453564064111
January 2008
Edition 1
Printed in the USA
Applicable to Model 860322
Philips Medical Systems shall not be
liable for errors contained herein or for
incidental or consequential damages in
connection with the furnishing,
performance, or use of this material. This
information is subject to change without
notice.
Philips Medical Systems
3000 Minuteman Rd.
Andover, MA 01810 USA
TEL: (978) 687-1501
CAUTION
THIS PRODUCT IS NOT INTENDED
FOR HOME USE. IN THE U.S.,
FEDERAL LAW RESTRICTS THIS
DEVICE TO SALE ON OR BY THE
ORDER OF A PHYSICIAN.
Trademarks
Windows is a registered trademark of
Microsoft Corporation.
All other brand and product names are
trademarks or registered trademarks of
their respective companies.
Warranty
This Philips product is warranted against
defects in materials and workmanship for
a period of two years from the date of
shipment. During the warranty, Philips
will, at its option, either repair or replace
products which prove to be defective.
Philips does not warrant that the operation
of this software, firmware, or hardware
shall be uninterrupted or error free.
No other warranty is expressed or
implied. Philips specifically disclaims the
implied warranties of merchantability and
fitness for a particular purpose other than
described herein.
Philips considers itself responsible for
effects on safety, reliability, and
performance only if:
– Assembly operations, extensions,
readjustments, modifications and
repairs are carried out by Philips
authorized personnel only.
– Electrical installation in the relevant
room complies with applicable
requirements.
– Equipment is used in accordance with
the instructions for use.
Limitations of Warranty
The foregoing warranty shall not apply to
defects resulting from:
– Improper or inadequate maintenance
by buyer.
– Buyer-supplied software or interfacing.
– Unauthorized modification or misuse.
– Operation outside of the environmental
specifications for the product.
– Improper site preparation.
– Improper maintenance.
Excluded from this warranty are
expendable supply items including, but
not limited to, electrodes, lead wires,
patient cables, and battery. This warranty
does not apply to any product that Philips
Medical Systems determines has been
modified or damaged by the customer.
Service Agreements
Periodic inspections and maintenance
service are recommended to ensure
effective operation. Call your Philips
Sales and Service Office for details
regarding a Customer Support Services
Agreement for your specific requirements.
Global Medical Device
Nomenclature (GMDN)
The 5-digit GMDN code adjacent to the
symbol is defined in the EN ISO 15225.
Unpacking and Storage
The equipment is packed in its own
reusable shipping container. When
unpacking the equipment, inspect the
carton for physical damage. Any damage
should be reported immediately to the
shipping company. Open the shipping
container and compare the contents to the
packing list. There is only one packing list
per shipment. If there are several parcels,
the shipping list is normally attached to
the largest container. If the packing list
does not agree with the items received,
contact a Philips Response Center. The
shipping container should be saved to
permit reshipment.
Medical Device Directive
DigiTrak XT, model 860322
with the requirements of the Medical
Device Directive 93/42/EEC and carries
the
The DigiTrak XT Holter Monitor is a battery operated solid state recorder, designed for up to
168 hours (7 days) continuous recording of ambulatory electrocardiograph (AECG) data. It
has the ability to detect and record pacemaker pulses according to the appropriate criteria for
AAMI pacer detection.
The DigiTrak XT is an AAMI Type I device and part of a conventional AECG monitoring
system where an ECG is recorded in Multi Media Card (MMC) memory within the recorder.
After the recording is complete, the DigiTrak XT recorder is connected via the docking station
to a USB port on the Holter computer. Follow the instructions provided with the Holter
application to download and analyze the recorded ECG data.
Indications for Use
The DigiTrak XT Holter recorder is intended for patients requiring ambulatory (Holter)
monitoring.
CAUTIONThe DigiTrak XT recorder is not intended for use on infants weighing under 10 kilograms
(22 pounds), as required by IEC 60601-2-47:2001.
