 4-3 OPERATOR entry, 4-4 orthogonal leads, Glossary-1
P
password, 4-2, Glossary-3 patient
preparing, 1-4
relaxing, 1-4 patient ID
configuration, 4-4
number, 3-3 patient module, SAECG, 1-1 , 5-1 placing electrodes, 1- 5 power on application, 4-2 preparing the patien t, 1-4 Preview Plus, Glossary-3 preview screen, 1-3, 2-3, 2-5, Glossar y-3 Print, 2-7, 2-8 Printing report... , 2-6 printing the report, 2-6, 2-8
Q
QRS
adjusting the onset and offset , 2-5
complex, 2-1
onset and offset, 2-1, 2-6, Glossary-3
peak, 4-5
R
reducing tension, 1-4 Reject, 2-2 relaxing the patient, 1-4 RL electrode, 1-5 ROOM entry, 4-4 R wave, 2-2
S
SAECG, 1-3, 1-5
acquiring, 2-1
application disk, 1-2
display, 2-7
editing a stored, 2-7
patient module, 1-1, 5-1, Glossa ry-3
printing a stored, 2-7 SAECG Configuration me nu, 4-3 SAECG report, 3-1-3
editing, 2-5 SAECG test, 5-2 scale
changing, 2-6
Index-2
Page 41
Select, 2-2, 2-8 Select the template.,2-1 SEX entry, 4-4 shorting plugs, 1-1, 1-2, 1-5 Signal-Averaged ECG, 1-1, 1-3, Glossary-4 signal averaging, G lossary-3 signal quality, 1-3, 1-4 softkeys, iii, 4-3, Glossary-4 Standby key, 1-1 Storing report..., 2-7 storing the report, 2-6, 2-7
T
template beat, 2-1, Glossary- 4
confirming, 2-2 Template screen, 2-2 troubleshooting, 5-1-2 T wave, 2-2
U
unfiltered lead
scale, 4-6
scaling, 4-6
waveforms, 3-3
V
vector
scale, 4-6
scaling, 4-6, Glossary-4
size, 3-3 Vector , 2- 6 viewing
another lead, 2-6
combined vector lead, 2-6
X
X+ electrode, 1-5 X- electrode, 1-5 X lead, 2-6 XLi, 1-3 XLi ECG, 1-1,1-3
Y
Y+ electrode, 1-5 Y- electrode, 1-5 Y lead, 2-6
Z
Z+ electrode, 1 -5 Z- electrode, 1-5 Z lead, 2-6
Index-3
Page 42
Index-4
Page 43
Page 44
3DUW1XPEHU0 3ULQWHGLQ86$ (GLWLRQ
'$"!(&&%'
Loading...