Do not forget to rescue the CAN piggy-back box from the
backside of the generator !
This is a decade comparison for an SCP –> Optimus R/F swap at a D96 system:
SCP Optimus function explanation
EZX 1 WBX 1 auxiliary release cassette EZX 2 WBX 2 auxiliary release cassette / tomo EZX 3 WBX 3 auxiliary release DSI EZX 4 WBX 4 auxiliary release DSI / tomo
EZX 21 WAX 1 auxiliary release free cassette (can also be a 2nd tube) EZX 22 WAX 2 auxiliary release wall stand (can also be a 2nd tube)
EZX 10 WBX 11 Bucky ready / system ready, select side fields cassette tray,
check whether ready signals from EZX8 SCP have to be used as well
EZX 8 WAX 11 tomo ready, Bucky – tomo remote switch-over,
Bucky ready >> check EZX10 SCP as well, side field selection in case there is a 2
EZX 15 WAX 21 tomo time input EZX 16 WAX 22 tomo trajectories
EZX 31 WBX 12 lock-in
EZX 33 (WBX 12) there is no kV-mA-manual operation possible like with the SCP,
only dose rate limiter function is working, but signal not provided
EZX 28 WBX 13 image intensifier format contacts
EZX 30 WBX 14 EXON signal
EZX 29 should not be present, Scopomat systems only
EZX35 WBX23 external APR EZX34 WBX 24 patient size correction
Other signals apart from the system CAN box:
SCP Optimus function explanation
EZX 19 EZX 4 emergency off EZX 37 WGX 8 tube temperature switch tube 1 EZX 38 WGX 9 tube temperature switch tube 2
nd
6028 collimator
D94-96_SCP_OM_RF_gen_swap.doc Oct - 06 – 2003 page 1 - 1