Such monitoring is most frequently used for the indications below:
Evaluation of symptoms suggesting arrhythmia or myocardial ischemia
Evaluation of ECG documenting therapeutic interventions in individual patients or groups
of patients
Evaluation of patients for ST segment changes
Evaluation of a patient's response after resuming occupational or recreational activities
(for example, after myocardial infarction or cardiac surgery)
Clinical and epidemiological research studies
Evaluation of patients with pacemakers
Reporting of time and frequency domain heart rate variability
Reporting of QT interval
v
Safety and Warning Information
Equipment Symbols
SymbolMeaning
0123
Attention: Consult accompanying documents
Symbol on label indicates Philips meets the applicable requirements
of the European directive 93/42/EEC
Type B equipment contains adequate protection against electrical
shock, particularly regarding the allowable leakage current and the
reliability of protective earth connection (when present)
Dispose of in accordance with the requirements of your country.
SN
REF
Date of manufacture
2000
China ROHS
Serial number
Not shown - catalogue number is DigiTrak XT
The 5-digit Global Medical Device Nomenclature (GMDN) code
adjacent to the symbol is defined in the EN ISO 15225.
viDigiTrak XT Recorder Instructions for Use
Precautions
Patient leads must be removed from electrodes before defibrillation.
When using Pacer Detect, be aware that false positive and false negative pacer detects
may occur.
–False positives — may result from poor electrode hookup or high noise conditions.
–False negatives — may occur with bipolar pacers due to a weak pacer pulse signal at
The presence of pacemaker signals in the ECG trace should not be considered true
representations of the actual pacemaker stimulus amplitude when viewing the ECG data.
Observe local laws for disposal of alkaline and lithium batteries.
Do not leave the battery in the recorder when it is not in use. Damage from corrosion
could result.
Use of rechargable batteries is not recommended.
For the best recording results, the patient should be instructed to avoid close proximity to
heavy electrical equipment or other sources of electromagnetic interference such as
electric blankets, heating pads, etc.
Precautions
the patient's skin surface.
The DigiTrak XT recorder should not be immersed in water. The patient should be
instructed not to wear the recorder in the shower or bath.
The DigiTrak XT recorder supports an Early Out feature that allows a trained individual to
stop a recording before the selected recording time has elapsed. Failure to follow this
procedure may result in the loss of all recorded ECG data.
This product consists of devices that may contain mercury, which must be recycled or
disposed of in accordance with local, state, or federal laws. Within this system, the
backlight lamps in the monitor display contain mercury.
Electrode Application
It is recommended that trained medical personnel handle the application of electrodes.
Use only electrodes designed for longer term Holter monitoring.
Proper preparation of the patient's skin is absolutely essential for obtaining a quality ECG
recording. Refer to your electrode provider or this guide for instructions on skin
preparation techniques.
Apply electrodes as shown in the diagrams of the Monitor Hookup Kits, the Electrode
Placement diagrams in this guide, or as instructed by a physician.
Materials Used
The DigiTrak XT recorder, belt clip, and patient cable do not contain any latex material.
DigiTrak XT Recorder Instructions for Usevii
Safety and Warning Information
Security Recommendations
To further strengthen the security and confidentiality of your patient records and system in
general, we recommend that you implement additional security measures, including:
Remind your patients that the recorder contains confidential patient data and they should
safeguard the recorder while it is in their possession.
Store the recorders in a secure location when not in use.
Delete patient data from the recorder after it has been downloaded to Holter. (See
“Deleting Patient Information” on page 1-19 for instructions.)
Conventions
This book uses the following text conventions:
WARNINGWarning statements describe conditions or actions that can result in personal injury or
loss of life.
CAUTIONCaution statements describe conditions or actions that can result in damage to the equipment or loss
of data. Caution statements alert the user that the clinician has the responsibility of determining
significance of results due to actions and varying factors present with each case.
NOTENotes contain additional information on using this product.
viiiDigiTrak XT Recorder Instructions for Use
1
Chapter 1Using the DigiTrak XT Recorder
The DigiTrak XT Holter Monitor is a battery operated solid state recorder, designed for up to
168 hour (7 day) continuous recording of ambulatory electrocardiograph (AECG) data. It has
the ability to detect and record pacemaker pulses according to the appropriate criteria for
AAMI pacer detection.
Features include:
Multi-day recording, up to 168 hours (7 days)
Easy navigation of user menus
Streamlined form factor and light weight for patient comfort
Workflow enhancements, featuring the ability to preprogram user demographics
information, on-screen lead status and gain adjustment
Reliability improvements, featuring a reinforced patient cable connector and power-on
self tests to check for battery life
Supportability enhancements, featuring easy removal and replacement procedures
Security enhancements, featuring soft encryption of data. Encryption is the conversion of
data into a form that cannot be understood by unauthorized people. Decrypting is the
process of converting encrypted data back into its original form so that it can be
understood.
Your Holter software must be at Version 2.9 or later in order to perform multi day scans
greater than 48 hours. See the Philips Holter Installation and Configuration Guide or the Philips Holter Instructions for Use for a complete list of system requirements.
Using the DigiTrak XT with Legacy Software
You must install the DigiTrak XT Compatibility software to use the DigiTrak XT with Holter
releases 2.7, 2.8, and 2.8.1. See the DigiTrak XT Compatibility Software Installation Instructions that came with the DigiTrak XT CD for more information.
1-2DigiTrak XT Recorder Instructions for Use
Accessing Documentation Updates at InCenter
Accessing Documentation Updates at InCenter
Updates to the documentation are provided periodically. To access the latest documentation,
visit the Philips InCenter website, at
If you do not yet have an InCenter user ID, register by clicking the link
software updates and documentation for cardiology products”
ID and password will be sent to the email address you provide, enabling access to
documentation updates, as well as free software updates, when available.
on the right side of the page. A user
“Click here for access to
Accessing Documentation at the Philips Website
Documentation for many Philips medical products can be downloaded from the Philips
website, at
AboutDisplays the following information about the recorder:
Product name
Serial number
Firmware version information
StartAfter configuring or reviewing all the settings, select the start screen
and press
Enter.
This will start the recording. During recording, the recorder displays
the current time and time remaining to record.
Navigating the Recorder Display
Use the following buttons to navigate the DigiTrak XT screens:
Navigating the Recorder Display
ButtonDescription
Left arrow. Moves from one tab to the next
and changes values within a field.
Right arrow. Moves from one tab to the next
and changes values within a field.
Down arrow. Moves from one field to the
next.
Up arrow. Moves from one field to the next.
Enter. Used to change settings and to save the
current settings.
Event marker. Used by the patient to record
events.
If scroll arrows appear on the display, it indicates additional fields that are located off the
screen. Use the up and down arrow keys to access these additional fields.
DigiTrak XT Recorder Instructions for Use1-7
Using the DigiTrak XT Recorder
Using the Belt Clip
To insert the recorder into the belt clip
Slide the belt clip over the battery end of the recorder until it snaps into place.
1
2Rotate the belt clip until it locks into the desired position.
To remove the recorder from the belt clip
XSlide the battery end of the recorder out of the belt clip.
Getting Started
The following sections describe how to set up the recorder for use.
NOTEIf the recorder is turned off for any reason, for example, to shut it down early, you must remove the
battery. To restart the recorder, insert the battery and press Enter. When the recorder is powered
on, default settings are in effect; any custom settings are lost. In addition, the previous ECG recording
is erased.
Setting the Recorder Date and Time (First Time Use only)
To set the recorder date and time
NOTEUse an alkaline battery for recordings up to 96 hours. Use a lithium battery for recordings over 96
hours.
1
Make sure a fresh AAA battery is in the recorder.
NOTEThe recorder enters sleep mode after 20 seconds of no activity. If this happens, simply press Enter
to turn the recorder on.
2
Use the arrow buttons to display the Date/Time tab.
3Press Enter.
4Use the up and down arrows to select the parameter to change (hour, minute, date, and so
on), and use the
5Press Enterto accept the changes.
6To start a recording, display the Start tab and then press Enter to begin recording.
7If you do not want to start a recording, remove the battery to shut down the recorder.
left and right arrows to change the setting of the parameter.
1-8DigiTrak XT Recorder Instructions for Use
Getting Started
Step One — Connecting the Recorder to the Dual Dock
You connect the recorder to the docking station to enter patient information and to download
data collected in the DigiTrak XT recorder to the Holter application. The data is transferred to
and from the Holter system via a USB docking station. The docking station is connected to the
PC through a USB port; you place the recorder in the docking station to transfer the data. The
dual dock refers to the docking station that accommodates both the DigiTrak XT and DigiTrak
Plus recorders.
Ensure the dual dock is connected to the Holter computer before proceeding.
To connect the dual dock to the Holter computer
XPlug the USB connector from the dual dock into an available USB port on the computer.
To connect the recorder to the dual dock
Remove the battery from the recorder.
1
2Remove the belt clip or pouch from the recorder, if attached.
CAUTIONDo not insert the recorder at a high angle or with the pins facing out. Inserting the recorder from a
high angle will result in bent pins.
3Gently slide the left side of the recorder straight into the dual dock station (at the angle
shown) with the pins facing the side (under the flap). The recorder snaps into place.
4Gently press down on the right side of the recorder to lock it into the dual dock station.
There should be light resistance.
DigiTrak XT Recorder Instructions for Use1-9
Using the DigiTrak XT Recorder
CAUTIONDo not force the recorder into the dual dock. If it does not easily snap into place, remove the
recorder and try installing it again.
USB connected appears on the recorder display when the recorder is properly connected
and the
Preloading message appears in the bottom right of the Holter application window
(if there is data on the recorder). The dual dock is now operational and you are ready to
enter or download data.
1-10DigiTrak XT Recorder Instructions for Use
Getting Started
Step Two — Entering the Patient’s Name (Optional)
To enter patient information
With the dual dock connected to the PC, insert the recorder into the dual dock.
1
2Launch the Holter program.
3Click Tools > Recorder Utility.
The Recorder Utility screen appears.
4Select Philips Recorders as the type.
5Enter the patient information (last name, first name, ID, and so on).
6Click Write to Philips Recorders.
After several seconds the data is written to the recorder.
NOTEAll patient information is displayed on the recorder if you entered it using the Holter 2.9 software.
Although previous versions of software store the patient information and display it on the report, only
the patient name is displayed on the recorder screen.
7
Remove the recorder from the dual dock.
DigiTrak XT Recorder Instructions for Use1-11
Using the DigiTrak XT Recorder
Step Three — Preparing the Patient
Before You Begin
Insert a fresh AAA battery with each patient. Be sure to observe the correct battery
polarity.
Instruct the patient not to tamper with the recorder, remove the battery, or disconnect the
patient cable.
To prepare the patient
Prepare the patient’s skin prior to applying the electrodes. Skin is a poor conductor of
1
electricity, so skin preparation is important in achieving good electrode-to-skin contact.
–If necessary, clip hair at the electrode sites (or shave sites, if needed).
–Clean and abrade the skin at the electrode site. Wash skin thoroughly with soap and
water.
–Dry the electrode sites briskly to increase capillary blood flow in the tissues and to
remove oil and skin cells.
2Attach the leads to the electrodes before placing them on the patient.
3Apply the electrodes by peeling them, one at a time, from the protective backing and
sticking them firmly to the patient’s skin. (Refer to Figure 1-1 for proper electrode
placement.) Press around the entire edge of each electrode to ensure they are secure. Make
sure the lead wires do not pull on the electrodes.
1-12DigiTrak XT Recorder Instructions for Use
Figure 1-1Electrode Placement
Getting Started
Red (-)
White (-)
Brown (+)
CAUTIONLead colors for EASI Holter hookups are different from those used for EASI telemetry hookups. Lead
placements are the same.
Electrode Placement
Green
Black (+)
E (Brown)Level of 5th intercostal space, midsternum
A (Black)Same level as E and I, left mid-axillary line
S (Red)Top of sternum (manubrium)
I (White)Same level as E and A, right mid-axillary line
Ground (Green)Center of sternum or any convenient location
Raw ChannelDescription
Channel 1E (+) to S (-)
Similar to MC V1, anterior view of the heart
Channel 2A (+) to S (-)
Similar to MC V6, a lateral view of the heart -- useful for ST
measurements
Channel 3A (+) to I (-)
CC6, similar to the inferior lead aVF – approximation suitable for ST
measurements
DigiTrak XT Recorder Instructions for Use1-13
Using the DigiTrak XT Recorder
Step Four — Checking the Lead Status and ECG Signal
Quality
To check the lead status
Make sure a fresh AAA battery is in the recorder.
1
The recorder turns on as soon as you put in a battery. The recorder performs a self-test and
the splash screen appears for a couple of seconds, with the message
displayed at the bottom of the screen. If the cable is not connected, the message No Cable
appears at the bottom of the screen and you cannot proceed.
2Insert the patient cable (lead set) into the recorder connector. Press firmly to be sure it is
seated properly.
3Press any button to enter the menus.
Press any key to start
If you entered patient information, it is displayed. Otherwise, the
NOTEAll patient information is displayed on the recorder if you entered it using the Holter 2.9 software.
Although previous versions of software store the patient information and display it on the report, only
the patient name is displayed on the recorder screen.
4
Use the arrow buttons to select the Lead Status tab.
5Check the diagram for loose connections.
Leads Status tab appears.
– indicates a lead has a good patient connection
– (flashing) indicates a lead does not have a good patient connection
–
Leads OK or Leads Error message displayed
6Use the arrow buttons to select the CH1, CH2, and CH3 tabs and check the ECG signal
quality.
1-14DigiTrak XT Recorder Instructions for Use
7Observe the screen for cable and lead errors as follows.
8When all the leads are properly connected, the ECG waveform appears.
9Use the arrow buttons to change the gains settings of the display, as shown in the
following table.
NOTEThe recording for all resolutions is always made at the 1X gain setting.
Getting Started
Table 1-2Gain Settings
Gain on screenMillivolt range
0.5X± 5 peak-to-peak
1X± 2.5 peak-to-peak
2X± 1.25 peak-to-peak
10
Proceed to Settings and select Pacer Detector, if applicable.
11 Use the arrow buttons to select the Start tab and press Enter to start the recording.
NOTEIf you do not start the recording, the DigiTrak XT automatically starts recording after 30 minutes,
provided there are no lead errors.
During a Recording
During a recording, the DigiTrak XT displays the date, current time, and the time remaining
for the recording as well as any lead or cable error, if present.
DigiTrak XT Recorder Instructions for Use1-15
Using the DigiTrak XT Recorder
Step Five — Removing and Shutting Down the Recorder
To retrieve the recorder
1
When the patient returns, remove the recorder from the patient.
The recorder automatically shuts down after the record time you set has expired. To shut
down the recorder before the specified time expires, see “Stopping a Recording (Early
Out)” on page 1-19.
2Remove the patient cable.
If you pre-loaded the patient information, the patient name appears on the recorder screen
until you download the data.
Step Six — Downloading Data from the Recorder
To download data from the recorder into the Holter system
Launch the Holter application.
1
2Connect the recorder to the Holter system as described on page 1-9. Be sure to remove the
battery from the recorder.
The text
Preloading appears in the lower right-hand corner of the PC screen.
CAUTIONDo not disconnect the recorder whenever the Preloading message is flashing. To cancel the
Preloading sequence, double-click the Preloading message and select Abort.
If you accidentally unplug the recorder during a preload sequence, restart the PC to clear the
DigiTrak XT device. If the recorder is unplugged prematurely, the ECG is corrupted.
3Leave the recorder connected until downloading of the data is complete.
You can perform an analysis while downloading data.
4When downloading of the data is complete, detach the recorder from the dual dock, then
scan and save the ECG.
5Repeat steps 1-3 for subsequent new patients.
6Refer to the Philips Holter Instructions for Use for information on performing a scan.
1-16DigiTrak XT Recorder Instructions for Use
Changing Recorder Settings
You can change various recorder settings (specified in Table 1-1) and save them (along with
the date and time settings) as the default through the
To change recorder settings
1
Select the Settings tab.
2Change the settings, as necessary.
3Select OK and press Enter to save the settings.
Saving Default Settings
To save the settings as the recorder default settings
1
Change the settings, as necessary. Change the date and time settings as necessary (located
on the
Date/Time tab).
2Select Save as default (located on the Settings tab) and press Enter to save the settings.
Changing Recorder Settings
Settings tab.
Pacemaker Detection
Pacemaker detection for the DigiTrak XT recorder is automatically defaulted to OFF. You
must turn it On for each procedure in which it is used or save the setting as part of the default
settings.
Pacemaker Threshold
The DigiTrak XT recorder allows the user to adjust the threshold for pacemaker spike
detection. The purpose of this feature is to raise or lower the threshold of the recorder
detection in order to eliminate spikes from a rate-modulated pacemaker or lower the threshold
to allow for better detection of bipolar pacemakers.
NOTEAdjusting the threshold on the DigiTrak XT recorder does NOT affect any threshold settings on the
patient’s pacemaker. The adjusting of the threshold settings are confined totally to the function of the
recorder.
You do not have to adjust the pacemaker recording threshold for all recordings.
Adjusting Pacemaker Thresholds
To adjust a pacemaker threshold
Select the Settings tab.
1
2Ensure that the Pacer Detector setting is ON.
3While viewing the ECG waveform, press Enter to display the pacemaker threshold
indicator on the right side of the screen.
The top position of the bar indicates maximum sensitivity. The indicator defaults to one
below the maximum sensitivity.
DigiTrak XT Recorder Instructions for Use1-17
Using the DigiTrak XT Recorder
NOTEThe Enter button toggles between the pacemaker threshold indicator and the gain setting.
4
Check to see if there are pacer pulses in the pacer detection channel on the bottom of the
screen.
5If there are many pacer pulses in the pacer detection channel and less or none in the ECG
channel, reduce the sensitivity of the pacemaker threshold indicator.
6If there are no pacer pulses in the pacer detection channel and there are pacer pulses in the
ECG channel, increase the sensitivity of the pacemaker threshold indicator.
7Use the arrow buttons to select the Start tab and press Enter to start the recording when you
are satisfied that the correct pacer pulses are being detected.
Registering an Event (Patient Event Marker)
The DigiTrak XT allows an easy way for the patient to record events. Once an event has been
marked, the patient must wait one minute to register another event.
To register an event
XPress Enter each time to mark an event.
The marked event time stamp appears on the recorder screen. After 20 seconds have
elapsed, the time stamp is cleared from the recorder screen.
1-18DigiTrak XT Recorder Instructions for Use
Stopping a Recording (Early Out)
The recorder supports an Early Out feature that allows a trained individual to stop a recording
before the selected recording time has elapsed.
To stop a recording in progress
1
Hold the left arrow and Enter buttons simultaneously.
A menu is displayed with the choices Stop Recording or Exit Menu.
2Select Stop Recording and press Enter.
The messages,
Shutting down please wait. Recording complete appear, the recording stops and
the recorder shuts down.
3If you do not want to shut down the recorder, select Exit Menu.
The recorder status screen appears and the recording continues.
Deleting Patient Information
Stopping a Recording (Early Out)
Patient information (including ECG data) is erased from the recorder the next time you turn on
the recorder. Once you press a key to enter the recorder’s menus, the message
Initializing Card
briefly appears on the screen. When this occurs, the MMC card is erased and any previous
patient and ECG information is deleted from the recorder.
If you are storing the recorder after downloading data, you can use the following procedure to
erase the patient data.
To delete patient information
Remove the battery from the recorder.
1
2Insert a battery into the recorder.
The recorder turns on as soon as you put in a battery. The recorder performs a self-test and
the splash screen appears for a couple of seconds, with the message
Press any key to start
displayed at the bottom of the screen. If the cable is not connected, the message No Cable
appears at the bottom of the screen and you cannot proceed.
3Insert the patient cable (lead set) into the recorder connector. Be sure it is seated properly.
4Press Enter.
The message
Initializing Card briefly appears on the screen. When this occurs, the MMC
card is erased and any previous patient and ECG information is deleted from the recorder.
DigiTrak XT Recorder Instructions for Use1-19
Chapter 1Service and Specifications
This chapter provides information about servicing your recorder, device specifications, and
available parts and accessories.
NOTEFor better protection of the recorder, we strongly recommend using the DTXT pouch (part #
989803153451 ). The pouch provides extra protection and cushioning.
Supplies & Parts
The following part can be ordered by contacting the Response Center.
DescriptionPhilips P/N
Battery door453564067201
DigiTrak XT Recorder Instructions for Use2-7
Service and Specifications
The following parts can be ordered by contacting your local Philips Medical Systems Sales
Office or your authorized Philips Medical Systems Dealer or Distributor.
DescriptionPhilips P/N
USB Dual Dock989803157511
Patient diaries (200)M4701A
24-inch patient cable (lead set)989803157481
36-inch patient cable (lead set)989803157491
54-inch patient cable (lead set)989803157501
DigiTrak XT hookup kit (1 patient diary, 1
M3730-62600
alkaline AAA battery, 1 alkaline AA battery, 5
electrodes)
Patient electrodes (300)M4706A
Belt clip989803158191
Belt clip (10-pack)989803158210
Replacing the Battery Door
To replace a broken battery door
Cut the rubber tether on the battery door.
1
2Remove any tether pieces from the recorder.
3Insert the tether of the new battery door through the recorder tether slot. Make sure the
small hook at the end of the tether pops out from underneath the recorder tether slot.
Recorder tether slot
4With the battery door fully closed, slide the door to release the locking mechanism.
5Lift the battery door at it’s widest point.
When installed correctly, the battery door automatically pivots about the tether. The
battery door will hang off the end of the recorder.
2-8DigiTrak XT Recorder Instructions for Use
Specifications
Functional
Channels3
Recorded amplitude resolution10 bits
RecordingFull disclosure
Download interfaceUSB
Sample rate175/sec maximum
Frequency response0.05Hz to 60Hz, @-3dB
Signal verificationLCD display
Event switchPress Enter
Pacemaker DetectionProgrammable on/off
Memory
Specifications
Capacity/Recording time256 MB up to 96 hours
512 MB up to 168 hours (7 days)
Recording typeMMC
Physical
Dimensions91.44 x 55.88 x 19.05mm
(3.60 x 2.20 x 0.75inches)
Weight with battery70 g. (2.5 oz.)
EnclosureMolded plastic (UL 94V-2)
Operating positionAny orientation
Electrical
Gain setting0.5X, 1X, 2X
Connector11 pin
Patient cable5 lead
Environmental
Operating temperature0°C to +45°C/32°F to 113°F
Non-operating temperature-10°C to +70°C/14°F to 158°F
Operating humidity10% to 95% (non-condensing)
Non-operating humidity5% to 95% (non-condensing)
DigiTrak XT Recorder Instructions for Use2-9
Service and Specifications
Battery
Type
Life
Warranty
(1) AAA Alkaline IEC-LR3 for recordings
up to 96 hours
(1) AAA Lithium for recordings longer than
96 hours
168 hours (7 days)
24 months from shipment
2-10DigiTrak XT Recorder Instructions for Use
Electromagnetic Compatibility (EMC)
Electromagnetic compatibility with surrounding devices should be assessed when using the
DigiTrak XT recorder.
An electronic device can either generate or receive electromagnetic interference. Testing for
electromagnetic compatibility (EMC) has been performed on the DigiTrak XT recorder
according to the international standard for EMC for medical devices (IEC 60601-1-2). This
IEC standard has been adopted in Europe as the European Norm (EN 60601-1-2).
The DigiTrak XT complies with the requirements of standard EN 60601-1, as follows:
The equipment is not suitable for use in the presence of a flammable anaesthetic mixture
with air or with oxygen or nitrous oxide.
IPX3 ordinary equipment (enclosed equipment with protection against spraying water).
The recorder will withstand exposure to liquids, such as rain, or splashing water. The
recorder will not survive full immersion or liquid spray under pressure. The patient should
be instructed not to wear the recorder in the shower or bath.
A
Appendix A
Internally powered equipment
Mode of operation - continuous operation.
The DigiTrak XT recorder should not be used adjacent to, or stacked on top of other
equipment. If the DigiTrak XT recorder must be used adjacent to or stacked on top of other
equipment, verify that the recorder operates in an acceptable manner in the configuration in
which it will be used.
Fixed, portable, and mobile radio frequency communications equipment can affect the
performance of medical equipment. For more information on troubleshooting, see
“Troubleshooting” on page 2-2.
The list of cables with which Philips claims compliance with emissions and immunity
requirements of IEC 60601-1-2 are listed on page 2-7.
A-1
Only use Philips Medical Systems replacement parts with the DigiTrak XT recorder. The use
of non-approved replacement parts may result in increased Radiated Emissions or decreased
Electromagnetic Immunity of the DigiTrak XT recorder.
Table A-1. Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The DigiTrak XT recorder is intended for use in the electromagnetic environment specified
in the table below. The customer or the user of the DigiTrak XT recorder should assure that
it is used in such an environment.
RF energy only for its internal
function. Therefore, its RF
emissions are very low and are
not likely to cause any
interference in nearby
electronic equipment.
Harmonic Emissions IEC
61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not Applicable
Not Applicable
A-2DigiTrak XT Recorder Instructions for Use
Electromagnetic Compatibility (EMC)
Table A-2. Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The DigiTrak XT recorder is intended for use in the electromagnetic environment specified
below. The customer or the user of the DigiTrak XT recorder should assure that it is used in
such an environment.
Immunity TestIEC 60601 Test
Level
Electrostatic
Discharge (ESD)
+/- 6 kV contact
+/- 8 kV air
IEC 61000-4-2
Electrical Fast
transient/burst
IEC 61000-4-4
+/- 2 kV for power
supply line
+/- 1 kV for input/
output lines
Surge IEC 610004-5
+/- 1 kV differential
mode
+/- 2 kV common
mode
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines IEC
<5% U
(>95% dip in UT) for
0.5 cycle
40% U
(60% dip in UT) for 5
cycles
T
T
61000-4-11
70% U
T
(>30% dip in UT) for
25 cycles
Compliance LevelElectromagnetic
Environment:
Guidance
CompliesFloors should be wood,
concrete, or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
N/AThe DigiTrak XT does
not have AC or DC
power lines.
N/AThe DigiTrak XT does
not have AC or DC
power lines.
N/AThe DigiTrak XT does
not have AC or DC
power lines.
<5% U
T
(>95% dip in UT)
for 5 seconds
Power frequency
3 A/mCompliesPower frequency
(50./60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the AC mains voltage prior to application of the test level.
DigiTrak XT Recorder Instructions for UseA-3
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Table A-3. Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The DigiTrak XT recorder is intended for use in the electromagnetic environment specified
below. The customer or the user of the DigiTrak XT recorder should assure that it is used in
such an environment.
Immunity TestIEC 60601
Test Level
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
Compliance LevelElectromagnetic Environment:
guidance
3 Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the DigiTrak XT,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
D = 1.2vP
D = 1.2vP 80 MHz to 800 MHz
D = 2.3vP 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and D
is the recommended separation
distance in metres (m).
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from surfaces, objects, and people.
Additional notes are on following page.
A-4DigiTrak XT Recorder Instructions for Use
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Electromagnetic Compatibility (EMC)
ADDITIONAL NOTES
A Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radios, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the DigiTrak XT recorder is used exceeds the applicable RF
compliance level above, the DigiTrak XT recorder should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the DigiTrak XT recorder.
B Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.
Table A-4. Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the DigiTrak XT recorder: for equipment and
systems that are not life-supporting
The DigiTrak XT recorder is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the DigiTrak XT
recorder can help to prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
DigiTrak XT recorder as recommended below, according to the maximum output power of
the communications equipment.
Rated Maximum Output
Power of Transmitter W
Separation Distance According to Frequency of
Transmitter (m)
150 KHz to
80 MHz
D = 1.2vP
80 MHz to
800 MHz
D = 1.2vP
800 MHz to 2.5
GHz
2.3vP
0.01.12.12.23
0.1.38.38.73
11.21.22.3
103.83.87.3
10012.012.023.0
For transmitters rated at a maximum output power not listed above, the recommended
separation distance D in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by the absorption and reflection from structures, objects, and people.
DigiTrak XT Recorder Instructions for UseA-5
Index
B
battery cover
replacing 2-8
C
calling for service 2-4
changing recorder settings 1-17
channels 1-6
cleaning the recorder 2-1
connecting the recorder to the dual
dock 1-9
contrast 1-6
D
default settings 1-6
deleting patient information 1-19
documentation
accessing at the Philips Website 1-3
updates at InCenter 1-3
downloading data from the recorder 1-16
dual dock 1-9
back view 1-5
changing settings 1-17
checking lead status and ECG signal
quality 1-14
cleaning 2-1
connecting to the dual dock 1-9
downloading data 1-16
entering patient information 1-11
features 1-1
front view 1-4
functions 1-6
indications for use i-v
navigating the display 1-7
safety information i-v
saving default settings 1-17
security recommendations i-viii
serial number 1-5
service tag 1-5
setting the recorder date and time 1-8
shutting down 1-16
specifications 2-9
stopping 1-19
supplies and parts 2-7
system requirements 1-2
troubleshooting 2-2
registering an event 1-18
resolution 1-6
S
safety information i-v
sample rate 1-6
saving default settings 1-17
security recommendations i-viii
serial number 1-5
service 2-4
service tag 1-5
setting the recorder date and time 1-8
settings
contrast 1-6
default settings 1-6
languages 1-6
pacemaker detection 1-6
record time 1-6
resolution 1-6
sample rate 1-6
shutting down the recorder 1-16
specifications 2-9
stopping a recording 1-19
supplies 2-7
system requirements 1-2